EECE 206 Page 1 of 4 Lab 3a: Oscilloscope and Signal Generator Use Laboratory Goals Familiarize students with the oscilloscope Familiarize students with the function generator Apply a signal to a resistive circuit and measure output frequency and amplitude Pre-lab reading Student Reference Manual for Electronic Instrumentation Laboratories by Stanley Wolf and Richard Smith, Copyright 1990. XYZ’s of Analog and Digital Oscilloscopes published by Tektronix, Copyright 1992. Equipment needed Lab notebook, pen Tektronix 2252 Digital Oscilloscope Tektronix FG 501A Function Generator 2 oscilloscope probes (already attached to the oscilloscope) 1 BNC/BNC test lead 1 BNC/EZ Hook test lead Parts needed Circuit breadboard Resistors (2), 1k Ohm, ¼ Watt Lab safety concerns Make sure before you apply power to a circuit, all connections are correct, and no shorted wires exist. Do not short the function generator signal and ground connections together Do not touch the circuit wiring while power is applied to it EECE 206 Page 2 of 4 Lab 3a: Oscilloscope and Signal Generator Use 1. Oscilloscope Operation Locate the Tektronix 2252 Digital Oscilloscope Press the brown POWER button to turn on the oscilloscope Press the STORE/RECALL SETUP button in the MENU section of the oscilloscope Press the button next to the RECALL option on the screen (this forces the oscilloscope to a default setting) Press the AC button in the COUPLING section of CH 1 (the button’s light turns off) Adjust the vertical POSITION knob for CH 1 so that the green signal line is displayed on the middle horizontal line of the screen Adjust the display knobs (A INTEN, FOCUS, READOUT), located below the screen, so that the signal and oscilloscope settings are bright and in focus Press the AC button again to enable CH 1 (the button’s light turns on) Clip the CH 1 probe to the PROBE ADJUST point (just to the lower right of the screen). A square wave calibration signal now appears on the screen Clip the probe’s ground clip to the silver part of the CH 3 input connector (this an easy place to attach the ground clip to the oscilloscope ground) Adjust the CH1 VOLTS/DIV knob so that the display now reads “.5V~” Adjust the A AND B SEC/DIV knob so that it now reads “.5ms” Press the VOLTMETER button in the MEASUREMENTS section of the oscilloscope Press the button next to the PK-PK option on the screen Record the peak-to-peak voltage of the square wave in your lab notebook Press the CLEAR DISPLAY button to clear the screen measurement Readjust the CH1 VOLTS/DIV knob so that the display reads “.2V~” Readjust the A AND B SEC/DIV knob so that it now reads “.2ms” Press the COUNTER/TIMER button in the MEASUREMENTS section of the oscilloscope Press the button next to the FREQ option on the screen Record the frequency of the square wave in your lab notebook Press the CLEAR DISPLAY button to clear the screen measurement Press TIME button in the MEASUREMENTS section of the oscilloscope Press the button next to the 1/SEC (i.e., Hz) option on the screen Turn the CURSORS POSITION knob so that the two cursors lines move to the left, and line up with the high-to-low transitions of the square wave Turn the CURSORS POSITION knob so that the right cursor line moves to the left, and lines up with the low-to-high transition of the square wave (notice the frequency reads 2x the original value, since only ½ the waveform is measured) Press the CLEAR DISPLAY button to clear the screen measurement EECE 206 Page 3 of 4 Lab 3a: Oscilloscope and Signal Generator Use Press the CH2 button in the VERTICAL/MODE section of the oscilloscope to enable CH2 (the button’s light turns on, and a signal line appears on the screen) Press the CH1 button in the VERTICAL/MODE section of the oscilloscope to disable CH1 (the button’s light turns off, and the square wave disappears) Adjust the vertical POSITION knob for CH 2 so that the green signal line is displayed on the middle horizontal line of the screen Press the AC button in the COUPLING section of CH 2 (the button’s light turns on) Unclip the CH1 probe and ground clip from the oscilloscope, and set it aside Disconnect the CH2 probe from the oscilloscope connection (twist the connector ¼ turn counter-clockwise, and gently pull the connector off) 2. Function Generator Operation Turn on the TM 504 module Locate the FG 501A Function Generator Press the white ~ (sine wave) output button to select a sine wave output Press the white FREE RUN button to select a continuous output Press the white –20dB button to decrease the voltage output of the generator Turn the gray MULTIPLIER dial on the generator to the 10^3 (i.e., 1kHz) setting Turn the AMPL knob fully counter-clockwise (i.e., output voltage is minimum) Connect the BNC/BNC test lead between the function generator OUTPUT and oscilloscope’s CH2 input Press the SOURCE button in the TRIGGER section of the oscilloscope to select CH2 as the signal source for triggering Adjust the CH2 VOLTS/DIV knob so that the display now reads “.5V~” Adjust the function generator AMPL knob until the amplitude of the CH2 oscilloscope signal is approximately 1Vpp (use the oscilloscope’s voltmeter function as you did above) Turn the FREQUENCY dial on the generator until the frequency of the CH2 oscilloscope signal is approximately 1kHz (use the oscilloscope’s counter/timer function as you did above) 3. Circuit Construction and Signal Measurement Remove the BNC/BNC test lead from the signal generator and the oscilloscope Connect the BNC/EZ Hook test lead to the signal generator Construct the two-resistor circuit as shown on the following page Connect the red and black EZ Hook ends of the test lead to the circuit input and ground connections respectively as shown in the photo EECE 206 Page 4 of 4 Input Lab 3a: Oscilloscope and Signal Generator Use R1 R2 Ground Voltage Divider Circuit with the Input Signal Connected Reconnect CH2 probe to the oscilloscope Clip the probe to the top of R2 and the ground clip to the bottom (ground) side of R2 Adjust the oscilloscope to measure the frequency and amplitude of the signal (use what you have learned about the oscilloscope!) Record the values in your lab notebook Experiment with the different features of the oscilloscope before ending the lab, and turning off the equipment Record your observations in your lab notebook 4. Analysis Write a brief summary report for this lab. Be sure to also include the following topic: Explain any difficulties you had with this lab. (Please include suggestions to improve the lab, if you have them). Before leaving the lab, take a few minutes to make sure all equipment and test leads are returned to your cabinet, and that you have cleaned up your work space.