TOF Analysis of Bay Water surrounding the Battleship Texas Presented at: SETAC 2013 South Regional Conference Work By: Sue D’Antonio, Joan Stevens, Andre Szczesniewski , Lynne Marshall, John Kimmel and Ken Helmreich Agilent 6230 TOF LC/MS TOF and QTOF techniques (non-target compounds) • Identifies and differentiates compounds that were previously “hidden” under background • Accurate mass TOF instruments: – Can determine empirical formula based on the mass of the compound in question (accurate mass) Novel data mining techniques Accurate Mass Databases, Mass Profile Processional Software • Discover more compounds/contaminants, faster and with greater precision 6/27/2013 API-TOF Ion Optical Design Two Stage Ion Mirror TOF Mass Analyzer Effective Flight Path Length of 2.0m Beam Shaper Optically Coupled Ion Detector API Ion Source Page 3 6/27/2013 Formulae for TOF Performance 922.0092 Resolution = (m/z)/w1/2 1000/0.1 = 10,000 W1/2 = 0.087 923.0121 922/0.087 = 10,597 921.5 922.0 922.5 923.0 923.5 924.0 924.5 m/z, amu +TOF MS: 5.474 to 5.547 min from steroid2TOFap42.wiff Agilent, subtracted (4.696 to ... 1.16e5 Mass accuracy (mass error) (1000.001-1000.000)/1000)x106=1 ppm 1.10e5 1.00e5 9.00e4 In te n s ity , c o u n ts measured - actual Mass accuracy = 106 actual Max. 1.2e5 counts. 289.2159 8.00e4 Epitestosterone Resolution 6890 7.00e4 6.00e4 meas = 289.2159 5.00e4 4.00e4 actual = 289.2162 271.2051 3.00e4 Therefore: 1 ppm @ 1000 m/z is 0.001Da (1 mDa) -1.06 ppm 2.00e4 269.1893 287.1997 1.00e4 0.00 120 140 160 180 200 220 240 260 m/z, amu 280 300 320 340 360 380 400 6/27/2013 How Much Accuracy is Needed? (Newer concept for LC/MS: ppm accuracy) Reserpine (C33H40N2O9) has a protonated ion at 609.28066 Single quad reports mass to +/- 0.1 = 165 ppm Number of possible formulas using only C, H, O & N: • 165 ppm 209 • 10 ppm 13 • 5 ppm 7 • 3 ppm 4 • 2 ppm 2 Accurate mass reduces number of candidates and risk of investing effort with the wrong molecule Page 5 6/27/2013 SPE Sample Prep 6 6/27/2013 7 6/27/2013 8 6/27/2013 Analysis of Water Samples 9 6/27/2013 Choose Your Elements: Find Compounds by Molecular Feature and Generate Formulae 10 6/27/2013 Molecular Feature Generator 11 6/27/2013 Compound Chromatogram 12 6/27/2013 Deductive Compound Choices 13 6/27/2013 14 6/27/2013 15 6/27/2013 Conclusions Utilizing Agilent’s Time of Flight accurate mass, high resolution, superior isotopic fidelity and the unique features of Mass Hunter software, such as Molecular Feature Extractor and Molecular Formula Generator, we were able to identify over 100 compounds found in the samples. 16 6/27/2013