A9800 Technical Specifications

advertisement
Technical Specifications - A9800 Anesthesia Workstation
Physical Specifications
Ventilator Operating
Dimensions
Specifications
Height
Width
Depth (without YOKE)
Weight (basic unit)
1410mm
920mm
760mm
Approximately 130kg
Modes of ventilation – standard
Standby, Manual, VCV;
Demo
Top Shelf
Weight limit
Width
Depth
Modes of ventilation – Options
25kg/55Ib
650mm
380mm
PCV,
SIMV-VC,
SIMV-PC,
PCV-VG,
PS(with Apnea backup)
Work surface
Height
Width
Depth
850mm
630mm
320mm
Ventilator parameter ranges
Tidal volume range
20 to 1500mL
(volume control modes)
5 to 1500mL (pressure
Flip-up side tray
Height
Weight limit
Width
Depth
control modes)
850mm
12kg
265mm
330mm
Incremental settings
increments of 5mL(Set
Vt below 100mL) or
10mL(when
Drawers (internal dimensions)
Quantity
Height
Width
Depth
between
3
130mm
490mm
350mm
100
Vt
and
1000mL) or 50mL(when
set
Vt
bigger
than
1000mL)
Pressure (Pmax) range
Casters
Diameter
Brakes
set
125mm
Individual locking front
casters
10 - 70 cmH2O
(volume control modes)
Pressure (ΔP) range
3 to 50 cmH2O
(increments of 1
Cylinder(optional)
Interface
Type
Number
cmH2O)
PISS
E
Maximum 3 cylinders
(SIMV-VC, SIMV-PC,
and PS modes)
Pressure (PTARGET)
5 to 70 cmH2O
range
(increments of 1
cmH2O)
(PCV and SIMV-PC
vent mode)
Page 1 of 9
Technical Specifications - A9800 Anesthesia Workstation
(increments of 1
cmH2O)
Rate
4 to 100 breaths per
(VCV, PCV, SIMV-VC,
minute for VCV, PCV,
SIMV-PC, PCV-VG and
PCV-VG;
PS vent modes)
2 to 100 breaths per
minute for SIMV-VC
Ventilator performance
and SIMV-PC;
Pressure range at inlet
2 to 60 breaths per
280kPa to 600kPa/
40.6psig to 87psig
minute for PS vent
mode (increments
Peak gas flow
of 1 breath per minute)
≥90 L/min + fresh gas
flow
Inspiratory/expiratory
4:1~1:8
Ventilator monitoring
ratio
(increments of 0.5)
Expiratory minute
(VCV, PCV and
volume range
0L~30L
PCV-VG vent modes)
Expiratory tidal volume
Inspiratory time
0.2s~5 seconds
0 mL~3000mL
range
(increments of 0.1
seconds)
O2%
18%~100%
SIMV-PC vent modes)
Peak pressure (Ppeak)
-20cmH2O~99cmH2O
OFF, 5%~60%
Mean pressure (Pmean)
-20cmH2O~99cmH2O
Plat pressure (Pplat)
-20cmH2O~99cmH2O
PEEP
-20cmH2O~
(SIMV-VC and
Inspiratory Pause Time
(VCV and SIMV-VC
vent modes)
Flow trigger
1 L/min~15L/min
(SIMV-VC,
SIMV-PC
99cmH2O
and PS vent modes)
Frequency
TSLOPE (Inspiratory
0s~2 seconds
Slope Time)
(increments of 0.1
0~110 breaths per
minute
seconds)
Trend chart
(PCV, SIMV-VC,
Continuous trend information together with time
SIMV-PC and PS vent
discrete events are stored and shown in the table,
modes)
including Ppeak, Pplat, Pmean, PEEP, Freq, VT,
MV, FiO2, etCO2, FiCO2, Agent1, Agent2, N2O,
Positive End Expiratory Pressure (PEEP)
MAC, FG-O2, FG-Air and FG-N2O. The left page
Type
Integrated,
shall include the 10 parameters and the remains
electronically controlled
shall be in the right page.
OFF, 3 to 30 cmH2O
Trend table
Range
Page 2 of 9
Technical Specifications - A9800 Anesthesia Workstation
The Setting shall include the patient type, the vent
mode and all the setting parameters.
The
Monitor
shall
include
all
Insp. HAL
the
Low: OFF, 0.1-8.3%
High: 0.1-8.4%
monitor
parameters in the trend data.
The machine shall remember the last 8 hours
Insp. ISO
High: 0.1-8.4%
trend data, and the interval shall be 30 seconds.
