Elemental Spectroscopy ICP-OES Content: ICP-OES • Fundamentals of ICP-OES • Instrument Components 2 Theory of Inductively Coupled Plasma Optical Emission Spectroscopy ICP Basics ICP is shorthand for ICP-AES or ICP-OES. What is ICP-AES? It is: Inductively Coupled Plasma Atomic Emission Spectrometer. What is ICP-OES? It is: Inductively Coupled Plasma Optical Emission Spectrometer. 4 Atomic Emission Theory • Atomic emission spectroscopy (AES or OES) uses quantitative measurement of the optical emission from excited atoms to determine analyte concentration • Analyte atoms in solution are aspirated into the excitation region where they are desolvated, vaporized, and atomised by a plasma 5 Atomic Emission Theory Inductively Coupled Plasma Atomic Emission Spectrometer Plasma 6 Polychromator Detector Excitation Excitation E Relaxation Ground State Excited State Electrons can be in their ground state (unexcited) or enter one of the upper level orbitals when energy is applied to them. This is the excited state 7 Atomic Emission A photon of light is emitted when an electron falls from its excited state to its ground state + hv Excited State 8 Ground State Photon Element Wavelengths • Each element has a unique set of wavelengths that it can emit <-- uv --> 180nm 9 1 <-- visible --> 400nm 2 3 4 800nm 5 Atomic Emission explained • Atomic Emission – the wavelength regions Spectral Region Vacuum UV 160 Ultra-Violet 190 Visible 360 Wavelength = nm Near IR 760 900 Lower wavelengths are shorter and have more energy, higher wavelengths e.g. in the Visible region, are longer and have less energy 10 Effect of Temperature on Emission Wavelength increasing -> 200 300 400 600 800 5000 k As Pb Mn Mg Cu Ca Ba Na Li K Mg Cu Ca Ba Na Li K Na Li K 3000 k 2000 k Ca 11 Emission sources • Flames • Arcs / Sparks • Direct Current Plasmas (DCP) • Inductively Coupled Plasmas (ICP) 12 Inductively Coupled Plasma (ICP) – source, plasma formation, plasma zones • Quartz torch surrounded by induction coil • Magnetic coupling to ionized gas • High temperature – equivalent to 10,000k 13 Plasma Advantages • High Temperature – allows for full dissociation of sample components • Argon is Inert – non reactive with sample • Linearity – analysis of samples from ppb to ppm range in the same method • Matrix tolerance – robust and flexible design with Duo and Radial options 14 Plasma Torch 15 Plasma Zones Plasma Zones observation region (mm) 25 6000 k 20 6500 k 15 7000 k RESIDENCE TIME ~ 2MS 8000 k 0 10000 k sample 16 TEMPERATURE ~ 2X NITROUS OXIDE ACETYLENE FLAME Instrument Components There are six basic components to an ICP 1. Sample Introduction 2. Energy Source 3. Spectrometer 4. Detector 5. Electronics 6. Computer and Software 17 Instrument Components 1. Sample Introduction 2. Energy Source 3. 6. Computer and Software 5. Spectrometer Electronics 4. 18 Detector Instrument Components 1. Sample Introduction The sample solution cannot be put into the energy source directly. The solution must first be converted to an aerosol. The function of the sample introduction system is to produce a steady aerosol of very fine droplets. 19 Instrument Components 1. Sample Introduction There are three basic parts to the sample introduction system. i. the Peristaltic pump draws up sample solution and delivers it to ii.the Nebulizer which converts the solution to an aerosol that is sent to iii. the Spray chamber which filters out the large, uneven droplets from the aerosol. 20 Instrument Components 1. Sample Introduction 21 i. the Peristaltic pump ii. the Nebulizer iii. the Spray chamber Concentric Nebuliser 22 Instrument Components 2. Energy Source = plasma The sample aerosol is directed into the center of the plasma. The energy of the plasma is transferred to the aerosol. The main function of the energy source is to get atoms sufficiently energized such that they emit light. 23 Instrument Components 2. Energy Source There are three basic parts to the energy source. i. the Radio frequency generator which generates an oscillating electomagnetic field at a frequency of 27.12 million cycles per second. This radiation is directed to ii.the Load coil which delivers the radiation to iii. the Torch which has argon flowing through it which will form a plasma in the RF field. 