Archives of Perinatal Medicine 14(2), 47-50, 2008
ORIGINAL PAPER
The illumination intensity in the neonatal intensive care unit
MAREK SZCZEPAŃSKI, MONIKA KAMIANOWSKA
Abstract
Objective: The aim of this study was the description of the Neonatal Intensive Care Unit light environment. Patient and methods: Using the
luxometer, light intensity (lx) in open as well as closed incubators in standard, reduced and minimal night illumination and applying
procedures reducing light transmission to neonate eyes was recorded. The measurements were performed in the NICU at the Department
of Neonatology of Medical University in Białystok. Results: Standard night illumination generated light intensity of 325 ± 5.0 lx, reduced 31
± 1.0 lx and minimal 1 ± 0.01 lx. Procedures reducing light transmission to neonate eyes reduced light intensity to 22 ± 1.0 lx in standard
illumination and to 3 ± 0.7 lx in reduced illumination. Conclusion: Minimal night illumination generates light intensity which is consistent with
recommendations for NICU, standard and reduced night illumination exceeds permissible levels. Applying illumination reducing procedures
in NICU, restricts night light transmission to neonate eyes, reducing the risk of visual system disorders and circadian rhythm disturbances
in the hospitalized neonates.
Key words: neonatal intensive care unit, illumination intensity
Introduction
The development of the sight organ starts at the end of
the 3rd week of intrauterine life. At the 5th week, the plate of
the lens is situated on the top of the primary optic vesicle, of
which the retina is formed. Nervous and glia elements are
transformed into the optical nerve. At about the 10th week of
gestation, the lids merge as two folds of superficial ectoderm
and the eye fissure arises only at about the 26th week. The
development of the eyes and sight is not finished at the moment of birth and is influenced by environmental optical stimuli. The axial length of the eyeballs, the cornea curvature
and the lens thickness change in infancy. Total myelinization
of the optic nerve may last up to the age of 2 years, and the
spot field up to the age of 5 [1, 2].
Light inducing visual sensation is electromagnetic radiation of 400-780 nm wave length. Illumination intensity, that is
luminous flux density on the illuminated surface, is expressed
in lux (lx) [3]. Light plays a major role in the postnatal development of the sight organ (eyeball growth, peripheral retina
parts development) and supports vision processes (visual
acuity, colour vision) [4].
Exposure to high illumination may lead to structural changes in the vision organ [5]. The changes, located most of all in
the external segments of the retina, and dependent on the time and continuity of the exposure, illumination and night activity, are generated in the mechanism of thermal, mechanical
and photochemical (oxidation) damage [6]. Deformation, vacuolization and fragmentation of rods and cones progress leading
to the complete destruction of the photosensitive layer of the
retina [6]. It has been suggested that degenerative changes
induced by the light of high intensity are developed in apoptosis mechanism [7]. According to Abramow et al. impaired
colour and space vision (without a dysfunction of rods) is observed in the neonate in the NICU [8]. In his earlier studies,
Hamer et. al. did not prove the influence of continuous light
Department of Neonatology, Medical University of Białystok, Poland
exposure on the functioning of retina sensor cells in the
neonate staying in the NICU [9]. According to Kennedy et. al.
and Roy et al. the preventive light reduction does not impair
the development of the retina [10] and [11]. Animal studies
prove an unfavourable effect of continuous light exposure on
the growth of the eyeball, leading to the impaired eye refraction [12]. Recent studies exclude the influence of light on
the incidence, intensity degree, and the course of retinopathy
in premature infants [13].
Circadian rhythms induced by the rotation of the Earth
generated by the light-induced biologic clock located in the
hypothalamic suprachiasmatic nuclei synchronise physiological processes of the organism during the 24hr light/dark cycle,
conditioning the health [14]. After birth, the circadian rhythm
of the fetus is regulated by the mother’s day/night activation.
It depends on the environmental illumination (regular and irregular light/dark cycles), feeding (breast milk, instant formula) and intrauterine age [15]. Rivkees’s et al. studies demonstrated a higher activity in premature infants during the day
than at night, in comparison with those staying continuously
in reduced illumination [14].