Delivery/monitoring accuracy
Volume delivery
Low: OFF, 0.1-8.3%
Insp. ENF
Low: OFF, 0.1-9.8%
High: 0.1-9.9%
< 100 mL = better than
10 mL
> 100 mL = better than
Insp. DES
High: 0.1-21.9%
10%
Pressure delivery
±10% or ±3cmH2O
Low: OFF, 0.1-21.8%
Insp. SEV
Low: OFF, 0.1-9.8%
High: 0.1-9.9%
PEEP delivery
±2cmH2O or ±10%
Apnea alarm
Volume monitoring
Mechanical ventilation
< 100 mL = better than
ON:
10 mL
Vt< 10 mL breath or
> 100 mL = better
Pmean<1 cmH2O
than10%
measured in 30
seconds when
Pressure monitoring
±10% or ±3cmH2O
Frequency≥6
Vt < 10 mL breath or
Alarm settings
Minute volume (MVexp)
Pmean<1 cmH2O
Low: 0 to 20 L/min
measured in 35
High:1 to 25 L/min
seconds when
Frequency<6
Low airway pressure
0~70cmH2O
Mechanical ventilation
OFF:
High pressure
10cmH2O~80cmH2O
< 10 mL breath
measured in 60
High Breath Rate
8-60bpm
seconds
Inspired oxygen (FiO2)
Low: 18 to 99%
Sustained airway
Mechanical ventilation
High: 21 to 100%
pressure
ON:
Paw>PEEP add 10
exhalant CO2 (etCO2)
Low: OFF, 0.1-9.8% or
cmH2O measured over
OFF, 1-74 mmHg
15 seconds
High: 0.1-9.9% or
Continuously
1-75mmHg
Mechanical ventilation
OFF:
Inspired CO2 (FiCO2)
High: 0.1-1.4% or
Paw>10 cmH2O
1-10 mmHg
measured over 15
Page 3 of 9
Technical Specifications - A9800 Anesthesia Workstation
seconds Continuously
Subatmospheric
Paw < -2 cmH2O
ExpAgent2)
Display graphics
pressure
Waveform of P-T, F-T,
V-T, CO2-T, Paw-V
Loop, V-Flow Loop
Alarm silence
120 to 0 seconds
countdown timer
Communication ports
RS-232 serial
port(DB9);
Ventilator components
RJ45 connector,
100-Base-TX, support
HL7 communication
Flow transducer
Type
protocol;
fixed orifice
USB.
(honeycomb)
Integrated safety functions
Location
Installed in breathing
In case of electricity and battery failure, manual
system
ventilation, gas delivery and agent delivery are
possible.
Oxygen Sensor
Positive pressure relief valve opens at about
Type
Galvanic fuel cell
Life Cycle
proximately 12 months
85cmH2O.
(Dependent on usage)
Anesthetic agent delivery
Color active matrix TFT
Delivery
with Touch screen
Vaporizer
A9040/A9050
15 inch diagonal
Type
Halothane, Enflurane,
Ventilator Screen
Display type
Display size
Isoflurane, Sevoflurane,
Pixel format
Display parameters
1024×768
Number of positions
2
All settings and alarm
Third parking position
Optional
parameters(including
Mounting
Selectatec○R manifold
VT, Freq., I:E, TINSP,
with interlocks
PEEP, FreqMIN, TP,
Trigger, PTARGET, ΔP,
TSLOPE, PEAK,
Gas monitor(optional)
MEAN,PLAT, FiO2,
FiCO2, etCO2, MAC,
Type
InspAgent1,
main stream / side
stream
ExpAgent1,
InspAgent2,
Module
Page 4 of 9
IRMA CO2; IRMA AX+
Technical Specifications - A9800 Anesthesia Workstation
Operating
Monitor Gas
CO2
Measurement range
0-15 vol%
ISA CO2: 0 to 50℃
Accuracy
0-15 vol%
ISA AX+: 5 to 50℃
(Standard conditions)
± (0.2 vol% + 2 % of
ISA CO2; ISA AX+
IRMA CO2: 0 to 40℃
IRMA AX+: 10 to 40℃
reading)
Storage
-20 to 70℃
Humidity
10 - 95%(operating,
Anesthesia Gas Module
Monitor Gases
non-condensing), 5 -
CO2,N2O,HAL,ISO,ENF
SEV;DES
100%(storage,
condensing)
Measurement range
CO2: 0-15 vol%
N2O: 0-100 vol%
Atm.pres.
525 - 1200 hPa(4572m)
HAL, ISO, ENF: 0-8
vol%
Calibration
Warm-up time
No routine calibration is
SEV: 0-10 vol%
required.
DES: 0-22 vol%
20 sec, full
Accuracy (Standard conditions)
specifications within 60s
CO2
0-10 vol% ±(0.2 vol%
+ 2 % of reading)
Rise Time
10-15 vol% ±(0.3 vol%
< 350 ms, typically
+ 2 % of reading)
Total system response
< 3 sec
time
N2O
±(2 vol% + 2 % of
reading)
Breath detect
Adaptive threshold,
minimum 1vol% change
HAL, ISO, ENF
in CO2 concentration.
