Supplementary material Sensitivity of ODR to change in saturation The methods section explains that for every unit change in optical density ratio (ODR), the calculated saturation decreases by 2.47 percentage points. This is about 1.93 times greater rate of change than for the fundus camera based Oxymap T1 oximeter, according to default manufacturer calibration. A smaller change in ODR for each unit change in saturation suggests that the SLO oximeter is less sensitive to change in saturation. The difference in sensitivity may be due to the different wavelengths used in the two systems. The fundus camera oximeter uses 570nm (isosbestic) and 600nm (sensitive to oxygen saturation) while the SLO uses 532nm (close to isosbestic) and 633nm (sensitive). The light extinction coefficient for a solution of a mixture of Hb and HbO2 can be written as follows: εsolution = εHbO2*SatO2+ εHb(1-SatO2)1 The optical density of a solution can be written as OD = εsolution*C*d, where C is the total concentration of hemoglobin and d is the distance, which the light has to travel through the solution. Now let us assume that we have two solutions of hemoglobin, with saturations SatO2-1 and SatO2-2. The difference in the OD at the sensitive wavelength is OD = ε*C*d (since C and d do not change in this example). Using the equation for εsolution above, we find (after some algebra) that ε= εsolution-1 - εsolution-2 = (εHbO2-εHb)*(SatO2-1 – SatO2-2) This means that for one percentage point change in saturation OD1% = (εHbO2-εHb)*C*d Now let us calculate the change in ODR, with one percentage point change in saturation. ODR1% = OD1% / ODisosbestic (because ODisosbestic does not change with saturation). This means that ODR1% = (εHbO2-εHb)*C*d / εiososbestic C*d = (εHbO2-εHb) / εiososbestic In short, the ODR change with 1 percentage point change in saturation is (εHbO2-εHb) / εiososbestic The following table compares the SLO wavelengths and the fundus camera wavelengths with data from Zijlstra et al.2 The table uses data for 632nm instead of 633nm and an average of extinction coefficients for HbO2 and Hb at (near-)isosbestic wavelengths. Table S1. Extinction coefficients for oxy- and deoxyhemoglobin (HbO2 and Hb) at wavelengths used for SLO oximetry and fundus camera oximetry.* The table also shows the theoretical change in ODR with 1 percentage point change in oxygen saturation. SLO wavelengths Fundus camera wavelengths εHbO2 0.129 0.961 εHb 1.022 3.759 εIsosbestic 9.92 11.6 -0.090 -0.241 ODR1% =(εHbO2-εHb) / εiososbestic *The Oxymap T1 fundus camera oximeter uses 570nm (isosbestic) and 600nm (sensitive) light. The SLO oximeter tested here uses 532nm (close to isosbestic) and 633nm (sensitive). The table uses data for 632nm instead of 633nm, which was unavailable. Table S1 shows that ODR should change about 2.7 times faster in the case of fundus camera with the same change in saturation. This is approximately in agreement with the ratio of the calibration coefficients (1.93). The above calculations assume that one wavelength in each system (532nm in the SLO and 570nm in the fundus camera) are completely insensitive to changes in saturation. In fact, this is close to being accurate for 570nm but there is considerable difference in εHbO2 and εHb at 532nm (an average is taken to calculate εIsosbestic in the table above). What difference does this make for oxygen sensitivity of the SLO oximeter? According to Ziljstra et al.2, the extinction coefficients at 532nm are 11.42L*mmol-1*cm-1 for HbO2 and 8.411L*mmol-1*cm-1 for Hb. Therefore, with increasing saturation, the OD at 532nm will increase slightly while the OD at 633nm decreases. This will increase the rate of change in ODR for the SLO with change in saturation and counteract somewhat the small sensitivity. It is not ideal to have only an approximately isosbestic wavelength. However, this fact may help to explain why the ratio of the calibration coefficients is 1.93 while the theoretical ratio is slightly higher when isosbesticity is assumed at 532nm. Basic data for retinal vessel oxygen saturation calculation for the 11 healthy subjects of the study and variability in ODs along the measured vessel segments Tables S2 and S3 below show the means of each variable along the measured vessel segment. An OD value is for example a mean of OD values many cross sections on each segment (38-162 cross sections, usually above 100). Comparison is made between the SLO oximeter and the Oxymap T1 fundus camera based oximeter. Table S2. Arterioles. Table shows means of each variable, for arterioles, along a measured vessel segment. The upper part shows mean values from the SLO for each individual (n=11). The lower part shows, for comparison, an average of corresponding values from 10 (different) healthy individuals, measured with a fundus camera based oximeter. Subject I0-532 I532 I0-633 I633 OD532 OD633 ODR