LIGHT OUTPUT CHARACTERISTICS OF SPRING HEALTH LED
BASED CURING LAMPS
By Noureddine Melikechi
Professor of Physics
Delaware State University
nmelikechi@gmail.com
Phone: 302 943 7213
Submitted to
Spring Heath Products, PA
August 2011
Experimental Setup
The power was measured with the Coherent Lab_Max Top Laser/Energer
meter.
The Diameter of the power head was smaller compared to the beamdiameter
of the LED.
The Data was fitter with the Erf function to find the beam waist.
We noticed that the beam waist did not change significantly for the position
of the power meter varying from 0mm to 4mm.
Lens_ 1: Are measruement without the plastic lens.
Lens _2: are measurements with the plastic lens.
Note: Lens_1_Knife_Edge_n (n refers to the distance in mm between the
light source and the detector.)
Figure: Experimental set up for the charaterizations of light sources
LENS_1 (no plastic lens)
LENS_2 (Plastic lens)
Spectrum