UREA UV Method (5 + 1 Reagent) METHOD and PRINCIPLE Enzymatic 2-Point determination in the UV-range, according to the following reactions: Urease Urea + H20 2NH3 + CO2 GLDH 2NH4+ + 2-Ketoglutarate + 2NADH 2L-Glutamate + 2NAD+ + 2H20 REFERENCE VALUES In Serum / Plasma Adults Global Women Men Children 1 - 13 Years 14 - 19 Years (GLDH = Glutamate dehydrogenase Urea/Creatinine-Ratio in Serum REAGENTS Composition (concentrations in the test) 25 – 40 [(mmol/l)/(mmol/l)] R1: R2: Standard : TRIS-Buffer( pH 7,8) 2-Oxoglutarate ADP Urease GLDH (Glutamatdehydrogenase) NADH 120 mmol/l 7 mmol/l 0,6 mmol/l 6 kU/l 1 kU/l 0,25 mmol/l 50 mg/dl (8,33 mmol/l) PRECAUTIONS - For in vitro diagnostic use only. The reagents contain < 0,95 g/L sodium azide. Avoid contact with skin and/or mucous membranes - STABILITY OF REAGENTS When stored at 2-8° C and protected from light, the reagents are stable until the expiry date stated on the label PREPARATION AND STABILITY OF WORKING REAGENT • One-reagent procedure Mix 5 volumes of R1 with 1 volume of R2. (Let monoreagent stand for 10 min before use) Stability : [mg/dl] [mmol/l] 17- 43 15 - 43 18 - 55 2,8 - 7,2 2,6 - 6,7 3,0 - 9,2 11 - 36 18 - 45 1,8 - 6,0 2,9 - 7,5 = 20 – 35 [(mg/dl)/(mg/dl)] 26 – 43 g/day (0,43 – 0,72 mol/day) In Urine Note: It is recommended for each laboratory to establish and maintain its own reference values. The given data here are only an indication. ASSAY PROCEDURES This reagent can be used manually (see method below) and on most analyzers. Applications are available on request. Wavelength : 340 nm, Hg 334nm, Hg 365 nm Temperature : 37°C Cuvette : 1 cm light path Read against distilled water • One-reagent procedure Working reagent Sample/Standard 1 mL 10µL Mix and read the change of absorption (A) between 30 seconds and 90 seconds: A = A 1 - A 2 • Two-reagent procedure 5 days at 15-25°C 4 weeks at 2-8°C Reagent R1 1000 µL Sample/Standard 10 µL Mix and incubate about 5 min , the add : Reagent R2 200 µL SAMPLES Serum, Plasma (EDTA / heparin) Urine diluted 1+100 with distilled water. (Do not use anticoagulants containing fluoride or ammonium ions). Mix and read the change of absorption (A) between 30 seconds and 90 seconds: A = A 1 - A 2 CALCULATION A Sample A Standard x C mg/dL g/L mmol/L C = 50 C = 0.5 C = 8.325 (C = standard concentration) For Urine note dilution, i.e. multiply by dilution factor Page 1/2 Greiner Diagnostic GmbH - Unter Gereuth 10 – D-79353 Bahlingen – Germany www.greiner-diagnostic.com BIBLIOGRAPHY CALIBRATORS & CONTROLS For the calibration of automated analyzers Greiner Multicalibrator is recommended, for quality control use Greiner normal and abnormal control, Unitrol I and Unitrol II. 1. 2. 3. PERFORMANCE DATA - Analytical range The reagent is linear from 2 – 400 mg/dL (0.33 – 66.6 mmol/L) in Serum / Plasma or 30 g/dL (5.0 mol/L) in urine. Thomas L. Clinical Laboratory Diagnostics. 1st ed. Frankfurt: TH-Books Verlagsgesellschaft; 1998, p. 374-7. Burtis CA, Ashwood ER, editors. Tietz Textbook of Clinical Chemistry. 3rd ed. Philadelphia: W.B Saunders Company; 1999. p. 1838. Talke H, Schubert GE. Enzymatische Harnstoffbestimmung in Blut und Serum im optischen Test nach Warburg. Klin Wschr 1965;43:174-5. SYMBOLS USED - Detection limit The detection limit is equal to 2 mg/dL For in vitro diagnostic medical use - Precision Within-run reproducibility N = 11 Mean mg/dL Control 1 42,4 SD mg/dL 0,76 CV % 1,79 Batch Code Use by Control 2 145 1,22 0,84 Patient 86,1 2,39 2,77 Between-run reproducibility N = 11 Mean mg/dL Control 1 42,4 Temperature limitation SD mg/dL 0,55 CV % 1,29 Control 2 144 1,50 1,05 Patient 86,5 1,44 1,66 - Correlation A comparative study has been performed between the Greiner method and another commercial reagent on 26 human serum samples. The parameters of linear regression are as follows: y = 1.000x + 1.450 mg/dl r = 0.999 INTERFERENCES - Ascorbic acid: no interference until 20 mg/dL - Bilirubin: no interference until 40 mg/dL - Hemoglobin: no interference until 500 mg/dL - Triglycerides: no interference up to 2000 mg/dL Page 2/2 Greiner Diagnostic GmbH - Unter Gereuth 10 – D-79353 Bahlingen – Germany www.greiner-diagnostic.com