I. Introduction 1.1 Transtibial Prostheses Over 40,000 transtibial, or below-knee, amputations are performed annually in the United States. While prosthetic limbs help restore mobility and normal appearance to the amputee, the skin and underlying tissue at the residual limb do not react well in load-bearing activities. The prosthetic socket plays a crucial role in assisting with load transfer and providing stability at the prosthetic interface. Thus, it is important that the socket is designed to optimally distribute the load to the residual limb to maximize comfort to the patient and ensure proper movement. A typical prosthetic contains a socket, foot, and liner as shown in Figure 1. Many studies have been done to elucidate the mechanism through which discomfort arises at the prosthetic interface in an attempt to develop effective sockets. However, patients continue to report pressure ulcers, blisters, cysts, edema, skin irritation, and dermatitis due to poor fitting of the socket (Polliack, Sieh, Craig, Figure 1: A prosthetic system consists of a foot, liner, and socket. Landsberger, McNeil, & Ayyappa, 2000). It has been demonstrated that knowledge of the biomechanical factors acting at the prosthetic interface is crucial to the development of comfortable sockets, such as the patellar tendon bearing (PTB) transtibial socket that transfers the load on pressure-sensitive regions to the pressure-tolerant regions of the residual limb (DeLisa, 2005). Since the 1960s, the PTB has internationally been the most commonly accepted standard transtibial socket (DeLisa, 2005). Another method aimed to increase comfort to the patient distributes pressure more evenly across the entire limb, which is employed by the total surface-bearing (TSB) socket. Regardless of the method, it is important to determine an accurate means of characterizing the mechanical interactions at the interface. Studies quantifying the stresses at the prosthetic interface have been conducted for over 40 years (Zhang, Turner-Smith, Tanner, Roberts, 1998). Computational modeling using finite element analysis is a common and simple method for predicting stresses and strains at the interface in that estimates of pressures at any region of the limb can be made, and elaborate experimental setups are not required. However, while convenient, computational models do not account for all potential factors at the interface, 1 specifically the biological tissue reactions to loading (Mak, Zhang, & Boone, 2001). Thus, computational modeling may simply serve best as a means of verification of experimental data. Experimental methods involving the use of sensors to quantify the stresses at the prosthetic interface are more desirable in that measurements obtained from actual trials incorporate all factors contributing to discomfort at the interface. However, devising an accurate measurement system remains a challenge. Many factors, such as the size of the sensor, the type of force transducer, and differences in pressure dynamics between types of movement, often compromise the accuracy of stress measurements (Mak et al., 2001). For example, too large of a sensor provides an inaccurate average of too large an area of the limb, while sensors that are too small produce undesirable edge effects (Mak et al., 2001). Thus, careful consideration of these factors is necessary in experimental procedures. Researchers have been interested in quantifying two primary stresses at the prosthetic interface: shear stress, resulting from frictional forces between the socket and skin, and interfacial pressure, resulting from load-bearing activities. While both types of stresses contribute to patient discomfort, most studies have predominantly focused on interfacial pressures, and few studies have successfully quantified shear stresses due to inaccurate measurement transducers (Zhang et al., 1998). Furthermore, it is thought that minimizing shear stresses may not be desired. One study demonstrated that smaller coefficients of friction effectively reduced shear stresses but led to an increase in normal stresses in order to support the same load, thus perpetuating damage to the underlying tissue of the residual limb (Mak et al., 2001). Thus, some amount of friction may be desirable in order to help support the load on the limb. Additionally, a gel liner, or the layer that fits in between the prosthetic socket and residual limb, is used to help reduce shear stresses and increase comfort (DeLisa, 2005). Thus, the predominant need in the development of effective transtibial socket lies in the accurate measurement of interfacial pressures. 1.2 The Surgical Clinic The Surgical Clinic in Nashville, TN specializes in fabricating prosthetic devices and sockets for amputees. Socket-fitting is conducted on a case-by-case basis, according to the patient’s qualitative feedback and established pressure-sensitive and pressure-tolerant regions of the residual limb. Yet, although each socket is tailored to the specific needs of the patient, reports of pain and discomfort are common, necessitating refitting procedures that constitute a hefty portion of the costs for each prosthesis. The process for constructing a transtibial socket at The Surgical Clinic begins with an evaluation of the patient for several factors, including skin type and integrity, limb measurements, range of motion, hand dexterity, fine motor skills, gross motor skills, cognition, etc. Then, the material for the gel liner is selected. Depending on the limb shape, off-the-shelf liners or custom-made liners can be used. The most common options include the urethane, thermoplastic elastomer, and silicone liners. After the gel liner is selected and fit to the patient, a cast is measured and constructed over the