1 Optimization of the Laser-Induced Thermotherapy Procedure for Liver Tumors 12/21/2010 Intro to Biofluids CRN 42697 Alli Dickey and Kelli Martino 2 Executive Summary Cancer is notorious for its destructive effect on the human body, capable of spreading uncontrollably from one area to another. Several treatments are available to reduce or eliminate tumor growth. Laser interstitial therapy (LITT) destroys tumor tissue by heating until the blood vessels within the tumor coagulate or clot the vessels in the tumor [A], stopping blood supply. Although laser interstitial thermotherapy effectively coagulates tumor tissue, it possesses potential for improvement of precision. By creating a robust model of the LITT process using COMSOL Multiphysics, the optimization of input parameters for the LITT procedure can be performed without difficulty and expensive in vivo experiments. The model consists of a 2cm spherical tumor [1] with a rectangular laser applicator assumed to be with a width of 0.05cm and a length of 1cm [2]. This model uses a temperature range of 333.15K (60°C) to 363.15K (90°C) [3, 4, 5] and a time domain of 120s (2 min) to 3300s (55min) [2]. Using these ranges, the optimal value of coagulated tumor tissue to damaged healthy tissue is 1cm to 0.033cm from the center of the tumor at 355K and 120s. Introduction Metastatic cancer, cancer that starts in one tissue then spreads to other areas of the body can contribute to malignant tumor growth [1]. Despite the liver’s ability to regenerate, a liver weakened by cancer poses a serious health risk to the patient because it is no longer able to perform vital functions such as toxin removal. Careful ablation of the tumor is necessary to maximize the destruction of tumor tissue while preserving as much of the healthy tissue as possible. Surgical removal of malignant tissue is a favored option, but only approximately 20% of patients are capable of receiving this treatment, leading to exploration of other methods [6]. LITT is an effective alternative in controlling tumor growth from liver metastatic cancer from the breast and colon [7, 2]. The problem schematic, set up, and construction of the model can be found in the Design Objectives. Following the description of the model the results of a parametric study are presented in Results and Discussion. Finally, conclusions and design recommendations are made. Design Objectives Modeling LITT on a tumor allows optimization of the duration of the procedure and temperature of the applicator edge that will maximize the ablated tumor tissue but minimize healthy tissue damage. Time and temperature of the applicator are optimized through a parametric study. Parametric studies allow the relationship between an output and input to be analyzed by isolating and varying each input and evaluating the output. This optimization is quantified by minimizing the amount of healthy tissue damage when the temperature at the edge of the tumor (r = 0.01000m) in the vertical direction is the temperature that coagulation occurs (323.15K or higher). The coagulated tumor tissue is when the distance from the center of the tumor (r) is less 3 than 0.01000, and has a temperature greater than 323.15K (50°C). Damaged healthy tissue is defined as the tissue with a temperature (T) between 323.15K (50°C) and 316.15K (43°C) when the distance from the center of the tumor (r) is greater than 0.01000m. If healthy tissue (r > 0.01000m) is coagulated (T > 323.15K) the time/temperature combination is considered undesirable. During optimization, the geometry, other boundary conditions, and other input parameters remain constant. These input parameters consist of the initial temperature of the liver and tumor at 310.15K, which is the average temperature of the body [8]. Also the heat flux is zero on the outside boundaries of the