Item Name : 300MA X-ray Machines with Computed Radiography system . Vide T. No.4.19A /APMSIDC/2014-15, Dated: 03-12-2014. Sl.No S. No. as per Tender specs Amendment requested By the firm M/s. Siemens Required Amendment Any X-ray equipment with 32KW will be delivering 200mAS. It directly relates to the time factor. The mAS range clinically suggested from 125mA onwards for achieving good image quality. The time range used dueing the exposure is only from 5mS to 0.9S for obtaining good quality X-Rays. mAS range 1.25200mAS in 23 steps. 1 Output should be 32 KW or more offering 500 mAS or more. 2 mA Range 40mA to 300mA or more 3 The time range should be 1 ms to 10 sec or more 4 It should have 5" or more LCD display for mA, KV & mAs. In Analog X-Ray units, only Electronic Control console is only available with most of the vendor. 5 It should have Anatomical programming radiography should be possible with user programmable up to 1000 or more settings. The APR feature is mainly available in the Digital Radiography system. In analog X-Ray system, this feature will be not available with most of the vendor. Kindly remove the point A-7 6 It should have one touch Auto calibration mode for mA & kv to perform periodic calibration & should also display waveforms for mA & KV with analysis. The Auto Calibration mode feature is mainly available in the High end Digital Radiography system. In analog X-ray system, this feature will be not available with most of the vendors. Kindly remove the point A-8 7 It should have provision to connect solid state ion chambers for Automatic exposure control The Automatic Exposure control mode feature is mainly available in the High end Digital Radiography system. In analog X-Ray system, this feature will be not possible for integrating with the existing X-Ray system. 8 It should have the possibility to connect & upgrade to Digital radiography Solution in future We can upgrade any analog xray system with External Detectors to Digital Radiography system. However the cost involved in the Hardware & Software for the up gradation should be borne by the purchaser. 125mA to 300mA The time range should be 5mS-0.9S Electronic Control console Kindly remove the point A-9 Remarks II. M/s. Agfa Healthcare India (P) Ltd Request you to separate both X-ray Machine & Computed Radiography as non of the Manufactures make both so that it will avoid taxation, additional expenses in transportation and creates comfort in post installation service support co ordination. Sl.No S. No. as per Tender specs Amendment requested By the firm M/s. Agfa Clinical Importance Remarks Healthcare India (P) Ltd 1 It should support the following plate sizes: (in inches) 14 x 17, 10 x 12, 8 x 10,15cm x 30 cm. It should support the following plate sizes: (in inches) and the image plates must be rigid type or image plates mounted on a tray for improved plate life. 14x17, 10x12, 8x10, 15cmx30cm. 2 Printer should have a resolution of above 300 dpi /ppi or more .and should Print the same resolution for all Radiological examinations It Should have single online tray with obtain of immediately replacement of any film sizes. Printer should have a resolution of above 500 dpi /ppi and should Print the same resolution for all Radiological examinations 3 It should have two online trays with capacity of printing all the film sizes available. The most Common reason for artifacts in image is Flexible plates as dust accumulating on rollers which creates marks on image plate when the image plate move on it. In some old generation digitisers image plates get bend inside the digiser, in that case plate gets worn out after a period. These issues are taken care in Rigid image plate Technology. Printing resolution is very important as all these machines are for PHC’s/CHC’s For any critical case study the patient can be sent to the nearest district hospital with the film for evaluation. With single tray, whenever the film size is changed the technician has to change the tray for each size and needs to be calibrated every time. Item Name : Ultrasound Scanner, T. No.4.19B /APMSIDC/2014-15, Dated: 03-12-2014. S.No. As given in Tender 1 Request amendment by the firm M/s. Trivitron Healthcare System should be 12" or more System monitor size should be LCD/TFT/LED monitor with 10” or above tilting function 2 Real time scanner –Scan depth Approximately 26 cm Scan depth is up to 24cm 3 ---2.0-6.0 MHz Convex Probe for Convex probe frequency 2.5 MHz to 6.0MHz. T.V Probe frequency 4.0 Mhz to 7.0Mhz 6 slices of TGC Mass patient data storage facility with at least 8G Bites(External or Internal) Battery facility is for 1 Hour (Should be External or Internal) 4 ---5.0-8.0MHz Transvaginal Probe 5 6 .8 slices of TGC Mass patient data storage facility with at least 8G bites 7 In-built battery with minimum 1 hour continuous scan facility should be supplied. Remarks 8 USB orts/VGA output output/Video Single USB port/VGA output/Video output. II) M/s. Esaote India (NS) Ltd., Sl. No. 1. Point No. of Specific ations 2 As mentioned in Tender Document Specifications 2 5 3. 