liner, either by hand or using 2 CAD-CAM digitizing. The negative cast is then modified by hand or using computer modeling before the positive check socket is constructed. The positive check socket can be made from a number of materials, which have been selected by the prosthetists at The Surgical Clinic based on qualitative characteristics of the material. For example, Polytol, a high-quality material that is polyurethane-based and adheres well to the skin is the most ideal material for use in constructing the socket. After the positive check socket is constructed and assessed in both static and dynamic conditions, the final laminated socket is fitted to the patient (A. Fitzsimmons, personal communication, January 10, 2011). The entire process is both time-consuming and costly, taking over a month and averaging around $2500 each time. The heavy reliance on qualitative assessments of material characteristics and patient feedback throughout the process results in patient discomfort and numerous refitting procedures down the line. As material costs are fixed, it is worthwhile to attempt to reduce the number of refittings necessary after a socket is made. In fact, The Surgical Clinic spends approximately $180,000 a year on refitting procedures. Thus, transforming the socket construction process from a qualitative one to a more quantitatively-based method is vital. 1.3 Goals In identifying our goals for the project, it was necessary to consider what could reasonably be accomplished in the given amount of time (around eight months). It was decided that a long-term study with one patient, rather than few trials with multiple patients, would provide more useful results. Thus, the overall goal was to develop a novel socket system that would optimally distribute the pressures on the residual limb for the patient in our study. If successful, the measurement system could then be implemented in constructing sockets for other amputees as well. In working towards the overall goal, several objectives were established. First, an appropriate and accurate system for acquiring pressure measurements needed to be designed. The measurement system needed to be reliable, as several trials were to be conducted with the patient, it needed to minimize interference with the prosthetic interface, and include an appropriate force transducer. Second, using the measurement system, the patient would be tested under several conditions—standing on both legs, standing on the prosthetic leg, and walking—and using all combinations of the gel liner and prosthetic foot in order to establish a pressure distribution across the prosthetic interface. Finally, after examination of the data, the socket system, comprised of the liner and foot that most effectively reduces the pressures on the limb and a newly constructed socket transferring the peak pressures from the pressure-sensitive regions to the pressure-tolerant regions of the residual limb, would be designed. The use of a systematic measurement procedure and subsequent socket redesign should effectively reduce The Surgical Clinic’s costs by lowering the number of refittings necessary for each amputee, while improving the utility and comfort of the prosthetic socket. 3 II. Methods 2.1 Determination of Data Acquisition Method After thorough research of the technology used to acquire force measurements, Force Sensing Resistors (FSRs) were determined to be the best device due to their thin structure and reliable decreases in resistance with increases in force. FSRs (Model 402), produced by Interlink Electronics (See Appendix A1.1) with a .5” diameter were chosen, because they were the largest size that did not contain undesirably cumbersome leads that could cause skin abrasions or inadvertently redistribute pressures within the socket. They were also small enough that the contact interface between the Figure 2: Current to Voltage Converter active area of the FSR and the subject’s skin would be approximately flat. This was necessary to avoid false triggering due to bending of the FSR. A current-to-voltage converter (Figure 2) was selected as the best method to convey changes in FSR resistance to recordable output voltages due to the direct proportionality between the two quantities, as calculated by the following equation: The FSR Integration Guide (“FSR Integration Guide,” 2010) stated that a VREF of 0.1V to 5.0V was optimal for the FSRs; VREF was selected to be 5.0V in the measurement circuit to minimize the signal to noise ratio of the recorded signal. Initially, gain resistors (RG) of 470Ω were chosen so that the expected maximum pressure, obtained from a literature review, would not exceed the maximum possible output voltage of the circuit (10V) (Smith, Michael, & Bowker, 2004). However, most of these gain resistors were switched to 680Ω after a preliminary test with the patient at The Surgical Clinic revealed that most of the regions tested in the socket needed higher gains to optimize sensitivity. 