5250 Respiratory Gas Monitor REF: 6051-0000-125 and 6051-0000-126 Service Manual 5250 RGM S PAW 10:15 Tue 01-Sep-98 CO2 40 CO2 mmHg 0 SpO2 1 P aw 20 /39 I/Et 0 SpO2 I O S G N L 90 80 90 LIMITS RR 12 mmHg O2 PR 67 I/E % 670 TV ml ADULT MV 18/ -1 MENU 99 % 48/4 4 Paw (6) SpO2 8.0 N 2O cm H2O L O2 51% CKT. 48% WAVE 6050 0005 264 Printed in USA 5250 Respiratory Gas Monitor REF: 6051-0000-125 and 6051-0000-126 Service Manual Important This manual is subject to periodic review, update, and revision. Customers are cautioned to verify that the information in this manual applies to the software and hardware present in the equipment. Attention! Consult the accompanying instructions before using this device. This device performs as described in this manual, and in accompanying labels and/or inserts, when assembled, operated, maintained, and repaired in accordance with the instructions provided. This product must be cleaned and checked periodically. Do not use a defective product. Parts that are broken, missing, plainly worn, distorted, or contaminated should be replaced immediately. If repair or replacement becomes necessary, request service advice from Datex-Ohmeda (information is listed on the back cover). Do not repair this product or any of its parts other than in accordance with written instructions provided by Datex-Ohmeda. The user of this device shall have the sole responsibility for any malfunction that results from improper use, faulty maintenance, improper repair, unauthorized service, damage, or alteration by anyone other than Datex-Ohmeda. The safety, reliability, and performance of this device can be assured only under the following conditions: • If the device has been used according to the accompanying operating instructions. • If fittings, extensions, readjustments, changes, or repairs have been carried out by Datex-Ohmeda’s authorized agents. • If it is used in buildings having ground equalization wiring that complies with relevant local standards and regulations. Trademarks Datex® and Ohmeda® are the property of Instrumentarium Corp. or its subsidiaries. Nafion is a registered trademark of E.I. DuPont deNemours and Co., Inc. Tygon is a registered trademark of Norton Peformance Plastics All other product and company names are the property of their respective owners. Text revised: March 2000 © 1998, 1999, 2000 Datex-Ohmeda, Inc. All rights reserved. Table of Contents 1/Overview 1.1 Introduction............................................................................................................................................ 1-1 Related information...............................................................................................................1-1 Technical competence.........................................................................................................1-1 1.2 Features..................................................................................................................................................... 1-2 1.2.1 Display panel.............................................................................................................................1-2 1.2.2 Gas sample measurement..................................................................................................1-2 1.2.3 Internal O2 measurement..................................................................................................1-2 1.2.4 Airway pressure measurement.......................................................................................1-2 1.2.5 SpO2 measurement...............................................................................................................1-3 1.2.6 Circuit O2 measurement....................................................................................................1-3 1.2.7 Flow volume measurement ..............................................................................................1-3 1.2.8 Alarm features ..........................................................................................................................1-3 1.2.9 Ventilator interface................................................................................................................1-3 1.3 Major components .............................................................................................................................. 1-4 1.3.1 Chassis assembly.....................................................................................................................1-4 1.3.2 Display panel assembly....................................................................................................... 1-4 1.3.3 Front panel assembly...........................................................................................................1-4 1.3.4 Rear panel assembly.............................................................................................................1-4 Line voltage selection ...........................................................................................................1-5 1.4 Specifications........................................................................................................................................ 1-5 Electromagnetic effects........................................................................................................1-5 EMC performance ..................................................................................................................1-5 Compliance with standards..............................................................................................1-6 General safety requirements ............................................................................................1-6 1.4.1 Basic monitor.............................................................................................................................1-6 Physical characteristics ....................................................................................................... 1-6 Compensation ...........................................................................................................................1-7 Sample flow rate .....................................................................................................................1-7 Environmental characteristics.........................................................................................1-7 Range and accuracy..............................................................................................................1-7 1.4.2 SpO2 option ...............................................................................................................................1-8 Sensor emitter wavelength ranges................................................................................1-8 Displays......................................................................................................................................... 1-8 1.4.3 Gas accuracy..............................................................................................................................1-8 1.4.4 Analog outputs..........................................................................................................................1-9 Connector pinouts ..................................................................................................................1-9 Channel assignments............................................................................................................1-9 Strip chart calibration........................................................................................................1-10 1.4.5 RS-232 communications....................................................................................................1-10 Connector pinout, type DB-9...........................................................................................1-11 1.5 Precautions........................................................................................................................................... 1-11 1.5.1 Warnings....................................................................................................................................1-11 1.5.2 Cautions......................................................................................................................................1-13 1.6 Safety guidlines................................................................................................................................. 1-15 1.7 Cleaning the monitor ..................................................................................................................... 1-16 i Table of Contents 2/Theory of Operations 2.1 Power input assembly ...................................................................................................................... 2-1 2.1.1 Connectors ..................................................................................................................................2-1 2.1.2 Toroid.............................................................................................................................................2-2 2.1.3 Fuses...............................................................................................................................................2-2 2.2 Power supply board........................................................................................................................... 2-2 2.2.1 General..........................................................................................................................................2-2 2.2.2 Main regulators ........................................................................................................................2-2 Major components for each channel...........................................................................2-2 2.2.3 Power monitor..........................................................................................................................2-3 2.2.4 Display preregulators............................................................................................................2-3 2.2.5 Frequency monitor................................................................................................................2-3 2.2.6 Infrared photometer source switch..............................................................................2-3 2.3 Motherboard........................................................................................................................................... 2-4 2.3.1 General description...............................................................................................................2-4 2.3.2 Rear panel connectors.........................................................................................................2-4 Protection.....................................................................................................................................2-4 Isolated RS-232..........................................................................................................................2-4 2.3.3 Display serial driver..............................................................................................................2-5 2.3.4 Configuration jumper SW1...............................................................................................2-5 2.4 Pneumatics description ................................................................................................................... 2-5 2.4.1 Routing the gas sample....................................................................................................... 2-5 2.4.2 Conditioning the gas .............................................................................................................2-5 2.4.3 Normal flow with O2............................................................................................................2-6 2.4.4 Normal flow without O2.....................................................................................................2-7 When the RGM is sampling inhaled gas ..................................................................2-7 When the RGM is sampling exhaled gas..................................................................2-7 2.4.5 Purge flow ...................................................................................................................................2-8 2.4.6 Manual/auto zero flow.........................................................................................................2-9 2.4.7 Analyzing the dry gas.........................................................................................................2-10 2.4.8 Exhausting the gas................................................................................................................2-10 2.5 Pneumatics interface board....................................................................................................... 2-11 2.5.1 General........................................................................................................................................2-11 2.5.2 Pressure and flow .................................................................................................................2-11 2.5.3 Water trap full sensor.........................................................................................................2-11 2.5.4 O2 amplifiers...........................................................................................................................2-12 2.5.5 Temperature ............................................................................................................................2-12 2.5.6 Solenoid drivers.....................................................................................................................2-12 2.5.7 Test points .................................................................................................................................2-12 2.5.8 Pump driver.............................................................................................................................2-13 2.6 IR gas analyzer.................................................................................................................................. 2-13 2.6.1 IR photometer (ACX-200)..................................................................................................2-13 ACX-200 measurement board........................................................................................2-13 IR photometer .........................................................................................................................2-14 Internal and external bus.................................................................................................2-16 Memory.......................................................................................................................................2-16 Reference voltages................................................................................................................2-16 Measuring electronics for CO2, N2O, and AA......................................................2-16 A/D conversion.......................................................................................................................2-17 D/A conversion.......................................................................................................................2-17 ii Table of Contents Timing of CO2, N2O, and AA signals .......................................................................2-17 Motor speed control.............................................................................................................2-17 Pressure measurement......................................................................................................2-17 Temperature measurement ............................................................................................2-17 2.7 Signal processor board.................................................................................................................. 2-19 2.7.1 General description.............................................................................................................2-19 2.7.2 Microprocessor .......................................................................................................................2-19 2.7.3 Memory.......................................................................................................................................2-19 2.7.4 Decoding ....................................................................................................................................2-19 2.7.5 Watchdog...................................................................................................................................2-19 2.7.6 Configuration input switches.........................................................................................2-20 2.7.7 Analog acquisition................................................................................................................2-20 RGM analog input assignments....................................................................................2-20 2.7.8 Analog output..........................................................................................................................2-21 2.7.9 TVX interface ..........................................................................................................................2-21 2.7.10 SpO2 oximeter (MINX) interface.................................................................................2-21 2.7.11 Status LEDs...............................................................................................................................2-21 2.7.12 Serial I/O...................................................................................................................................2-22 2.7.13 Other inputs.............................................................................................................................2-22 2.7.14 Test points.................................................................................................................................2-22 2.8 SpO2 oximeter (MINX) board.................................................................................................. 2-23 2.8.1 Description................................................................................................................................2-23 2.8.2 Power supply...........................................................................................................................2-23 2.8.3 Control inputs.........................................................................................................................2-23 2.9 Display controller board............................................................................................................... 2-24 2.9.1 General description.............................................................................................................2-24 2.9.2 Processor and interfacing.................................................................................................2-24 2.9.3 System memory .....................................................................................................................2-24 2.9.4 Communicating to the signal processor..................................................................2-24 2.9.5 Watchdog circuit...................................................................................................................2-24 2.9.6 Programmable logic............................................................................................................2-25 2.9.7 Power regulation...................................................................................................................2-25 2.9.8 Audio............................................................................................................................................2-25 2.9.9 Touch screen interface......................................................................................................2-25 2.9.10 Video ............................................................................................................................................2-26 Graphic CPU...........................................................................................................................2-26 Video memory and video output circuits...............................................................2-26 2.9.11 Test points.................................................................................................................................2-26 2.10 Display matrix assembly .............................................................................................................. 2-27 2.10.1 Input signal descriptions..................................................................................................2-27 2.10.2 Input connector configuration (J1) ..............................................................................2-28 2.10.3 Optical characteristics .......................................................................................................2-28 3/Checkout and Calibration 3.0 General...................................................................................................................................................... 3-1 3.0.1 Special tools required for checkout and calibration procedures...............3-1 3.1 Functional checkout.......................................................................................................................... 3-2 iii Table of Contents 3.2 Dip switch and jumper positions............................................................................................... 3-3 3.2.1 On the signal processor board ........................................................................................3-3 3.2.2 SW1 on the motherboard...................................................................................................3-4 3.3 Power-up self-test................................................................................................................................. 3-4 3.3.1 Power-up functions................................................................................................................3-5 3.4 Power supply checks ........................................................................................................................ 3-5 3.5 Pneumatic system checks .............................................................................................................. 3-6 3.5.1 Sample flow ...............................................................................................................................3-6 3.5.2 Pump leakage............................................................................................................................3-8 3.5.3 System leakage .........................................................................................................................3-8 3.5.4 Pump vacuum ..........................................................................................................................3-8 3.5.5 Zero valve test...........................................................................................................................3-8 3.5.6 Purge valve test ........................................................................................................................3-9 3.5.7 O2 valve test...............................................................................................................................3-9 3.5.8 Water trap sensor test...........................................................................................................3-9 3.5.9 Purge and sample line blocked test...........................................................................3-10 3.6 Calibration checks........................................................................................................................... 3-10 3.6.1 Barometric pressure check..............................................................................................3-10 3.6.2 Barometric pressure calibration...................................................................................3-11 Pressure calibration.............................................................................................................3-11 3.6.3 PAW check.................................................................................................................................3-13 3.6.4 Airway pressure calibration...........................................................................................3-14 3.6.5 Gas analyzer check—zero and span calibration..................................................3-15 Zero and span calibration...............................................................................................3-15 3.6.6 Circuit O2 check ...................................................................................................................3-16 3.7 SpO2 module checks (if installed)........................................................................................ 3-17 3.8 Patient flow sensor check (if installed)............................................................................... 3-18 3.8.1 Flow threshold test...............................................................................................................3-18 3.8.2 Reverse flow detection simulation..............................................................................3-20 3.9 Sample flow calibration............................................................................................................... 3-20 3.10 Internal temperature calibration............................................................................................ 3-21 3.11 Analog output check ...................................................................................................................... 3-22 Channel assignments..........................................................................................................3-23 3.12 RS-232 loop back hardware check....................................................................................... 3-24 3.13 Leakage current and ground resistance checks ............................................................ 3-24 4/Messages and Troubleshooting 4.1 Messages ................................................................................................................................................... 4-1 4.2 Troubleshooting information........................................................................................................ 4-4 4.2.1 Power-up self-test.....................................................................................................................4-5 System error messages.........................................................................................................4-5 4.2.2 Using service screens...........................................................................................................4-6 Service screen...........................................................................................................................4-6 4.2.3 Enter Calibration Constants screen..............................................................................4-9 Procedure to recall barometer factory calibration constants ........................4-9 Pneumatics Service Screen............................................................................................4-10 iv Table of Contents 4.3 Troubleshooting tables.................................................................................................................. 4-11 4.3.1 Nonfunctioning monitor or power supply.............................................................4-11 4.3.2 TVX/patient flow sensor problems.............................................................................4-11 4.3.3 Communication fail or signal processor failure..................................................4-12 4.3.4 O2 sensors/O2 calibration problems.........................................................................4-13 4.3.5 Pneumatics problems.........................................................................................................4-14 4.3.6 Gas analyzer/calibration problems............................................................................4-15 4.3.7 Display assembly problems............................................................................................4-16 4.3.8 SpO2 oximeter board problems...................................................................................4-16 4.3.9 Intermittent problems or unusual behavior..........................................................4-17 5/Repair Procedures 5.1 Maintenance schedule..................................................................................................................... 5-1 5.2 Service policy........................................................................................................................................ 5-2 5.2.1 Obtaining service......................................................................................................................5-2 5.3 Cover removal and replacement................................................................................................ 5-3 5.4 Front panel.............................................................................................................................................. 5-4 5.4.1 SpO2 connector replacement..........................................................................................5-4 5.4.2 Power switch replacement................................................................................................5-5 5.4.3 Display panel removal and replacement..................................................................5-5 5.4.4 Display cable replacement................................................................................................5-5 5.4.5 Display panel disassembly ................................................................................................5-6 5.4.6 Display controller board replacement........................................................................5-7 5.4.7 Display matrix panel replacement ...............................................................................5-7 5.4.8 Display touch panel assembly replacement ...........................................................5-7 5.4.9 Display panel reassembly ..................................................................................................5-8 5.5 Rear panel............................................................................................................................................... 5-8 5.5.1 Fan replacement .....................................................................................................................5-8 5.5.2 Rear panel I/O board replacement..............................................................................5-9 5.6 Photometer replacement................................................................................................................. 5-9 5.7 Toroid with power input module........................................................................................... 5-10 5.7.1 Toroid assembly replacement .......................................................................................5-10 5.7.2 Secondary fuse replacement..........................................................................................5-12 5.8 Circuit board removal and replacement............................................................................ 5-13 5.8.1 Power supply board assembly......................................................................................5-13 5.8.2 Signal processor board ......................................................................................................5-14 Battery replacement ............................................................................................................5-15 5.8.3 SpO2 oximeter board .........................................................................................................5-16 5.8.4 Measurement board ............................................................................................................5-17 5.8.5 Motherboard ............................................................................................................................5-17 5.9 Pneumatics system ............................................................................................................................. 5-18 5.9.1 Pneumatics tubing assembly removal and replacement..............................5-18 5.9.2 Zero solenoid valve replacement ................................................................................5-20 5.9.3 Purge solenoid valve replacement..............................................................................5-20 5.9.4 Flow restrictor replacement ...........................................................................................5-20 5.9.5 O2 solenoid valve replacement....................................................................................5-21 5.9.6 Zero inlet kit replacement...............................................................................................5-21 5.9.7 Pump replacement ..............................................................................................................5-21 5.9.8 Scrubber replacement.......................................................................................................5-22 v Table of Contents 6/Illustrated Parts List 6.1 5250 main unit assembly .............................................................................................................. 6-2 6.1.1 Parts list ........................................................................................................................................ 