Compressive Fatigue Behaviour of a Polymer-Based Osteochondral Scaffold Y-H. Hsua, C Luptona, J. Tonga, A. Cosseyb, A. Auc a Mechanical Behaviour of Materials Laboratory, School of Engineering, University of Portsmouth, UK b c Queen Alexandra Hospital, Portsmouth, UK Smith & Nephew, Advanced Surgical Devices, Andover, MA, USA Abstract Cyclic loading from daily activities or from intensive exercise can lead to increased risk of fracture. Implants designed for load bearing purposes, such as repair of articular cartilage and underlying subchondral bone in knees must have the necessary resistance to fatigue loadings. To date, virtually no studies have been reported on the fatigue behaviour of osteochondral scaffolds, despite damage due to repeated loading is of considerable interest for tissue repair purposes. In this study, a polymer-based osteochondral scaffold has been studied under cyclic loading conditions. Multi-step cyclic tests have been carried out with increasing compressive loads in a phosphate buffered saline solution at 37oC. Changes in secant modulus and residual strain accumulation are monitored. Morphological parameters of the scaffold are determined using micro-computed tomography. The secant modulus and the number of cycles to failure of the scaffold are obtained and compared with those of human trabecular bone (Topolinski et al, 2011). 1. Introduction A variety of biological and synthetic scaffolds have been developed for articular cartilage and subchondral bone repair purposes.1-11 However, scaffolds often lack the physical structure and mechanical properties necessary to sustain long-term applications.12 Although biocompatibility and bio-related issues have been studied widely, evaluation of mechanical properties has been an area of underdevelopment, with testing parameters and conditions varied greatly; and testing under controlled physiological loading conditions similar to those in vivo particularly lacking. Evaluation of the mechanical properties of scaffolds is not always done in aqueous solution at 37°C, although the mechanical properties of hydrated scaffold samples are known to be much lower than those of dry samples.4-6 13 A number of scaffolds and bones were tested in tension,8, 14, 15 which bears little resemblance to the in vivo loading environment primarily in compression. Human bone is subjected to a variety of loading patterns during daily routine activities. A typical loading condition is repeated or cyclic loading as experienced in walking or running. Bone response under such loading conditions may be referred to as fatigue behaviour.16, 17 Fatigue damage is a process of gradual mechanical degradation caused by repeated loading at stress or strain much lower than those required to fracture in a single application of force. As a result, damage due to cyclic loading is of significant interest for design and application of synthetic scaffolds. In a conventional fatigue test, specimens are subjected to a constant amplitude load till failure to determine the number of cycle to failure at a given load range. A number of studies based on this method have been carried out on the fatigue behaviour of bones.15, 18-30 Some studies have also been carried out for synthetic biopolymers.31, 32 Using this method a great number of samples may be needed to obtain the fatigue behaviour under variable stress amplitudes. Stepwise-testing has since been developed which would allow the generation of fatigue data with only a small numbers of samples.33, 34 In stepwise-tests, samples are subject to loads of a range of amplitudes in each a number of cycles are allowed to elapse, thus providing an accelerated testing regime with increased data generation. parameter obtained from such tests is the evolution of secant modulus One important 21, 33, 35 which is an indication of damage accumulation.30, 36 Polymer-based osteochondral scaffolds are synthetic resorbable implants designed to replace worn-out cartilage surfaces, restoring mobility and relieving joint pain. Although used in some clinical cases,37-39 there is a lack of published data on the biomechanical properties of the implant. Knowledge of its mechanical behaviour is essential in order to obtain accurate prediction of stresses and deformation in clinical applications, where the structural integrity of the implant is particularly important when cyclic loading is considered. The motivation of the study is to evaluate the fatigue behaviour of an osteochondral scaffold using the step-wise testing method and to compare the results with those from human bone in terms of stress-strain response and secant modulus. Morphological