Tender_Advt_no_441_2014 - Chittaranjan National Cancer Institute

advertisement

Chittaranjan National Cancer Institute

37 S. P. Mukherjee Road, Kolkata 700026

Tender Advt. no. 441 /2014

The Director, CNCI, invites Sealed Tenders from the original manufacturer/authorised dealer for purchase of the following instrument:

1.

Protein(vertical) and Nucleic Acid(Horizontal) Gel Electrophoresis unit with

Power Pack.

Last date of submitting bids :- -18-08-2014 upto 2pm

Opening of bids: - 18-08-2014-----at 3.30pm

Interested bidders are requested to visit CNCI Website: www.cnci.org.in

for detailed tender notice/information.

Director

For Publication in CNCI Website: www.cnci.org.in

Dual Gel Mini Electrophoresis unit with complete module for elcectroblottting in a common gel tank and Nucleic Acid (Horizontal) Gel Electrophoresis system with Power Pack.

Specifications for Dual Gel Mini Electrophoresis systems

1. Vertical electrophoresis should be comprised with a tank, electrode module, casting module and optional blotting module.

2. This unit should be built by injection mode technology with anti-leak rubber and clasprelease design*

3. Guarantee for leakage free while gel casting. Spacers separately and combs load different volume protein samples.

4. Electrode module should be able to run two gels simultaneously and the blotting module should share the same gel tank.

5. Gel size (mm): 100 x 80 (W x H)

6. Wet Transfer Blot Module; 110 x 75 (W x H)

7. Spacer thickness (mm): 1.0

8. Wet Transfer Blot Module; Sponge size (mm): 110 x 75 (W x H)

9. Applicable number of samples: 8 to 15

10. Number of gels: 2 gels simultaneously

11. Wet transfer blots module with transfer capacity of 2 gels simultaneously

12. Buffer requirement for transfer (ml): 1450

13. Blot module Cassette dimensions (mm): 140 x 11x 77 (w x d x h)

14. Electrode module w/ platinum wire: For 2 gels operation

15. Casting module: 2 sets for gel casting simultaneously

16. Casting base and 2 no silicone seals

17. Glass plate: 5 sets

18. 2 no 1 mm thick 8/12 sample combs

19. Physical dimensions (mm): 200 x 130 x 165 (W x D x H)

20. 2 no. of cassettes with 8 fibre pads

21. Certification for Blot Module: CE

Specifications for Nucleic Acid (Horizontal) Electrophoresis

The submarine cell should be comprised with gel tank, gel tray, top lid, casting module and red/black L type electrode. Unit built by injection mode technology is preferable; must be leakage free while gel casting.

1. Should be provided with Fluorescent Ruler

2. Tray Size (mm): 65 x 70

3. Number of teeth: one 10 x 1.0mm teeth and one 15 x 1.0mm teeth fixed-height comb

4. Thickness of comb (mm): 0.75, 1.0, 1.5

5. Sample throughput (ml): Max. 30

6. Buffer Volume (ml): 200 to 300

7. Cable: Safety black & red cable

8. Physical dimensions (mm): 200 X 95 X 85 (W x D x H)

9. Certification: CE

10. Should be compatible with Power Supply from variety of manufacturers

Specifications for Power Pack

1. Voltage automatic; Current automatic

2. Output Terminals: 4 sets

3. Current adjust steps: 4-500 mA in 1 mA

4. Voltage in 1 volt steps: 10-300 V

5. Output power: 90W max

6. Operating Temperature: 0-40°C

7. Timer: 0-999 mins

8. Display: 3 digit 7 seg LED with Touch Pad and Alarm

9. Input Power: 120/230V, 50/60Hz

10. Physical dimensions (mm): 240 x 320 x 73 (WxDxH)

11. Constant current or power setting. Automatic crossover,

12. Overload protection, short circuit protection

13. Certification: CE

Kolkata based after sale service with proven track record and User list for all the above instruments.

Terms and conditions

1. Cost of tender documents Nil ,EMD Rs. 2400.00 for the Electrophoresis apparatus with power pack is payable in the form of Demand Draft drawn in favour of the Director,

CNCI, Kolkata ,payable at Kolkata. Earnest money will be refunded to the unsuccessful bidders as per rule. No quotation will be considered without EMD.

Earnest money deposited shall be liable to be forfeited if the successful bidder withdraws his tender after opening.

. Tender documents may be downloaded from the official web site : www.cnci.org.in

2. Prices should be quoted in Indian Currency . Rates will be inclusive of VAT and other taxes, only net price is to be quoted.

3 . Quotations should have validity of minimum 90 Days.

4. Bids received after the specified date and time of receipt will not be considered.

5. Non-fulfilment of the conditions of executing the contract by the supplier would constitute sufficient ground for annulment of the award and forfeiture of EMD.

6. No variation of price will be allowed during the period of contract.

7. All supply of goods must conform to the specification.

8. Place of submission of the tender document:-Department of Cancer Chemoprevention

Room No.201,Research Wing of CNCI,Kolkata.-700026.

9. Last date of submission of tender is upto14.00 hrs. on 18-08-2014

10. Tenders will be opened at 15.30 Hrs. on 18-08-2014 in presence of representatives of the Bidders. Bids received after the specified date and time of receipt will not be considered..

In the event of the date being declared as a closed holiday for purchaser’s office, the due date for submission of bids and opening of bids will be the following working day at the appointed time.

11. Every effort has been made to put forth general specifications in this tender document.

If inadvertently, any of the specifications drawn happens to match with the specifications of any particular firm’s product only, in respect of critical parameters, then it will not automatically mean that this particular firm’s offer is only technically suitable.

In general, the specifications offered by other firms will be assessed in their own entirety to ascertain whether or not the broad functions in general expected of the equipments of the purchaser and accordingly the offers would be considered based on prudent assessment of the purchaser.

12. Price Payment Terms.

Payment will be made only after satisfactory installation and the date of acceptance certificate issued by the purchaser’s representative for the respective delivery.

13. The tenderer must have supplied this machine in State/ Central Govt./Autonomous

Body or any Institute.. list of current users in and around Kolkata is to be submitted.

14. The tenderer must have authorised service centre within the city limit of Kolkata/

Howrah, for providing after-sales service of the machine in the hospital. The address, fax,

Telephone No. should be clearly mentioned on the offer.

15. With regard to interpretation of terms and conditions, the decision of CNCI shall be final and binding to the Bidder..

16. CNCI reserves the right to accept or reject any or all quotations without assigning any reason .

17. Settlement of any dispute emanating from the resultant contract is subject to the jurisdiction of Kolkata only.