6-2 6.1.2 Illustration ...................................................................................................................................6-3 6.2 Access panel assembly..................................................................................................................... 6-4 6.2.1 Parts list ........................................................................................................................................ 6-4 6.2.2 Illustrations, front and back......................................................................................6-5, 6-6 6.3 Display filter PCA ............................................................................................................................... 6-7 6.3.1 Parts list ........................................................................................................................................ 6-7 6.3.2 lllustration...................................................................................................................................6-7 6.4 Display assembly ................................................................................................................................. 6-8 6.4.1 Parts list ........................................................................................................................................ 6-8 6.4.2 Illustration ...................................................................................................................................6-8 6.5 Rear panel assembly ......................................................................................................................... 6-9 6.5.1 Parts list ........................................................................................................................................ 6-9 6.5.2 Illustration ...................................................................................................................................6-9 6.6 Signal processor board.................................................................................................................. 6-10 6.6.1 Parts list ......................................................................................................................................6-10 6.6.2 Illustration .................................................................................................................................6-10 6.7 Pneumatics assembly..................................................................................................................... 6-11 6.7.1 Parts list ......................................................................................................................................6-11 6.7.2 Illustration .................................................................................................................................6-12 6.7.3 Pneumatics interface board parts list .......................................................................6-12 6.7.4 Pneumatics routing..............................................................................................................6-13 6.8 Photometer and O2 sensor assembly.................................................................................... 6-14 6.8.1 Parts list ......................................................................................................................................6-14 6.8.2 ACX 200 illustration ............................................................................................................6-14 6.9 Motherboard........................................................................................................................................ 6-15 6.9.1 Parts list ......................................................................................................................................6-15 6.9.2 Illustration .................................................................................................................................6-15 6.10 Toroid assembly ................................................................................................................................ 6-16 6.10.1 Parts list ......................................................................................................................................6-16 6.10.2 Illustration .................................................................................................................................6-16 6.11 Water trap sensor assembly ....................................................................................................... 6-17 6.11.1 Parts list ......................................................................................................................................6-17 6.11.2 Illustation...................................................................................................................................6-17 6.12 SpO2 (MINX) oximetry assembly .......................................................................................... 6-18 6.12.1 Parts lists....................................................................................................................................6-18 6.12.2 Illustration .................................................................................................................................6-18 6.13 Upgrade kits ........................................................................................................................................ 6-18 6.14 Special service tools........................................................................................................................6-18 6.15 Optional service parts.................................................................................................................... 6-19 6.16 Additional parts................................................................................................................................. 6-19 A/Material Safety Data Sheets vi List of Figures Figure Description Page 2-1 2-2 2-3 2-4 2-5 2-6 2-7 2-8 SW1 motherboard.....................................................................................................................2-5 Normal flow with O2...............................................................................................................2-6 Normal flow without O2........................................................................................................2-7 Purge flow to clear blocked line.......................................................................................2-8 Manual/auto zero flow ...........................................................................................................2-9 ACX-200 measuring board block diagram...............................................................2-14 ACX-200 measuring board timing diagram ............................................................ 2-18 SW1 signal processor configuration switch............................................................ 2-20 3-1 3-2 3-3 3-4 3-5 3-6 3-7 3-8 3-9 3-10 3-11 3-12 3-13 3-14 3-15 3-16 3-17 3-18 SW1 signal processor configuration switch...............................................................3-3 SW1 motherboard.....................................................................................................................3-4 Select Service mode ................................................................................................................3-6 Service screen .............................................................................................................................3-7 Pneumatics Service Screen.................................................................................................3-7 Calibration setup....................................................................................................................3-11 Calibration screen..................................................................................................................3-11 Barometric Pressure Calibration screen ...................................................................3-12 Airway pressure test setup................................................................................................3-13 Airway Pressure Calibration screen............................................................................3-14 Cal gas connection.................................................................................................................3-15 Flow threshold test setup...................................................................................................3-18 Select adult patient type from Setup Screen.........................................................3-19 Activate reverse flow alarm on Setup Screen .......................................................3-20 Sample Flow Calibration screen................................................................................... 3-20 Flowmeter and flow restrictor connections............................................................ 3-21 Analog output diagnostic screen ................................................................................... 3-22 Analog output connector....................................................................................................3-23 4-1 4-2 4-3 4-4 Service screen .............................................................................................................................4-6 Enter Calibration Constants screen ................................................................................4-9 Pneumatics Service Screen..............................................................................................4-10 Communications, signal processor, or power supply fail and LED status.................................................................................................................4-12 5-1 5-2 5-3 5-4 5-5 5-6 5-7 5-8 5-9 5-10 Cover removal and replacement......................................................................................5-3 SpO2 connector replacement.............................................................................................5-4 Display panel disassembly...................................................................................................5-6 Photometer replacement.......................................................................................................5-9 Toroid assembly removal...................................................................................................5-10 Ground wire connection ....................................................................................................5-11 Power supply fuse replacement ....................................................................................5-12 Power supply assembly removal/replacement.....................................................5-13 Signal processor, SpO2, and measurement circuit board removal...........5-16 Pneumatics assembly replacement .............................................................................5-18 vii viii 1/Overview This chapter contains: • An overview of the features and major components of the 5250 Respiratory Gas Monitor (RGM). • Complete specifications for the monitor’s components. • A list of the precautions you must follow when using or servicing this monitor. • Safety guidelines to follow when servicing this monitor. • Instructions for cleaning the monitor and its components. 1.1 Introduction This service manual describes the following 5250 RGM models: RGM with agent (REF: 6051-0000-126) RGM with agent and SpO2 (REF: 6051-0000-125) Related information The information and instructions in this manual assume you are familiar with the operation of the 5250 RGM. Refer to the 5250 RGM Operation and Maintenance Manual for details regarding the features, functions, and use of the monitor. For information on oximeter sensor application and cleaning (reusable sensors only), see the instructions for the sensor. Replacement parts whose part numbers are listed in this manual can be ordered from Datex-Ohmeda. If you need to reference printed circuit board schematics and components, purchase and refer to the information contained in the 5250 RGM PCA Drawing Service Kit. See chapter 6 for part numbers. Technical competence CAUTION: Only qualified service personnel should perform the procedures described in this manual. Only Datex-Ohmeda service personnel or competent individuals who are experienced with servicing medical devices of this nature should perform the procedures described in this manual. 1-1 5250 RGM Service Manual 1.2 Features 1.2.1 Display panel The 5250 RGM uses a 4 x 8-in. amber electroluminescent flat display panel. The panel has infrared (IR) touch controls for using the RGM’s software operations and is removable for easy access and viewing. The RGM software program includes several service screens to aid in troubleshooting. 1.2.2 Gas sample measurement The RGM uses the principle of IR absorption to measure the amount of CO2, N2O, and anesthetic agent present in a gas sample. A small gas sample is continuously drawn into the pneumatic system through the sample inlet, and then routed through various components for conditioning (water separator, filters), analysis (photometer), flow control (pump, valve), and finally exhausted through a fitting (sample exhaust) on the back panel of the RGM. The signal processor board monitors the flow within the pneumatic system (differential pressure sensor on the pneumatics interface board). A sudden drop in the flow or barometric pressure, indicating a blocked line or condensate in the sample line, causes the microprocessor to switch pneumatic valves, thereby shifting the sample flow through the water purge path of the pneumatics line to clear the condensate from the sample line. The IR analyzer (photometer and measurement board) determines the content of CO2, and N2O, and agent in the gas sample and sends the waveform information to the signal processor as a digital serial stream. The stored values are continuously updated and displayed by the software program as numeric values, continuous waveforms, trend data, and respiratory rate data. The RGM determines a breath rate from the CO2 activity. 1.2.3 Internal O 2 measurement An internal O2 sensor measures the O2 concentration of the sample gas. The response time of the sensor cartridge is very slow, so the signal processor triggers the solenoid sample valve alternately to sample inspired and expired gases. The presence of CO2 in the sample triggers the collection of exhaled O2 concentrations; while the absence of CO2 triggers the collection of inhaled O2. For safety reasons, the solenoid is normally open to the O2 sensor when it is not energized. This allows the RGM to continuously measure and display the average O2 concentrations. 1.2.4 Airway pressure measurement The pressure transducer on the pneumatics interface board measures the airway pressure (PAW) when the patient circuit is connected to the PAW inlet. 1-2 1/Overview 1.2.5 SpO2 measurement The SpO2 sensor provides continuous, real time SpO2, and pulse rate readings. Trend information is available on the display. The RGM determines the patient's arterial oxygen saturation (SpO2) and pulse rate by measuring the absorption of selected wavelengths of light. The light generated in the sensor passes through the tissue and is converted to an electronic signal by the photodetector. (Some light is absorbed by the tissue.) The electronic signal passes to the oximeter and is amplified. The oximeter's circuitry processes the signal, converting the light intensity information into SpO2 and pulse rate values. The patient data then appear. 1.2.6 Circuit O 2 measurement The circuit O2 sensor monitors the O2 concentration at an external sensor cartridge. The O2 sensing cartridge produces a voltage that is proportional to the O2 concentration (partial pressure) at its detecting surface. The signal processor board converts the sensor signal into a corresponding percent O2. The O2 values then appear. 1.2.7 Flow volume measurement The RGM uses a turbine vane transducer cartridge and a sensor clip, placed in the patient breathing circuit, to convert gas flow into electrical pulses. A clip-on optical coupler sensor fastens to the transducer cartridge and connects to the RGM with a flexible, extendible, coiled cord. The signal processor analyzes the signals from the transducer to determine tidal and minute volumes, whose values appear on the screen. 1.2.8 Alarm features The RGM generates various alarms and messages to indicate conditions of the gas sample, SpO2 sensor, O2 sensor, flow sensor, and of the RGM itself. The RGM evaluates each condition, compares it to an operator-selected limit or to a system default value and generates audible alarms, visual alarm indicators, and alphanumeric messages indicating the cause of the alarm, which may be patient (gas sample) or RGM related. 1.2.9 Ventilator interface The RGM can be used with Datex-Ohmeda 7800, 7810 (referred to as 78xx in the following chapters), and 7900 ventilators. An RS-232 cable connects the ventilator to the RGM and transmits values (TV, MV, PAW [min, max, inspiratory pause], and circuit O2) calculated by the ventilator to the RGM, which displays the values received from the ventilator. 1-3 5250 RGM Service Manual 1.3 Major components The RGM is made up of four major assemblies: the chassis, the display panel, the front panel, and the rear panel. 1.3.1 Chassis assembly • Toroid assembly • Power supply assembly • Motherboard • Pneumatics assembly • Pneumatics interface board • ACX photometer • ACX measurement board • Signal processor board • SpO2 oximeter (MINX) board • Signal processor software module 1.3.2 Display panel assembly • The display panel and extension cord • Display processor board • Display matrix assembly • Touch panel • Display processor software module 1.3.3 Front panel assembly • Gas sample inlet • PAW inlet • SpO2 connector • Power switch • Display release • Water separator assembly 1.3.4 Rear panel assembly 1-4 • Cooling fan • Gas sample exhaust • Software module access cover • Equipotential ground connector • Power input receptacle (line filter, fuses, voltage selector) • I/O panel PCA 1/Overview Line voltage selection The 5250 RGM covers four (100, 120, 220, 230-240) line voltage applications. • As received in the USA, the RGM is set up to operate from a 120 V AC power source (50/60 Hz). If the RGM is ordered with the correct country option, the line voltage and fuses will be set appropriately for the destination country. Note: If the local line voltage in your location is 230 V, always use the 240 V setting on the voltage drum that is in the power module. • To use the RGM from an alternate power source, change the fuses, the voltage selector switch (drum), and the power cord to match the application. Refer to the 5250 RGM Operation and Maintenance Manual. CAUTION: Make sure the selected monitor voltage agrees with the local voltage available. 1.4 Specifications All specifications are based at sea level and subject to change without notice. Electromagnetic effects Indications that the 5250 is experiencing electromagnetic interference include the following: • Variations in the waveform display. • Sudden increases or decreases in the waveform height that do not correlate to the physiological condition of the patient. • Messages that cannot be resolved by the instructions found in this Operator’s Manual. • The display of dashes on numeric LEDs when a valid physiological signal is present. This interference may be intermittent. Careful correlation between the effect and its possible source is important. The 5250 will not display any of these indications if it is used within its intended electromagnetic environment as described in IEC/EN 60601-1-2 (1993). EMC performance The 5250 complies with the requirements of IEC/EN 60601-1-2 (1993): Electromagnetic compatibility - Requirements and test. The following EMC standards were applied to verify conformance: Immunity: IEC/EN 61000-4-2 (1995), 8 kV air, 3 kV contact IEC/EN 61000-4-3 (1995), 3 V/m IEC/EN 61000-4-4 (1995), 1 kV power, 500 V I/0 IEC/EN 61000-4-5 (1995), 2 kV line to earth, 1 kV line to line Emissions: IEC/EN 55011 Group 1, Class B (1997) The 5250 was tested with an RS-232 cable attached (no printer or CPU) when operating on AC power. 1-5 5250 RGM Service Manual Compliance with standards The presence on the monitor of any symbol described below indicates compliance with the standard represented by that symbol. 0197 Medical Device Directive 93/42/EEC of the European Union for a class I (with a measuring function), IIa, IIb, or III device. Medical Device Directive 93/42/EEC of the European Union for a class I device. Medical electrical equipment classified with respect to electric shock, fire, and mechanical hazards only, in accordance with the Canadian Standards Association CAN/CSA C22.2 No. 125 (1984). Medical electrical equipment classified in the US and Canada with respect to electric shock, fire, and mechanical hazards only, in accordance with the Canadian Standards Association CAN/CSA C22.2 No. 601.1 (1990) and Underwriters Laboratories Inc. UL 2601-1 (1997). General safety requirements The 5250 complies with the requirements of IEC/EN 60601-1 (1988) Part 1: General requirements for safety of medical electrical equipment, including Amendment 1 (1991) and Amendment 2 (1995). Type B applied part. Type BF applied part. IPX1 Protected against vertically falling water drops. Type of protection against electric shock: Class I Mode of operation: Continuous 1.4.1 Basic monitor Physical characteristics Main chassis Size: 12.4 in. W x 7.5 in. H x 16 in. L Weight: Less than 25 lbs. Power-fuse rating 100 Vac 50/60 Hz 1.0A T1.5A/250 V 120 Vac 50/60 Hz 1.0A T1.5A/250 V 220-230/240 Vac 50/60 Hz 0.5A T1.25A/250 V Voltage tolerance +10 to -15 percent 1-6 1/Overview Compensation CO2, N2O, and O2 for barometric pressure. CO2 for O2, and N2O Sample flow rate Sample flow rate = 190 ±40 mL/min Environmental characteristics Operating Temperature 15º to 40 °C Humidity 0 to 95% RH, noncondensing Pressure 500 to 800 torr Storage and transportation Temperature -5º to +50 °C Humidity 0 to 95% RH, noncondensing Pressure 5 to 20 psia Range and accuracy Measured parameter Displayed parameter Range Accuracy *Response time Resolution O2 FiO2 0 to 100% ±3.0% O2 (<60%) 20-60 sec. 1% 20 sec. 1% ±5.0% O2 (≥60%) O2 Circuit Ckt O2 Flow Tidal volume Pediatric: Minimum flow of 5 LPM in breath 50 to 200 mL Adult: Minimum flow of 8 LPM in breath 80 to 200mL 200 to 800mL 800 to 2500 Minute 0 to 50.0 L volume Pressure 0 to 59% 60 to 100%* ±3.0% O2 ±5.0% O2 ±40mL May take up to 1 minute to adjust to new patient, then every breath 5mL ±40mL ±10% or 40mL ±10% Sum of tidal volumes 5 sec. or end of for last minute breath 5mL 10mL 50mL 0.1 L Waveform bar graph 5 to 100 LPM ±10% Sampling time 10 milliseconds settling time 1 second 0.4 LPM on analog output Waveform -20 to 120 cm H2O ±5 cm H2O 500 msec 1 cm H2O Mean 0 to 99 cm H2O 1-7 5250 RGM Service Manual 1.4.2 SpO2 option SpO2 pleth waveform: Has no units SpO2 accuracy (within 1 standard deviation): 80 to 100% ± 2% 60 to 79% ± 3% Below 60% unspecified Pulse rate range: 40 to 235 bpm Pulse rate accuracy: ±1.7% of reading Sensor emitter wavelength ranges Red LED peak wavelength range: 650 to 665 nm Infrared (IR) LED peak wavelength range: 930 to 950 nm Average power: ≤ 1 mW Displays The displayed SpO2 and pulse rate values are updated every 2 seconds. The plethysmographic waveform sweep is updated every 6 seconds (fast rate— 12.5 mm/sec.) or 12 seconds (slow rate—6.5 mm/sec.). 1.4.3 Gas accuracy Internal pressure range: 450 to 850 mmHg Operating temperature range: 15 to 40 °C CO2: 1-8 Measuring range: 0 to 8% 8.1 to 15% Zero drift: ±0.2% CO2 /24 hr unspecified Gain drift: ±0.2% CO2 /24 hr unspecified Temperature zero drift: ±0.2% CO2 /10 °C unspecified Temperature gain drift: ±0.2% CO2 /10 °C unspecified Response time: ≤400 ms @ 200 mL/min unspecified Linearity: ±0.3% CO2 unspecified N 2O : Measuring range: 0 to 100% Zero drift: ±2% N2O/24 hr Temperature zero drift: ±2% /10 °C Temperature gain drift: ±3% /10 °C Response time: ≤400 ms @ 200 mL/min 1/Overview Agent: Measuring range: Gain stability: Zero drift: 0.0 to 5% halothane, enflurane, isoflurane ±0.4% halothane/24 hr ±0.6% halothane/24 hr Linearity: Temperature zero drift: Temperature gain drift: Response time: ±0.2% halothane ±0.6% halothane/10 °C ±0.4% halothane/10 °C ≤500 ms @ 200 ml/min Agent: Measuring range 0.0 to 5% desflurane 5.1 to 18.0% 18.1 to 30.0% Temperature gain drift: ±0.4% desflurane/10 °C ± 0.1% unspecified Response time: ≤500 ms Agent: Measuring range 0.0 to 8% sevoflurane 8.1 to 15.0% Temperature gain drift: ±0.4% sevoflurane/10 °C unspecified Response time: ≤500 ms 1.4.4 Analog outputs Connector pinouts CAUTION: Maximum voltage. No more than 5 V should appear on any pin of the analog output connector. Channel assignments Channel Scale 1 2 3 4 5 6 7 66.6 mV/% 10 mV/% (V x 140)-20 = cm H2O 10 mV = 1 L/min 66.6 mV/% (Desflurane: 33.3 mV/%) 10 mV/% 0 to 0.5 V with 16.1 mV resolution CO2 waveform N2O waveform PAW waveform Flow waveform Agent waveform O2 inspired/expired Plethysmograph 1-9 5250 RGM Service Manual The analog output accuracy is ±3.5 percent (absolute), linearity ±0.5 percent, with an update rate of 10 mS and a resolution of 3.9 mV. Note: Output source impedance 100 Ω. Strip chart calibration 1. Select MENU, SETUP, VIEW ALL, and REC CAL. 2. Connect the Strip Chart recorder to the analog output connector. Pinout of the connector is shown in this section. 3. Select ZERO. All channels will be set to zero volts. Adjust the strip chart recorder. 4. Select ONE. All channels will be set to 1 volt (full scale). Adjust the strip chart recorder. 5. Press EXIT three times to return to the display screen. 1.4.5 RS-232 communications The following data will be output to the RS-232 port at the configured time interval. At the top of each page the following title will be output with 58 lines of data formatted under each column. Data communication is at 1200 baud, odd parity, 7 bits per character, 1 stop bit or 9600 baud, no parity, 1 stop bit, depending on setup configured. <ff> * :RG CO2 O2 INT PAW FLOW CKT Bar Agent X * :RG I E RR I E I E M TV MV N2 O O 2 Pres SpO2 PR I E <cr> :RG xx.x xx.x xxx xxx xxx xxx xxx xxx xxxx xx.x xxx xxx xxx xxx xxx xx.x xx.x <cr> Value indexes: 1 2 3 4 * 6 7 8 9 10 11 12 13 14 15 16 17 These channel labels are printed at the top of each page or when new agent is selected. 1. 2. 3. 4. 5. 6. 7. 8. 9. 10. 11. 12. 13. 1-10 5 CO2 inspired in % CO2 expired in % CO2 respiratory rate in breaths per minute O2 inspired in % O2 expired in % PAW inspired in cm H2O PAW expired in cm H2O PAW mean in cm H2O Tidal volume in milliliters Minute volume in liters Nitrous N2O in % Patient circuit O2 in % Barometric pressure in torr in sample chamber. Can be used to convert % CO2 to mmHg if desired. CO2 mmHg = CO2% x Barometric pressure /100. CO2 kPa = CO2 mmHg /7.5 1/Overview 14. SpO2 %. Only output if option installed. 15. SpO2 pulse rate in beats per minute from SpO2 board. Only output if option is installed. 16. Agent inspired in % (maximum 15%). Title line will change to indicate type of agent configured. “H” indicates halothane, “I” indicates isoflurane, “E” indicates enflurane, “D” indicates desflurane, “S” indicates sevoflurane, and “N” indicates no agent. 17. Agent expired in % (maximum 15%). Connector pinout, type DB-9 Pin Function 2 3 7 9 Transmitted data from monitor Received data into monitor Ground External alarm silence 1.5 Precautions Two types of precautionary notes are used in this manual: • Warnings indicate a potentially harmful situation that may cause injury to the patient or operator. • Cautions indicate a condition that may lead to equipment damage or malfunction. 1.5.1 Warnings Biohazard When handling materials that may have come into contact with patient exhalant or fluids, follow approved procedures for contamination control. Dispose of filter cartridges, water bottle contents, and sample lines in a receptacle designated for patient waste. Electrical shock hazard Before cleaning or repairing the monitor, power it off and disconnect it from AC mains power. Measure the leakage current whenever an external device is connected and at the completion of repair procedures. Leakage current must not exceed 100 microamperes. Do not touch any exposed wiring or conductive surfaces while the cover is removed. The voltages present when electrical power is connected to the RGM can cause serious injury or death. Never wear a grounding wrist strap when working on an energized monitor. The display generates potentially dangerous voltages (high voltage pulses up to 225 V ac). Do not touch the driver-controller circuit board during operation. 1-11 5250 RGM Service Manual Flammability and electrical shock hazard For continued protection against fire hazard and electric shock, replace fuses only with fuses of the same type and rating shown on the rear panel of the 5250 RGM. Patient safety Never test or perform maintenance on the monitor while it is being used to monitor a patient. The correct use of the oximetry function of this monitor is to measure only arterial oxygen saturation (SpO2) and pulse rate. • A pulse oximeter does not measure respiration and should never be used as a substitute for an apnea monitor or as the primary monitor for infants being monitored for apnea. • A pulse oximeter may be used during sleep studies with adults only to gather information regarding SpO2 and pulse rate. Patient safety (sensors) To prevent patient injury or equipment damage, use only Datex-Ohmeda sensors identifed for use with this monitor. For complete information about the safe and appropriate use of a sensor, consult the instructions for the sensor. Prolonged monitoring or patient condition may require you to periodically change the sensor site. If there is evidence of blistering, skin erosion, or ischemic skin necrosis (such as skin discoloration or reddening), change the site immediately. Otherwise, change the site as specified in the instructions for the sensor. Orient the flow sensor in the breathing circuit as indicated by the markings on the sensor clip. The arrows must point in the direction of the flow in the breathing circuit. If the clip is not mounted correctly, the 5250 RGM will not operate properly. After completing simulation testing, be sure to return the sensor to its correct orientation in the breathing circuit. The arrows should point away from the patient. Data validity Use only Datex-Ohmeda calibration gas of 1.75% halocarbon-23 (CHF3)/freon or equivalent 6% CO2, 40% N2O, 50% O2 and 2.25% argon. Use of the wrong calibration gas kit will result in improper calibration and the RGM will display incorrect gas concentrations. In the presence of alcohols, ketones, or other organic hydrocarbon vapors—in the sample line or patient circuit, and ethyl alcohol in the patient’s bloodstream—the 5250 RGM may not indicate accurate readings of anesthetic agent. The oxygen monitoring portion of the 5250 RGM monitor should be calibrated at the same temperature at which it will be used to monitor oxygen delivery in the patient circuit. Operation at temperatures other than those present during calibration may result in readings outside of the stated accuracy for the monitor. When the ambient temperature changes, we recommend recalibrating the monitor for maximum accuracy. Refer to the O2 sensor information sheet for more details. Keep the circuit O2 sensor attached to the monitor to assure accurate O2 readings. If detached, the sensor must be reattached and stabilized for up to 1 hour before recalibrating. 1-12 1/Overview Failure of operation Perform the checkout procedures before using the monitor with a patient. If the monitor fails any test, it must be removed from use until it has been repaired and checked for correct operation. System interconnection Accessory equipment connected to the RS-232 serial connector must be certified according to the current version of the respective IEC/EN standards (e.g., IEC 60950 for data processing equipment and IEC/EN 60601-1 for medical equipment). All configurations shall also comply with IEC/EN 60601-1-1. Anyone who connects additional equipment to the RS-232 port connector configures a medical system, and is therefore responsible that the system complies with the requirements of IEC/EN 60601-1-1. If in doubt, call your local authorized service office, as listed on the back cover of this manual. The 5250 RGM is referred to as an IEC 60601/B device in the summary of situations table contained in IEC/EN 60601-1-1. Service A damaged membrane separator may result in a contaminated agent photometer. Once contaminated the agent photometer assembly cannot be serviced and must be replaced. 1.5.2 Cautions Static sensitivity Internal electronic components are susceptible to damage by electrostatic discharge. To avoid damage when disassembling the monitor, observe the standard precautions and procedures for handling static-sensitive components. Cleaning When cleaning the exterior of the RGM, follow these precautions: • Do not autoclave, pressure sterilize, or gas sterilize the 5250 RGM. • Use cleaning solution sparingly. Do not immerse the monitor in liquid. Excessive solution may flow into the monitor and damage internal components. • Do not touch or rub the display area with abrasive cleaning compounds or any material that could damage the surface. • Do not use organic-, petroleum-,or acetone-based solutions, or other harsh solvents to clean the display area or the unit. These substances attack the device’s materials and device failure may result. To prevent permanent damage to the internal sensor and voiding its warranty, do not clean the sample chamber of an RGM with agent. Service and repair Only qualified service personnel should perform the procedures described in this manual. Servicing this product in accordance with the service manual should be undertaken only by qualified service personnel having the proper knowledge, tools, and test equipment. 