Mini SDS-PAGE Gel Protocol for the Novex Surelock Cell Note: For best results, use MilliQ or nanopure equivalent water to prepare all buffers and samples. Preparing 1X NuPAGE SDS PAGE Running Buffer Components 20X NuPAGE MOPS or MES Running Buffer Nanopure water Amount for 0.8 L 40 ml 760 ml The 1X can be made several days in advance. Note: If the 20X or 1X buffer is turning yellow, the glycine is degrading and so it is time to properly dispose of it. Sample Preparation 1) Prepare the samples: 10-Well Gels-maximum loading volume 25 µl/well (1.0 mm thick) & 37 µl/well (1.5 mm thick) 12-Well Gels-maximum loading volume 20 µl/well (1.0 mm thick) 15-Well Gels- maximum loading volume 15 µl/well (1.0 mm thick) & 25 µl/well (1.5 mm thick) Sample Water NuPAGE LDS Sample Buffer 2.5 μl (4X) NuPAGE Reducing Agent (10X)* 1.0 μl Total Volume 10 μl 2) Prepare 1X sample buffer to fill the empty wells: Water NuPAGE LDS Sample Buffer (4X) NuPAGE Reducing Agent (10X)* Total Volume 3.75 μl 5 μl 6.25 μl 10 μl 1.5 μl 15 μl 2.0 μl 20 μl 2.5 μl 25 μl 4.0 μl 40 μl 650 μl 250 μl 100 μl 1000 μl Note: *Add the Reducing Agent within 1 hr of loading. Do not freeze sample after adding. Reoxidation of samples occur during storage and produce inconsistent results. 3) Heat the samples at 70°C for 10 minutes for optimal results. Do not boil your samples. Boiling the sample at 100°C in SDS containing buffer results in proteolysis (Kubo, K. Anal Biochem 1995, 351353). Note for people who are working with membrane proteins: Do not boil your sample because it may cause your proteins to aggregate. Instead, load your sample immediately after adding the loading buffer. 1 Preparing a Precast Gel for Electrophoresis 1) 2) 3) 4) 5) Remove the precast gel from the storage pouch. Rinse the outside of the gel thoroughly with distilled water. Remove the tape from the slot on the back of the gel cassette. In one fluid motion, gently remove the comb. Rinse the wells thoroughly with either: distilled water or 1X running buffer. Invert the gel and shake to remove excess buffer/water. Repeat twice. If necessary, straighten the sides of the wells. Assembly of the Surelock Cell 1) Lower the Buffer Core into the Lower Buffer Chamber so that the negative electrode fits into the opening in the gold plate on the Lower Buffer Chamber. 2) Insert the Gel Tension Wedge into the cell behind the buffer core. Make sure the Wedge is in the unlock position. 3) Insert gel cassettes into the Lower Buffer Chamber by place one cassette in front of the Buffer Core. Place the other gel or Buffer Dam behind the Buffer Core. The shorter plate/”well” side of the cassette needs to face towards the Buffer Core (inside). The slot at the bottom of gel must face towards the buffer in the Lower Buffer Chamber (outside). 4) Pull forward on the Gel Tension Lever until the lever comes to a firm stop. Pull lever this direction. 5) Fill the upper/inner chamber with 200 ml of the appropriate Running Buffer. Make sure there is enough buffer to completely cover the sample wells. Check to see that the buffer is not leaking from the inner chamber into the outer chamber. 6) Add 500 μl of the NuPAGE Antioxidant to upper/inner chamber buffer just prior to loading your samples. Mix thoroughly. ( Note: The Antioxidant prevents sample reoxidation and maintains the proteins in a reduced state. In the neutral pH environment of NuPAGE gels, DTT and β-mercaptoethanol tend to remain at the top of the gel, and do not co-migrate with the sample. The Antioxidant migrates with the proteins during electrophoresis, and protects disulfide bonds and sensitive amino acids (e.g. methionine and tryptophan) from oxidizing. Sample Loading 1) Using a gel loading pipette tip, load the molecular weight standards (usually in lane 1). Note: All Molecular Weight Standards are “not” compatible with the NuPAGE gels. Use only the standards listed below in the Supply Table or ones approved by Life Technologies. 2 2) Load your samples into the appropriate wells by lowering the pipette tip to the bottom of the sample well and slowly pipet the sample into well. Be careful not to contaminate another well with the sample or puncture the bottom of the well. 3) Load 20 µl 1X Sample loading buffer into any remaining empty wells. This promotes uniform running of the stacking front. Running Gel(s) 1) Fill the Lower Buffer Chamber with the remaining 600 ml of Running Buffer by pouring it in the gap between the Gel Tension Wedge and the back of the Lower Buffer Chamber. 2) Place the lid on the Mini-Cell. Make sure the lid is properly seated or no power will go through the MiniCell. 