Running head: THE RELIABILITY OF MANUAL MUSCLE TESTING THROUGH THE COMPARISON OF THE ISOKINETIC BIODEX TESTING ABOUT THE KNEE JOINT The Reliability of Manual Muscle Testing Through the Comparison of Isokinetic Biodex Testing About the Knee Joint tJamie Hale, Samantha Namm, Jane Nguyen, Jeren Marquecho-Riley, Onofre Sunga California State University of San Marcos Running head: THE RELIABILITY OF MANUAL MUSCLE TESTING THROUGH THE COMPARISON OF THE ISOKINETIC BIODEX TESTING ABOUT THE KNEE JOINT Abstract Manual muscle testing (MMT) is often used to test an individual’s muscle strength or the ability of a specific muscle to perform against resistance. MMT may not be reliable and could generate more antagonist activity. The purpose of this study was to compare EMG activity of knee extension and flexion of the vastus medialis, vastus lateralis, rectus femoris, and biceps femoris. EMG is commonly used to help diagnose neuromusculoskeletal dysfunction. The Biodex is currently the gold standard for measuring muscle strength. Subjects were 12 college students between the ages of 18-30 with no specific requirements. Surface EMG was placed on 4 muscles listed above and subjects performed a total of 12 tests (3 manual extension, 3 manual flexion, 3 biodex extension, and 3 biodex flexion) all about the knee joint. Using a T-test ran with the standard deviations collected over a 2000 point series we found a significant difference for one muscle, (Vastus Lateralis) during knee flexion (P=.029,P< .05). It was predicted these values may occur as these are the antagonist muscles for knee flexion and the leg may contract these muscle more during manual testing in an attempt to stabilize the leg. It was hypothesized that the muscle manual test will be less reliable when compared to the isokinetic dynamometer (Biodex) and will elicit greater antagonist muscle activity. Introduction Manual muscle testing (MMT) is often used to assist in diagnosing neuromusculoskeletal dysfunction. Mild symmetrical muscle weakness is still difficult to detect despite MMT common clinical usage due to that age, weight, height, and gender affect strength (Schmitt et al., 2008). Three distinct muscles of the leg—such as the quadriceps and hamstrings—are tested (Schmitt et al., 2008). There are several limitations to MMT. One noted is if the tester maintains consistency in the starting position. If this angle varies by more than a few degrees or half of an inch, the subject’s line of force will alter and substitution of coactive muscle functions may take place resulting in skewed EMG measurements (Schmitt et al., 2008). Another is if the tester applies force too quickly and overpowers the subject. MMT assesses strength so the tester must apply an increasing rate of force until the muscle begins to give (Schmitt et al., 2008). An experienced tester will be able to sense this difference. The test is over once the muscle is in motion. Muscle activity during maximal isokinetic contraction is measured using electromyography (EMG). The antagonist EMG measurements range from 10%-50% to nearly 60% of the muscle when acting as an agonist during contraction (Kellis et al., 1997). However, because there is no linear relationship between EMG activity and muscle moment, this places a limitation on using normalized EMG measurements as a representation of antagonist muscle function. To keep the knee joint stabilized during isokinetic, concentric knee extensions, hamstrings and quadriceps coactivation must take place (Kellis et al., 1997). Running head: THE RELIABILITY OF MANUAL MUSCLE TESTING THROUGH THE COMPARISON OF THE ISOKINETIC BIODEX TESTING ABOUT THE KNEE JOINT In a comparative study of isokinetic dynamometry and MMT of ankle dorsal and plantar flexors and knee extensors and flexors, 108 patients’ strength, who had either diabetes or alcoholic liver cirrhosis, were compared to the control group, who were made up of 90 healthy individuals (Andersen et al., 1997). The comparison found that MMT underestimated the severity of muscle weakness than the isokinetic dynamometer (Andersen et al., 1997). This may be a result of several limitations. Conversely, another research study found a moderate to strong Pearson/Spearman correlation in the strength measured by MMT (Bohannon, 2004). Yet, the sensitivity of MMT is lacking. It was found that with as little as 50% of muscle strength measured, subjects were classified as normal which leads to that MMT cannot clinically diagnose damage to the muscle because it cannot distinguish differences (Bohannon, 2004). MMT can be used for screening deficits in strength. MMT used in conjunction with the dynamometer proved to be very effective in identifying strength deficits. The isokinetic dynamometer is generally accepted as the gold standard in measuring position, torque and velocity. A study was done to assess the dynamometer’s mechanical reliability and validity. It was found that the concentric velocity was valid up to 300 deg/sec and that this equipment can consistently reproduce measurements across trials (Drouin et al., 2004). With these two tests so widely accepted and used, the question of how accurate is muscle manual testing arises. We hypothesize that the muscle manual test will be less reliable when compared to the isokinetic dynamometer, and will elicit greater antagonist muscle activity. Methods Subjects were 12 college students between the ages of 18-30. 