1-13 5250 RGM Service Manual Disposal Dispose of this medical device and its packing materials in accordance with local procedures and regulations. Destroy malfunctioning flow cartridges to prevent their inadvertent use. Software The 5250 RGM software consists of two cartridges that must be replaced in sets to ensure software compatibility. The replacement procedures are located in the 5250 RGM Operation and Maintenance Manual. Pneumatic system To prevent damage to the 5250 pneumatic system, use only ventilators and anesthetic equipment that limit the pressure at the patient connection port to 12.5 kPa (125 cm H2O) in compliance with IEC/EN 60601-2-13 (1997). Follow the normal ventilator power-off sequence before you power off the 5250. General U.S. Federal law and Canadian laws restrict this device to sale by or upon the order of a licensed medical practitioner. After replacing the internal O2 sensor, perform the calibration procedure in section 3.6.6 to verify the RGM is working properly. Allow up to 1 hour to stabilize the sensor. Avoid storing O2 sensors outside the temperature range of -5 °C to 50 °C (+23 °F to 122 °F). O2 sensors must have shorting clips or be connected to the RGM when in storage or not in use. Make sure the selected monitor voltage agrees with the local voltage available. Never tamper with the set screws in the flow cartridge. Such action will render the cartridge unusable. Pressure in excess of 10 psi above atmospheric pressure could damage the PAW transducer or sample system barometric pressure transducer in the RGM. System calibrations may be erased if power is applied while the signal processor software module is not in place. When replacing either the microprocessor or the EPROM on the SpO2 oximeter (MINX) board, be sure to use a chip extractor to avoid damaging the components. Maximum voltage. No more than 5 volts should appear on any pin of the analog output connector. Do not attempt to repair or replace the photometer pressure transducer (B1). Each pressure transducer is specifically calibrated with selected resistors for use on a given board. 1-14 1/Overview 1.6 Safety guidelines Before you start any procedure that involves disassembly of the monitor, review these guidelines to ensure the proper and safe completion of the procedure. WARNINGS: • Patient safety. Never test or perform maintenance on the monitor while it is being used to monitor a patient. • Electrical shock hazard. Before cleaning or repairing the monitor, turn it off and disconnect it from AC mains power. • Biohazard. When handling materials that may have come into contact with patient exhalant or fluids, follow approved procedures for contamination control. 1. Power off and disconnect the unit from AC mains power. 2. Disconnect any pulse oximeter sensor from the monitor. 3. Disconnect the patient sample line and scavenger line from the monitor, empty the water bottle, and clean the monitor—see section 1.7. 4. Read and follow each step of the procedure. Give special attention to all warnings and cautions. While working on the monitor, wear safety eyeglasses and gloves when handling any component that may have come in contact with the patient’s exhalant gas or fluids. A mask and gown are also recommended. Important: After repairs are complete, test the monitor as directed in each procedure to verify that is is functioning properly. If the cover is removed for any test or repair procedure, always perform the Functional checkout (section 3.1) and Leakage current and ground resistance checks (section 3.13) before returning the unit for patient monitoring. Note: We recommend that the 5250 RGM Operation and Maintenance Manual be available for your reference. 1-15 5250 RGM Service Manual 1.7 Cleaning the monitor You must clean the monitor, • Before you start any test or repair procedure that involves disassembly of the monitor. • Before you send it to Datex-Ohmeda for repair. Equipment • Safety eyeglasses or face guard and waterproof gloves (gown and mask are also recommended) • Paper towels or soft cloth; cotton swab • Mild soap and water solution or cool, liquid cleaning solution, such as 70% isopropyl alcohol, or equivalent WARNINGS: Biohazard • Dispose of filter cartridges, water bottle contents, and sample lines in a receptacle designated for patient waste. • When handling materials that may have come in contact with patient exhalant or fluids, follow approved procedures for contamination control. WARNING: Electrical shock hazard. Before cleaning or repairing the monitor, turn it off and disconnect it from AC mains power. CAUTIONS: Cleaning. • Do not autoclave, pressure sterilize, or gas sterilize the 5250 RGM. • Use cleaning solution sparingly. Do not immerse the monitor in liquid. Excessive solution may flow into the monitor and damage internal components. • Do not touch or rub the display area with abrasive cleaning compounds or any material that could damage the surface. • Do not use organic-, petroleum-, or acetone-based solutions, or other harsh solvents, to clean the display area or the unit. These substances attack the device's materials and device failure may result. • To prevent permanent damage to the internal sensor and void its warranty, do not clean the sample chamber. 1. Turn off the monitor and disconnect the power cord from AC mains power. 2. Dispose of the sampling tube assemblies and/or PAW Tygon® tubing. 3. Remove the water trap bottle and filter cartridge, and dispose of any contents in a receptacle designated for patient waste. 4. Using a fine mist, spray the external surface of the RGM and the internal and external surfaces of the water bottle. Do not spray directly into the speaker or ventilation holes. 5. After 5 minutes, wipe the exterior surface of the monitor with a damp paper towel or soft cloth. Saturate a cotton swab with 70 % isopropyl alcohol and gently wipe the display area. Rinse the bottle with water. The monitor should be completely dry before use. 1-16 2/Theory of Operation Important: All references in the following descriptions are found in the 5250 Service Kit, PCA Drawings, part number 6050-0005-384. 2.1 Power input assembly The power input assembly consists of: • Power input module • Power switch • Toroid • Secondary fuses The power switch (SW1) on the front panel of the RGM connects the power input module to the toroid (T1) through connectors J2 and P1. 2.1.1 Connectors Connectors TB and FB connect the secondary windings of the toroid to the power supply board. The table below lists the corresponding connectors between the input assembly and the power supply. * Input assembly Power supply V ac Wire color TB5 TB4 E1 E2 0 (GND) 0 (GND) Brn Brn FB5 FB4 E5 E6 22.0 22.0 Red Orn TB3 TB2 E7 E8 0 (GND) 0 (GND) Blk Blk FB3 FB2 E9 E10 9.5 9.5 Grn Blu TB1 FB1 E11 E12 11.5 (RS-232) (RS-232) ]* Vio Vio These float with respect to system GND and can be measured only as a differential pair. 2-1 5250 RGM Service Manual 2.1.2 Toroid The toroid is a power-efficient step-down transformer, used in this unit for space and size considerations. 2.1.3 Fuses The primary fuses are located in the power input module. To replace the primary fuses, refer to the 5250 RGM Operation and Maintenance Manual. The secondary fuses are 5 x 20-mm fast blow types. A fuse connects to each secondary winding of the toroid. To replace the secondary fuses, refer to 5/Repair Procedures. 2.2 Power supply board 2.2.1 General The power supply board provides outputs of +5 V dc, -5 V dc, +15 V dc, and -15 V dc for all the boards in the RGM. In addition, it supplies +5V_ISO for the RS-232 connection. It also supplies +6.6 V dc and +15 V dc to the display. The power supply has a circuit that monitors the incoming power for an adequate voltage level for proper operation, and it employs conventional linear regulation to provide power free of switching transients. The inputs to the power supply are fused on a terminal strip connector. Another circuit passes a line frequency logic signal to the signal processor, which is used to synchronize the SpO2 (MINX) oximeter 2.2.2 Main regulators The regulators VR1, VR3, VR4, and VR5 are for main system operation while the VR6 is an isolated regulator for external serial communications.. Major components for each channel Channel Filter Regulator SCR LED +5 C13 VR4 CR13 CR18 -5 C16 VR5 CR22 CR19 +15 C18 VR3 CR2 CR16 -15 C15 VR1 CR3 CR17 V_ISO C20 VR6 CR27 CR14 Each regulator has an over-voltage protection circuit consisting of a zener diode and resistor to set a threshold and an SCR, which will short out the regulator if the threshold is exceeded. The main regulators have indicator LEDs that indicate the supply is on. Use a voltmeter to verify the supply is within specification. A decoupled +5-V output is provided (by L1 and C11) for the analog power for the SpO2 oximeter (MINX) board 2-2 2/Theory of Operations The +5-V regulator (VR4, an LT1084) is a low-overhead voltage part, with the transformer winding chosen for minimum overhead at line voltage of 103 V ac. Since the -5-V regulator is powered from of the same winding, a greater ratio of capacitance (C16) to current reduces ripple, and therefore provides the extra overhead for the standard regulator. 2.2.3 Power monitor The power monitor circuit has two functions. One is to monitor the main +5-V dc regulator's unregulated input to tell the signal processor when power has been shut off, giving it enough time to shut down in an orderly way. The circuit's output is routed to the watchdog chip, U18, on the signal processor board, which compares this to a 1.3-V internal reference. The watchdog chip will trip when the +5 V unregulated drops below a threshold, which varies from 6.9 to 8.4 V depending on revision levels and signal processor boards. This allows at least 12 mS of continuing operation before the regulator drops out. The other function of the circuit is to ensure that the voltage being monitored will not exceed the +5.5 V dc maximum input voltage to the watchdog chip. 2.2.4 Display preregulators Because the display operates at the end of a long cord, preregulation on the power supply minimizes the voltage drop for the final regulators in the display. The preregulator consists of a PNP pass transistor (Q1) controlled by an NPN transistor (Q4) collector driving the base of the PNP. A REF01 (U1) provides a 10-V reference to the circuit. Feedback is returned to the emitter of the NPN transistor, with diode (CR4) for temperature compensation. 2.2.5 Frequency monitor One of the transformer windings, E6, used for the 15-V outputs runs to a simple one-transistor amplifier to produce a square wave that tracks the polarity of the input sine wave. This is routed to the signal processor board, which measures the cycle time, and the signal processor board then informs the SpO2 oximeter (MINX) board of the current line frequency. 2.2.6 Infrared photometer source switch Q8 switching ON provides a current path to ground. This turns on the infrared (IR) source bench lamp in the photometer in response to the LAMP CTL signal from the photometer measurement board. Q7 and the surrounding resistor network provide a 0 or 15 V input to Q8 in response to a LAMP CTL +4 to +15 V ON command or 0 to -15 V OFF command. This range is necessary to accommodate the ACX analyzer’s CMOS logic output levels. 2-3 5250 RGM Service Manual 2.3 Motherboard 2.3.1 General description The motherboard interconnects the signal processor, the SpO2 oximeter (MINX) board, the IR photometer analyzer boards, the display, and the rear panel connectors. The system power enters the motherboard at J10 and is distributed to the boards and display. Serial drivers to the display and to the RS-232 port are also located on the motherboard. This board is not bussed, so each daughter board has a specific socket: J7 J8 J9 Signal processor board SpO2 oximeter (MINX) board Measurement board 2.3.2 Rear panel connectors The rear panel flow sensor connector is compatible only with the new TVX cable, which provides cable shielding. The RS-232 channel operates through a 9-pin DB-type connector J2, which is wired as follows: 2 3 5, 7 9 Transmit Receive Common Alarm silence, transition high to low. The analog output (J5) is wired with channels 1through 7 on consecutive pins running clockwise from the gap at the top of the connector and from the common on the center pin. (See section 1.4.4, Analog outputs.) Protection With the exception of the isolated RS-232 port, all of the rear panel connectors are protected against electrostatic discharge by 5-V transorbs (CR1-CR13). The TVX inputs have 4.7KΩ in series (R6, R9), while the analog outputs have 100Ω in series with each output and with the common (analog ground). The TVX interface also includes a multi-hole ferrite block (L1) in series with each lead to limit electromagnetic emissions. Isolated RS-232 The serial communications path is optically isolated by U3 for transmission and by U2 for reception. This allows the device attached to the RS-232 port to operate from a different ground reference. U1, operating from an isolated 5-V output of the power supply, converts between 5V logic levels and RS-232 levels. This IC uses capacitors C1 and C4-C6 to raise the 5 V to ±10 V. An isolated channel also provides the alarm silence function. 2-4 2/Theory of Operations 2.3.3 Display serial driver A driver (U4) converts the TTL/CMOS level serial string from the signal processor board into differential RS-422/485 levels for communications with the display through J11. 2.3.4 Configuration jumper SW1 A jumper pad (SW1) provides sense inputs for configuration information to the signal processor. Spare Spare ACX 200 Figure 2-1. SW1 motherboard 2.4 Pneumatics description The pneumatics system has four basic functions: • Drawing a gas sample from the patient breathing circuit via a gas sample tube. • Conditioning the gas by removing any water or other liquids as well as particulate matter. • Routing the clean, dry gas through the appropriate sensors for analysis. • Exhausting the gas to a scavenger system or back into the breathing circuit . 2.4.1 Routing the gas sample The RGM routes the sample gas through the pneumatics system four ways: • Normal flow with O2 cell sampling. • Normal flow with O2 cell bypassed. • Purge flow. • Manual/auto zero flow. 2.4.2 Conditioning the gas Before being analyzed, the gas sample is conditioned by the water separator. The membrane water separator extracts water and any other fluids from the gas sample. If the water trap bottle is full, a sensor in the water trap shuts down the pump, and the RGM displays the WATER TRAP FULL message. The RGM resumes normal operation after the water trap bottle is emptied. 2-5 5250 RGM Service Manual 2.4.3 Normal flow with O 2 Zero Filter Yellow 0.010 CO2 Scrubber Zero Valve O2 Gas Analyzer NC Patient Sample Inlet 90% Flow NO COM COM Nafion NO NC 0.04 in.ID 4 in. (CO2) or 20 in. (Agent) Water Separator O2 Valve Barometer Purge Valve NO COM 10% Flow Hydrophilic Filter Tubing Coil (for flow measurement) NC Green 0.005 Hydrophobic Filters (2) Aqua 0.007 Sample Exhaust Flow Sensors Pump Bacterial Filter Figure 2-2. Normal flow with O2 The O2 solenoid is de-energized (normally open position). The sample gas flows through the O2 sensor. The O2 sensor is used to sense both inhaled and exhaled gas. The flow rate is measured before the gas is exhausted through the pump and out of the machine. The sample flow rate ranges from 150 to 230 ml/min. The system is designed so that, during normal operation, 90 percent of the gas sample passes through the gas analyzer and O2 sensor while 10 percent flows through the water separator bottle and then through the purge branch of the pneumatics. 2-6 2/Theory of Operations 2.4.4 Normal flow without O 2 Zero Filter Yellow 0.010 CO2 Scrubber Zero Valve O2 Gas Analyzer NC Patient Sample Inlet NO COM 90% Flow COM Nafion NO NC 0.04 in.ID 4 in. (CO2) or 20 in. (Agent) Water Separator O2 Valve Barometer Purge Valve NO COM 10% Flow Hydrophilic Filter Tubing Coil (for flow measurement) NC Green 0.005 Hydrophobic Filters (2) Aqua 0.007 Sample Exhaust Flow Sensors Pump Bacterial Filter Figure 2-3. Normal flow without O2 The level of CO2 in the sample gas triggers the operation of the O2 valve. When the RGM is sampling inhaled gas The signal processor energizes the O2 valve to bypass the O2 sensor when the gas analyzer detects CO2 in the system. The presence of CO2 identifies exhaled gas. The previously inhaled sample gas is then trapped in the O2 detector while the detector slowly responds to the trapped sample. The valve returns to normal flow when the analyzer detects the lack of CO2 in the system once again. When the RGM is sampling exhaled gas The previous process is reversed. The O2 sensor is bypassed when the lack of CO2 is detected, trapping the exhaled gas in the O2 sensor. This process allows the comparatively slow O2 sensor to alternately sample inhaled sample gas for several breaths then exhaled sample gas for several breaths. 2-7 5250 RGM Service Manual 2.4.5 Purge flow Zero Filter Yellow 0.010 CO2 Scrubber Zero Valve O2 Gas Analyzer NC Patient Sample Inlet NO COM COM Nafion NO NC 0.04 in.ID 4 in. (CO2) or 20 in. (Agent) Water Separator O2 Valve Barometer Purge Valve NO COM 100% Flow Hydrophilic Filter Tubing Coil (for flow measurement) NC Green 0.005 Hydrophobic Filters (2) Aqua 0.007 Sample Exhaust Flow Sensors Pump Bacterial Filter Figure 2-4. Purge flow to clear blocked line If the flow sensor detects a drop of 40 ml/min from the baseline flow or more than 90 torr below atmospheric pressure, the RGM enters a purge-flow mode. The zero valve and purge valve will energize (NC position). This opens up the gas analyzer to the atmosphere and directs 100 percent of the sample gas through the water bottle. The system tests the flow rate every 5 seconds. If the flow rate doesn’t return to normal within 25 seconds, the SAMPLE LINE/FILTER BLOCKED or NO FLOW message appears. No flow is defined as less than 120 mL/min in the manual/auto zero state. 2-8 2/Theory of Operations If the system bounces back and forth between purge flow and normal, the SAMPLE FILTER BLOCKED message appears. This means the membrane filter cartridge in the water separator is probably the cause of the line blockage and should be replaced. WARNING: Service. A damaged membrane separator may result in a contaminated agent photometer. Once contaminated, the agent photometer assembly cannot be serviced and must be replaced. 2.4.6 Manual/auto zero flow Zero Filter Yellow 0.010 CO2 Scrubber Zero Valve O2 Gas Analyzer NC Patient Sample Inlet 90% Flow NO COM COM Nafion NO NC 0.04 in.ID 4 in. (CO2) or 20 in. (Agent) Water Separator O2 Valve Barometer Purge Valve NO COM 10% Flow Hydrophilic Filter Tubing Coil (for flow measurement) NC Green 0.005 Hydrophobic Filters (2) Aqua 0.007 Sample Exhaust Flow Sensors Pump Bacterial Filter Figure 2-5. Manual/auto zero flow The manual/auto zero flow mode is used to zero calibrate the gas analyzer. You can enter this mode manually through the service screens. The system automatically enters this mode periodically to compensate for the inherent drift of the analyzer. In this mode the zero valve is energized, and purge and O2 valves are de-energized. 2-9 5250 RGM Service Manual 2.4.7 Analyzing the dry gas The dry gas passes through the zero solenoid valve to a series of filters, valves, and detectors: IR analyzer Measures the percentage of N2O, CO2 and agent in the gas sample. An IR photometer and measurement board measure the amount of energy absorbed by the sample gas. The result is transmitted to the signal processor board and then to the display. O2 sensor Measures the percentage of O2 in the gas sample. This sensor is measured by the pneumatics interface board. Tubing coil A length of coiled tubing. Flow is determined by the flow sensor, which measures the gas pressure drop across the coil. Flow sensor Measures the flow rate. Located on the pneumatics interface board. Barometer Measures the barometric pressure in the sample system. The pressure transducer is located on the measurement board. CO2 scrubber Removes CO2 from atmosphere before zero calibration of the analyzer. Nafion® tubing Evaporates condensed moisture and humidity of sample gas. Zero filter Removes unwanted particles from the zero gas sample. Zero valve Normally open (not energized) to direct sample gas to the analyzer. When energized, the valve switches position to allow atmospheric gas to enter the pneumatic system. O2 valve Normally open (not energized). Bypasses the internal O2 sensor when energized. Purge valve Normally open. Energizes during purge flow. Yellow restrictor Simulates the resistance of the patient tubing for the zero inlet. Green restrictor Produces the 90% and 10% gas flow proportioning at normal gas flow. Aqua restrictor Sets the overall flow rate for the system. Hydrophilic filter Removes particulate matter, but allows water to pass. Hydrophobic filter Protects the pump. Allows no water to pass. Pump Creates suction for the pneumatics system. Bacterial filter Filters exhaust gas. 2.4.8 Exhausting the gas After passing through the pump, the gas exhausts through the bacterial filter to the sample exhaust outlet on the rear panel of the RGM. 2-10 2/Theory of Operations 2.5 Pneumatics interface board 2.5.1 General This board conditions signals for the O2, pressure and flow sensors, plus the temperature and water-trap-full sensors. It also contains the drivers for the pneumatic valves and the sample pump. 2.5.2 Pressure and flow The board has two differential pressure transducers (XD1 and XD2) with two opamp differential amplifiers (U6A and U6D; U6B and U6C). One transducer has a port removed, yielding a gauge transducer system, which is used for monitoring PAW. The other is used differentially, to monitor pressure drop across a flow restrictor to allow the system to read sample flow. Both transducers are nominally rated for 1 psi, and have a nominal transfer ratio of 15 mV/PSI with a reference of 10 V from the signal processor. The amplifiers have a gain of 333, yielding a final transfer ratio of 5 V/psi. The Service screen indicates the voltages and values measured by these sensors. CAUTION: Pressure in excess of 10 psi above atmospheric could damage the PAW transducer or sample system barometric pressure transducer in the RGM. 2.5.3 Water trap full sensor The water level is sensed by a plastic rod with a conical end that reflects a light beam if the cone is in air. A narrow beam IR LED and phototransistor are PCboard mounted in the separator block above the rod, along with the pull-up resistor for the phototransistor. The sensing circuitry consists of: • An adjustable, 10-kHz square-wave oscillator (U4A, R9, R19, and C9). R19 is adjusted to give maximum AC voltage output at U1B with no fluid in the bottle. • An LED driver (via U4B-E and R8) that provides sufficient current for the IR LED. • A unity gain buffer (U1A) to keep the phototransistor from being loaded down. • An inductor (L1) that prevents DC offset changes due to changes in ambient light. • A 10-kHz band-pass filter (U1B, C1, C2, R1, and R2). • A charge-pump type detector (C5, CR1, CR2, C10, and R10). • A Schmitt trigger (U4F) to insure a valid logic level output (trap full) when the fluid collection bottle is full. A logic high corresponds to WATER TRAP FULL status. The Service screen indicates the status of the sensor. 2-11 5250 RGM Service Manual 2.5.4 O2 amplifiers The amplifier (U3A) for the circuit O2 sensor has a load resistor of 681 Ω and a noninverting gain of 122. One side of the thermistor in the cell holder is connected to the cell anode, which is the ground of the amplifier circuit. The other side of the thermistor connects to the 49.9 Ω resistor R14. This applies a temperature variable load to the cell in parallel with the 681 Ω fixed load (R15). The 10 V offset, applied through R34, shifts the signal to the full -10 V to +10 V range of the A to D converter. Thus, 10-V output is 0 percent O2, also defined as no cell installed. Note: The internal O2 sensor amplifier has a similar configuration, using U3B. The voltages and values measured on the internal and circuit O2 sensors are identified in the Service screen. 2.5.5 Temperature U2 is a temperature-controlled current source, with a transfer function of 1 uA/K. The current from the 10-V reference through R6 offsets for room temperature (about 300 K), and amplifier U5B yields a voltage output of 200 mV/degree, centered around approximately 28 °C. The 10-V reference of the temperature sensor comes from the ADC of the signal processor board. The voltages and values measured by the temperature sensor are indicated in the Service screen. This circuit is used to generate the DEVICE OVERHEATED alarm if the internal temperature of the RGM exceeds 50 °C. 2.5.6 Solenoid drivers FETs Q2-Q5 switch the solenoid valves in response to logic level signals from the signal processor board. Zener diodes CR5, 7, 9, and 11 provide voltage drop from the 15 V supply, to the 12-V solenoids. Diodes CR4, 6, 8, and 10 absorb any back EMF when the transistor is turned off. Note: This board includes one spare solenoid valve driver. 2.5.7 Test points The pneumatics interface board contains seven test points: Test point Nominal conditions Other conditions TP1 PAW 0 +0.1 @ 0 cm H20 6.5 +1 @ 100 cm H20 TP2 Analog ground Ground TP3 Flow 3.5 +0.2 @ 230 mL/min ≈ 0 @ 0 mL/min TP4 Trap full 0 (empty) ≈ 5 full TP5 Temperature 0 +1 @ 28 °C TP6 Circuit O2 Internal O2 -7.0 to -8.5 @ 21% O2 -2 to +8 @ 100% O2 -7.0 to -8.5 @ 21% O2 -2 to +8 @ 100% O2 TP7 2-12 2/Theory of Operations 2.5.8 Pump driver The gas sampling pump is operated by a variable voltage controlled by the signal processor. Transistor Q1 acts as a current buffer for unity gain amplifier U5A, which tracks the input control voltage. The transistor in the feedback loop of the op-amp eliminates the junction drop at the output. Diode CR3 insures a feedback path should the op-amp receive a negative input. 2.6 IR gas analyzer The IR gas analyzer consists of the IR photometer and the measurement board: IR photometer In the photometer, a beam of IR energy passes through the gas sample. An IR detector measures the amount of energy absorbed. Measurement board Signal processing electronics demodulate the detector's output to acquire a DC signal, which corresponds to the content of each gas in the sample. Gas concentrations and barometric pressure values are passed as serial, digital data to the signal processor. The measurement board plugs into the motherboard; and the photometer mounts to the chassis behind the front access cover. 2.6.1 IR photometer (ACX-200) The ACX-200 photometer with five-agent capability measures: sevoflurane, desflurane, enflurane, halothane, isoflurane; CO2, and N2O. ACX-200 measurement board The measuring electronics can be divided into a few functional blocks, which are described as follows (see Figure 2-6). The ACX measurement board controls gas measurements. It converts the photometer signal to digital data, calculates results, and transmits it to the main CPU board. The board contains, in addition to the 80C51FA processor, EPROM, RAM, and EEPROM, several analog and digital I/O functions. It passes these values as serial data to the signal processor board. 2-13 5250 RGM Service Manual 2 Oxygen Sensor Input Figure 2-6. ACX-200 measuring board block diagram IR photometer The IR photometer is a dual-path type. The photometer detects IR light absorption through two separate cells in the measuring chamber: a reference cell and a sample cell. The sample gas is measured by determining the ratio between the intensity of IR light that is transmitted through the reference cell and through the sample cell. The two cells are located next to each other. The sample cell is in series with the gas sample system; the reference cell is open only to the inside of the photometer. The photometer contains a gas absorber that ensures that the reference cell is essentially free of CO2, N2O, and the agent. A revolving disc, or wheel, located between the measuring chamber and the detector, controls the light passing through the photometer. Five optical filters are located around the outer edge of the wheel. They are spaced 72 degrees apart and are in line with the sample and reference cells. Each filter passes a specific wavelength of IR light that is associated with each gas to be measured (CO2, N2O, and agent). 2-14 2/Theory of Operations IR light first passes at the CO2 absorption wavelength through the reference cell, then through the sample cell, and is then blocked completely. This sequence is repeated for IR light at the N2O absorption wavelength and then the agent absorption wavelength. After passing through the filters, the IR light is focused by a mirror (part of the detector assembly) onto an IR detector. The detector measures the light levels for each gas. For each revolution of the wheel, this process produces 10 light levels specific to the gas content (two for each gas) and a baseline reference signal (one for each gas) when the IR light is completely blocked off (Dark). Gas Light level CO2 Reference (Ref. CO2) Sample (Meas. CO2) Dark (Dark) N2O Reference (Ref. N2O) Sample (Meas. N2O) Dark (Dark) Agent Reference (Ref. Agent) Sample (Meas. Agent) Dark (Dark) The reference and sample cells are located next to one another so that the leading edge of the filters passes over the reference cell first, and the trailing edge of the filters passes over the sample cell last. The leading portion of the resulting waveform is associated with the reference cell; the trailing portion is associated with the sample cell. An optical sensor in the photometer detects light from a reflective surface on the wheel once every revolution. The pulses from this sensor synchronize the electronics on the measurement board to the signal from the IR detector. A thermistor measures the temperature within the photometer to compensate for thermal effects. The IR detector, optical sensor, and stabilizing diode are mounted on a preamplifier board, which is an integral component of the photometer. This board amplifies the signals from the detector, the sensor, and the thermistor, and converts them to a low-impedance level before routing them to the measurement board for further processing. An EEPROM in the photometer is programmed with parameters specific to that photometer. The measurement board uses these parameters to adjust its circuits for that specific photometer. 2-15 5250 RGM Service Manual Under normal operation, the measurement board supplies a signal (lamp) to the motherboard, which controls power to the IR source. If the measurement board fails to receive the expected signals from the photometer, it assumes that a photometer failure may have caused the wheel to stop spinning. It removes the lamp signal to avoid overheating and damage to the wheel. Internal and external bus The processor D1 has access to the measurement board peripherals (memory, A/D converter, D/A converters, etc.) via an internal bus. For communication between the CPU board and the measurement board, there is an external bus in connector X1. The external bus is driven by D21 (data lines), D3 (address lines) and D18 (read and write lines). Memory Memory components include 64K x 8K-bit program memory EPROM (D4), 32K x 8K-bit low current CMOS RAM (D22) powered by a RAM Upgrade Kit or a jumper on the motherboard, and EEPROM (D2) for permanent calibration values and setup memory. Reference voltages Reference voltages are established by the A/D converter (D14) reference voltage output (REFOUT, pin 8). This +10 V voltage is buffered by A2D. -10 V reference voltage is obtained by inverting and buffering +10 V with amplifier A2C. Measuring electronics for CO2, N2O, and AA CO2, N2O, and anesthetic agent measurement is accomplished by measuring each of these gases from the reference and measuring chambers of the ACX-200 photometer. The gas signals are transmitted from the ACX-200 photometer assembly through connector X2 pin 2 of the ACX measuring board and applied to the reference input of a D/A converter (D8). D8 is controlled by the microprocessor and is used for automatic gain control. The output current from D8 is proportional to the incoming signal and the gain is established for each gas (CO2 reference and measuring, N2O reference and measuring, and AA reference and measuring) by software. The signals are converted to a voltage by A24D and amplified by A24A, then applied to capacitor C30, which removes the DC offset. The dark level is established on C30 when the synchronous switch A28A is closed. Each signal is sampled by the peak detection circuit, consisting of A24C, V26, A28B, R97, R201, and C15. When the peak voltage of a signal is sampled, the switch A28B is open, sending the signal through V26, which acts like a diode. The peak signal is then applied to the capacitor C15. C15 is brought down to ground potential between signal peaks when A28B is closed and the dark signal is transmitted to it. The voltage peak of each gas (both measure and reference) is applied to an instrumentation amplifier (A24B) then to the input of a multiplexer (D23). D23 separates the signal to each of its components (CO2 reference and measuring, N2O reference and measuring, and AA reference and measuring). For CO2, the offset voltage is subtracted from the reference signal at A13B. 2-16 2/Theory of Operations Each gas signal, the temperature compensation signal, and the pressure signal are transmitted to D13, which serves as a demultiplexer whose output is applied to an A/D converter (D14) through an instrumentation amplifier. A/D conversion A/D conversion is made with a 12-bit A/D converter (D14). Input signal is multiplexed with D10 and D13. After conversion is completed, signal ADCRDY rises to +5 V. D/A conversion D/A conversion is made with a 12-bit D/A converter (D11). D12 multiplexes the analog output to 8 sample and hold circuits. Two of these are used to drive offset voltages for N2O and CO2 measurement. The others are used for external analog outputs (CO2,OUT, N2OOUT, VOLC, O2OUT, etc.). Timing of CO 2, N2O, and AA signals A timing pulse is produced when light is reflected to a phototransistor from a reflectorized surface on the filter wheel. The pulse produced is shaped by A28 on the preamplifier board and transmitted to port 3 of the microprocessor on the ACX measuring board. The processor produces the necessary address information to cause the PAL (D15) to produce the control pulses for the synchronous switches A28A and A28B (dark and clear). Motor speed control The speed of the stepper motor in the ACX-200 photometer is controlled with MOTOR-signal from the processor. This signal is buffered by D6. Pressure measurement CAUTION: Do not attempt to repair or replace the photometer pressure transducer (B1). Each pressure transducer is specifically calibrated with selected resistors for use on a given board. The pressure transducer (B1) measures the sampling system pressure after the photometer. Voltage reference V1, resistors R17, R108, R89, and amplifier A31C supply the pressure measurement bridge with 4 mA current. The pressure signal is amplified with A31A and A31B. The output of A31A corresponds to pressures 400 to 900 mmHg and is within -9.5 V and +9.5 V range. Temperature measurement Temperature measurement exitation voltage for the photometer stabilizing diode is fed from +10 V through resistor R104. The stabilizing diode voltage is proportional to photometer temperature. This voltage is amplified with A31D. 2-17 5250 RGM Service Manual Figure 2-7. ACX-200 measuring board timing diagram 2-18 2/Theory of Operations 2.7 Signal processor board 2.7.1 General description The signal processor board is a general purpose computer for analog acquisition, signal processing, and communications. It features a 16-channel analog multiplexer feeding a 12-bit analog-to-digital converter (ADC), and 8 (digital-toanalog converters (DACs) with 8-bit resolution. The board includes four channels of serial I/O. 2.7.2 Microprocessor The microprocessor (U13) is a NEC V25+ integrated processor with parallel I/O (32 lines, shared with other functions), two serial I/O ports with independent baud rate generators, two timers, and an interrupt controller. The crystal frequency is 14.3181 MHz, which gives a 5-MHz system clock on CKOUT, P0 bit 7. The parallel port pins are used for system configuration, oximeter (MINX) interfacing, TVX clip interfacing, and analog input multiplexer control. 