parameters of the scaffold were also obtained and compared with those of human bone. 2. Materials and methods 2.1 Specimen A commercial scaffold TRUFIT (Smith and Nephew) was used in the study. It is composed of polylactide-co-glycolide (PLG) copolymer with calcium-sulfate and polyglycolide (PGA) fibres. The dual-layer design of the implant contains both a cartilage and a bone phase. Cylinders of the two layers scaffolds were sectioned to obtain a single layer of TRUFIT Bone, with a diameter of 10.5mm, a length of 9 mm and an aspect ratio of 0.86. The aspect ratio was selected based on the information obtained previously,40-42 where reported aspect ratios are generally less than 1. 33, 43-45 The aspect ratio and dimensions of the TRUFIT Bones are in a similar range of those of human trabecular bones in Topolinski’s study. 33 The samples were soaked in phosphate buffered saline (PBS) solution overnight prior to testing. 2.2 Morphological Parameters The microstructure of the specimens was examined prior to fatigue testing using micro-focus computed tomography (CT X-Ray Inspection System, Metris X-Tek Systems Ltd). The MicroCT scanner was operated at a voltage of 51 kV, current of 160 μA and a voxel size of 17-20 μm. Bone volume density (BV/TV), mean thickness of the trabeculae in the specimen (Tb.Th) and the mean distance between trabeculae (Tb.Sp) were obtained by processing the micro-CT images using Microview software. 2.3 Compression fatigue testing Fatigue testing was carried out on a Bose ElectroForce 3200 Testing system equipped with an environmental chamber to which PBS solution was filled and controlled at a temperature of 37˚C. Each specimen was placed carefully at the centre of the platens inside the environmental chamber. During the testing, the specimens were unconstrained33 between the platens of the testing machine as fixation of the specimens to the test platens may increase stiffness 46 Five TRUFIT Bones were tested under load control26, 27 with a stepwise increase of maximum stress whilst keeping the minimum stress constant, following a testing protocol of Topolinski et al. 33 A preload of 5 N was applied first to ensure good end contact and this preload was subsequently kept constant as the minimum load throughout the test. The maximum load started from 10 N with a gain of 10 N at successive steps, as shown in Figure 1. The maximum load was kept constant during a period of 500 cycles and increased to the next level thereafter. The loading scheme was adopted from Topolinski et al33 to allow easy comparison with the results from human trabecular bone.33 The loading waveform was sinusoidal at a frequency of 1 Hz. Engineering stress and strain were calculated by using the applied load, displacement and the original dimension of the specimens, where lateral deformation under compression was not considered. The changes in the secant modulus and the maximum and minimum strains with fatigue loading were recorded. The change in the secant modulus is considered indicative of damage accumulation in the specimen30, 36 in fatigue studies of bones from the literature.15, 18, 21, 23-29, 33-35, 47 The secant modulus at a given cycle was defined as Δσ/Δ = (σmax-σmin)/(max- min), determined from the stress-strain curve. Figure 2 illustrates this as the slope of the line connecting the lowest and the highest point of a stress-strain loop. The secant modulus and strains were measured throughout the test. The cyclic stress-strain curves were also compared with those obtained under monotonic loading conditions. 3. Results and discussion Table 1 shows the morphometric values of the TRUFIT Bone and human trabecular bone.33 The average BV/TV of the scaffold (0.343) is well within the range of human trabecular bone (0.076-0.460). The trabecular thickness of the scaffold is slightly lower (0.086mm) than that of human trabecular bone (0.105-0.268mm); whilst the trabecular spacing (0.166mm) is smaller than that of human (0.424-1.829mm). These indicate that the microstructure of the scaffold consists of thinner but more densely packed struts than that of human trabecular bone. Table 1 also includes the maximum stress corresponding to the maximum secant modulus. Samples with higher values of the maximum stress indicate higher fatigue resistance. It is clear that the fatigue resistance of the scaffold at an average of 0.462MPa is much lower than the range (0.859-12.28MPa) from human trabecular bone. A large variation in the results from human bones may be due to the varied sources (e.g., multiple donors with individual characteristics and pathologies).33 In contrast, synthetic TRUFIT