Item Name : 300MA X-ray Machines with Computed Radiography

advertisement
Item Name : 300MA X-ray Machines with Computed Radiography system .
Vide T. No.4.19A /APMSIDC/2014-15, Dated: 03-12-2014.
Sl.No S. No. as per Tender
specs
Amendment requested
By the firm M/s. Siemens
Required Amendment
Any X-ray equipment with
32KW will be delivering
200mAS. It directly relates to the
time factor.
The mAS range clinically
suggested from 125mA onwards
for achieving good image
quality.
The time range used dueing the
exposure is only from 5mS to
0.9S for obtaining good quality
X-Rays.
mAS range 1.25200mAS in 23 steps.
1
Output should be 32 KW
or more offering 500 mAS
or more.
2
mA Range 40mA to
300mA or more
3
The time range should be
1 ms to 10 sec or more
4
It should have 5" or more
LCD display for mA, KV &
mAs.
In Analog X-Ray units, only
Electronic Control console is
only available with most of the
vendor.
5
It should have Anatomical
programming
radiography should be
possible with user
programmable up to 1000
or more settings.
The APR feature is mainly
available in the Digital
Radiography system. In analog
X-Ray system, this feature will
be not available with most of the
vendor.
Kindly remove the
point A-7
6
It should have one touch
Auto calibration mode for
mA & kv to perform
periodic calibration &
should also display
waveforms for mA & KV
with analysis.
The Auto Calibration mode
feature is mainly available in the
High end Digital Radiography
system. In analog X-ray system,
this feature will be not available
with most of the vendors.
Kindly remove the
point A-8
7
It should have provision
to connect solid state ion
chambers for Automatic
exposure control
The Automatic Exposure control
mode feature is mainly available
in the High end Digital
Radiography system. In analog
X-Ray system, this feature will
be not possible for integrating
with the existing X-Ray system.
8
It should have the
possibility to connect &
upgrade to Digital
radiography Solution in
future
We can upgrade any analog xray system with External
Detectors to Digital Radiography
system. However the cost
involved in the Hardware &
Software for the up gradation
should be borne by the
purchaser.
125mA to 300mA
The time range should
be 5mS-0.9S
Electronic Control
console
Kindly remove the
point A-9
Remarks
II. M/s. Agfa Healthcare India (P) Ltd
Request you to separate both X-ray Machine & Computed Radiography as non of the
Manufactures make both so that it will avoid taxation, additional expenses in transportation and
creates comfort in post installation service support co ordination.
Sl.No S. No. as per
Tender specs
Amendment requested
By the firm M/s. Agfa
Clinical Importance
Remarks
Healthcare India (P) Ltd
1
It should support
the following plate
sizes: (in inches)
14 x 17, 10 x 12, 8
x 10,15cm x 30
cm.
It should support the
following plate sizes: (in
inches) and the image plates
must be rigid type or image
plates mounted on a tray for
improved plate life. 14x17,
10x12, 8x10, 15cmx30cm.
2
Printer
should
have a resolution
of above 300 dpi
/ppi or more .and
should Print the
same resolution
for all Radiological
examinations
It Should have
single online tray
with obtain of
immediately
replacement
of
any film sizes.
Printer
should
have
a
resolution of above 500 dpi
/ppi and should Print the same
resolution for all Radiological
examinations
3
It should have two online
trays with capacity of
printing all the film sizes
available.
The most Common reason for
artifacts in image is Flexible
plates as dust accumulating on
rollers which creates marks on
image plate when the image
plate move on it. In some old
generation digitisers image
plates get bend inside the
digiser, in that case plate gets
worn out after a period. These
issues are taken care in Rigid
image plate Technology.
Printing resolution is very
important as all these machines
are for PHC’s/CHC’s For any
critical case study the patient can
be sent to the nearest district
hospital with the film for
evaluation.
With single tray, whenever the
film size is changed the
technician has to change the tray
for each size and needs to be
calibrated every time.
Item Name : Ultrasound Scanner,
T. No.4.19B /APMSIDC/2014-15, Dated: 03-12-2014.
S.No. As given in Tender
1
Request amendment by the firm
M/s. Trivitron Healthcare
System should be 12" or more System monitor size should be
LCD/TFT/LED monitor with 10” or above
tilting function
2
Real time scanner –Scan depth
Approximately 26 cm
Scan depth is up to 24cm
3
---2.0-6.0 MHz Convex Probe for
Convex probe frequency 2.5
MHz to 6.0MHz.
T.V Probe frequency 4.0 Mhz to
7.0Mhz
6 slices of TGC
Mass patient data storage facility
with at least 8G Bites(External
or Internal)
Battery facility is for 1 Hour
(Should be External or Internal)
4
---5.0-8.0MHz Transvaginal Probe
5
6
.8 slices of TGC
Mass patient data storage facility
with at least 8G bites
7
In-built battery with minimum 1
hour continuous scan facility
should be supplied.
