Supplemental Material Supporting Information A Uniform Field Ion Mobility Study of Melittin and Implications of Low-Field Mobility for Resolving Fine CrossSectional Detail in Peptide and Protein Experiments Jody C. May and John A. McLean Department of Chemistry; Center for Innovative Technology, Vanderbilt Institute for Integrative Biosystems Research and Education; Vanderbilt Institute of Chemical Biology; Vanderbilt University, Nashville, TN Email: john.a.mclean@vanderbilt.edu Supplemental Material Details of the Helium Collision Cross Section Measurements The procedure for calculating the collision cross section (CCS) from experimental uniform field measurements involves three primary steps: (1) extraction of raw drift times from the ion mobility arrival time distributions (ATD), (2) correcting the raw drift times for the non-mobility time components, and (3) conversion of corrected drift times into the ion-neutral CCS using the Mason-Schamp relationship [1], which is used in the following form: 3βπβππ 2π πΆπΆπ = ( 16βπ ) β (π π΅ βπ 1 2 1 ππππ +ππππ 2 πβπ‘π πππ βππππ πΏ2 ) β(π ) β( β 273.15 π π β 760) (Equation 1) Here, Z is the integer charge state of the ion (unitless), ec is the constant for elementary charge (1.60217657 x 10-19 C), N is the gas number density (molecules/m3), kB is the Boltzmann constant (1.3806488 x 10-23 J•K-1), mion is the ion mass (Da), and mgas is the helium gas mass (4.0026029 Da). For step 1 of the CCS workflow, an ion mobility spectrum is extracted from a narrow m/z region and specific drift time ranges containing peak features of interest are then selected and processed for centroid time values. In the case where the peak resolution is insufficient for the software to calculate a centroid value, the peak centroid is determined by visual inspection, which allows the CCS of low abundance peak features to be estimated at a cost of reduced measurement precision. While somewhat subjective, we find that manual determination of peak centroids still provides good CCS precision (well below 2%). In step 2, these raw drift times are compiled for a range of drift fields (Table S1) in order to determine the non-mobility time component of the experiment. Because the non-mobility time component arises from the ion’s transit through several pressure and voltage regimes throughout the instrument (and thus has both ion mobility and mass time-offlight contributions), this value cannot be accurately determined a priori and so the stepped drift field method is used. It was found that the higher fields resulted in an abrupt loss of resolution for closely spaced peak features, and so only the seven lowest field values (1.9, 3.2, 4.5, 5.8, 7.1, 8.3, and 9.6 V/cm) were subsequently used for purposes of the drift time correction and CCS calculation. This observation is discussed in the manuscript. For each peak feature, the raw drift times (y-axis) are plotted against the inverse drift voltages (1/V, x-axis) and the time intercept is extrapolated from a linear fit to the data. In all cases, the coefficient of determination (R2) of the linear fit of drift time vs. 1/V was better than 0.999, suggesting that the experimental precision is high and that the ion system is governed by low field conditions where the drift field does not affect the ion mobility. The time intercept extracted from this procedure represents the time ions spend outside of the ion mobility drift region and is subtracted from the raw drift time in order to determine the actual ion mobility drift time of each peak feature. The rationale for this analysis is that the ion drift time will approach zero in the limit of infinite electric field (1/V = 0), leaving only the non-mobility time component. For the final step (step 3), once the actual drift times are determined, the CCS can be calculated directly from the Mason-Schamp relationship (equation 1). Supplemental Material Drift Fields Investigated and their Corresponding Voltage and E/N Values Drift Field (V/cm): 1.9 3.2 4.5 5.8 7.1 8.3 9.6 10.9 12.2 Drift Voltage (V): 151 251 351 451 551 651 752 852 952 E/N (Td): 1.5 2.6 3.6 4.6 5.6 6.6 7.6 8.7 9.7 E/N values correspond to 4.000 ± 0.001 Torr and 306.5 ± 0.5 K. Table S1. A summary of the drift fields investigated in this work. The corresponding drift voltages and E/N values (expressed as Townsend units, Td) are also provided. In uniform field