pH and ICP Techniques Clemson Scoop 2 checks One blank Double scoop 6 samples Tag set with lab number, check, and blank positions pH quality control checks 1 2 3 Two QC check samples Two samples at beginning, middle, and end of set Cross Check electrodes #1 and #2 Quality control row ck Check both QC check values at beginning and within set Check all double scooped samples at beginning, middle, and end of set Buffer run - checks Buffer run - blank (1:1, water:buffer) End of run verification Friday ◦ clean out electrode filling solution ◦ fill with water Ultrasonic cleaner Store in buffer pH 7 for the weekend Maintenance Monday ◦ Take water out of electrode ◦ Refill with KCl filling solution Verify flow of KCl through filter Refresh buffer solutions Electrode (from Fisher) ◦ Accumet rugged bulb pH combination electrode ◦ 13-620-183A Maintenance – cont. 5.3 5.25 5.2 5.15 5.1 5.05 5 0 20 40 60 pH Check values 80 100 120 140 Check B data since April 1, 2010 115 continuous reads over a 2.5 month period Rinse pH < 4.5 and > 7.0 NAPT value 5.15 Lab average 5.15 stdev 0.036 Two checks within the set One blank ICP quality control checks Sample entry Probe – Leeman Labs ◦ Phone: 603-886-8400 ◦ Part number 902-00017 Check all tubing for cleanliness and kinks Change pump winding tube when needed Adjust pump winding clamp - look for smooth flow as sample travels though tube Aspirate yttrium stock solution Look for the tip of the red bullet Adjust nebulizer gas flow rate TJA Nebulizer gas flow ARCOS ARCOS Keep clean ◦ Take apart to rinse out nebulizer and spray chamber parts TJA – use cleaning wire ARCOS ◦ SeaSpray Nebulizer ◦ Twister Spray Chamber ◦ Cross Flow ARCOS TJA Nebulizer and spray chamber ARCOS ARCOS – clean as needed – salt buildup TJA – clean ~once a year Torch TJA 25 seconds P K Time (seconds) using pump speed of 60 rpm Setting the flush time ARCOS - Crossflow Transient scan – run standard Check to see when the peak intensities stabilize Clemson – 35 second total flush ◦ 25 seconds at 60 rpm ◦ 10 seconds at 30 rpm 18 seconds 25 seconds Crossflow spray chamber Twister spray chamber Transient scan on ARCOS P – 100 ppm Ca – 100 ppm K – 100 ppm Na – 50 ppm 25 seconds 45 seconds Time in seconds using pump speed of 60 rpm Transient scan going to rinse ARCOS - Crossflow Options for changing flush time ◦ ◦ ◦ ◦ Spray chamber Tubing length Pump speed Sample loop Ca - 40 seconds Pump speed 175 rpm Run time scan Check time for stability Clemson soil method ◦ pump speed 200 rpm ◦ flush time 30 seconds Setting the flush time - TJA Ca – 396.847 Ca – 315.887 Saturated signal Look at several recommended lines Clemson - Ca line – 317.933 for soil Line selection – ARCOS Peak position Set peak position Select clean area for background subtraction Look for uncontaminated blank Cu – 324.754 Blank correction blank Method development - ARCOS Run profile when needed On the AROCS - check intensity of P158.599 line If significant UV intensity drop – clean window and/or lens to optics Check new standard against old Read reference sample at start, mid point, and end of run Verify 2 check samples and blank within set Instrument profile and data verification