Remarks
8
USB orts/VGA
output
output/Video Single USB port/VGA
output/Video output.
II) M/s. Esaote India (NS) Ltd.,
Sl.
No.
1.
Point
No. of
Specific
ations
2
As mentioned in
Tender Document
Specifications
2
5
3.
7
2.0-6.0 MHz Convex
Probe
4
8
Scanning modes B, M,
B/M, B/B
5
10
Minimum 800 frames
cine loop memory
6
17
Mass patient data
storage facility with at
least 8G bites
7
22
ISO Certification, CE
approval
System should be 12”
or more
LCD/TFT/LED
monitor with tilting
function
Real time scanner Scan
Depth approximately
26 cms
Modifications required as
by M/s. Esaote India (NS)
Ltd.,
Clinical
Importance
System should be 12” or
more LCD/TFT/LED
monitor.
As tilting of
monitor is
confined to one
particular
company.
Real time scanner – Scan
Scan Depth upto
Depth approximately 26
26 cms and +/-2
cms and +/-2 cms will be
cms will be
acceptable
acceptable.
2.0-6.0 MHz Convex Probe More flexibility
+/- 1 MHz will be
in choosing the
acceptable.
frequency.
Scanning modes B, M,
Preferable PW
B/M, B/B and PW mode
mode for
will be preferred
scanning of
OB/GYN patients
for heart beat.
Minimum 256 frames cine Minimum 256
loop memory
frames is ideal for
B/W systems.
Minimum patient data
Minimum storage
storage facility is 500MB
will be 500MB
and optional up to 500G
Bites HD.
ISO Certification, CE and
USFDA is very
USFDA approval
proven
certification in the
world.
III) BPL Medical Technologies Private Limited.
Sl.No
As given in Tender
Request amendment by the firm M/s.
BPL Medical Technologies Prt Ltd
1
1. Should be a portable
ultrasound system with
wide application for
Abdomen, GYN\OB,
Urology, Small Parts,
Orthopediac, Cardiac etc
2. System should be 12" or
more LCD/TFT/LED
monitor with tilting
function .
Minumum 800 frames cine
loop memory
As per your specification point 1 says
portable ultrasound system and as per
your point 2 it says monitor should be
12" or more. So we would like to bring
to your notice that bigger the monitor
size the portability will be an issue. As
per our understanding the portable
ultrasound machines are seen with
maximum 10" monitor (CRT/LCD).
2
its reads as 800 frames cineloop
memory In general the retrival of the
frames from the memory in portable b/w
ultrasound machine is about 80 -100
Remarks.
Remarks
frames/sec (max) to see the moment of
the organ. So 800 frames will be too
high to and some times it take more time
to retrive the data from memory. As per
our understanding 100 frames will be
more appropriote for clinical studies in a
faster way in portable b/w ultraasound
machine.
3
Image repetition rate not
less than 150 frames / sec
(higher preferred)
it reads as repetition rate not less than
150 frames /sec. We would like to draw
your attention that for performing
applications like Gynac, OBG, Small
parts in b/w ultrasound scanner 100
frames/sec would be ideal.
4
Multi calipers facility
with electronic
presentation for
distances, area,
circumference,
ellipse,hip angel,
OB/Gyn measurements,
Cardiac, Urology,
Orthopedic and Vascular
measurement, volume
and trackball
measurement facility
the machine should be useful for
cardiac, Urology and orthopedic
applications also. In this regard we
would like to know more about the
requirement of the software’s, hardware
and the probes.
5
In-built battery with
minimum 1 hour
continuous scan facility
should be supplied.
. in-built battery with mim 1 hour
backup is asked for.
But as per our understanding on usage
of portable b/w ultrasound in-built
battery will add weight
to the machine. Being asked for portable
machine kindly give an option for
external battery source
also.
Item Name : Semi Auto Analyzer,
Vide T. No.4.19D /APMSIDC/2014-15, Dated: 03-12-2014.
S.No. As given in Tender
Request amendment by the
firm M/s. IRIS Healthcare
Technologies Pvt,Ltd.
1
Hard Glass Pyrex Cuvvette
Fexible Cuvvette System:
Flow cuvvette
Remarks
11) M/s. Transasia Bio-Medicals Ltd.,
Sl.No As given in Tender
1
polygonal multi standard (calibration
curve)
2
No: of Deltas : from 2 to 20
3
CE certification
Request amendment by
the firm M/s. Transasia
Bio-Medicals Ltd.,
Clarification with an
example for a
hemistry/analyte/Parameter
Clarification with an
example for a chemistry
/Analyte/Parameter. Is the
delta for time or for
absorbance or for
temperature?
Since it comes under IVD
Guidelines Self Certification
is acceptable-Needs
confirmation.
Remarks.
Download