ion mobility, the characteristic ion mobility energy parameter, E/N, can be directly calculated using the following equation: πΈ π π = πΏβπ β 1017 (Equation 2) Here V is the drift voltage (V), L is the length of the drift region (78.1 cm), and N is the gas number density (molecules/cm3), determined from the ideal gas law. The 1017 multiplier in Equation 2 converts V·cm2 into Td units (10-17 V·cm2) [2]. For this study, resolution of finer detail within peptide peak features is observed below an E/N value of ca. 4 Td, although successfully operating at these corresponding low field values is contingent on ability to obtain good ion statistics, which in turn depends on the abundance of the ion as well as the sensitivity of the instrumentation utilized. Supplemental Material Helium Collision Cross Section (Å2) Melittin Ion Form 0% MeOH 100% H2O 25% MeOH 75% H2O 50% MeOH 50% H2O 75% MeOH 25% H2O 100% MeOH 0% H2O [M+H]1+ 1A m/z 2845.76 469 ± 30 (3) a 2A [M+2H]2+ 2B m/z 1423.38 2C 2D 355.0 ± 1.1 (7) 385.1 ± 3.6 (7) 449.9 ± 2.9 (7) -- 355.0 ± 2.6 (7) 390.3 ± 2.9 (7) 449.6 ± 3.8 (7) -- 356.1 ± 1.0 (7) 393.8 ± 1.7 (7) 450.8 ± 3.0 (7) -- 359.1 ± 1.8 (7) 403.7 ± 1.5 (7) 451.5 ± 4.9 (7) -- -419.7 ± 8.9 (6) -469.7 ± 13.5 (7) [M+3H]3+ m/z 949.25 3A 3B 3C 405.9 ± 1.0 (7) 488.4 ± 1.5 (7) 523.0 ± 1.7 (7) 407.2 ± 0.7 (7) 487.8 ± 1.1 (7) 523.3 ± 1.6 (7) 404.7 ± 1.0 (7) 486.6 ± 0.9 (7) 523.6 ± 0.4 (7) 405.4 ± 0.7 (7) 487.6 ± 1.6 (7) 525.8 ± 1.1 (7) 407.8 ± 4.8 (7) 489.0 ± 7.4 (7) 526.6 ± 7.7 (7) [M+4H]4+ m/z 712.20 4A 4B 4C 541.9 ± 0.8 (7) 593.5 ± 2.7 (4) -- 543.2 ± 0.4 (7) 595.3 ± 1.2 (5) -- 540.8 ± 0.7 (7) 595.1 ± 1.3 (6) -- 540.6 ± 0.7 (7) 611.0 ± 3.2 (7) 657.5 ± 4.0 (7) 541.7 ± 7.1 (7) 602.9 ± 16.4 (7) -- [M+5H]5+ m/z 569.96 5A 5B 5C 5D 605.3 ± 5.6 (7) 631.7 ± 1.8 (7) --- 595.9 ± 4.9 (7) 628.4 ± 1.0 (7) --- 601.1 ± 2.1 (6) 602.0 ± 8.7 (7) 623.2 ± 0.9 (6) 626.3 ± 1.0 (7) 645.3 ± 12.9 (6) -670.0 ± 5.6 (7) -- 605.7 ± 6.8 (7) 629.9 ± 9.6 (7) 646.0 ± 7.1 (7) 669.0 ± 11.0 (7) -- 485 ± 10 (5) b -- -- a Ions generated by MALDI. The CCS was measured on a periodic DC-field drift tube instrument [3]. Ions generated by MALDI. The CCS was measured on a home-built drift tube instrument in the authors’ laboratory, which has been described previously [4]. b Table S2. A summary of CCS measurements for all signifcant peak features observed for various charge states of melittin. Singly-protonated melittin CCS measurements were obtained using MALDI, and categorized based on solvent preparation conditions. Values highlighted in bold represent the most abundant peak feature for the corresponding solvent and charge state. All CCS values exhibit an RSD of better than 2%. Supplemental Material Precision and Accuracy of Helium Collision Cross Section Measurements The IM-MS instrumentation used in this study was originally developed to support nitrogen-based CCS measurements. In order to determine the precision and accuracy for obtaining helium CCS values, a series of tetraaklylammonium (TAA) salts were analyzed. The helium CCS values of these salts have previously been measured using a confining-RF drift tube instrument [5]. Helium CCS results for the TAA salts are summarized in Table S3. The percent relative standard deviation (% RSD) for all CCS values is better than 1%, with a corresponding agreement to literature to be better than 0.5% based on the relative difference to the mean. These results are consistent with the performance of the instrument for nitrogen studies [6], and suggest good precision and accuracy can also be expected for the helium CCS measurements. Helium Collision Cross Sections (Å2) for Tetraaklylammonium Salts Ammonium Salt Mass (Da) CCS (Å2), This work a % RSD CCS (Å2), Literature b % Difference, Relative c TAA2 130.16 64.9 ± 0.2 (6) 0.3% 65.9 (0.0%) 0.4% TAA3 186.22 89.0 ± 0.4 (6) 0.4% 88.9 (0.1%) 0.0% TAA4 242.28 111.8 ± 0.6 (6) 0.5% 111.2 (0.0%) 0.1% TAA5 298.35 134.1 ± 0.8 (6) 0.6% 133.5 (0.0%) 0.1% TAA6 354.41 155.6 ± 0.9 (6) 0.6% 154.9 (0.2%) 0.1% TAA7 410.47 175.2 ± 1.1 (6) 0.6% 174.5 (0.5%) 0.1% TAA8 466.54 192.6 ± 1.6 (6) 0.8% 194.3 (0.0%) 0.2% TAA10 578.66 223.8 ± 1.8 (6) 0.8% -- -- a. Values in parenthesis represent the number of measurements. Values obtained from reference [5]. Values in parenthesis are the % RSD. c. The difference relative to the mean of the two values, expressed as a percentage. b. Table S3. A