Official reprint from UpToDate® www.uptodate.com ©2011 UpToDate® High-frequency ventilation in adults Authors Anthony J Courey, MD Robert C Hyzy, MD Section Editor Polly E Parsons, MD Deputy Editor Kevin C Wilson, MD Last literature review version 18.3: September 2010 | This topic last updated: August 2, 2010 INTRODUCTION — High-frequency ventilation (HFV) is a form of mechanical ventilation that combines very high respiratory rates (>60 breaths per minute) with tidal volumes that are smaller than the volume of anatomic dead space [1]. The indications for HFV are limited and its use should be coordinated by clinicians who have experience using HFV. This topic review describes the different types of HFV, as well as patient selection, efficacy, and potential harms. Alternative modes of mechanical ventilation are described separately. (See "Modes of mechanical ventilation".) TYPES OF HFV — There are four basic types of HFV: high frequency jet ventilation, high frequency oscillatory ventilation, high frequency percussive ventilation, and high frequency positive pressure ventilation (figure 1). High frequency jet ventilation — High frequency jet ventilation (HFJV) refers to HFV delivered using a jet of gas (figure 1). It is initiated by inserting into the lumen of the endotracheal tube a small (14 to 16 gauge) cannula, which is connected to a specialized ventilator. An initial pressure of approximately 35 pounds per square inch (psi) drives the jet of gas from the cannula with an initial respiratory rate of 100 to 150 breaths per minute and an inspiratory fraction less than 40 percent (figure 2). The inspiratory fraction is the inspiratory time divided by the sum of the inspiratory and expiratory times. Applied positive endexpiratory pressure (PEEP) can be added if needed. An arterial blood gas should be measured approximately 15 minutes after the initiation of HFJV: • • • Appropriate adjustments when the arterial carbon dioxide tension (PaCO2) is elevated include: increasing the driving pressure in 5 psi increments to a maximum of 50 psi, increasing the inspiratory fraction in 5 percent increments to a maximum of 40 percent, increasing the frequency in 10 breaths per minute increments to a maximum of 250 breaths per minute, or adding an another mode of mechanical ventilation (see "Modes of mechanical ventilation") Appropriate adjustments when the PaCO2 is low include: decreasing the driving pressure in 5 psi decrements, decreasing the inspiratory fraction in 5 percent decrements to a minimum of 20 percent, or decreasing the frequency in 10 breaths per minute decrements to a minimum of 100 breaths per minute Appropriate adjustments when the arterial oxygen tension (PaO2) is low include: adding applied PEEP in 3 to 5 cmH2O increments, increasing the driving pressure by 5 psi increments to a maximum of 50 psi, or increasing the inspiratory fraction in 5 percent • increments to a maximum of 40 percent Appropriate adjustments when the PaO2 is high include: decreasing the fraction of inspired oxygen (FiO2) or decreasing applied PEEP A respiratory rate of approximately 150 breaths per minute is generally required during HFJV. Use of ultrahigh frequency jet ventilation (180 to 400 breaths per minute) in patients with ARDS has been reported, although the study had numerous important limitations [2]. HFJV always requires sedation and usually requires pharmacologic paralysis also. (See "Sedative-analgesic medications in critically ill patients: Selection, initiation, maintenance, and withdrawal" and "Use of neuromuscular blocking medications in critically ill patients".) High frequency oscillatory ventilation — High frequency oscillatory ventilation (HFOV) uses an oscillatory pump to deliver a respiratory rate of 3 to 15 Hertz (up to 900 breaths per minute) through the endotracheal tube (figure 1). This rate is so fast that the airway pressure merely oscillates around a constant mean airway pressure. The respiratory rate is set directly by the clinician. The mean airway pressure is set by adjusting the inspiratory flow rate and an expiratory back pressure valve (similar to applied PEEP) [3]. Some pumps allow the mean airway pressure to be set directly. The constant mean airway pressure maintains alveolar recruitment, avoids low endexpiratory pressures, and avoids high peak airway pressures. It also impacts oxygenation. Specifically, a higher mean airway pressure is associated with better oxygenation. HFOV induces a higher mean airway pressure than most modes of mechanical