The influence of light on physiological processes in neonates, evaluated by many authors is not explicit. According to
Jung, the cycled lighting used for 10 days increases the neonate’s body mass and saturation, decreases the heart beat, not
affecting the respiration rate, behaviour stability and the body
weight [16]. Brandon et al. observed a beneficial influence of
the cycled light on the weight gain in neonates [17]. Boo et al.
did not confirm these observations [18].
Neonates hospitalised in the NICU are exposed to natural
and artificial illumination (illumination of the ward room, procedural illumination, phototherapy, ophthalmoscopy) [19]. The
physical properties of light, physiological factors (lid thickness, opening of the eyes, pupil diameter), and environmental
factors (type of incubator, methods reducing illumination in-
48
M. Szczepański, M.Kamianowska
tensity used, location in the NICU room) affect the amount of
light that the neonate is exposed to [19] and [20]. The examinations suggest that an increase in the level of illumination
intensity decreases the frequency of the opening of the lids in
neonates, which may impair the process of vision development
(sight fixation) and lead to phototoxicity [21]. The aim of the
study was the description of illumination in the NICU at night.
Material and methods
Illumination intensity was measured by a luxometer L-0A
(producer – PPUH SONOPAN, Poland) and expressed in lux
(lx). The examinations were performed in one of the NICU
rooms of the Department of Neonatology of the Medical University of Białystok, Poland.
There were 4 incubators placed peripherally (two at each
side of the room) sized 5.9 × 4.6 m The walls of the room
were covered with tiles to the height of 1.6 m, the floor had
a covering, two windows sized 150 × 200 cm were screened
with roller-blinds (impregnated linen), the desk and cabinets
were placed peripherally. Standard illumination of the room included 3 fluorescent lamps placed under the ceiling turned on
with the separate switcher. Procedural illumination (100 W)
and the bedside lamp were the additional source of light. The
measurements were performed after twilight. The lightning
was switched on for about 30 min before measurement and
the photoelectric receiver was illuminated for at least 5 min
with the measured light intensity (according to PN – 84 E –
02033 Interior illumination with electric lighting). Each measurement was done three times, the mean and standard deviation were calculated. The measurements were carried out in
the closed incubator (ISOLETTE, Air – Shield, Vickers, GB)
and the open incubator (Babytherm 8000, Dräger, Medizintechnik GmbH, Germany) situated at 0.6 m from the wall.
illumination (extra source of light) and phototherapy. The
procedures reducing the light amount that the neonate was
exposed to, were applied (the change in a doll’s position in the
incubator, a one- or four-ply cover on the incubator, eyepatches for phototherapy) (a producer PRO – OPTICS, INC.,
PALATINE, IL, USA).
Results
The illumination intensity measured in the NICU, in the
open and closed incubator at the standard, reduced and minimal night lighting is presented in Table 1.
Table 1. Illumination intensity in the NICU at various types
of illumination used (lx)
Type of illumination
Closed incubator
Open incubator
Standard illumination
325 ± 5.0
200 ± 5.8
Reduced (night)
31 ± 1.0
31 ± 2.1
Minimal illumination
1 ± 0.01
1 ± 0.01
It is worth mentioning that the structure of the open incubator provided darkening in the field of measurement, due to
which the result obtained (200 ± 5.8 lx) was significantly lower
in comparison with the results taken in the closed incubator
(325 ± 5.0 lx).
The amount of light that the neonate is exposed to is modified by its positioning in the incubator and under the incubator cover. In the conditions of both standard and reduced
illumination, if the incubator is uncovered, the illumination
intensity taken at the height of the neonate’s eyes is lower
when the doll is positioned on the side in comparison with the
supine position on its back. However, the cover on the incubator reduces more significantly the amount of light the neonate is exposed to from above than from the side (Tab. 2).
