Biomaterial_Lecture 8

advertisement

BIOMATERIALS

ENT 311/4

Lecture 8

BIOLOGICAL TESTING OF BIOMATERIALS

Prepared by: Nur Farahiyah Binti Mohammad

Date: 25 th August 2008

Email : farahiyah@unimap.edu.my

Teaching Plan

METALLIC

BIOMATERIALS

DELIVERY

MODE

LEVEL OF

COMPLEXITY

COURSE OUTCOME

COVERED

Describe and assess in vitro and in vivo tissue compatibility of biomaterials in biological environment.

Lecture 

Knowledge

Repetition

Application

Analysis

Evaluation

Ability to explain and evaluate the biocompatibility of biomaterials utilized as implants or contact devices with human tissue.

2

1.0 Introduction

Biomaterials must be evaluated to determine if they are biocompatible and will function in a biologically appropriate manner in the in vivo environment.

Biomaterial will be evaluated under in vitro and in vivo conditions.

3

1.0 Introduction

Evaluation under in vitro conditions can provide rapid and inexpensive data on biological reaction.

However, the question must always be raised – will the in vitro test measure parameters relevant to what will occur in the much complex in vivo environment? → In vivo evaluation

4

1.0 Introduction

Protocol for biomaterial test provide by:

American Society for Testing Material

(ASTM)

International Standards Organization (ISO)

Government agencies, e.g., the FDA

5

2.0 In vitro Assessment of

Tissue Compatibility

The evaluation of biomaterials by method that use isolated, adherent cells in culture to measure cytotoxicity and biological compatibility.

Cytotoxicity means to cause toxic effect.

6

2.0 In vitro Assessment of

Tissue Compatibility

2.1 Background Concepts

(Things that observe)

2.1.1 Toxicity

Toxic material is defined as a material that release a chemical in sufficient quantities to kill cell either directly or indirectly through inhibition of key metabolic pathway.

Number of cells that are effected is an indication of dose and potency of the chemical.

7

2.0 In vitro Assessment of

Tissue Compatibility

2.1.2 Delivered and Exposure Doses

Delivered dose refers to the dose of the agent that is actually absorbed by the cell.

Exposure dose is the amount of cytotoxic agent delivered to the test system.

Example: if an animal is exposed to an atmosphere containing a noxious substance

(exposure dose), only a small portion of the inhaled substance will be absorbed and delivered to the internal organ organs and cell.

8

2.0 In vitro Assessment of

Tissue Compatibility

2.1.3 Solubility Characteristic

Test of dissolution of materials

Investigate either it stimulate the intended clinical application or may create desirable or undesirable degradation products.

9

2.0 In vitro Assessment of

Tissue Compatibility

2.2 ASSAY METHODS

Three primary in vitro cell culture cytotoxicity assay are:

2.2.1 Direct Contact Test

2.2.2 Agar Diffusion

2.2.3 Elution Test

10

2.0 In vitro Assessment of

Tissue Compatibility

2.2.1 Direct Contact

A near-confluent monolayer of L-929 mammalian fibroblast cells is prepared in a

35mm diameter cell culture plate.

The culture medium is removed and replaced with 0.8 ml of fresh culture medium.

Specimen of negative or positive controls and the test article are carefully placed in prepare culture and incubated for 24 hour.

11

2.0 In vitro Assessment of

Tissue Compatibility

Live cells adhere to the culture plate and are stained by the cytochemical stain.

Toxicity is evaluated by the absence of stained cells under and around the periphery of the specimen.

12

2.0 In vitro Assessment of

Tissue Compatibility

2.2.2 Agar Diffusion Test

A near-confluent monolayer of L-929 is prepared in a 60mm diameter plate.

The culture medium is removed and replaced with a culture medium containing

2% agar.

After the agar has solidified, specimen of negative and positive controls and the test article are placed on the surface of the same prepared plate and the culture incubated for at least 24 hours.