2.2 FSR Calibration 4 Each FSR was calibrated in order to convert the voltages obtained from the testing to pressure values. After the necessary calibration ranges were determined from a preliminary test with the subject amputee, ranges of fifteen different weights were selected for each gain resistor. Circuits using the 470Ω resistor were calibrated with weights ranging from 100g to 4500g, and circuits using the 680Ω resistor were calibrated with weights ranging from 100g to 3500g (See Appendix A1.2 and A1.3). These calibration weights, borrowed from the Vanderbilt Physics Department, were placed sequentially in a basket weighing 250g that was balanced on a rubber stopper and centered on the FSR. The circular rubber stopper had a diameter approximately equal to that of the circular active area of the FSRs, which allowed the force from the calibration weight to be evenly distributed across Figure 3: The set-up for calibration. The basket contained the free weights. the region of the active region (Figure 3). A voltage reading was acquired five times at each mass value for each FSR. After the first 1200 calibration measurements demonstrated that the standard deviations between each repeated mass value were typically less than 0.1V, only three voltage readings were taken. The mass of the weights was then converted to pressure using the following equations: 𝐹 𝐴 𝑃 = , 𝐹 = 𝑚 ∗ 𝑔, 𝐴 = 𝜋 ∗ 𝑟 2 𝑃 = 𝑃𝑟𝑒𝑠𝑠𝑢𝑟𝑒, 𝐹 = 𝐹𝑜𝑟𝑐𝑒, 𝑚 = 𝑚𝑎𝑠𝑠, 𝐴 = 𝑎𝑟𝑒𝑎, 𝑟 = 𝑟𝑎𝑑𝑖𝑢𝑠, 𝑔 = 𝑔𝑟𝑎𝑣𝑖𝑡𝑦 (9.8 𝑚 ) 𝑠2 Pressure(kPa) Voltage was then plotted as a function of 400 350 300 250 200 150 100 50 0 y = 4.7439x2 + 9.0666x + 29.264 R² = 0.9953 pressure for each of the FSRs (Figure X). A linear relationship was fairly accurate for the lower values of pressure but was less accurate at higher values. It was determined that the pressure and voltage had a polynomial (power of 2) relationship due to the high R2 value for this relationship (R2 was 0.9953 for the FSR in 0 2 4 Voltage(V) 6 Figure 4: Pressure as a function of Voltage. A second degree polynomial relationship is seen. 8 Figure 4). An equation from the calibration curve of each FSR was determined and used to convert voltages 5 to pressures for the data acquired from testing. A calibration curve was made for each FSR to account for the variability between each FSR. 2.3 Testing Subject One volunteer amputee participant was used in the study. This participant was a middle-aged male who has been assisting in the production of prosthetic sockets for several years. He was chosen due to his prior knowledge of prosthetics and ability to easily switch between the socket and leg combinations used in the study. The participant is a K-4 Medicare level, which means he has “the ability or potential for prosthetic ambulation that exceeds basic ambulation skills, exhibiting high impact, stress, or energy levels” (“K Levels,” 2005). Due to time constraints, more subjects could not be tested. Unlike most transtibial amputees, the test subject was capable of bearing high pressures at the distal end of the limb and preferred to have high pressures in that region. When developing the new socket, pressures were to be minimized in this region to make the design fundamentals applicable to a larger proportion of transtibial amputees. 2.4 Materials A typical lower limb prosthesis consists of three components: a foot, liner, and socket. The three most common feet used in industry today are the Vertical Shank ESF, Flex-Foot & Multiaxial Ankle, and SACH foot. The three most common liners are the thermoplastic elastomer, urethane, and silicone liner (Figure 5). The Vertical Shank and Flex-Foot have moderate shock absorption capacities, and the urethane and silicone liners are thicker than the thermoplastic elastomer liner. Before designing a new socket, Figure 5: The three most common liners from left to right: urethane, thermoplastic, and silicone. the best liner and foot combination had to be determined. In order to improve the current design, the best option currently available to patients had to be determined as described in section 2.8. 2.5 Regions of Interest After researching the regions of the transtibial amputee residual limb, it was determined that data should be obtained for both pressure tolerant and pressure sensitive regions of the limb. Eight regions were tested due to the limitations because there are only eight available channels on the NI data acquisition interface. Mr. Fitzsimmons aided in the selection of these regions based on his previous experience. Four of these regions shown in Figure 6 are pressure-sensitive regions: the distal end of the limb (2), the fibula head (6), the distal tibia (7), and the tibial tubercle (8). The other four regions shown in Figure 6 are pressure tolerant regions: the center of the gastrocnemius muscle belly (1), the medial 6 tibial flare (3), the mid patella tendon (4), and the tibialis anterior (5). The FSRs were secured to the participant’s skin with Scotch tape during testing. Mr. Fitzsimmons placed the FSRs during our first trial due to his knowledge of the anatomy of the residual limb (See Appendix A1.4) Pictures of the regions (see Figure 6) were taken and referred to in all later trials to Figure 6: A FSR was placed on each of the eight labeled regions. The pictures in order from left to right show the right lateral (outer) view, the anterior view, and the left lateral (inner) view of the residual limb. A portion of the distal end is shown in the picture on the right. The posterior location of the gastrocnemius muscle belly is not shown. ensure proper placement of the FSRs in the same regions for every trial. 2.6 Setup One breadboard was used for throughout the testing and calibration procedures. The breadboard contained eight individual circuits (current-to-voltage converters), with each circuit powered by a power supply. The output of each circuit was read out by LabVIEW using a BNC cable to connect the circuit to the NI Instruments Box. Each FSR was connected to its corresponding circuit using seven feet of 22 AWG, RoHS compliant wire. The electrical connections to the FSR leads were made with conductive silver epoxy from MG Chemicals, since solder would have damaged the FSRs. Since this connection was fragile, each FSR-to-wire connection was sealed with a thin, smooth layer of high strength nonconductive marine epoxy. An initial attempt was made to establish an easily transportable LabVIEW program using the National Instruments CompactRIO system. In addition to its mobility, the CompactRIO was