The X-Ray Microscopy Facility Project at the ESRF Jean Susini and Ray Barrett European Synchrotron Radiation Facility, BP220, F-38043 Grenoble cedex, France Abstract. A beamline dedicated to X-ray imaging and spectro-microscopy in the 0.26.0keV energy range is under construction at the ESRF. This beamline is installed on a low beta straight section equipped with three phaseable undulators and will consist of two branch lines: the first, a scanning microscope including various contrast modes; the second, an imaging full-field microscope using Zernike phase contrast. The scanning microscope will be equipped with 2 fixed-exit high resolution monochromators (crystal/multilayer monochromator and plane grating monochromator). Both microscopes will use high resolution zone plates as focusing elements. The design of this beamline is discussed with emphasis on the optical layout and on heat load management. 1 Introduction Third generation synchrotron sources produce a beam of unprecedented quality: the extremely low emittance coupled with high brilliance together with the versatility of new insertion devices, offer the capability to control brightness, spectrum, polarisation, coherence and size of the beam. This means that X-ray microscopy techniques which have been extensively used in the soft X-ray region [1] can now be extended, with the anticipation of very high performance to higher photon energies. This will enable new investigations: study of thicker specimens, access to K absorption edges of elements of major interest in the biological and materials sciences, in particular from Potassium to Chromium, access to M and L edges of heavy metals (i.e. Au or Ag) for specimen labelling, and the use of X-ray fluorescence for trace element mapping. To exploit these characteristics an X-ray microscopy beamline is being built at the European Synchrotron Radiation Facility, whose source properties offer the required coherence and brilliance [2]. Approved in 1993, the project started at the end of 1994, the first beam will be delivered in December 1996 and the beamline will be open to external users by the end of 1997. Scanning and imaging microscopes will be built on two independent branch lines. In parallel, collaboration contracts have been placed between the ESRF and two European laboratories for the fabrication of high energy zone plates [3–6]. After a brief overview of the ESRF source properties, the conceptual design of the X-ray microscopy beamline will be presented. The problem of building a soft X-ray beamline which should cover an energy range (0.3–6 keV) on a high energy machine will be addressed. Finally, the optical layout will be described in more detail. 2 The Source The ESRF is a high energy machine with an electron energy of 6.02 GeV and -9 -11 emittances (1% coupling) of εh = 4 x 10 mrad (horizontal) and εv= 4 x 10 mrad I - 46 J. Susini and R. Barrett (vertical). Currently, two types of straight sections, low or high β, are available. The considerations taken into account in the choice of the source type are summarised in Table 1. Table 1. Considerations involved in choice of the straight section. High β Straight section Low β Optical functions β =27.0m, β =13.3m h v β =0.58m, β =3.14m h v Source parameters σ =330µm, σ =23µm h v σ’ =12.1µrad,σ’ =7.5µrad h v - Total power can be limited by the use of an aperture. - Symmetric beam - Requires small optics. σ =59µm, σ =11µm h v σ’ =83µrad, σ’ =8.2µrad h v - Better phase-space matching - Wider beam @ 30m ∅lower power density Pros ∅possibility for simultaneous operation of the two branch lines - Asymmetric beam - Requires longer horiz. deflecting optics - Phase-space mismatch - High power density Cons 5 10 19 On-axis spetral brilliance (ph/s/mm 2 /mrad 2 /0.1%bw) N = 20 4 10 K=4 2100W 19 N = 60 K=1.5 887W K=2.0 1580W K=1.0 394W K=1.8 n=3 K=0.8 252W K=1.5 n=3 3 10 19 K=5.5 4000W K=0.6 142W 2 10 19 x5 K=0.4 63W x5 1 10 19 x5 x5 0 0 1000 2000 3000 4000 5000 6000 7000 Energy (eV) Fig. 1. On-axis spectral brightness vs. energy for various K values. The integrated total power emitted by the IDs is given. The X-Ray Microscopy Facility Project at the ESRF I - 47 A number of imaging experiments require at least partially coherent illumination. Therefore, the main criterion for the choice of the source is the flux per coherent phase-space volume which can also be expressed as the number of wave modes contained in the photon beam. The low beta straight section, ID21, of the ESRF storage ring has been chosen as the source best suited for our applications due to the higher phase-space density (proportional to λ/σ.