John Siller Laboratory 5 UV_Vis Spectroscopy Introduction Absorbance of wavelengths in the ultraviolet and visible spectra can be used to determine the concentration of a particular analyte in solution. Generally the more analyte in a solution corresponds to more absorption of the electromagnetic radiation and therefore the less transmittance is allowed through the solution. UV-Vis spectroscopy relies on the principle that π-electrons and non-bonding electrons can absorb radiation and become excited. This absorption of UV and visible light is used to determine the concentration of a solution once it is compared to the absorption through a blank containing none of the analyte. UV-Vis spectroscopy is used to analyze a number of different analytes from transition metal ions and organic molecules to large biological macromolecules. In this lab the concentrations of unknown solutions of aspirin containing salicylic acid in hexane were determined using calibration curves prepared with concentrations of 10 ppm, 20 ppm, 40 ppm, and 50 ppm. In addition to UV-Vis Spectroscopy this lab also focused on spectrofluoroscopy. When a molecule absorbs electromagnetic radiation and is excited it can shift down to the lowest vibrational state of the excited state and then release its energy falling down to the ground state. When this occurs there is the chance that a photon is released and the sample fluoresces. Measuring the fluorescence of a sample is another way of quantitative determining the concentration of the analyte. In this lab the concentration of unknown quinine solutions was determined using a calibration curve with standards of 1 ppm, 2 ppm, 4 ppm, 6ppm, and 8 ppm. The quinine was dissolved and diluted in sulfuric acid and run through our spectrofluorometer. Procedure: UV-Vis Turn on instrument Open VisionPro software and allow startup diagnostics Select the proper scan Adjust the wavelength, scanning, and lampsettings Set the baseline correction Prepare blanks and samples Load blanks first Set the menu to peak pick Load sample and run using the run icon repeating for multiple samples Results: Trial results are listed in lab notebook. Each standard concentration was run three times, and then the unknown was run three times. Below is the calibration curve calculated by the means of the three trials at each concentration. Concentration Trial 1 Trial 2 Trial 3 Mean (ppm) Absorbance Absorbance Absorbance 10 0.497 0.492 0.495 0.494667 20 0.862 0.842 0.852 0.852 40 0.998 0.991 0.989 0.992667 50 1.265 1.271 1.260 1.265333 Unknown 1.147 1.136 1.150 1.144 Table 1: Salicylic acid content determined using absorbance on the UV-Vis spectrometer 1.4 Calibration Curve for UV-Vis 1.2 Absorbance 1 0.8 y = 0.0168x + 0.3966 R² = 0.9166 0.6 0.4 0.2 0 0 10 20 30 40 50 60 Concentration (ppm) Figure 1: Calibration curve of salicylic acid absorption analyzed using the UV-Vis spectrometer Using the calibration line y=0.0168x+0.3966 and our average absorbance value of 1.144 for our aspirin unknowns we calculated the unknown as 44.5 ppm. This calibration curve is not as good as it could have been. This could be due to inaccuracy in our standards or from the use of different cuvettes. For the analysis on the spectrofluorometer of quinine in tonic water standards at concentrations of 1, 2, 4, 6, and 8 ppm were prepared. The quinine standards were diluted using sulfuric acid. For each standard and sample multiple peaks at different wavelengths were observed and recorded. The averages of these peak intensities are included in the table below. A full list of peak intensities is recorded in my lab notebook. 341 nm Concentration 1 2 4 6 8 Trial 1 Trial 2 Trial 3 10.631 10.671 11.256 73.174 74.658 74.027 16.222 14.815 15.264 78.686 75.542 75.116 31.748 29.143 27.572 Mean 10.85267 73.953 15.43367 76.448 29.48767 448 nm 1 280.638 291.422 283.754 285.2713 2 519.088 515.072 515.252 516.4707 4 995.766 993.242 991.648 993.552 6 1015.189 1015.189 1015.189 1015.189 8 1015.189 1015.189 1015.189 1015.189 683 nm 1 2.309 2.395 2.705 2.469667 2 13.003 13.236 13.203 13.14733 4 3.626 3.314 3.195 3.378333 6 12.999 13.013 13.077 13.02967 8 6.632 4.818 6.619 6.023 883 nm . 1 13.714 14.083 13.423 13.74 2 24.464 24.372 24.187 24.341 4 46.235 46.18 46.139 46.18467 6 55.514 55.694 55.56 55.58933 8 84.563 84.516 84.467 84.51533 Table 2: Calibration data for quinine in sulfuric acid 1200 1000 341 nm 800 448 nm 683 nm 600 883 nm 400 Unknown y = 9.5873x + 4.6073 R² = 0.9792 200 Calibration 883nm 0 0 2 4 6 8 10 Figure 2: Calibration data and curve for quinine in sulfuric acid This data is rather inconsistent and a good calibration curve can only be made using the peaks recorded at 883 nm. For some reason the peaks for 448 nm plateaued after 4 ppm and remained constant for 6 ppm and 8 ppm at 1015.189. The optimal wavelengths for excitation and emission were supposed to be 343 nm and 448 nm respectively. These data points were not consistent enough to be used for a calibration curve and therefore I needed to make one using the 883 nm peaks. Using this curve, y=9.5873x+4.6073, and the average intensity of 25.41 we can calculate the concentration for our unknown tonic water. The result is a concentration of 3.13 ppm. Conclusion: Spectroscopic techniques such as UV-Vis and fluorescence are very useful for determining quantitative information for a known sample. Concentrations can be determined using calibration curves prepared with known standards. A wide range of industrial applications use these techniques for quality assurance and testing. It is important to learn the mechanics behind these instruments and become familiar with running samples on them. This lab went fairly well with nothing disastrous taking place, although our results could have been better. Improvement would rely on running the samples more and troubleshooting any problems that arise. The preparation of calibration standards is the most important part of the procedure when running these instruments. The quality of these standards determines the effectiveness of the instrument and the quality of the results. Without good standards that can be used to create good calibration curves the effectiveness of these instruments is nullified. Our data could have been better and could be improved by redoing our calibration standards until they formed a good useable calibration curve, but due to time constrains this was not possible. For such apparently simple instruments this lab was complex and the errors we received cannot be explained easily.