liver and continuity is assumed for the internal boundaries such as the tumor/tissue boundary. The applicator temperature is varied from 333 to 363K [3, 4, π 5], but the heat source is determined arbitrarily to be 5000π3 . Time is another varied parameter with a range of 120 (2min) to 3,300 (55min) seconds [2]. All boundary conditions and input parameters can be found in Appendix A. These conditions were used for the 2-dimensional model seen in Figure 1to simulate the LITT treatment. Blood Vessels Figure 1. A 2-dimensional Model Depicting the Subdomains Affected by the Heat from the Applicator A vertical slice of the liver is used as the geometry for the model. The minor axis of the liver slice is measured in the horizontal direction of the chest to back at 0.105m, while the major axis is measured in the vertical direction from head to feet at 0.125m [M]. For simplicity, the tumor was assumed to be in the approximate center of the liver with respect to the sagittal plane (as seen in Figure 2). This assumption removes the need to consider the temperature distribution 4 outside of the liver. If the tumor is assumed to be near edge of the liver, surrounding organs, arteries, and veins would have to be considered. Sagittal Plane Approximate location of tumor model Figure 2. Location of Tumor in Abdomen. [1] Since tumors with diameters of 2cm or less are best for LITT [1], the tumor is assumed to be spherical with a diameter of 2cm. Table 1 shows that a 2cm tumor can be found in 3 stages of liver cancer [1]. Table 1. Stages of Cancer that Have 2cm Tumors [1] Cancer Stage Number Diameter of Tumor(s) 1a, 1b 2 Single Tumor Any Size Single Tumor any Size or Multiple Tumors No Larger Than 5cm in Diameter Multiple Tumors with at least One that is Greater Than 5cm Across 3a The tumor size is directly related to the size of the laser applicator (shown in Figure 1) [2]. Laser applicators vary in length from 1cm to 3cm, [2] so a length of 1cm is assumed. The width is arbitrarily chosen as 0.5 cm because the laser applicators are usually thin. Results and Discussion A sensitivity analysis and an accuracy comparison with a published model determined the model to be credible. Inconsistency of values of input parameters such as density, thermal conductivity, and specific heat from a literature search proved the need to confirm insensitivity. A sensitivity analysis assessed the parameters shown in Tables 2 and 3. 5 Table 2. Properties and Their Ranges for a Sensitivity Analysis. Material Blood Tumor Liver Applicator ρ (kg/m^3 ) 1000-1200 1040-1060 1040-1060 2100-2300 Properties [8] k (W/mK) N/A 0.47-0.58 [10] 0.5122-0.5737 [8] 1.25-1.45 C_p (J/kgK) 3200-3400 3500-3890 3500-3890 650-750 The given ranges for the density and specific heat for the tumor and liver were chosen based on the densities and specific heats of other organs. The thermal conductivity values of the liver and tumor were maintained from the stated sources. The values for blood and the laser applicator properties were chosen arbitrarily. The ranges for the heat source and blood perfusion rate were also chosen arbitrarily and can be seen in Table 3. Table 3. Properties and Their Ranges for a Sensitivity Analysis Parameter Range Laser Source Term(W/m^2) 5000-9000000 V_b (ml_blood/(ml_tissue*s)) 0.0002-0.0006 After analyzing the minimum and maximum values for each input parameter, all of the parameters except for the heat source and tumor density proved to have extremely little (0.5% or less) or no effect on the solution. The maximum and minimum temperatures did not vary at all when changing the specified variables, except for the 18K increase in initial temperature from the increase in heat source. The temperature distribution only varied by a maximum of 5.92% and 1.973% for the differences in the range of the heat source and tumor