σ’). The principle disadvantage of this source is the large horizontal beam divergence which gives rise to a very flat elliptical beam shape which does not match the spherical acceptance of the zoneplate. The ID21 straight section is 4.8m long and is equipped with three identical 1.6m long, 80mm period, phaseable insertion devices (ID) allowing the number of poles to be changed between 20, 40 and 60. The first harmonic is tuneable from 0.2 to greater than 4.5 keV corresponding to a gap range from 15 to 80mm. Due to the flexibility of the ESRF strategy of constructing undulators in three independent sections, the total emitted power can be maintained at a level which can be easily handled by the first mirror without affecting significantly the coherent brilliance : a 4.8m device would be used for the highest energy experiments, while a single 1.6m ID would be used for soft X-ray experiments. In diffraction limited imaging conditions, the factor 9 in the peak brilliance lost by the fact than only 1 ID is used is largely compensated by the effect of the lateral coherence width increase with wavelength at a given distance. In all cases, the total power through an aperture of 5x5 mm2 placed at 28m from the source is always maintained below 700W. 3 Optical Layout The main design features of the optical layout of the beamline are : - Preservation of the high coherent brightness of the source. - Scanning (STXM) and imaging (TXM) microscopes are to be built on two independent branch lines which could be used simultaneously : the STXM requires only a fraction of the beam (corresponding to a few wave modes) with no loss of performance, the remaining being used by the TXM. Wide bandpass, horizontally deflecting, X-ray multilayers are used to steer the beam on the TXM branch line. The Bragg angle of 4 degrees allows a large physical separation between the two endstations. - On the STXM branch, a small pinhole aperture (5µm to 50µm) will be used as a secondary source, set at about 1m from the zone-plate. This aperture and the microscope are mechanically linked to the same support in order to minimise relative movements (mechanical vibrations) between the two components. Even in an experiment requiring spatial coherence, the optimum amount of phasespace to accept is rather more than a single mode and involves a trade-off of high coherence for diffraction limited focusing- against flux -which is required for observation of weak contrast phenomena and fluorescence experiments. Moreover, an overfilling of the aperture used as secondary source and the zoneplates allows the effects of beam instability to be minimised. The aperture size and the distance between the pinhole and the microscope are variable to accommodate a wide variety of microscopy, spectro-microscopy and related coherence experiments over a large energy range. I - 48 - J. Susini and R. Barrett The beamline will be operated in windowless mode in order to maximise the photon flux at the lower energies and to minimise the degradation of the inherent coherent brightness of the source. 3.1 Total Power Management by Use of two X-Ray Mirrors It is important to limit the heat load on those optical components which strongly influence the energy resolution, namely the focusing mirrors and the two monochromators. Consequently, a 2-bounce horizontally deflecting mirror device consisting of two parallel silicon mirrors is used for : i) Power filtering: 90% of the unwanted part of the spectrum is absorbed by the first mirror. The maximum integrated power is damped from 700W to less than 50W after two reflections. ii) Suppression of the higher-order harmonics of the insertion device: the cutoff angle can be tuned from 5 to 20 mrad allowing harmonic rejections better than 10-3 for any energy ranging from 1 to 6 keV. At these small grazing angles the position of the exit beam is effectively fixed. iii) Separation of the bremsstrahlung radiation from the synchrotron radiation: the combination of the mirror device with a bremsstrahlung stop and a collimator allows a tremendous reduction of the shielding required for the downstream beam transport. If the “pink” beam is efficiently “cleaned” and collimated there is no shielding required downstream of the collimator. This strategy has not only an obvious financial impact upon the beamline design cost but also allows relatively free access to the microscope during the experiments. Furthermore, horizontally deflecting grazing incidence mirrors have several advantages : i) preservation of the vertical emittance of the source which determines both the degree of coherence of the photon beam and the energy resolution of the downstream monochromators. ii) The photon beam in the horizontal plane is very wide. Therefore, the heat load along the mirror length is much more homogeneous than it would be in the case of a vertically deflecting mirror and, the thermal gradient along the mirror length is minimised. The two first mirrors are side water cooled. A gallium-indium layer between the mirror and the copper block provides both a good thermal contact and an efficient mechanical decoupling. The first mirror which removes much of the unwanted power, is 130 mm thick while the mirror is cooled along a 10 mm strip on the top. Finite-element thermo-mechanical simulations show that this cooling geometry allows the thermal deformation to be kept below 10 µrad rms for a maximum absorbed power of 700 W. The X-Ray Microscopy Facility Project at the ESRF I - 49 3.2 Layout of the STXM Branchline The optical design was influenced by several constraints: For spectroscopic applications the very wide energy range cannot be covered with a single device, therefore 2 high resolution monochromators are planned. It is intended that the operations necessary to change between these two basic operating modes (see Fig. 2) should be automated as much as possible and allow imaging over the full spectral range to be achieved for the same sample with a