2.7.3 Memory Memory consists of static RAM package location (U8) and a ROM card (U6). U8 is connected to a lithium battery for non-volatile RAM. The ROM card is replaceable without removing the board from the RGM. Note: Never power up the unit with the ROM card removed. 2.7.4 Decoding Two programmable logic chips (PALs U14 and U16) perform the external I/O decoding, using simple combinatorial logic upon address and processor control lines to enable the I/O functions. Another PAL (U24) performs the same functions for memory accesses. 2.7.5 Watchdog The watchdog chip U18 requires pulses every 50 mS on its watchdog input (WDI) to inhibit an auto-reset function. These pulses are output on P2 bit 7 of the V25 CPU. A watchdog fail reset may be seen when bit #8 is lit (see Section 2.7.11, Status LEDs). This chip also has a power-fail comparator and reference, which monitors PWR MON on its PFI pin. This comparator outputs a signal to the CPU nonmaskable interrupt (P1 bit 0) if the voltage falls below 1.3 V. The monitored signal from the power supply will allow operation for at least 12 mS after this threshold has been passed. This circuit also contains a lithium battery (B1), which supplies power to the Vout pin if Vcc falls below the battery voltage. This powers the time of day clock (U23) and the nonvolatile memory (U8). When Vcc falls below 4.65 V (±0.1 V), two other signals, PDCTL and LO_V, disable U8 and U23 respectively. Two additional control lines, RESET and ~RESET, are used for U8 and U23. 2-19 5250 RGM Service Manual 2.7.6 Configuration input switches The V25 port labeled PT on the schematic is for input only and features an adjustable threshold. It is used to identify the system configuration with four switches for option status (refer to 3/Checkout and Calibration for more detail). Dip switch Position Indication 1 Closed/right (ON) Open/left (OFF) SpO2 board not installed. SpO2 board installed. 2 Closed/right (ON) Open/left (OFF) Normal position (Not used). 3 Closed/right (ON) Open/left (OFF) Calibration constants are locked. Calibration constants are unlocked. 4 Closed/right (ON) Normal position (this switch is not used). Not used Calibration constants Not used SpO2 board Off On Figure 2-8. SW1 signal processor configuration switch 2.7.7 Analog acquisition Fifteen external signals plus internal test signal (ATEST) are multiplexed by IC U20. The enable signal (MXEN) outputs the signal, selected by MX0-3, which is then stored by capacitor (C49) while the voltage level is converted by the ADC AD674 (U19). This capacitor is buffered by a fast, low-bias current op amp (U22) to prevent loading by U19. The ADC converts the voltage and sends the data to the processor as two bytes. The input range is ±10 V, for a scale factor of 4.88 mV/bit. The converter reference is buffered by op-amp U21 and output to the pneumatics interface board. The multiplexer channel is controlled by the V25 P0 bits 0 to 3, and enabled by P0 bit 4. The conversion starts by executing an I/O write to 1CH. Conversion complete is tested on P1 bit 7 where a 1 indicates conversion is in process. Data is then read at two successive addresses in I/O memory. RGM analog input assignments 2-20 Test point Code Function Test point Code Function TP1 5 PAW U20, pin 11 8 DAC loop back TP3 4 Flow 7 Ext. O2 TP4 6 Int. O2 10 Temperature TP16 1 MINX pleth 2/Theory of Operations 2.7.8 Analog output The board has two 8-bit DACs U10 and U15. Each DAC has four voltage outputs. The DACs share a 10-V reference: IC U9. Of the 8 voltage outputs, 7 are divided by 10, R9-R15 and R2-R8, providing the general purpose analog outputs with a 0-to-1 voltage range. Channel 6 is also routed to the analog input multiplexer U20 as a test signal (ATEST). Channel 7 is buffered by a unigain op amp and routed to the pneumatics interface board as a variable pump drive signal (PUMPB). Each output is unity buffered by op amps U4 and U5. Capacitors C3, 10, 12, 5, 6, 13, and 15 prevent DAC glitches from appearing at the outputs. 2.7.9 TVX interface The TVX connects to the microprocessor through Schmitt buffers U12A (TVXA) and U12B (TVXB). TVXB goes to interrupt line (INT1), where it can either directly interrupt the processor or generate a macro service call to store the time of the event in a buffer. The other bit for direction sensing may be read at P1 bit 5. The interrupt bit may be read at P1 bit 2. 2.7.10 SpO2 oximeter (MINX) interface The decode PAL U16 decodes an address and the read/not write (R/~W) signals from the processor, and generates the enable signal (MINX), which with R/~W control bidirectional buffer U17. It also generates ~STB (for data OUT to the oximeter) and ~ACK (for data IN from the oximeter), which coordinate the transfer of data. The processor reads ready status (IBF and ~OBF) at P1 bits 6 and 1 respectively. ~OBF may also generate and interrupt (INT0) to insure timely reception of data from the oximeter. Saturation data and the oximeter status/control information occur through the parallel interface provided by the bus transceiver U17. A patient pulse signal from the oximeter is received at P2 bit 4 as a short (about 16 mS) logic pulse, which software must sample regularly to sense. This signal triggers the pulse beep. The ADC reads the patient plethysmograph, which is then sent to the display. 2.7.11 Status LEDs A 10-LED bar chip (U1) indicates the software and system status. Also see Figure 44 for LED bar status during powerup. Bit # 0 1 2 3 4 5 6 7 8 9 2-21 5250 RGM Service Manual Bits Status 0-4 General software status Bits 0-3 are CO2 breath phase bar graph. Bit 4 is PAW breath phase. 5 O2 solenoid 6 Purge solenoid 7 Zero solenoid 8 Watchdog failure 9 Spare solenoid valve 2.7.12 Serial I/O In addition to the serial ports on the microprocessor, a dual-channel multimode IC (U11) provides independent baud rate generators, and operates in asynchronous or synchronous modes. This device drives the external RS-232 port. It is addressed in I/O space. The timing signal (CKOUT) for U11 comes from the V25 clock out labeled P0 bit 7. 2.7.13 Other inputs The following additional inputs are provided on port P2: ~ALM_SI bit 5 Alarm silence WTRFL bit 3 Indicates that the reservoir on the water separator is full. FREQ_MN bit 2 A square-wave signal that tracks the polarity of the AC power line into the RGM and measures the line frequency. 2.7.14 Test points The table below lists the signal processor’s test points and functions: 2-22 TP1 PAW TP11 TVX pulse (buffered) TP2 A/D ref out TP12 Digital ground TP3 Flow TP13 Reset capacitor TP4 Internal O2 TP14 Sample/hold (S/H) out TP5 Pump voltage TP15 Analog ground TP6 5 MHz TP16 MINX plethysmograph TP10 +5V TP20 Buffered reset 2/Theory of Operations 2.8 SpO 2 oximeter (MINX) board 2.8.1 Description The Datex-Ohmeda SpO2 (MINX module) oximeter is a single surface-mounted circuit board that provides a method of monitoring noninvasive pulse oximetry. As used in the 5250 RGM, the oximeter board is programmed to use parallel command strings. Data can be output in serial, parallel, or analog formats. Status messages are provided in serial, parallel, or logic-level formats. Logic-level control inputs configure the board in stand-alone systems. The oximeter board uses the Intel 80C196KA 16-bit microprocessor. This microprocessor has a high level of integration and an internal 10-bit ADC. The microprocessor and EPROM are socketed for easy replacement and code enhancement. These components are the only user-serviceable parts. CAUTION: When replacing either the microprocessor or the EPROM on the SpO2 oximeter (MINX) board, be sure to use a chip extractor to avoid damaging the components. 2.8.2 Power supply The oximeter board receives three separate voltage supplies: Name Use Voltage Current Ripple Pin +5 Logic +5V ±5% 30 mA Type 36 mA Max 100 mV 31A,C +V Analog +5V 53 mA Type 64 mA Max 20 mV 29C -V Analog -5V 115 mA Type 138 mA Max 50 mV 30C The +V and -V analog supplies share a common return path called AGND. The logic return is called DGND. AGND and DGND are tied together at the microprocessor on the module. Note: The logic supply must be within ± 5 percent of +5 V. The + analog must be within + 100 mV of the +5 logic supply. The absolute magnitude of the -V supply must be within ± 5 percent of +V. 2.8.3 Control inputs The oximeter board has 14 control input lines that connect to the 82C55A I/O port lines (configured by software as inputs). If all control inputs are selected low (worst case), an additional 6.5 mA current is required from the +5V power supply. The oximeter board communicates through both the serial and parallel ports. Commands may be sent to configure performance parameters, override defaults, and initiate diagnostic or calibration routines. 2-23 5250 RGM Service Manual 2.9 Display controller board 2.9.1 General description The display controller board is a microprocessor-based interface for controlling the electroluminescent (EL) display output, the audible output, and the operator input from the IR touch frame. There are two audio alarm signal sources, one programmable device for all normal system output (patient warnings and alarms) and a second for total system failure. 2.9.2 Processor and interfacing The processor used in the display controller is the V25 (U21), which is the same as the processor used in the signal processor board. This IC has several built-in functions that minimize the amount of support circuitry needed for the system. 2.9.3 System memory The system RAM (U22) used is a CMOS 32 k static part. The ROM is on a plug-in ROM card (J5). A 10-MHz oscillator (Y2) provides the timebase for all the V25 functions. 2.9.4 Communicating to the signal processor For communicating to the signal processor board, the controller uses an RS-422 serial channel. U18 translates between the V25 TTL serial string and the differential RS-422 signals. R48-50 bias and terminate the incoming signal. 2.9.5 Watchdog circuit A ADM 8699 chip (U17) provides the supervisory functions such as reset and watchdog timer. R24 and R40 allow overriding the watchdog function by permitting U17 pin 11 to be tied to 1.8 V, while insuring that pin 11 does not float. The processor must change the state of pin 11 at least every 50 mS to prevent the watchdog from being triggered. If the watchdog is triggered, it will: 2-24 • Attempt to restart the processor by pulsing its reset pin. • Enable a system fail audio alarm (generated by U14). • Operate the yellow alarm LED in the touch frame via Q2. • Shut down the display matrix. 2/Theory of Operations 2.9.6 Programmable logic A key part of the circuit is the programmable logic device, U33, which carries out three major functions: • Performs all the memory decoding for mapping the V25 memory • Generates all the timing and control signals for the video RAM • Generates various clock frequencies used in the video control circuitry and for the alarm sound IC U33 has its own 16-MHz oscillator, U1, which provides the clock for the output frequencies and for the state machine, which does the video timing functions. 2.9.7 Power regulation Power is brought onto the board loosely regulated to 6.5 V and 15 V, to allow minimum heat generation. The 6.5 is regulated to 5 V for the logic circuits, and the 15 V operates the display and the audio power amplifier. Two low dropout regulators provide the +5-V supply with minimum power dissipation. One regulator supplies the sound IC to isolate it from the digital noise of the rest of the system; the other supplies the rest of the circuits. 2.9.8 Audio A single-chip sound synthesizing IC (U2) generates complex audio alarm tones and has software-controllable volume levels. For the system-failure alarm, an additional audio tone circuit operates independently of the V25. A 4-MHz clock from U33 provides the time base for the internal operations of sound synthesis. A two-stage RC low-pass filter smooths the switching transients from the synthesizer and mixes the two outputs (one for musical sounds and one for percussion sounds) with the system fail-sound generator, U14. U16 and its related resistors and capacitors drive the speaker that is mounted on the display case. 2.9.9 Touch screen interface For the user interface, the controller board contains a circuit that interprets the inputs from the IR touch frame built into the bezel around the EL display. This circuit uses a 80C32 microprocessor (U3) and associated components to drive LEDs along the bottom and right edge of the touch frame, read analog values from the phototransistors along the top and left edge of the frame, and sense if and where a light beam is broken. The V25 communicates to the 80C32 via TTL level serial interface. The red and yellow alarm LEDs are also mounted on the touch frame, but are controlled by the V25. 2-25 5250 RGM Service Manual 2.9.10 Video Graphic CPU The 72123 advanced graphic display controller (AGC II), U30, controls the display memory, and generates display synchronizing signals. It also provides high-speed drawing operations for the host CPU (V25) such as line, rectangle, and circle drawing, painting, copying, and rotating. The V25 accesses the AGDC by two separate chip selects: • CSIR for accessing the internal registers • CSDM for accessing the display memory The V25 can affect the display memory contents in three ways: • By directly addressing the display memory just as it does RAM or ROM. In this case, the AGDC acts as an interface between the 8-bit V25 bus and the 16-bit video memory bus. The hardware interface employs chip-select signal CSDM and the handshake signal READY. The AGDC then uses V25 address line A0 to generate either its lower byte enable output (DLBE) or its upper byte enable output (DUBE), depending on whether the address is odd or even. • Indirectly by using the AGDC to perform high-level functions. A feature of these functions is that the parameters given to the AGC II can be specified as either actual display memory addresses, or more conveniently, as X,Y coordinates relative to a previously set up 0,0 or origin point that is defined as an address and particular bit. • By using the V25s DMA mode in conjunction with the AGDCs DMA port (one of the internal registers). Data can be transferred in either direction between system memory and the display memory using handshaking lines DMARQ from the V25, and DMAAK from the AGDC. Video memory and video output circuits The video memory is made up of devices U23-U24. The AGDC a uses programmable logic chip (U33) to control the RAMs and the output shifting through two types of display memory cycles: • Read cycles: normally used for manipulating memory data. • Write cycles: normally used for manipulating memory data. 2.9.11 Test points The following are the display processor test points: 2-26 • TP1 Ground • TP2 Logic level scope-sync for touch panel scan 2/Theory of Operations 2.10 Display matrix assembly The EL matrix displays high resolution graphics and text in a self-contained enclosure. The display consists of the following: EL panel A thin-film display mounted with an elastic spacer to an electronic board containing the high voltage circuits required to drive the EL panel. The EL panel has 512 transparent column electro-des crossing 256 row electrodes in an X-Y fashion. Light is emitted when AC voltage is applied at a row-column intersection. Driver board Flexible conductors connect the EL panel to the driver board. A 20-lead cable connects the driver board to a controller board attached to the frame. The controller consists of timing logic, drive voltage waveform generators, and a voltage converter. The voltage converter converts input DC voltages required for the matrix display. The input signal and input power connection for the matrix display are made through a 16-pin flat cable connector on the controller. Controller Drive logic signals are processed and drive voltage pulses are generated and controlled on the controller board. The power supply input voltages are +5 V and +15 V dc. WARNING: Electrical shock hazard. The display generates potentially dangerous voltages (high voltage pulses up to 225 V AC). Do not touch the driver-controller circuit board during operation. 2.10.1 Input signal descriptions The four input signals required for the display are at 5V TTL logic levels. Signal Symbol Description Video data VID Signal that supplies the pixel information to the system. A logic 1 at the VID input corresponds to a lit pixel on the display. Video clock VCLK The VCLK signal shifts data present on the VID line into the display system. VCLK is active on the rising edge. Horizontal sync HS The horizontal sync signal HS selects the 512 bits of data from the VID lines. Vertical sync VS Data of the first 256 lines after VS rising edge will be shown on the display. Logic supply voltage Vcc1 Supply voltage for interface logic. Display supply voltage Vcc2 Supply voltage converted to required internal supply voltages. VID data loaded during the last 512 VCLK pulses preceding the falling edge of the HS signal are displayed on the selected line. The lines to be displayed are selected sequentially by the first 256 HS pulses following the rising edge on the VS pulse. 2-27 5250 RGM Service Manual 2.10.2 Input connector configuration (J1) Pin Signal Pin Signal 1 Vcc2 9 VS 2 Vcc2 10 GND 3 Vcc1 11 HS 4 Vcc1 12 GND 5 GND 13 VCLK 6 GND 14 GND 7 GND 15 VID 8 GND 16 GND 2.10.3 Optical characteristics 2-28 Display area (H x W) 97.7 x 195.2 mm (3.85 x 7.69 in.) Matrix configuration 512 x 256 pixels Pixel pitch 0.38 mm (15 mil) Pixel size 0.28 mm (11 mil) square Pixel luminance 125 cd/m2 (37 fL) typical 80 cd/m2 (23 fL) min Luminance variation (within the display) <35% Color broadband yellow, peak at 580 nm Viewing angle >140 deg 3/Checkout and Calibration 3.0 General This chapter provides • Checkout procedures to determine whether the RGM is functioning properly. • Calibration procedures. WARNING: Failure of operation. Perform the checkout procedures before using the monitor with a patient. If the monitor fails any test, it must be removed from use until it has been repaired and checked for correct operation. Before returning the RGM to clinical use, perform the leakage current and ground resistance checks (see section 3.13). If the RGM fails to meet any of the checks, refer to 4/Messages and Troubleshooting. Note: Some checkout procedures require you to partially disassemble the machine. Refer to 5/Repair Procedures for disassembly instructions. 3.0.1 Special tools required for checkout and calibration procedures WARNING: Data validity. Use only Datex-Ohmeda calibration gas of 1.75% Halocarbon-23 (CHF3)/freon or equivalent, 6% CO2, 40% N2O, 50% O2 and 2.25% argon. Use of the wrong calibration gas kit will result in improper calibration and the RGM will display incorrect gas concentrations. The following tools are needed to perform checkout procedures: • Calibration gas kit. • Calibrated air flowmeter with restrictor, 0 to 600 cc/min (6024-0000-006) Accuracy is ±2.0% of the reading or ±division, whichever is greater. • Pressure gauge 0 to 150 cm H2O (0-100 mmHg) (0380-1500-112). • DB9 male connector (pins 2 and 3 shorted) (6050-0001-709). • Needle-nose pliers or hemostats. • Source of dry clinical oxygen at 100% concentration. • DIN Connector, 8-pin (6006-0000-184). • Oscilloscope or chart recorder (optional). • Calibration bag only (6050-0000-226). • Flow restrictor only (6027-0000-126). 3-1 5250 RGM Service Manual • Digital multimeter with a minimum of 3 digits, 0.1% DC accuracy. • Absolute barometric pressure meter 600 to 800 mmHg ±0.25 mmHg, or gauge pressure meter 0 to 100 mmHg range. • Leakage current tester. • Extender board 96-pin DIN (32 x 3) Eurocard connector (0237-2135-300). 3.1 Functional checkout Follow this procedure before using a new monitor, or one that has been returned from repair or service, for patient monitoring. WARNING: Failure of operation. Perform the checkout procedures before using the monitor with a patient. If the monitor fails any test, it must be removed from use until it has been repaired and checked for correct operation. This procedure provides an initial checkout procedure for monitors after they are unpacked and before first use. 1. Inspect all of the accessories of the RGM for damage. Replace any broken or damaged accessories. Datex-Ohmeda recommends that only genuine replacement parts manufactured or sold by Datex-Ohmeda be used for all repairs. 2. Inspect the exterior of the RGM for damage. Check all the connectors and controls and replace any broken or damaged parts. 3. Press the Display Release latch to slide the display panel to the right. Empty and replace the water trap bottle before proceeding with the checkout procedure. 4. Turn on the RGM and allow it to warm up for 5 minutes. 5. Verify that the fan turns freely and blows air out the rear of the unit when power is turned on. 6. During poweron the RGM performs several internal checks. The alarm tone should sound and the red and yellow LEDs should flash. If a problem exists, an alarm message appears. Refer to 4/Messages and Troubleshooting for a description of the alarm messages. 7. Check the leakage current and ground resistance (see section 3.13). 8. Block sample input and verify that purge occurs. If not, see section 3.5.3. 9. Check the barometric pressure (see section 3.6.1). 10. Zero and span calibrate the monitor, (see section 3.6.5). 11. Check the patient circuit O2 sensor (see section 3.6.6). 12. If installed, check the SpO2 module (see section 3.7). 13. When the airway pressure line is connected to the PAW inlet and the ventilator is on, pinch the airway pressure line to block the flow of air. Verify that the PAW reading remains stable for 5 seconds. 3-2 3/Checkout and Calibration 3.2 Dip switch and jumper positions 3.2.1 On the signal processor board WARNING: Electrical shock hazard. Before cleaning or repairing the monitor, power it off and disconnect it from AC mains power. See section 5.3 for cover disassembly instructions. Remove the top cover of the RGM. Facing the front of the RGM, find the signal processor board on the far left side. A DIP switch (SW1), which is accessible from the top of the board, identifies the optional boards installed in the RGM. The switches are numbered: 1 is the closest to the front of the monitor; 4 is closest to the rear panel. Check that the switches are positioned as required for the features present on the RGM. DIP switch Position Indication 1 Closed/right (ON) Oximeter board not installed. Open/left (OFF) Oximeter board installed. 2 Closed/right (ON) Normal position. 3 Open/left (OFF) Calibration constants are locked. (Normal position.) Closed /right (ON) Calibration constants are unlocked. Closed/right (ON) Normal position (this switch is not used). 4 Not used Calibration constants Not used SpO2 board Off On Figure 3-1. SW1 signal processor configuration switch Note: Except in the case of locking calibration constants, DIP switches and jumpers are read only during poweron. Recycle the power after changing a switch setting. 3-3 5250 RGM Service Manual 3.2.2 SW1 on the motherboard Jumper Position Indication 1 On the right. ACX-200 IR analyzer installed. 2 Unused install jumper on left or open. 3 Unused install jumper on left or open. Note: Except in the case of locking calibration constants, DIP switches and jumpers are read only during poweron. Recycle the power after changing a switch setting. A jumper pad (SW1) provides sense inputs for configuration information to the signal processor. Location of the jumper depends on whether the RGM is an agent or non-agent monitor, and may need to be set after replacing the motherboard. Spare Spare ACX 200 Figure 3-2. SW1 motherboard 3.3 Power-up self-test WARNING: Electrical shock hazard. Do not touch any exposed wiring or conductive surfaces while the cover is removed. The voltages present when electrical power is connected to the RGM can cause serious injury or death. Never wear a grounding wrist strap when working on an energized monitor. During poweron, the RGM performs several internal checks. The alarm tone sounds and the red and yellow LEDs on the front panel flash. If a problem exists, an alarm message appears. Refer to 4/Messages and Troubleshooting for a description of the alarm messages. Turn on the RGM and observe the functions as outlined in the following table. The times listed are approximations. Warmup could take up to 3 minutes. 3-4 3/Checkout and Calibration 3.3.1 Power-up functions Time Display processor function Signal processor function At 0 to 1 second Turn on red and yellow LEDs. Start RAM test. Turn on LEDs 0 and 7. Start RAM test. At 1 to 2 seconds Turn off LED 0. Run initial sequence. Delay 1.8 seconds. A 6- to 10-second delay occurs to wait for barometric pressure data from photometer. At 3 seconds Read local barometric pressure. Turn on pump. Wait 3 seconds. At 4 seconds Turn off red and yellow LEDs. Run initial sequence. At 5 seconds Display all pixels on screen. At 6 seconds Display power-up self test. Test analog system, vacuum, and sample Request configuration from the flow rate. Start 3-second purge. signal processor. At 9 seconds At 12 seconds Complete initialization. Display monitor main screen. Send configuration to display processor. At 18 seconds Start auto zero of internal O2 cell: 22 seconds. At 50 to 180 seconds Photometer data available. At 300 seconds Start auto zero of photometer, (2 to 9 seconds). Repeats as necessary. 3.4 Power supply checks WARNING: Electrical shock hazard. Do not touch any exposed wiring or conductive surfaces while the cover is removed. The voltages present when electrical power is connected to the RGM can cause serious injury or death. Never wear a grounding wrist strap when working on an energized monitor. 1. Turn off the power and disconnect the power cord; see section 5.3 for cover disassembly instructions. 2. Remove the cover; see section 5.3. 3. Reconnect the power cord and turn on the monitor. 4. Verify that the LEDs on the power supply board are lit. Channel LED Filter Regulator SCR +5 CR18 C13 VR4 CR13 -5 CR19 C16 VR5 CR22 +15 CR16 C18 VR3 CR2 -15 CR17 C15 VR1 CR3 3-5 5250 RGM Service Manual 5. 6. Using a DVM, with fine point test leads, check for the following voltages from the power supply board with reference ground J1 pin 4. Connector Voltage Description J1 Pin 8 +5 ±0.25 V DC Digital supply J1 Pin 9 +15 ±0.75 V DC Positive analog supply J1 Pin 10 -15 ±0.75 V DC Negative analog supply J1 Pin 12 -5 ±0.25 V DC Analog (MINX) supply J1 Pin 13 +6.6 ±0.4 V DC Prereg display digital, loaded J1 Pin 17 +15 V DC ± 0.75 Prereg display matrix, loaded J2 Pin 2 +12 ±0.75 V DC Fan supply Check for the following voltage from the power supply with reference ground J1 pin 2. Connector Voltage Description J1 Pin 1 +5 ±0.25 V DC RS-232 isolated 3.5 Pneumatic system checks 3.5.1 Sample flow 1. Select MENU from the display screen. 2. Select SETUP from the menu. 3. Select Service mode from the Setup Screen and select “on.” The Service screen appears. Note: Monitors with SpO2 have a two-page Setup screen. The Service mode option is on page 2. NEXT PAGE Setup Screen on Circuit O2 Rev. flow det Default Display RS-232 device Display labels Service mode Language Trend Clear N2O display Wave Type Patient type 90 VIEW ALL Figure 3-3. Select Service mode 3-6 CLOCK SET EXIT 3/Checkout and Calibration Agent ACX-200 Version Display Version SpO2 Version Signal Processor Version Agent Barometric Pressure Breath volume flow Breath volume data Cal constants CO2 Carbon Dioxide Internal Temperature SpO2 Pleth SpO2 Pulse Rate Saturation O2 N2O Nitrous Oxide O2 Circuit Oxygen O2 Internal Oxygen PAW Pressure Airway Pump Voltage Sample flow rate Oxygen valve status Purge valve status Water trap full status Zero valve status ENTER CAL 1.20 6.00 S 8.007 0.00 627.5 --0 0 0.10 24.5 ------0.0 0.0 20.6 0.1 6.00 201 0 0 0 0 Service 0402 0000 ACX status hex ACX error status hex CPU type V25+ (%) 0.00 (torr) 627.5 (ml/breath) 0.0 (liters/minute) Uncorrected 0 (min 25.6us cnts/flip) 1=locked 92 Run Time Hours (%) 0.10 (degrees C) 0.28 (volts) 0.45 (volts) (Beats/min) (%) (%) -0.8 (%) -10.00 (volts) (%) -8.00 (volts) (cmH2O 0.01 (volts) (volts) 153 (dac counts) (ml/min) 3.15 (volts) 0=oxygen cell active 1=in purge state 1=full 1=in zero state PNEUMA TESTS CAL EXIT Figure 3-4. Service screen 4. Select PNEUMA from the Service screen. The Pneumatics Service Screen appears. Pneumatics Service Screen Barometric Pressure CO2 Carbon Dioxide Last zero error O2 Internal Oxygen Oxygen valve status Pump Voltage Purge valve status Sample flow rate Sample pump status Water trap full status Zero valve status 597.3 0.00 none 20.6 0 6.00 0 201 1 0 0 Zero Input Sample Inlet CO2 N2O Agent (%) 0=oxygen cell active (volts) 1=in purge state (ml/min) 0=off 1=full 1=in zero state O2 O2 597.3 0.00 0 -8.02 153 3.13 Converse value (volts) (dac counts) (volts) Purge Sample Zero Sample Exhaust Valve ON (torr) (%) OFF Sensors UP Valve DOWN Sensor ZERO Valve PURGE Pump O2 EXIT Figure 3-5. Pneumatics Service Screen 5. Read sample flow rate on the screen. 6. Connect the 96-in. (243.84-cm) patient sample tube to the sample inlet. 7. Connect a calibrated flowmeter to the sample exhaust. 8. Measure the flow at the sample exhaust for 30 seconds. The flow must be greater than 150 ml/min and less than 230 ml/min. 9. Check that the flow on the calibrated flowmeter matches the flow recorded on the Pneumatics Service Screen ±20 ml/min. If the flows do not match, see section 3.9 for sample flow calibration. 3-7 5250 RGM Service Manual 3.5.2 Pump leakage 1. Connect a 0–600 cc/min flowmeter to the sample exhaust. 2. While in the Pneumatics Service Screen, pinch the inlet tubing to the pump (tube closest to pneumatics board) with an electrically insulated blunt-nose pliers or tube-clamping device. 3. Measure the sample flow rate on the flowmeter. 4. Greater than 10 ml/min flow on the flowmeter indicates an excessive pump leak. Replace the pump. 5. Remove any tubing clamps and flowmeter previously installed. 3.5.3 System leakage 1. Block the sample inlet on the front of the RGM. 2. Measure the sample flow rate on the flowmeter. 3. RGMs with greater than 10 ml/min flow over the pump leakage recorded above have excessive leakage in the pneumatic circuit and must be repaired. Refer to Figure 4-4, Pnematics Service Screen. 3.5.4 Pump vacuum 1. While on the Pneumatics Service Screen, switch off the pump and record the barometric pressure. 2. Switch on the pump. 3. Block the sample inlet. Wait 30 seconds and record the barometric pressure. 4. The change must be at least 180 Torr. The pressure should not fluctuate by more than 15 Torr. 3.5.5 Zero valve test 1. Use the sample flow rate on the Pneumatics Service Screen. 2. Open the zero valve to zero input position. 3. Record the sample flow rate. 4. Block the sample inlet and verify that the sample flow does not change by more than 20 ml/min. If this test fails, refer to 4/Messages and Troubleshooting for bad zero valve, or resistive scrubber. Note: If you have just replaced the scrubber cartridge, be sure to remove the plug in the top of the scrubber cartridge (inserted for shipping purposes only) otherwise, it will obstruct flow through the auto zero path. 3-8 5. Block the sample inlet and pinch the line between the auto zero filter and zero valve. RGMs with greater than 10 ml/min flow over the pump leakage recorded in the previous pump leakage section have excessive leakage and must be repaired. Refer to section 4.3.4, Pneumatics problems, to determine the source of the leak. 6. Remove any tubing clamps previously installed. 3/Checkout and Calibration 3.5.6 Purge valve test 1. Use the sample flow rate on the Pneumatics Service Screen. 2. Set the purge valve to the purge position and the zero valve to the zero input position. 