Table 2. Illumination intensity depending on positioning a doll,
with application of the incubator cover (lx)
Closed incubator
Standard illumination
incubator without the cover
incubator with the cover
one-ply fabric
four-ply fabric
Reduced illumination
incubator without the cover
incubator with the cover
Position on
the back
Position on
the side
325 ± 5.0
253 ± 6.0
47 ± 7.0
22 ± 1.0
108 ± 9.6
100 ± 3.0
31 ± 1.0
3 ± 0.7
24 ± 2.5
18 ± 1.1
Fig. 1. Method of illumination intensity measurement
The photoelectric receiver of the luxometer was placed
on a doll’s head at the level of its eyes.
The measurements were performed in the conditions of
standard illumination of the NICU room (3 fluorescent lamps
switched on), reduced illumination (1 fluorescent lamp) minimal illumination (1 bedside lamp on the desk) and procedural
An increase in the thickness of the fabric used to sew the
incubator cover reduces the standard illumination intensity
from 47 ± 7.0 lx in case of one-ply fabric and 22 ± 1.0 lx, when
four-ply fabric was used, when the doll was positioned on the
back position. When the doll was positioned on the side position, the incubator cover reduces the standard illumination
from 108 ± 9.6 lx to 100 ± 3.0 lx respectively (Tab. 2).
The illumination intensity in the neonatal intensive care unit
Applying extra light sources (procedural lighting or lamps
for phototherapy) used in the NICU increases markedly
the amount of the light the neonate’s eyes are exposed to
(Tab. 3).
Table 3. Illumination intensity with using extra light sources (lx)
Standard illumination
Procedural illumination
without eye-patches
Phototherapy
with eye-patches
Open incubator
200-325*
533-1280**
3860-8560***
120-211***
* with regard to the type of the incubator
** with regard to the type of light source
*** with regard to a model of the lamp used for phototherapy
Eye-patches used in phototherapy reduce the illumination
intensity from 3860-8560 lx to 120-211 lx depending on the
model of the lamp used.
Discussion
The role of illumination in caring for hospitalised neonates (it maintains and influences the development of vision
processes, regulates the circadian rhythm, enables professional care and parents’ visits) and taking into consideration the
harmful effect of too bright light on the development of the
vision organ and the circadian cycle, delineate the schemes for
the recommended illumination in the NICU [4] and [5].
In the 1980s, the level of illumination intensity ranging
from 600 to 1000 lx was commonly accepted [18]. Proving the
harmfulness of high intensity illumination on the process of
the vision organ development caused the reduction in the recommended illumination intensity to the level of 1-2 lx in the
NICU, permitting at the same time an individual increase in
the illumination for observation and procedures [22]. Taking
into consideration the significance of the circadian rhythm for
the physiological development of the neonate hospitalised in
the NICU in 1998 in Great Britain, 50-100 lx was taken as
a mean level in the indirect care and 1000 lx in the intensive
therapy with the reduction in the illumination to 5 lx between
8 p.m. and 8 a.m. [19]. Similarly, in 1997 American Academy
of Pediatrics (AAP) recommended introducing regular day/
night cycles of illumination in the NICU [15]. Developing
a physiological circadian rhythm in premature babies requires
illumination intensity of 100-150 lx for 8 hours daily with
a reduction in the evening and at night [22]. The concept of
illumination assumes the individualisation of lighting underlying beneficial effects of the natural light [23].
Robinson et al. examined the conditions of illumination in
7 NICUs in Great Britain in 1990, proving a mean level of the
illumination intensity at night (8 p.m.-8 a.m.) 348 lx (range192690 lx), and during the day (7.a.m.-8 p.m.) 470 lx (236-905 lx).
These results were lower than those obtained earlier by Landry et. al. (285-1485 lx) [19]. The further reduction in the illumination in the NICU was recorded by Bullough et al. showing
the illumination intensity of 184 lx during the day (1 p.m.) and
49
at night (12.30 a.m.) – 34 lx [24]. In our study, the standard
illumination of the NICU room had the intensity of 325 lx,
reduced – 31 lx, minimal – 1 lx. According to the standards
accepted in the Department of Neonatology of the Medical
University of Białystok, referring the neonate’s minimal care
and therapy, minimal illumination with the individualised
illumination intensity in case of necessary procedures, was
applied at night.
