Clin. Cardiol. 12, 412-414 (1989) Runaway Pacemaker: A Still Existing Complication and Therapeutic Guidelines H. MICKLEY, M.D., c. ANDERSEN, M.D., L.HEDEMANNNIELSEN, M.D. Departments of Cardiology B and Clinical Physiology, University Hospital, Odense, Denmark Summary: Runaway pacemaker is a rare, but still existing potential lethal complication in permanent pacemakers. Within 4 % years, we saw two cases of runaway pacemaker in patients with multiprogrammable, VVI pacemakers (Siemens-Elema, Model 668). In both cases a pacemaker-induced ventricular tachycardia (rate 240260 beatdmin) was documented. One patient died. Runaway pacemakers must be exchanged as soon as possible. Until this can be accomplished, different emergency maneuvers should be tried. As documented in the cases presented, placing a magnet over the pacemaker may result in a lower, more physiological pacing rate. Reprogramming the pulse generator to a lower output or the use of external chest wall overdrive stimulation may also be successful, but these procedures require the presence of an adequate escape rhythm. If this is not the case or the former maneuvers have failed, an external pacemaker may be connected to the permanent pacing lead. Thereafter, the lead can be safely cut. As an alternative, a temporary transvenous pacing lead may be established prior to disconnecting the permanent pacing lead. Key words: pacemaker failure, runaway pacemaker Introduction In a runaway pacemaker the pacing rate is inappropriately high due to battery failure or malfunction. In older Address for reprints: Hans Mickley, M.D. Depanment of Cardiology B University Hospital DK-5000 Odense C, Denmark Received: December 24, 1988 Accepted: April 1, 1989 permanent pacemakers pacing rates from 100 to 2 100/min have been described,'-3 and the result may be a pacemaker-induced tachycardia. Despite improved safeguards in design and evolution of protection or security circuits in modem pacemakers, the runaway phenomenon has not been eliminated. As the mortality rate of a runaway pacemaker-induced ventricular tachycardia is 30-40%,2-5 knowledge of this rare, but still existing complication should be widespread and immediate therapeutic guidelines are needed in order to improve the poor prognosis. Case Report 1 In 1979 an 81-year-old woman suffering from seconddegree sinoatrial block had a VVI pacemaker implanted. As the patient had syncope and battery depletion was approaching, the pulse generator was exchanged in May 1982 with a Siemens-Elema multiprogrammable VVI pacemaker, Model 668. Lead measurements revealed a chronic stimulation threshold at 3.4 V (pulse duration 0.5 ms) and the resistance was 522Cl. Because of the rather high stimulation threshold, output was programmed to 10 V. At follow up in August 1982, the rate was 70/min, stimulation threshold 3.1 V (pulse duration 0.72 ms), and output unchanged at 10 V. In December 1982, the patient was brought to the emergency room with clinical cardiac arrest. The ECG showed an irregular pacing rate of 210-240/min with 1: 1 ventricular capture (Fig. 1, top). DC defibrillation with 300 J did not affect the tachycardia. When a magnet was placed over the battery, the pacing rate fell to 70/min (Fig. 1, bottom), however resuscitation failed, mainly due to cerebral anoxia. Measurement of the pacemaker unit following postmortem explantation revealed an amplitude of 1.4-2.0 V and a runaway rate of 220-300/min, when connected to a load of 500 Q. Case Report 2 In December 1981, a 63-year-old man suffering from sinus node dysfunction with extreme bradycardia had an H . Mickley et a/.: Runaway pacemaker 413 Siemens3 FIG 1 (Top) ECG o n arrival showing a pacemaker-controlled tachycardia with ventncular rate up to 210-240/min (50 mm/s). (Bottom) After placing a magnet over the pulse genelator, the pacemaker rate declined to 701min (50 mm/s) .- , ~ ~# J - 2-’ ~ ” la&-*>-, ,,1 j /11/ /, , ,I,. .,,_ ~ ,r * . , ~ y , FIG 2 The ECG shows a pacemaker-controlled tachycardia wlth ventricular rate 260/min. sources, which allows a hermetically sealed case. Nevertheless, as documented in the present case reports, runaway pacemaker is still a serious complication, and the failing pacing unit must be removed as soon as possible. Unfortunately, literature gives only scant information on the emergency maneuvers which should be tried until surgical replacement of the defective pulse generator can be done. Treatment with antiarrhythmic agents or DC shock is ineffecti~e.~,’ A magnet placed over the pacemaker, as illustrated in both the cases presented, resulted in a more physiological pacing rate. Although one of the patients died, this 83-year-old woman had suffered from clinical cardiac arrest for several minutes, before the magnet maneuver was tried. Measurement of pacemaker output after explantation revealed an amplitude which should have been too low to stimulate the right ventricle. We have no explanation for this phenomenon, but the DC defibrillation might have damaged the pacemaker further.’ If a magnet fails to inhibit the pacemaker-induced t a ~ h y c a r d i a , ~ ,other ~ - ’ ~ emergency treatment should be tried. Reprogramming the pulse generator to lower outp ~ t , ~ .or I l partial inhibition of the pacemaker by external Discussion chest wall overdrive stimulation has been s u ~ c e s s f u l . ~ ~ ~ In older pacemakers runaway used to be the result of Both these methods require the presence of an adequate escape rhythm which may not always be the leakage of tissue fluid into the pulse g e n e r a t ~ r . ~ . ~ . spontaneous ~ However, the incidence of the runaway phenomenon has c a ~ e . ~Chest . ~ , wall ~ stimulation also demands that the declined, and the reason for this is generally believed to pulse generator is able to sense the external ~ t i m u l i . ~ he the incorporation of electrical safeguards built into the If emergency intervention still is not successful, and if newer pacemakers, added to the use of lithium power the patient is hernodynamically unstable, the permanent unipolar VVI pacemaker implanted. In February 1982, because of skeletal muscular stimulation due to a defective coating, the pacemaker was replaced by a SiemensElema, multiprogrammable VVI pacemaker, Model 668. In April 1987 the stimulation threshold was I .3 V (pulse duration 0.74 ms). In July 1987 the patient was admitted to the local hospital. The ECG showed a pacemakerinduced tachycardia 260/min with 1:1 ventricular capture (Fig. 2), but the patient was conscious with blood pressure 90/60 mmHg. A magnet placed over the pulse generator immediately reduced the pacing rate to a fixed rate at 70- 120lmin. The patient was hernodynamically stable and was transferred to a pacemaker center, where the pacing unit was immediately replaced. The electrode had no defects with stimulation threshold 1.9 V (pulse duration 1 .O ms), and a resistance of 815 9.Measurements of the failing pacemaker unit after explantation revealed fixed rate pacing at 70/min, output 4.6 V, and no runaway phenomenon. Revision of the pacemaker by the manufacturers revealed a “leak current” of the security circuit, and this was sufficient to induce the high pacing rate. 414 Clin. Cardiol. Vol. 12, July 1989 pacing lead should be connected to an external pacemaker, after which the lead can be safely cut.4.5.9Alternatively, a temporary transvenous pacing system should be established prior to cutting the permanent If equipment for transthoracic pacing is available,” this may also be applied when disconnecting the permanent pacing system. If the patient is hemodynamically stable, pacemaker exchange can be performed in a more ambient atmosphere, and if the procedure is done in a sterile manner, the permanent lead may be used again for implantation of a new permanent pacing unit.9 Recent reports illustrate that life-threatening arrhythmias may result from therapeutic inadiation of newer multiprogrammable pacemakers which employ complementary metal oxide semiconductors (CMOS). These appear to be more Sensitive to ionizing radiation than the bipolar semiconductors used in older pacemakers. I s Although runaway pacemaker is a rare complication in modern pacemakers, the phenomenon still occurs. Therefore, all physicians must be familiar with the emergency treatment protocol to follow when confronted with this potentially lethal complication. References 1 . Inoue H, Ueda K, Ohkawa S, Mifune J, Sugiura M: Runaway pacemaker. A case report with a runaway rate of 2100 ppm. PACE 2, 608 (1979) 2. Shettigar UR, Kleiger RE, Krone R, Geda AS: Runaway pacemaker with demonstration of supernormal period of excitation. Chest 65, 227 (1974) 3. Wallace WA, Abelmann WH, Normann JC: Runaway demand pacemaker. Report, in vitro reproduction, and review. Am Thoruc Surg 9 (3), 209 (1970) 4. Odabashian HC, Brown DF: “Runaway” in a modem generation pacemaker. PACE 2, 152 (1979) 5. Solow E, Bacharach B, Chung EK: Runaway pacemaker. Unpredictable pacemaker failure. Arch Intern Med 139, 1190 (1979) 6 . Bramowitz AD, Smith JW, Eber LM, Berens SC, Bilitch M, Grechko M: Runaway pacemaker: A persisting problem. J Am Med Assoc 228, 340 (1974) 7. Rubenstein JJ, Laforet EG: Pacemaker runaway following intermittent output failure. PACE 6, 645 (1983) 8. Bluhm G, Sundstedt C-D: Runaway pacemaker inhibited by external overdrive stimulation: A case report. PACE 7, 440 (1984) 9. Campo A, Nowak R, Magilligan D, Tomlanovich M: Runaway pacemaker. Ann Emer Med 12, 32 (1983) 10. Van Gelder LM, El Gamal MIH: Externally induced irreversible runaway pacemaker. PACE 4, 578 (1981) 1 1 . Zoll PM, Zoll RH, Falk RH, Clinton JE, Eitel DR, Antman EM: External non-invasive temporary cardiac pacing: Clinical trials. Circulation 71 (5) 937 (1985) 12. Katzenberg CA, Marcus F, Heusinkveld RS, Mammana RB: Pacemaker failure due to radiation therapy. PACE 5, 156 (1982) 13. k w i n AA, Surago CF, Schwade JG, Abitol AA, Margolis SC: Radiation induced failures of complementary metal oxide semiconductor containing pacemakers: A potentially lethal complication. Int J Rud Oncol Biol Phys 10, 1967 (1984) 14. Quertemous T, Megahy MS, Das Gupto DS, Griem MD: Pacemaker failure resulting from radiation damage. Rudiology 148, 257 (1983) 15. Ademec R , Haefliger JM, Killisch JP, Niederer J , Jaquet P: Damaging effect of therapeutic radiation on programmable pacemakers. PACE 5, 146 (1982)