chosen based on its ability to interface with most computers and the high number of available channels for recording the signals from each FSR. It was essential that our measuring system be able to move easily because we planned on testing at Figure 7: Output from all eight channels seen in Labview program. The data is outputted as voltage vs. time. The Surgical Clinic. However, numerous problems were encountered regarding the 7 implementation, setup, and reliability of this device. Thus, the NI BNC Connector Block used in Vanderbilt’s Biomedical Engineering Instrumentation laboratory was used to digitally acquire the output signals of the FSRs. As shown in Figure 7, a LabVIEW program was created to display and save the voltage versus time data from each test. The large window of the program displayed the original signal from the FSRs. The windows beneath displayed the inverted signals by FSR grouping so that it was possible to view the results of a particular FSR and quickly assess if the results and determine if it malfunctioned. These signals were recorded at 1 kHz so that there would be adequate single sampling for the portions of walking trials where pressures changed very quickly. These trials were saved in a format that could be opened in excel to apply the calibration curves to the voltage versus time data. An excel template file was created to automatically convert the voltage data into pressures, create a graph of the trial data, and calculate the necessary statistics for analysis. 2.7 FSR Replacement Each time that an FSR broke during testing, as determined from viewing the LabVIEW module, the particular FSR was replaced with a new FSR. Each new FSR was calibrated, and trials conducted with a broken FSR were repeated. 2.8 Trial Procedure and Analysis To determine the optimal combination of liner and foot to be used in the new socket system, pressure recordings were acquired for each of the nine possible combinations of three liners and three feet. The sockets used for each of these trials were made of the same material and had the same shape, but minor adjustments were made due to the thickness of the liners, which was not significant enough to alter the results. For each combination, three trials each were recorded for the patient walking two mph on a treadmill, standing on the prosthetic limb only, standing on both legs, and standing on the normal leg only with the prosthetic leg suspended. At least one trial was recorded with the subject wearing the liner without the prosthetic limb attached before putting on the leg for the walking and standing tests. Since the FSRs are prone to significant long-term drift but are stable over short periods of time, the subject removed the prosthetic limb in between each trial to reset the FSRs. Furthermore, the standing trials were limited to three seconds in length, and the walking trials were limited to nine seconds in length. Once the trials were recorded, they were copied into an excel template file that converted the data into pressures and calculated the minimum pressure, maximum pressure, and average pressure recorded for each FSR during the trials. These summaries are included in the appendix for the standing trials. The statistics for the walking trials were manually extracted. The peak pressures for each FSR during each gate cycle were recorded and compiled. The average pressure was calculated from an established point on the first recorded gait cycle to the equivalent point on the last gate cycle for each walking trial. 8 Similar procedures were followed to measure the redistribution of pressures in the sockets that were redesigned after analyzing the data acquired from testing the first nine combinations of sockets and liners. A minimum of five trials for each new socket were conducted with the subject standing on both legs for three seconds and walking at two mph for nine seconds. This was also done for the optimal socket and foot combination with an original test socket and with the socket, liner, and foot that the subject normally wears. The average of a 0.25 second segment of the data for each standing trial was calculated to assess the persistent distribution of forces during normal stance. A 0.25 second segment was used to avoid large fluctuations due to the subject shifting weight while standing. The average and maximum pressures for the walking trials were manually acquired by the same methods used when analyzing the first nine combinations of sockets and liners. 2.9 New Socket Materials When designing the new sockets there were two main factors of the socket that were manipulated: contour and material. In order to relieve pressure in certain regions or add pressure in other regions the contour can be changed by cutting some of the socket away, which will relieve pressure in that specific area because there will no longer be material there. It will also serve to add pressure in other areas. Typically the contour of the socket is manipulated towards the top of the socket since cut outs cannot be made in the middle or at the bottom of the socket while maintaining suction. The materials can be manipulated throughout the socket. The socket can have rigid and soft regions in specific areas in order to add or relieve pressure. Our new sockets included manipulation of both contour and material. 2.9.1 New Socket 1 (NS1) The data acquired from the trials with the participant’s current prosthesis and the combination that was determined the best from the foot and liner testing showed that the pressure-sensitive regions had some of the highest pressures. Therefore, the goal was to decrease Figure 8: New Socket 1 (NS1) with void visible in the lower left area of the socket. pressure in the pressure-sensitive regions by transferring the pressure to the pressure-tolerant regions. In the first design, a void was created at the distal end of the limb (Figure 8) due to the high pressures that were consistently acquired in this region throughout the trials, which would give the limb more area to dissipate pressure. Also, the socket had both an inner flexible region made of thermaline soft and an outer region consisting of carbon fiber. 