bones have more consistent morphometric values. Although a relationship between bone structure and fatigue strength for human bone was reported,33 it is not possible to obtain such a relation for the scaffold due to the lack of variety of the microstructure features. Typical stress-strain curves of a sample tested under the multi-step loading scheme (LS1 to LS8) are shown in Figure 3. Each block, shown in a different colour, represents the response to a single loading step with 500 loading cycles. A distinctive characteristic at all load levels is the accumulation of residual strain, or cyclic creep, with the increase of loading cycles. Greater strain ratchetting appears to be associated with higher load levels, when non-linear deformation becomes more evident in the larger hysteresis loops under LS7 and LS8. This is consistent with the behaviour observed in trabecular bones.26, 29, 35, 47 The evolution of the secant modulus (as defined in Figure 2) with cycles is shown in Figure 4(a) for the five samples tested. In all cases, it appears that an increase in secant modulus with cycle persisted for most of the loading steps, particularly the early ones. Only towards later at higher load levels reductions in secant modulus become evident. The initial hardening seems to be somewhat unexpected, certainly in contrast to the continuing increase in the residual strain with cycle, as shown in Figure 4(b), where the maximum and the minimum compressive strains of the five samples are recorded. Interestingly samples with lower residual strains (A, B, C) tend to have higher secant modulus (Figure 4(a)). Both maximum and minimum strains increased with the number of cycles during the entire test and the gap between the maximum and the minimum strains is roughly constant for each sample, indicating that the strain range stays constant. In addition, the strain rate increases with the increase of the number of cycles. It is known that residual strain or cyclic creep plays an important role in material failure due to the accumulated plastic deformation.30, 35 Increase in strains with the number of cycles have been generally observed when testing bones under cyclic loading, 27, 34 although increasing strain rate during the entire test was found in bovine trabecular bone. 34 Three stages of deformation during compression fatigue tests have been reported for human trabecular bone,27 including a transient behaviour characterised by rapid strain increase within the initial load cycles, saturation of strain and an acceleration towards final catastrophic failure. When secant modulus of the scaffold is compared with those of human trabecular bones,33 as shown in Figure 4(c), much lower secant modulus and total number of cycles to failure were obtained for the scaffold, although the patterns of evolution with cycle appear to be similar between the two. The initial hardening of the bone is also consistent with the results under conventional constant-amplitude fatigue testing. Linde and Hvid51 reported that in human trabecular bones the stiffness increases until a stress level about 50% of ultimate stress is reached followed by decreasing stiffness, although others reported decreasing in modulus from the beginning to the end.18, 22, 24-27, 52, 53 Michel et al35 studied the fatigue behaviour of bovine trabecular bones under load control using a sinusoidal compressive load profile at a frequency of 2 Hz. The results showed that the pattern of modulus change with the number of cycles was associated with the load level. At low cyclic load levels the modulus increased initially followed by a rapid drop in the final stage. At high cyclic loads a continuous decrease in modulus from the beginning was found. The authors35 suggested that both creep and damage accumulation may be responsible for fatigue failure of trabecular bone, i.e. bone fails by creep under high cyclic loads whilst by microcrack damage accumulation at low cyclic loads. A more recent study suggested, however, that creep effects are negligible in fatigue loading cases for trabecular bone.54 An initial increase in modulus has been observed for TRUFIT at most of the load levels except the highest levels, which appears to be consistent with the observation in some of the reports.35 and 51 Stress-strain loops of first cycles at the beginning of each fatigue loading step are presented in Figure 5(a). Each centre of the loops was moved to the origin (0,0) then the points at the top of each loop were linked to form the initial cyclic stress-strain curve. Figure 5(b) shows the final cyclic stress-strain curve which was obtained similarly using the loops