DNA Quantitation System:

advertisement
6112/‫ل‬/1 ‫عطاء رقم‬
‫ كلية العلوم‬/‫توريد وتركيب وتشغيل أجهزة خمربية‬
:‫املواصفات اخلاصة‬
Item
1
quantity
Description
Real Time PCR System with Notebook, Open System
Features:
5 channels multiplexing for discrimination of up to five targets in a single reaction well*
Strong flexibility thanks to the 96-well format suited for standard PCR microplates, test tubes
and strips, Optimal signal/noise ratio, Light emitting diodes (LED)
detection using multiple exposure method
Technical Characteristics :
Thermal block format
96 test tubes of 0.2ml (12x8)
Range of thermal block temperature control
0C…100C
Resolution of temperature setting
0.1C
Absolute accuracy of temperature maintenance ±0.2C
Uniformity of thermal block temperature
±0.15C
Hot cover Temperature
105C ±1C
Excitation source
Light –emitting diode (LED)
Detector
CCD ( Charge coupled device )
Number of the fluorescence measurement
5
Excitation/detection wave length
470/252, 532/570, 585/633, 633/670, 690/750
Threshold sensitivity of each of the channels for solution of standard fluorophores
1
2
1
Analytical Balances
Capacity g : 210
Readability mg : 0,1 ( 0.0001g)
Linearity mg : ± 0,2
Repeatability mg : ± 0,05
functions and characteristics








3
LCD display with small decimal digits
Membrane keyboard, water proof and solvent resistant, easy to use with TARE,
ON/OFF, PRINT and MODE keys
Indication of the reached stable weight
Bar-graph indicator of dosage and remaining capacity of the balance
Parameters configurable by menu: reading in g (grams), lb (pound), oz (ounce), ct
(carats), pcs (pieces), % (percentage)
Full scale automatic calibration with external mass
Weighing underneath the balance
Selectable response time “fast/slow” (see table)
1
Autoclaves
Specification :
Control
Chamber Volume (L)
Sterilization Temp.
(℃)
Temperature 1)
Melting Temp. (℃/℉)
Safety
Timer
Microprocessor PID Control
50 / 1.8
110 to 123
60 to 100
Over Temperature Protector / Pressure Safety Valve
/ Over Current Pr tector
Warning Alert (Over Temp. / Low Temp. / Low
Heat / Sterilization Fail / Low Water Level)
0 to 999 min
Standard modes (Decontamination, Solids
Sterilization, Liquids Sterilization, Melting)
Program modes (Sterilization 1, Sterilization 2,
Liquid Sterilization, Melting)
Programs
4
pH/Temperature Bench Meter
This pH/temperature bench meter provides intuitive software and a well-organized LCD
display, making operation quick and simple. It also offers a functional design and reliable
calibration results, which are tracked on the display, ensuring precise measurements time
after time. This meter measures the pH and the temperature of a sample and includes a builtin 2-in-1 plastic electrode, temperature sensor, and a pH buffer
Specification
Calibration
2 Points at least
1 Predefined Buffer Group
(4.01, 7.00, 10.01)
Error Limits
±0.01 pH
±0.5° C
Power Supply
Range
110 to 240 V/50 Hz, DC 12 V
0.00 to 14.00 pH
1
Resolution
5
0 to 100° C
0.01 pH
0.1° C
REFRIGERATED BENCH TOP CENTRIFUGE
Maximum speed : 4.100 rpm Swing-out / 14.000 rpm Angle
Maximum RCF : 3.045 xg Swing-out / 18.405 xg Angle
Maximum capacity : 4 x 200 ml Swing-out / 24 x 1,5/2 ml Angle
Temperature range : -9°C / +40°C (+4°C at max. speed )
Programmable microprocessor control system
Digital displays for prog. no, speed, rcf, time, temperature, acceleration and braking rates
10 program memories
Timer : 1 - 99 minutes and hold position
10 acceleration / 10 braking rates
