Introduction to Ultrasound Physics Vassilis Sboros Medical Physics and Cardiovascular Sciences University of Edinburgh Transverse waves •Water remains in position Disturbance traverse producing more wave along the path Disturbance travel at 90o of water movement, hence transverse Longitudinal wave •Particles remains in position Disturbance travel at 0o of particle movement, hence longitudinal Sound-Mechanical wave Generated by piezoelectric crystals Single reflection Sound-Mechanical wave Frequency • • • • • 1Hz = 1 cycle per second Sound 20 Hz – 20 kHz Ultrasound > 20kHz Diagnostic Ultrasound 1-50 MHz Ultrasound Therapy 40kHz-1MHz Some definitions • • • • • • • • • Wavelength Phase Velocity of sound Acoustic impedance Reflection Scattering Refraction Absorption Attenuation Wavelength λ λ= c ν For ctissue = 1540 m/s ν=1MHz, λ=1.54mm ν=3MHz, λ=0.51mm ν=10MHz, λ=0.15mm c : velocity of sound (ms-1) ν : frequency (Hz) cair = 330 m/s λ=0.33mm Phase a) b) Angle of cycle rotation Phase difference between identical waves Pressure • Positive – compression, negative – rarefaction • Units 1 Pa = N / m2 Intensity (time) Units W / m-2 Velocity of sound c κ c= ρ κ : stiffness (Pa) ρ : density (Kg/m3) cair = 330 m/s cwater = 1480 m/s ctissue = 1540 m/s cfat = 1450 m/s cblood = 1570 m/s cbone = 3500 m/s Acoustic impedance Z p Z = = ρc u p : pressure (Pa) u : particle velocity (m/s) Reflection p2 Z1 − Z 2 ρ1c1 − ρ 2 c2 = = p1 Z1 + Z 2 ρ1c1 + ρ 2 c2 pmuscle / pblood = 0.03 pfat / pmuscle = 0.10 pbone / pmuscle = 0.64 pmuscle / pair = 0.99 Reflection a) b) c) Smooth surface Small particle Rough surface Scattering General case for reflection λ >> particle size = Rayleigh scattering λ ~ particle size = Mie scattering λ << particle size = reflection Refraction Attenuation Attenuation = scattering + absorption Absorption = conversion to heat Intensity decays exponentially Frequency dependant Interference Multiple ultrasound sources a) Constructive interference – waves in phase b) Destructive interference – waves in antiphase Plane disk transducer Intensity (space) Frequency Spectrum a) b) Time domain Frequency domain (FFT) Nonlinear propagation At high ultrasound pressure • Time domain – asymmetrical pattern • Frequency domain (FFT) – Harmonic frequencies Bibliography • McDicken W.N. Diagnostic Ultrasonics Churchill Livingstone New York 1991. • Barnett E., Morley P. Clinical Diagnostic Ultrasound Blackwell Scientific Publications, Oxford 1985. • Meire H.B., Cosgrove D.O., Dewbury K.C., Farrant P. Clinical Ultrasound a comprehensive text: Abdominal and General Ultrasound Vol.2 Churchill Livingstone New York 2001. The Engineering of Ultrasound Imaging Vassilis Sboros Medical Physics and Cardiovascular Sciences University of Edinburgh Transducer Engineering Piezoelectric materials • • Positive Voltage = compression Synthetic ceramic - Lead Zirconate Titanate (PZT) – – – – • High sensitivity High acoustic power Easy to micromachine Impedance 20x tissue Thickness = λ/2 - resonance – Resonance due to internal reflection – Determines transmit frequency Transducer Engineering – Backing layer • PZT Impedance 20x tissue – Duration of pulse difficult to control due to internal ringing • Backing layer = absorber – High impedance – Reduces ringing Transducer Engineering – Matching layer • PZT Impedance 20x tissue – Only 20% of energy transmitted to tissue • Matching layer = impedance matching – Impedance lower than PZT and higher than tissue – Remove some ringing • 1 layer 2x sensitivity – λ/4 thickness – Constructive interference towards tissue – Destructive interference towards PZT Transducer Engineering – Frequency bandwidth vs sensitivity • High sensitivity = specific dimensions for Backing, PZT and Matching layers – Frequency band is narrow – Resolution low • >1 Matching layers – Decreasing impedance • Bandwidth 2x (60% to 120%) – Little loss in sensitivity 1D – Single Plane disk transducer 2D beams – Array transducers a) b) c) d) e) Linear Curvilinear Trapezoidal Sector Radial Transducer Engineering – Lens • Single element – Focus has high sensitivity and resolution • Linear Array – Electronically in scan plane – Only in elevation plane • Phased Array – Mild in scan plane – Stronger in elevation plane Linear Array Transducers • 128 elements – Binary processing • Choice of frequency – Penetration vs resolution or attenuation vs frequency • Dimensions ~ 1/f – ~1.3λ width per element (83mm @3MHz) – ~30λ height - elevation (15mm @3MHz) Linear Array Transducers • Active group of elements – Finite beam per element – Transmit fixed (~20) – Receive (<20 to >20 as depth increases) – Electronic focus Linear Array Transducers • Transmit Electronic Focus – Transmission timing – One focus – Controllable Linear Array Transducers • Receive Electronic Focus – – – – – Electronic delay Depth ~ element number Multiple foci Not controllable/automatic High resolution at all depths Linear Array Transducers Transmit Multiple focus Linear Array Transducers 1.5D array for improved elevation focus Linear Array Transducers Transmit Apodization Curvilinear Array Transducers • Sector scanning – Wider field – Linear array structure – Active element number reduced Poorer resolution Phased Array Transducers • Sector scanning – Narrow acoustic window • Narrower elements – All elements used (transmit and receive) – Shorter near field per element – Wider far field per element – Beam steering ±45o Linear/Phased Array Transducers Compounding – Reduction of noise Persistence – Reduction of frame rate Matrix Array Transducers Endocavity Array Transducers a) Curvilinear – transvaginal b) Curvilinear – Transvaginal, transrectal c) Bi-plane – Transrectal (prostate) d) Phased array – Transoesophageal (heart) Intravascular Array Transducers • • • Curvilinear/convex 360o High frequency (30MHz) Vessel wall phantom A-mode (transmission) A-mode Eye A-mode B-mode scanning Eye B-mode B-mode Formation of B-mode image B-mode B-mode Transmit gain and power B-mode Time gain compensation (TGC) Compensate for attenuation B-mode Analogue to digital conversion limited values – memory binary system sampling rate (40MHz) digital processing B-mode Digital signal Rectification Enveloping B-mode Compression Accommodate in the image low and high echoes B-mode Image memory B-mode Interpolation Linear? B-mode Reading of image memory to form display Gray scale Ultrasound Imaging Modes • Real-time 2D imaging – Good spatial resolution – Good temporal resolution – Good Penetration Heart scan Ultrasound Imaging Modes • 3D and 4D – Good spatial resolution – Poor temporal resolution – OK Penetration Foetal scan Heart scan Doppler Ultrasound Pete Hoskins and Vassilis Sboros Medical Physics and Cardiovascular Sciences University of Edinburgh Doppler ultrasound • • • • Principles of Doppler CW/PW Doppler Doppler systems (spectral, duple, colour) and controls Principles of contrast imaging Doppler effect Controls Doppler system patient Doppler effect Change in pitch is proportional to speed of source Change in pitch = fS - fO Doppler shift = fd = fS - fO Speed = v fd ~ v Doppler ultrasound Transducer Blood Transmission T Scattering R Reception R Case 1. Blood stationary Transmission T Scattering R Reception R fr = ft Case 2. Blood moving away from transducer Transmission T Scattering R Reception R fr < ft Case 3. Blood moving towards transducer Transmission T Scattering R Reception R fr > ft General case ft v fr = ft + fd fd = 2 ft v/c Some values • Transmit frequency • Speed of sound • Speed of blood • Doppler shift • Hear Doppler signal 4 MHz 1540 m/s 1 m/s = 5194 Hz Doppler ultrasound Transmission ft Scattering Reception fr Doppler ultrasound ft + fd ft θ v fd = 2 ft v cos θ/c Cosine function 1.0 Cosine 0.8 0.6 0.4 0.2 0.0 0 10 20 30 40 50 60 70 80 90 Angle (degrees) 80ο 60ο 40ο Some more values • • • • Transmit frequency Velocity Angle Speed of sound 3-5 MHz 0-3 m/s 40-80 degrees 1540 m/s • Doppler frequency shift 0-15 kHz • Audio range 0-20 kHz • Can hear Doppler shift frequencies Doppler systems • Spectral display • Colour flow Spectral display Frequency shift (kHz) baseline Time (s) Colour flow ‘Triplex’ display Summary of systems and main controls • 2 main types of system are – Spectral Doppler – Colour flow • main controls for spectral Doppler adjust: – position of sensitive region – beam direction – spectral Doppler display • main controls for colour flow adjust: – size and depth of colour box – beam direction – colour display Spectral Doppler Frequency shift (kHz) baseline Time (s) Spectral Doppler - continuous wave (CW) Display R T Doppler signal processor Transducer Sensitive region • Separate transmit and receive elements • Emits ultrasound continuously • Receives ultrasound continuously • Doppler signals from sensitive region Stand alone CW Doppler system: features • • • • No B-mode image No depth discrimination Use for vessels at defined location Use for vessels with characteristic waveform shapes • Obstetric applications - umbilical arteries • Peripheral vascular application - carotid, lower limb CW spectral Doppler examples Arcuate artery Internal iliac External iliac Umbilical 2 vessels in beam Pulsed wave (PW) Doppler systems Doppler signal processor Display Sensitive region Gate depth • Emits ultrasound in pulses Gate length • Depth discrimination • Sensitive region depth and length set by user Stand alone PW Doppler system - features • • • • No B-mode image Depth discrimination