density respectively. The laser heat source term proved to have a large effect on the maximum temperature, but the laser source term only had a 2.714% or 6.18K difference in temperature distribution. This could be the reason the COMSOL model is about 6K off of the published model because they do not state what value of heat source they use. Since the temperature distribution was the main concern, the differences of up to 5.92% proved the difference in parameters values do not significantly change the solution. The x and y direction were also evaluated to determine the direction that was most sensitive to the laser applicator. Since the edge of the tumor in the ydirection is closer the laser applicator, the edge of the tumor in the y-direction reached the temperature of coagulation before the edge of the tumor in the x-direction. Because the coagulation or damage of healthy tissue is undesirable, the y-direction was considered before the x-direction for evaluation of the parametric study. These results can be seen in Appendix C. A published model [3] was also used to verify the validity of this model. The COMSOL model was altered to have the input parameters and geometry as the published model for a more accurate comparison. The temperature distribution was compared using a contour plot of the temperature distribution, which proved the COMSOL model to be accurate within +/- 6 degrees. 6 These results can be seen in Appendix C. This difference is reasonable because the COMSOL model does not take radiation and varying paramters like the other model. After confidence was gained in the COMSOL model, the parametric study was performed with the parameters and ranges shown in Table 4. Table 4. Parameters for Parametric Study with Ranges Parameter Ti of Applicator (K) Time Domain Range 333-363 [6,7,8] 2min-55min[1] This time range was evaluated at the minimum and maximum temperatures to determine the amount of healthy tissue damage at a radius of coagulation of 0.01000m. The radius of coagulation was determined by using graphical data similar to Figure 3 to evaluate where the temperature was 323.15K. When the temperature of coagulation reached 0.01000m, the optimal value for the specific set of parameters was achieved. The healthy tissue damage was then determined by finding the distance from the center in the y-direction where the temperature is 316.15K (the temperature where tissue damage begins). This evaluation was used to evaluate both time and temperature and is shown in Figure 3 with the optimal results. Healthy Tissue Damaged Above this line (T>323.15) Tissue is Coagulated Above this line (T>316.15) Tissue is Damaged Figure 3. Temperature Distribution from Center of Tumor in Y-direction Using the Optimal Parameters The model was first tested at the minimum temperature for the minimum and maximum times. If neither the minimum nor maximum time produced the desired radius of coagulation, 0.01000m, intuition of a time that would lead to that value was tested. If the estimated time did not result in the desired amount of coagulated tissue, a revised estimation was made based on the results. This process continued until a value of 0.01000m of coagulated tissue was determined. 7 This resulted in the radius of coagulation reaching 0.01000m (the edge of the tumor in the ydirection) after 2050 seconds for the minimum temperature. This was determined to be the new maximum amount of time necessary to coagulate the tumor in the y-direction because a time of above 2050s would result in coagulation of healthy tissue. This procedure was also used for evaluation of the time domain at the maximum temperature, which resulted in the coagulation of the tumor before the minimum time of 120seconds. Therefore, the temperature of 363.15K had to be reduced to determine the temperature where coagulation