Supporting Information Available See supplemental Supplemental discussion on detection of N-heterocyclic compounds Generation of formulas was performed with our samples using Agilent Technologies MassHunter qualitative software, which uses discrete mathematics and chemical theory (e.g. valence rules) to produce likely molecular formulas of unknown spectral peaks [Darland et al., 2008; Meija, 2006; Sweeney, 2003]. While simultaneous evaluation of our field blanks determined several software and/or sampling media artifacts, the presence of several N-heterocyclic compounds appeared likely (S/N > 10 relative to field blanks). Following manual extraction of the m/z equivalent to the [M + H]+ ion of several of the N-heterocycles (N-het) and the corresponding isotopes (due to the natural occurrence of 13 C and 15 N) , injection of an authentic standard was performed to determine retention times and relative isotope responses. Using [M + H] + as the quantitation ion, a 5-level calibration was performed using an internal standard (phenylalanine-d8) method, producing a linear response (R2 > 0.99) and good reproducibility (Table 1). The current study calculated the concentrations of N-het with a multi-level calibration using authentic standards and an aromatic amino acid internal standard (phenylalanine-d8) that was spiked onto filters prior to extraction. The concentrations for the three archived samples (stored at – 50 oC until reinjection) discussed in the article were determined using a same-day 5-level calibration response factor with post-sample calibration checks to confirm instrument response. This response factor was further applied to estimated N-het concentrations for all samples (N = 30) from the initial sample analysis and evaluate relative correlations with FAA concentrations. The estimated concentrations were only used for correlative analysis with FAA and are not reported as mass concentration values due to quality control considerations. Only two hydroxy-pyridine isomers (2-HP, 3-HP) were evaluated with authentic standard in this study, therefore the concentration estimate for 3-HP did not rule out the potential contribution from 4-HP. The method development presented here suggests that future work can account for N-het compounds with usage of an alternate (N-het) internal standard (e.g. quinoline-d7), potentially improving method accuracy and precision. For example, on-going work in our laboratory has indicated similar or improved instrumental sensitivity when using quinoline-d7 as the internal standard. A more in-depth evaluation of this topic is currently underway and will be described in a forthcoming publication. Additional MS/MS transitions were confirmed for norharmane (169.0760 m/z) and harmane (183.0917 m/z) in both ambient samples and authentic standards and include the following daughter fragments: 89.0391 m/z, 116.0500 m/z, 141.0570 m/z, and 142.0650 m/z. A partial list of likely fragment structures can be found at the METLIN tandem MS database (http://metlin.scripps.edu). 3-HP displayed both the [M + H – H2O]+ (78.0344 m/z) and [M + H – CO]+ (68.0508 m/z) daughter fragments in ambient samples and authentic standards, while 2-HP was not detected in ambient samples and displayed the [M + H – H2O]+ (78.0344 m/z) fragment more dominantly at 20V collision energy with an authentic standard. The second most abundant fragment was observed at 51.0242 m/z for 2-HP, which is in agreement with tandem MS spectra found at the MassBank spectral repository (www.massbank.jp). Supplementary information on MARGA QA, raw data adjustments, processing of IN concentrations and average MARGA compound concentrations during the sampling period Mass flow controlled MARGA air flow rates were verified weekly by measuring the flow rate at the atmospheric inlet using a NIST traceable primary standard (DryCal DC-LITE flowmeter, Bios International Corporation, Butler, NJ). MARGA inlets and air sampling tubing were cleaned with DDI water and dried with zero grade air weekly. Following the completion of the experiment, a liquid blank was analyzed by sampling the absorption solution with the air pumps disconnected and denuder inlets sealed. Additionally, an external standard test was performed by replacing the absorption solution with a known liquid standard containing NH4+, NO3-, and SO42- with the air pumps disconnected and denuder inlets sealed, to verify analytical accuracy as controlled by the LiBr internal standard. Prior to calculation of final air concentrations, raw data were adjusted based on differences between the measured and MARGA recorded air flow rates, liquid blanks and biases in calibration standard checks. The average measured flow rates were 16.48 l min-1 and 15.98 l min-1 for MARGA unit (MU) 1 and MU 2, respectively. The MARGA calculates concentrations based on a flow rate of 16.7 l min-1, therefore the raw data concentrations were adjusted by 