Respiratory rate
0 - 150 bpm ± 1 bpm
±(0.15 vol% + 5 % of
reading)
SEV
±(0.15 vol% + 5 % of
reading)
Compensation
Automatic for
atmospheric pressure,
DES
temperature and
±(0.15 vol% + 5 % of
reading)
spectral interference
Airway adapters (main stream)
Adult/Pediatric
6 mL dead space
Infant
1 mL dead space
Electrical specifications
Power and battery backup
Power input
CO2 Module
100-240V, 50/60Hz,
Max. 10A
Page 5 of 9
Technical Specifications - A9800 Anesthesia Workstation
of N2O with reduction in
Battery type
Internal re-chargeable,
O2 pressure
sealed lead acid,
Backup power
24VDC, 4AH
Supply failure alarm
Range: 185 to 215 kPa
Demonstrated battery
O2 flush
Range: 25 to 75 L/min
backup time under
Charge time
typical operating
Flowmeter (Electronic)
conditions is 120
O2 ranges
0 to 10 L/min;
minutes when fully
N2O ranges
0 to 10 L/min
charged
Air ranges
0 to 12 L/min
Approximately
Integrated safety functions
8hours(in running status
O2 and N2O Ratio regulating valve
or standby mode)
guarantees a minimum O2 concentration of 21% in
an O2/N2O mixture.
Power cord
5m/16.4ft
N2O cut-off if O2 fresh gas valve is closed or if O2
flow is less than 0.2 L/min.
Outlets
4 outlets on back, 1.5A
Total flow tube
Range
Pneumatic specifications

Auxiliary common gas outlet(optional)
Connector
ISO
22mm
OD
and
0-15SLM
Air/O2 blender with flowmeter
Flow Range
1-15L/min
O2 concentration
21%-100%
Flow indicator
Flow tube
15mm ID
Auxiliary gas output(optional)
Gas supply
Gas type
Pipeline input range
O2, N2O, Air
Gas
Oxygen
Pressure
280-600kPa
Flow rate
Max.90L/min
280kPa to 600kPa
/41psi to 87psi
Breathing circuit
Pipeline connections
NIST/DISS
Cylinder input
O2, N2O, Air pin index
specifications
yokes (PISS)
Primary regulator
250kPa/36psig
CO2 absorbent canister
nominal output
Absorbent capacity
O2 controls
Ports and connectors
Method
Proportionate decrease
Exhalation
Page 6 of 9
3000mL
22mm OD ISO 15mm
Technical Specifications - A9800 Anesthesia Workstation
ID taper
Circuit volume
6.9 L Vent Mode
(including absorber and
Inhalation
Bag port
22mm OD ISO 15mm
bellows)
ID taper
5.4 L Bag Mode
22mm OD
Expiratory resistance
0.57 kPa
under manual condition
Airway pressure gauge
Scale range
–20 to 100 cmH2O
Inspiratory resistance
Bag-to-Ventilator switch
Type
0.36 kPa
under manual condition
Manual lever with signal
feedback
Expiratory resistance
0.47 kPa
under automatic
Control
Controls ventilator and
direction of breathing
Inspiratory resistance
gas within the circuit
under automatic
0.35 kPa
Integrated Adjustable Pressure Limit
(APL) valve
Note: According to ISO8835-2, test under peak
Range
0 to 70 cmH2O
With patient circuit and wye piece added
Tactile knob
10 cmH2O and above
Heating system(optional)
indication at
flow 60L/min, fresh gas 10L/min.
Temperature
32-40℃
fmainnmainmainMaintaine
Adjustment range of
0 to 30 cmH2O (0 to
rotation
180°)
30 to 70 cmH2O (180 to
288°)
dMaintainedMaintainedMainta
Materials
ined Controlled to 31
All materials in contact with exhaled patient gases
– 35°C
are autoclaveable, except mechanical pressure
gauge and O2 cell.
Accuracy
<30 cmH2O:±3
All materials in contact with patient gas are free of
cmH2O;
natural rubber latex.
≥30 cmH2O:±15% of
set value;
Anesthetic gas scavenging
Breathing circuit parameters
Compliance in Bag
3.15 mL /cmH2O
mode
Mechanical
Automatically
ventilation mode
compensates for
System(AGSS)
Size
445×142×95 (height x
width x depth)
compression losses
within the absorber and
Weight
2.25Kg
Type of disposal
Low-flow disposal
bellows assembly
Page 7 of 9
Technical Specifications - A9800 Anesthesia Workstation
system
system
(non-condensing)
Extract flow
35L/min~50L/min
Barometric
Pressure relief device
Pressure compensation
System storage
opening to the
Temperature
–10 to 60℃
Relative Humidity
15 to 90%
53 to 106kPa
atmosphere
Filter
Stainless screen with
( non-condensing)
hole diameter of 150μm
Barometric
Spillage
<100mL/min
Electromagnetic compatibility
Immunity
Max. constant flow
50 to 106kPa
50L/min
Complies with all
requirements
of EN 60601-1-2
Max. intermittent flow
35L/min
Emissions
CISPR 11 group 1 class
B
Directive 93/42/EEC
Environmental
specifications
System operation
Temperature
10 to 40℃
Relative Humidity
15 to 90%
CE mark in this manua
product with CE mark.
Concerning Medical Devices
Concerning Medical Devices
Oricare Inc.
Address: 1900 AM Drive
Quakertown, PA 18951 USA
Tel:
+1-215-538-2470
Fax:
+1-215-538-2578
E-mail:
info@oricaremed.com
Page 8 of 9
Technical Specifications - A9800 Anesthesia Workstation
Page 9 of 9
Download