summary of CCS measurements obtained for the TAA salts. A corresponding comparison of the relative percent difference of measured values to ones reported in the literature are also provided. Supplemental Material Figure S1. An IM-MS spectrum of melittin electrosprayed from pure aqueous solvent. The HP-621, HP-921, HP-1221, and HP-1521 ions which appear in the spectrum are mass calibrants (Agilent ESI tune mixture). The mass spectrum does not exhibit any signficant ion signal between 1700 and 3200 (range not shown). Supplemental Material Details of the Experiments for Evaluating the Presence of Isobaric Aggregates The existence of aggregates (multimers) which are isobaric with the mass-to-charge can confound the interpretation of peptide and protein ion mobility data [7]. Since multimer formation is directly linked to solution concentration, a series of experiments were conducted at three different melittin peptide concentrations (1, 10, and 100 μM). These comparisons were conducted in pure water, as aqueous conditions have explicitly been shown to promote the formation of melittin aggregates [8]. Narrow m/z windows were isolated for each observed melittin charge state ([M + nH]n+, where n = 2 to 5) and these IM-MS spectra were examined for evidence of multimers. Based on an analysis of the mass spectral data, there is no evidence that multiplycharged multimers exist isobaric to the melittin charge states, as the isotope spacing for each ion species correspond to their presumed charge state (Figures S2-S4). The rationale here is that isobaric multimers would exist at higher charge state values. For example, m/z 1423.38 could arise from the [M + 2H]2+ ion, but also from the [2M + 4H]4+ or the [3M + 6H]6+ ions. Since mass spectrometry measures the mass-to-charge ratio of the ion, the presence of these higher charge state ions would be clearly indicated in the isotope spacing. This does not preclude the possibility that such multimers exist and dissociate prior to MS analysis, as multimers would be noncovalently-linked and thus subject to a weak dissociation energy threshold. An additional analysis was conducted in which the ion mobility ATDs from each peptide concentration was overlaid and examined for evidence of concentration-specific changes in individual peak features. The relative abundance and shape of peak features for all charge states remains largely unaffected by changes in solvent concentration across the two orders-of-magnitude investigated (Figure S5), suggesting that there is no significant contribution to multimers up to 100 μM peptide concentration. A slight decrease in the peak height of the “2A” peak feature is observed at higher concentration, although this is interpreted as not significant as the total abundance of the [M + 2H]2+ ion mobility ATD is low (ca. two orders-of-magnitude lower than the [M + 3H]3+ ATD based on peak height). Thus, there is no direct evidence of isobaric multimers in this study. To be certain, the peptide concentration was limited in this study to 10 μM, which represents the lowest concentration from which sufficient ion mobility spectra could be obtained for all charge states investigated. Supplemental Material Figure S2. IM-MS spectra obtained from 1 μM sample concentration of melittin peptide in pure aqueous solvent. Spectra represent IM-MS windows for the (A) [M + 2H]2+, (B) [M + 3H]3+, (C) [M + 4H]4+, and (D) [M + 5H]5+ ions. All spectra are obtained from a single experimental run with ion mobility separations at 3.6 Td (4.5 V/cm in 4.000 Torr helium drift gas). At this low concentration, traces of a singly-charged interferent is observed in the [M + 4H]4+ spectra (C), and this is considered when extracting the ATD and obtaining drift time measurements for the CCS. Ions in panels (A) and (D) are below the limits of detection at this sample concentration. Supplemental Material Figure S3. IM-MS spectra obtained from a 10 μM sample concentration of melittin peptide electrosprayed from pure aqueous solvent. IM-MS windows represent the (A) [M + 2H]2+, (B) [M + 3H]3+, (C) [M + 4H]4+, and (D) [M + 5H]5+ ions. All spectra are obtained from a single experimental run with ion mobility separations conducted at 3.6 Td (4.5 V/cm in 4.000 Torr helium drift gas). At this concentration, there is sufficient abundance of all four ion charge states for ATD feature comparisons and CCS