6 males and 6 females,during a manual muscle test, performed 3 trials each with maximum effort for extension upon the knee joint, and 3 trials for a maximum effort for flexion upon the knee joint. Surface electromyographies (EMG) were placed on the vastus medialis, vastus lateralis, rectus femoris, and biceps femoris carefully in hopes of gathering sufficient muscle activity during each exercise. Subjects were then asked to do the same procedure while strapped into the isokinetic biodex system 3 machine for maximum effort through extension and flexion upon the knee joint. The Biodex was set at a rate of thirty degrees per second for each movement involved, flexion and extension. There were a total of 12 movements during this effort on achieving optimal muscle activity, for both tests with extension and flexion at three trials each. The first two trials on each muscle exercise were practice runs for the subjects to understand the effects of each movement and the third was recorded for our study. All subjects were informed of experimental procedures and gave their consent before participating. All procedures were performed within the span of an hour on the same day, so no second visits were held to promote variability in their muscles. Running head: THE RELIABILITY OF MANUAL MUSCLE TESTING THROUGH THE COMPARISON OF THE ISOKINETIC BIODEX TESTING ABOUT THE KNEE JOINT Equipment Isokinetic biodex system 3 machine with over the shoulder straps, waist strap, right legged arm with ankle strap. Surface EMG, with 2 pads on each muscle listed as well as a ground surface emg placed on the patella. Consistent manual muscle tester with knee at an approximate 45 degree angle start position for both extension and flexion. Data Interpretation The data from the EMG machine for the 4 muscles gave the muscle contraction from the maximum force output for each individual muscle during each individual trial. The first trial was considered a warm up trial, and the second trial was used as a practice trial, and the third trial for each movement was used towards the data. For each subject, the data was taken from the burst of each muscle during the 2 different movements on both the manual test and biodex test. The standard deviation was taken from the moment of greatest muscle force for about 2 seconds at the rate of 1000 Hz per second (2000 Hz). Then a T-test was applied between the Biodex extension and the manual extension as well as the Biodex flexion to the manual flexion testing. This gave a value, which correlated the data against each other giving a percentage of the likelihood that in those 2 separate tests the muscles acted in the same way for each muscle in all subjects. A 5% difference between the manual muscle test and the biodex test for each muscle is ideal in the support of the null hypothesis. This determined if there was a small percentage that the muscles acted in the same way for both tests. If there was a less than five percent difference then the tests were different, accepting the null hypothesis. If there was a greater than five percent difference the tests were relatively the same, declining the null hypothesis that they are different. The mean was then taken for each different muscle and each separate test in each different subject. This data collected was put into a bar graph to determine the correlation between the 2 tests as well as a series error bar to illustrate a graphical representation of the variability of data and use of graphs to indicate the error in a reported measurement. See graphs for demonstration. Results Data from the third trial of each subject was used to find the standard deviations. The standard deviations were calculated by using data over a period of 2000 series points. EMG data was collected at 1000 HZ for both the biodex test and the manual test. The means of EMG (taken in volts) indicating muscle contraction was taken for 4 muscles. The mean volts recorded for the Vastus Medialis was similar for Biodex extension and manual extension (M=.144, M=.148). The mean found for biodex flexion and for the manual test for this muscle was also slightly similar (M= .031,M=.021). Vastus Lateralis averages were the same for extension (M= .148 and M=.148) but showed differences showed in flexion (M=.027, M=.012). Means for Rectus Femoris Running head: THE RELIABILITY OF MANUAL MUSCLE TESTING THROUGH THE COMPARISON OF THE ISOKINETIC BIODEX TESTING ABOUT THE KNEE JOINT were similar for extension (M=.167,M=.183) while flexion showed slight differences (M=.027, M= .012). Lastly, averages for Biceps Femoris muscles were similar (M=.027,M=.024) for extension and flexion (M=172,M= .168). Using a T-test ran with the standard deviations collected over the 2000 series points we found only a significant difference between the two test types for one muscle, the Vastus Lateralis when assessed for knee flexion (P=.029,P< .05). The P-value for the Vastus Medialis was close to showing a significant difference at (P=.07 ,P< .05). Discussion The purpose of this study was to compare EMG activity of knee extension and flexion of the vastus medialis, vastus lateralis, rectus femoris, and biceps femoris. The results suggest that manual testing is as reliable as the Biodex testing when performing knee flexion and extension. The manual and Biodex tests