3) With the power off, insert the electrical leads into the lead adapters in the Bio-Rad PowerPac HC. Caution: You must use HV lead adapters when using the Bio-Rad power supply with the Novex Surelock Mini-Cell; otherwise you will cook your gel. HV Lead Adapters 4) Apply power to the Mini-Cell and begin electrophoresis: Standard Method 200 V until the dye front reaches the bottom Low Voltage Method-Membrane Proteins and Bands for Mass Spectrometry Analysis 40 V for 10 min (allows better stacking of the proteins) 100 V until the dye front reaches the bottom Note: The low voltage method reduces the amount of heat which is generated during electrophoresis resulting in sharper bands. For further cooling, the gel maybe ran in a coldroom/coldbox. Don’t prechill the running buffer since colder temperatures will cause the SDS to come out of solution. Gel Removal 1) 2) 3) 4) At the completion of the run, turn off the power supply and disconnect the electrical leads. Remove the tank lid and unlock the Gel Tension Lever. Remove the gel cassettes from the Mini-Cell. Lay the gel with the bottom slot/long plate facing down. Insert the Gel Knife into the gap between the two plastic plates to separate each of the three bonded sides of the cassette. 5) Gently push down on the Gel Knife to separate the plates. You will hear a cracking sound which means you have broken the bonds which hold the plates together. Repeat on each side of the cassette until the plates are completely separated. Caution: Be careful while inserting the Gel Knife between the gel plates or you may cut into the gel. 3 6) Carefully remove the top plate. The gel may adhere to either side. Discard the plate without the gel. 7) Hold the cassette plate over a container with water or transfer buffer with the gel side facing towards the container. If the gel is stuck to the plate with the bottom slot, use the Gel Knife to push the gel foot through the slot. The gel should be in contact with the fluid, and with some with gentle agitation, the gel should fall into the fluid. If the gel is stuck to the short plate, place the gel into the fluid and agitate gently until the gel separates from the plate. You may need to use Gel Knife to carefully loosen the lower corner of the gel and then allow the gel to peel away from the plate. Note: If the gel will be transferred, the gel foot will need to be cut off with a Gel knife. 8) Discard the running buffer into the proper hazardous waste container. 9) Rinse the Mini-Cell and all its components with distilled, deionized water. References The protocol above was based in part, on the following literature from Life Technologies: • XCell SureLock Mini-Cell Manual (MAN0000739, Rev. Date 6 May 2010) • NuPAGE Technical Guide (MAN0003188, Rev. Date 29 October 2010) Written by Melissa Sondej (version 4/17/12) 4 Supplies Below lists the supplies needed. Gels NuPAGE 4-12% Bis-Tris Gel, 10 well, 1.0 mm, 25 µl NuPAGE 4-12% Bis-Tris Gel, 10 well, 1.5 mm, 37 µl NuPAGE 10% Bis-Tris Gel, 10 well, 1.0 mm, 25 µl NuPAGE 12% Bis-Tris Gel, 10 well, 1.0 mm, 25 µl NuPAGE 4-12% Bis-Tris Gel, 12 well, 1.0 mm, 20 µl NuPAGE 10% Bis-Tris Gel, 12 well, 1.0 mm, 20 µl NuPAGE 12% Bis-Tris Gel, 12 well, 1.0 mm, 20 µl NuPAGE 4-12% Bis-Tris Gel, 15 well, 1.0 mm, 15 µl NuPAGE 4-12% Bis-Tris Gel, 15 well, 1.5 mm, 25 µl * Gels are also sold in a 10 pack. Reagents NuPAGE MOPS SDS Running Buffer (20X) NuPAGE MOPS SDS Running Buffer (20X) NuPAGE Sample Reducing Agent (10X) NuPAGE Antioxidant NuPAGE LDS Sample Buffer Protein Standards BenchMark Unstained Protein Ladder (Sypro Ruby) Novex Sharp Unstained Protein Standard (Sypro Ruby) Novex Sharp Prestained Protein Standard (Western) Stain Sypro Ruby Fluorescent Protein Stain (10 ng sensitivityovernight) Gelcode Blue Stain Reagent (25 ng sensitivityovernight) Homemade Coomassie Blue (100 ng sensitivity) Imperial Protein Stain (>10 ng sensitivity) Size Vendor Catalog Number 2 gels 2 gels 2 gels 2 gels 2 gels 2 gels 2 gels 2 gels 2 gels Life Technologies Life Technologies Life Technologies Life Technologies Life Technologies Life Technologies Life Technologies Life Technologies Life Technologies NP0321PK2 NP0335PK2 NP0301PK2 NP0341PK2 NP0322PK2 NP0302PK2 NP0342PK2 NP0323PK2 NP0336PK2 500 ml 500 ml 250 μl 15 ml 10 ml Life Technologies Life Technologies Life Technologies Life Technologies Life Technologies NP0001 NP0002 NP0004 NP0005 NP0007 2 x 250 µl 2 x 250 µl 2 x 250 µl Life Technologies Life Technologies Life Technologies 10747012 LC5801 LC5800 200 ml Life Technologies S12001 500 ml Thermo Scientific Pierce 24590 200 ml Thermo Scientific Pierce Bio-Rad 161-0495 2.5 L 4L EMD/VWR EMD/VWR EMD-AX0073-9 EM-MX0475-1 1L Oriole Fluorescent Gel Stain Fixer/Destain Solution Acetic Acid Methanol 5 24615