CHAPTER 4 DC Meter. School of Computer and Communication Engineering, UniMAP Prepared By: Amir Razif b. Jamil Abdullah EMT 113: V-2008 1 4.0 DC Meters. 4.1 Introduction to Meters. 4.2 Analogue Meter 4.3 Introduction to DC Meters. 4.4 D’Arsonval Meter Movement in DC Meters. 4.5 Ayrton Shunt. 4.6 Voltmeter Loading Effect. 4.7 Ammeter Insertion Effect. 4.8 Ohmmeter. 2 4.1 Introduction to Meters. A meter is any device built to accurately detect and display an electrical quantity in a form readable by a human being. (i) Pointer (analogue). (ii) Series of lights (analogue). (iii) Numeric display (digital). In this chapter students will familiarized with the d’Arsonval meter movement, its limitations and some of its applications. Electrical meters; (i) DC, AC average quantities: -Voltmeter -Ammeter -Ohmmeter (ii) AC measurements: -Oscilloscope 3 Cont’d… A meter is any device built to accurately detect and display an electrical quantity in a form readable by a human being. In the analysis and testing of circuits, there are meters designed to measure the basic quantities of voltage, current, and resistance. Most modern meters are "digital" in design, meaning that their readable display is in the form of numerical digits. Older designs of meters are mechanical in nature, using some kind of pointer device to show quantity of measurement. The first meter movements built were known as galvanometers, and were usually designed with maximum sensitivity in mind. Figure 4.1: Galvanometer. Figure 4.2: Voltmeter 4 Cont’d… Galvanometer A very simple galvanometer may be made from a magnetized needle (such as the needle from a magnetic compass) suspended from a string, and positioned within a coil of wire. Current through the wire coil will produce a magnetic field which will deflect the needle from pointing in the direction of earth's magnetic field. An antique string galvanometer is shown in Figure 4.1. The term "galvanometer" usually refers to any design of electromagnetic meter movement built for exceptional sensitivity, and not necessarily a crude device such as that shown in Figure 4.1. Practical electromagnetic meter movements can be made now where a pivoting wire coil is suspended in a strong magnetic field, shielded from the majority of outside influences. Such an instrument design is generally known as a permanent-magnet, moving coil, or PMMC movement. 5 4.2 Analogue Meters. The analogue meters are mostly based on moving coil meters. The typical structure consists of a wire wound coil placed between two permanent magnets, Figure 4.3. When current flows through the coil in the presence of a magnetic field, a force is exerted on the coil; F = Bil This force is directly proportional to current flowing in the coil. If the coil is free to rotate, the force causes a deflection of the coil that is proportional to the current. By adding an indicator (e.g. needle) and a display, the level of current can be measured. Figure 4.3: Analogue Meter 6 Cont’d… For a given meter, there is a maximum rated current that produces full-scale deflection of the indicator; FSD rating. By adding external circuit components, the same basic moving coil meter can be used to measure different ranges of voltage or current. Most meters are very sensitive. That is, they give full-scale deflection for a small fraction of an amp (A) for example a typical FSD current rating for a moving coil meters is 50 μA, with internal wire resistance of 1 kΩ. With no additional circuitry, the maximum voltage that can be measured using this meter is 50 x 10-6x 1000W = 0.05V. Additional circuitry is needed for most practical measurements. 