Resident Physics Lectures Plain Film Tomography George David Associate Professor Department of Radiology Medical College of Georgia Body Section Tomography • Not CT • Body Section Imaging • Goal keep plane of interest in focus blur all other plans » enhances contrast • Popularity decreasing because of CT MRI Tomography History • Predates CT by decades • Used to be popular for inner ear studies • Still used at MCG for IVP’s George David Tomography Blurring • Blurring accomplished by synchronous movement of tube & film tube & bucky physically connected by rod rod pivots around fulcrum tube moves one direction film moves in other direction Conventional Tomography Blurring • Image produced on film • Objects above or below fulcrum plane change position on film & thus blur George David Tomography Blurring • the further from the fulcrum an object is, the more it blurs! Tomography Blurring • objects shaped & oriented in direction of motion elongate do not blur Tomographic Blurring • Blurring improved by use of complex motions circular Hypocycloidal tri-spiral elliptical Complex Blurring • The more complex the blurring motion the sharper the tomographic image » better blurring the more expensive the machine $ George David Tomo Patient Doses • Tomo can be high dose procedure several rads not unusual • one image per cut with film cuts at many levels routinely employed to find cut of interest each cut exposes entire field • Can do multiple images per cut with digital receptor George David Thickness of Cut • Same as thickness of region in focus • Determined by angle tube moves • Larger angles yield thinner cuts more motion = more blurring • Smaller angles yield thicker cuts 0o (stationary) yields conventional film Thickness of Cut Larger Angle; Thinner Cut Smaller Angle; Thicker Cut Cut Thickness • Approximate cut thickness for linear tomo Tomo Angle Thickness of Cut (mm) (degrees) ------------------------------------------------------0 Infinite (non-tomo) 2 31 4 16 6 11 10 6 20 3 35 2 50 1 George David Grids & Tomo • Linear grid used • Grid lines parallel with table for linear tomo • Grid must change orientation (rotate) in complex motion tomo eliminates grid cutoff George David