ventilation. The tidal volume (also called amplitude) is small during HFOV, usually less than or equal to the anatomic dead space. The amplitude depends on the endotracheal tube size and respiratory frequency: a smaller amplitude results when the endotracheal tube is small or the respiratory frequency is high [4]. High frequency percussive ventilation — High frequency percussive ventilation (HFPV) combines HFV plus time cycled, pressure-limited controlled mechanical ventilation (ie, pressure control ventilation, PCV). It can be conceptualized as HFOV oscillating around two different pressure levels, the inspiratory and expiratory airway pressures [5]. HFPV improves oxygenation, improves ventilation, and lowers airway pressures (peak, mean, and end-expiratory), compared to other modes of mechanical ventilation. (See "Modes of mechanical ventilation", section on 'Pressure-limited ventilation'.) HFPV is possible because of a device called a phasitron. The phasitron is an inspiratory and expiratory valve located at the end of the endotracheal tube. High-pressure gas drives the phasitron to deliver small tidal volumes at a high frequency (200 to 900 beats per min), superimposed on the inspiratory and expiratory airway pressures of PCV. The PCV is typically delivered at a respiratory rate of 10 to 15 breaths per min. HFPV does not require pharmacologic paralysis. In addition, it clears secretions more effectively than other types of HFV [5]. High frequency positive pressure ventilation — High frequency positive pressure ventilation (HFPPV) is rarely used anymore, having been displaced by the types of HFV discussed above. HFPPV is delivered through the endotracheal tube using a conventional ventilator whose frequency is set near its upper limits (figure 1). PATIENT SELECTION — There are no universally accepted indications for HFV. Its use has been described in a variety of clinical situations, including ALI/ARDS, bronchopleural fistula, inhalational injury, blunt trauma induced ARDS, and head injuries complicated by high intracranial pressure [5-8]. • • ALI/ARDS — The theoretical benefit of using HFV in patients with ALI/ARDS relates to the small tidal volumes. A strategy of low tidal volume ventilation has been proven in randomized trials to improve mortality, possibly due to decreased alveolar distension and ventilatorassociated lung injury. Although the trials did not use HFV, many clinicians suspect that HFV confers a similar benefit. Until this is proven, HFV should not be considered routine care for patients with ALI/ARDS. HFV is used by some clinicians when there is persistent hypoxemia during the first three days of mechanical ventilation despite maximal conventional therapy, although the data to support this are limited [9,10]. (See "Mechanical ventilation in acute respiratory distress syndrome", section on 'Low tidal volume ventilation' and "Ventilator-associated lung injury".) Bronchopleural fistula — HFJV is approved by the United States Food and Drug Administration for ventilating patients in whom a large and persistent bronchopleural fistula exists. However, the likelihood that HFJV will allow the bronchopleural fistula to close is unpredictable [6,7]. While HFJV may promote fistula closure by limiting alveolar distension, this may be outweighed in some patients by increased plateau airway pressure (alveolar pressure), decreased oxygenation, or worse hypercapnia [7]. (See "Management of bronchopleural fistula in patients on mechanical ventilation".) HFV should be avoided in patients with obstructive lung disease. The high respiratory rate used for HFV shortens the expiratory time, which can cause auto-PEEP and related sequelae. (See "Positive end-expiratory pressure (PEEP)", section on 'Auto (intrinsic) PEEP'.) EFFICACY — This section describes the clinical evidence related to the different types of HFV. Generally speaking, there is evidence that HFOV and HFPV improve oxygenation, although neither has been shown to improve clinical outcomes (eg, mortality, duration of mechanical ventilation, or length of ICU stay). HF jet ventilation — There is little moderate or high quality data evaluating the efficacy of HFJV in adults. One trial randomly assigned a heterogeneous group of 309 patients with acute respiratory failure to receive HFJV or volume-limited mechanical ventilation [11]. There was no significant difference in mortality or the duration of ICU stay. (See "Modes of mechanical ventilation", section on 'Volume-limited ventilation'.) In