Finger Detector test for high intensity 1 0 3 2 5 4 7 6 9 8 11 10 13 12 15 14 Finger detector: 15 plactic strips of ~14 mm width readout with 16 PMT's Examples of rates supported by the finger detector, measurements made during the 104Sn Coulex experiment with 124Xe beam at 700 MeV/u: Example 1: Setting for 104Sn. On this run slits at S1 were closed at ±10 mm. The average rate on 1 strip was at most ~580 kHz, it was derived from the average number of hits on 1 channel on the multihit TDC (time window of 5µs). The instant rate on 1 strip, derived from the mean value of the time difference between 2 hits in 1 channel (see fig 1), was at most ~890 kHz Fig. 1: Time difference between 2 hits if the multihit TDC for 1 channel in ns. The mean value give the mean instant rate on the strip Fig. 2: ToF measured with scintillators SC21 and SC41 (left) compared with ToF measured with Finger and SC41 (right). It looks like if the SC21 was damaged. Fig 3: Identification plot A/Q vs Z calculated with the finger detector (right). (103000 events 90600 in the plot) compared with the identification plot made with SC21 (left) Example 2: Setting for 104Sn. On this run slits at S1 were opened at ±30 mm. The average rate on 1 strip was at most ~850 kHz The instant rate on 1 strip was at most ~1350 kHz HV on SC21 scintillator was off. Again the same pictures as example 1: Fig. 4: Time difference between 2 hits if the multihit TDC for 1 channel in ns. The mean value give the mean instant rate on the strip Fig. 5: ToF measured with Finger and SC41 (HV on SC21 was off) Fig 3: Identification plot A/Q vs Z calculated with the finger detector. (on this run 2 PMT's so 3 strips were not working properly so the countrate is not correct)