7 2.0-6.0 MHz Convex Probe 4 8 Scanning modes B, M, B/M, B/B 5 10 Minimum 800 frames cine loop memory 6 17 Mass patient data storage facility with at least 8G bites 7 22 ISO Certification, CE approval System should be 12” or more LCD/TFT/LED monitor with tilting function Real time scanner Scan Depth approximately 26 cms Modifications required as by M/s. Esaote India (NS) Ltd., Clinical Importance System should be 12” or more LCD/TFT/LED monitor. As tilting of monitor is confined to one particular company. Real time scanner – Scan Scan Depth upto Depth approximately 26 26 cms and +/-2 cms and +/-2 cms will be cms will be acceptable acceptable. 2.0-6.0 MHz Convex Probe More flexibility +/- 1 MHz will be in choosing the acceptable. frequency. Scanning modes B, M, Preferable PW B/M, B/B and PW mode mode for will be preferred scanning of OB/GYN patients for heart beat. Minimum 256 frames cine Minimum 256 loop memory frames is ideal for B/W systems. Minimum patient data Minimum storage storage facility is 500MB will be 500MB and optional up to 500G Bites HD. ISO Certification, CE and USFDA is very USFDA approval proven certification in the world. III) BPL Medical Technologies Private Limited. Sl.No As given in Tender Request amendment by the firm M/s. BPL Medical Technologies Prt Ltd 1 1. Should be a portable ultrasound system with wide application for Abdomen, GYN\OB, Urology, Small Parts, Orthopediac, Cardiac etc 2. System should be 12" or more LCD/TFT/LED monitor with tilting function . Minumum 800 frames cine loop memory As per your specification point 1 says portable ultrasound system and as per your point 2 it says monitor should be 12" or more. So we would like to bring to your notice that bigger the monitor size the portability will be an issue. As per our understanding the portable ultrasound machines are seen with maximum 10" monitor (CRT/LCD). 2 its reads as 800 frames cineloop memory In general the retrival of the frames from the memory in portable b/w ultrasound machine is about 80 -100 Remarks. Remarks frames/sec (max) to see the moment of the organ. So 800 frames will be too high to and some times it take more time to retrive the data from memory. As per our understanding 100 frames will be more appropriote for clinical studies in a faster way in portable b/w ultraasound machine. 3 Image repetition rate not less than 150 frames / sec (higher preferred) it reads as repetition rate not less than 150 frames /sec. We would like to draw your attention that for performing applications like Gynac, OBG, Small parts in b/w ultrasound scanner 100 frames/sec would be ideal. 4 Multi calipers facility with electronic presentation for distances, area, circumference, ellipse,hip angel, OB/Gyn measurements, Cardiac, Urology, Orthopedic and Vascular measurement, volume and trackball measurement facility the machine should be useful for cardiac, Urology and orthopedic applications also. In this regard we would like to know more about the requirement of the software’s, hardware and the probes. 5 In-built battery with minimum 1 hour continuous scan facility should be supplied. . in-built battery with mim 1 hour backup is asked for. But as per our understanding on usage of portable b/w ultrasound in-built battery will add weight to the machine. Being asked for portable machine kindly give an option for external battery source also. Item Name : Semi Auto Analyzer, Vide T. No.4.19D /APMSIDC/2014-15, Dated: 03-12-2014. S.No. As given in Tender Request amendment by the firm M/s. IRIS Healthcare Technologies Pvt,Ltd. 1 Hard Glass Pyrex Cuvvette Fexible Cuvvette System: Flow cuvvette Remarks 11) M/s. Transasia Bio-Medicals Ltd., Sl.No As given in Tender 1 polygonal multi standard (calibration curve) 2 No: of Deltas : from 2 to 20 3 CE certification Request amendment by the firm M/s. Transasia Bio-Medicals Ltd., Clarification with an example for a hemistry/analyte/Parameter Clarification with an example for a chemistry /Analyte/Parameter. Is the delta for time or for absorbance or for temperature? Since it comes under IVD Guidelines Self Certification is acceptable-Needs confirmation. Remarks.