UGC DAE Consortium for Scientific Research Kolkata Centre (Formely Inter University Consortium for DAE Facilities) LB/8, Sector-III, Bidhan Nagar, Kolkata-700098 Ph. No. 033-2335-1866 Fax No. 033-2335-6543, 2335-7008. Specifications of Charged Particle detectors. 1 A (A.i) (A.ii) B Detector Telescope for light charged particle spectroscopy 2 Double Sided Trapezoid detector Si detectors – fully depleted transmission type detectors of : 3 nos each Thickness ~ 200 m & area 100 mm2 2 3 nos each Thickness ~ 500 m and area ~150 mm Partially depleted Stop detectors with depletion depths of 2 3 nos each mm and area 200 mm2 All the detectors are with radial mounted micro-dot connectors(F) with expected resolution ~20 keV or better for ~ 5.4 MeV alpha particles 6 nos o DSSD (Double-Sided Silicon Detector) 60 wedge detector Active Area: 80 cm² No of Channels: 24 Side Length: ~ 16.0 cm Window: Type 2M (<1µm) Subtends 60° (17 mm to 135 mm radius) Connector: DL 50-way 16 junction (radial) strips, pitch ~ 6.5 mm 8 ohmic sectors, each subtending ~ 7° 2 guard rings 24 grounds Package: PCB Transmission (1.6 mm thickness) Detector thickness: 150 µm Full Depletion (FD): ~ 30V max Total Leakage Current (FD+30V) (20ºC): 3 uA max Operating Voltage: FD to FD+30V Total Alpha Resolution (Junction): 300 KeV FWHM 1 µS TC max Total Alpha Resolution (Ohmic) 300 KeV FWHM 1 µS TC max Minimum Acceptance level: 100% (all channels operational) Thickness 150 m 3 Silicon Pad detector Depletion depth of 100 m. Active area is 40.0 (mm) x 15.0 mm Chip dimensions 43.0 mm x 18.0 mm Appropriate guard ring to be provided Preferably packed as a ready to mount PCB with minimal dead Layer / volume, or in a chip format. Vendors to specify the expected performance (energy resolution). 12 nos, 4 CsI(Tl) detectors for use in vacuum 25 nos CsI(Tl) detectors (integral assembly with photo-diode and preamplifier) for charged particle detection to be used in vacuum (preferred heat dissipation of 50 mW per unit. Vendors may quote the expected heat dissipation.) 20 x 20 x 2 mm Csl(Tl) Scintillation Detector with a 2 -3 micron aluminized mylar entrance window. Coupled via a light guide to a 10x10 mm active area PIN photodiode with a size of 12.7 x 14.5 mm (S3590-08 or eq) The photodiode is coupled to a hybrid charge sensitive preamplifier The charge sensitive pre-amplifier should have a 50 impedance The pre-amplifier should be operated from a + 6 - +12 V power supply The detector should have an operating bias of around 50 V , with a current of a maximum 1nA and the maximum permissible operating voltage should be +100V The energy resolution should be 5-6% (fwhm) for 241Am alpha particles. The entire assembly is enclosed/ encapsulated in a compact housing . Vendors may also quote for an alternate (square) shaped encapsulation. The connections should be with provided with flying leads for signal, power supply (pre-amplifier and bias) and a ground and a maximum of 50 cm cable length (minimum 20 cm in length). It is preferred that the cables be brought out from the rear. However, the vendors may provide for an alternate configuration for the cables. The rear housing plate should have 4 M6 blind tapped holes to couple the housing to the mounting structure.