Safety features : Imbalance detection, lid interlock, chamber guard ring, motor
over temperature , rotor over speed, diagnostic system
1
ACCESSORIES
Puller for rotor removal
Swing-out rotor 4 x 200 ml, 4.100 rpm, 3.045 xg
Set of 4 buckets
Set of 4 buckets with sealing cap
Set of 4 200 ml PP bottles with cap
Set of 4 inserts 1 x 100 ml
Set of 4 inserts 1 x 50 ml conical
Set of 4 inserts 7 x 15 ml
Set of 4 inserts 3 x 15 ml conical
Set of 4 inserts 7 x 5/7 ml
1
1
1
1
1
1
Angle Rotor 30 x 15 ml 4.100 rpm 3.007 xg
Adaptor for 1,5/2 ml (Ø11mm) tubes , (Should be ordered according to the capacity of the
rotor)
Adaptor for 13 x 100 mm vacuumed / non-vacuumed tubes (Should be ordered according
to the capacity of the rotor)
Adaptor for 13 x 75 mm vacuumed / non-vacuumed tubes (Should be ordered according
to the capacity of the rotor)
6
7
1
1
1
1
1
1
1
Angle rotor 24 x 1,5/2 ml 14.000 rpm, 18.407 x g
Adaptor 500 / 800 μl
Adaptor 200 μl PCR
Adaptor 250 / 400 / 700 μl
1
1
1
1
Angle rotor 6x50 ml 9.000 rpm, 9.418 xg
Set of 6 adaptors 1x15 ml conical
Set of 6 rubber base adaptors for 50 ml conical tubes
1
Microtitre rotor 4.100 rpm, 2.349 xg
1
Fluorometer Dual-channel fluorometer
1
Equipped with two fluorescence channels for nucleic acid and protein quantitation:
Blue fluorescence channel: Excitation 495nm shortpass (wavelengths up to 495nm), emission
510–580nm
Red fluorescence channel: Excitation 640nm shortpass (wavelengths up to 640nm), emission
660–720nm
Imager System
Imager software with USB connection to PC
Built-in, 21 x 26cm pullout UV-transilluminator
With Imager Analysis software
Viewing window for gel observation with standard 560 nm amber filter.
1
Specification :
UV light
312(302) nm
Camera
Minimum 5.0 MP
Image Storage
PC
Connection
USB to PC
Display
LED
Viewing
Window
For gel observation
Light
Optional
Accessory
White light (standard)
White light table (power built-in)
470 nm Blue light (power built-in)
Safety
USB port
CE, FCC
for PC connection
Accessories
Optical EtBr filter (620 nm) for SmartView Pro Imager
Optical SYBR Green filter (520 nm) for SmartView Pro
8
Hotplate Mignatic Stirrers
Stirrers Bar set
1
1
9
Incubated Shakers
Digital incubating benchtop shakers
Operating Features
General Control
touch screen LCD controller with easy icons and logical menus.
-volatile set point memory after power interruption.
-in RS-232 port and USB port for external control and data collection.
Shaking Control
Speed range: 10 to 500 rpm.
Timer range: 1 min. to 999 min.
-free operation.
- Automatic shaking speed adjustment in case of excessive workload such as unbalanced load
placement, unusual vibrations caused from unstable floor or external shock.
-current, stalled platform check
Specification:
Range (℃ )
Temperature
Fluctuation 1)
at 37℃
in flask (±℃)
Variation 1)
at 37℃
in flask (±℃)
Amb. +5 to 80
0.1 / 0.18
0.5 / 0.90
1
Shaking System
Motion
Type
Amplitude
Size
(mm / inch,
dia.)
Speed
Range
(RPM)
Accuracy
Timer
Max. Load
(kg )
11
TOPLOADING BALANCES
Capacity g : 210
Readability g : 0,001
Orbital
19.1 / 0.75
10 to 500
±1% of set speed (>100rpm) / ±1
(<100rpm)
1 min. to 999 min.
10 / 22.0 at 500 rpm
18 / 39.7 at 400 rpm
1
Linearity g : ± 0.002
Repeatability mg : ± 0.001
Response time : ~ 3 sec.
Pan mm: Ø 110
Calibration mass : External 1
11
WATER DISTILLER
- Capacity: 3.5 lt/hr. at least.
- Stainless steel boiler and condenser
- Stainless steel heater
- Distillation led
- Manometer for water pressure
- Warning led for shortage of water or overheating
1
Download