measurements. Supplemental Material Figure S4. IM-MS spectra obtained from a 100 μM sample concentration of melittin peptide in pure aqueous solvent. Spectra represent IM-MS windows for the (A) [M + 2H]2+, (B) [M + 3H]3+, (C) [M + 4H]4+, and (D) [M + 5H]5+ ions. All spectra are obtained from a single experimental run with ion mobility separations at 3.6 Td (4.5 V/cm in 4.000 Torr helium drift gas). The isotope spacing observed in each mass spectra are indicative of the assigned charge state, indicating no contribution of ions originating from highercharged oligomers. Supplemental Material Figure S5. An overlay of ion mobility spectra obtained at 4.6 Td (5.8 V/cm) for different sample concentrations in pure aqueous solvent. Each panel represents a different charge state of melittin. All spectra are converted to CCS and normalized to the same maximum abundance to facilitate comparison. Significant peak features observed at this solution composition are noted. For all charge states, there is no significant change in the ATDs across two orders of magnitude in sample concentration, suggesting no strong contribution from multimers at these concentrations. The lower apparent abundance of the 2A feature at higher concentration is not reproducible at the other drift fields, which is attributed to the unreliability of peak height for comparing abundances of poorly-resolved features. Supplemental Material Figure S6. A comparison of the number of samples (bins) across the [M + 2H]2+ ion mobility ATD for (A) 9.7 Td (12.2 V/cm), and (B) 3.6 Td (4.5 V/cm). Measurements at higher drift fields results in compression of the mobility ATD to a narrow time range, which subsequently results in less sampling bins across the ATD. The panels at right expand the region of a mostly-resolved peak feature (2C), indicating ca. 16 bins across this peak at 9.7 Td. This is approximately half of the bins at 3.6 Td, however, this sampling resolution is still sufficient to discern details within the ATD. While not explicitly explored in this study, the issue of insufficient sampling at higher fields can be experimentally evaluated. Using the present commercial IM-MS, the total bins across each ion mobility spectrum is adjustable by varying the ion mobility acquisition rate (denoted as “IM Transient Rate” in the Masshunter acquisition software). By selecting different IM acquisition rates, the ATDs can be directly compared for evidence of insufficient sampling. Supplemental Material References [1] Mason, E. A., Schamp, H. W., Mobility of Gaseous Ions in Weak Electric Fields. Annals of Physics 1958, 4, 233-270. [2] Huxley, L. G. H., Crompton, R. W., Elford, M. T., Use of the Parameter E/N. British Journal of Applied Physics 1966, 17, 1237-1238. [3] Fernandez-Lima, F. A., Blase, R. C., Russell, D. H., A Study of Ion-Neutral Collision Cross Section Values for Low Charge States of Peptides, Proteins, and Peptide/Protein Complexes. International Journal of Mass Spectrometry 2010, 298, 111-118. [4] Sundarapandian, S., May, J. C., McLean, J. A., Dual Source Ion Mobility-Mass Spectrometer for Direct Comparison of Electrospray Ionization and MALDI Collision Cross Section Measurements. Analytical Chemistry 2010, 82, 3247-3254. [5] Campuzano, I., Bush, M. F., Robinson, C. V., Beaumont, C., Richardson, K., Hyungjun, K., Kim, H. I., Structural Characterization of Drug-like Compounds by Ion Mobility Mass Spectrometry: Comparison of Theoretical and Experimentally Derived Nitrogen Collision Cross Sections. Analytical Chemistry 2011, 84, 1026-1033. [6] May, J. C., Goodwin, C. R., Lareau, N. M., Leaptrot, K. L., Morris, C. B., Kurulugama, R. T., Mordehai, A., Klein, C., Barry, W., Darland, E., Overney, G., Imatani, K., Stafford, G. C., Fjeldsted, J. C., McLean, J. A., Conformational Ordering of Biomolecules in the Gas Phase: Nitrogen Collision Cross Sections Measured on a Prototype High Resolution Drift Tube Ion Mobility-Mass Spectrometer. Analytical Chemistry 2014, 86, 2107-2116. [7] Gillig, K. J., Chen, C. H., Critical Examination of Gas-phase Protein Conformation/Multimer Ion Formation by Electrospray Ion Mobility-Mass Spectrometry. Analytical Chemistry 2013, 85, 2177-2182. [8] Bazzo, R., Tappin, M. J., Pastore, A., Harvey, T. S., Carver. J. A., Campbell, I. D., The Structure of Melittin. A 1H-NMR Study in Methanol. European Journal of Biochemistry 1988, 173, 139-146.