HPP Activity 55v1 Give me an A(-scan). Give a B(-scan). What does that spell? Exploration We are now in a position to understand how an ultrasound image is created. There are several kinds of ultrasound imaging systems. The earliest device, which also forms the basis for later systems, is called the A-scan. The "A" stands for amplitude. Let's get an idea of how the A-scan works by considering an echo measuring experiment. You may have done this experiment in a previous unit. The set-up is shown below. Figure 1. This shows an air-filled tube, closed at one end, a microphone, and an oscilloscope. We can snap our fingers and record the snap and the echo, after the sound pulse travels the length of the tube and back. Try the simulation out. GE 1. The figure below shows an example of the simulated oscilloscope display from an experiment done at room temperature. The vertical axis shows the amplitude of the signal and the horizontal axis shows the time (each square on the display is 2 milliseconds wide). Activity Guide 2010 The Humanized Physics Project Supported in part by NSF-CCLI Program under grants DUE #00-88712 and DUE #00-88780 HPP Activity 55v1 2 1. What change could you make to this picture so that it can show the object position encountered by the sound pulse? 2. How would the display change if a tube 2.4 m long (50% longer) were used instead. 3. Suppose the tube had been filled with an additional barrier, at 0.80 [m] from the end. If the barrier allowed transmission of sound, as well as reflection, how might the oscilloscope output look when the echo experiment is done? Invention The basic idea behind the A-scan is to use timing data on sound echoes to fix the distance of objects from the ultrasound transducer. By assuming a constant speed of sound, we can calculate the distance to an object by using the time between the initial pulse (the snap in this case) and the echo. If t is the time between sound emission and detection of the echo, and v is the speed of sound, then the distance, D, to the object is D v t 2 (1) Equation (1) can be used to convert the amplitude versus time picture given on the oscilloscope to an amplitude versus position picture. An A-scan instrument would do this calculation automatically. The A-scan gives a one-dimensional picture of what is in front of the transducer. It tells the observer where there are interfaces along a line drawn out from the transducer. Activity Guide 2010 The Humanized Physics Project HPP Activity 55v1 3 Application GE 2. Suppose you must design instrumentation for making an A-scan image of the eye. 1.What would you assume the speed of sound is in the eye? Explain. 2.Will your assumption produce any errors in the A-scan image? Hint: Will the speed of sound be the same in all tissues in the eye? If so, make an order of magnitude estimate of the (systematic) error introduced. Show your work/ 3. Obtain some anatomical information on the eye and estimate how much time the A-scan instrument must accommodate (wait for an echo) to obtain an image of the various interfaces inside the eye? Show your work. Invention Usually a physician would prefer at least a two-dimensional picture of a body's internal structure, rather than the simple one-dimensional picture of interface locations along a line that the A-scan gives. If the transducer that produces the A-scan can be moved, then two-dimensional information can be constructed. A B-scan instrument can do this. The "B" stands for brightness. The amplitude of an echo from the A-scan is converted into the brightness of a point on a CRT display or possibly a computer terminal. The larger the amplitude of the echo, the brighter is the spot displayed. View the Flash simulation UltraSoundImagingwithoutObject2.swf. This simulation allows you to compare the A-scan and B-scan screens. Note that the transducer and the data it generates is exactly the same for both scans. The data is just presented in a more visual format by the B-scan. By moving the transducer, a two-dimensional image of the interfaces can be constructed. GE 3. 1. Why might it be easier to produce a 2D image more easily from a series of B-scans than from a series of A-scans? View the flash animation UltraSoundImagingwithObject2.html. This simulation shows how the B-scan information can be built up into a 2D image by moving the transducer. Activity Guide 2010 The Humanized Physics Project HPP Activity 55v1 4 Instead of moving the transducer by hand, as you do in the simulation, a B-scan instrument would either move (or simply rotate) the transducer using an automated motor, or would use an array of transducers that are spatially separated. Application The B-scan image shows a two-dimensional picture of interfaces within the body. We would expect that interfaces between the same two tissue types should lead to the same brightness on the image display. Unfortunately, this is not the case and a correction must be applied to the ultrasound instrument. GE 4. 1. What happens to the intensity of sound waves as they travel through a medium? Think about how well you can hear a normal conversation when you are standing next to the people talking and when you are 10, 20, 30, or more feet away. What will happen to the brightness of the echoes for interfaces that are farther away from the transducer? 2. Suggest a way to correct the ultrasound image for this problem. Ultrasound imaging devices, including the ultrasonic motion detector you may have used in the past, correct for the attenuation of the echo amplitude. The ultrasound transducer uses the piezoelectric effect to both generate the sound pulse and to detect the echo. When an oscillating voltage is applied to a piezoelectric crystal, the crystal expands and contracts at the same frequency as the applied voltage producing a sound wave. Conversely, when a sound wave strikes a piezoelectric crystal, the crystal expands and contracts because of the pressure differences. The crystal generates a voltage with an amplitude that is related to the amount of deformation caused by the sound wave and a frequency that is the same as the sound wave. It is that voltage that is amplified to produce the ultrasound image. Activity Guide 2010 The Humanized Physics Project