Lab 6: Gas Chromatography Flame Ionization Detector INTRODUCTION The gas chromatography instrument is a very useful instrument for organic chemistry. The GC is chromatography instrument that is meant for separation of volatile organic compounds. When using the GC for separation, it imperative that one analyze that sample before running it through a GC. In order for a sample to be run on a GC it must be volatile, in that sense, it must have a very low boiling point, preferably able to evaporate at room temperature stemming a little bit above and below that criteria. When samples are ran on the GC that information is relayed to the researcher via a chromatograph. Using a chromatograph one can obtain a wide range of information. Also, one can calculated concentration and retention times. The GC has two characteristic detectors, (1) is the mass spec analyzer, and (2) a flame ionization detector. Both detectors are unique. The FID analyzer detector was the detector used in this experiment. PROCEDURE - Made sure the helium, compressed air and hydrogen tanks are on and reading 80psi. Then made sure the desktop knobs were turned on, and ready to go. Also, made sure the flame on the GC was lit. Using the computer, turned on the GC software, and made sure the GC-FID software was activated. Checked the instruments parameters, and made sure all parameters were set to the right setting and the appropriate method was selected. Sample Preparation - The syringes were cleaned with acetone before and after each use. - Five organic solvents were obtained and labeled. - 0.1mL (10uL) of standard sample with 10uL of hexane (solvent). - Using the syringe, took up 1uL sample, and 1uL of air. - Sample was ready for injection. GC Analysis Procedure minutes. - When syringe was ready, injected the sample in the GC-FID and ran the sample for ten After all standard were ran, ran an unknown mixture. Special Notes The chromatograph of the standard should all have two peaks, (1) for hexane, and (2) another peak for the analyte, which is whatever we dissolved in the sample. RESULTS RETENTION TIMES OF SAMPLE STANDARDS & UNKNOWNS (cm) PEAK 1 PEAK 2 PEAK 3 PEAK 4 PEAK 5 Unknown 1.20 cm 1.90 cm 2.75 cm 6. 70 cm --- Undecane 1.20 cm 5.40 cm 6.75 cm --------- Nonane 1.25 cm 2.9 cm ------------- N-Dodecane 1.30 cm 6.65 cm ---------------- Decane 1.25 cm 4.30 cm ---------------- Octane 1.25 cm 2.00 cm --------------- The data obtained from the chromatograph using the FID, showed the retentions times displayed in the table above. The unknown mixture had 4 characteristic peaks that showed to have the components of 1.25 cm (hexane solvent peak), 1.90 cm (Octane), 2.75 cm (Nonane), and 6.70 cm (Ndodecane). The data showed to be a little more accurate each spectra obtain for each standard had mostly two peaks which corresponded to the solvent peak and the sample standard. The only sample that showed to have contamination was Undecane because three peaks were obtained. The peaks showed to be consistent with 1.20 -1.25cm being the solvent peak. Overall, results seemed to be way better than the GC-MS. CONCLUSION The unknown sample contained Octane, hexane, nonane, and N-dodecane. I feel more confident with this decision because when using the GC-MS the spectrum obtained seemed messed up and gave funky results. However, using the GC-FID seemed to produce better results. The spectra weren’t skewed in any way, although some standard did show some type of contamination it seemed minor. All the data showed to have a solvent peak of about 1.20 -1.25 cm and one other peak that was characteristic for that particular sample standard. I was able to look at the data and easy pick out which sample were present in the unknown because the retention times were very similar. When related the GC-MS and GC-FID I think for this experiment the GC-FID showed better results because separation was very good and the resolution between to peaks were good as well. The GC-MS spectrum showed peaks that were much skewed. The peaks didn’t separate well. I think that in reality both instruments are good at separating volatile organic solvents. However, I think the GC-MS might have had something wrong with it and that could have been a reason why it was not working properly.