reaches 0.01000m. To evaluate the maximum temperature, temperature was then varied at 120s. This resulted in a maximum temperature of 355K instead of 363K because anything larger than 355K would lead to coagulation of healthy tissue. Temperature was further optimized using the same method for time. After testing different time/temperature combinations, the optimal parameters were decided by which minimized the amount of healthy tissue damage. The optimal settings were then determined to be 355K at 120s, which only damaged 0.00330m of healthy tissue. With these values, the amount of tumor coagulated was 0.0100m in the y-direction and 0.00868m in the xdirection. There was about 0.00330m of damaged healthy tissue in the y-direction and 0.0020m in the x-direction. The temperature distribution for the optimal settings along the y-axis can be seen in Figure 3, and the overall temperature distribution can be seen in Figure 4. Max =355K Y - Direction Tumor Laser Applicator Arrows indicate Temperature Decrease decrease Min =310.15K Figure 4. Temperature Distribution Throughout the Tumor and Liver Tissue Conclusions and Design Recommendations A finite element analysis was performed with COMSOL to determine the optimal time and temperature. The evaluated ranges consisted of 120s to 3300s [2] for time and 333K to 363K [3,4,5] for temperature. The optimal coagulated tissue to damaged healthy tissue ratio was 8 approximately 1cm of tumor killed to 0.033cm of healthy tissue damaged at 355K (82°C) and 120s. An important aspect to be evaluated is the transient heat transfer during the beginning and end of the procedure. Since the laser cannot rapidly cool down to body temperature (315.15 or 37°C) or rapidly heat to 355K, the transient heat transfer could affect the optimal parameters. This heat transfer could be altered by changing the size and shape of the laser applicator. Other complexities could be added to the model. Instead of assuming the tumor to be in the center, the tumor could be assumed to be near the edge of the liver. The heat transfer analysis would then require the assessment of the temperature distribution to surrounding organs, arteries, and veins. Heat due to blood perfusion brings the model closer to a realistic situation. Despite blood perfusion larger arteries exist near the liver such as the hepatic portal vein and hepatic artery. A large artery or vein could introduce connective heat flux that could affect the temperature distribution of the tumor. 9 Appendix A. Mathematical Model Geometry Laser interstitial thermotherapy utilizes heat transfer to coagulate cells. The geometry can be simplified to a 2-dimensional axisymmetric domain. Figure A1. Model Schematic. A vertical slice of the liver is used as the geometry for the model. The minor axis of the liver slice is measured in the horizontal direction of the chest to back at 0.105m, while the major axis is measured in the vertical direction from head to feet at 0.125m. A spherical tumor is assumed with a diameter of 0.02m. The applicator dimensions are 0.01000m in height and approximately 0.005m in width. Governing Equations Although the human body functions off of different types of physics, the model represents heat transfer. The Cartesian form of the Energy Conservation Equation is used in the model. ππ ππ‘ π2π π π2 π 1 = ππΆ [ππ₯ 2 + ππ¦ 2 ] + ππΆ π π π ππ where π [π3 ] = density of the tissue π½ πΆπ [ππβπΎ] = specific heat capacity of the tissue (1) 10 T [K] = temperature t [s] = time π k [πβπΎ] = thermal conductivity of the tissue π Q [π3 ] = heat generation To make the model more realistic, blood perfusion is added to the physics. The heat generated by blood perfusion added to the heat source equals the heat generation within the domain. Therefore the heat source and blood perfusion must be defined. πβπππ‘ π ππ’πππ = πΌπΌ0 π −πΌπ₯ ππππππ πππππ’π πππ = ππ πΆππ ππ (ππ − π) 1 where πΌ [π] = specific absorption rate of the laser π πΌ0 [π2 ] = laser power intensity ππ ππ [π3 ] = density of blood π½ πΆππ [ππβπΎ] = specific heat of blood π½ ππ [π ] = dermal blood perfusion rate ππ [K] = arterial blood temperature π [K] = temperature of tissue Initial Conditions Human body temperature self regulates and is normally at 310.15K. Initially, all tissues including blood can be assumed to start at this temperature. πππππ‘πππ = 310.15πΎ, t = 0s Boundary Conditions Surrounding the liver is human tissues which are assumed to be at normal body temperature. Therefore, there is no heat flux at the outer edge of the liver. As a part of the optimization the temperature of the applicator edge is varied. πΜ |π₯=0.0525π,π¦=0.625π = 0 π π2 ππ = Temperature of Applicator Edge 11 Input Parameters The following parameters are needed for the COMSOL model. Constants π½ Cp_b = 3300 [ππβπΎ] T_a = 310.15 [K] ππ Rho_b = 110 [π3 ] 1 V_bp = 0.024/60 [π ] Subdomain Tumor π k = 0.566 [πβπΎ] ππ ρ = 1050 [π3 ] π½ Cp = 3800 [ππβπΎ] Applicator π k = 1.38 [πβπΎ] ππ ρ = 2203 [π3 ] π½ Cp = 703 [ππβπΎ] π Q = Heat Source [π3 ] Liver π k = 0.566 [πβπΎ] ππ ρ = 1050 [π3 ] π½ Cp = 3590 [ππβπΎ] π Q = Q_bp [π3 ] Boundary Applicator Edge Ti = Temperature of Applicator Edge [K] Tedge_liver = 310.15 [K] 12 Global Expressions π Q_h = 13000 [π3 ] π Q_bp = rho_b*Cp_b*V_bp*(T_a-T) [π3 ] Solver Parameters Times Range(120,10,3300) 13 Appendix B. Model Verification Liver Tumor Mesh The subdomains from smallest to largest are the applicator, the tumor, and the liver. Figure B1. Converged Mesh To reduce discretization error, a convergence study shows the mesh converges around 11,000 degrees of freedom. Several other meshes with less degrees of freedom were plotted to show the convergence shown in Figure B2. Convergence Study 336 334 Temperature 332 (K) 330 328 326 0 2000 4000 6000 8000 10000 12000 Degrees of Freedom Figure B2. Convergence Study Using Temperature at the Tumor Tissue and Healthy Tissue Boundary 14 Accuracy Check To verify the accuracy of our model it was compared to a published model. Comparing Figures B3 and B4 revealed that the model was fairly accurate in predicting the temperature distribution throughout the model domains. Geometries, boundary and initial conditions were replicated as well. Figure B3. Temperature Distribution Through the Tumor and Liver in the Math Model [3] Figure B4. Temperature Distribution Through the Tumor and Liver in LITT Model 15 Appendix C Part 1: Sensitivity Analysis Table C1. Table of Contents for Appendix C Part 1 and Results Material Applicator Blood Properties Liver Tumor Sensitivity analysis Max Diff. Property (%) ρ (kg/m^3 ) N/A C_p (J/kgK) N/A k (W/mK) N/A Q (W/m^2) 5.92% ρ (kg/m^3 ) N/A C_p (J/kgK) 0.410% V_b (ml_b/(ml_t*s)) N/A T_a [K] 0.349% ρ (kg/m^3 ) 0.37% C_p (J/kgK) N/A k (W/mK) N/A ρ (kg/m^3 ) 1.973% C_p (J/kgK) 0.41% k (W/mK) N/A Max Diff. (K) N/A N/A N/A 18.555 N/A 1.304 N/A 1.101 1.169 N/A N/A 6.187 1.311 N/A Fig. # C1 C2 C3 C4 C5 C6 C7 C8 C9 C10 C11 C12 C13 C14 Page # 15 16 16 17 17 18 18 19 19 20 20 21 21 22 *N/A is for no noticeable difference **All Results achieved with Specified Input Parameters and Boundary Conditions for a Time of 120s, an Applicator Ti of 333K, and Q=5000W/m^3 unless stated. Figure C1. Sensitivity Analysis for Applicator Density. 335 Temperature [K] 330 325 320 Applicator Rho max 315 Applicator Rho min 310 305 0.00E+00 2.00E-02 4.00E-02 6.00E-02 Distance from Center in y-direction [m] 8.00E-02 16 Figure C2. Sensitivity Analysis for Specific Heat for Applicator. 335 Temperature [K] 330 325 320 Applicator Cp Max 315 Applicator Cp min 310 305 0.00E+00 2.00E-02 4.00E-02 6.00E-02 8.00E-02 Distance from Center in y-direction [m] Figure C3. Sensitivity Analysis for Thermal Conductivity of Applicator. 