revealed similar muscle activation output. When MMT and the Biodex EMG output data’s mean and standard deviation values were compared, the Biodex testing numbers were slightly higher, showing that the Biodex testing showed more contraction activity when flexion and extension were assessed. During knee extension in both tests, the biceps femoris showed a low EMG score. It was hypothesized that the muscle manual test will be a less reliable method of testing when compared to the isokinetic dynamometer (Biodex), and will elicit greater antagonist muscle activity because of instability. This study showed that the vastus lateralis and vastus medialis during knee flexion for the two different conditions, did in fact have more contraction due to instability from the results having a significant difference value of (P=.029,P< .05) and (P=.07 ,P< .05). However, after analyzing the mean for contractions in these two muscles it was recognized that this significant difference came with the increase contraction of the muscles during biodex testing rather than the manual tests proving the hypothesis incorrect. All other results for the other 3 muscles during different conditions of this study did not support the hypothesis because the “P-values” were above a 5% difference. These findings are applicable to an individual who may have been afflicted by an injury or some sort of deficiency. By having a MMT screening run, it will be able to determine which muscle was most affected. MMT is also easily accessible rather than its counterpart, the Biodex for the reason that they are expensive and not portable. MMT is considered a reliable screening test from this study, but it has some limitations. One limitation is if the tester maintained consistency in the start position for each trial. If the start position angle differed by more than a few degrees than the subject’s line of force would also differ, making the muscles that are supposed to perform work change and skew the EMG data (Schmitt et al., 2008). Some disadvantages in the usage of Surface EMG include only being able to record superficial muscles and large muscle groups and may not be selective enough for detailed or smaller muscles. Surface EMG is more prone to crosstalk, the electrical activity from neighboring muscles, as well as movement artifact and contact pressure fluctuations that can hinder the impulse of an individual Running head: THE RELIABILITY OF MANUAL MUSCLE TESTING THROUGH THE COMPARISON OF THE ISOKINETIC BIODEX TESTING ABOUT THE KNEE JOINT muscle unit. This may have affected some results. The biodex chair was used for both tests to reduce a knee angle variable that may affect contraction quantity. The same tester was also used for all manual tests to reduce differences. Another problem that may affect the results was that an isometric test was done during manual testing where a concentric and eccentric test was performed with the biodex providing resistance throughout full range of motion. The stable nature of the biodex allowed the antagonist muscles to better reflect the co-contraction that naturally occurs in the leg during flexion and extension, thus would seem to be more reliable in determining force output of the leg as a whole. However, this study would require more testing with the use of a force sensor to determine accuracy of force output of the upper leg muscles. The rest periods between trials also could have increased from 2 minutes to 5 minutes to ensure full recovery. The slight differences in muscle activity may have been due to electrode placement or a learning effect created by subjects being unfamiliar with the biodex machine and thus exerting more than exerted on a human tester . Overall, the manual test compared to the biodex test resulted in similar means per muscle with only a slight difference seen in antagonist muscle groups during flexion. The study did not show a significant amount of difference in muscle contraction activity and the manual test would appear accurate disregarding overall force output. A test determining force output would most likely be required to develop a rehabilitative program for an injured subject but a manual test seems to be indicative of a difference in specific muscle contraction. The results that were in this study collected were unsupportive of the hypothesis, proving that a manual muscle test with surface EMG is a valid form for assessing muscle strength and ability. Running head: THE RELIABILITY OF MANUAL MUSCLE TESTING THROUGH THE COMPARISON OF THE ISOKINETIC BIODEX TESTING ABOUT THE KNEE JOINT Works Cited Andersen, H., Jakobsen, J. (1997). A Comparitive Study of Isokinetic Dynamometry and Manual Muscle Testing of Ankle Dorsal and Plantar Flexors and Knee Extensors and Flexors. Bohannon, R. (2004, Nov. 4). Manual Muscle Testing: Does It Meet the Standards of an Adequate Screening Test?. Drouin, J., Valovich-McLeod, T., Shultz, S., Gansneder, B., Perrin, D. (2004). Reliability and Validity of the Biodex System 3 Pro Isokinetic Dynamometer Velocity, Torque, and Position Measurements. Kellis, E., Baltzopoulos, V. (1997). The Effects of Antagonist Moment On the Resultant Knee Joint Moment During Isokinetic Testing of the Knee Extensors. Schmitt, W., Cuthbert, S. (2008, June 2). Common Errors and Clinical Guidelines for Manual Muscle Testing: “The Arm Test” and Other Inaccurate Procedures.