Walsh-Sukys et al. examined the effects of the reduced
illumination, which is beneficial for the preterm neonate’s development, patients’ safety, medical staff’s perception and the
costs of hospitalisation in the NICU of level III reference.
Introducing the individualised illumination for each neonate
caused the reduction in illumination from 193 lx to 12 lx, individual extra illumination, which could be compared to traditional (202 lx), was provided if necessary. The costs of the
project were low. About 60% of the medical staff regarded new
conditions to be beneficial for both neonates and caretakers.
No increase in the number of unwanted accidents, malpractice, accidental extubations, infections or in mortality rate
was reported [22].
Phototherapy applied in the NICU generates the high illumination intensity. It is recommended to use eye-patches,
mechanically stopping the light in order to protect the vision
organ against damage (deformation, vacuolization, fragmentation of the rods and cones) [25] and [26]. Robinson et. al. proved that when using various protective patches for the eyes,
the illumination was reduced up to 2% - 10% at the wavelength
of 700 nm [25]. In Chin’s et al. study transmission of the light
wavelength of 460 nm did not exceed 0.04% when eye-patches
were used [26]. In the study, we reduced the illumination
intensity using phototherapy eye-patches to 2.5-3.1% depending on a model of the lamp used.
Lee et. al. evaluated the influence of the incubator cover
on the reduction of the light that the neonate was exposed to
[27]. They proved that the incubator cover in dark colours
reduced more of the light inside the incubator than the cover
in light colours [27]. The further reduction in the illumination
transmitted to the neonate can be achieved by an increase in
thickness of the fabric used to sew the cover. When an incubator is covered, the neonate’s supine position on the back is
favorable (22 ± 1.0 lx at a supine position on the back, and 100
± 4.7 lx at the position on the side). In the uncovered incubator,
a position on the side is more favorable (325 ± 5.0 lx in the position on the back and 253 ± 6.0 lx at the position on the side).
The reduction in the illumination that the neonate is
exposed to in the incubator correlates with the thickness of
the wall in the incubator. According to Kardson, an increase
in the wall thickness, to diminish heat loss in the neonate,
reduces the therapy light that the neonate is exposed to
during more than 40% [28]. In our study, the illumination
intensity was lower in the open incubator (200 ± 5.8 lx) in
comparison with the intensity in the closed incubator (325 ±
5.0 lx). It resulted from the shadow of the open incubator
structure cast on the measurement field.
50
M. Szczepański, M.Kamianowska
Conclusion
Minimal night illumination generates the light intensity
in accordance to the recommendations for the NICU. Application of the procedures reducing the illumination intensity
in the NICU decreases night light transmission to the
neonate’s eyes, which is especially important in the physiological development of the vision organs and circadian rhythm
of neonates hospitalised in the NICU.
References
[1] Prost M.E. (1998) Problems in children’s ophthalmology. Biblioteka Pediatry 31, Warszawa: PZWL, pp.13-21.
[2] Bartel H. (2004) Embriology. Warszawa: PZWL, pp. 442-52.
[3] Encyclopaedia of Metrology. (1989) Warszawa: WNT, pp. 241
and 405.
[4] Rea M. (2004) Lighting for caregivers in the neonatal intensive
care unit. Clin. Perinatol. 31: 229-42.
[5] Fielder A.R., Moseley M.J. (2000) Environmental light and the
preterm infant. Semin. Perinatol. 24: 291-8.
[6] Logvinov S.V., Potapov A.V., Varakuta E.Y. et al. (2003)
Dynamics of structural changes in the retina during long-term
exposure to bright light. Bull. Exp. Biol. Med. 136: 411-4.
[7] Wenzel A., Grimm C., Samardzija M. et al. (2005) Molecular
mechanisms of light-induced photoreceptor apoptosis and neuroprotection for retinal degeneration. Prog. Retin. Eye Res. 24:
275-306.
