ME 322: Instrumentation Lecture 4 January 26, 2016 Professor Miles Greiner Lab Guidelines and grading, pressure gages, Manometers, Instrument sensitivity Announcements • HW assignments, reading and Labs on Website • http://wolfweb.unr.edu/homepage/greiner/teaching/MECH322Instrumentation/ • Turn in HW 1 now – Use ME 322 student number, Not name • HW 2 due Monday 2/1/2015 – Use ME 322 student number and Lab Day to make it easier to return in lab, if necessary. • Joseph Young will hold office hours – 9-10:50 AM (right after this class) on the Monday, Wednesday, of Friday before HW is due in ME Kitchen – This Friday • Go to PE 2 for lab this week – Lab 2 - Statistical Analysis of UNR Quad Measurements – Download and read instructions Lab 2 Instructions and Lab Guidelines Lab Grading • Participation Grade Assigned by lab assistant, based on: – Working in 2-person lab groups (assigned in lab) – Being on time (responsibility to lab partner and TA) – Being prepared (read lab instructions, in future labs bring prepared Excel and LabVIEW files) – Participating and interacting with your partner in setting up experiment, acquiring data, checking consistency, retaking data if necessary, writing report, and cleaning up. – Being professional and patient (treat partner and TA’s well) • Report Grade Assigned by grader, based on: – Following directions and answering questions – Calculation results (based on your measurements), and – Clear tables and plots (formatting) Plots Formats Good ο • • • • • Bad ο Clear Engineering Communication Variable name, symbol and [units] on axes Make plot, labels and symbols big enough to see clearly Avoid redundant frames that make the plot smaller These features are graded Tables Bad ο Good ο Standard Average Deviation Average Standard Deviation Long Side Length, L [ft] 464 81 L 464 81 Short Side Lenth, S [ft] 135 23 S 135 23 Area, A [ft2] 64000 20000 A 64000 20000 Cost, C [$] 1110 350 C 1110 350 • Font size, variable name, symbols, units • Grids, centered • Clarity, simplicity Pressure • Generally can’t see it • Need gages (gauges) to – Qualitatively know when it changes – Quantitatively measure it • Units: Force per area – [Pa] = [N/m2]; [kPa] = [1000 Pa]; [MPa] = [106 Pa] – [psi] = [lbf/in2], [1 atm] = 101.3 kPa • Gage Pressure – Amount a vessel’s pressure is above its surrounding’s P – PGAGE = P – PATM PATM Some Gages • Bourdon Tube • Diaphragm • Manometers – Vertical – Inclined – Water Head, h Instrument Transfer Function • Sometimes call Calibration curve • Relationship between the instrument Reading R (h or output) and its Measurand M (pressure or quantity being measured) Reading – Not always linear Measurand (quantity being measured) U-Tube Manometer Sensed π1 Fluid Density π2 Manometer Fluid Density Reading π = π1 + ππ πβ = π2 + ππ πβ βπ = π1 − π2 = ππ − ππ πβ πΌπ ππ << ππ πβππ: βπ = ππ πβ Measurand Fluid Height when DP = 0 π Reading π ππ ππ Fluid Air (1 ATM, 27°C) 1.774 Water (30°C) 995.7 Hg (27°C) 13,565 • The reading can be given in units of inches or cm of water column [in WC] or [cm WC] – Or inches or cm of mercury or alcohol (manometer fluid) • Transfer function β = βπ ππ π (Reading β versus Measurand βπ, linear) Transfer Function Reading, h [cm] Low rm (alcohol) High rm (mercury) βπ β= ππ π Measurand, βP [kPa] • Instrument sensitivity describes how much the Reading changes for a small change in the measurand – π= ππ ππ = π ππππ = 1 ππ π – Increases as ππ decreases (small change in P produces larger h) Inclined-Well Manometer AW P1 DP = 0 h1 h2 q AT P2 R β = β1 + β2 β2 = π π πππ β1 π΄π€ = π π΄π‘ π΄π‘ β1 = π π΄π€ π΄π‘ π΄π‘ β= π + π π πππ = π + π πππ π΄π€ π΄π€ • By measuring R we can find h. Generally π > β • Use a scale on the measuring tube to measure π directly Transfer Function π΄π‘ βπ = π1 − π2 = ππ πβ = ππ ππ π πππ + π΄π€ βπ = ππ π π πππ + If π ≈ 30°, 1 And π πππ ≈ β« 2 π΄π‘ π΄π€ R π΄π‘ π΄π€ πβππ: βπ = ππ ππ π πππ 1 πππππ πππ πΉπ’πππ‘πππ: π = βπ ππ ππ πππ • Sensitivity π = ππ ππ = π ππππ = 1 ππ ππ πππ • S increases as ππ and π decreases – Easy to vary instrument sensitivity Inclined Manometer Transfer Function 1 π = βπ ππ ππ πππ Reading, R [cm] Small q Large q Measurand, βP [kPa] General “Linear” Instrument Characteristics π πππ₯ Reading βπ πππ ππππ ππ‘ππ£ππ‘π¦: ππ π= ππ Measurement π πππ₯ ≡ Maximum Reading (i. e. tube length) βππππ ππππ₯ βπ πππ ≡ Readability (smallest change in reading that can be detected, i.e. tick mark, digit) ππππ₯ = π πππ₯ π βππππ = ≡ Instrument Range (Maximum measurand for which instrument can be used) βπ πππ ≡ Instrument Resolution (smallest detectiable change in measurand) π Sensitivity affects both Resolution and Range • In general, it is not hard to change sensitivity, i.e. ππ 1 π= = ππ ππ ππ πππ – Increasing S improves resolution – But decreases range ππππ₯ βππππ π ππ΄π = π • Resolution as a fraction of Range βπ πππ βππππ βπ πππ π = = π