Melissa C. Martin, M.S., FAAPM, FACR
AAPM Spring Clinical Meeting
March 20, 2012 Dallas, TX
Phantom Submission
Slide provided by Doug Pfeiffer
Phantom Submission
Slide provided by Doug Pfeiffer
Phantom Submission
Slide provided by Doug Pfeiffer
Phantom Site Scanning Form -
Page 1
Slide provided by Doug Pfeiffer
Phantom Site Scanning Data
Form - Page 2
Slide provided by Doug Pfeiffer
Phantom Images
Slide provided by Doug Pfeiffer
Phantom Images
Slide provided by Doug Pfeiffer
Phantom Images
Slide provided by Doug Pfeiffer
Phantom Scoring
Slide provided by Doug Pfeiffer
CT # Accuracy
Slide provided by Doug Pfeiffer
Contrast-Noise Ratio
Slide provided by Doug Pfeiffer
Contrast-Noise Ratio
Slide provided by Doug Pfeiffer
CT # Uniformity
Slide provided by Doug Pfeiffer
Dosimetry
Table 1 - Common Mistakes
For mA row – entering mAs, effective mAs or mAs/slice
Help site understand difference between these
And that they are not all equivalent mA ≠ mAs mAs ≠ eff. mAs mAs ≠ mAs/slice
Dosimetry
Siemens – eff. mAs (effective mAs)
Eff .
_ mAs
= mA * rot _ time
Pitch mA
=
Eff .
_ mAs * Pitch rot _ time
Philips – mAs/Slice (similar definition to eff. mAs) mAs / slice
= mA * rot _
Pitch time mA
=
( mAs / Slice rot _
) * time
Pitch
Toshiba and GE use mA, time , Pitch as separate values
Dosimetry
Common Mistakes include:
Reporting mAs or eff. mAs or mAs/slice in Table
1
Then using mAs or eff. mAs when performing
CTDI measurements
Example: 200 eff. mAs, pitch .9, rot. time = 0.5 sec
In this case, mA = 360
Should perform CTDI measurement with 180 mAs
Spreadsheet will use pitch 0.9 and correct for values of effective mAs
Dosimetry
Common Mistakes include:
If site does this incorrectly, spreadsheet will have incorrect values
If they perform acquisition with 200 mAs
And then use N,T and I such that a pitch of 0.9 results, then CTDIvol reported will be too high
If Pitch < 1, CTDIvol reported will be too high
If Pitch > 1, CTDIvol reported will be too low
Dosimetry
Common Mistakes - Reporting collimation incorrectly
Admittedly this can be confusing for some scanners
Example: Siemens Sensation 64
Scanner user interface says 64 x 0.6 mm
Scanner uses z-flying focal spot, which double samples on zaxis of anode to obtain 2X images.
Actual beam width is N=32, T = 0.6 mm
For Pitch 1.0, table travel will be 19.2 mm/rotation
Site sometimes list N=64, T=0.6mm, I= 38.4mm (-> Pitch 1)
In spreadsheet, this yields CTDIvol that is half what it should be
Dosimetry
Common Mistakes - Reporting collimation incorrectly
Consult ACR CT Accreditation website for FAQs and clarifications http://www.acr.org/accreditation/computed/ct_faq.asp
x
Dosimetry
Slide provided by Doug Pfeiffer
Dosimetry
Slide provided by Doug Pfeiffer
Dosimetry
Slide provided by Doug Pfeiffer
Dosimetry
Slide provided by Doug Pfeiffer
Dose Levels
Slide provided by Doug Pfeiffer
Dosimetry
Slide provided by Doug Pfeiffer
Dosimetry
Slide provided by Doug Pfeiffer
Dosimetry
Slide provided by Doug Pfeiffer
Pass/Fail Criteria
Slide provided by Doug Pfeiffer
Phantom Deficiencies
Slide provided by Doug Pfeiffer
Phantom Deficiencies
Slide provided by Doug Pfeiffer
Phantom Deficiencies
Slide provided by Doug Pfeiffer
Phantom Deficiencies
Slide provided by Doug Pfeiffer
Slide provided by Doug Pfeiffer
Toshiba CT Scanners
Aquilion
Scanners (32, 64 or 320)
Slide provided by Erin Angel, Toshiba Medical Systems
Toshiba Aquilion 32 or 64 slice
CT Scanners
The maximum number of axial images able to be acquired simultaneously in a single rotation (Nmax) is 4 .
The minimum rotation time for a full 360 rotation can be found in the product data sheet or via loading any helical protocol and choosing the lowest available 360 rotation time (note: grayed-out times are not available).
Slide provided by Erin Angel, Toshiba Medical Systems
Toshiba Aquilion 32 or 64 slice
CT Scanners
Slice thicknesses available in helical mode include:
0.5mm, 1.0mm, 2.0mm, 3.0mm, 5.0mm, 7.0mm, and 10.0mm.