of the final cycles at the end of each loading step. These cyclic stress-strain loops at the first and the final cycles do not appear to differ significantly (Figure 6, CSS_1 and CSS_f), suggesting stress softening may not be significant. Figure 6 shows a comparison of cyclic and monotonic stressstrain curves for a typical specimen of TRUFIT. Here LS1_0 to LS5_0 are monotonic stressstrain curves whilst CSS_1 and CSS_f are the first and the final cyclic stress-strain curves. The monotonic curves are the loading parts of a stress-strain curve at the beginning of the each loading step. The slopes of these curves are the modulus of elasticity. A comparison of the mean elastic modulus of the stress-strain curves is shown in Figure 7. The elastic modulus of monotonic stress-strain curves increase with the number of loading steps (a) and the moduli of cyclic curves are generally higher than those of monotonic ones, suggesting stress hardening, consistent with the trend shown in Figure 4(a). However, the elastic modulus of first cyclic stress-strain curve is higher than that of final cyclic stress-strain curve, suggesting damage accumulation due to fatigue. 4. Conclusions Multi-step cyclic tests have been carried out on TRUFIT Bones to study the fatigue behaviour of the scaffold under increasing compressive cyclic loading conditions. Morphological parameters of TRUFIT scaffold have also been obtained using computed tomography and compared with those of human trabecular bone. The results show an increase in secant modulus and stress hardening during the initial steps of fatigue loading, consistent with that observed in human trabecular bone. The modulus and the number of cycle to failure of the scaffold are, however, significantly lower than those of human trabecular bone. The elastic modulus of monotonic stress-strain curves increases with the number of loading steps and the moduli of cyclic curves are generally higher than those of monotonic curves. Progressive increase in residual strains has been observed in the scaffold for the entire test duration, indicating that residual strain accumulation, or cyclic creep, may be the predominant driving force leading to failure of the scaffold. Multi-step tests seem to be useful in assessing the essential fatigue behaviour of scaffolds. These tests allow the reduced number of samples for a reliable analysis of fatigue properties of biomaterials. Acknowledgements The authors would like to thank Smith & Nephew for the provision of samples and University of Portsmouth for financial support. References 1. Toolan BC, Frenkel SR, Pachence JM, Yalowitz L, Alexander H. Effects of growth-factorenhanced culture on a chondrocyte-collagon implant for cartilage repair. Journal of Biomedical Materials Research 1996;31:273-80. 2. Athanasiou K, Schmitz JP, Agrawal CM. The effects of porosity on in virto degradation of polylactic-acid-polyglycolic acid implants used in repair of articular cartilage. Tissue Engineering 1998;4(1):53-63. 3. Lu L, Zhu X, Valenzuela RG, Currier BL, Yaszemski MJ. Biodegradable polymer scaffolds for cartilage tissue engineering. Clinical Orthopaedics and Related Research 2001;391s:S251S70. 4. Slivka MA, Leatherbury NC, Kieswetter K, Niederauer GG. Porous, resorbable, fiberreinforced scaffolds tailored for articular cartilage repair. Tissue Engineering 2001;7(6):767-80. 5. Harley BA, Lynn AK, Wissner-Gross Z, Bonfield W, Yannas IV, Gibson LJ. Design of a multiphase osteochondral scaffold. II. Fabrication of a mineralized collagen–glycosaminoglycan scaffold. Journal of Biomedical Materials Research Part A 2010;92(3):1066-77. 6. Harley BA, Lynn AK, Wissner-Gross Z, Bonfield W, Yannas IV, Gibson LJ. Design of a multiphase osteochondral scaffold III: Fabrication of layered scaffolds with continuous interfaces. Journal of Biomedical Materials Research Part A 2010;92(3):1078-93. 7. Oka M, Ushio K, Kumar P, Ikeuchi K, Hyon SH, Nakamura T et al. Development of artificial articular cartilage. Proceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine 2000;214:59-68. 8. Bera B. Development of artificial articular cartilage. S¯adhan¯a. 2009;34(5):823-31. 9. Oka M, Noguchi T, Kumar P, Ikeuchi K, Yamamuro T, Hyon SH et al. Development of an artificial articular cartilage. Clinical Materials 1990;6:361-81. 10. Hung CT, Lima EG, Mauck RL, Taki E, LeRoux MA, Lu HH et al. Anatomically shaped osteochondral constructs for articular cartilage repair. Journal of Biomechanics 2003;36(12):1853-64. 11. Sherwood JK, Riley SL, Palazzolo R, Brown SC, Monkhouse DC, Coates M et al. A threedimensional osteochondral composite scaffold for articular cartilage repair. Biomaterials 2002;23(24):4739-51. 