24 Instrument Components 2. Energy Source i. the Radio Frequency generator ii. the Load coil iii. the Torch 25 Plasma Configuration • Axial • Radial • Axial and Radial 26 Radial or Axial Configuration • Radial design – Robust, fewer interferences • Petrochemical • Metallurgy • Axial design – best sensitivity, lowest detection limits • Environmental • Chemical 27 Axial Advantage • Much more light available. This gives you the opportunity to achieve Lower Detection Limits than Radial Plasma • BUT- unfortunately, you also get... • More Matrix Interferences • Slightly Reduced Dynamic Range 28 Duo viewing • Axial view plasma looks down the central channel of the plasma, this provides the best sensitivity and detection limits • DUO – this is an axially configured plasma that also allows for radial view through a hole in the side of the axial torch 29 Dual View Optics Radial view Axial view 30 Instrument Components 3. Spectrometer Once the atoms in a sample have been energized by the plasma, they will emit light at specific wavelengths. No two elements will emit light at the same wavelengths. The function of the spectrometer is to diffract the white light from the plasma into wavelengths. 31 Simultaneous Optics – Echelle Spectrometer Detector Grating Prism ICP-Source 32 Instrument Components 3. Spectrometer There are several types of spectrometers used for ICP. Regardless of type, all of them use a diffraction grating. CID Detector Focusing Mirror Echelle grating Prism Shutter Collimating Mirror 33 Slit (dual) For the iCAP, an echelle spectrometer is used. The components in this spectrometer are shown at left. iCAP Optics - Polychromator • High resolution • 7pm @ 200nm • High image quality & low stray light • aberration compensation over whole CID • High energy throughput • double pass prism • All lines on chip • anamorphic magnification • Stable • thermal insulation & heater control to 0.10C 34 Instrument Components 4. Detector Now that there are individual wavelengths, their intensities can be measured using a detector. The intensity of a given wavelength is proportional to the concentration of the element. The function of the detector is to measure the intensity of the wavelengths. 35 Charge Injection Device Array Detector • >291,600 addressable silicon-based photo detectors • Full Spectrum Imaging • Random Access Integration (RAI) • Inherently Anti-blooming – Non Destructive Readout (NDRO), allows the S/N ratio to be improved by repeatedly reading each pixel 36 Instrument Components 4. Detector 37 The detector is a silicon chip that is composed of many individual photoactive sections called “picture elements”. These picture elements, or pixels, will build up charge as photons impinge on them. Individual pixels are of a size such that they can be used to measure individual wavelengths. Emission lines appear as points of light 800 nm 178 nm 38 740 nm 177 nm Readout Subarray - CID Intensit y 28 by 28 mm detector element Wavelength 39 What you get Full, continuous wavelength coverage; never miss an analyte 40 Power and flexibility • Rapid qualitative analysis • Ability to analyze for elements in the future without rerunning samples • Fingerprinting • Matrix or spectral subtraction 41 Instrument Components 5. Electronics The output from the detector is processed by a set of electronics. The electronics control the detector as well as collect the readings from the pixels The function of the electronics is to measure and process the output of the detector. 42 Instrument Components 6. Computer and Software The software, via a computer, controls and runs the instrument. Not only are measurements made but the other five components of the instrument are controlled and monitored by the computer and software, The function of the computer and software is to operate, monitor, and collect data from the instrument. 43 ICP Basics ICP Performance • Typical analysis time for ICP is ~2-3 minutes. This includes flush time, multiple repeats, printing, etc. (Analysis time is independent of the number of elements being determined) • Typical precision, amongst repeats within an analysis, is ~0.5% • Typical drift is ≤ 2% per hour • Typical detection limits are ~ 1-10 parts per billion 44