2.9.2 New Socket 2 (NS2) 9 The second socket was designed with the same goal as the first socket: to decrease the pressure in the pressure sensitive regions. However, different materials were used this time, with less rigid material at the top (See Appendix A1.5). The material in NS2 was not cut in the same manner as in NS1 in the fibula head, as the first socket had significantly increased the pressure on that region in an attempt to relieve some of the pressure on the distal end of the limb. The flexible material in this socket was proflex with silicone, which is much more flexible than the thermaline soft used in the first socket. This design also had an extra “cushion” layer of a urethane insert that would be laminated in the final design between the soft and rigid layers. 2.9.3 New Socket 3 (NS3) The third socket established the importance of the contour of the socket. It was constructed using the same material as was used in NS1, but a different liner was used. However, it was found that the urethane and the silicone liners were basically the same. In NS1 the rigid layer extends to almost the topmost point on the socket, but in NS3 the rigid layer stops much lower down the socket (See Appendix A1.6). III. Results Figure 9: A sample graph automatically generated by the excel template file shows pressure (kPa) over time (s) for each of the eight FSRs during a trial with the subject walking two mph. As shown in Figure 9, the devised pressure measurement system was able to record rapidly changing pressure values. The sampling rate of the LabVIEW program and the NI box was set at 1 kHz, 10 and the rise time of the FSRs is less than 3 microseconds. Thus, the pressure measurement system was able to acquire maximum forces in regions where the peak values lasted for only a few milliseconds. It was important to attain the true peaks without under-sampling, since the comparison of maximum recorded pressures was a key component for assessing the differences between the transtibial prosthetic Pressure (kPa) systems. 450 400 350 300 250 200 150 100 50 0 Urethane Silicone Thermoplastic Elastomer 3.1 Determination of the Optimal Combination of Foot and Liner The maximum pressures for each gait cycle were recorded for all 27 walking trials with the nine combinations of three feet and three liners. The averages of all the peak pressures for a Figure 10: Averages of Maximum Walking Pressures (kPa) for Each Liner. Error bars depict one standard deviation. Numeric values in Appendix Table A2.1 particular liner with the three different feet were compiled and compared. From Figure 10, it is evident that the thin thermoplastic elastomer liner caused significantly higher pressures in the distal tibia. The comparison showed that the urethane and silicone liners had similar values in each anatomical region, and that 400 the thermoplastic liner also The averages of all the peak pressures for a particular foot with the three different values were analyzed by the same Pressure (kPa) caused higher pressures to be recorded in the MPT. Flex-Foot & Multiaxial Ankle Sach Vertical Shank ESF 350 300 250 200 150 100 50 0 procedure. Based upon the standard deviations, there were no significant differences between the feet. The large standard Figure 11: Averages of Maximum Walking Pressures (kPa) for Each Foot. Error bars depict 1 standard deviation. Numeric values in Appendix Table A2.2 11 deviations displayed in Figure 11 were primarily due to variations caused by the three liners used with each foot. This can be confirmed by viewing the results in the Appendix for each of the nine combinations. Since the optimal combination of foot and liner would reduce pressures in the pressure-sensitive regions of the residual limb, the combination with the lowest total pressure for the distal end of the limb, fibula head, distal tibia, and tibial tubercle was chosen. The total pressures from the silicone liner and flex foot were lower than the total pressures with the other feet and liners. As persistent pressure is responsible for many of the problems such as rashes that can develop within a socket, the average pressures in each region were also analyzed. The results of the average pressures during walking were concordant with the maximum pressures results. Once again, the thermoplastic liner was associated with much higher pressures in the distal tibia and mid patella tendon. There were minimal differences noticed for the feet, since a majority of the differences were due to the selection of the liner. The average pressures were also much lower than the maximums in most regions. *Averages shown in bold; respective standard deviations shown in grey. 12 *Averages shown in bold; respective standard deviations shown in grey. Pressures were also recorded for each of the nine combinations with the subject standing stationary. The data (See Appendix Tables A2.11-A2.19) for trials with the subject standing on the normal leg with the prosthetic leg suspended and trials with the subject standing with the liner only show that there are near-zero pressures in these situations. The data (See Appendix Tables A2.11-A2.19) also show that the trials for the subject standing on both legs demonstrate very similar trends to the trials for the subjects standing on the prosthetic leg only. The primary difference between these