[8] Abramov I., Hainline L., Lemerise E. et al. (1985) Changes in
visual functions of children exposed as infants to prolonged illumination. J. Am. Optom. Assoc. 56: 614-9.
[9] Hamer R.D., Dobson V., Mayer M.J. (1984) Absolute thresholds
in human infants exposed to continuous illumination. Invest.
Ophthalmol. Vis. Sci. 25: 381-8.
[10] Kennedy K.A., Ipson M.A., Birch D.G. et al. (1997) Light reduction and the electroretinogram of preterm infants. Arch. Dis.
Child. Fetal Neonatal. Ed. 76: F168-73.
[11] Roy M.S., Caramelli C., Orquin J. et al. (1999) Effects of early
reduced light exposure on central visual development in preterm infants. Acta Paediatr. 88: 459-61.
[12] Li T., Howland H.C. (2003) The effects of constant and diurnal
illumination of the pineal gland and the eyes on ocular growth
in chicks. Invest. Ophthalmol. Vis. Sci. 44: 3692-7.
[13] Phelps D.L., Watts J.L. (2001) Early light reduction for preventing retinopathy of prematurity in very low birth weight infants.
Cochrane Database Syst. Rev. 1: D000122.
[14] Rivkees S.A., Mayes L., Jacobs H. et al. (2004) Rest-activity patterns of premature infants are regulated by cycled lighting. Pediatrics 113: 833-9.
[15] Mirmiran M., Ariagno R.L. (2000) Influence of light in the
NICU on the development of circadian rhythms in preterm infants. Semin. Perinatol. 24: 247-57.
[16] Jung I.S. (2005) Effects of cycled lighting on body weight, physiological variables and behavioral states in low birth weight
infants. Taehan Kanho Hakhoe Chi. 35:143-53.
[17] Brandon D.H., Holditch-Davis D., Belyea M. (2002) Preterm
infants born at less than 31 weeks' gestation have improved
growth in cycled light compared with continuous near darkness.
J. Pediatr. 140:192-9.
[18] Boo N.Y., Chee S.C., Rohana J. (2002) Randomized controlled
study of the effects of different durations of light exposure on
weight gain by preterm infants in a neonatal intensive care unit.
Acta Paediatr. 91: 674-9.
[19] Robinson J., Moseley M.J., Fielder A.R. (1990) Illuminance of
neonatal units. Arch. Dis. Child. 65: 679-82.
[20] Robinson J., Fielder A.R. (1992) Light and the immature visual
system. Eye 6: 166-72.
[21] Moseley M.J., Thompson J.R., Levene M.I. et al. (1988) Effects
of nursery illumination on frequency of eyelid opening and state
in preterm neonates. Early Hum. Dev.18: 13-26.
[22] Walsh-Sukys M., Reitenbach A., Hudson-Barr D. et al. (2001)
Reducing light and sound in the neonatal intensive care unit: an
evaluation of patient safety, staff satisfaction and costs. J. Pe-
rinatol. 21: 230-5.
[23] White R.D. (2004) Lighting design in the neonatal intensive
care unit: practical applications of scientific principles. Clin.
Perinatol. 31:323-30.
[24] Bullough J., Rea M.S., Stevens R.G. (1996) Light and magnetic
fields in a neonatal intensive care unit. Bioelectromagnetics 17:
396-405.
[25] Robinson J., Moseley M.J., Fielder A.R. et al. (1991) Light transmission measurements and phototherapy eyepatches. Arch. Dis.
Child., 66: 59-61.
[26] Chin K.C., Moseley M.J., Bayliss S.C. (1987) Light transmission
of phototherapy eyeshields. Arch. Dis. Child. 62: 970-1.
[27] Lee Y.H., Malakooti N., Lotas M. (2005) A comparison of the
light-reduction capacity of commonly used incubator covers.
Neonatal Netw. 24: 37-44.
[28] Karsdon J., Schothorst A.A., Ruys J.H. et al. (1986) Plastic blan-
kets and heat shields decrease transmission of phototherapy
light. Acta Paediatr. Scand. 75: 555-7.