12. Suh J-K, Arøen A, Muzzonigro TS, Disilvestro M, Fu FH. Injury and repair of articular cartilage: Related scientific issues. Operative Techniques in Orthopaedics 1997;7(4):270-8. 13. Slivka MA, Leatherbury NC, Kieswetter K, Niederauer GG. In vitro compression testing of fiber-reinforced, bioabsorbable, porous implants. In: Mauli Agrawal C, Parr JE, Lin ST, editors. Synthetic Bioabsorbable Polymers for Implants. Kansas City, USA2000. p. 124-35. 14. Messner K. Hydroxylapatite supported Dacron plugs for repair of isolated full-thickness osteochondral defects of the rabbit femoral condyle: mechanical and histological evaluations from 6-48 weeks. Journal of Biomedical Materials Research 1993;12:1527-32. 15. Zioupos P, Casinos A. Cumulative damage and the response of human bone in two-step loading fatigue. Journal of Biomechanics 1998;31:825-33. 16. Martin RB. Fatigue microdamage as an essential element of bone mechanics and biology. Calcified Tissue International 2003;73:101-7. 17. Taylor M, Tanner KE. Fatigue failure of cancellous bone: a possible cause of implant migration and loosening. The Journal of Bone & Joint Surgery. 1997;79-B:181–2. 18. Pattin CA, Calert WE, Carter DR. Cyclic mechanical property degradation during fatigue loading of cortical bone. Journal of Biomechanics 1996;29(1):69-79. 19. O’Brien FJ, Taylor D, Lee TC. Microcrack accumulation at different intervals during fatigue testing of compact bone. Journal of Biomechanics. 2003;36:973-80. 20. Yeni YN, Fyhrie DP. Fatigue damage-fracture mechanics interaction in cortical bone. Bone. 2002;30(3):509-14. 21. Cotton JR, Winwood K, Zioupos P, Taylor M. Damage rate is a predictor of fatigue life and creep strain rate in tensile fatigue of human cortical bone samples. Journal of Biomechanical Engineering 2005;127:213-9. 22. Fleck C, Eifler D. Deformation behaviour and damage accumulation of cortical bone specimens from the equine tibia under cyclic loading. Journal of Biomechanics 2003;36:179-89. 23. Yamamoto E, Crawford RP, Chan DD, Keaveny TM. Development of residual strains in human vertebral trabecular bone after prolonged static and cyclic loading at low load levels. Journal of Biomechanics 2006;39:1812-8. 24. Zioupos P, Wang XT, Currey JD. The accumulation of fatigue microdamage in human cortical bone of two different ages in vitro. Clinical Biomechanics. 1996;11:365-75. 25. Moore TLA, Gibson LJ. Fatigue microdamage in bovine trabecular bone. Journal of Biomechanical Engineering 2003;125:769-76. 26. Moore TLA, Gibson LJ. Fatigue of bovine trabecular bone. Journal of Biomechanical Engineering 2003;125:761-8. 27. Dendorfer S, Maier HJ, Taylor D, Hammer J. Anisotropy of the fatigue behaviour of cancellous bone. Journal of Biomechanics 2008;41:636-41. 28. Carter DR, Hayes WC. Compact bone fatigue damage – I. Residual strength and stiffness. Journal of Biomechanics 1977;10:325-37. 29. Rapillard l, Charlebois M, Zysset P. Compressive fatigue behavior of human vertebral trabecular bone. Journal of Biomechanics 2006;39:2133-9. 30. Dendorfer S, Maier HJ, Hammer J. Fatigue damage in cancellous bone: an experimental approach from continuum to micro scale. Journal of the Mechanical Behavior of Biomedical Materials 2009;2(1):113-9. 31. Palissery V, Taylor M, Browne M. Fatigue characterization of a polymer foam to use as a cancellous bone analog material in the assessment of orthopaedic devices Journal of Materials Science: Materials in Medicine 2004;15(1):61-7. 32. Meyer RW, Pruitt LA. The effect of cyclic true strain on the morphology, structure, and relaxation behavior of ultra high molecular weight polyethylene Polymer 2001;42:5293-306. 33. Topolinski T, Cichanski A, Mazurkiewicz A, Nowicki K. Study of the behavior of the trabecular bone under cyclic compression with stepwise increasing amplitude. Journal of the Mechanical Behavior of Biomedical Materials 2011;4(8):1755-63. 34. Guillén T, Ohrndorf A, Tozzi G, Tong J, Christ H-J. Compressive fatigue behavior of bovine cancellous bone and bone analogous materials under multi-step loading conditions. Advanced Engineering Materials 2012;14(5):B199-B207. 35. Michel MC, Guo X-DE, Gibson LJ, McMahon TA, Hayes WC. Compressive fatigue behavior of bovine trabecular bone. Journal of Biomechanics 1993;26(4-5):453-63. 36. Zioupos P, Gresle M, Winwood K. Fatigue strength of human cortical bone: Age, physical, and material heterogeneity effects. Journal of Biomedical Materials Research Part A 2008;86A(3):627-36. 37. Carmont MR, Carey-Smith R, Saithna A, Dhillon M, Thompson P, Spalding T. Delayed incorporation of a Trufit Plug: perseverance is recommended. Arthroscopy: The Journal of Arthroscopic and Related Surgery 2009;25(7):810-4. 