ππππ₯ π πππ₯ πππ₯ π ≡ % of full scale that can be detected. βπ ππΌπ = π Instrument Repeatability • Will an instrument give the same reading every time it is exposed to the same Measurand? • Why not? – Transfer function may drift with time. So at a later time the readings may shift to consistently higher (or lower) values than before. • Referred to as Systematic or Calibration Error – Random variations of uncontrolled inputs (such as RF (radio frequency) noise, orientation of instrument, humidity, hysteresis), may lead to Random variations of the Reading. • Referred to as Random Errors or Imprecision Calibration Drift Hysteresis General Instrument Desired Input (Measurand M) i.e. length, pressure temperature Controlled i.e. temperature, orientation Instrument Undesired Inputs x1, x2, x3, x4, x5,… Output Reading R (deflection, number of steps, needle angle) Uncontrolled i.e. RF frequency, Average walking stride length, Value approved from above or below General Transfer Function • Reading R = fn(M, x1, x2,…, xn) • How to find the Transfer Function? – Theory: • good for simple device (manometer) • Done in other classes • Only includes the effects you model (at best)! – Calibration: • Controlled measurement process • Measure reading (R) while exposing instrument to a range of measurands (M) that are being measured by a reliable standard (used to determine M). • Includes affects you don’t anticipate Standard Instrument Under Test M (units) R(Units) -------------------------------- Reading, R (units) Calibration Correlation Measured Data “Best Fit” Line “Expected” Transfer Function Measurand, M (units) • Expose instrument to a range of Measurands M that are measured by a reliable standard. Record Standard output M and Instrument output Reading R • Correlate Reading R with Measurand M (least squares fit) – May not be linear • Uncontrolled inputs can cause R to have undesired variations of the same M! • The size of the variation is a measure of the instrument impression – random, inconsistent output • Systematic errors can be removed using calibration, but random errors cannot! How to reduce Measurement Imprecision? • Improve the control of undesired inputs, and/or • Use a different instrument that is less sensitive to uncontrolled inputs Lab 3 Pressure Transmitter Calibration Proximity Sensor or IT PHI PLO • Dwyer Series 616 Pressure Transmitter • Measurand: Pressure difference between HI and LO ports, – DP = PHI – PLO – Power must be supplied to pins 1 and 2 (10-35 VDC) • Sensor: diaphragm with a strain or proximity gage – Output may be affected by orientation due to gravity (undesired) • The Output or “Reading” is current, IT [mA], measured by a Digital Multimeter (DMM), pins 3 and 4 End 2016 Transmitter Characteristics • See Lab 3 website, ~$150 • Use different diaphragm thickness or flexibilities to get different sensitivities to vary Full Scale (FS) range and resolution • Output Signal: 4 to 20 mA (h = 0 to FS, linear) • Accuracy: 616: ±0.25% F.S. • Stability: ±1% F.S./yr. • Two models in lab: – 616-1: FS = 3 in WC; Accuracy = 0.0025*3 = ±0.0075 in WC – 616-4: FS = 40in WC; Accuracy = 0.0025*40 = ±0.1 in WC – What does this accuracy mean (what is its confidence level)? Manufacturer’s Inverted Transfer Function • Output Signal: 4 to 20 mA (h = 0 to FS, linear) • Measurand vs Reading – Needed to use the gage • Pressure head: – β = πΉπ πΌ−4ππ΄ 16 • Pressure Difference: – Δπ = ππ»πΌ − ππΏπ = ππ€ πβ ππ π 2.54 ππ 1π πΌ−4ππ΄ = 998.7 3 9.81 2 πΉπ π π πππβ 100 ππ 16 • What is the confidence level (probability) that the real pressure head is within the interval h ± Accuracy? • this is what we will determine in Lab 3 • Manufacturer’s Transfer Function (R vs M) 16 ππ΄ – πΌ= β + 4 ππ΄ πΉπ – We will try to confirm this in Lab 3 Calibration Transfer Function M (units) R(Units) -------------------------------Scatter – Uncontrolled inputs instrument Correlate R (units) Data Uncertainty M (units) Lab 2 Plots and Table Average Standard Deviation Long Side Length, L [ft] 430 110 Short Side Lenth, S [ft] 120 30 Area, A [ft2] 56,000 19,000 Cost, C [$] 980 330 Lab 2 Questions • Is stride length F “highly correlated” with height, H? • Does the distribution of the cost estimates look the way you expect? – Do any of the cost estimates look “out of place?” • How can you interpret the sample mean and standard deviation of the cost estimates in terms of the distribution? • Are the measured values of L and S “highly” correlated? – Should they be? What does this mean? • If you budget the amount of your cost estimate, then – You are only 50% sure to have enough to cover quad (be above the average value, which we assume is the most accurate estimate) – How much money should you budget to be 90% sure to have enough Manometer Reading Rm Transfer Function Plot Actual Outputs Best fit line Manufacturer Specified Performance Applied Pressure hT