two trials is that higher pressures were recorded in each respective region when the testing subject stood on the prosthetic leg only (See Appendix A2.3-A2.4) than when the subject stood on both legs (See Appendix A2.11A2.19). This was due to the prosthetic leg bearing more weight in the former situation. Figures 12 and 13 show that the distal end of the limb must accommodate much higher pressures than the other regions of the socket when the transtibial amputee is standing. Any new socket design would have to reduce the high pressure in this region. Once again, the thermoplastic elastomer liner was associated with higher pressures in the distal tibia, and there was more variation due to the type of liner used than the type of foot used. The results when standing and walking both showed that there were high pressures in the distal end of the limb and the distal tibia. The other two regions sensitive to pressure also had pressures high enough that they could be significantly reduced with a new socket design. The pressures in both standing and dynamic situations were normally lower with the urethane and silicone liners than with the thermoplastic elastomer liner. Ultimately, Table R4 was used to select the silicone liner and Flex-Foot to be used in the redesigned socket system. 13 600 Urethane Silicone 500 Thermoplastic Elastomer Pressure (kPa) 400 300 200 100 0 Figure 12: Average Pressures (kPa) for Each Liner When Standing on the Prosthetic Leg Only. Errors bars depict one standard deviation. Numeric values in Appendix Table A2.3 Figure 13: Average Pressures (kPa) for Each Foot When Standing on the Prosthetic Leg Only. Errors bars depict one standard deviation. Numeric values in Appendix Table A2.4 3.2 Testing of New Prosthetic Systems The test socket (SF) with 500 450 SF CN NS1 NS2 the silicone liner and the flexible NS3 foot with a multi-axial ankle was Pressure (kPa) 400 selected as a baseline to compare 350 300 improvements in force 250 redistribution for the newly 200 designed sockets. The 150 combination normally used (CN) 100 by the test subject was also used 50 as a baseline for measuring 0 Distal End of Limb Fibula Head Distal Tibia Tibial Tubercle Figure 14: Maximum pressures in regions sensitive to pressure when walking. Errors bars depict one standard deviation. improvement. These two sockets and the three new sockets (NS) were compared on the basis of measurements recorded with the subject standing on both legs and with the subject walking 2 mph. 14 The averages of the maximum pressures for each gait cycle were compiled in Appendix Table A2.20. Since the success of 200 SF CN NS1 NS2 the new sockets was NS3 180 dependent on reducing the Pressure (kPa) 160 pressure in the regions of the 140 residual limb that are 120 100 sensitive to pressure, these 80 regions were compared in 60 Figure 14. This figure 40 shows that the pressures in 20 the distal end of the limb and 0 Distal End of Limb Fibula Head Distal Tibia Tibial Tubercle the distal tibia were reduced dramatically by the voids in Figure 15: Average pressures in regions sensitive to pressure when walking. Errors bars depict one standard deviation. those regions of socket NS1. These decreases were determined to be significantly reduced by observing that the pressures for these regions for socket NS1 were many standard deviations away from the pressures for these regions in the baseline sockets SF and CN. Likewise, there was a detrimental significant increase in pressure for socket NS1 in the fibula head region and a non-significant increase in pressure in the tibial tubercle region. Pressures for sockets NS2 and NS3 were also reduced 450 in the sensitive regions. 400 However, many of these were noticed for the persistent average pressures when walking, but the respective average pressures were typically considerably lower than they were for the Pressure (kPa) significant. Identical trends CN NS1 NS2 NS3 350 beneficial pressure reductions were not SF 300 250 200 150 100 50 0 Distal End of Limb Fibula Head Distal Tibia Tibial Tubercle Figure 16: Average pressures in regions sensitive to pressure when standing on both legs. Errors bars depict one standard deviation. maximum pressure values. Both walking pressure metrics depict overall pressure reductions in sockets NS1-3. 15 The average pressure results for the pressure tolerant regions for both walking and standing trials were recorded in Appendix Tables A2.21 and A2.22, respectively. Figure 16 depicts the average pressures observed in only the pressure-sensitive Table 4: Sum of Recorded Pressures (kPa) in Sensitive Regions regions when the subject stood evenly on both legs. Similarly to the walking trials, socket NS1 had the Walking Maximums SF 916 CN 984 NS1 714 NS2 615 NS3 643 Walking Averages 416 422 321 274 283 Standing Averages 721 679 497 543 482 lowest average pressure of all the sockets for the distal end of the limb, but had the highest pressures of all the sockets for the fibula head and tibial tubercle. The only significant reductions in pressure were for sockets NS1 and NS3 in the distal end of the limb. However, Table 4 demonstrates that the sum of the pressures in the sensitive regions decreased between the baseline sockets (SF & CN) and the new Table 5: Percentage of Total Recorded Pressure Located in Tolerant Regions sockets. This reduction was consistent for the peak pressures while walking, the average pressures while walking, and the average pressures while standing. The complete table for both pressure-sensitive and pressure-tolerant regions from which Tables 4 and 5 are extracted can be found in Appendix Table A2.23. Walking Walking Standing Maximums Averages Averages SF 31.8 32.8 25.7 CN 31.0 32.8 27.7 NS1 37.3 40.7 40.4 NS2 36.8 40.9 33.2 