38. Dhollander AAM, Liekens K, Almqvist KF, Verdonk R, Lambrecht S, Elewaut D et al. A pilot study of the use of an osteochondral scaffold plug for cartilage repair in the knee and how to deal with early clinical failures. Arthroscopy: The Journal of Arthroscopic and Related Surgery 2012;28(2):225-33. 39. Pearce CJ, Gartner LE, Mitchell A, Calder JD. Synthetic osteochondral grafting of ankle osteochondral lesions. Foot and Ankle Surgery 2012;18:114-8. 40. DiSilvestro MR, Suh J-KF. A cross-validation of the biphasic poroviscoelastic model of articular cartilage in unconfined compression, indentation, and confined compression. Journal of Biomechanics 2001;34:519-25. 41. Huang C-Y, Stankiewicz A, Ateshian GA, Mow VC. Anisotropy, inhomogeneity, andtension– compression nonlinearity of human glenohumeral cartilage in finite deformation. Journal of Biomechanics 2005;38:799-809. 42. Linde F, Nobrgaard P, Hvid I, Odgaard A, Soballe K. Mechanical properties of trabecular bone. Dependency on strain rate. Journal of Biomechanics 1991;24(9):803-9. 43. Carter DR, Hayes WC. The compressive behavior of bone as a two-phase porous structure. The Journal of Bone and Joint Surgery (Am) 1977;59(7):954-62. 44. Korhonen RK, Laasanena MS, Toyras J, Rieppob J, Hirvonena J, Helminen HJ et al. Comparison of the equilibrium response of articular cartilage in unconfined compression, confined compression and indentation. Journal of Biomechanics 2002;35:903-9. 45. Park S, Hung CT, Ateshian GA. Mechanical response of bovine articular cartilage under dynamic unconfined compression loading at physiological stress levels. OsteoArthritis and Cartilage 2004;12:65-73. 46. Linde F, Hvid I. The effect of constraint on the mechanical behaviour of trabecular bone specimens. Journal of Biomechanics 1989;22(5):485-90. 47. Haddock SM, Yeh OC, Mummaneni PV, Rosenberg WS, Keaveny TM. Similarity in the fatigue behavior of trabecular bone across site and species. Journal of Biomechanics 2004;37:181-7. 48. Caler WE, Carter DR. Bone creep-fatigue damage accumulation. Journal of Biomechanics 1989;22:625-35. 49. Topolinski T, Cichanski A, Mazurkiewicz A, Nowicki K. Study of the behavior of the trabecular bone under cyclic compression with stepwise increasing amplitude. Journal of the Mechanical Behavior of Biomedical Materials In Press. 50. Fray ME, Altstadt V. Fatigue behaviour of multiblock thermoplastic elastomers. 1. Stepwise increasing load testing of poly(aliphatic/aromatic-ester) copolymers. Polymer. 2003;44:4635-42. 51. Linde F, Hvid I. Stiffness behaviour of trabecular bone specimens. Journal of Biomechanics 1987;20(1):83-9. 52. Schaffler MB, Radin EL, Burr DB. Mechanical and morphological effects of strain rate on fatigue of compact bone. Bone 1989;10:207-14. 53. Zioupos P, Wang XT, Currey JD. Experimental and theoretical quantification of the development of damage in fatigue tests of bone and antler. Journal of Biomechanics 1996;29:989-1002. 54. Moore TLA, O’Brien FJ, Gibson LJ. Creep does not contribute to fatigue in bovine trabecular bone. Journal of Biomechanical Engineering 2004;126(3):321-9. Figure1. The loading scheme of a stepwise fatigue test (Fmax=10N; Fmin=Fpreload=5N), following the protocol of Topolinski et al, 2011.33 Figure 2. Secant modulus (Ei) is defined as the stress range divided by the strain range at a given cycle i, which is determined by the slope of the line connecting the minimum and the maximum stress values of the cycle. Figure 3. Stress-strain curves from a multi-step fatigue test. Figure 4. (a) The evolution of secant modulus as a function of cycles for TRUFIT Bone (b) the development of the maximum and minimum strains with cycle and (c) comparison of secant modulus of TRUFIT Bone and human trabecular bone.33 Figure 5. Cyclic stress-strain responses: (a) Loops of first cycles at the beginning of each loading step; (b) loops of the final cycles at the end of each loading step. Figure 6. Comparison of the cyclic and the monotonic stress-strain curves of a typical sample. Figure 7. The elastic modulus of (a) monotonic curves; and (b) cyclic stress-strain curves. Table 1. Morphometric values and maximum stresses of TRUFIT bone and human trabecular bone. 33 Bone volume density Trabecular thickness Trabecular spacing TRUFIT Bone BV/TV Tb.Th (mm) Tb.Sp (mm) Max Stress (MPa) (corresponding to Emax) A 0.359 0.089 0.161 0.547 B 0.340 0.082 0.158 0.328 C 0.353 0.087 0.159 0.547 D 0.328 0.089 0.184 0.337 E 0.333 0.084 0.169 0.552 Human trabecular bone [33] 0.076 - 0.460 0.105 - 0.268 0.424 - 1.829 0.859 - 12.280