AbstractID: 9701 Title: Comparison of Peripheral Dose Measurements to those

advertisement
AbstractID: 9701 Title: Comparison of Peripheral Dose Measurements to those
Calculated by the Peridose Software Program for 6 MV X-rays from a Varian 2100 EX
Accelerator
Peridose1,2 , a free software calculation program, assists users in approximating the dose
to critical organs at large distances from external beam treatment fields. Such an estimate
is very useful for instance when a pregnant patient or a patient with a pacemaker is being
considered for treatment. The software’s calculation is based on measured data and
largely takes into account X-ray energy, depth of critical organ, distance from the
treatment fields, field geometry, scatter from cerrobend blocks, scatter from the
collimator, wedge attenuation and scatter, head leakage, and couch attenuation. The
program distinguishes between treatments involving either orthogonal or tangential
beams. We have tested the software for distances up to 40 cm from a 6 MV, 10x10 cm of
a Varian 2100EX accelerator and depths up to 12 cm. Calculated results agreed with
measured results to within 30% at distances greater than 10 cm from an open beam. This
agreement is well within the 60% (two standard deviations) accuracy claimed by the
author of the software.
1. van der Giessen, Radiotherapy and Oncology: 58, 2, (2001), pages 209-213
2. Peridose is currently available from the Medical Physics Software Exchange list
maintained by Carter Schroy.
Download