7 4.3 Introduction DC Meters. The meter movement will have a pair of metal connection terminals on the back for current to enter and exit. Meter; (i) Polarity sensitive. (ii) Polarity-insensitive. Most meter movements are polarity-sensitive, one direction of current driving the needle to the right and the other driving it to the left. Some meter movements are polarity-insensitive, relying on the attraction of an un-magnetized, movable iron vane toward a stationary, current-carrying wire to deflect the needle. Such meters are ideally suited for the measurement of alternating current (AC). A polarity-sensitive movement would just vibrate back and forth uselessly if connected to a source of AC. 8 Cont’d… An increase in measured current will drive the needle to point further to the right. A decrease will cause the needle to drop back down toward its resting point on the left. Most of the mechanical meter movements are based on electromagnetism ; electron flow through a conductor creating a perpendicular magnetic field, A few are based on electrostatics; the attractive or repulsive force generated by electric charges across space. 9 Cont’d… (a) Permanent Magnet Moving Coil (PMMC). Figure 4.4: Permanent Magnet Moving Coil (PMMC) Meter Movement. In the PMMC-type instruments, Figure 4.4. Current in one direction through the wire will produce a clockwise torque on the needle mechanism, while current the other direction will produce a counter-clockwise torque. 10 Cont’d… (b) Electrostatic Meter Movement. In the electrostatics, the attractive or repulsive force generated by electric charges across space, Figure 4.5. This is the same phenomenon exhibited by certain materials; such as wax and wool, when rubbed together. If a voltage is applied between two conductive surfaces across an air gap, there will be a physical force attracting the two surfaces together capable of moving some kind of indicating mechanism. That physical force is directly proportional to the voltage applied between the plates, and inversely proportional to the square of the distance between the plates. Figure 4.5: Electrostatic Meter Movement. 11 Cont’d… The force is also irrespective of polarity, making this a polarityinsensitive type of meter movement. Unfortunately, the force generated by the electrostatic attraction is very small for common voltages. It is so small that such meter movement designs are impractical for use in general test instruments. Typically, electrostatic meter movements are used for measuring very high voltages; many thousands of volts. One great advantage of the electrostatic meter movement, however, is the fact that it has extremely high resistance, whereas electromagnetic movements (which depend on the flow of electrons through wire to generate a magnetic field) are much lower in resistance. 12 Cont’d… Some D'Arsonval movements have full-scale deflection current ratings as little as 50 µA, with an (internal) wire resistance of less than 1000 Ω. This makes for a voltmeter with a full-scale rating of only 50 millivolts (50 µA X 1000 Ω). Figure 4.6: Voltmeter. 13 4.4 D’Arsonval Meter Movement in DC Meter. The basic d’Arsonval meter movement has only limited usefulness without modification. By modification on the circuit it will increase the range of current that can be measured with the basic meter movement. This is done by placing the low resistance in parallel with the meter movement resistance Rm. The low resistance shunt (Rsh) will provide an alternate path for the total meter current I around the meter movement. The Ish is much greater than Im. Where Rsh = resistance of the shunt Rm = internal resistance of the meter movement (resistance of the moving coil) Ish = current through the shunt Im = full-scale deflection current of the meter movement 14 Figure 4.7: D’Ársonval Meter Movement I = full-scale deflection current for the ammeter Used in Ammeter Circuit Cont’d… The voltage drop across the meter movement is Vm = ImRm Since the shunt resistor is in parallel with the meter movement, the voltage drop across the shunt is equal to