another trial, seven patients who were already receiving a traditional mode of mechanical ventilation for respiratory failure complicated by a bronchopleural fistula were randomly assigned to either switch to HFJV or continue their mode of ventilation [7]. There was no significant difference in the size of the chest tube leak (a measure of the bronchopleural fistula), but the HFJV group developed worse oxygenation and hypercapnia after switching to the HFJV. HF oscillatory ventilation — Most studies of HFOV have been performed in adults with ALI/ARDS [12-15]. In the largest multicenter trial, 148 patients with ALI/ARDS were randomly assigned to undergo mechanical ventilation using HFOV or pressure control ventilation (PCV) [12]. The PCV settings targeted a tidal volume of 6 to 10 ml/kg of actual body weight. The HFOV group had a lower mortality rate that was not statistically significant (37 versus 52 percent). The same group also had a significantly higher mean airway pressure and PaO2/FiO2 ratio, although these differences did not persist beyond 24 hours. A meta-analysis of six randomized trials (365 patients), including the trial just described, found that adults with ALI/ARDS who received HFOV had significantly lower hospital mortality or 30-day mortality than those who received conventional mechanical ventilation alone (39 versus 49 percent, RR 0.77, 95% CI 0.61-0.98) [15]. A limitation of this meta-analysis was that some of the trials that were included did not use low tidal volume ventilation in their control groups, which could bias the results in favor of HFOV. When trials that allowed tidal volumes ≥8 mL/kg were excluded and the meta-analysis repeated, there was a trend toward lower mortality among patients who received HFOV (RR 0.67, 95% CI 0.44-1.03). These results indicate that the repeat metaanalysis was too small to exclude or confirm a clinically important effect and additional trials are necessary to compare the effects of HFOV and low tidal volume ventilation on mortality. (See "Mechanical ventilation in acute respiratory distress syndrome", section on 'Low tidal volume ventilation'.) The effects of HFOV combined with another intervention have also been evaluated. Generally speaking, HFOV improves oxygenation when combined with inhaled nitric oxide or recruitment maneuvers, [16,17]. It may also prevent worsening of hypoxemia when a patient returns to the supine position following prone ventilation [18]. None of these combinations have been shown to improve important clinical outcomes and all of the studies had significant methodologic limitations. It was hypothesized that the duration of mechanical ventilation using a conventional mode prior to HFOV correlates with mortality [16]. However, a meta-analysis of nine studies (two randomized trials and seven observational studies) found no such relationship, even after confounding variables were considered [19]. Only the oxygenation index was independently associated with mortality. (See "Oxygenation and mechanisms of hypoxemia", section on 'Oxygenation index'.) HFOV at a respiratory rate greater than 6 Hz may be required because the usual respiratory rate of 3 to 6 Hz results in airway pressures that are potentially not lung protective. The feasibility of this approach was demonstrated by a single center, prospective cohort study of 30 patients with ARDS who were receiving HFOV after failing conventional lung protective ventilation [20]. Among the patients whose respiratory rates exceeded 6 Hz (range 6 to 15 Hz), most were able to meet their oxygenation (PaO2 55 to 80 mmHg) and ventilatory goals (pH 7.25 to 7.35). The effect of this strategy on mortality is not known; as a result, we cannot recommend this approach as a rescue modality. (See "Mechanical ventilation in acute respiratory distress syndrome", section on 'Low tidal volume ventilation'.) HF percussive ventilation — HFPV improves both oxygenation and ventilation without hemodynamic instability or clinically evident pulmonary barotrauma [5,21,22]. It may also decrease intracranial pressure in patients with head injuries [5]. • • A trial randomly assigned 35 patients with inhalational injury to undergo HFPV or volume-limited mechanical ventilation [21]. The HFPV group had significant improvement in the PaO2/FiO2 ratio of the initial 72 hours, compared to the volumelimited ventilation group. An uncontrolled trial of 54 patients with ALI/ARDS demonstrated improved oxygenation, decreased physiologic shunting, and decreased peak airway pressures