3. Block the sample inlet. RGMs with flows greater than 10 ml/min over pump leakage recorded in the previous pump leakage section must be repaired. Refer to section 4.3.4, Pneumatics problems, to determine the source of the leak. 3.5.7 O2 valve test 1. Use the sample flow rate on the Pneumatics Service Screen. 2. Set the zero valve to the zero input position, the purge valve to the purge position, and the O2 valve to the sensor-active position (status=0). 3. Pinch the purge line, which is the bottom-most line (silicone tubing) out of the back of the water separator. At the same time, pinch the single line from the O2 valve to the O2 sensor housing. RGMs with flows greater than 10 ml/min over pump leakage mentioned earlier in this section must be repaired. Refer to section 4.3.4, Pneumatics problems, to determine the source of the leak. 4. Bypass the O2 sensor by placing the O2 valve in the open (status=1) position. 5. Pinch the purge line, as before, and at the same time pinch the line between the O2 valve and the plastic tee. RGMs with flows greater than 10 ml/min over pump leakage recorded in step 2 above must be repaired. Refer to section 4.3.4, Pneumatics problems, to determine the source of the leak. 3.5.8 Water trap sensor test Note: The pump must be off during the water trap sensor test. WARNING: Biohazard. When handling materials that may have come into contact with patient exhalant or fluids, follow approved procedures for contamination control. 1. While on the Pneumatics Service Screen, turn off the pump by selecting OFF. 2. To access the water trap bottle, press the Display Release and slide the display panel to the right. 3. Carefully remove the water trap bottle. Carefully dispose of any contents in a receptacle designated for patient waste. 4. Fill the trap bottle with water and submerge the tip (“V” portion) of the trapfull plastic rod. Be careful not to let the water touch the tubing next to the sensor rod. 5. Verify that the water trap sensor displays a Water trap full status of “1.” 6. Remove, empty, and replace the water trap bottle and verify that the sensor displays a water trap full status of “0.” 3-9 5250 RGM Service Manual 3.5.9 Purge and sample line blocked test WARNING: Biohazard. When handling materials that may have come into contact with patient exhalant or fluids, follow approved procedures for contamination control. 1. Press EXIT twice to select the display screen. 2. Connect the standard 96-in. (243.84-cm) patient sample tube to the sample inlet. 3. Briefly block the flow into the RGM at the end of the sample tubing until the PURGING message appears on the screen. • a. A purge must occur within 2 seconds, and the RGM must not display the message SAMPLE LINE BLOCKED. A purge can be detected by an audible click in the purge valve. • b. An advisory two-tone alarm sounds at this time. 4. Block the flow into the RGM at the end of the sample tubing. Within about 25 seconds the RGM must display the message SAMPLE LINE/FILTER BLOCKED. A warning alarm sounds. Unblock the line and the alarm condition should clear. 5. While wearing waterproof gloves, remove the filter cartridge and block the top, center “silver” port of the membrane separator assembly. Within 25 seconds the RGM must display the message SAMPLE LINE/FILTER BLOCKED, and the advisory alarm sounds. 6. Unblock the line and the alarm condition should clear. Note: Purging is disabled in certain service screens to allow for testing and calibration. 3.6 Calibration checks 3.6.1 Barometric pressure check 1. Obtain and record local barometric pressure (not corrected to sea level). Note: Be sure that the true local barometric pressure used for comparison is not corrected to sea level. Absolute barometric pressure is strongly affected by altitude. A typical absolute barometric pressure at sea level is 760 mmHg, but at 5,000 ft (1,600 m) a typical absolute barometric pressure is 615 mmHg. 3-10 2. Select MENU from the Display Screen 3. Select SETUP from the menu. 4. Select Next Page. 5. Select Service mode from the Setup Screen. 6. Select PNEUMA from the Service screen. The Pneumatics Service Screen appears. 7. Select OFF from the Pneumatics Screen. This will turn off the vacuum pump. 8. If the displayed barometric pressure is more than 5 mmHg from the true local barometric pressure, perform the next procedure, Barometric pressure calibration. 3/Checkout and Calibration 3.6.2 Barometric pressure calibration Notes: • Be sure that the true local barometric pressure used for comparison is not corrected to sea level. Absolute barometric pressure is strongly affected by altitude. A typical absolute barometric pressure at sea level is 760 mmHg, but at 5,000 ft (1,600 m) a typical absolute barometric pressure is 639 mmHg. • To convert inches of Hg to mmHg, use this formula: Inches Hg x 25.4 = mmHg Sample Inlet Sample Exhaust 5250 Syringe 60 cc or larger Pressure Gauge Figure 3-6. Calibration setup Pressure calibration Note: Calibration constants must be unlocked (set signal processor's dip switch 3, to On or closed position) before barometric calibration. Relock calibration constants at the end of this procedure. 1. Select MENU from the main screen. 2. Select SETUP from the menu screen. 3. Select Next Page. 4. Select Service mode and touch to display the Service screen. 5. Select CAL to display the Calibration screen. Calibration For each type of calibration, follow the directions on the menu screen. BARO CAL PAW CAL SAM FLOW EXIT Figure 3-7. Calibration screen 3-11 5250 RGM Service Manual 6. Select BARO CAL from the Calibration screen. The Barometric Pressure Calibration screen is then displayed. Barometric Pressure Calibration Barometric Pressure Sample pump status Measurement range -93.3 To 624.7 (torr) 0 0=off 2924.7 (torr) Current set barometric pressure 624.7 650.0 (torr) Using the CLEAR and number icons, enter the uncorrected to sea level barometric pressure. Then press the LOW icon. Apply 100 torr above ambient pressure and press the HIGH icon. 0 1 CLEAR 2 3 4 5 LOW 6 7 8 HIGH 9 EXIT Figure 3-8. Barometric Pressure Calibration screen Note: Units of torr and mmHg for most practical purposes are identical. 7. Leave the inlet/exhaust ports open to ambient pressure. 8. Using the CLEAR and number icons, enter the local barometric pressure (not corrected to sea level) in mmHg. Store this value by touching the LOW key. Note: If the message INVALID LOW appears, go back to step 6. 9. Connect the sample exhaust to the sample inlet. Refer to Figure 3-6. 10. Connect a tee to the loop. 11. Connect a pressure gauge and open syringe to the tee. 12. Apply pressure via the syringe until the gauge stabilizes at 100 mmHg. Note:: If the pneumatics system is not tight enough to hold a stable pressure with a syringe, remove the RGM cover (see section 5.3, Cover removal and replacement) and connect the pressure gauge and syringe directly to the silicone tube that leads to the barometric pressure transducer (barometer) on the gas analyzer measurement board. A second option if the pneumatics system is not tight enough to hold a stable pressure is to use 100 mmHg vacuum for set LOW and ambient pressure to set HIGH. If this option is used, substitute the following steps 7, 8, and 12 for those in the procedure: 7. Apply 100 mmHg of vacuum. 8. Enter 100 mmHg less than ambient pressure. 12. Open the sample exhaust and inlet ports to ambient pressure. 13. Holding this pressure to within +1 mmHg, press the HIGH key. Pressing either LOW or HIGH automatically resets the calibration values unless you get an error message. 3-12 3/Checkout and Calibration Notes: • If the INVALID HIGH CALIBRATION message appears, check that the connections match Figure 3-6 and go back to step 8. • If the MUST DO LOW FIRST message is displayed, repeat this procedure starting at step 6. • If repeated attempts to calibrate barometric pressure are ineffective, set M to 26000 and B to 0 using the Enter Calibration Constants screen and try calibration again. See section 4.2.3, Enter Calibration Constants screen. 14. Press EXIT to return to the previous screen. The values in the following table are to be used only as a point of reference. Use them to check the uncorrected atmospheric pressure reading received from a local source (airport, etc.) to verify that it is truly an uncorrected (to sea level) value. The numbers used in this table should never be used to calibrate a unit. Altitude (elevation above sea level in feet) Estimated uncorrected barometric pressure reading at 25 °C 0 760 500 747 1000 734 1500 722 2000 709 3000 685 4000 662 5000 639 6000 617 7000 596 8000 576 3.6.3 PAW check 1. While viewing the Service screen, connect a squeeze bulb or syringe in parallel with a 0 to 150 cm H2O (0 to 100 mmHg) pressure gauge to the PAW inlet of the RGM. (A bulb is shown for the procedure in Figure 3-9.) Figure 3-9. Airway pressure test setup 3-13 5250 RGM Service Manual 2. Apply 0, 50, and 100 cm H2O (0, 36.8, and 73.6 mmHg) ±1 mmHg to the pressure inlet. The PAW display and pressure gauge reading should be within ± 5 cm H2O (+4 mmHg) of the gauge accuracy. If not, perform the next procedure, airway pressure calibration. 3. Exit to the display screen. 3.6.4 Airway pressure calibration 1. Go to the Service Screen and select CAL. 2. From the Calibration screen, select PAW CAL. The Airway Pressure Calibration screen appears. Airway Pressure Calibration PAW Pressure Air Way -1.2 (cmH2O) -0.08 (volts) Measurement Range -131.6 to 171.5 (cmH2O) Apply 0 cmH2O to the pressure inlet and touch the SET ZERO icon. Apply 60 cmH2O to the pressure inlet and touch the SET SPAN icon. SET ZERO SET SPAN EXIT Figure 3-10. Airway Pressure Calibration screen 3-14 3. Expose the PAW inlet to ambient atmospheric pressure and touch the SET ZERO key. The new value is displayed on the screen. 4. Connect a squeeze bulb or syringe and a pressure gauge in parallel with the PAW inlet. Refer to Figure 3-9 in section 3.6.3, PAW check. 5. Apply 60 cm H2O (44.1 mmHg) above ambient atmospheric pressure to the pressure inlet and touch the SET SPAN key. The new value is displayed on the screen. 6. Expose the PAW inlet to ambient atmospheric pressure. The PAW display should read 0 ±5 cm H2O (+4 mmHg). 7. Apply -10, 50, and 100 cm H2O (-7.4, 36.8, 73.6 mmHg) below/above ambient atmospheric pressure to the pressure inlet. The PAW display should read within ± 5 cm H2O (±4 mmHg). 8. Press EXIT to return to the previous screen. 3/Checkout and Calibration 3.6.5 Gas analyzer check—zero and span calibration Zero and span calibration WARNINGS: Data validity. Use only Datex-Ohmeda calibration gas of 1.75% Halocarbon-23 (CHF3)/freon or equivalent, 6% CO2, 40% N2O, 50% O2, and 2.25% argon. Use of the wrong calibration gas kit will result in improper calibration and the RGM will display incorrect gas calibrations. 1. From the display screen select MENU and then select CALIBRATE INTERNALS. Touch the ZERO icon on the Calibration Internal Sensors screen. After about 30 seconds, the ZERO COMPLETE message is displayed After 15 seconds, the values shown are: Freon – 0.0%, ±0.1% CO2 – 0.1%, ±0.1% N2O – 0.0%, ±1% O2 – 21.0%, ±1% Refer to 4/Messages and Troubleshooting. 2. Use the proper Datex-Ohmeda calibration gas (as indicated on the screen). Connect the sample tube from the sample inlet to the female luer connector of the calibration bag as shown in Figure 3-11. Connect the other end of the calibration bag to the can of calibration gas. 1 2 3 4 5 Figure 3-11. Cal gas connection 1 2 3 4 5 3. Calibration gas Brass restrictor Calibration bag Sample line Sample inlet Once the pumping action of the RGM has evacuated the calibration bag, the RGM automatically generates an alarm tone, and the message BEGIN SPAN CALIBRATION appears. 3-15 5250 RGM Service Manual 4. Press down on the valve stem of the gas canister until the calibration bag fills but is not pressurized; then touch the SPAN key on the Calibrate Internal Sensors screen. Notes: 5. • Allow the RGM to draw gas from the bag, but do not let the RGM totally evacuate the bag before span calibration is completed. Feed small amounts of gas into the bag to prevent the bag from being emptied. Continue this until the span calibration is complete. • If the calibration bag does not fill up, either the canister is low, the flow restrictor is blocked, or the bag has a leak in it. After about 25 seconds, the message SPAN COMPLETE appears. Verify that the values shown are the same as those on the calibration gas canister, with the following tolerances: Halocarbon 23 (CHF3)/Freon ± 0.1% CO2 ± 0 .1% N2O ± 1% O2 ± 1% If the Span calibration is unsuccessful, the INVALID SPAN message appears on the screen, and you should repeat the span calibration. If necessary, refer to section 4.3.5, Gas analyzer/calibration problems. 6. Disconnect the calibration gas from the sample inlet. 7. Select EXIT from the Calibrate Screen to return to the display screen. 8. Store the calibration bag in a location where it will be safe from puncture. 3.6.6 Circuit O 2 check WARNINGS: Data validity. • The oxygen monitoring portion of the 5250 RGM monitor should be calibrated at the same temperature at which it will be used to monitor oxygen delivery in the patient circuit. Operation at temperatures other than those present during calibration may result in readings outside of the stated accuracy for the monitor. When the ambient temperature changes, we recommend recalibrating the monitor for maximum accuracy. Refer to the O 2 sensor information sheet for more details. • Keep the circuit O2 sensor attached to the unit to assure accurate O2 readings. If detached, the sensor must be reattached and stabilized for up to 1 hour before recalibrating. CAUTION: Avoid storing O2 sensors outside the temperature range of -5 ºC to 50 ºC (+23 ºF to 122 ºF). O2 sensors must have shorting clips or be connected to the RGM when in storage or not in use. Note: Before checking the patient circuit O2 sensor, turn the RGM on and allow the O2 sensor output to stabilize for at least 10 minutes. The 10-minute stabilization assumes that the cartridge was removed from its sealed protective packaging just before calibration, or that the sensor had been connected to the RGM without interruption. 3-16 3/Checkout and Calibration 1. Connect the patient O2 sensor to the rear panel O2 connector. 2. Switch the RGM on and allow it to warm up for 10 minutes. 3. Select MENU from the display screen. 4. Select CALIBRATE CKT O2. The Calibrate Circuit O2 Sensor screen appears. 5. Expose the sensor to room air for a minimum of 2 minutes. 6. Check for 21% O2 ±3%. Note: The value may be blanked on the display screen if the circuit oxygen has not been enabled through the Setup Screen. 7. If the sensor is not within calibration, refer to the patient circuit O2 sensor calibration procedure in the 5250 RGM Operation and Maintenance Manual. 8. Apply approximately 5 l/min of 100% O2 to the circuit O2 sensor for 2 minutes to flush the room air from the sensor housing. 9. The CKT O2 display should read 100% O2 within ± 3%. If the sensor is not within calibration, refer to the patient circuit O2 sensor calibration procedure in the 5250 RGM Operation and Maintenance Manual. 10. After exposing the sensor to room air once again, the sensor should return to 21% ±3% within 2 minutes. 3.7 SpO 2 module checks (if installed) WARNING: To prevent patient injury or equipment damage, use only Datex-Ohmeda sensors identified for use with this monitor. For complete information about the safe and appropriate use of a sensor, consult the instructions for the sensor. 1. Check the sensor for foreign materials such as tape or cotton. Remove any substances that may interfere with the transmission of light between the emitter and detector. 2. Verify that a durable sensor opens and closes smoothly. 3. Check that the sensor is the correct model (see the user instructions for the sensor) before connecting it to the RGM. 4. Connect a sensor to the SpO2 sensor connector. Check that the connection is firm. 5. Turn on the RGM. Check that the sensor’s red emitter LED is on. 6. Attach the sensor to either a finger or an ear. Check that the sensor cable is not twisted. Verify the following: • A saturation value (SpO2) of greater than 90 percent is displayed. • A clean plethysmographic waveform depicting a pulse is displayed. • A digital pulse rate (PR) is displayed. • A beep occurs with every pulse. The tone frequency depends on the saturation value. High frequencies indicate high saturation values. Note: This beep tone can be disabled from the Setup Screen. 3-17 5250 RGM Service Manual 7. 8. 9. Verify that the high and low alarm limits for SpO2 and the pulse rate work (if enabled) by setting the high and low alarm limits beyond the indicated readings. Ensure that: • The alarm tone sounds. • The alarm light flashes. Remove the sensor from the finger or the ear and verify that: • The alarm message PROBE OFF PATIENT is displayed and the waveform and digital values are not displayed. • Three high alarm tones occur when the sensor is removed. Pressing alarm silence stops the audio. Unplug the sensor from the RGM and verify that the alarm message NO SPO2 PROBE appears on the screen. 10. Reconnect the sensor to the RGM and verify that the PROBE OFF PATIENT alarm message is present. Notes: • The alarm volume level is selected through the Setup Screen and remains as set until changed by the operator. • The high/low alarm limits are defaulted at power on and remain so until changed by the operator. 3.8 Patient flow sensor check (if installed) 3.8.1 Flow threshold test Use this check when you suspect a malfunction in either the flow transducer (TVX) or clip due to inconsistent or low readings for tidal volume. 1. Connect the TVX flow sensor from the back panel of the RGM to the common gas outlet of an anesthesia system or to a calibrated flow source known to be functioning properly. Note: Use 4 to 8 ft of tubing between the source and sensor to stabilize flow. 4 to 8 feet of tubing Anesthesia System or Calibrated Flow Source Figure 3-12. Flow threshold test setup 3-18 3/Checkout and Calibration 2. On the RGM display screen, select MENU, then • Select SETUP. • Select Patient type from the Setup Screen. • Select adult. NEXT PAGE Setup Screen Circuit O2 Rev. flow det Default Display RS-232 device Display labels Service mode Language Trend Clear N2O display Wave Type adult pediatric Patient type 90 VIEW ALL CLOCK SET EXIT Figure 3-13. Select adult patient type from Setup Screen 3. Set the gas flow from the flow source at 6.0 L/min. 4. Select Service mode and press “on.” Read the number for L/min in the righthand column. If the RGM does not agree with the calibrated air source to within ±2 L/min, replace the cartridge and repeat this test. CAUTIONS: 5. • Destroy malfunctioning flow cartridges to prevent their inadvertent use. • Never tamper with the set screws in the flow cartridge. Such action will render the cartridge unusable. If the RGM still does not agree with the calibrated air source, try replacing the sensor clip, then repeat the test. If the unit still fails the test, there is a leak in the test circuit or there is a malfunction in the RGM and repairs are required (see 4/Messages and Troubleshooting). 3-19 5250 RGM Service Manual 3.8.2 Reverse flow detection simulation 1. Remove the flow sensor cartridge from the breathing circuit. 2. To activate the reverse flow alarm on the RGM Setup Screen, • • Press Rev. flow det. Press “on.” NEXT PAGE Setup Screen Circuit O2 Rev. flow det Default Display RS-232 device Display labels Service mode Language Trend Clear N2O display Wave Type off on Patient type 90 VIEW ALL CLOCK SET EXIT Figure 3-14. Activate reverse flow alarm on Setup Screen 3. Flow gas through the sensor cartridge in the reverse direction. An intermittent low tone sounds, the yellow LED flashes and the REVERSE FLOW message appears and indicates a reverse flow alarm condition. Note: In this case the reverse flow is simulated for testing purposes only. WARNING: Patient safety. After completing simulation testing, be sure to return the sensor to its correct orientation in the breathing circuit. The arrows should point away from the patient. 3.9 Sample flow calibration 1. Select CAL from the Service screen to display the Calibration screen. 2. Select SAM FLOW to display the Sample Flow Calibration screen. Sample Flow Calibrat ion Sample flow rate Sample pump status Measurement range 192 1 -649 To (ml/min) 2.89 0=off 646 (m l/ min) (volts ) Press th e SET ZERO icon. A dd a flow restr ict or t o sample inlet and a f low mete r to sample exhaust. Af t er adjustin g restr icto r unti l f low met er reads 1 50 ml ± 15 ml, press SET SPAN. SET ZERO SET SPAN Figure 3-15. Sample Flow Calibration screen 3-20 EXIT 3/Checkout and Calibration 3. Connect a calibrated air flowmeter, p/n 6024-0000-006, (0-600 mL/min) to the sample exhaust outlet on the back of the RGM or to the sample inlet. Accuracy is ±2.0 percent of the reading or ±division, whichever is greater. Flowmeter 6024-0000-006 150 mL/min. Variable flow restrictor To Sample Exhaust Figure 3-16. Flowmeter and flow restrictor connections 4. Press the SET ZERO key. 5. Note that the sample pump switches off for this procedure. Wait for the ZERO CALIBRATION COMPLETE message to flash and the sample pump status to display 1 (ON). 6. Measure the flow at the sample outlet. Adjust the flowmeter flow restrictor until the flow is 150 mL/min ±15 mL/min. Then press the SET SPAN key. See Figure 3-15. Note: A flow restrictor can be placed either on the sample inlet or exhaust and achieve the same effect. 7. After SPAN calibration, the flow reading on the calibration screen should be the same as the external calibrated air flowmeter within ±30 mL/min. 8. Press EXIT to return to the previous screen. 3.10 Internal temperature calibration Internal temperature calibration affects only whether the DEVICE OVERHEATED message appears. If you are getting a false positive DEVICE OVERHEATED message, perform this procedure. 1. Remove the top cover from the RGM. (See section 5.3.) 2. Allow RGM to stabilize to ambient temperature for 20 to 30 minutes. 3. Unlock cal constants using DIP switch #3 on the signal processor. 4. Go to the Enter Calibration Constants screen (see Figure 4-1). 5. If the internal temperature displayed in °C is not accurate ±5 °C, subtract the reading on the screen from the current room temperature. Multiply this by ten and add it to the present temp B calibration constant. 6. Enter the resultant B as the new B and verify that the temperature shown is accurate ±5 °C. For example, if the ambient temperature is 25.2 and the monitor shows 35.1, (25.2 - 35.1) x 10 = -99. Add -99 to the present B value for internal temperature. 7. Lock Cal constants using DIP switch #3. 3-21 5250 RGM Service Manual 3.11 Analog output check Run the analog output check when you suspect a problem with the analog output subsystem or when calibrating a strip chart recorder. 1. Select MENU from the display screen. 2. Select SETUP from the menu. 3. Select Service Mode from the Setup Screen and press “on.” The Service screen appears. 4. Select TESTS to view the Diagnostic Selection screen, where you have the following choices: • Clear run time hours • Analog output • RS232 • EXIT 5. Attach an 8-pin DIN connector to the analog output jack. 6. Select ANALOG OUTPUT. 7. To run the DAC test, select ADDRESS. Chart Recorder Calibration/Analog Diagnostic ADDRESS – Outputs .1 V on dac channel 1, .2 V on dac channel 2, etc. RAMP – Generates 0-1 Volt ramp on all channels ZERO – outputs 0 V to all channels ONE – outputs 1 V to all channels ADDRESS RAMP ZERO ONE EXIT Figure 3-17. Analog output diagnostic screen 8. 3-22 Use a DVM to measure the output voltage of each pin (with respect to common) of the analog output connector as indicated in the following table. 3/Checkout and Calibration Pin Voltage in ADDRESS test 1 0.1 V DC ±25 mV 2 0.2 V DC ±25 mV 3 0.3 V DC ±25 mV 4 0.4 V DC ±25 mV 5 0.5 V DC ±25 mV 6 0.6 V DC ±25 mV 7 0.7 V DC ±25 mV 8 Common GND 1 7 2 6 5 3 4 Figure 3-18. Analog output connector Optional: 9. Connect an oscilloscope or chart recorder to measure the output from each pin. 10. Use the ZERO key to adjust the chart recorder to 0 V. 11. Use the ONE key to adjust the chart recorder to 1 V, full scale. 12. Select RAMP. During the ramp test, all seven pins should have a saw tooth wave range 0 to 1 V. One ramp takes approximately 3 seconds. The ramp should appear smooth with 1 jump from 1 to 0 V. CAUTION: Maximum voltage. No more than 5 volts should appear on any pin of the analog output connector. Channel assignments Channel Scale 1 CO2 waveform 66.6 mV/% 2 N2O waveform 10 mV/% 3 PAW waveform (V x 140)-20 = cm H2O 4 Flow waveform 10 mV –1 L/min 5 Agent waveform 66.6 mV/% (Desflurane: 33.3 mV/%) 6 O2 inspired/expired 10 mV/% 7 Plethysmograph 0-0.5 V with 16.1 mV resolution The analog output accuracy is ±3.5 percent (absolute), linearity ±0.5 percent, with an update rate of 10 mS and a resolution of 3.9 mV. Note: Output source impedance is 100 ohms. 3-23 5250 RGM Service Manual 3.12 RS-232 loop back hardware check Run this check if you have trouble communicating with either the 78xx ventilator or a printer attached to the RS-232 port. 1. From the RGM Setup Screen, select Service mode and touch “on.” 2. Select TESTS from the Service Screen to display the Diagnostic Selection screen. 3. Connect a loop back jumper (DB9 connector with pins 2 to 3 shorted) to the RS-232 port on the rear of the machine. Note: The connector must be installed before selecting the test. 4. Select RS232 from the Diagnostic Selection screen. 5. Run at least ten passes of the diagnostic. Note: The Diagnostic Pass Count numbers may not be consecutive due to the speed of the test (for example, 3, 6, 7, 9, 10, . . ., etc.) 6. Press EXIT to complete the test. 7. Remove the loop back jumper. 3.13 Leakage current and ground resistance checks 3-24 1. With the power cord connected only to the RGM, measure the resistance through the power cord ground prong to all exposed metal on the chassis. The measured resistance must not exceed 0.15 Ω. 2. Measure the enclosure leakage current of the RGM following the instruction supplied with your leakage current tester. The leakage current must not exceed 50 µA for 120 V units or 100 µA for 240 V units. 3. Record the results for reference in future resistance/leakage tests. A significant change may indicate an impending failure. 4/Messages and Troubleshooting This chapter contains • A list of messages you may receive on the screen. • Troubleshooting information. • Specific troubleshooting procedures. 4.1 Messages Message Category Range Default ACX DECODE TASK FAIL System failure alarm na na ACX XMIT TASK FAIL System failure alarm na na AGENT DETECTED Single-tone emergency na >0.8% halothane ANALOG FAIL System failure alarm na na AUTO ZERO IN PROGRESS Silent advisory na na CALIBRATE TASK FAIL System failure alarm na na COMMUNICATIONS FAIL System failure alarm na na CO2 APNEA Quiet emergency** 20 to 30 seconds 30 seconds DEVICE OVERHEATED Advisory na na DISPLAY CPU FAIL System failure alarm na na DISPLAY DECODE TASK FAIL System failure alarm na na DISPLAY DIAG TASK FAIL System failure alarm na na DISPLAY RAM FAIL System failure alarm na na DISPLAY ROM CHECKSUM FAIL System failure alarm na na DISPLAY TASK FAIL System failure alarm na na DISPLAY TOUCH PANEL FAIL System failure alarm na na DISPLAY TREND TASK FAIL System failure alarm na na DISPLAY VIDEO RAM FAIL System failure alarm na na DISPLAY XMIT TASK FAIL System failure alarm na na EVENT TASK FAIL System failure alarm na na GAS ANALYZER SATURATED Warning na na ** Three advisory tones (one high tone and one low tone) followed by an emergency tone (three high tones). 4-1 5250 RGM Service Manual Message Category Range Default GAS ANALYZER WARM-UP Silent advisory na na HIGH CIRCUIT O2 Warning 18 to 100% OFF HIGH Et CO2 Emergency 0 to 15% OFF HIGH EXPIRED AGENT Emergency 0 to 15% * HIGH FiCO2 Emergency 0 to 15% OFF HIGH INSPIRED AGENT Emergency 0 to 15% * HIGH INSPIRED O2 Emergency 18 to 100% OFF HIGH N2O Emergency 0 to 100% 80% HIGH SUSTAINED PAW Emergency 10 to 30 cm H2O 30 cm H2O HIGH MINUTE VOLUME Warning 0.2 to 50.0 L OFF HIGH PAW Warning 0 to 120 cm H2O 80 cm H2O HIGH PULSE RATE Warning 80 to 250 bpm OFF HIGH SpO2 Warning 70 to 99% OFF HIGH TIDAL VOLUME Warning 50 to 2500 mL OFF INSUF LIGHT DETECTED Advisory na na INTERFERENCE ON SpO2 Silent advisory na na LOW CIRCUIT O2 Emergency 18 to 100% 18% LOW Et CO2 Emergency 0 to 15% OFF LOW Expired AGENT Emergency 0 to 10% * LOW Inspired AGENT Emergency 0 to 10% * LOW INSPIRED O2 Emergency 18 to 100% 18% LOW MINUTE VOLUME Warning 0.2 to 50.0 L OFF LOW PULSE RATE Warning 40 to 150 bpm OFF LOW QUALITY SpO2 SIGNAL Emergency na na LOW SpO2 Emergency 50 to 100% 90% LOW Sustained PAW Warning -20 to 20 cm H2O OFF LOW TIDAL VOLUME Emergency 50 to 2500 mL OFF NO BREATH DATA System failure alarm na na NO FLOW Advisory na <100 mL flow NO SpO2 PROBE NO VACUUM Silent advisory na na Advisory na na PNEUMATIC TASK FAIL System failure alarm na na PURGING Advisory na na RECALIBRATE BAROMETER * 4-2 Advisory na na Default depends on agent selected. Refer to Anesthetic Agent display in 5250 RGM Operation and Maintenance Manua l. 4/Messages and Troubleshooting Message Category Range Default RECALIBRATE CIRCUIT O2 Advisory na na RECALIBRATE GAS ANALYZER Advisory na na RECALIBRATE PAW Advisory na na REVERSE FLOW Warning fixed OFF SAMPLE FILTER BLOCKED Advisory na na SAMPLE LINE/FILTER BLOCK Advisory na <100 mL flow WATER TRAP FULL, PUMP OFF Warning na na SERIAL DEVICE ERROR Advisory na na SERIAL TASK FAIL System failure alarm na na SERVICE TASK FAIL System failure alarm na na SET NEW AGENT LIMITS Silent advisory na na SIGNAL CPU FAIL System failure alarm na na SIGNAL DECODE TASK FAIL System failure alarm na na SIGNAL DIAG TASK FAIL System failure alarm na na SIGNAL RAM FAIL System failure alarm na na SIGNAL ROM CHECKSUM FAIL System failure alarm na na SIGNAL TREND TASK FAIL System failure alarm na na SpO2 INOPERATIVE Advisory na na SpO2: INSUF LIGHT DETECTED Silent advisory na na SpO2 PROBE FAILURE Advisory na na SpO2 PROBE ID ERROR Advisory na na SpO2 PROBE OFF PATIENT Emergency*** na na SUB ATMOSPHERIC PAW Emergency fixed -15 cm H2O TARGET CHECK TASK FAIL System failure alarm na na VENTILATOR COMM FAIL Silent advisory na na VENT–CHECK GAS SUPPLY Advisory na na VENT–CHECK O2 PROBE Advisory na na VENT–HIGH CIRCUIT O2 Warning Set on 78xx na VENT–HIGH PAW Emergency Set on 78xx na VENT–HIGH SUSTAINED PAW Emergency fixed na VENT–LOW CIRCUIT O2 Emergency Set on 78xx na VENT–LOW MINUTE VOLUME Warning Set on 78xx na VENT–LOW PAW Emergency Set on 78xx na VENT–REVERSE FLOW Warning fixed na VENT–SUBATMOSPHERIC PAW Emergency fixed na *** This alarm is acknowledgeable by pressing alarm silence. 4-3 5250 RGM Service Manual Message Category Range Default VENT–TIDAL VOLUME APNEA Quiet emergency** na na VENT–TV APNEA ALARM OFF Advisory na na VENT–VOLUME SENSOR FAIL Advisory na na VENT–VOL MONITOR STANDBY Advisory na na ** Three advisory tones (one high tone and one low tone) followed by an emergency tone (three high tones). 4.2 Troubleshooting information Refer to the 5250 RGM Operation and Maintenance Manual and the previous chapters of this manual to learn how a properly operating RGM functions and to determine what the RGM is not doing correctly. Refer to this chapter and the following chapters in this manual to help locate and correct a malfunction. If you are not familiar with the information in these troubleshooting sections, refer to 2/Theory of Operation to better understand the RGM functions and refer to 5/Repair Procedures for the proper disassembly of the RGM components. CAUTION: Static sensitivity. This 5250 RGM’s electronic components that are susceptible to damage by electrostatic discharge. When disassembling the RGM, • Work at a static-control workstation and wear a static-control wrist strap to discharge accumulated static charges. • Use static-safe tools (tools with conductive or dissipative handles). Observe the safety precautions associated with using conductive tools in a high-voltage environment. • Handle the circuit boards (replacement and defective) by their nonconductive edges. Use antistatic containers for transport. Let the symptoms and the test results be your guide. 4-4 • If only a single function is not working, concentrate on the circuit that controls that function. • If several symptoms are present, locate a circuit that is common to all the affected functions. • If the solution to the problem is not obvious, check the power supply circuits first––observe the LEDs on the power supply board. • Perform a visual inspection of the RGM. • Verify that all controls are set properly. • Look for any obvious conditions that may be the cause of the malfunction: loose connections, broken or burnt (discolored) components. • Verify that all circuit boards are properly seated in the motherboard. 4/Messages and Troubleshooting 4.2.1 Power-up self-test When the RGM is first turned on, it tests several major functional areas of its internal circuitry. As each test passes, it performs the next test in line until an error is detected and displayed, or until all tests pass and the message POWER UP SELF TEST IN PROGRESS disappears. If a test fails, a corresponding error message (indicating which test has failed) appears in the upper-right corner of the display as follows: System error messages Error messages (alphabetically) Repair action Analog fail Replace the signal processor board. COMMUNICATIONS FAIL See comm fail troubleshooting. DISPLAY CPU FAIL Replace the display controller board. DISPLAY RAM FAIL Replace the display controller board. Display ROM checksum fail Replace the display software cartridge. DISPLAY . . . TASK FAIL Replace the display software cartridge, or replace the display controller board. DISPLAY TOUCH PANEL FAIL Ensure that the unit was not powered on while touching the screen or with any object blocking the touch panel. Replace the touch frame, or Replace the display controller board. DISPLAY Video RAM Fail Replace the display controller board. INVALID . . . ZERO or span See gas analyzer troubleshooting. INVALID O2 ZERO or span See O2 sensor troubleshooting. NO BREATH DATA See communications troubleshooting. No flow See pneumatics troubleshooting. No Vacuum See pneumatics troubleshooting. Signal CPU fail Replace the signal processor board. Signal RAM fail Replace the signal processor board. Signal ROM checksum fail Replace the signal processor software cartridge. SIGNAL . . . TASK FAIL Replace the signal processor software cartridge, or Replace the signal processor board. 