the voltage drop across the meter movement. That is, Vsh = Vm The current through the shunt is equal to the total current minus the current through the meter movement:, Ish = I – Im Knowing the voltage across, and the current through, the shunt allows us to determine the shunt resistance as Rsh = Vsh/Ish = ImRm/Ish = (Im/Ish)(Rm) = Im/(I – Im)*Rm W 15 Example 4.1: D’Arsonval Movement. A D'Arsonval meter movement having a full-scale deflection rating of 1 mA and a coil resistance of 500 Ω: Ohm's Law (E=IR), determine how much voltage will drive this meter movement directly to full scale, Solution: E I *R E (1mA) * (500W) E 0.5V . 16 Example 4.2: D’Arsonval Meter. Calculate the value of the shunt resistance required to convert a 1-mA meter movement, with a 100 W internal resistance, into a 0 - to 10 mA ammeter. Solution: Calculate Vm. Vm I m Rm 1mA * 100W 0.1V Vm is in parallel with Vsh. Vsh Vm 0.1V KCL I sh I I m 10mA 1mA 9mA . Rsh Vsh 0.1V 11.11W I sh 9mA 17 4.5 Ayrton Shunt. The purpose of designing the shunt circuit is to allow to measure a current I that is some number n times larger than Im, Figure 4.8. The number n is called a multiplying factor and relates total current and meter current as the Ayrton Shunt. I nI m Substituting for I in previous equation, yields Rm I m Rm Rsh (nI I m ) (n 1) Advantage: (i) it eliminates the possibility of the meter movement being in the circuit without any shunt resistance. (ii) May be used with a wide range of meter movements. Figure 4.8: Aryton Shunt. 18 Cont’d… The individual resistance values of the shunts are calculated by starting with the most sensitive range and working toward the least sensitive range. The shunt resistance is, Rsh Ra Rb Rc On this range the shunt resistance is equal to Rsh and can be computed by the equation, R Rsh m (n 1) The equation needed to compute the value of each shunt, Ra, Rb, and Rc, can be developed from the circuit in Figure 4.8. Since the resistance Rb + Rc is in parallel with Rm + Ra, the voltage across each parallel branch should be equal and can be written as VRb VRc VRa VRm 19 Cont’d… In current and resistance terms we can write or ( Rb Rc )( I 2 I m ) I m ( Ra Rm ) I2(Rb + Rc) - Im(Rb + Rc) = Im[Rsh-(Rb + Rc)+Rm] Multiplying through by Im on the right yields I2(Rb + Rc) - Im(Rb + Rc) = ImRsh- Im(Rb + Rc)+ImRm This can be rewritten as Rb+ Rc = Im (Rsh+ Rm)/I2 Having already found the total shunt resistance Rsh, we can now determine Ra as Ra = Rsh – (Rb + Rc) The current I is the maximum current for the range on which the ammeter is set. The resistor Rc can be determined from Rc = Im(Rsh+ Rm)/I3 The resistor Rb can now be computed as, Rb = (Rb + Rc) – Rc 20 Example 4.3: Aryton Shunt. Compute the value of the shunt resistors for the circuit below. I3 = 1A, I2 = 100 mA, I1 = 10 mA, Im = 100 uA and Rm = 1K Ohm. Solution: The total shunt resistance is found from Rsh Rm 1KW 10.1W n 1 100 1 This is the shunt for the 10 mA range. When the meter is set on the 100-mA range, the resistor Rb and Rc provide the shunt . The total shunt resistance is found by the equation. Rb Rc I m ( Rsh Rm ) I2 (100uA) * (10.1W 1KW) 1.01W 100mA 21 Cont’d… The resistor Rc , which provides the shunt resistance on the 1-A range can be found by the same equation, however the current I will now be 1A. I (R R ) Rc m sh m I3 (100uA) * (10.1W 1KW) 0.101W 1A The resistor Rb is found from the equation below; Rb ( Rb Rc ) Rc 1.01W 0.101W 0.909W The resistor Ra is found from; Ra Rsh ( Rb Rc ) 10.1W (0.909W 0.101W) 0.909W Verify the above result. Rsh Ra Rb Rc 9.09W 0.909W 0.101W 10.1W . 