minimum of manual realignment. Fig. 2. Various optical configurations of the beamline I - 50 J. Susini and R. Barrett 3.2.1 Crystal Monochromator (2.0–6.0 keV) A fixed exit 2-crystal monochromator which can be equipped with Silicon and multilayers will be used in the 2–6 keV range (Fig. 3). The Bragg angle is tuneable -4 -2 from 3 to 70 degrees. The energy resolution is about +5.10 for Silicon and 10 – -3 10 for multilayers. The optics will be water cooled. In order to ensure a good The X-Ray Microscopy Facility Project at the ESRF I - 51 stability of the outgoing beam during an energy scan, the monochromator is positioned as close as possible to the entrance aperture of the microscope (about 1 m). In this operation mode the source is imaged onto the pin-hole plane via a sagitally focusing cylindrical mirror in the horizontal plane and a vertically focusing spherical mirror. 3.2.2 Plane Grating Monochromator (0.2–1.5 keV) The grating monochromator design (see Fig. 4) associates plane gratings, which are tuned by a simple rotation, with a vertically focusing spherical mirror (VFM). In this design it is the source (conjugated by the VFM) which acts as the entrance slit of the monochromator. The pinhole apertures for the STXM act as the exit “slits” of the grating monochromator and are fixed at 8.23m from the VFM. In the horizontal plane, a sagitally focusing mirror (HFM) is used to image the source onto the same pinhole apertures. Three holographic gratings have been optimised to cover the full energy range [7]. In order to achieve a good overall efficiency, in particular, at high energies, the deviation angle has been set to 6 degrees. Assuming slope errors of the order of 1µrad on each of the vertically deflecting components the average calculated resolving power of the PGM should be around 6000 over the operating energy range. 3.2.3 Stigmatic Design The exit pinhole of the grating monochromator, used as secondary source for the scanning microscope, being fixed over the full energy range, the different focusing mirrors must have the correct radius of curvature. This is particularly critical because the short distance between the zone-plate and the source (pinholes) makes the microscope performance very sensitive to astigmatism [8]. The maximum tolerable longitudinal separation between the vertical and horizontal sources (corresponding to the foci of the focusing mirrors) is only 10cm. Therefore, the vertical focusing mirror is bendable (bimorph mirror) in order to match its image plane exactly with those of the horizontally focusing mirrors. A beam position monitor will be set at the secondary source plane [9]. 3.2.4 The Scanning Microscope The STXM is conceived to work with zone plate type optics which currently offer the best proven performance for this type of application. The development of high resolution zone plates capable of efficiently focusing X-rays up to 6keV has been addressed in two different collaborative research programs established with laboratories at King’s College London and Göttingen [3-6]. Clearly a single zone plate will be incapable of providing efficient focusing over the entire spectral range available and one of the challenging aspects of the microscope design has been to facilitate the remote exchange of the zone plate allowing close to optimum focusing conditions regardless of the operating energy. The X-ray microscope will be placed -4 -6 in an environmental chamber allowing operation in air, helium or vacuum (10 –10 mbar). The entire microscope can be moved along the beam axis relative to the fixed exit pinhole which acts as the secondary source. This movement allows the pinhole– zone plate distance to be varied from 0.3 to 1.5 m and allows the illumination conditions of the zone plate to be adapted to the experiment. Taking into account the mirror reflectivities, monochromator band pass and undulator characteristics the 9 7 estimated photon flux in a 50nm probe is of the order of 10 –10 ph/s. I - 52 J. Susini and R. Barrett Fig. 4. Parameters of the plane grating monochromator and associated focusing mirrors The X-Ray Microscopy Facility Project at the ESRF I - 53 The wide spectral operating range of the microscope is attractive for spectromicroscopy. Whilst in its simplest form this might consist of taking multiple images of a single sample region at different incident energies, an interesting extension is to perform highly spatially resolved XAS scans on small regions of the sample. The spatial resolution of this mode is potentially limited by the probe size, convergence and the sample thickness, but requires careful mechanical design due to the energy dependence of the zone plate focal length. The sample will be scanned using a combination of piezo driven flexure and mecha2 nical stages giving a total scan area of 10 x 10 mm . The current aim is to approach pixel rates of 1 kHz. A manual sample rotation will be available, primarily for fluorescence mode imaging, but with the possibility of future upgrade to motorised movement for micro-tomography measurements The microscope design is intended to offer maximum flexibility for the use of various different detector types. Currently it is planned for absorption measurements to use alternatively, proportional gas detectors, PIN photodiodes and avalanche photodiodes. A high energy resolution Germanium solid state detector will also be available for fluorescence measurements. A standard sample holder has been designed upon similar principles to that currently used on the full field imaging microscope at the ALS [11]. This kinematically mounted system should allow regions of interest to be identified and recorded on a standard light microscope prior to transfer into the X-ray microscope and rapidly aligned to the probe scan. It is intended for the same holder to be used on the TXM allowing rapid transfer between the two microscopes. The microscope will be controlled using essentially standard ESRF VME based electronics running OS9 with a user interface running on Unix workstations. 