NS3 36.6 38.1 32.8 Table 5 shows that the proportion of the total recorded pressure in pressure-tolerant regions increased for sockets NS1-3. This indicates that a portion of the pressure was redistributed from sensitive to tolerant regions. The Socket Comfort Score (SCS) is a number that the patient assigns to the socket based upon his Table 6: Socket Comfort Score Socket SF CN NS1 NS2 NS3 Comfort Score 5.0 7.0 6.5 7.0 8.0 *Scale of 0 being most uncomfortable to 10 being completely comfortable level of comfort from 0-10, with 0 being most uncomfortable and 10 being completely comfortable. It has been shown that the reported SCS is a consistent and reliable assessment method. (Hanspal, Fisher & Nieveen, 2003). Table 6 shows that the participant found one of the new sockets (NS 3) to be more comfortable than both his current socket and the best foot/liner combination socket. NS2 was rated better than the best foot/liner combination socket and comparable to the patient’s current socket. NS3 was rated more comfortable than the best foot/liner combination but worse than the patient’s current socket. 16 3.3 Safety Considerations When developing the force measurement system, design considerations were made to reduce the risk of electrical shock to the testing subject via the powered FSRs. The power to the measurement circuit containing the FSRs was limited to ± 12V with a power supply. The voltage from the measurement circuit to the FSRs was limited further to ±5V by a voltage regulator. The test subject was also protected from this minimal voltage by using insulated RoHS-compliant wires to connect the FSRs to the measurement circuit. The conductive connection between the leads of the FSRs and readout wires was coated in a strong, non-conductive marine epoxy. The testing subject was protected from skin irritation by ensuring that all items placed between the subject’s skin and the liner were very thin. It was affirmed that the subject did feel pressure from the wires or the FSRs. The connection between the FSRs and the wires were made with smooth layers of epoxy rather than the standard bulky lead to wire connectors. Scotch tape was used to affix the FSRs to locations on the skin rather than using other adhesives that could be irritants. The subject was also protected from falling by ensuring that there were always chairs or handrails nearby for support and by holding the FSR wires perpendicular to the direction of walking when using the treadmill. The prosthetic system used fabrication materials and methods commonly used in the prosthetic industry. This ensured that all of the feet, liners, and sockets were made from biocompatible materials. These materials have proven material strength and durability properties that reduce the risk of breakage. 3.4 Costs The new socket design is intended for transtibial amputees and will be able to be used by many patients. Even though testing was performed on one individual, the advantages of the design can easily be altered to apply the overall fundamentals and ideas of the design to several other patients. As current socket design is based on a qualitative process, this design will have a significant time- and cost-benefit for The Surgical Clinic. The main materials needed for the project were FSRs, circuit components, and materials to fabricate each socket. This budget reflects the production of three transtibial sockets. This does not reflect the cost of the sockets that were already designed, the feet, or the liners that were already available at the Surgical Clinic prior to the start of our project. Appendix A2.27 shows that the total cost involved in our project was approximately $900. The typical lifespan of the socket is approximately one year. The main variables that affect this are the age of the patient, activity level of the patient, type of skin and bony prominences, and diseases such as diabetes, Peripheral Vascular Disease (PVD), and arthritis. In general if a patient is healthier and more active, his/her socket will demonstrate wear sooner. The older sicker patients can have sockets lasting five years (A. Fitzsimmons, personal communication, April 22, 2011). 17 Our research could increase the lifespan of a socket for a young, active patient by double or triple by identifying problem regions before the entire fabrication process and accommodating for them to make the socket more comfortable for a longer period of time and decrease associated problems, such as skin breakdown. The area this research will affect the most is cost maintenance, as the lifespan of the socket is not as important as maintenance. When a socket is uncomfortable, it is necessary for the patient to go back to the Clinic and have adjustments made. While no new materials are necessary for maintenance, there is a significant time cost in the form of gas money, time off work, and lost time. If an initial appropriate fit with definable parameters can be established for each patient, the cost maintenance can be reduced, which will prove beneficial for both the patient and The Surgical Clinic. IV. Conclusions Our final designs were successful in achieving all three of the original objectives. The FSRs, circuit, and NI box proved to be an accurate measurement system to acquire pressures on the residual limb of an amputee during standing and walking conditions. After testing nine combinations of standard liners and feet used in the prosthetic industry today, it was determined that the silicone liner and the flex foot reduced the pressures in the pressure-sensitive regions and the total pressure experienced on the residual limb compared to the other feet and liners. After this was determined, three new sockets were designed in an attempt to account for the shortcomings