13

2.0 In vitro Assessment of

Tissue Compatibility

This assay also contain red stain in the agar mixture, which allows ready visualisation of live cells.

Healthy cells retain red stain.

Dead or injured cells do not retain neutral red and remain colourless.

Toxicity is evaluated by the loss of the stain under and around the periphery of the specimens.

14

2.0 In vitro Assessment of

Tissue Compatibility

2.2.3 Elution Test

An extract of the material is prepared by using 0.9% sodium chloride or serum-free culture medium.

The extract is placed on prepared nearconfluent monolayer of L-929 mammalian fibroblast cells.

Toxicity is evaluated after 48 hours.

Live or dead cells may be distinguished by the use of histochemical or vital stain as agar diffusion test method.

15

Direct contact

Agar

Diffusion

Advantages and Disadvantages of Cell Culture Methods

Advantages

Eliminate extraction preparation

Zone of diffusion

Target cell contact with material

Mimic physiological conditions

Standardize amount of test material or test indeterminate shapes

Can extend exposure time by adding fresh media

Eliminate extraction preparation

Zone of diffusion

Better concentration gradient of toxicant

Can test one side of a material

Independent of material density

Disadvantages

Cellular trauma if material moves

Cellular trauma with high density materials

Decreased cell population with highly soluble toxicants

Requires flat surface

Solubility of toxicant in agar

Limited exposure time

Risk of absorbing water form agar

16

Elution

Advantages and Disadvantages of Cell Culture Methods

Advantages

Separate extraction from testing

Dose response effect

Extend exposure time

Disadvantages

Additional time and step

17

3.0 In Vivo Assessment of

Tissue Compatibility

The goal of in vivo assessment of a biomaterial, prosthesis or medical devices is:

 to determine that the device performs as intended and presents no significant harm to the patient or user.

In vivo test for assessment of tissue biocompatibility are chosen to stimulate end-use applications.

18

3.0 In Vivo Assessment of

Tissue Compatibility

To facilitate the selection of appropriate tests, medical devices with their components of biomaterial can be categorized by:

The nature of body contact of the medical device

Duration of contact of the medical device

19

3.0 In Vivo Assessment of

Tissue Compatibility

Medical device categorization by tissue contact and contact duration

Tissue Contact

Surface devices

External communicating devices

Implant devices

Contact duration

Skin

Mucosal membrane

Breached or compromised surface

Blood path

Tissues/Bone/dentin communicating

Circulating blood

Tissue/bone

Blood

Limited, ≤ 24 hours

Prolonged, ≥ 24 hours and < 30 days

Permanent, >30 days 20

3.0 In Vivo Assessment of

Tissue Compatibility

1.

2.

3.

4.

5.

6.

7.

8.

In vivo test for tissue compatibility

Sensitization

Irritation

Intracutaneous reactivity

Systemic toxicity (acute toxicity)

Subcronic toxicity (subacute toxicity)

Genotoxicity

Implantation

Hemocompatibility

21

3.0 In Vivo Assessment of

Tissue Compatibility

9.

10.

11.

12.

13.

Chronic toxicity

Carcinogenicity

Reproductive and developmental toxicity

Biodegradation

Immune responses

22

3.0 In Vivo Assessment of

Tissue Compatibility

1. Sensitization

Sensitization test estimate the potential for contact sensitization to medical devices or materials.

Symptom of sensitization are often seen in skin.

Sensitization is a immune system response to chemicals

23

3.0 In Vivo Assessment of

Tissue Compatibility

2.

Irritation

Irritant test emphasize utilization of extracts of biomaterials to determine the irritant effects of potential leachables

Irritation is a local tissue inflammation response to chemical.

3.

Intracutaneous (intradermal) reactivity

Determine the localized reaction of tissue to intracutaneous injection of extracts of medical devices, biomaterials, or prosthesis in the final product form.

