AbstractID: 7107 Title: A direct method to measure scatter and primary component in exit megavoltage radiotherapy beams Scatter from the patient in the megavoltage treatment beam exiting the patient is difficult to eliminate. It introduces artifacts in megavoltage CT reconstruction. Also, knowledge of the scatter component in the exit beam is required in order to perform accurate dose reconstruction, which is a process whereby the dose into a patient CT representation at the time of treatment is inferred from the exit dose detected. We developed two complementary methods to measure the scatter and primary photon dose at the detector. One or several MLC leaves are used to block a portion of a fan (slit) beam emitted from a linac. A constant thickness solid water phantom placed at the isocenter is used to scatter photons towards a detector. Thus, the photons scattered into the detector behind the closed leaf can be directly measured. By changing the position of the closed leaves, we can measure the profile of scatter at the detector. In the other method, only one or several leaves are opened. The primary photon dose is measured directly and the scatter is obtained by subtracting the primary from an exposure with all leaves open. The work will be used to reduce artifacts in megavoltage CT reconstruction and to verify the process of dose reconstruction. These methods were tested on two different radiotherapy systems: a Varian linac and MLC system with film used to measure scatter; and a 4-MV Orion linac with a NOMOS-MLC and xenon detector. On each system, both methods for estimating scatter exhibited similar distributions.