3.4 Layout of the TXM Branchline The main difficulty of the implementation of a TXM branchline using the ID21 source is associated with the relatively small emittance and the relatively large energy range. The consequent problems of creating a suitable condenser system for the microscope are discussed elsewhere [10]. Zone plate optics will be used for the objective lens with similar characteristics to those planned for use in the scanning microscope. It is probable however, for reasons of imaging dose, that a stronger emphasis will need to be placed upon the efficiency of these zone plates, potentially at the expense of resolution. It is planned initially to use a channel cut Si 111 monochromator for this branch thus allowing narrow bandpass imaging to be performed. The microscope will enable absorption contrast (bright field), dark field and Zernicke phase contrast imaging to be performed. The estimated flux at the detector plane for bright field imaging with a Si 111 monochromator is of the order 10 of 10 ph/s. Cryo-cooling is foreseen to reduce the effects of beam damage and thus permit multiple imaging of dose sensitive samples such as required in tomographic measurements. It is anticipated that the primary detectors to be used will be CCD cameras. It is probable that in the longer term two such detectors will be available depending upon the imaging energy to be used. In the case of the lower energies this will probably be a back-thinned CCD using direct conversion whilst for higher energies (to preserve dynamic range) a phosphor coupled system is envisaged. 4 Conclusions The conceptual design of the X-ray microscopy facility at the ESRF has been described. Priority has been given to the spatial resolution (100–50nm), resolving I - 54 J. Susini and R. Barrett 9 power of the monochromators (E/∆E §±IOX[DWWKHVDPSOHSRVLWLRQ – 7 10 ph/s in a probe of 50nm diameter for the STXM) and flexibility (trade-off of energy resolution for flux, multiple contrast modes, two microscopes). The beamline will exploit the potential of the imaging applications in the 2–6 keV energy region which are, so far, relatively unexplored. In particular, the access to the K and L edges and emission lines of the medium elements will offer new capabilities: higher penetration depth, fluorescence contrast. For many samples, cryo-microscopy will be necessary to cope with the very high flux and the related radiation damage. Acknowledgements The authors are very grateful to R. Baker, G. Marot and R. Guetta for their for their very active involvement in the design of the beamline components, E. Delcamp and F. Polack for the optimisation of the parameters of the grating monochromator. We thank C. Buckley and B. Niemann for a number of very constructive discussions. References 1 2 3 4 5 6 7 8 9 10 11 J. Kirz, C. Jacobsen, and M. Howells, Soft X-Ray Microscopes and their Biological Applications, Quarterly Reviews of Biophysics, 28 (1), (1995). M.R. Howells, R. Burge, C.J. Buckley, A. Miller, G. Morrison, D. Rudolph, G. Schmahl, J. Thieme and M. Vollbrecht, Conceptual design for an X-ray microscopy facility at the ESRF, ESRF Technical Report, (1992). B. Kaulich, Nanostructures with high aspect ratios for phase zone plates for a phase contrast X-ray microscope at ESRF for wavelengths around 0.3nm, Proceedings of X-ray Microscopy IV, editors A.I. Erko and V.V. Aristov, Chernogolovka (1994). B. Kaulich, Phase zone plates for X-ray microscopy at the ESRF, this conference. P. S. Charalambous and A. Firsov, Optimization of the process parameters for the fabrication of high resolution diffraction optical elements, Proceedings of X-ray Microscopy IV, editors A.I. Erko and V.V. Aristov, Chernogolovka (1994). P. Charalambous and R.E. Burge; Zone plate fabrication at King’s College, London, this conference. E. Delcamp, B. Lagarde, F. Lagarde and J. Susini, Optimization software for X-UV monochromators, Proceedings SPIE conference, 2856, Denver (1996). McNulty, A. Khounsary, Y.P. Feng, Y. Qian, J. Barraza, C. Benseon, and S. Shu, A beamline for 1-4 keV microscopy and coherence experiments at the Advanced Photon Source, Rev. Sci. Instrum., 66 (9), September 1995. G.S. Dermody, C.J. Buckley, J. Susini, and R. Barrett, Design considerations for a prototype beam position monitor for the X-ray microscopy beamline ID21 at the ESRF; this conference. W. Meyer-Ilse, X-ray microscopy in Berkeley, this conference. S. Oestreich and B. Niemann, Design of a condenser for an X-ray microscope on a low-β section undulator at the ESRF, this conference.