Use for vessels at defined location Use for vessels with characteristic waveform shapes • Transcranial Duplex system B-mode + PW Doppler = Duplex Duplex system - features • B-mode and PW Doppler • depth discrimination • all cardiovascular applications • basis for all modern Doppler systems System components and signal processing Display R T Doppler signal processor Tissue Tissue Blood Blood T Received signal Tissue Blood Amplitude R From tissue (Clutter) From blood 4.999 5.000 5.001 Frequency (MHz) Tissue Blood 5.002 Amplitude Demodulation 4.999 5.000 5.001 Frequency (MHz) 5 .002 -1000 Demodulation removes underlying transmit frequency 0 Frequency (Hz) 1000 2000 Lost blood signal High pass filter -1000 0 1000 2000 -1000 0 1000 Frequency (Hz) Filter frequency thresholds Filtering removes the clutter signal 2000 Amplitude 10ms Time Spectrum analysis Doppler frequency Time Spectrum analysis estimates all the frequencies present in the Doppler signal Transducer Received signal 4.999 5.000 5.001 5 .002 Signal processor Demodulator High pass filter Doppler signal -1000 Spectrum analysis Display Spectral display 0 1000 2000 Frequency (MHz) Cut-off filter Filter low End diastolic flow Filter high Loss of end diastolic flow Typical filter values • Obstetrics • Vascular • cardiology 80-100Hz (little arterial movement) 150-200 Hz (some arterial pulsation) 300Hz+ (valves and myocardium) Pulsed wave (PW) Doppler Doppler signal processor Sensitive region Gate depth Gate length Display Doppler signal CW PW Aliasing • Upper limit to detected velocity measured using PW Doppler Max Doppler frequency shift CW Doppler signal PW Doppler signal (lots of samples) PW Doppler signal (2 samples/wavelength) PW Doppler signal (not enough samples) Aliasing Aliasing • Doppler frequency shift estimated correctly when: – at least 2 samples per wavelength – prf > 2 fd • Maximum Doppler frequency shift which can be estimated is half the prf – fd(max) = prf/2 Waveforms in disease • Local disease (Atherosclerosis) • Downstream disease (placental disease) Atherosclerosis Jet Turbulence Quantification 1. Peak velocity Max velocity Measurement of blood velocity I. Transducer θ v v = c fd 2ft cos θ Measurement of blood velocity II. Measurement of blood velocity III. Standard table Diameter stenosis (%) 0 Peak systolic velocity (cm/s) < 90 0 - 15 < 100 15 - 50 < 125 50 - 80 > 135 80 - 99 > 230 Downstream disease Uterine artery Fetus Placenta Spiral/arcuate arteries Abnormal placental development leads to increase in resistance to flow Umbilical waveforms Quantification 2. Waveform shape. Max Mean Min Resistance index (RI) = (max-min)/max Pulsatility index (PI) = (max-min)/mean Estimation of RI Peak systolic marker End diastolic marker Controls for CW, PW and duplex – position of sensitive region (PW, duplex) • gate length, gate depth – beam direction (PW, duplex) • Beam steering angle – spectral Doppler display (CW, PW, duplex) • • • • • gain Filter level Velocity scale Time scale Baseline – Measurement (duplex) • Beam-vessel angle Colour flow Colour flow image • Display of 2D flow image superimposed on B-mode image Colour boxes Sector Colour box Linear array Colour box • Image built up line by line • Each line consists of adjacent sample volumes Colour flow system components Display B-scan processor Colour flow processor Spectral Doppler processor Beamformer Transducer Transmitters Colour flow processor Demodulator Clutter filter Doppler statistic estimator Post processor Blood tissue discriminator Clutter filter clutter blood Frequency (MHz) Frequency (MHz) Frequency estimation • Fast Fourier Transform (64-128 data points) – full frequency spectrum • Autocorrelator (3 data points) – mean frequency – variance – power Post-processor • ‘Persistence’ or ‘Frame-averaging’ – Reduces noise – ‘lag’ in image High persistence Low persistence Value = 0.4 frame 1 + 0.3 frame 2 + 0.2 frame 3 + 0.15 frame 4 + 0.10 frame 5 Value = 0.6 frame 1 + 0.4 frame 2 Blood-tissue discriminator Colour image (mean Doppler frequency) B-mode image Blood-tissue discriminator Colour image (mean Doppler frequency) B-mode image No blood tissue discriminator With blood tissue discriminator Colour modes Colour processor Variance Mean frequency Colour Doppler Power Power Doppler Mean frequency: red-blue scale Mean frequency + variance: red-blue + green Power: no B-mode in colour box Power: with B-mode in colour box Angle dependence θ θ θ Colour Doppler angle dependence Power Doppler angle dependence Angle dependence Doppler amplitude 40o 60o 90o Doppler frequency Clutter filter Angle dependence Aliasing Aliasing Doppler amplitude 3m/s 4m/s 1m/s 2m/s 3m/s Doppler frequency Aliasing limit Aliasing limit Jet Recirculation