22 Example 4.4 (T2-2005): Aryton Shunt. Figure below is an Aryton Shunt circuit. Given that R1 = 0.5 W, R2 = 6.5 W, R3 = 55.5 W, Rm = 1 kW, Im=100 mA and n = 16. Calculate the value of, I1, I2, I3 and I4. Solution: Rsh Rm 1KW 66.67W n 1 16 1 Find Rsh R1 R2 R3 I3 I3 I m ( Rsh Rm ) I3 I m ( Rsh Rm ) R1 R2 R3 100 mA(66.67W 1KW) 0.5W 6.5W 55.5W I 3 1.707 mA Can be easily derived 23 R1 R2 I2 I2 I m ( Rsh Rm ) I2 I m ( Rsh Rm ) R1 R2 100 mA(66.67W 1KW) 0.5W 6.5W I 2 15.238mA I2 I3 R1 I m ( Rsh Rm ) I1 I1 I m ( Rsh Rm ) R1 100 mA(66.67W 1KW) 0.5W I 1 313.33mA I1 I4 R1 R2 R3 R4 I4 I m ( Rsh Rm ) I4 I m ( Rsh Rm ) R1 R2 R3 R4 100 mA(66.67W 1KW) 0.5W 6.5W 55W 4.17W I 4 1.60mA .. Rsh R1 R2 R3 R4 66.67W 24 4.6 Voltmeter Loading Effect. (a) D’Arsonval Meter Movement Used in a DC Voltmeter. The basic d’Arsonval meter movement can be converted to a dc voltmeter by connecting a multiplier Rs in series with the meter movement, as in Figure 4.10 below. The multiplier will extend the voltage range of the meter to limit current through the d’Arsonval meter movement to a full scale deflrction current. The value of the multiplier resistor can be found by determine the sensitivity. Sensitivity 1 (W / V ) I fs Figure 4.10: The d’Arsonval meter movement used in the DC voltmeter. 25 Cont’d… Voltmeter Design. Consider a moving coil meter with FSD rating of 1 mA and coil resistance, Rc, of 500 Ω. The maximum voltage required to produce FSD is 0.5 V. The voltage range is increased by adding a series resistor, Figure 4.11: Voltmeter. The voltage that can be applied to the – and + terminals before FSD current flows, is then increased to: VFSD I FSD ( Rc Rm ) Rm is called a multiplier resistor because it multiplies the working range of the meter. 26 Cont’d… For a given required FSD voltage, say VFSD, the multiplier resistance, Rm, is chosen as: VFSD Rc Rm I FSD For example, to provide a voltmeter with FSD reading of 10 V with the given meter (IFSD = 1 mA, Rc= 500 Ω): V 10V Rm FSD Rc 500W 9.5kW 1mA I FSD With exactly 10 V applied, there will be exactly 1 mA of current flowing, thereby producing full-scale deflection. There is only the maximum allowed voltage of 0.5V dropped across the moving coil meter. The scale of the meter must be changed to indicate the new range of the circuit. 27 Example 4.5: Voltmeter Sensitivity. Calculate the sensitivity of 100 uA meter movement which is to be used as dc voltmeter. Solution. The sensitivity is compute as Sensitivity, S 1 1 10kW / V I fs 100mA . The unit of sensitivity is expressed as the value of multiple resistance for 1-V range. To calculate the value of the multiplier for voltage range greater than 1V, multiply the sensitivity by the range and subtract the internal resistance of the meter movement. Rs S * Range Internal Re sis tan ce 28 Example 4.6: Sensitivity Voltmeter Range. Calculate the value of multiplier resistance on the 50-V range of a dc voltmeter that used a 500-uA meter movement with an internal resistance of 1kW. Solution. Figure 4.12: The Volmeter. The sensitivity of 500uA movement is, Sensitivity, S 1 1 kW 2 I fs 500mA V The value of the multiplier Rs is calculated as, Rs S * Range Internal Re sis tan ce . Rs 2kW * 50V 1kW 99kW V 29 Cont’d… (b) Voltmeter Loading Effect. When the voltmeter is used to measure the voltage across a circuit component the voltmeter circuit itself is in parallel with the circuit component. The parallel combination of two resistor is less than either resistor alone. The resistance seen by the source is less with the voltmeter connected than without. The voltage across the source is less when the voltmeter is connected. This effect is called voltmeter loading. The resulting error is called loading error. 30 Example 4.7: Voltmeter Loading Effect. In the voltmeter loading effect experiment, the two different voltmeters are used to measure the voltage across resistor RB in the circuit Figure 4.13 below, The meters are as follow, Meter A: S = 1 k W /V, Rm = 0.2 k W , range = 10V Meter B: S = 20 k W /V, Rm = 1.5 k W , range = 10V Calculate, (i) Voltage across RB without any meter connected across it. (ii) Voltage across RB when meter A is used. (iii) Voltage across RB when meter B is used. (iv) Error in voltmeter readings. Solution. Figure 4.13: The Voltmeter Loading Effect. 