LCLS CDR 2/17/16 9.2.3.2 Warm dense matter and plasma 9.2.3.2.1 Requirements The Warm Dense Matter (WDM) experimental program studies dense matter rapidly heated to temperatures exceeding 10 eV/atom. Initial studies will use the LCLS X-FEL to heat thin samples, whose properties will then be measured by an array of diagnostics placed around the sample. These diagnostics include an x-ray spectrometer, and a Fourier Domain Interferometer (FDI) driven by a short-pulse optical laser. Interpretation of these measurements requires knowledge of the x-ray photon flux incident on the sample and the x-ray photon flux transmitted through the sample. The experiment requires the use of the highest energy LCLS X-FEL photons, at 8.275 keV, in order to have enough penetrating power to uniformly heat the sample from front to back. The required flux density at the sample is TBD photons/centimeter2. The heated volume needs to have a radius of > 5 microns in order to create a large enough sample for study. Since the minimum required flux density is TBD x the peak flux density of the 8.275 keV beam in the near hall, the experiment requires x-ray focusing optics to achieve the required flux densities. The use of focusing optics adds additional requirements for alignment, which include the ability to image the spot at the position of the sample and the capability of running at high repetition rates at reduced intensity during the alignment process. -1- LCLS CDR 2/17/16 250 mm aperture Incident Beam Monitors Back-scatter x-ray spectrometer100 mm thick sample Imaging detector FDI Laser FEL Beam Focusing Optic Variable beam attenuator FDI Spectrometer 50-100 mm aperture Imaging detector (on sample holder) Outgoing Beam Monitor Figure 9.2.3.2.1, A schematic of the warm dense matter experiment Figure 9.2.3.2.1 shows a schematic of the WDM experiment including its auxiliary hardware. The X FEL enters from the left and first passes through an aperture of slightly larger diameter which blocks most of the non FEL light, then through an adjustable attenuator, an incident beam intensity monitor and on to the focusing optic. The focusing optic is a variant of a blazed Fresnel zone-plate operating like a thin lens in visible optics. After passing though the lens, the light passes through another incident beam monitor, which serves to measure the intensity of the light directed to the sample. A smaller aperture, closer to the sample, removes stray light scattered by the optic. The samples are arranged in a grid on a movable stage at the position of the focal spot. The sample stage also holds an imaging detector that can be moved into position for alignment and characterization of the focal spot. The diagram shows the back-scatter x-ray spectrometer that measures the x-ray spectra emitted by the sample and the path of the optical FDI laser beam used to characterize the expansion of the sample. X rays passing through the sample are measured by the downstream beam monitor and imaging detector. -2- LCLS CDR 2/17/16 9.2.3.2.2 Optical Design and Components 9.2.3.2.2.1 Focusing System 9.2.3.2.2.1.1 Optical design The WDM experiment will be situated in the 4th shielded room in the near hall. Table 1 shows the distances that are relevant for the optical design. The shielded rooms are 13 meters long, but the PPS beam-stop limits the length available to experimenters to 10.6 meters. The sample is located approximately in the center of the room, 6.2 meters from the upstream wall. It turns out that the flux requirements can be met with lenses whose focal lengths are longer than the distance from the sample to the upstream wall, but in the interests of minimizing interference between experiments, we have chosen a position for the lens that is as far as possible from the sample, but still within the confines of the WDM shielded room. Table 9.2.3.2.1. Distances and specifications associated with the warm dense matter optical components Distances From To units Upstream wall Upstream wall Downstream wall PPS Beam Stop 13.00 10.61 m m Upstream wall Upstream wall Lens Sample 1.20 6.20 m m End of undulator Source at 8.275 Kev Upstream wall End of undulator 72.11 19.50 m m Source at 8.275 Kev Lens Lens Sample 92.80 5.00 m m 4.74 131.12 5.72 m microns FWHM microns FWHM 2.74 6.34 microns FWHM microns FWHM Optic Specification Lens Focal length Beam size at lens position Diffraction limited spot Thin lens image size Approximate focused spot size The lens, placed 1.2 meters from the upstream wall, forms an image of the X-FEL source at the sample. The source-to-lens and lens-to-image distances determine its focal -3- LCLS CDR 2/17/16 length. The X-FEL source point is at the position of the waist in the Gaussian model which is located one Rayleigh length upstream of the exit of the undulator. As the table shows, the resulting source-to-lens distance is 92.8 meters, and the lens-to-image