335 330 325 320 Applicator k max Applicator k min 315 310 305 0.00E+00 2.00E-02 4.00E-02 6.00E-02 8.00E-02 17 Figure C4. Sensitivity Analysis for Heat Source of Applicator. 355 350 Temperature [K] 345 340 335 330 325 Q min 320 Q min 315 310 305 0.00E+00 2.00E-02 4.00E-02 6.00E-02 8.00E-02 Distance from Center in y-direction [m] The maximum heat source causes the initial temperature to increase by 18.55K, but the temperature at the edge of the applicator remains the same. Although the heat source appears to have a large effect, it only alters the values by a maximum of 5.92% or 6.813K around the edge of the tumor. To verify the optimal results, values were tested with the optimal heat source and still yielded the same optimal settings. Also the heat source is hard to very realistically, so it will be dependent on the temperature of the laser applicator. Figure C5. Sensitivity Analysis for Blood Density. 335 Temperature [K] 330 325 320 Blood Rho max 315 Blood Rho min 310 305 0.00E+00 2.00E-02 4.00E-02 6.00E-02 Displacement from Center in y-direction [m] 8.00E-02 18 Figure C6. Sensitivity Analysis for Specific Heat of Blood. 335 Temperature [K] 330 325 320 Blood Cp max Blood Cp Min 315 310 305 0.00E+001.00E-022.00E-023.00E-024.00E-025.00E-026.00E-027.00E-02 Distance from Center in y-direction [m] The differences in the specific heat of blood for blood perfusion still prove the specific heat does not significantly alter the solution. The sensitivity analysis determined that the difference in values for specific heat result in a 0.410% or 1.304K difference. Figure C7. Sensitivity Analysis for Blood Perfusion Rate. 335 Temperature [K] 330 325 320 V_b max 315 V_b min 310 305 0.00E+00 2.00E-02 4.00E-02 6.00E-02 Displacement from Center in y-direction [m] 8.00E-02 19 Figure C8. Sensitivity Analysis for Arterial Temperature. 335 Temperature [K] 330 325 320 T_a max 315 T_a min 310 305 0.00E+00 2.00E-02 4.00E-02 6.00E-02 8.00E-02 Distance from Center in y-direction [m] Although the arterial temperature results in a change in temperature, the 0.349% or 1.101K difference is not significant enough to rule the solution sensitive. Figure C9. Sensitivity Analysis for Liver Density. 335 Temperature [K] 330 325 320 Liver rho max Liver rho min 315 310 305 0.00E+00 2.00E-02 4.00E-02 6.00E-02 8.00E-02 Distance from Center in y-direction [m] Much like the sensitivity analysis for the arterial temperature, the temperature difference of 0.37% or 1.169K did not produce enough evidence to consider the solution sensitive. 20 Figure C10. Sensitivity Analysis for Specific Heat of Liver. 335 Temperature [K] 330 325 320 Liver Cp max Liver Cp min 315 310 305 0.00E+001.00E-022.00E-023.00E-024.00E-025.00E-026.00E-027.00E-02 Displacement from Center in y-direction [m] Figure C11. Sensitivity Analysis for Thermal Conductivity of Liver. 335 Temperatre [K] 330 325 320 Liver k max 315 Liver k min 310 305 0.00E+00 1.00E-02 2.00E-02 3.00E-02 4.00E-02 5.00E-02 6.00E-02 7.00E-02 Displacement from Center in y-direction [m] 21 Figure C12. Sensitivity Analysis for Tumor Density. 335 Temperature [K] 330 325 320 Tumor rho max 315 Tumor rho min 310 305 0.00E+00 2.00E-02 4.00E-02 6.00E-02 8.00E-02 Distance from Center in the y-direction [m] The tumor density was the second most influencing parameter with a difference of 1.973% of 6.187K, but this result is still not significant. Figure C13. Sensitivity Analysis for Specific Heat of Tumor. 335 Tempertaure[K] 330 325 320 Tumor Cp max 315 Tumor Cp min 310 305 0.00E+00 2.00E-02 4.00E-02 6.00E-02 8.00E-02 Displacement from Center in y-direction [m] Although the liver and tumor are assumed to the same specific heat, only the tumor displayed a difference in temperature. This difference was 0.41% or 1.311K, which is not enough to influence a solution. The specific heat of the tumor was a factor only because it the temperature is much larger in the tumor than the healthy tissue. 