after changing the mode of ventilation to HFPV [22]. HARMS — HFV is not risk free. The high respiratory rate shortens the expiratory time, potentially causing auto-PEEP and dynamic hyperinflation. The plateau airway pressure (alveolar pressure) and mean airway pressure are likely to increase if auto-PEEP and dynamic hyperinflation develop, elevating the risk of pulmonary barotrauma and hemodynamic instability. This occurs despite a lower peak airway pressure conferred by the smaller tidal volumes. In one trial, the risk of pulmonary barotrauma or hemodynamic instability was the same for patients receiving HFV compared to those receiving an alternative mode of mechanical ventilation [2]. (See "Pulmonary barotrauma during mechanical ventilation" and "Physiologic and pathophysiologic consequences of mechanical ventilation", section on 'Hemodynamics'.) There are also complications unique to type of HFV. As an example, HFJV is associated with necrotizing tracheobronchitis, endotracheal tube mucus inspissation, and variability of cardiac output [23]. Proper gas humidification reduces the likelihood of necrotizing tracheobronchitis or endotracheal tube mucus inspissation. SUMMARY AND RECOMMENDATIONS • • • • • High-frequency ventilation (HFV) combines a very high respiratory rate with tidal volumes that are smaller than the volume of anatomic dead space. (See 'Introduction' above.) There are four types of HFV: highfrequency jet ventilation (HFJV), high-frequency oscillatory ventilation (HFOV), high-frequency percussive ventilation (HFPV), and highfrequency positive pressure ventilation (HFPPV). (See 'Types of HFV' above.) There are no universally accepted indications for HFV. Its use has also been described in a variety of clinical situations. HFV should be avoided in patients with obstructive lung disease. (See 'Patient selection' above.) There is evidence that HFOV and HFPV improve oxygenation, although neither has been conclusively shown to improve clinical outcomes (eg, mortality, duration of mechanical ventilation, or length of ICU stay). (See 'Efficacy' above.) HFV is not risk free. Potential harms include intrinsic positive endexpiratory pressure (auto-PEEP), dynamic hyperinflation, and related sequelae (eg, pulmonary barotrauma, hemodynamic instability). In addition, there are specific risks associated with each type of HFV. (See 'Harms' above.) Use of UpToDate is subject to the Subscription and License Agreement. REFERENCES 1. Standiford, TJ, Morganroth, ML. High-frequency ventilation. Chest 1989; 96:1380. 2. Gluck, E, Heard, S, Patel, C, et al. Use of ultrahigh frequency ventilation in patients with ARDS. A preliminary report. Chest 1993; 103:1413. 3. Fessler, HE, Derdak, S, Ferguson, ND, et al. A protocol for highfrequency oscillatory ventilation in adults: results from a roundtable discussion. Crit Care Med 2007; 35:1649. 4. Hager, DN, Fessler, HE, Kaczka, DW, et al. Tidal volume delivery during high-frequency oscillatory ventilation in adults with acute respiratory distress syndrome. Crit Care Med 2007; 35:1522. 5. Salim, A, Martin, M. High-frequency percussive ventilation. Crit Care Med 2005; 33:S241. 6. Carlon, GC, Ray C, Jr, Klain, M, McCormack, PM. High-frequency positive-pressure ventilation in management of a patient with bronchopleural fistula. Anesthesiology 1980; 52:160. 7. Bishop, MJ, Benson, MS, Sato, P, Pierson, DJ. Comparison of highfrequency jet ventilation with conventional mechanical ventilation for bronchopleural fistula. Anesth Analg 1987; 66:833. 8. Eastman, A, Holland, D, Higgins, J, et al. High-frequency percussive ventilation improves oxygenation in trauma patients with acute respiratory distress syndrome: a retrospective review. Am J Surg 2006; 192:191. 9. Mehta, S, Granton, J, MacDonald, RJ, et al. High-frequency oscillatory ventilation in adults: the Toronto experience. Chest 2004; 126:518. 10. David, M, Weiler, N, Heinrichs, W, et al. High-frequency oscillatory ventilation in adult acute respiratory distress syndrome. Intensive Care Med 2003; 29:1656. 11. Carlon, GC, Howland, WS, Ray, C, et al. High-frequency jet ventilation. A 12. 13. 14. 15. 16. 17. 