31 (i) Voltage across RB without any meter connected across it. VRB E RB 5kW (30V ) 5V R A RB 25kW 5kW (ii) Voltage across RB when meter A is used. RS S * Range Rm 1kW (10V ) 1kW 9kW V The parallel combination of RB and meter A, Re1 ( RB )( RTA ) (5kW)(10kW) 3.33kW RB RTA 5kW 10kW Therefore the voltage reading obtained with meter A, determine by the voltage divider equation is VRB Re1 3.33kW E (30V ) 3.53V Re1 R A 3.33kW 25kW 32 (iii) Voltage across RB when meter B is used. The total resistance that meter B presents to the circuit is RTB S * Range 20kW (10V ) 200kW V The parallel combination of RB and meter B is Re 2 R B RTB (5kW)(200kW) 4.88kW R B RTB 5kW 200kW the voltage reading obtain with meter B, determine by use of the voltage divider equation Re 2 4.88kW V E (30V ) 4.9V RB Re 2 R A 4.88kW 25kW (iv) Error in voltmeter readings. Voltmeter A error ` 5V 3.53V *100% 29.4% 5V Voltmeter B error 5V 4.9V *100% 2% . 5V 33 4.6 Ammeter Insertion Effect. We frequently overlook the error caused by inserting an ammeter in a circuit to obtain a current reading. All ammeters contain some internal resistance. By inserting the ammeter in the circuit means increase the resistance of the circuit and result in reducing current in the circuit. Refer to the circuit in Figure 4.14, Ie is the current without the ammeter. Suppose that we connect the ammeter in the circuit (b), the current now becomes Im due to the additional resistance introduced by the ammeter. Figure 4.14: (a) Expected Current Value in a Series Circuit (b) Series Circuit with Ammeter. 34 Cont’d… From the circuit; Placing the meter in series result in; E Ie R1 E Im R1 Rm Divide the above equations yields; Im R1 Ie R1 Rm Insertion error, Im 1 *100% Ie 35 Example 4.8: Ammeter Insertion Effects. A current meter that has an internal resistance 78 Ohm is used to measure the current through resistor Rc in Figure 4.14. Determine the percentage of error of the reading due to ammeter insertion. Solution. Look back into the circuit from terminal X and Y. 36 Solution. The Thevenin equivalent resistance. Rth Rc Ra Rb Ra Rb 1KW 0.5 KW 1.5 KW The ratio of meter current to the expected current is, Im R1 1.5KW 0.95 Ie R1 rm 1.5KW 78W Solving for Im yields, Insertion I m 0.95I e Im error 1 *100% 5.0% Ie . 37 4.7 Ohmmeter. The d’Arsonval meter movement can be used with the battery and resistor to construct a simple ohmmeter. Figure 4.15 is the basic ohmmeter circuit, I fs E R z Rm Introduce Rx between point X and Y so that we can calculate the value of resistance. E I R z Rm R x Figure 4.15: Basic Ohmmeter Circuit. 38 Cont’d… E /( Rz Rm Rx ) ( R z Rm ) I I fs E /( Rz Rm ) ( R z Rm R x ) P represent the ratio of the current I to the full scale deflection P ( R z Rm ) I I fs ( Rz Rm Rx ) Figure 4.16: Basic Ohmmeter Circuit with Unknown Resistor Rx Connected Between. 39 Example 4.9: Ohmmeter. A 1mA full-scale deflection current meter movement is to used in an ohmmeter circuit. The meter movement has an internal resistance, Rm, of 100 Ohm, and a 3-V battery will be used in the circuit. Mark off the meter face for reading resistance. Solution. Value of Rz, which will limit current to full-scale deflection is, Rz E Rm I fs 3V Rz 100Ohm 2.9 KOhm 1mA Value of Rz, with 20% full-scale deflection is, Rx R z Rm ( R z Rm ) P 2.9 KW 1.0 KW (2.9 KW 1.0 KW) 0.2 12 KW 40 Cont’d… Value of Rz, with 40% full-scale deflection is, Value of Rz, with 50% full-scale deflection is, Value of Rz, with 75% full-scale deflection is, The ohmmeter is nonlinear due to the high internal resistance of the ohmmeter. Rx Rx Rx R z Rm ( R z Rm ) P 3 KW (3KW) 0.4 4.5 KW R z Rm ( R z Rm ) P 3 KW (3KW) 0.5 3 KW R z Rm ( R z Rm ) P 3 KW (3KW) 0.75 1KW 41