22 Figure C14. Sensitivity Analysis for Thermal Conductivity of the Tumor . 335 Temperature [K] 330 325 320 Tumor k max 315 Tumor K min 310 305 0.00E+00 2.00E-02 4.00E-02 6.00E-02 Distance from Center in y-direction [m] 8.00E-02 23 Appendix C Part 2: Parametric Study Results Lines highlighted in yellow denote the optimal values for that particular set of parameter values. Green denotes the overall optimal values. Table C2. Table of Contents for Appendix C Part 2 Q[W/m3] C16 C17 C18 Varying Parameter Ti of Applicator Ti of Applicator Ti of Applicator 9000000 9000000 5000 Ti of Applicator [K] 343.65-360.15 333.15-363.15 333.15-363.15 C19 Ti of Applicator 5000 333.15-363.15 120 C20 C21 C22 Ti of Applicator Ti of Applicator Ti of Applicator 9000000 5000 9000 333.15-355.06 333.15-355.15 333.15-355.15 120 1080 200 C23 Ti of Applicator 5000 333.15-355.15 125 Figure Time [s] Optimum Value 120 2050 2050 355.06[K] 333.15[K] 333.15[K] 355.00 & 355.15[K] 355.06[K] 333.15[K] 343.15[K] 354.15 & 355.15[K] Table C16. Damaged and Coagulated Tissue in the Y-direction for Maximum Heat Source and Time with Varying Temperature Find temp that can be used with Qmax at 120s Temp of Appl. (K) 360.15 356.81 355.06 353.27 343.65 Laser Source (Q=9E6 W/m^3, Time = 120s) Distance from Center in y dir (m) for Damaged Tissue Coagulated Tissue N/A 0.01048 N/A 0.01010 N/A 0.01000 N/A 0.00990 N/A 0.00929 Table C16 displays the maximum temperature that can be used for the maximum time and heat source in order to have the coagulated tissue stay within 0.0100m from the center of the tumor. Table C7. Damaged and Coagulated Tissue in the Y-direction for Maximum Heat Source and Maximum Time with Varying Temperature Ti of Applicator (K) 333.15 363.15 Laser Source (Time=2050s, Q=9e6W/m^3) Distance from Center in y dir (m) for Damaged Tissue Coagulated (316.15K) Tissue(323.25K) 0.01840 0.01000 0.02775 0.01900 24 Table C18. Damaged and Coagulated Tissue in the Y-direction for Minimum Heat Source and Maximum Time with Varying Temperature Ti of Applicator (K) 333.15 363.15 Laser Source (Time=2050s, Q=5000W/m^3) Distance from Center in y dir (m) for Damaged Tissue Coagulated (316.15K) Tissue(323.25K) 0.01850 0.01000 0.02800 0.01910 Table C19. Damaged and Coagulated Tissue in the Y-direction for Minimum Heat Source and Time with Varying Temperature with Optimum Paraeters Ti of Applicator (K) 333.15 354.15 354.85 355.00 355.15 355.50 360.15 363.15 Laser Source (Time=120s, Q=5000W/m^3) Distance from Center in y dir (m) for Damaged Tissue Coagulated (316.15K) Tissue(323.25K) 0.00776 0.00804 0.01320 0.00995 0.01328 0.00997 0.01329 0.00998 0.01330 0.01002 0.01332 0.01004 0.01360 0.01050 0.01377 0.01080 Table C20. Damaged and Coagulated Tissue in the Y-direction for Minimum Time and Maximum Heat Source with Varying Temperature Ti of Applicator (K) 333.15 355.06 Laser Source (Time=120s, Q=9e6 W/m^3) Distance from Center in y dir (m) for Damaged Tissue Coagulated (316.15K) Tissue(323.25K) 0.01042 0.00805 0.01326 0.01000 Table C13 and C14 show 0.0100m of the tumor tissue is coagulated around 333.15K even though different heat sources are used. Also Tables C15 and C16 show the same relationship with 0.0100m of the tumor tissue destroyed around 355K. 25 Table C21. Further Testing for Optimization of Damaged and Coagulated Tissue in the Y-direction Ti of Applicator (K) 333.15 355.15 Laser Source (Time=1080s, Q=5000W/m^3) Distance from Center in y dir (m) for Damaged Tissue Coagulated (316.15K) Tissue(323.25K) 0.01668 0.00969 0.02277 0.01564 The time of 1080s was chosen and evaluated because it was the average between 120s and 2050s. This time was determined to be insignificant because the damaged