BIO 224
Intro to Molecular and Cell Biology
Microscopes are tools frequently used in cell biology
Type of microscope used depends on the specimen
being viewed and magnification desired
Many types of microscopes are available
Light microscopes remain the basic tools of cell
biologists
Early microscopes used by Hooke and others were
simple light microscopes
Earliest microscopes were capable of up to 300 X
magnification
Early scientists viewed and documented an impressive
array of specimens
Bacteria
Human cells
Paramecia
Modern microscopes are capable of 1000X
magnification (mag)
Most cells range from 10um to 100um diameter,
easily seen with light scope
Some large organelles can be visualized as well
Fine structural details can’t be viewed
Requires higher resolution
Ability of a microscope to distinguish items separated by small
distances
More important than magnification
Limit of light microscopes about 0.2 um based on equation
Determined by two factors
Wavelength of visible light: λ
Numerical aperture of the lens: NA
Calculated by an equation: (0.61 λ) ÷ NA
Wavelength fixed at 0.5 for light microscope
NA is found by equation NA=η sinα
η is refractive index of the medium which light travels through between
the specimen and the lens (1.0 for air, 1.4 with immersion oil)
Angle α is half the width of cone of light collected by the lens
(maximum value is 90)
Highest possible value of NA is 1.4
Maximum theoretical resolution (also called resolving
power) had been achieved by late 1800s, no increases
expected
Several types used routinely
Bright-field microscopy is simplest
Light passes directly through the cell
Ability to see parts depends on absorption of visible
light by cell components
Usually requires use of dyes
Tissues usually fixed (preserved) prior to staining
Requires cells be killed
Histology labs routinely examine fixed and stained
tissues
Phase-contrast and differential interference-
contrast microscopy commonly used for living
cells
Achieve contrast due to variations in thickness in
cell parts
Speed of light drops as it passes through
intracellular structures, altering its phase
compared to light in cytoplasm
Changes in phase converted to differences in
contrast
Allows for improved images of live, unstained cells
Video and cameras can be added to enhance features
of microscopes
Has allowed visualization of movement of organelles
along microtubules
Location of certain molecules can be seen using labels
and dyes
Fluorescence microscopy is used for studying
intracellular distribution of molecules
Molecules of interest labeled with fluorescent dye
Used in living or fixed cells
Fluorescent dye absorbs light at one wavelength and
emits it at a second
Filters detect the wavelength of light the dye emits
Fluorescent labeled antibodies often used to detect
specific proteins
Green fluorescent protein (GFP) of jellyfish used to see
proteins inside living cells
Protein can be expressed in cells and viewed with
microscope
Other related proteins with blue, yellow, or red emissions
also available
Other methods developed to follow interactions of GFPlabeled proteins within living cells
Fluorescence recovery after photo-bleaching (FRAP)
Region of interest bleached by exposure to high-intensity
light
Unbleached molecules move into bleached region
Allows detection of rate of movement of protein within cell
Fluorescence resonance energy transfer
(FRET)
Two proteins coupled to fluorescent dyes, like two
GFP variants that emit different wavelengths of
light
Light emitted by one GFP variant excites the
second
If proteins labeled by the two GFP variants interact
within cells, the fluorescent molecules will be near
each other and light emission will occur
Conventional microscopy produces blurred and out-of-
focus images
Can be improved by deconvolution: a computer analyzes
images obtained by different depths of focus and generates
a sharper image
Confocal microscopy can be used to allow images of
increased contrast and detail
Obtained from fluorescence from a single point in specimen
Light produced by laser focused on specimen at a certain depth
Fluorescence emitted passes through pinhole aperture before
hitting detector
Allows only light emitted from plane of focus to reach detector
Results in sharper image from scanning across image
Series of images may be used for 3D image of sample
Multi-photon excitation microscopy
Alternative to confocal microscopy
Can be applied to living cells
Specimen illuminated with wavelength needing
absorption of two or more photons to excite fluorescent
dye
Photons only excite fluorescent dye at point in the
specimen where input beam is focused
Fluorescence only emitted from plane of focus
Allows for 3D resolution without need of pinhole
aperture
Minimal damage, allowing 3D image of living cells
More powerful than light microscopy
Developed in 1930s, first used on biological specimens
in 1940s and 1950s
Higher resolution due to wavelength of electrons
Practical limit of resolution for biological specimens is
1 to 2nm
Over 100X improvement on light scopes
Transmission electron microscopy (TEM)
Similar to observation of stained cells with bright field
scope
Specimens fixed and stained with heavy metal salts
Scatter electrons to provide contrast
Electron beam passed through specimen, focused to
form image on fluorescent screen
Electrons encountering heavy metal are deflected,
causing stained areas to be dark
Can use positive or negative staining
3D views can be obtained using electron tomography
Uses computer to generate 3D image from 2D scans over range of
directions
Metal shadowing is technique used to see surface of
structures or molecules
Evaporated metal sprayed on specimen from angle so surfaces
facing source are more heavily coated
Creates shadow effect, allowing 3D look
Freeze fracture used in studies of membrane structure,
usually interior faces
Specimens frozen in liquid N2 and fractured by knife blade
Followed by shadowing with Pt and dissolving biological
material with acid
Produces metal replica of sample surface
Freeze etching is variation the allowing visualization of
external surface of membranes along with interior
Scanning electron microscopy
Used to provide 3D image of cells
Surface of specimen coated with heavy metal
Beam of electrons used to scan across specimen
Electrons scattered or emitted from specimen surface
collected by detector
Resolution only about 10nm, restricted to whole cells
rather than macromolecules or organelles
Disclaimer
This workforce solution was funded by a grant awarded under the
President’s Community-Based Job Training Grants as
implemented by the U.S. Department of Labor’s Employment and
Training Administration. The solution was created by the grantee
and does not necessarily reflect the official position of the U.S.
Department of Labor. The Department of Labor makes no
guarantees, warranties, or assurances of any kind, express or
implied, with respect to such information, including any
information on linked sites and including, but not limited to,
accuracy of the information or its completeness, timeliness,
usefulness, adequacy, continued availability, or ownership. This
solution is copyrighted by the institution that created it. Internal
use by an organization and/or personal use by an individual for
non-commercial purposes is permissible. All other uses require
the prior authorization of the copyright owner.