4-5 5250 RGM Service Manual 4.2.2 Using service screens Service screen Access the Service screen by the following touch sequence: 1. MENU 2. SETUP (Additional information about RGM configuration is found on the VIEW ALL screens under SETUP. In particular, the status of the configuration jumpers and switches is shown there.) 3. NEXT PAGE (if multipage setup screen) 4. SERVICE MODE 5. ON The main service screen provides a summary of the overall state of the machine. This screen shows the various software versions (SpO2 version appears only if the function is present) and the readings from all of the sensors. Readings here, and on the Enter Calibration Constants screen are formatted in both units (i.e., %CO2) and volts. The “Breath volume data” indicates uncorrected tidal volume of last breath. This number will only update after “breath” start and end have been detected. The SpO2 data is the same as that on the normal monitoring screen, except the pleth data is a periodic snapshot of the voltage from the oximeter board. Agent ACX-200 Version Display Version SpO2 Version Signal Processor Version Agent Barometric Pressure Breath volume flow Breath volume data Cal constants CO2 Carbon Dioxide Internal Temperature SpO2 Pleth SpO2 Pulse Rate Saturation O2 N2O Nitrous Oxide O2 Circuit Oxygen O2 Internal Oxygen PAW Pressure Airway Pump Voltage Sample flow rate Oxygen valve status Purge valve status Water trap full status Zero valve status ENTER CAL PNEUMA Figure 4-1. Service screen 4-6 1.20 6.00 S 8.007 0.00 627.5 --0 0 0.10 24.5 ------0.0 0.0 20.6 0.1 6.00 201 0 0 0 0 Service 0402 0000 ACX status hex ACX error status hex CPU type V25+ (%) 0.00 (torr) 627.5 (ml/breath) 0.0 (liters/minute) Uncorrected 0 (min 25.6us cnts/flip) 1=locked 92 Run Time Hours (%) 0.10 (degrees C) 0.28 (volts) 0.45 (volts) (Beats/min) (%) (%) -0.8 (%) -10.00 (volts) (%) -8.00 (volts) (cmH2O 0.01 (volts) (volts) 153 (dac counts) (ml/min) 3.15 (volts) 0=oxygen cell active 1=in purge state 1=full 1=in zero state TESTS CAL EXIT 4/Messages and Troubleshooting From this screen the following additional screens can be accessed: Screen Function ENTER CAL Allows operator to view and alter the calibration constants directly. PNEUM Allows direct control over the pump and solenoids to aid in diagnosing problems with the pneumatics system. TESTS Accesses tests of the analog output and the RS-232 port. CAL Allows calibration of the barometer, PAW, and sample flow, which are less frequently required. “ACX Status” is the hex or decimal representation of binary data indicating status bit condition of the gas analyzer. The number is the sum of the following powers of 2 with the indicated meanings. ACX status Hex Decimal Description 8000 32768 Gas measuring disabled (motor or lamp off) 4000 16384 Motor not running (out of sync, or accelerating) 2000 8192 EEPROM writing not complete 1000 4096 Initial AGC control 800 2048 Gas sampling disabled 400 1024 Open gas circuit 200 512 Occlusion 100 256 Averaging on 80 128 00= halothane, 10=isoflurane, 40 64 01=enflurane 20 32 Temperature sample frozen 10 16 Pressure sample frozen 8 8 Agent sample frozen 4 4 N2O sample frozen 2 2 O2 sample frozen 1 1 CO2 sample frozen Some of these values (such as occlusion and open gas circuit) are not relevant in the RGM, because it has its own pneumatics sensing and control. For example, a normal ACX status would be 1090 (44216 ), which means open gas circuit, oxygen sample frozen, and enflurane selected (1024 + 64 + 2). 4-7 5250 RGM Service Manual ACX error status Hex Description Recommended action 0001 RAM test fail Fatal malfunction, requires service 0002 EPROM checksum Fatal malfunction, requires service 0004 external RAM test fail Fatal malfunction, requires service. Check for +5 V on the measurement board RAM, D22, pin 28. 0008 EEPROM measurement board error Fatal malfunction, requires service 0010 EEPROM sensor board error Host device tried to write to sensor EEPROM, requires service. 0020 Automatic gain control error Fatal malfunction, requires service 0040 Analog digital converter error One or more gases out of gain, Signal Amplitude error. Cycle power; if bit still set, requires service. 0080 Digital to analog converter error One or more gases out of gain. Measurement reference Shape error. Cycle power, if bit still set, requires service. 0100 ACX-200 internal auto zero in progress Internal adjustment of reference voltage. No action required unless this bit stays on. 0200 Sync pulse from motor none 0400 Motor none 0800 Pressure out of range Pressure out of range, if everything else O.K., requires service. 1000 Open gas circuit (Note: This is normal for ACX) none 2000 Zero valve open none 4000 Software none 8000 Watchdog none The gas analyzer is considered to have failed if any of the bits in the mask 08FF are set in the ACX error status. 4-8 4/Messages and Troubleshooting 4.2.3 Enter Calibration Constants screen The ENTER CAL page provides a summary of most signal inputs and all system calibration factors. The value column shows the input value in units with software limits imposed; for example, CO2 value is shown as a percent. The volts column is the raw input signal voltage; except that CO2, N2O, agent, and barometric pressure are raw digital values from the ACX-200 measure-ment board. System calibration factors slope (m) and intercept (b) are displayed in the last two columns. Procedure to recall barometer factory calibration constants The procedure to recall factory calibration should be used only as a temporary measure if you do not have proper equipment for calibration. It is absolutely necessary that the monitor be properly calibrated as soon as possible to ensure that the unit meets specifications. 1. Unlock the calibration constants by setting DIP switch #3 on the signal processor board; see section 3.2, Dip switch and jumper positions. 2. Set the cursor to the line you wish to default (either m or b), then press DEFAULT LINE twice. 3. Lock the calibration constants using DIP switch #3. 4. Cycle power to the RGM and verify that no error message other than RECALIBRATE appears. Note: For internal temperature and sample flow, no RECALIBRATE message appears. 5. Go to the Enter Calibration Constants screen. Highlight the line that was defaulted, either m or b, and press Enter. This will turn off the RECALIBRATE message. Note: The screens below contain default calibration constant values. Default values for agent, barometric pressure, CO2, and N2O are different for every gas analyzer. DEFAULT LINE Enter Calibration Constants Agent Barometric Pressure CO2 Carbon Dioxide Internal Temperature N2O Nitrous Oxide O2 Internal Oxygen PAW Pressure Airway Sample flow rate value volts 0.00 -0.12 617.6 617.6 0.14 0.14 23.3 0.04 0.1 0.1 18.6 -8.27 0.1 -10.00 -1.8 -0.12 300 6.27 0 1 2 3 4 5 zero 0 0 63446 0 0 345 0 0 0 6 m 33982 31800 10269 2000 4602 3492 6080 6064 2592 7 b 26220 63850 33623 65267 42880 0 0 64220 64885 8 9 U s e up /do wn a rro ws to s e lec t m or b. U s e CLEA R and n umb er s t o en te r v a lue . Pr e s s EN T ER to s et high ligh te d m a nd b. P re s s D EF AUL T LI NE t o r e que s t de fa ult v al ues fo r h igh ligh ted m an d b to be l oad ed on t he ne x t po wer u p. Af te r D EF AUL T LI NE or DEF AU LT AL L c y c le pow er . CLEAR ENTER DEFAULT ALL EXIT Figure 4-2. Enter Calibration Constants screen 4-9 5250 RGM Service Manual Pneumatics Service Screen The Pneumatics Service Screen provides an overview of the status of the pneumatic system and allows manual control of the valves and the pump. The barometric pressure, sample flow rate, valve status, and water trap full status are displayed. Purging is suppressed while in this screen. Valves and pump are restored to normal operation when exiting this screen. Pneumatics Service Screen Barometric Pressure CO2 Carbon Dioxide Last zero error O2 Internal Oxygen Oxygen valve status Pump Voltage Purge valve status Sample flow rate Sample pump status Water trap full status Zero valve status 597.3 0.00 none 20.6 0 6.00 0 201 1 0 0 Zero Input Sample Inlet CO2 N2O Agent (%) 0=oxygen cell active (volts) 1=in purge state (ml/min) 0=off 1=full 1=in zero state O2 597.3 0.00 0 -8.02 153 3.13 Converse value (volts) (dac counts) (volts) Purge Sample Zero Sample Exhaust Valve ON O2 (torr) (%) OFF Sensors UP Valve DOWN Sensor ZERO Valve PURGE Pump O2 EXIT Figure 4-3. Pneumatics Service Screen Option Function Option Function On Turns on pump Zero Toggles zero valve Off Turns off pump Purge Toggle purge valve Up Raises pump voltage O2 Toggles O2 valve Down Lowers pump voltage Exit Returns to main service screen Zero error codes No flow Insufficient flow for zero Vacuum high Too much vacuum for zero Noise high Patient waveform seen during zero Zero inlet Barometric pressure did not return to ambient N2O error N2O zero error CO2 error CO2 zero error Agt/ O2 error O2 zero error Agt error Agent zero error On this screen, the pump can be toggled ON and OFF. Also, all valves (zero, purge, and O2) can be individually controlled. The best method for leak testing is to set the purge valve on, pump on, zero valve off, and to block the sample inlet. Then to localize a leak, use hemostats or needle-nose pliers to pinch the tubing from the pump, to the purge valve, then to the line, then to the water separator, then to the zero valve, then to the gas analyzer, then to the O2 sensor. When the flowmeter drops to zero there isn't a leak. When the flowmeter doesn't go to zero, you've located the leak. 4-10 4/Messages and Troubleshooting 4.3 Troubleshooting tables 4.3.1 Nonfunctioning monitor or power supply Condition Cause(s) Recommended action(s) Totally dead; no fan, display, or sounds. Blown primary fuses on rear panel power input module. Replace fuses with correct type and rating. Improper line voltage setting. Correct voltage input setting on rear panel power input module Bad front panel power switch. Unplug monitor from AC power and check power switch continuity. Blown thermal fuse in transformer. Unplug monitor from AC power and check continuity of yellow wires on transformer. If open circuit, replace toroid assembly. Blown secondary fuses. Note: Fuses for +5 and +15 V tend to blow as pairs. Replace secondary fuses with correct type and rating. Bad power supply board. With power off, unplug all cables into power supply board. If LEDs are still dim with no load, replace the power supply board. Bad system board loading down supply. With power off, unplug all DC cables into power supply board (and display cable into motherboard). Turn on power adding one cable connection at a time to isolate which board overloads the supply. Replace the bad board. Bad DC power cables from the power supply board. Check connections and cable continuity All power supply board LEDs are dark. Some power supply board LEDs are dark. Some power supply board LEDs are dim. All power supply board LEDs are lit. Could also be caused by a short. 4.3.2 TVX/patient flow sensor problems Condition Cause(s) Recommended action(s) Reads too low Bad TVX flow sensor. Replace TVX flow sensor cartridge. Numerous REVERSE flow alarms Leaking exhalation check valve in anesthesia circuit. Replace exhalation check valve. Using pediatric mode on adult patient. Go to SETUP menu and select adult patient type. Bad TVX flow sensor clip connection. Check connection of sensor clip cable to rear panel. Failed signal processor board. Replace signal processor board. Incorrect TVX flow sensor clip. Order and install new clip. No numbers appear on main screen 4-11 5250 RGM Service Manual 4.3.3 Communication fail or signal processor failure If the COMMUNICATIONS FAIL message appears, remove the top shroud and look at the LED bar on the signal processor board: Bit # 0 1 2 3 4 5 6 7 8 9 RAM test Bits 0-3 4 5 6 7 8 9 Status CO2 breath bar graph PAW breath indicator O2 solenoid Purge solenoid Zero solenoid Watchdog failure Spare solenoid valve Watchdog fail 1 2 3 4 5 6 7 8 9 10 Any State If poweron halts with the LED bar's lights 1 and 8 set, the RAM test has failed and the signal processor board should be replaced. Flashing If, during poweron, the LED bar's light 8 flashes slowly and repeatedly, a watchdog failure has occurred and the signal processor board, signal processor ROM cartridge, or power supply should be replaced. Figure 4-4. Communications, signal processor, or power supply fail and LED status Condition Cause(s) Recommended action(s) COMMUNICATION FAIL message and the system halts at RAM test, or the watchdog continually flashes. Signal software module or signal processor board is bad. Replace the signal software module or the signal processor board. +15 V is out of specifications. Power supply failure. Check the +15 V supply or replace the power supply. AC line voltage is too high or low. Check if the message disappears when the line voltage is at the nominal value. Replace the toroid. Lithium coin battery may be bad. Check the battery voltage, measure top of battery (under tape) compared to ground (TP12). If voltage is less than 3.0 V, replace lithium coin battery. Hardware problem on reset line of 72001, on signal processor board. Replace signal processor board. RECALIBRATE BAROMETER, RECALIBRATE GAS ANALYZER, RECALIBRATE CIRCUIT O2, or RECALIBRATE PAW 4-12 4/Messages and Troubleshooting Condition Cause(s) Recommended action(s) NO BREATH DATA, SIGNAL CPU FAIL, SIGNAL RAM FAIL, SIGNAL ROM CHECKSUM FAIL, SIGNAL … TASK FAIL Signal software module or the signal processor board is bad. Replace the signal software module or the signal processor board. ANALOG FAIL Analog power supply may be bad. If analog +15 V power supplies are not within +0.75 V, replace power supply board. Signal processor board ADC or DAC has failed. Replace signal processor board. The unit is unable to communicate with the device connected to the RS-232 port. Check the RS-232 connections (check for bent pins, too). SERIAL DEVICE ERROR Check RS-232 device screen for proper device selection. Check RS-232 power. See section 3.4. Perform RS-232 loop back test. If test fails, replace signal processor board, rear panel filter board or motherboard. 4.3.4 O2 sensors/O2 calibration problems Condition Cause(s) Recommended action(s) Sudden erratic changes in reading, (especially when changing sensor orient-ation), or repeated difficulty calibrating Possible bubbles in O2 sensor or sensor is bad. Replace the O2 sensor. INVALID O2 ZERO Bad cable or connection to O2 sensor. If ENTER CAL screen shows about -10 V, check cable continuity and pins on O2 sensor housing. INVALID O2 SPAN O2 zero calibration is invalid or has not been done. First do zero calibration. Bad O2 sensor. Replace O2 sensor. No circuit O2 readings appear on main screen SETUP selection not entered for circuit O2. Go to SETUP screen and select circuit O2. O2 readings near zero Bad cable or connection to O2 sensor. If ENTER CAL screen shows about -10 V, then check cable continuity and pins on O2 sensor housing. Clean contacts on O2 sensor housing. Continual downward drift of O2 values during operation. Newly installed O2 sensor has not stabilized. Leave O2 sensor installed for an hour, then recalibrate. 4-13 5250 RGM Service Manual 4.3.5 Pneumatics problems Condition Cause(s) Recommended action(s) INVALID ZERO or INVALID AUTO ZERO Pneumatic error. See zero error codes in Section 4.2.3. Auto zero valve, purge valve, or scrubber failure. Unit requires service. Note: An occasional (less than twice a day) N2O, CO2, or agent error is normal. LAST ZERO ERROR Indicates failure in the previous invalid zero. Retest for zero calibration. NO VACUUM Severe internal pneumatic leak. Isolate and repair leak. Bad internal barometric pressure calibration. Check barometric pressure calibration. Recalibrate barometric pressure if necessary. If recalibration is unsuccessful, replace gas analyzer/measurement board assembly. Pump drive circuit. Reseat measurement board. Bad pump. Replace vacuum pump. Bad internal flow sensor calibration. Check sample exhaust flow with flowmeter. Recalibrate flow sensor if necessary. If recalibration is unsuccessful, replace pneumatics board. Inadequate sample flow because of blocked filter or restrictor. Locate source of restriction. Replace filters and restrictors until adequate flow returns. Inspect inlet luer and tubing to membrane separator. Severe internal leak between the pump and flow sensor. Isolate and repair leak. Bad pump. Replace vacuum pump. Water trap bottle full of water, or sensor tip dirty. Empty bottle and clean plastic rod tip with soap and water. Misalignment or failure of trap full sensor emitter or detector. Disassemble membrane separator assembly and replace trap full sensor assembly. SAMPLE FILTER BLOCKED Clogged sample filter cart-ridge or blocked/ kinked tube in gas analyzer tubing. Replace sample filter and check tubing through gas analyzer and O2 sensor. PURGING with every inspiration Kink or blockage in tubing to O2 sensor. Replace tubing around O2 sensor. SAMPLE LINE BLOCKED Blocked patient sample tube, clogged sample filter cartridge, or blocked line on membrane separator assembly. Replace sample tube and filter cartridge. Examine membrane separator assembly for blockages. NO FLOW NO VACUUM and NO FLOW messages together WATER TRAP FULL 4-14 4/Messages and Troubleshooting 4.3.6 Gas analyzer/calibration problems Condition Cause(s) Recommended action(s) INVALID CO2 ZERO, Possibly invalid zero. Retest zero calibration. INVALID N2O ZERO INVALID AGT ZERO, or GAS ANALYZER INOPERATIVE No filament voltage to photometer. Check the m and b values. No power to the measurement board RAM. See no filament voltage condition below. Open filament on photometer. Check for 5.0 V, +0.5 V on pin 28 of RAM, D22. Zero valve failure. Disconnect photometer cable (J4) from power supply. If filament resistance is not 4.0 Ω, +2.0 Ω, replace gas analyzer assembly. Use valve checkout procedure (in 3/Checkout and Calibration) to test valve function. Replace valve if failed. INVALID CO2 SPAN, INVALID N2O SPAN, or INVALID AGT SPAN Possibly invalid zero. Retest zero calibration. Incorrect or bad gas mixture. Check the m and b values. Sample line blocked. Check gas concentration against values listed on CALIBRATE screen. Unblock or replace the sample line. Unable to select agent SW-1 on motherboard set incorrectly. Set switch to match photometer type. Unable to select desflurane, sevoflurane Failed measurement board or photometer. Replace measurement board or replace photomoter. CO2 reads correctly in percent, but low in mmHg or kPa Barometric pressure reading low. Check barometric pressure calibration. Recalibrate if necessary. If unable to recalibrate, replace gas analyzer assembly No filament voltage (5.0 V, +0.5 V) across screw terminals of photometer Bad lamp control circuit on power supply board. If lamp control input (J1-20) of power supply is 5.0 V, +0.5 V and no filament voltage appears, replace the power supply board. Bad motherboard to power supply cable. If no lamp control voltage appears (J1-20) of power supply, check DC power cable continuity. Heater wires Check heater wires. Bad measurement board. Using extender card, measure lamp control voltage (pin 13B) of measurement board. If value is not 5.0 V, +0.5 V, replace gas analyzer assembly. Bad photometer. Replace photometer. 4-15 5250 RGM Service Manual 4.3.7 Display assembly problems Condition Cause(s) Recommended action(s) Display is blank and constant high alarm tone is heard. The display watchdog circuit is active. Replace the display software module or the display processor board. Display is blank and normal power up tones are heard. Display matrix power supply is bad. Unplug main DC power cable from power supply. If power supply J1-17 is not 15 +1 V (unloaded) then replace power supply board. Display cable or display matrix ribbon cable is bad. Check cable continuity and replace defective cable(s). Display matrix panel is bad. Replace display matrix panel. During the first 2 minutes of poweron, fine horizontal lines appear through the display. Normal powerup test of video RAM. No action required. DISPLAY TOUCH PANEL FAIL Monitor powered on with obstructed display screen. Switch off the power and switch it back on. Display controller board has failed. Replace the display controller board. Display touch panel frame has failed. Replace the touch panel frame. DISPLAY CPU FAIL, Display software module or DISPLAY RAM FAIL, the display controller board DISPLAY ROM CHECKSUM FAIL, is bad. DISPLAY . . . TASK FAIL, or DISPLAY VIDEO RAM FAIL Replace the display software module or the display controller board. 4.3.8 SpO2 oximeter board problems Condition Cause(s) Recommended action(s) Board is present but no SpO2 field appears on main screen Configuration dip switch on signal processor board not set or inactive. Set correct dip switch setting (switch 1 OFF) and cycle power. SPO2 INOPERATIVE Oximeter board is bad. Replace oximeter board. Signal processor board is bad. Replace signal processor board. SpO2 pulse beep does not sound and heart icon does not flash Bad pulse port bit connection from MINX to signal processor. Replace oximeter board or signal processor board or motherboard. SpO2 pulse beep does not sound and heart icon does flash SETUP selection for SpO2 pulse beep volume set to off. Go to SETUP screen and select SpO2 pulse beep volume. No plethysmographic waveform Bad analog connection from oximeter board to signal processor. Replace oximeter board or signal processor board or motherboard. 4-16 4/Messages and Troubleshooting 4.3.9 Intermittent problems or unusual behavior Condition Cause(s) Recommended action(s) Sudden switch to NO AGENT or to GAS ANALYZER WARMUP Signal processor board watchdog reset. Replace signal processor board. Intermittent NO FLOW and NO VACUUM on powerup Slow starting pump. Replace pump. Intermittent subatmospheric PAW Bad ribbon cable from signal processor to pneumatics board. Replace ribbon cable. DEVICE OVERHEATED Vent holes blocked. Remove top cover and inspect fan and lint buildup in chassis and in display assembly. Vacuum clean. Fan failure. Check fan voltage (12 V) on power supply board connector J2. Replace fan if voltage is good. Check for obstructions that may prevent operation. Remove any obstructions. Display blanking on warm running unit Display software module or display controller board bad. Replace display software module or display controller board. High agent readings or AGENT DETECTED message at start of case Traces of isopropyl alcohol, if used to clean breathing circuits. Flush all traces of alcohol from breathing circuit before using on a patient. 1% to 2% N20 reading appears when no N20 is present An interaction between the anesthetic agent and N20 absorbtion wavelengths in the gas analyzer may infrequently produce this phenomenom; it is most likely to be seen with isoflurane as the agent and is clinically insignificant. No action required. 4-17 5250 RGM Service Manual Notes 4-18 5/Repair Procedures This chapter contains: • Maintenance schedule. • Datex-Ohmeda service policy. • Procedures for removing and replacing parts. Important: If the cover is removed for any test or repair procedure, always perform the Functional checkout (section 3.1) and Leakage current and ground resistance checks (section 3.13) before returning the unit for patient monitoring. 5.1 Maintenance schedule Interval Function As needed Replace or clean airway adapter and sample tubing. Replace membrane filter cartridge. Replace flow transducer (TVX) cartridge. Daily or trap full message Clean and disinfect water trap bottle and optical sensor cone. Every 30 days Perform 02 sensor linearity check (see 5250 RGM Operation and Maintenance Manual). Daily to monthly Check zero and span calibration (see 3/Checkout and Calibration). Quarterly to yearly Vacuum clean the inside of RGM and display assemblies. Yearly Perform service checkout procedure. Replace internal filters and tubing assembly (see section 5.9.1, Pneumatics tubing assembly removal and replacement). Replace auto zero scrubber (see section 5.9.8, Scrubber replacement). Two years Replace O-ring seals on water trap bottle. Replace O-ring seals on membrane separator assembly. Four years Replace signal processor board battery. 5-1 5250 RGM Service Manual Note: Refer to the 5250 RGM Operation and Maintenance Manual for the following: • • • • • • • • • Software replacement Signal processor software replacement Display panel software module replacement Sample filter cartridge replacement Internal O2 sensor replacement Circuit O2 sensor assembly and cartridge replacement Flow transducer cartridge replacement Flow sensor clip replacement Line fuse replacement 5.2 Service policy Warranty repair and service must be performed by Datex-Ohmeda. When the warranty is not applicable, repairs are made by Datex-Ohmeda at the current list price for replacement parts plus a reasonable labor charge. Do not use malfunctioning equipment. Make all necessary repairs or have the equipment repaired by a Datex-Ohmeda service representative. For all repairs, use only genuine replacement parts and service kits sold by Datex-Ohmeda. Parts listed in this manual may be repaired or replaced by a competent, trained person who has experience in repairing devices of this nature. 5.2.1 Obtaining service To order parts or obtain assistance, contact Datex-Ohmeda or your authorized service office (see the back cover of this manual). Packaging and return procedure If you are instructed to ship the monitor to Datex-Ohmeda or to an authorized service office for repair, follow these steps: 1. Clean the equipment as described in section 1.7 of this manual. The unit must be thoroughly dry before you pack it for shipment. 2. Package the monitor carefully for shipment (in the original shipping container, if possible). 3. When you call for assistance, verify the shipping requirements. You may be required to include the following items: • • • • • 4. 5-2 A letter describing the problem in detail. Person to contact for questions about necessary repairs (name, telephone or fax number, and country). Ship to and Bill to information. Warranty information (a copy of the invoice or other applicable documentation). Purchase order number for tracking purposes or to cover repairs (if out of warranty). Ship the monitor as directed by your service office. 5/Repair Procedures 5.3 Cover removal and replacement Refer to section 6.1.2 for the 5250 Main unit assembly illustration. WARNING: Electrical shock hazard. Before cleaning or repairing the monitor, power it off and disconnect it from AC mains power. WARNING Biohazard. When handling materials that may have come into contact with patient exhalant or fluids, follow approved procedures for contamination control. CAUTION: Static sensitivity. Internal electronic components are susceptible to damage by electrostatic discharge. To avoid damage when disassembling the RGM, observe the standard precautions and procedures for handling static-sensitive components. Figure 5-1. Cover removal and replacement 1. Turn off the power and disconnect the power cord. 2. Set up the static-control workstation and use it whenever the top cover is removed from the RGM. 3. Press the Display Release latch and slide the display panel to the right. Empty the fluid bottle on the membrane water separator into a receptacle designated for patient waste, replace the bottle, and latch the Display Release. 4. Remove the two side mounting screws and washers for the cover. Note: Save all hardware for reuse. 5. Remove the two rear mounting screws and lock washers for the cover. 6. Slide the cover toward the back of the RGM. Remove the cover. 7. Remove the three top-plate mounting screws and remove the plate. 8. To replace the cover, reverse the order of steps 3 through 7 above. 9. Removal of the cover requires an electrical safety inspection be performed prior to using the unit, see Section 3.13. 5-3 5250 RGM Service Manual 5.4 Front panel Refer to the illustrations in section 6.2.2, Access panel assembly. 5.4.1 SpO2 connector replacement Mounting screw Figure 5-2. SpO2 connector replacement 1. Remove the RGM cover assembly. See section 5.3. 2. Remove the three screws from the access panel. 3. Remove the three mounting screws for the front panel from the left and two from the front right side access bezel. 4. Remove the SpO2 cable connector from the oximeter board. 5. Remove the three mounting screws, two from the SpO2 bracket assembly and one from the interface board (PCA, SpO2 with cable). 6. Remove the toroid ring with the board and cable connected. 7. Remove the old SpO2 connector from the front panel by removing the locknut. 8. Place the new connector in position and mount it with the locknut; finger tighten only. 9. Reinstall the SpO2 interface board and toroid. 10. Reinstall the SpO2 bracket assembly. 11. Reconnect the cable to the oximeter board. 12. Reinstall the front panel and top cover. 5-4 5/Repair Procedures 5.4.2 Power switch replacement 1. Remove the RGM cover assembly. See section 5.3. 2. Remove the three screws from the access panel. 3. Remove the three mounting screws for the front panel from the left and two from the front right side access bezel. 4. Remove the two mounting screws for the left bracket assembly. 5. The new power switch comes with a harness. Disconnect connector (P1) on the power cord between the power switch and the input power module. Replace the power switch/connector assembly only. 6. The switch is press fit to the front panel. To remove the switch, use a wideblade screwdriver to push each side of the protruded switch assembly out of the front panel. 7. Reconnect connector P1. 8. Reinstall the left bracket, front panel, access panel, and top cover assemblies. 5.4.3 Display panel removal and replacement Refer to the illustration in section 6.4.2. 1. If access to the display cable assembly is required, remove the RGM cover assembly. See section 5.3. 2. Remove the display panel from the RGM. • Press the Display Release push button. • Slide the display panel to the right until the alignment marks in the trap area and on the upper right side of the display line up. 3. Gently lift the display straight up and pull the bottom away from the chassis. 4. Reattach the display panel to the RGM. 5. • Coil the cord in the cord storage box. • Insert the top of the display, positioning the alignment marks at the top. • Gently push the bottom towards the chassis. Slide the display to the left until the latch clicks. 5.4.4 Display cable replacement Refer to the illustration in section 6.4.2. WARNING: Electrical shock hazard. Before cleaning or repairing the monitor, power it off and disconnect it from AC mains power. 1. Remove the display panel as described in section 5.4.3. 2. Remove the toroid enclosure on the left of the cord storage box (will pry off with a small screwdriver). 3. Press the connector latch release and remove the connector. 4. Remove the four mounting screws from the display panel. 5-5 5250 RGM Service Manual 5. Carefully separate the front and back portions of the display panel about an inch. Note the ribbon cables connecting the two sections. 6. Fold open the two halves of the display so the short ribbon cable is not strained. 7. Remove the screw connecting the cable shield wire to ground. 8. Remove the grommet for the display cable and remove the cable. 9. Replace with the new cable, making sure to replace the grommet and connector J4. 10. Carefully reassemble the front and back portions of the display panel. 11. Replace the four mounting screws and hardware from the display panel. 12. Reinstall the toroid enclosure by snapping it into position. 5.4.5 Display panel disassembly Refer to the illustration in section 6.4.2. Revision level Figure 5-3. Display panel disassembly WARNING: Electrical shock hazard. Before cleaning or repairing the monitor, power it off and disconnect it from AC mains power. 5-6 1. Turn off the power and disconnect the power cord. 2. Remove the display panel. (It is not necessary to disconnect the coil cord). 3. Remove the four mounting screws from the display panel. 4. Carefully separate the front and back portions of the display panel about an inch. Note the ribbon cables connecting the two sections. 5. Fold open the two halves of the display so the short ribbon cable is not strained. 5/Repair Procedures 6. Remove the speaker connector (J3) from the display processor board. 7. Carefully remove the ribbon cable connector (J3) from the display processor board by grasping the connector and pulling away from the board. 8. Carefully remove the ribbon cable connector (J1) from the display processor board by releasing the latch on the connector and pulling away from the board. 9. Carefully remove the ribbon cable connector (J2) from the display processor board by releasing the latch on the connector and pulling away from the board. 5.4.6 Display controller board replacement 1. Disassemble the display panel as described in section 5.4.5. 2. Remove all connectors from the display processor board. 3. Remove the mounting screws for the circuit board. 4. Remount the circuit board with the mounting screws. 5. Reconnect all connectors to the display processor board. 6. Reassemble the display panel as described in section 5.4.9 5.4.7 Display matrix panel replacement 1. Disassemble the display panel as described in section 5.4.5. 2. Using a #1 Phillips screwdriver, remove the four 4-40 screws retaining the leftand right-hand display brackets. 3. Remove the brackets and EL panel from the bezel. 4. Transfer the two mounting brackets to the new display matrix. 5. Replace the display matrix and mount it with the four bracket mounting screws. 6. Carefully replace the ribbon cable from the display matrix boards to the connector on the display processor board. 7. Reassemble the display panel as described in section 5.4.9 5.4.8 Display touch panel assembly replacement Refer to Section 6-4. 