24

3.0 In Vivo Assessment of

Tissue Compatibility

4.

Systemic toxicity (acute toxicity)

Estimate the potential harmful effects in vivo on target tissues and organs away from the point of contact with either single or multiple exposure to medical devices or biomaterials.

Acute toxicity is considered to be the adverse effects occurring after administration test sample within 24 hours.

25

3.0 In Vivo Assessment of

Tissue Compatibility

5.

6.

Subacute toxicity

Focuses on adverse effect occuring after administration of a single dose or multiple doses of a test sample per day during a period of from 14 to 28 days.

Subcronic toxicity

 adverse effect occuring after administration of a single dose or multiple doses of a test sample per day given during a part of the life span, usually 90 days but not exceeding 10% of the life span of the animal.

26

3.0 In Vivo Assessment of

Tissue Compatibility

7.

Genotocity

Genocity tests are carried out if in vitro test results indicate potential genotoxicity.

The in vitro assay should cover three levels of genotoxicity effects:

DNA destruction

Gene mutation

Chromosomal aberrations (abnormality)

27

3.0 In Vivo Assessment of

Tissue Compatibility

8.

Implantation

Implantation test assess the local pathological effects on the structure and function of living tissue induced by a sample of a material or final product at site where it is surgically implanted.

28

3.0 In Vivo Assessment of

Tissue Compatibility

9.

Hemocompatibility

This test evaluate effect on blood and/or blood component by blood contacting medical devices or materials.

From the ISO standard prospective, five test categories for hemocompatibility evaluation:

Thrombosis (blood coagulation)

Coagulation

Platelets

Haematology

Immunology

29

Alternative scenario that can be applied for interpreting results of blood-material interaction assay

Alternate interpretation

Result implying poor blood compatibility

Evaluation method

Result implying good blood compatibility

Alternate interpretation

Many platelet adhere, but the platelets are not activated and form passivating natural biological layer on the surface

The thrombus layer forms a nonreactive natural biological film on the surface

Many adherent platelets

Surface coated with adherent thrombus

Measure platelet adhesion

Measure the mass of adherent thrombus

Released factors stimulated desirable endothelial cell growth

Extensive platelet granule release

Measure the platelet granule release

No adherent platelets

No adherent thrombus

No release

Platelets aggregate and embolize downstream

Thromus detaches and embolizes downstream.

Therefore it not seen on the surface

Release actually occurs but its diluted by the flowing blood

30

3.0 In Vivo Assessment of

Tissue Compatibility

10.

Carcinogenity

This test determine the tumorigenic potential of medical devices and biomaterial.

11.

Reproductive and Developmental

Toxicity

These test evaluate the potential effects of medical devices and biomaterials on reproductive function, embryonic development and prenatal and postnatal development.

31

3.0 In Vivo Assessment of

Tissue Compatibility

12.

Biodegradation

This test determine the effects of biodegradation materials and its biodegradation products on the tissue response.

This test focus on:

Amount of degradation during a period of time

The nature of the degradation products

The origin of the degradation product

Leachable in adjacent tissue and in distant organ.

32

3.0 In Vivo Assessment of

Tissue Compatibility

13.

Immune response

Immune response evaluation is not a component of the standards currently in vivo tissue compatibility assessment.

However, ASTM, ISO and FDA currently have working groups developing guidance documents for immune response evaluation.

Synthetic material are not generally immunogenic

However, immune response evaluation is necessary with modified natural tissue implant such as collagen.

33

Advantages and Limitation of

Biocompatibility Test

Test/Assay Advantages Limitations

In vitro tests Quick turnover (days), high throughput screening, standardized with appropriate protocols

Relevance to in vivo

In vivo test Provide multi-system interactions, more comprehensive than ioutcome inconsistent

In vitro test

Relevance to clinical use questionable, low turnover (week to months), high cost and low throughput, animal use concerns, outcome can be difficult to interpret

34

Download