distance is 5 meters, requiring a lens focal length of 4.74 meters and giving a magnification of 0.05. -4- LCLS CDR 2/17/16 Figure 9.2.3.2.2. Details of the lens design Figure 9.2.3.2.2 shows details of the lens design. The lens is carved into the face of a C disk and mounted over a hole drilled through a 25.4 mm diameter Cu mount. The C disk is 650 microns thick except in the center where it thins down to 400 microns. The active portion of the lens is 200 microns in diameter and consists of 6 concentric grooves machined to a maximum depth of 18.8 microns. The plot of the beam profile at the lens, figure 2g, shows that the 200 micron lens diameter nicely captures most of the beam. The shape of the grooves was determined by calculating, at the position of the lens, the phase change necessary to convert the diverging Gaussian X-FEL beam from the undulator to a converging Gaussian waveform whose waist is at the sample position. The radial phase profile was converted to a depth profile by multiplying by the optical constant for C which, at 8.275 KeV, is 18.8 microns / 2 radians phase change ( with respect to vacuum.) The stage used to position the lens will have positioning precision of < 5 microns and angular precision of < 1 mRad. 9.2.3.2.2.1.2 lens fabrication Diamond tool Diamond tool carves lens pattern here Reference flat Al rod Al rod viewed in cross section Figure 9.2.3.2.3. Lens manufacturing technique The C lens will be machined from a C rod by single-point diamond turning as illustrated in figure 9.2.3.2.3. The tool first machines the end of the rod flat to provide a reference surface. Then a disk of material is removed from the center providing a recessed area for the protection of the lens. Then the lens profile is machined into the base of the recess. -5- LCLS CDR 2/17/16 Glue diamond turned rod end to glass Glass slide with surface polished to level of B 4 C Mounting Block Shim B 4C Polish back of lens to level of B 4 C Figure 9.2.3.2.4. Mounting the lens After diamond turning, the end of the rod is carefully sliced off and its reference flat is glued onto the polished glass surface of a specially prepared thinning fixture illustrated in figure 4. The thinning fixture consists of a glass slide mounted on a mounting block on top of a Cu shim whose thickness is equal to the desired final thickness between the bottom of the lens, and the reference flat. Alongside of the blocks are a set of B4C (boron carbide) sticks whose surfaces have been previously polished flat and parallel. Because the B4C is much harder than the glass it is easy to polish the surface of the glass so that it is level with the B4C. The reference surface of the lens is then glued to the glass surface with a very thin glue. The shim is removed, lowering the mounted lens so that the desired back surface of the lens is in the plane of the B4C sticks. The lens is then polished to this level and removed for mounting over the hole in the Cu mount. 9.2.3.2.2.1.3 Lens Survivability. C was chosen as the material for the lens based on its capability to withstand the energy density of the full X FEL beam. Table 2 is a list of the lowest Z materials amenable to single-point diamond-turning. The table gives the optical constants, the calculated dose from the peak of the full beam at the lens position, and the dose needed to bring the material up to its melting temperature. Cu and Al are ruled out because the doses to these materials are enough to bring them to their melting temperature. C has both a low dose and high melting point, and therefore will survive the full beam intensity. -6- LCLS CDR 2/17/16 The dose to polystyrene is also very low, but it's large phase shift length results in too large an aspect ratio for single-point diamond-turning. Table 9.2.3.2.2. Doses to low-Z materials amenable to diamond turning l = 0.15 nm 1 pi phase shift atten dose dose dose length length at lens to mp* through mp micron micron eV/atom eV/atom eV/atom Cu 3.27 24.1 0.23 0.21 0.35 Al 9.34 85.7 0.09 0.09 0.2 C 9.41 997 0.004 0.86 0.86 9.2.3.2.2.2 Apertures Survivability is also an issue in the design of the two apertures. The basic concept is to utilize a laminate consisting of 4 mm of B4C, 150 microns of Al, and 200 microns of Ta. As shown in table 9.2.3.2.2, this laminate has sufficient absorption to block x-rays up to the 3rd harmonic. Furthermore the B4C attenuates the direct X-FEL beam enough to prevent damage to the Al which further attenuates the beam enough to prevent damage to the Ta. -7- LCLS CDR 2/17/16 Optional hi-pass Be filter 4 mm B4C 150 mm Al 200 mm Ta 1 mm 500 mm 250 mm 100 mm Fixed 1 mm apetrure Selectable Aper tures onm oving stage Figure 9.2.3.2.5. Apertures for the warm dense matter experiment A series of holes having diameters from 1 mm down to 100 microns will be drilled through the laminate which will then be mounted on a movable stage that provides