1. Disassemble the display panel as described in section 5.4.5. 2. Remove the speaker and associated foam tape from with the bezel. 3. Using a #1 Phillips screwdriver, remove the four 4-40 screws retaining the leftand right-hand display brackets. 4. Remove the brackets and EL panel from the bezel. 5. Remove the touch panel assembly from the front bezel. 6. Drop in the new touch panel replacement. Ensure that the two LEDs at the bottom of the panel align with the two holes in the bottom of the bezel. 5-7 5250 RGM Service Manual 7. View the display from the front for alignment within the bezel. Confirm that the short flex cable is oriented towards the top. 8. Reassemble the display brackets and screws. 9. Place the EL bracket (loose) assembly into the bezel and carefully tighten the four 4-40 screws. 10. Check the touch panel alignment from the front side of the bezel. 11. Slide the speaker into position using the foam tape as a wedge between it and the bracket. 12. Reassemble the display panel as described in the following section. 5.4.9 Display panel reassembly 1. Reconnect ribbon cable connector (J1) to the display processor board. 2. Reconnect ribbon cable connector (J2) to the display processor board. There should be an audible click. 3. Reconnect ribbon cable connector (J6) to the display processor board. There should be an audible click. 4. Reconnect the speaker connector (J3) to the display processor board. 5. Carefully reassemble the front and back portions of the display panel. 6. Replace the four mounting screws for the display panel. 5.5 Rear panel Refer to the illustration in section 6.5.2. 5.5.1 Fan replacement 5-8 1. Remove the RGM cover assembly. See section 5.3. 2. Remove the two top-mounting and two bottom-mounting screws and lock washers for the back panel and pull it away from the RGM slightly. 3. Disconnect the fan power connector. 4. Remove the two mounting nuts for the fan assembly. 5. Replace the fan assembly and mount it with the two mounting nuts. 6. Reconnect the fan power connector. 7. Place the back panel in position on the RGM and mount the top portion with two mounting screws. Install the two bottom-mounting screws, and then install the two bottom-mounting screws. 8. Replace the RGM cover assembly. See section 5.3. 5/Repair Procedures 5.5.2 Rear panel I/O board replacement 1. Remove the RGM cover assembly. See section 5.3. 2. Remove the two top-mounting and two bottom-mounting screws for the rear panel and pull it away slightly from the RGM. 3. Disconnect the fan power connector. 4. Disconnect the rear panel I/O board ribbon cables from J1 and J2 on the motherboard. 5. Remove the mounting screws and nuts, then remove the rear panel I/O board. 6. Replace and remount the I/O board. 7. Reconnect the rear panel I/O board ribbon cables to J1 and J2 on the motherboard. 8. Reconnect the fan power connector. 9. Place the rear panel in position on the RGM and mount the top portion with two mounting screws, and then install the two bottom-mounting screws. 10. Replace the RGM cover assembly, see section 5.3. 5.5 Photometer replacement Refer to illustrations in sections 6.8.2 and 6.8.3. Figure 5-4. Photometer replacement 1. To remove the RGM cover assembly, see section 5.3. 2. Remove the ribbon cable from the measurement board. 3. Remove the two red wires (connector J4) from the power supply board. 4. Remove the bracket screws retaining the photometer. 5. Remove the sample tubing from the photometer. 6. Install the new photometer onto the bracket. 7. Reinstall sample tubing to the sample chamber. 5-9 5250 RGM Service Manual 8. Reinstall the two red wires to the power supply board. 9. Install the ribbon cable into the measurement board. 10. Go to the Enter Calibration Constants screen of the service screens. 11. Use arrows to set the highlighted box to column “m” and the row to agent. 12. Unlock calibration constants (see section 3.2, Dip switch and jumper positions on the signal processor board section) and press the DEFAULT LINE option twice. 13. Repeat steps 11 and 12 for barometric pressure, CO2, and N20. 14. Cycle power on and off. The message RECALIBRATE BAROMETER should appear. 15. Return to the Enter Calibration Constants screen. 16. Set the highlighted box as before and press Enter. 17. Relock calibration constants and perform the calibration checks in section 3.6.2, Barometric pressure calibration, and 3.6.5, Gas analyzer check—zero and span calibration. 5.7 Toroid with power input module Refer to the illustration in section 6.10.2. 5.7.1 Toroid assembly replacement Figure 5-5. Toroid assembly removal 5-10 5/Repair Procedures Note: The toroid assembly includes the bracket, power input module, fuse block, and terminal block (and cable to front panel switch), which are replaced at the same time. Save all mounting hardware for installation of the new assembly. 1. Remove the RGM cover assembly. See section 5.3. 2. Remove the two top-mounting screws and two bottom-mounting screws from the back panel and disconnect the fan power connector. 3. Disconnect J1 from the power supply board. 4. Disconnect J3 from the power supply board. 5. Disconnect J4 from the power supply board. 6. Remove the remaining terminal strip wire connections for the power supply board. 7. Remove the two bottom-mounting screws from the power supply assembly bracket. 8. Remove the two power supply board bracket screws from the right side of the chassis. 9. Lift up and remove the power supply board assembly. 10. Remove the three mounting nuts for the power input module toroid assembly bracket. 11. Remove the two mounting nuts from the power input module. 12. Disconnect the front panel connector cable to the power switch cable. Feed the cable back through the unit. 13. Remove the ground wire connection from the module. A6-GND (Rear Panel) A5-P4 (Power Module) A5-E1 Figure 5-6. Ground wire connection 14. Remove the toroid assembly and power input module. 15. Install a new toroid assembly and power input module in the RGM. 16. Feed the cable to the power switches for the new assembly. Reconnect to the front panel switch cable. 17. Install the new power module on the back panel with the two mounting nuts and lock washers. 5-11 5250 RGM Service Manual 18. Connect the ground wire to the RGM chassis. See Figure 5-6. 19. Install the three mounting nuts securing the toroid bracket assembly to the RGM base. 20. Replace the power supply board assembly making sure all connections to the toroid terminal strip are correct. Reconnect J1, J3, J4, and the fan power connection. 21. Reinstall the back panel and secure the two top screws and two bottom screws. 22. Reinstall the back panel and secure the two top screws and two bottom screws. 23. Check out the RGM as described in section 3.1, Functional checkout, before returning to service. 5.7.2 Secondary fuse replacement Figure 5-7. Power supply fuse replacement 5-12 1. Disconnect the power cord from the RGM. 2. Check the power supply fuses in the fuse block next to the power supply circuit board. 3. Use a fuse puller to remove the old fuse. 4. Replace the fuse with the same type and rating. Note: 5 x 20 mm metric fuses are used. See section 6.10.1. 5/Repair Procedures 5.8 Circuit board removal and replacement 5.8.1 Power supply board assembly CAUTION: This circuit board is subject to damage by static electricity. Handle only at an approved static-control workstation. 1. Remove the RGM cover assembly. See section 5.3. 2. Disconnect the fan power connector. 3. Disconnect J1 from the power supply board. 4. Disconnect J3 from the power supply board. 5. Disconnect J4 from the power supply board. 6. Remove the remaining terminal strip wire connections for the power supply board. 7. Remove the two power supply assembly bracket mounting screws from the bottom of the RGM for the power supply assembly. 8. Remove the two power supply board screws from the right side of the chassis. 9. Lift up the power supply board assembly and remove. Figure 5-8. Power supply assembly removal/replacement 10. When replacing the power supply board assembly, make sure all connections to the terminal strip are correct. 11. Reconnect J1, J3, and J4. 5-13 5250 RGM Service Manual 5.8.2 Signal processor board CAUTION: This circuit board subject to damage by static electricity. Handle only at approved static-control workstation. 1. Enter the Service Mode screen and access the Enter Calibration Constants screen. Note: It is desirable to move the calibration constants from the old to the new board. This can only be done if the old board will communicate with the display. So, if there is no COMMUNICATIONS FAIL message, access the Enter Calibration Constants screen and write down the “m” and “b” columns for all channels except internal temperature. To reach the Enter Calibration Constants screen from the main screen, touch MENU, then SETUP, then SERVICE MODE, then “on,” and finally ENTER CAL. 2. Record the values from the columns titled m and b for the following parameters: m b Agent ------------------------- ------------------------- Barometric pressure ------------------------- ------------------------- CO2 carbon dioxide ------------------------- ------------------------- N2O nitrous oxide ------------------------- ------------------------- Internal temperature ------------------------- ------------------------- O2 internal oxygen ------------------------- ------------------------- O2 circuit oxygen ------------------------- ------------------------- PAW pressure airway ------------------------- ------------------------- Sample flow rate ------------------------- ------------------------- 3. Power off the RGM and disconnect the power cord. 4. Remove the signal processor board. 5. • Remove the RGM cover assembly. See Section 5.3. • Viewed from the front, the signal processor board is located at the far left side of the RGM. • Remove the two top-mounting screws for the circuit board. • Remove the ribbon cable from connector (J2) at the top of the circuit board by carefully opening the latches. • Carefully lift the signal processor board up and out of the RGM. Place the board on a static-controlled surface or in an anti-static package. Transfer software to the new processor board, or if required, install new software onto the signal processor board. Note: If installing new software on the signal processor board, also install new software in the display module (see the 5250 RGM Operation and Maintenance Manual). 5-14 5/Repair Procedures 6. Install the board into the RGM. • Reconnect J2 and reinstall the two mounting screws and lock washers for the board. • When replacing the top plate, ensure that the measurement board and SpO2 board align with the grooves in the hold-down block. Check that no tubing or ribbon cables are pinched under the plate. 7. Poweron the RGM, proceed to the Setup Screen. 8. Unlock the calibration constants by setting DIP switch #3 closed. Verify that the View All Set Up or main Service screen indicates calibration constants are unlocked. 9. Proceed to the Enter Constants Calibration screen. 10. Use the up or down arrows to move the cursor and enter the calibration values recorded from the old board (previous step 3). Enter new values for each line by touching Clear, typing the numbers and touching Enter. 11. After all values are entered, relock calibration constants and perform calibration checks in section 3.6.2, Barometric pressure calibration and section 3.6.5, Gas analyzer check—zero and span calibration. Note: For messages such as RECALIBRATE GAS ANALYZER or RECALI-BRATE FLOW, use the calibration procedures described in this manual. After repairs are complete, perform the checkout procedure as specified in the 5250 RGM Operation and Maintenance Manual and retest the RGM leakage current as described in section 3.13, Leakage current and ground resistance checks. Battery replacement Use this procedure if all calibration constants, configurations/setup choices, and time/date information are lost. Note: If the battery clip is loose, calibration constants can be lost. Verify that the clip tightly secures the battery.) 1. Disconnect all power to the RGM; then remove the signal processor board. 2. Connect a DVM between the terminal labeled B1 (+lead, which is located beneath the gray adhesive battery cover) and TP12 (Grd). 3. The battery must measure at least 3.0 V. If it is below 3.0 V, replace the battery, recalibrate the RGM, barometric pressure, PAW , flow, gases, and enter the proper time and date information. To obtain a new battery, order part number 6050-0001-459. After replacing the battery and calibrating the respective parameters, power off the RGM. Power on the RGM and verify that no RECALIBRATE messages appear. 5-15 5250 RGM Service Manual Measurement Board SpO2 Board Signal Processor Board Figure 5-9. Signal processor, SpO2, and measurement circuit board removal 5.8.3 SpO2 oximeter board CAUTION: This circuit board subject to damage by static electricity. Handle only at approved static-control workstation. Refer to Figure 5-9 and to the illustration in section 6.12.2. 5-16 1. Remove the RGM cover assembly. See section 5.3. 2. For easier access to the SpO2 board, remove the ribbon cable connector from the top of the signal processor circuit board by rotating the latches toward the side of the RGM. 3. Disconnect the SpO2 harness connector (J2) from the circuit board. 4. Carefully lift the SpO2 board up and out of the RGM. Place the board on a static-controlled surface or in an anti-static package. 5. Replace the SpO2 board. 6. Reconnect J2 for the SpO2 board 7. Reconnect J2 on the signal processor board. 5/Repair Procedures 5.8.4 Measurement board Refer to Figure 5-9. CAUTION: This circuit board is subject to damage by static electricity. Handle only at an approved static-control workstation. 1. Remove the RGM cover assembly. See section 5.3. 2. For easier access to the measurement board, remove the ribbon cable connector from the top of the signal processor board by rotating the latches toward the side of the RGM. 3. Remove the ribbon cable connector from the top of the measurement board by pushing the latches away from the connector. 4. Remove the plastic tubing from the pressure sensor on the measurement board. 5. Carefully lift the measurement board up and out of the RGM. Place the board on a static-controlled surface or in an anti-static package. 6. Replace the measurement board. 7. Reconnect the cable connector (X1) for the measurement board. 8. Reconnect J2 on the signal processor board. 9. After repairs are complete, perform the checkout procedure as specified in the 5250 RGM Operation and Maintenance Manual. 5.8.5 Motherboard Refer to Figure 5-9 and to the illustration in section 6.9.2. CAUTION: This circuit board subject to damage by static electricity. Handle only at approved static-control workstation. 1. Remove the RGM cover assembly. See Section 5.3. 2. Remove the two top-mounting and two bottom-mounting screws and lock washers for the rear panel and pull it slightly away from the RGM. 3. Disconnect the connectors (J1, J2) attached to the rear panel I/O board. 4. Disconnect the fan power connector. 5. Swing the rear panel assembly toward the left side where the ground lug connection is still intact. 6. Remove the signal processor board, SpO2 board, and the measurement board as described in earlier procedures. 7. Disconnect J6, J10, and J11 from the motherboard. 8. Remove the nine mounting screws from the motherboard. 9. Carefully slide the motherboard out of the back of the RGM. Place the board on a static-controlled surface or in an anti-static package. 10. Replace the motherboard using the mounting screws to mount the board. 11. Reconnect J6, J10, and J11. 5-17 5250 RGM Service Manual 12. Replace previously removed boards and reconnect all interconnections. Note: Switch SW-1 should match that of the old board. See Figure 3-2 for the proper switch position. 13. Reconnect the ribbon cables (J1, J2) from the rear panel I/O board. 5.9 Pneumatics system 5.9.1 Pneumatics tubing assembly removal and replacement Note: To assist you in installing the replacement tubing, carefully note or mark the tubing locations as you disconnect them, and mark each piece of removed tubing on the pneumatic diagram. Pneumatics Interface Board Figure 5-10. Pneumatics assembly replacement Refer to Figures 5-10 and the illustrations in section 6.7. Removing the pneumatics assembly 5-18 1. Remove the RGM cover assembly. See section 5.3. 2. Remove the ribbon cable connector (J1) from the top of the pneumatics interface board by rotating the latches upwards. 3. Disconnect the cable plugs at J2, J3, J4, J5, J6, and J7 from the circuit board. 4. Use a Phillips screwdriver to remove the two mounting screws from the pneumatics assembly from the base of the RGM. 5/Repair Procedures 5. Disconnect the tubing at the sample exhaust barb. 6. Carefully lift and support the pneumatics assembly about 2 in. above the base. 7. Remove the tubing from the two ports of the flow sensor. 8. Remove the tubing from the port of the PAW sensor. 9. Disconnect the tubing from the photometer measurement board barometric pressure transducer. To remove the access panel, remove the attachment screws on each side. 10. One at a time, replace the five pieces of tubing on the access panel (sample inlet, PAW, and three on the membrane separator) and discard the old tubing. 11. Disconnect and discard the tubing between the photometer and the auto zero valve. Disconnect the tubing at the NC barb on the auto zero valve. 12. Disconnect and discard the tubing between the purge valve and the O2 sensor housing. 13. Remove the pneumatics assembly from the unit. Replacing the pneumatics assembly 1. Using a Phillips screwdriver, loosen the cable clamp used to capture the auto zero valve. 2. Transfer the pneumatics interface board to the new bracket assembly. If not included with this kit, mount the board with the five retaining screws. 3. Returning to the main unit, remove and discard the tubing between the photometer and the O2 sensor housing. 4. Using the diagram provided with the pneumatics tubing kit, complete the pneumatic connections within the unit with the replacement tubing provided. • Replace the tubing from the scrubber inlet. • Unfasten the hook and loop tape strap and remove the expired scrubber. • Install the new scrubber cartridge by attaching the tubing and then fastening it in place with hook and loop tape. Note: Be sure to remove the cap or plug inserted in the top of the scrubber cartridge (inserted for shipping purposes only); otherwise, flow through the auto zero path will be obstructed. 5. When connecting the Nafion tubing between the photometer and the auto zero valve, wind the Nafion around the photometer before connecting to the inlet. 6. Reconnect the tubing to the PAW pressure, barometric pressure, and sample flow sensors if not previously installed. 7. Mount the assembly to the base of the RGM, taking special care not to crush or pinch the sump line tubing beneath the bracket. 8. Reconnect the cable plug connections at J2, J3, J4, J5, J6, J7, and J1 to the pneumatics board. 9. Ensure that no tubing protrudes into the vanes of the rear panel fan. 10. Perform the pneumatics system checkout. See section 3.5. 11. Replace the RGM cover assembly; see section 5.3. 5-19 5250 RGM Service Manual 5.9.2 Zero solenoid valve replacement 1. To remove the RGM cover assembly, see section 5.3. 2. Remove the three tubing connections for the air zero solenoid valve. See Figure 6.7.4 for reconnections. 3. Loosen the mounting screw and cable clamp for the zero solenoid valve enough to slide it forward. 4. Unplug J6 from the pneumatics board assembly. 5. Use a fine-blade screwdriver or other small tool to release pin latches from the body. Remove the old solenoid valve pins. 6. Insert the new solenoid valve into pins 5 and 6 of the connector body (polarity is not important) 7. Slide the new zero solenoid valve in the plastic bracket and tighten the mounting screw. 8. Reconnect the three tubing connections for the zero solenoid valve. 9. Perform the zero valve test in section 3.5, Pneumatic system checks. 5.9.3 Purge solenoid valve replacement 1. To remove the RGM cover assembly, see section 5.3. 2. Remove the two mounting screws for the pneumatics interface board from the bottom side of the RGM. 3. Carefully lift the pneumatics interface board up about 2 in. and rest it on the metal frame supporting the fuse terminal strip. 4. Remove the three tubing connections for the purge solenoid valve. 5. Remove the two mounting screws for the purge solenoid valve. 6. Unplug J6 from the PCA. 7. Use a fine-blade screwdriver to release the pin latches from the body. Remove the pins. 8. Install the new purge solenoid valve and mount it on the metal plate with the two mounting screws. 9. Insert the pins from the new solenoid valve into the connector body plug J6 pins 3 and 4 (pin polarity is not important). 10. Install connector J6 to the pneumatics interface board. 11. Reconnect the three tubing connections for the purge solenoid valve. 12. Mount the pneumatics board assembly to the base of the RGM with the two mounting screws. 13. Perform the purge valve test in section 3.5, Pneumatic system checks. 5.9.4 Flow restrictor replacement 5-20 1. To remove the RGM cover assembly, see section 5.3. 2. Note directional arrow on inline flow restrictor. 3. Disconnect tubing from both ends of flow restrictor 4. Install new restrictor with directional arrow oriented as noted in step one. 5/Repair Procedures 5.9.5 O2 solenoid valve replacement 1. To remove the RGM cover assembly, see section 5.3. 2. Note the three tubing connector positions and remove. 3. Use a Phillips screwdriver to remove the solenoid valve from the top of the O2 sensor housing. 4. Unplug the connector at J4 on the pneumatics interface board. 5. Use a fine-blade screwdriver to release the pin latches from the body. Remove the pins. 6. Insert the pins from the new solenoid valve into the connector body and install valve onto the O2 sensor housing. 7. Install connector pins 5 and 6 of J4 to the pneumatics interface board (pin polarity is not important). 8. Perform the O2 valve test in section 3.5, Pneumatic system checks. 5.9.6 Zero inlet kit replacement 1. To remove the RGM cover assembly, see section 5.3. 2. Cut the tie wrap that holds the zero filter in place. 3. Remove the filter and tubing from the zero valve. 4. Reconnect the new tubing and filter to the zero valve. 5. Secure the filter with the tie wrap from the kit. 5.9.7 Pump replacement 1. To remove the RGM cover assembly, see section 5.3. 2. Disconnect connector J3 from the pneumatics interface board. 3. Remove the two screws in the bottom that hold the pneumatic assembly and it will tilt back. 4. Remove the mounting screws for the pump cover. 5. Carefully remove the pump from the housing. 6. Note the tubing connections. Remove the 2 tubing connections from the pump. 7. Reconnect the two tubing connections to the new pump. Make sure the outlet connection on the pump is connected to the sample exhaust. The arrow points away from the pump toward the outlet connection. 8. Install the new pump and reconnect connector J3 to the pneumatics assembly. Note: Do not over tighten the screws—tighten only until the mounting foam (if present) is compressed approximately 30 to 50 percent. 9. Perform the pneumatic leak, pump leak, pump vacuum, and sample flow tests in section 3.5, Pneumatic system checks. 10. Replace the RGM cover assembly. 5-21 5250 RGM Service Manual 5.9.8 Scrubber replacement 1. Remove the tubing from the scrubber inlet. 2. Unfasten hook-and-loop tape strap and remove the expired scrubber. 3. Replace exhausted scrubber cartridge by attaching the tubing and then fastening in place with hook-and-loop tape. Note: Be sure to remove the cap or plug in the top of the scrubber cartridge (inserted for shipping purposes only); otherwise, this will obstruct flow through the auto zero path. 4. 5-22 Perform the zero valve test in section 3.5, Pneumatic system checks. 6/Illustrated Parts List Section Page 6.1 5250 main unit assembly...........................................................................................6-2 6.1.1 Parts list ..................................................................................................................6-2 6.1.2 Illustration .............................................................................................................6-3 6.2 Access panel assembly................................................................................................6-4 6.2.1 Parts list ..................................................................................................................6-4 6.2.2 Illustrations, front and back...............................................................6-5, 6-6 6.3 Display filter PCA...........................................................................................................6-7 6.3.1 Parts list ..................................................................................................................6-7 6.3.2 lllustration.............................................................................................................6-7 6.4 Display assembly............................................................................................................6-8 6.4.1 Parts list ..................................................................................................................6-8 6.4.2 Illustration .............................................................................................................6-8 6.5 Rear panel assembly ....................................................................................................6-9 6.5.1 Parts list ..................................................................................................................6-9 6.5.2 Illustration .............................................................................................................6-9 6.6 Signal processor board.............................................................................................6-10 6.6.1 Parts list ...............................................................................................................6-10 6.6.2 Illustration ..........................................................................................................6-10 6.7 Pneumatics assembly................................................................................................6-11 6.7.1 Parts list ...............................................................................................................6-11 6.7.2 Illustration ..........................................................................................................6-12 6.7.3 Pneumatics interface board parts list ................................................6-12 6.7.4 Pneumatics routing.......................................................................................6-13 6.8 Photometer and O2 sensor assembly...............................................................6-14 6.8.1 Parts list ...............................................................................................................6-14 6.8.2 ACX 200 illustration .....................................................................................6-14 6.9 Motherboard...................................................................................................................6-15 6.9.1 Parts list ...............................................................................................................6-15 6.9.2 Illustration ..........................................................................................................6-15 6.10 Toroid assembly ...........................................................................................................6-16 6.10.1 Parts list ...............................................................................................................6-16 6.10.2 Illustration ..........................................................................................................6-16 6.11 Water trap sensor assembly..................................................................................6-17 6.11.1 Parts list ...............................................................................................................6-17 6.11.2 Illustration ..........................................................................................................6-17 6.12 SpO2 (MINX) oximetry assembly ......................................................................6-18 6.12.1 Parts lists.............................................................................................................6-18 6.12.2 Illustration ..........................................................................................................6-18 6.13 Upgrade kits ...................................................................................................................6-18 6.14 Special service tools...................................................................................................6-18 6.15 Optional service parts and PCA Drawing Service Kit...........................6-19 6.16 Additional parts............................................................................................................6-19 6-1 5250 RGM Service Manual 6.1 5250 main unit assembly 6.1.1 Parts list Item 1 2 3 4 5 6 7 8 9 10 11 12 13 14 N/S N/S N/S N/S Part number Screws for top plate assembly, 6-32 x 0.375 SEMS (3) Cable assembly, power supply.......................................................6050-0001-111 Cable assembly, processor................................................................6050-0004-059 Cable assembly, pneumatic power..............................................6050-0001-112 Cable assembly, external O2............................................................ 6050-0004-060 Oximeter, PCA (MINX) .......................................................................0380-0800-075 Power supply assGembly...................................................................6050-0005-374 Signal processor board........................................................................6050-0001-439 Pneumatics interface board .............................................................6050-0001-442 Motherboard PCA ..................................................................................6050-0004-053 Photometer (ACX-200)..........................................................................6050-0002-639 Measurement board..............................................................................6050-0002-640 Toroid assembly, complete ...............................................................6050-0002-482 Top plate......................................................................................................6050-0001-078 Shroud, top.................................................................................................6050-0005-381 Support, foam...........................................................................................6050-0002-066 Hardware kit, fasteners .......................................................................6050-0005-383 Power cords: Commonly Used In: 6-2 Socket Type Australia, China 6030-0000-001 Canada, Japan Latin America, USA 0208-0943-300 Continental Europe 6030-0000-006 Italy 6030-0000-002 United Kingdom 6050-0002-259 6/Illustrated Parts List 6.1.2 5250 main unit assembly illustration 5 9 4 Barometric Pressure Transducer 2 1 3 PL 3 14 7 6 13 (beneath power supply) 8 12 10 11 6-3 5250 RGM Service Manual 6.2 Access panel assembly 6.2.1 Parts list Item 1 3 4 5 6 7 8 10 12 13 14 15 16 17 19 N/S N/S 21 22 23 24 6-4 Part number Access panel, with door, switch, release button...................6050-0005-379 Left bracket kit .........................................................................................6050-0005-372 Assembly, water separator ................................................................6050-0004-458 Door, access ..............................................................................................6050-0005-382 Cable assembly, power switch ......................................................6050-0001-573 Screw, SST PPH 6-32 x 0.375 Screw, SST PPH 4-40 x 0.44 Screw, SST PPH 4-40 x 0.375 Lever, Display Release arm..............................................................6050-0000-873 Button, Display Release push ..........................................................6050-0000-872 Spring, compression (Display Release)......................................6029-0000-038 Membrane filter, 5 pack ....................................................................6050-0001-379 Screw STLZ, 4-20 x 0.25 plastite Tube silicone 1/16 ID...........................................................................6026-0000-017 Tubing coex PVC/PE 0.043 ID.........................................................6026-0000-041 Label, SpO2................................................................................................6050-0005-169 Non SpO2....................................................................................................6050-0005-170 SpO2 filter PCA with cable ..............................................................6050-0005-376 SpO2 socket, white.................................................................................0380-0800-119 Luer, female with tubing pin ..........................................................6050-0001-597 Fitting, 1/8 barb to 1/8 barb.............................................................6050-0001-328 6/Illustrated Parts List 6.2.2 Access panel illustration, front 23 24 21 22 12 15 10 4 8 6 7 14 13 1 5 6-5 5250 RGM Service Manual 6.2.2 Access panel illustration, back Sample inlet tube connected at back of water separator. See view A-A Sample inlet PAW inlet Three full turns (four passes through the center) SpO2 Item 17 (REF) Item 31 (REF) SpO2 cable assembly installed .25 ± .067 .50 (REF) Detail "D" Scale: none VIEW B-B 17 19 Sample inlet tubing (installed) SpO2 cable assembly (installed) 16 VIEW A-A PAW inlet tubing (installed) 17 Water trap flow cable assembly (installed) Display release arm (installed) Compression spring (installed) Push button (installed) 6-6 6/Illustrated Parts List 6.3 Display filter PCA 6.3.1 Parts list Item Part number 1 Display filter PCA................................................................................... 6050-0002-011 2 Screw SST PPH 4-40 x 0.375 SEMS 6.3.2 Display filter PCA illustration 6-7 5250 RGM Service Manual 6.4 Display assembly 6.4.1 Parts list Item 1 2 3 4 5 6 7 8 9 10 11 12 N/S N/S N/S Part number Display assembly, complete.............................................................6050-0005-371 Label, display release...........................................................................6050-0001-169 Touch screen panel w/o filter, w/o EPROM ..........................6050-0001-463 Display cord, coiled with connector............................................6030-0000-005 Strain relief, display cable ................................................................6032-0000-040 Cover, EPROM Screw, 4-40 x 0.25 blk plastic (for EPROM cover) Replacement bezel kit (with labels)............................................6050-0005-390 Display matrix, EL panel....................................................................6050-0004-065 Bracket, 2 used.........................................................................................6050-0005-295 Screw, 4-40 x 0.375 SEMS (4) Display processor PCA ........................................................................6050-0005-375 Display rear cover kit with brackets (item 9) and hardware (item 10) ......................................................................6050-0005-380 Cable, display processor.....................................................................6050-0005-378 Display filter with gasket....................................................................6050-0004-605 Speaker assembly................................................................................... 