Slice thicknesses available in axial mode include:
0.5mm, 1.0mm, 2.0mm, 3.0mm, 4.0mm, 6.0mm,
8.0mm, 12mm (16mm, 24mm, and 32mm are also available for non-clinical applications).
Available beam energies * are
80 kVp, 100 kVp, 120 kVp, and 135 kVp
Slide provided by Erin Angel, Toshiba Medical Systems
Toshiba Aquilion 16 slice CT
Scanners
The maximum number of axial images able to be acquired simultaneously in a single rotation (Nmax) is 4 .
The minimum rotation time for a full 360 rotation can be found in the product data sheet or via loading any helical protocol and choosing the lowest available 360 rotation time (note: grayed-out times are not available).
Slide provided by Erin Angel, Toshiba Medical Systems
Toshiba Aquilion 16 slice CT
Scanners
Slice thicknesses available in helical mode include:
0.5mm, 1.0mm, 2.0mm, 3.0mm, 5.0mm, 7.0mm, and 10.0mm.
Slice thicknesses available in axial mode include:
0.5mm, 1.0mm, 2.0mm, 3.0mm, 4.0mm, 6.0mm,
8.0mm, 12mm (16mm, 24mm, and 32mm are also available for non-clinical applications).
Available beam energies * are
80 kVp, 100 kVp, 120 kVp, and 135 kVp
Slide provided by Erin Angel, Toshiba Medical Systems
Toshiba Scanners - Location of
Technical Parameters
Toshiba Scanners - Scan Detail
Tab
Slide provided by Erin Angel, Toshiba Medical Systems
Slide provided by Erin Angel, Toshiba Medical Systems
Scan Details for Axial Scanning
Slide provided by Erin Angel, Toshiba Medical Systems
Recon Details Tab
IEC Pitch Value and Table
Speed
Always use the IEC definition of pitch when reporting to ACR
IEC pitch can be calculated either by:
Dividing the helical pitch (HP) by the number of detector channels being used (N)
Clicking on the HP button on the Scan Details tab to find the IEC pitch value (ex: PF 0.828). The IEC pitch value is called “Pitch Factor” on your Aquilion scanner, also denoted PF.
Slide provided by Erin Angel, Toshiba Medical Systems
IEC Pitch Value and Table
Speed
Slide provided by Erin Angel, Toshiba Medical Systems
IEC Pitch Values and Table Speed
Detailed Pitch
Row Helical Pitch (HP) IEC Pitch Factor
16 11
32
0.688
21 0.656
64 41 0.641
Slide provided by Erin Angel, Toshiba Medical Systems
IEC Pitch Values and Table Speed
Standard Pitch
Row Helical Pitch (HP) IEC Pitch Factor
16 15
32
0.9375
27 0.84375
64 53 0.828
Fast PitchRowHelical Pitch (HP)IEC Pitch Factor
(PF)16231.437532451.4062564951.484
Slide provided by Erin Angel, Toshiba Medical Systems
IEC Pitch Values and Table Speed
Fast Pitch
Row Helical Pitch (HP) IEC Pitch Factor
16 23
32
1.4375
45 1.40625
64 95 1.484
ACR Tips for Toshiba CT’s
Have the scanner calibrated for air and water, at all relevant kVps, prior to testing
Make sure you scan the ACR phantom with the small (S) Scan FOV for the image quality tests. Scanning a small size phantom with a large size FOV can lead to CT number and uniformity errors.
For best results be sure the phantom is properly centered, utilizing the lasers and a
ACR Tips for Toshiba CT’s
Have your “Average Clinical Parameters” chart filled out in advance and be certain to scan the phantom with the corresponding parameters.
Before proceeding with phantom image quality testing, ensure your protocols are optimized for dose while obtaining adequate image quality and that radiation doses do not exceed ACR guidelines.
For proper landmarking, note where the center of the slice will be when you prescribe a scan in axial or helical mode
Slide provided by Erin Angel, Toshiba Medical Systems
Finding the Center Slice Location
Axial Mode: If a scan is prescribed, for example, from 24mm to 26mm in 4 x 0.5mm axial mode, the center slice location will be as follows:
Slide provided by Erin Angel, Toshiba Medical Systems
Helical Mode: If a scan is prescribed, for example, from 24mm to 26mm in 4 x 0.5mm helical mode, the center slice location will be as follows (note: this is different from axial mode):
Slide provided by Erin Angel, Toshiba Medical Systems
Objectives
CONCLUSIONS
For all scanners, make sure that the facility has had service perform preventive maintenance recently.
Review the protocols used with the technologist prior to performing phantom scans and dosimetry.
Contact the vendor help desk for further clarification on a particular scanner.
If at all possible, burn the CD’s for ACR Submission prior to leaving the facility. Do not depend on staff at the facility to correctly burn the CD’s.
Contact Information:
Melissa C. Martin, M.S., FACR,
FAAPM
Therapy Physics Inc.
879 West 190th St., Ste 400
Gardena, CA 90248 e-mail:
Melissa@TherapyPhysics.com
website: www.TherapyPhysics.com
Phone: 310-612-8127