18. prospective randomized evaluation. Chest 1983; 84:551. Derdak, S, Mehta, S, Stewart, TE, et al. High-frequency oscillatory ventilation for acute respiratory distress syndrome in adults: a randomized, controlled trial. Am J Respir Crit Care Med 2002; 166:801. Bollen, CW, van Well, GT, Sherry, T, et al. High frequency oscillatory ventilation compared with conventional mechanical ventilation in adult respiratory distress syndrome: a randomized controlled trial [ISRCTN24242669]. Crit Care 2005; 9:R430. Mentzelopoulos, SD, Roussos, C, Koutsoukou, A, et al. Acute effects of combined high-frequency oscillation and tracheal gas insufflation in severe acute respiratory distress syndrome. Crit Care Med 2007; 35:1500. Sud, S, Sud, M, Friedrich, JO, et al. High frequency oscillation in patients with acute lung injury and acute respiratory distress syndrome (ARDS): systematic review and metaanalysis. BMJ 2010; 340:c2327. Mehta, S, MacDonald, R, Hallett, DC, et al. Acute oxygenation response to inhaled nitric oxide when combined with high-frequency oscillatory ventilation in adults with acute respiratory distress syndrome. Crit Care Med 2003; 31:383. Ferguson, ND, Chiche, JD, Kacmarek, RM, et al. Combining high-frequency oscillatory ventilation and recruitment maneuvers in adults with early acute respiratory distress syndrome: the Treatment with Oscillation and an Open Lung Strategy (TOOLS) Trial pilot study. Crit Care Med 2005; 33:479. Demory, D, Michelet, P, Arnal, JM, et al. High-frequency oscillatory 19. 20. 21. 22. 23. ventilation following prone positioning prevents a further impairment in oxygenation. Crit Care Med 2007; 35:106. Bollen, CW, Uiterwaal, CS, van Vught, AJ. Systematic review of determinants of mortality in high frequency oscillatory ventilation in acute respiratory distress syndrome. Crit Care 2006; 10:R34. Fessler, HE, Hager, DN, Brower, RG. Feasibility of very high-frequency ventilation in adults with acute respiratory distress syndrome. Crit Care Med 2008; 36:1043. Reper, P, Wibaux, O, Van Laeke, P, et al. High frequency percussive ventilation and conventional ventilation after smoke inhalation: a randomised study. Burns 2002; 28:503. Hurst, JM, Branson, RD, DeHaven, CB. The role of high-frequency ventilation in post-traumatic respiratory insufficiency. J Trauma 1987; 27:236. Angus, DC, Lidsky, NM, Dotterweich, LM, Pinsky, MR. The influence of high-frequency jet ventilation with varying cardiac-cycle specific synchronization on cardiac output in ARDS. Chest 1997; 112:1600. GRAPHICS Modes of high frequency ventilation Modes of high frequency ventilation. With high frequency positive pressure ventilation (top), high-pressure conditioned gas (a) is delivered during inhalation and flows predominantly through an endotracheal tube (b) to the patient with partial escape to the atmosphere. During exhalation, the gas exits through an optional one-way valve (c). With high frequency jet ventilation (middle), conditioned high-pressure gas enters from a cannula (a) at a selected level along the endotracheal tube or trachea (c). This gas entrains additional conditioned gas (b) by the Venturi effect. During exhalation the gas exits passively through an optional one-way valve (d). With high frequency oscillation (bottom), the piston or diaphragm (a) oscillates while fresh conditioned gas (bias flow) enters (b) and exhaust gas exits (c) at a balanced constant rate. The bias flow ports can be positioned anywhere along the path from the external tip of the endotracheal tube to within the trachea itself. Initial jet ventilator settings * If the patient was hypoxic with controlled mechanical ventilation, driving pressure should be adjusted as necessary to achieve a mean airway pressure equal to that present during controlled mechanical ventilation. If minute ventilation under those conditions is not at least equal to that present during controlled mechanical ventilation, driving pressure should be further increased until this condition is met before obtaining arterial blood gas levels. If the patient was hypercapnic with controlled mechanical ventilation, driving pressure should be adjusted as necessary to achieve a minute ventilation two times that present during controlled mechanical ventilation. If mean airway pressure exceeds 110 percent of that present during controlled mechanical ventilation before obtaining that minute ventilation, arterial blood gas levels should be obtained at that point before further increasing driving pressure. In individual patients, changes in rate may have opposite effects on PaCO2. • Careful monitoring of peak, mean, and end-expiratory airway pressure as well as minute ventilation and oximetry should be performed when any change is made in ventilatory parameters. Only experienced personnel should use high-frequency jet ventilation. Redrawn from: Standiford, TJ, Morganroth, ML, Chest 1989; 96:1383.