of the silicone liner, Flex foot, and socket combination and the participant’s current liner/foot/combination. While the study did not culminate in the design of one single socket, the three final sockets that were designed collectively established attributes that would be beneficial in the design of future sockets. In accordance with its design, New Socket 1, the socket made of thermaline soft and carbon fiber with a void at the distal end, most significantly reduced the pressure on the distal end of the limb. The more flexible proflex with silicone material and unique cut of New Socket 2 allowed for greater reduction in three of the pressure-sensitive regions of the limb but not the distal end of the limb. While New Socket 3, the socket made of the same material as New Socket 1 but with a different contour, did not reduce pressures in three of the pressure-sensitive regions quite as much as New Socket 2 did, it demonstrated the significance of contour when compared to the pressures generated from using New Socket 1. Thus, the desirable characteristics of each socket can be taken into consideration in the design of a new socket. For example, the material used to construct New Socket 2 can be used with the contour of New Socket 3 with a void as was used in Socket A to decrease the pressure in targeted regions. Additionally, as the silicone gel liner and flex foot were determined to be most effective in reducing pressure on the various regions of the residual limb, this liner-foot combination can be incorporated into future socket systems. 18 The Socket Comfort Score (SCS) is a number that the patient assigns to the socket based upon his level of comfort from 0-10 with 0 being most uncomfortable and 10 being completely comfortable. It has been shown that the reported SCS was consistent and reliable assessment method. (Hanspal, Fisher & Nieveen, 2003). It was found that both new socket 2 and new socket 3 had better or comparable Socket Comfort Scores than both the patient’s current socket and the best foot and liner combination (silicone liner with flex foot). New socket 1 had a better SCS than the best foot and liner combination (silicone liner with flex foot), but a worse SCS than the patient’s current socket. It has been found that the new socket designs have been successful in increasing patient comfort in addition to decreasing the pressure experienced in the pressure sensitive regions and distributing the pressures better throughout the socket. V. Recommendations 5.1 Future Work This study lends itself well to future work in socket design because of the accurate pressure measurement system in place and the success of the new sockets in various aspects. Because of the success of new socket 1’s ability to reduce the pressure at the distal end and the success of changing the materials in new socket 2, future studies should look to incorporate the elements of successful redesign of the sockets in this study. By using proflex with silicone and creating additional voids, an even greater decrease in pressure should be observed. Testing should be expanded to include additional patients. This should include patients of varying ages and activity levels in order to be able to apply the results to a larger general population. Testing situations should also include additional activities such as walking up and down stairs and ramps. In addition, assessment of the comfort levels of the new sockets should occur after the patient has used the socket for a duration of several days. Another variable that can affect the pressures on the residual limb is the suspension system used, so further tests can analyze the difference in pressures between different suspension systems. 5.2 Ethical and Societal Issues All of our testing included the use of a human subject. This subject volunteered to assist in our project and was never pressured into testing. The purpose of our project and the tasks required of him were explained. Throughout testing the safety and comfort of the subject was our primary concern and efforts were made to ensure that the subject was not fatigued or in any type of pain. 19 References DeLisa, J. (2005). Upper and Lower Extremity Prosthetics. Physical medicine and rehabilitation: principles and practice (4 ed., p. 1335). Philadelphia: Lippincott Williams & Wilkins. Dou, P., Jia, X., Suo, S., Wang, R., & Zhang, M. (2006). Pressure distribution at the stump/socket interface in transtibial amputees during walking on stairs, slope and nonflat road. Clinical Biomechanics, 21, 1067-1073. FSR Integration Guide. (2010, September 23). FSR Integration Guide. Retrieved March 1, 2011, from interlinkelectronics.com/sites/default/files/94-00004A_FSR_Integration_Guide.pdf K levels / For patients with a trans-tibial amputation. (2005, September 21). Home - Yanke Bionics. Retrieved April 24, 2011, from http://www.yankebionics.com/patients_and_caregivers/what_to_expect/for_patients_wit h_a_trans_tibial_amputation/k_levels Hanspal, R. S., Fisher, K. & Nieveen, R. (2003). Prosthetic socket fit comfort score. Disability and Rehabilitation, 25 (22), 1278-1280. Mak, A., Zhang, M., & Boone, D. (2001). State-of-the-art research in lower-limb prosthetic biomechanics-socket interface: A review. Journal Of Rehabilitation Research And Development, 38(2), 161-174. Polliack, A., Sieh, R., Craig, D., Landsberger, S., McNeil, D., & Ayyappa, E. (2000). Scientific validation of two commercial pressure sensor systems for prosthetic socket fit. Prosthetics and Orthotics International, 24, 63-73. Smith, D. G., Michael, J. W., & Bowker, J. H. (2004). Atlas of amputations and limb deficiencies: surgical, prosthetic, and rehabilitation principles (3rd ed.). Rosemont, IL: American Academy of Orthopaedic Surgeons. 20