IR, Raman, & IR Microscopy Lab Introduction: Infrared spectroscopy is a type of instrumentation that uses the wavelengths found in the infrared region of the electromagnetic spectrum to determine a particular compound. One of the most widely used types of infrared spectroscopy is known as the Fourier Transform Infrared Spectroscopy, which is named after the mathematical algorithm it uses to interpret the data it receives. This instrument works by passing shining a beam of infrared energy through a chemical sample, where some of the wavelengths of light are absorbed by the sample whereas others simply pass through, or transmitted. Since each chemical compound absorbs/transmits individually, no two spectrums are similar and therefore each compound has a “molecular fingerprint” of frequencies of that compounds molecular bonds/structure. Raman Spectroscopy is another type and it utilizes a laser as an energy source through the sample and the scattered light is then collected from the sample. The instrument measures the vibrational frequencies that are characteristic to each particular molecule, bond, etc. From these frequencies, the type of compound can then be identified. Purpose: There are two procedures in this particular lab, each with its own individual purpose. In the first FTIR procedure, the purpose of the experiment is to first learn the basics of how the IR operates by running various organic samples through the instrument and identifying the resulting peaks. After identifying the peaks of the organic samples, mixtures of these samples will then be run in order to produce a calibration curve, from which the identity of an unknown mixture will be derived. The second procedure in this experiment Instrumental Analysisuses the FTIR microscope, in which the purpose is to understand the basics of operating the instrument and then to run unknown fiber samples through the IR and uses the observations and data collected to determine what the fibers might be. Instrumental Procedure: Varian 3100 Excalibur Series FT-IR 1. 2. 3. 4. 5. Make sure that the FTIR is on (should not be turned off but have to check) The black lever between the FT-IR and the Raman should be pushed down Open the software by double-clicking ‘Varian Resolutions Pro’ icon From ‘current scan’ drop down menu at bottom of screen, select ‘Basic IR Scan.’ From ‘collect’ menu, select ‘rapid scan’ and make sure parameters are set (under ‘electronics’ tab) to speed- 20KHz, filter-5, UDR- 2, and save range- custom 4000700 cm-1. 6. Select the optics tab and set parameters to: a. IR source – Mid-IR b. Beam – Internal c. Detector –DTGS d. Beamsplitter – KB r Broadband e. Accesory – None f. ATR Crystal –None 7. 8. 9. 10. g. Optical Filter –None h. Aperture – Not installed Click setup Center burst should read between 1 and 5. Click ‘ok.’ To run background scan, click ‘background’ and make sure no sample is in the compartment. To run sampel, click ‘collect’, the in rapid scan menu, click’scan’ Varian 600 UMA FT-IR Microscope 1. 2. 3. 4. 5. Fill the Microscope detector with liquid nitrogen Insert ATR crystal To set up software, open software by click ‘varian resolutons pro’ From ‘current scan’ drop down, select method file ‘microscope’ From collect menu, select rapid scan, check that parametres are set up same as above procedure states 6. Run a background scan and then run the samples Experimental Procedure: FTIR: 1. Obtain three different organic compounds from the available stock samples. 2. Run each sample individually to identify the characteristic peaks for each. 3. Run a 1:1:1 ratio solution of each and identify the characteristic peaks. 4. Repeat #3 for a 2:1:1. 1:2:1, and 1:1:2 ratios. 5. Make an unknown sample and attempt to identify the ratio of each organic compound Data: 1. FTIR: Compound 2-propanol Toluene 2-butanone 1:1:1 ratio Key Peak Peak Description 3000-3500 (broad peak) 2800-2900 2800-3100 1500 + 1600 (twin peaks) 790 2800-3000 ~1700 3000-3500 (broad peak) 2800-3000 ~1700 1500 + 1600 (twin peaks) ~790 OH alcohol CHsp2/sp3 CHsp2/sp3 Alkene (aromatic) mono sub. Benzene CHsp2/sp3 C=O OH alcohol CHsp2/sp3 C=O Alkene (aromatic) mono sub. Benzene 2:1:1 ratio 1:2:1 ratio 1:1:2 ratio Unknown ratio 3100-3400 (broad peak) 2800-3000 ~1700 1500 + 1600 (twin peaks) ~790 2800-3000 ~1700 1500 + 1600 (twin peaks) ~790 2800-3100 ~1700 1500 + 1600 (twin peaks) ~790 2800-3100 ~1700 1500 + 1600 (twin peaks) ~790 OH alcohol CHsp2/sp3 C=O Alkene (aromatic) mono sub. Benzene CHsp2/sp3 C=O Alkene (aromatic) mono sub. Benzene CHsp2/sp3 C=O Alkene (aromatic) mono sub. Benzene CHsp2/sp3 C=O Alkene (aromatic) mono sub. Benzene Unknown Ratio Determination: 0:1:2 (2-propanol:toluene:2-butanone) 2. Raman Compound Key Peak Peak Description 2-propanol 3100-3000 tall peak ~3000 ~900 3000-3100 tall peak 1600 ~900 OH alcohol CHsp2/sp3 mono sub benzene OH alcohol C=O ortho sub benzene Toluene methyl salicylate 3. IR Microscopy a. The IR microscope was not working properly, so we were not able to complete this part of the procedure. Conclusion: Overall, I was quite confident in the overall results obtained from this experiment. Both the IR and Raman results were easily obtained, and the results from each test were conclusive and easily analyzed as each spectrum corresponded to its respective compound. They were easily utilized to compare relative concentrations of each compound in each mixture as well as to ultimately determine the mixture of compounds that was prepared for us. However, the only real trouble experience with this experiment was just the fact that the microscope was not functioning properly, and therefore that part of the procedure were not able to be completed.