1.4% (MU1) and 4.5% (MU2), respectively. Liquid blanks for all the compounds (SO42-, NO3-, and NH4+) ranged from 0-0.15 µg m-3 for both MU1 and MU2. External standard accuracy % for SO 42and NO3- was very good ranging from 104.5%-111.2% for MU1 and 103.5%-110.7% for MU2. However, the external standard accuracy % was not as good for NH4+, with values ranging from 77.3% -85.5 % for MU1 and 75.6%-78.9% for MU2. The poorer performance of the MUs in measuring NH 4+ is likely due to consumption of NH4+ by bacteria in the MUs. Consequently, NH3 and NH4+ were adjusted for the external standard results as an offset. To determine the extent to which NH3 and NH4+ concentrations were affected by bacterial consumption, concentration trends between the denuder/filter pack and the MARGA were examined. In addition for NH 4+, NH4+/ SO42- ratios were also examined. One-hour concentrations for each MU were averaged to correspond with ON sampling periods. Hourly detection limit values were 0.04 µg m-3 for SO2, 0.1 µg m-3 for HNO3, 0.05 µg m-3 for NH3, 0.06 µg m-3 for SO42-, 0.1 µg m-3 for NO3-, and 0.05 µg m-3 for NH4+. Precision between the two duplicate MARGA units was assessed by calculating the absolute relative percent difference between each corresponding sampling period, and then taking the median value. The precision was < 10% for SO2, SO42- and NH4+, and <25% for NH3, HNO3, and NO3-. The average concentrations of compounds measured during the ON sampling period by the MARGA were 539 ng-N m-3 for NH3, 185 ng-N m-3 for HNO3, 706 ng-N m-3 for NH4 and 108 ng-N m-3 for NO3-. SO42- average concentrations during the ON sampling period was 2568 ng m-3. Figure S1. FAA percent distribution for subset of samples (N=17) processed for CAA. Figure S2. HNO3 + NO3 least squares correlation with percent serine (ser) + glutamine (glut) + threonine (threo). Figure S3. Winter sample results. Figure S4. N-het extracted ion chromatograms (EIC) of PM2.5 collected at RTP. Supplemental Table S1. List of target AA with compound formula, molecular weight (MW), quantitation ion, and internal standards. Method details available in Samy et al (2011). Name Formula MW (g mol-1) Quantitation Ion [M + H]+ Internal Standard Glycine C2H5NO2 75.03 76.0401 Alanine C3H7NO2 89.05 90.0554 Serine C3H7NO3 105.04 106.0502 Glycine-2,2-D2 Alanine2,3,3,3-D4 Serine-D3 Proline C5H9NO2 115.06 116.0708 Valine C5H11NO2 117.08 118.0864 C4H9NO3 119.06 C3H7NO2S Threonine Cysteine 1 Hydroxyproline Isoleucine Leucine Asparagine Ornithine 1 1 1 MW (g mol-1) 77.08 Quantitation Ion [M + H]+ 78.0526 93.12 94.0803 108.11 109.0689 Histidine-U-13C6 161.15 162.0958 125.2 126.1362 120.0655 Valine-D8 Serine-D3 108.11 109.0689 121.02 122.0271 Cysteine-D2 123.17 124.0394 C5H9NO3 131.05 132.0654 Glycine-2,2-D2 77.08 78.0526 C6H13NO2 131.09 132.1020 Leucine-D3 134.19 135.1204 C6H13NO2 131.09 132.1021 134.19 135.1204 C4H8N2O3 132.05 133.0605 Leucine-D3 Serine-D3 108.11 109.0689 138.2 139.1342 136.12 137.0630 108.11 109.0689 C5H12N2O2 132.09 133.0970 Aspartic Acid C4H7NO4 133.04 134.0446 Glutamine1 C5H10N2O3 146.07 147.0758 Ornithine-D6 Aspartic Acid2,3,3-D3 Serine-D3 Lysine C6H14N2O2 146.10 147.1128 Arginine-D7 Glutamic Acid C5H9NO4 147.05 148.0598 Glutamic Acid2,3,3,4,4-D5 152.16 153.0909 Methionine C5H11NO2S 149.05 150.0578 Methionine-D3 152.23 153.0763 Histidine C6H9N3O2 155.07 156.0761 161.15 162.0958 Phenylalanine C9H11NO2 165.08 166.0853 Histidine-U-13C6 PhenylalanineD8 173.24 174.1352 Arginine C6H14N4O2 174.11 175.1184 181.25 182.1621 Tyrosine C9H11NO3 181.07 182.0799 173.24 174.1352 Tryptophan1 C11H12N2O2 204.09 205.0957 Arginine-D7 PhenylalanineD8 PhenylalanineD8 173.24 174.1352 Cystine C6H12N2O4S2 240.02 241.0291 Cysteine-D2 123.17 124.0394 1 added to NIST mixture. 182.1621 Supplemental Table S2 Speciated compound results for RTP, North Carolina. Amino Acid Glycine Alanine Aspartic Acid Arginine Glutamic Acid Serine Glutamine Threonine Valine Proline Tyrosine Asparagine Hydroxyproline Methionine Histidine Phenylalanine Leucine Isoleucine Lysine Ornithine Total PM2.5 FAA (ng m-3) 3.9±2.6 1.4±0.9 1.4±0.8 1.3±0.9 0.7±0.7 0.7±0.5 0.4±0.7 0.4±0.2 0.2±0.1 0.2±0.1 0.1±0.1 0.1±0.1 0.1±0.1 Trace Trace Trace ND ND ND ND 11±6 Minimum (ng m-3) Maximum (ng m-3) 0 0.3 0.2 0.4 0.2 0.1 0 0.1 0 0 0 0 0 0 0 0 0 ND ND ND 2 12.1 4.3 3.8 3.1 3.9 1.8 3.6 1.1 0.4 0.5 0.3 0.5 0.2 0.4 0.3 0.2 0.1 ND ND ND 31 %FAA 35.66 12.46 12.37 11.71 6.73 6.50 4.08 3.40 1.43 1.41 1.02 1.01 0.52 0.32 0.52 0.35 0.15 0.00 0.13 0.00 100 PM2.5 CAA (ng m-3) 8.2±5.3 18.2±9.1 2.1±1.4 Trace 7.7±5.5 0.9±0.8 -3.1±2.3 0.6±0.3 0.9±0.6 1.9±1.9 -Trace 0.7±0.8 0.1±0.1 0.4±0.2 0.6±0.3 0.3±0.2 Trace Trace 46±21 Minimum (ng m-3) 0 5.1 0 0 2.9 0 -0.3 0.2 0 0 -0 0 0 0.1 0.3 0.1 0 0 14 Maximum (ng m-3) 18.0 37.2 5.6 0.3 26.7 3.3 -8.3 1.5 1.9 5.7 -0.7 3.2 0.4 0.8 1.7 0.8 0.5 0.4 92 %CAA 17.74 39.65 4.54 0.04 16.70 2.04 -6.74 1.21 1.92 4.21 -0.09 1.48 0.28 0.93 1.37 0.70 0.31 0.04 100