tissue was larger than the 0.00330m of damaged tissue when the time is 120s, the heat source is 5000 π π3 , and the temperature is 355K. Table C22. Further Testing for Optimization of Damaged and Coagulated Tissue in the Y-direction Try Time around 120s Ti of Applicator (K) 333.15 343.15 355.15 Laser Source (Time=200s, Q=9000W/m^3) Distance from Center in y dir (m) for Damaged Tissue Coagulated (316.15K) Tissue(323.25K) 0.01200 0.00841 0.01358 0.00981 0.01501 0.01143 Table C22 shows that a smaller time of 200s still does not conclude in less than 0.00330 of damaged healthy tissue. Table C23. Further Testing for Optimization of Damaged and Coagulated Tissue in the Y-direction Try Time closer 120s Ti of Applicator (K) 333.15 353.75 354.15 355.15 Laser Source (Time=125s, Q=5000W/m^3) Distance from Center in y dir (m) for Damaged Tissue Coagulated (316.15K) Tissue(323.25K) 0.01063 0.00871 0.01333 0.00997 0.01335 0.00999 0.01343 0.01005 Table C19 shows that 120s is the optimal time resulting in 0.01330m of healthy tissue being damaged, which is smaller than the 0.00335m or 0.00343m of healthy tissue damaged shown in the table above. 26 Appendix D: References [A] "infarction." Encyclopædia Britannica. 2010. Encyclopædia Britannica Online. 21 Dec. 2010 <http://www.britannica.com/EBchecked/topic/287454/infarction>. [1] Time, By This. "Pancreatic Cancer." American Cancer Society :: Information and Resources for Cancer: Breast, Colon, Prostate, Lung and Other Forms. Web. 05 Oct. 2010. <http://www.cancer.org/Cancer/PancreaticCancer/DetailedGuide/pancreatic-cancer>. [2] Mack, M. G., R. Straub, K. Eichler, O. Sollner, T. Lehnert, and T. J. Vogl. "Breast Cancer Metastases in Liver: Laser-induced Interstitial Thermotherapy--Local Tumor Control Rate and Survival Data." Radiology 233.2 (2004): 400-09. Print [3] Fasano, Antonio, Dietmar Homberg, and Dmitri Naumov. "On a Mathematical Model for Laser-induced Thermotherapy." Applied Mathematical Modeling 34 (2010): 3831-840. Print. [4] Merkle, Elmar M., John R. Haaga, Jeffery L. Duerk, Gretta H. Jacobs, Hans-Juergen Brambs, and Jonathan S. Lewin. "MR Imaging-Guided Radiofrequency Thermal Ablation of the Pancreas in a Porcine Model with a Modified Clinical C-Arm System." Radiology (1999): 461-67. Print. [5] Mueller-Lisse, Ullrich G., Martin Thoma, Sonja Faber, Andreas F. Heuck, Rolf Muschter, Peter Schneede, Ernst Weninger, Alfons G. Hofstetter, and Maximilian F. Resier. "Coagulative Interstitial Laser-induced Thermotherapy of Benign Prostatic Hyperplasia: Online Imaging with a T2-weighted Fast Spin-Echo MR Sequence— Experience in Six Patients." Radiology (1999): 373-79. Print. [6] Vogl, Thomas J., Katrin Eichler, Stefan Zangos, and Martin G. Mack. "Interstitial Laser Therapy of Liver Tumors." Medical Laser Application 20.2 (2005): 115-18. Print. [7] Vogl, T. J., R. Straub, K. Eichler, O. Sollner, and M. G. Mack. "Colorectal Carcinoma Metastases in Liver: Laser-induced Interstitial Thermotherapy--Local Tumor Control Rate and Survival Data." Radiology 230.2 (2004): 450-58. Print. [8] Datta, Ashim K., and Vineet Rakesh. An Introduction to Modeling of Transport Processes: Applications to Biomedical Systems. Cambridge, UK: Cambridge UP, 2010. Print. [9] ONISHI, NORIMITSU. "Japan, Seeking Trim Waists, Measures Millions." The New York Times. 13 June 2008. Web. 07 Oct. 2010. <http://www.nytimes.com/2008/06/13/world/asia/13fat.html>. [10] MuΜller, Gerhard J., and AndreΜ Roggan. Laser-induced Interstitial Thermotherapy. Bellingham, WA: SPIE Optical Engineering, 1995. Print. [M] Wolf, Douglas C. "Evaluation of the Size, Shape, and Consistency of the Liver -- Clinical Methods -- NCBI Bookshelf." National Center for Biotechnology Information. Butterworth 27 Publishers, 1990. Web. 16 Nov. 2010. <http://www.ncbi.nlm.nih.gov/bookshelf/br.fcgi?book=cm&part=A3049>.