6050-0005-377 6.4.2 Display assembly illustration 2 7 8 9 9 12 11 10 5 6 9 10 4 1 6-8 3 6/Illustrated Parts List 6.5 Rear panel assembly 6.5.1 Parts list Item 1 2 3 4 5 6 7 8 9 10 11 Part number Rear panel assembly, complete .....................................................6050-0005-373 Fan assembly ............................................................................................6050-0001-093 PCA rear panel I/O ...............................................................................6050-0004-066 Cover, Rear panel EPROM 5250 w/ hardware kit ..............6050-0005-680 Fitting, barb-to-barb...............................................................................6050-0001-328 Conn potential equilization..............................................................6027-0000-018 Tape foam 1/4-in. thick Nut SST #6-32 Nut M6 x 1 THRD nickel-plated Washer SST INT-LCK SCR SST PPH 4-40 x 0.375 SEM Nut, STLZ SST HEX 4-40 SM PATT 6.5.2 Rear panel illustration 6-9 5250 RGM Service Manual 6.6 Signal processor board 6.6.1 Parts list Item Part number Signal processor board .....................................................................................6050-0001-439 Battery, 3 V, lithium coin ...............................................................................6050-0001-459 6.6.2 Signal processor board illustration TP4 TP12 6050-0001-024 TP1 TP3 TP5 TP2 B1 TP7 TP13 TP8 TP6 TP9 TP14 TP20 TP15 TP11 TP10 TP16 TP17 TP19 6-10 TP18 6/Illustrated Parts List 6.7 Pneumatics assembly 6.7.1 Parts list Item 2 3 5 15 16 19 20 21 22 23 24 N/S N/S Part number Pneumatics assembly with PCA *................................................6050-0001-443 Pneumatics assembly without PCA *.........................................6050-0001-999 3-way auto zero valve assembly .....................................................6050-0002-435 3-way valve assembly (for purge or O2 valves).......................................................................6050-0001-115 Connector plug, 8-position (valve connector) .........................6006-0000-133 Vacuum pump w/connector ..........................................................6050-0003-068 (see illustration in section 6.74) Pump sound damping kit (w/tightening instructions.) (includes damping foam base, and pump foam base block)......................................................................6050-0003-531 Complete pneumatic tubing kit *.................................................6050-0001-664 (see illustration in section 6.74) Filter, 5-micron (hydrophilic-purge line) ..................................6015-0000-013 Filter, 0.45-micron hydrophobic.....................................................6015-0000-059 (Available in pneumatic tubing kits above) Filter, nebulizer.......................................................................................6015-0000-005 Restrictor, 0.007 in. (aqua) .................................................................6027-0000-079 Restrictor, 0.005 in. (lime green) ....................................................6027-0000-077 Pneumatics interface board .............................................................6050-0001-442 Auto zero pneumatic kit (tubing) (includes all items connected to NC port of auto zero valve in illustration in section 6.74).......................6050-0001-611 * The long Nafion tubing is not included in the tubing kits. It is not needed. 6-11 5250 RGM Service Manual 6.7.2 Pneumatics assembly illustration 6.7.3 Pneumatics interface board parts list Item Part number Pneumatics interface board complete.....................................................6050-0001-442 6-12 6/Illustrated Parts List 6.7.4 Pneumatics routing illustration 6-13 5250 RGM Service Manual 6.8 Photometer and O 2 sensor assembly 6.8.1 Parts list Item Part number 1 Bench, ACX-200, (without measurement board)..................6050-0002-639 Board, ACX-200, measurement board.........................................6050-0002-640 2 Screw, M4 x 0.38 .....................................................................................6019-0000-102 3 Fastener strap for scrubbers 4 Connector, 6-position............................................................................6006-0000-142 5 3-way valve assembly (O2 valve)....................................................6050-0001-115 6 O2 sensor housing assembly ............................................................ 6050-0001-095 7 Cable assembly internal O2 module...........................................6050-0001-102 8 Photometer/O2 replacement tubing kit.....................................6050-0001-588 9 Photometer power cable assembly 10 Cap, sensor housing (See note below)........................................6050-0001-129 N/S O2 sensor cartridge................................................................................0237-2034-700 N/S O-ring, small, internal O2 housing................................................6016-0000-039 N/S O-ring, large, internal O2 housing.................................................6016-0000-042 N/S Scrubber assembly, auto zero.........................................................6050-0001-387 Note: The PC board in the old cap must be transferred to the new cap. 6.8.2 ACX-200 assembly illustration 1 4 3 2 5 9 10 7 6-14 6 6/Illustrated Parts List 6.9 Motherboard 6.9.1 Parts list Item Part number Motherboard PCA................................................................................................6050-0004-053 6.9.2 Motherboard illustration JJ77 J8 J11 FF ON OFF O SW1 J9 J10 J2 PCA, MOTHERBOARD, CE, 5250 6050-0003-678 SERIAL NO. XXXXXX 6-15 J1 5250 RGM Service Manual 6.10 Toroid assembly 6.10.1 Parts list Item 2 3 4 5 6 7 Part number Toroid assembly complete ................................................................6050-0002-482 Terminal block.........................................................................................6015-0000-015 Fuse block...................................................................................................6015-0000-016 Receptacle, power..................................................................................6021-0000-010 Fuse, 5 x 20 mm, 5 A, slow (for 15V +8 V)................................6012-0000-011 Fuse, 5 x 20 mm, 2.5 A, slow (for +15 V)....................................6012-0000-010 Fuse, 5 x 20 mm, 0.25 A, slow (for VISO)..................................6012-0000-008 Components for AC power input receptacle (not shown) Fuse holder, 1c x c, 2 required ....................................................................0208-1034-300 Fuse holder, 5 x 20 mm, 2 required..........................................................0208-1036-300 Fuse 5 x 20 mm, 1.25 A, slow, 250 V (2 required) (for 220/240 V AC use) ........................................................................6012-0000-051 6.10.2 Toroid assembly illustration 6 5 7 6-16 6/Illustrated Parts List 6.11 Water trap sensor assembly 6.11.1 Parts list Item * N/S 1 2 3 4 5 6 7 * Part number Complete water trap sensor with membrane filter............6050-0001-322 Label for separator.................................................................................6050-0001-872 O-ring for bottle (2 required) ...........................................................6016-0000-032 Fitting, 0.054 (1 pin in back plate) ................................................6050-0001-315 Fitting, 0.038 (2 pins in back plate) ..............................................6050-0001-316 Water separator body (membrane)..............................................6050-0002-028 O-ring for 0.054 and 0.038 fittings (6 required)......................6027-0000-090 Fluid bottle.................................................................................................6050-0000-847 Sensor assembly (for membrane separator) ...........................6050-0001-321 Cartridge filter (package of 5)..........................................................6050-0001-379 Also order label for separator. 6.11.2 Water trap sensor assembly illustration 2 Places Note: Tighten screws to 2 in-lbs. torque 6-17 5250 RGM Service Manual 6.12 SpO 2 (MINX) oximetry board 6.12.1 Parts list Item Part number SpO2 (MINX) oximetry board.......................................................................0380-0800-075 SpO2 software, U26.............................................................................................0380-0800-126 6.12.2 SpO2 (MINX) oximetry board illustration J2 U24 Processor U26 EPROM J1 6.13 Upgrade kits SpO2 conversion kit: Used to add SpO2 monitoring capability to a 5250 RGM. Upgrade kit Part number SpO2 conversion kit............................................................................................6050-0001-220 6.14 Special service tools and PCA Drawing Service Kit Item Part number 5250 RGM PCA Drawing Service Kit.......................................................6050-0005-384 Extender card (32 x 3) .......................................................................................0237-2135-300 Flowmeter with restrictor ...............................................................................6024-0000-006 Restrictor only........................................................................................................6027-0000-126 Analog output connector .................................................................................6006-0000-184 DB9 serial loopback tester...............................................................................6050-0001-709 Pressure manometer..........................................................................................0380-1500-112 6-18 6/Illustrated Parts List 6.15 Optional service parts Item Part number TVX clip assembly, 16 ft ..................................................................................6050-0002-206 RS-232 cable, 78xx/7900 interface ..............................................................6050-0001-629 6.16 Additional parts Miscellaneous Item Part number Sample lines (package of 10).....................................................73318 or 6026-0000-009 Sample lines (package of 100) ................................................733188 or 6026-0000-037 ET tube adapter, Straight-T metal ISO (one per package).........................................6050-0000-478 Straight-T (package of 100).....................................................................6027-0000-073 Elbow (package of 100) ............................................................................6027-0000-072 Critical care adapters Package of 5....................................................................................................6027-0000-070 Package of 10 .................................................................................................6027-0000-059 Package of 100...............................................................................................6027-0000-071 Pediatric/neonatal adapters 3.0 mm (package of 10) ......................6027-0000-065 Replacement filters for critical care adapters (10-pack)................6027-0000-060 Pressure sensor line tee with 0.63 cm (0.125-in.) barbs .................6027-0000-005 Pressure-sensing tee adapter .........................................................................6050-0000-456 Tubing 0.63 cm (0.125-in.) ID x 244 cm (96 inches)..........................6026-0000-014 Scavenger adapter kit, 19-mm tee M/F with 0.63 cm (0.125-in.) barb and 306 cm (120 in.), 0.63 cm (0.125-in.) ID tubing......0237-2124-870 Cartridge, Filter (package of 5).....................................................................6050-0001-379 Package of 25 .................................................................................................6050-0001-772 Package of 50 .................................................................................................6050-0001-669 Nasal cannula CO2 sampling line (package of 10)...........................6002-0000-046 Return adapter kit, sample exhaust to patient circuit tee, 22 mm M/F with 0.63 cm (0.125-in.) barb ....................................6050-0000-002 5250 RGM Service Manual (English)........................................................6050-0005-264 5250 RGM Operation and Maintenance Manual, English...............................................................................................................6050-0005-158 French ...............................................................................................................6050-0005-266 German .............................................................................................................6050-0005-268 Italian .................................................................................................................6050-0005-270 Spanish .............................................................................................................6050-0005-272 O2 measurement Item Part number Patient circuit O2 sensor kit (cartridge not included) ....................0237-2030-700 O2 in-airway tee adapter, 22 mm................................................................6050-0001-222 O2 tee adapter, 22 mm......................................................................................0212-0763-100 Dome adapter kit..................................................................................................0236-0035-800 O2 sensor cartridge .............................................................................................0237-2034-700 6-19 5250 RGM Service Manual Pressure sampling Item Part number Pressure sensing installation kit for GMS absorber .........................0236-6152-870 Volume measurement Item Part number Flow transducer cartridge (package of 10)............................................0237-2228-870 Flow sensor clip assembly 244 cm (96 in.)...............................................................................................6050-0001-853 488 cm (192 in.) ............................................................................................6050-0002-206 Datex-Ohmeda calibration gases Item Part number Agent calibration gas canister with calibration bag........................6050-0005-392 Agent calibration gas canister: 6% CO2, 1.75% halocarbon-23 (CHF3), 50% O2, 40% N2O, 2.25% argon..............................................................................6016-0000-127 Agent calibration gas kits, bulk (6-pack) with bag ............................6050-0005-393 Calibration gas bag only ..................................................................................6050-0000-226 Misc., brackets, fuse kits, etc. Item Part number Display mounting bracket kit, Dovetail style, for Modulus II® Plus, and Excel.........................6050-0001-450 Pole style, 3/4 in. to 1/8 in. diameter range..................................6050-0001-642 Fuse kit, 100/120 V and 220/240 V.............................................................6050-0001-059 Auto zero sample scrubber............................................................................6050-0001-387 Water trap bottle with/label..........................................................................6050-0000-847 Water trap O-rings...............................................................................................6016-0000-032 Cable, RS-232 interface, 2 m (6 ft) ..............................................................6050-0001-629 Connector, analog output ................................................................................6006-0000-184 6-20 Appendix A/Material Safety Data Sheets (MSDS) A. O 2 sensor (page 1 of 3) SECTION I – GENERAL INFORMATION Trade Name: Company's Name: Company's Street: Company's City: Company's Country: Company's Emerg Ph #: Company's Info Ph #: Record No. For Safety Entry: Tot Safety Entries This Stk#: Status: Date MSDS Prepared: Safety Data Review Date: MSDS Serial Number: OXYGEN SENSOR C/S CITY TECHNOLOGY LIMITED WALTON RD PORTSMOUTH P061SZ UK 01-1705-325511 01-1705-325511 001 001 SMJ 01JUL89 07JUL95 BXYLV SECTION II – INGREDIENTS/IDENTITY INFORMATION Proprietary: Ingredient: Ingredient Sequence Number: NIOSH (RTECS) Number: CAS Number: OSHA PEL: ACGIH TLV: NO ACETIC ACID, POTASSIUM SALT; (4N SOLUTION POTASSIUM ACETATE), LEAD (Pb) PURE 01 AJ3325000 127-08-2 & 7439-92-1 N/K (FP N) 8 ML (MFR) SECTION III – PHYSICAL/CHEMICAL CHARACTERISTICS Appearance And Odor: Boiling Point: Vapor Pressure (MM Hg/70 F): Vapor Density (Air=1): Evaporation Rate And Ref: Solubility In Water: COLORLESS LIQUID, ODORLESS 230F,110C N/A N/A N/A MISCIBLE A-1 5250 RGM Service Manual A. O2 sensor (page 2 of 3) SECTION IV – FIRE AND EXPLOSION HAZARD DATA Flash Point: Lower Explosive Limit: Upper Explosive Limit: Extinguishing Media: Special Fire Fighting Proc: Unusual Fire And Expl Hazrds: N/A N/A N/A USE MEDIA SUITABLE FOR SURROUNDING FIRE (FP N). WEAR NIOSH/MSHA APPROVED SCBA & FULL PROTECTIVE EQUIPMENT (FP N). NONE. SECTION V – REACTIVITY DATA Stability: Cond To Avoid (Stability): Materials To Avoid: Hazardous Decomp Products: Hazardous Poly Occur: Conditions To Avoid (Poly): YES NONE. NONE KNOWN. NONE KNOWN. NO NOT RELEVANT. SECTION VI – HEALTH HAZARD DATA LD50-LC50 Mixture: Route Of Entry - Inhalation: Route Of Entry - Skin: Route Of Entry - Ingestion: Health Haz Acute And Chronic: Carcinogenicity - NTP: Carcinogenicity - IARC: Carcinogenicity - OSHA: Explanation Carcinogenicity: Signs/Symptoms Of Overexp: Med Cond Aggravated By Exp: Emergency/First Aid Proc: NONE SPECIFIED BY MANUFACTURER. NO NO NO ACUTE: NONE KNOWN. CHRONIC: NOT APPLICABLE. NO NO NO NOT RELEVANT. SEE HEALTH HAZARDS. NONE KNOWN. INHALATION: REMOVE TO FRESH AIR. SUPPORT BREATHING (GIVE OXYGEN OR ARTIFICIAL RESPIRATION) (FP N). EYES: WASH WITH WATER FOR AT LEAST 15 MINUTES. SKIN: WASH WITH WATER. INGESTION: DRINK WATER. SECTION VII – PRECAUTIONS FOR SAFE HANDLING AND USE Steps If Matl Released/Spill: Neutralizing Agent: Waste Disposal Method: Precautions-Handling/Storing: Other Precautions: A-2 KEEP IN PLASTIC BAG OR CONTAINER. NONE SPECIFIED BY MANUFACTURER. DISPOSE WITH OTHER NON-REUSABLE MATERIAL. DO NOT BURN OR INCINERATE. DISPOSE OF IN ACCORDANCE W/LOCAL, STATE & FEDERAL REGULATIONS (FP N). HANDLE WITH CARE TO AVOID CONTACT WITH ANY LEAKAGE. NONE. A/Material Safety Data Sheets A. O2 sensor (page 3 of 3) SECTION VIII – CONTROL MEASURES Respiratory Protection: Ventilation: Protective Gloves: Eye Protection: Other Protective Equipment: Work Hygienic Practices: Suppl. Safety & Health Data: USE NIOSH/MSHA APPROVED RESPIRATOR APPROPRIATE FOR EXPOSURE OF CONCERN (FP N). NONE SPECIFIED BY MANUFACTURER. IMPERVIOUS GLOVES (FP N). ANSI APPRVD CHEM WORKERS GOGGS (FP N). NONE NEEDED. EMERGENCY EYEWASH & DELUGE SHOWER MEETING ANSI DESIGN CRITERIA (FP N). NONE SPECIFIED BY MANUFACTURER. NONE SPECIFIED BY MANUFACTURER. SECTION IX – LABEL DATA Label Required: Technical Review Date: Label Date: Label Status: Common Name: Chronic Hazard: Signal Word: Acute Health Hazard-None: Contact Hazard-None: Fire Hazard-None: Reactivity Hazard-None: Special Hazard Precautions: Protect Eye: Protect Skin: Protect Respiratory: Label Name: Label Street: Label City: Label Country: Label Emergency Number: YES 07JUL95 07JUL95 G OXYGEN SENSOR C/S NO NONE X X X X ACUTE: NONE KNOWN. CHRONIC: NOT APPLICABLE. Y Y Y CITY TECHNOLOGY LIMITED WALTON RD PORTSMOUTH PO61SZ UK 01-1705-325511 The information set forth in this Material Safety Data Sheet is furnished free of charge for use by qualified employees of the user. All such information is furnished for the independent investigation and verification thereof by the user. NO GUARANTEE OR WARRANTY (INCLUDING MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE) OF ANY KIND IS MADE WITH RESPECT TO SUCH INFORMATION OR THE ACCURACY OR RELIABILITY THEREOF, OR WITH RESPECT TO THE PRODUCT COVERED BY SUCH INFORMATION. The manufacturer assumes no liability for any damages (whether incidental, consequential, special or otherwise) whatsoever arising out of or in connection with the use of such information or product, and all such use shall be at the user’s sole risk. A-3 5250 RGM Service Manual B. Agent cal gas (page 1 of 4) (1.75% halocarbon 23/freon, 6% carbon dioxide, 50% oxygen, and 40% nitrous oxide, balance Argon) SECTION I – IDENTIFICATION Product name: 4% Halocarbon 23, 6% Carbon Dioxide, 50% Oxygen, 40% Nitrous Oxide Scott Medical Products 4/3/91 Route 611 North, Plumsteadville, PA 18949 Plant manager at (215) 766-8861 Halocarbon 23/freon, Carbon Dioxide, Oxygen, Nitrous Oxide in Argon Gas mixture Supplier’s name: Address: Emergency phone no: Chemical name and synonyms: Chemical family: SECTION II – HAZARDOUS INGREDIENTS Material Halocarbon 23 Carbon Dioxide Oxygen Nitrous Oxide Argon Volume% 1.75 6 50 40 2.25 CAS No. 75-46-7 124-38-9 7782-44-7 10024-97-2 7440-37-1 OSHA PEL S/A 5000 ppm 90/91 NE S/A S/A ACGIH S/A 5000 NE S/A S/A MAC S/A 5000 NE S/A S/A Other STEL S/A 30000 NE S/A S/A S/A = Simple asphixiant NE = None established SECTION III – PHYSICAL DATA Appearance: Odor: Physical state: Vapor pressure: Vapor density (air = 1.0): Boiling point (C): Solubility in water % by weight: Appearance and odor: Specific gravity (H2O=1): Evaporation rate: Colorless Slightly sweetish taste and odor. Gas N/A 1.38 Gas 4.89 cm3/100 cm3 Colorless and odorless Gas Gas SECTION IV – FIRE AND EXPLOSION HAZARD DATA Flash point and method: Flammable limits: Extinguishing media: Special fire fighting procedures: Unusual fire and explosion hazards: A-4 Nonflammable LFL: N/A Nonflammable UFL: N/A Use what is appropriate for surrounding fire. Wear self-contained breathing apparatus and full protective clothing. Keep fire-exposed cylinders cool with water spray. If possible, stop the product flow. Cylinder rupture may occur under fire conditions. High oxidizing atmosphere increases the likelihood of fires. Oxygen vigorously supports combustion. A/Material Safety Data Sheets B. Agent cal gas (page 2 of 4) (1.75% halocarbon 23/freon, 6% carbon dioxide, 50% oxygen, and 40% nitrous oxide, balance Argon) SECTION V – REACTIVITY DATA Stability: Incompatibility (conditions to avoid): Materials to avoid: Hazardous decomposition: Hazardous polymerization: Stable under normal storage conditions. Avoid storage in poorly ventilated areas, or near a heat source. Avoid oils, greases, and flammable materials. Reacts violently with phosphine, hydrazine, ethers, alcohol’s hydrogen sulfide, and hydrocarbons. Carbon dioxide, being weakly acidic, reacts with alkaline materials to form carbonates and bicarbonates. Avoid contact with combustible materials or reducing agents. Nitrous oxide above 565 ºC decomposes to nitrogen and oxygen. None. Will not occur. SECTION VI – HEALTH HAZARD DATA Emergency Overview High pressure oxidizing gas Vigorously accelerates combustion Can increase respiration and heart rate. Can cause anesthetic effects. Routes of entry: Acute effects of overexposure: Chronic effects: Medical conditions aggravated by overexposure: Other effects of overexposure: Carcinogencity (U.S. only): Emergency and first aid procedures: Inhalation Exposure to carbon dioxide at 1-4% concentrations results in increased respiratory volume. Concentrations greater than 4% produce labored breathing and is dangerous for even a few minutes. Extended exposure to oxygen at higher pressures may be hazardous. A variety of central nervous system effects may result from breathing oxygen greater than 2 atm. Inhalation of small amounts may produce feelings of euphoria which may disguise sleepiness or loss of coordination associated with lack of oxygen. May cause anesthetic effects. Diminished mental alertness. None known. None known. None NTP - No IARC monographs - No OSHA regulated - No Remove to fresh air. Give oxygen and/or support breathing as required. Seek medical attention. Eye contact: none Skin contact: none Note to physician: none A-5 5250 RGM Service Manual B. Agent cal gas (page 3 of 4) (1.75% halocarbon 23/freon, 6% carbon dioxide, 50% oxygen, and 40% nitrous oxide, balance Argon) SECTION VII – SPILL OR LEAK PROCEDURE Clean up procedures: Evacuate and ventilate area. Remove leaking cylinder to exhaust hood or safe outdoor area. Shut off source if possible and remove source of heat. None. Specialized equipment: SECTION VIII – PERSONAL PROTECTION INFORMATION Engineering controls: Provide adequate general and local exhaust ventilation to maintain concentration below exposure limits. Safety glasses. None. Self-contained breathing apparatus in case of emergency or nonroutine use. Local exhaust recommended Safety shoes when handling cylinders. Eye protection: Skin protection: Respiratory protection: Ventilation: Other protective equipment: SECTION IX – SPECIAL PRECAUTIONS Precautions to be taken in handling: Precautions to be taken in storing: Secure cylinder when using to protect from falling. Use suitable hand truck to move cylinders. No part of cylinder should be exposed to temperatures above 52 ºC (125 ºF). Use only oxy-gen cleaned equipment. Avoid oils and greases. Oxygen reacts violently with hydrocarbons particularly at high pressure. Store in well ventilated areas away from combustibles. Keep valve protection cap on cylinders when not in use. SECTION X – TOXICOLOGICAL INFORMATION Lethal concentration: (LC50): Teratogenicity: Mutagenicity: None established. N/A N/A Lethal dose 50 (LD50): N/A Reproductive effects: N/A SECTION XI – ECOLOGICAL INFORMATION No adverse ecological effects are expected. SECTION XII – DISPOSAL CONSIDERATIONS Waste disposal method: A-6 Dispose of non-refillable cylinders in accordance with Federal, state, and local regulations. Allow gas to vent slowly to atmosphere in an unconfined area of exhaust hood. If cylinders are the refillable type, return cylinders to supplier with any valve outlet plugs or caps secured and valve protection caps in place. A/Material Safety Data Sheets B. Agent cal gas (page 4 of 4) (1.75% halocarbon 23/freon, 6% carbon dioxide, 50% oxygen, and 40% nitrous oxide, balance Argon) SECTION XIII – TRANSPORT INFORMATION Concentration: DOT description (U.S. only): 1.75 - 50% Proper shipping name: Compressed gas, oxidizing, N.O.S. Hazard class: 2.2 (non-flammable gas) Identification number: UN3156 Reportable quantities: none Labeling: Non-flammable gas Class 2, 1 0 Cylinders should be transported in an upright position in a well ventilated truck. ADR/RID (EU only): Special precautions: SECTION XIV – REGULATORY INFORMATION OSHA: TSCA: SARA: EU number: Number in annex 1 of DIR 67/548: EU classification: R: S: Process Safety Management: Materials are not listed in Appendix A of 29 CFR 1910.119 as highly hazardous chemicals. Materials are not listed in TSCA inventory. Threshold planning quantity for this mixture is 10,000 lbs. HN/A Mixture is not listed in annex 1. N/A SA, AS 9, 17A, 23 SECTION XV – OTHER INFORMATION Other precautions: Protect containers from physical damage. Do not deface cylinders or labels. Cylinders should be refilled by qualified producers of compressed gas. Shipment of a compressed gas cylinder which has not been filled by the owner or with his written consent is a violation of federal law (49 CFR). The information set forth in this Material Safety Data Sheet is furnished free of charge for use by qualified employees of the user. All such information is furnished for the independent investigation and verification thereof by the user. NO GUARANTEE OR WARRANTY (INCLUDING MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE) OF ANY KIND IS MADE WITH RESPECT TO SUCH INFORMATION OR THE ACCURACY OR RELIABILITY THEREOF, OR WITH RESPECT TO THE PRODUCT COVERED BY SUCH INFORMATION. Scott Medical Products assumes no liability for any damages (whether incidental, consequential, special or otherwise) whatsoever arising out of or in connection with the use of such information or product, and all such use shall be at the user’s sole risk. A-7 5250 RGM Service Manual Notes A-8 1315 West Century Drive Louisville CO 80027-9560 USA 1 303 666 7001 Corporate Office Datex-Ohmeda Division Instrumentarium Corp. PO Box 900 FIN-00031 Helsinki, Finland Tel 358 10 394 11 Fax 358 9 146 3310 Asia/Pacific China Datex-Ohmeda Pte. Ltd. Room B416, COFCO Plaza 8 Jianguomennei Avenue Beijing 100005, PR China Tel 86 10 6526 9773 Fax 86 10 6526 0653 Datex-Ohmeda Pte. Ltd. Room 1708, Yunlong Mansion No. 122 Luoguo Street Chengdu 610017, PR China Tel 86 28 661 4424 Fax 86 28 676 2703 Datex-Ohmeda Pte. Ltd. Room 1602, GIE Tower 403 Huan Shi Dong Road Guangzhou 510095, PR China Tel 86 20 8732 2521 Fax 86 20 8732 2518 Datex-Ohmeda Pte. Ltd. Room 2509, Lippo Plaza No. 222 Huaihai Road (M) Shanghai 200021, PR China Tel 86 21 5382 5657 Fax 86 21 5382 1619 Datex-Ohmeda Pte. Ltd. Room 809, Truroll Plaza Wusheng Road Wuhan 430033, PR China Tel 86 27 8571 2536 Fax 86 27 8571 2655 India Datex-Ohmeda (India) Pvt. Ltd. Block EP & GP, Sector V Plot XI-16, Salt Lake City Calcutta 700091, India Tel 91 33 357 4002 Fax 91 33 357 4001 Indonesia Datex-Ohmeda Pte. Ltd. Wisma Danamon Aetna Life 19th Fl. Jin. Jend Sudirman Kav. 45-46 Jakarta 12930, Indonesia Tel 62 21 575 0864 Fax 62 21 575 0865 Japan Vietnam United Kingdom Datex-Ohmeda K. K. TRC Annex 9F 6-1-1 Heiwajima, Ohta-ku Tokyo 143-0006, Japan Tel 81 3 5763 6801 Fax 81 3 5763 6838 Datex-Ohmeda Pte. Ltd. 522G Nguyen Tri Phuong St. Ho Chi Minh City, Dist. 10 Vietnam Tel 848 865 5875 Fax 848 862 5501 Datex-Ohmeda K. K. Technical Center TRC A Bldg. AE 4-8 6-1-1 Heiwajima, Ohta-ku Tokyo 143-0006, Japan Tel 81 3 5763 6850 Fax 81 3 5763 6852 Australia Datex-Ohmeda Ltd. Ohmeda House 71 Great North Road Hatfield Hertfordshire AL9 5EN England Tel 44 1707 263570 Fax 44 1707 260191 Korea Datex-Ohmeda Pte. Ltd. 10th Floor, Sam Sung Building 36–1, Yoido-Dong, Youngdeungpo-Ku Seoul, Korea Tel 82 2 786 7421 Fax 82 2 786 7420 Malaysia Datex-Ohmeda Pte. Ltd. Level 2 Bangunan O’Connor 13 Jalan 223 46100 Petaling Jaya Selangor, W. Malaysia Tel 60 3 754 7872 Fax 60 3 757 6948 Singapore Datex-Ohmeda Pte. Ltd. 152 Beach Road #12–05/07 Gateway East Singapore 189721 Tel 65 391 8618 Fax 65 291 6618 Taiwan & Philippines Datex-Ohmeda Pte. Ltd. 2nd Floor, No. 85 Chien-Kuo North Road, Sec. 2 Taipei, Taiwan Republic of China Tel 886-2 2515 0457 Fax 886-2 2501 9136 Thailand Datex-Ohmeda Pte. Ltd. 12th Floor (Unit F) Grand Amarin Tower 1550 New Petchburi Road Makasan, Rajathevi Bangkok 10320, Thailand Tel 66 2 207 1012/13 Fax 66 2 207 1014 Datex-Ohmeda Pty. Ltd. Units 1 & 2 149 Arthur Street PO Box 356 Homebush NSW 2140 Australia Tel 61 132 229 Fax 61 297 461796 Europe France Datex-Ohmeda S. A. S Parc de Pissaloup, BP 10 8 Avenue Jean d’Alembert F-78191 Trappes-Cedex, France Tel 33 1 30 68 60 00 Fax 33 1 30 68 60 01 Datex-Ohmeda S. A. S 17 rue Jean-Elysée Dupuy F-69410 Champagne au Mont d’Or, France Tel 33 1 30 68 60 00 Fax 33 4 78 43 26 58 Germany Datex-Ohmeda GmbH Auf der Hohe 49 Gewerbehof 49 D-47059 Duisburg, Germany Tel 49 2065 691 247 Fax 49 2065 691 255 Italy Datex-Ohmeda S. p. A. Via Cassanese 100 20090 Segrate Milan, Italy Tel 39 2 21693431 Fax 39 2 26926226 Netherlands Datex-Ohmeda B. V. Kantemarsweg 18 Post Box 22 3870 CA Hoevelaken Netherlands Tel 31 33 253 5404 Fax 31 33 253 7223 Spain Datex-Ohmeda S. L. C/Manuel Tovar 26 28034 Madrid, Spain Tel 34 1 334 26 00 Fax 34 1 358 12 84 Latin America, Caribbean Datex-Ohmeda Latin America 10685 N. Kendall Drive Miami FL 33176, USA Tel 1 305 273 9940 Fax 1 305 273 4382 Middle East Datex-Ohmeda Middle East Operations PO Box 5527 Dubai, United Arab Emirates Tel 97 14 822653 Fax 97 14 822659 North America Canada Datex-Ohmeda (Canada) Inc. 1093 Meyerside Drive, Unit 2 Mississauga, Ontario L5T 1J6 , Canada Tel 1 800 268 1472 Tel 1 905 565 8572 Fax 1 905 565 8592 United States Customer Service Datex-Ohmeda, Inc. Ohmeda Drive PO Box 7550 Madison WI 53707-7550, USA Tel 1 800 345 2700 Fax 1 608 221 4384 Technical Support Datex-Ohmeda, Inc. Three Highwood Drive Tewksbury MA 01876, USA Tel 1 800 345 2755 Equipment Service Center Datex-Ohmeda, Inc. 1315 West Century Drive Louisville Co 80027-9560, USA Tel 1 800 345 2755 Sales and Service Datex-Ohmeda, Inc. Three Highwood Drive Tewksbury MA 01876, USA Tel 1 800 635-6099 Fax 1 978 640 0469 1/00 (388/F)