SatO2 1 2 3 4 5 6 7 8 9 10 11 SLO Average 1034 1051 840 636 876 1226 956 1021 1543 1345 1295 1075 477 555 460 415 515 647 579 622 772 689 731 587 1120 1311 880 631 1356 1065 1378 935 1311 1306 1527 1165 935 1141 791 598 1234 939 1210 861 1099 1086 1324 1020 0.332 0.274 0.256 0.184 0.228 0.273 0.215 0.206 0.298 0.288 0.248 0.255 0.078 0.060 0.046 0.024 0.041 0.055 0.057 0.032 0.076 0.080 0.062 0.055 0.236 0.219 0.173 0.127 0.179 0.200 0.264 0.134 0.255 0.277 0.249 0.210 85.5 89.8 101.2 112.5 99.5 94.5 78.6 110.8 80.7 75.5 82.2 91.9 Fundus camera Fundus camera average I0-570 I570 I0-600 I600 OD570 OD600 ODR SatO2 932 458 2112 1759 0.315 0.083 0.263 90.3 *Note: The pixels from the oximeter do not correspond to the same retinal dimensions. Width (pixels) 11.4 9.4 10.2 11.4 7.7 12.0 12.9 12.0 11.9 11.3 11.3 11.0 Width (pixels) 15.4 Table S3. Venules. Table shows means of each variable, for venules, along a measured vessel segment. The upper part shows mean values from the SLO for each individual (n=11). The lower part shows, for comparison, an average of corresponding values from 10 (different) healthy individuals, measured with a fundus camera based oximeter. Subject I0-532 I532 I0-633 I633 OD532 OD633 ODR SatO2 1 2 3 4 5 6 7 8 9 10 11 Average 639 999 897 591 903 1228 1070 1104 1538 1322 1334 1057 368 475 478 340 483 543 554 471 671 536 635 505 1224 1314 898 609 1410 1013 1400 990 1321 1386 1582 1195 1014 1009 748 514 1098 785 1050 712 977 1025 1276 928 0.227 0.319 0.272 0.237 0.271 0.350 0.281 0.359 0.358 0.390 0.321 0.308 0.081 0.114 0.079 0.073 0.108 0.110 0.124 0.143 0.130 0.131 0.093 0.108 0.370 0.359 0.288 0.309 0.399 0.315 0.422 0.398 0.363 0.336 0.292 0.352 52.3 55.0 72.7 67.5 45.2 65.9 34.6 45.4 54.1 60.8 71.8 56.8 Fundus camera Fundus camera average I0-570 I570 I0-600 I600 OD570 OD600 ODR SatO2 923 330 2165 1243 0.448 0.241 0.538 55.2 *Note: The pixels from the oximeter do not correspond to the same retinal dimensions. Width (pixels) 11.0 11.3 11.3 19.7 12.0 11.8 16.5 19.6 13.7 14.8 11.5 13.9 Width (pixels) 17.8 Tables S4 and S5 show the variability in ODs along the measured vessel segments (coefficient of variation = (SD/ mean)*100%) for arterioles and venules seperately. Table S4. Arterioles. Table shows the variability in ODs for arterioles along measured vessel segments (coefficient of variation = (SD/ mean)*100%). Upper part of the table shows values for the 11 healthy individuals, measurements with the SLO. The lower part shows an average and median from 10 (different) healthy individuals, measured with the fundus camera based oximeter. Subject 1 2 3 4 5 6 7 8 9 10 11 Mean Median Fundus camera Mean Median OD532 Mean OD532 SD 0.332 0.274 0.256 0.184 0.228 0.273 0.215 0.206 0.298 0.288 0.248 0.046 0.054 0.090 0.032 0.050 0.070 0.066 0.076 0.068 0.061 0.037 0.059 0.061 OD570 SD OD532 OD633 OD633 Coeff. var. Mean SD (%) 13.8 0.078 19.8 0.060 35.3 0.046 17.4 0.024 22.1 0.041 25.7 0.055 30.6 0.057 37.0 0.032 22.9 0.076 21.3 0.080 14.9 0.062 23.7 22.1 OD570 Coeff. var. (%) 0.052 17.0 0.050 16.5 OD600 SD 0.011 0.014 0.022 0.005 0.009 0.014 0.017 0.027 0.017 0.020 0.009 0.015 0.014 OD633 Coeff. var. (%) 13.7 24.0 46.9 21.7 22.2 26.0 30.3 84.4 22.8 24.6 14.8 30.1 24.0 OD600 Coeff. var. (%) 0.018 22.5 0.017 23.1 Table S5. Venules. Table shows the variability in ODs for venules along measured vessel segments (coefficient of variation = (SD/ mean)*100%). Upper part of the table shows values for the 11 healthy individuals, measurements with the SLO. The lower part shows an average and median from 10 (different) healthy individuals, measured with the fundus camera based oximeter. Subject OD532 Mean 1 2 3 4 5 6 7 8 9 10 11 Mean Median Fundus camera Mean Median OD532 SD 0.227 0.319 0.272 0.237 0.271 0.350 0.281 0.359 0.358 0.390 0.321 0.060 0.048 0.072 0.056 0.035 0.080 0.069 0.094 0.047 0.048 0.049 0.060 0.056 OD570 SD OD532 OD633 OD633 Coeff. var. Mean SD (%) 26.3 0.081 14.9 0.114 26.4 0.079 23.5 0.073 12.9 0.108 22.8 0.110 24.5 0.124 26.2 0.143 13.1 0.130 12.3 0.131 15.2 0.093 19.8 22.8 OD570 Coeff. var. (%) 0.044 10.0 0.041 8.4 OD600 SD OD633 Coeff. var. 0.014 0.016 0.025 0.016 0.014 0.025 0.030 0.037 0.019 0.017 0.014 0.021 0.017 16.6 14.1 31.6 21.6 12.7 23.1 24.0 26.1 14.9 13.2 14.6 19.3 16.6 OD600 Coeff. var. (%) 0.023 9.8 0.021 8.7 As can be seen from the tables, both the isosbestic and the sensitive wavelength yield more variable data for the SLO, compared to the fundus camera (variability along the vessels). However, the 633nm does not yield clearly more variable data then the 532nm. References 1. Harris A, Dinn RB, Kagemann L, Rechtman E. A review of methods for human retinal oximetry. Ophthalmic Surg Lasers Imaging 2003;34:152-164. 2. Zijlstra WG, Buursma A, Van Assendelft OW. Visible and near infrared absorption spectra of human and animal haemoglobin - Determination and application. Utrecht: VSP; 2000:368.