both rotation and translation of the laminate. A second, fixed, laminate having a single 1 mm diameter hole keeps light from passing through all but a single hole in the movable laminate as shown in figure 9.2.3.2.5. The ability to rotate the laminate is necessary because of the large aspect ratio of the holes. Using a downstream intensity monitor, and starting with the largest diameter hole, the movable laminate will be rotated into a position that maximizes the signal. The laminate will be shifted to the next smaller diameter hole and rotated again to achieve highest intensity downstream. This process will be repeated with successively smaller holes until the hole of the desired diameter is positioned and aligned. The stages used to position the apertures will have positioning precision < 10 microns and angular precision of < 1 mRad. We note that, because the first harmonic of the spontaneous emitted undulator spectrum has the same divergence as the coherent line, and both lines overlap in space, only even harmonics and the off axis intensity of odd harmonics will be stopped in apertures. Edges interfering with the coherent beam will cause diffraction patterns, so that apertures may be restricted to >2 times the local coherent beam size.. -8- LCLS CDR 2/17/16 9.2.3.2.2.3 Attenuator The WDM experiment requires a local attenuator for calibration and to prevent damage to sensitive components during alignment. Preventing damage at the focused spot requires the highest attenuation. Table 9.2.3.2.3 shows an estimate of the dose to a selection of materials at the focal spot in comparison to their tolerable doses (10% of melting temperature.) To reduce the dose to Cu to a tolerable level requires an attenuation of at least 10-4. Table 9.2.3.2.3. Estimated dose to given materials at the focal spot Material Cu Si Al C B4C polystyrine Be dose at focused spot eV/atom Tolerable dose eV/atom Needed attenuation Needed Thickness B4C mm 207.68 115.91 82.11 2.90 1.93 1.93 0.97 0.021 0.026 0.009 0.086 0.068 0.010 0.023 1.0E-04 2.2E-04 1.1E-04 3.0E-02 3.5E-02 5.2E-03 2.4E-02 16.3 14.9 16.1 6.2 5.9 9.3 6.6 The table also shows the thickness of B4C required to provide the needed attenuation. Macroscopic thicknesses of B4C make good attenuators for 8.275 KeV radiation and, as discussed in the section on apertures, can withstand the full (unfocused) beam intensity. Table 9.2.3.2.4. Thickness of Boron carbide required for attenuation desired attenuation 0.1 0.2 0.3 0.4 0.5 0.6 0.7 0.8 0.9 1 -9- B4C thickness mm 4.1 2.8 2.1 1.6 1.2 0.9 0.6 0.4 0.2 0.0 desired attenuation 1.E-01 1.E-02 1.E-03 1.E-04 1.E-05 1.E-06 1.E-07 1.E-08 1.E-09 1.E-10 B4C thickness mm 4.1 8.1 12.2 16.3 20.4 24.4 28.5 32.6 36.7 40.7 LCLS CDR 2/17/16 Relative calibration of the downstream intensity monitors is best done with a combination of variable attenuators that provide both linear and logarithmic variations in intensity. Table 4 shows the sets of B4C thicknesses needed to vary the attenuation linearly from 0.1 to 1 and logarithmically from 10-1 to 10-10. Log attenuator 4 mm Beam B4C 1 cm 32.6 mm 4 mm Linear attenuator Figure 9.2.3.2.6. A linear attenuator The attenuators will be fashioned from single plates of B4C milled in a staircase pattern to the thicknesses specified in the table as shown in figure 9.2.3.2.6. The linear and logarithmic attenuators will be mounted on separate translation stages allowing all combinations of linear and logarithmic attenuation to be applied. The attenuator translation stages will provide motion in the X and Y directions with a precision of < 1 mm. 9.2.3.2.2.4 Beam intensity monitors The beam intensity monitors are required to measure the absolute flux incident on the samples and the amount of flux transmitted through the samples. These monitors will be of the simple ion chamber type as described in the common diagnostics section (9.2...). - 10 - LCLS CDR 2/17/16 9.2.3.2.3 Beam Diagnostics 9.2.3.2.3.1 Beam intensity monitors The beam intensity monitors are required to measure the absolute flux incident on the samples and the amount of flux transmitted through the samples. These monitors will be of the simple ion chamber type as described in the common diagnostics section. 9.2.3.2.3.2 Imaging detectors The WDM experiment requires imaging detectors to assist in the alignment and to determine the size of the heated volume. Beam transport simulations indicate that a spatial resolution of <5 microns is necessary for these purposes. Furthermore it is desirable to obtain the entire image in a single shot. It would also be desirable if the detector could take the full beam intensity but since the alignment could be done in an attenuated beam this is not necessary. 9.2.3.2.4 Layout and other equipment The equipment necessary to initiate the experiments is found in the requirements document, , specifically as a layout and hardware list. - 11 -