Joey’s Sputtering SOP. Updated: 5 May 2005 by William R. Evans Important: Please READ through all of the instructions on this and all other SOP’s which you are planning to use before you do anything. If you have not already done so: Record your name(s), the date, the material being Figure 0: The Run Book and Run Sheets. sputtered, and substrate(s) on the run sheet. [See Fig. 0] Once you have reached as low a pressure as you can with just the turbo and roughing pumps: 6. Open the cryo pump. To open cryo pump: a. Make sure air valve is open. (The air valve is open when the black handle is parallel to the tube. It is closed when the black handle is perpendicular to the tube.) b. Turn on pneumatic switch (light switch-looking thing). [See Fig. 4] (Cryo pump gate will open.) Figure 4: Air Valve and Pneumatic Switch Figure 5: Cryo Pump c. Close air valve. Gate Valve Handle d. Turn off pneumatic switch. (The cryo pump gate valve handle will go slack.) [See Fig. 5] After you have reached a base pressure, record what the base pressure was on the run sheet. 1. Check to see that all plugs and connections are secure and safe (including RF connection to the chamber). (Usually these are just fine.) 2. Turn on coolant water in back. (2 valves) (Usually this is off.) [See Fig. 1] Figure 1 3. Turn on the power switch in the back of panel APCS-3. [See Fig. 2] Figure 2 4. Turn on the following switches: [See Fig. 3] On CEX-2: a. “Power” Switch On AMNPS-1: b. “Power On” Button On APCS-3: (Nothing) On the Loading/Tuning Box: (Nothing) Bottom Panel: d. “Control” Switch When you turn the “Control” Switch on, the buttons on Source A will light up. e. “Line” Switch The “Line” Switch is behind the metal cover. On Source A: c. “Power On” Button (White Button) (This button should already be on, since you turned on the switches on the Figure 1 Bottom Panel. The Blue “RF Off” Button should also be already lit up.) (In other words, turn on all the power switches on the front except the HV Power Switch, the Auxiliary Control Switch, the RF On Switch, and the Incident Control Switch.) Make sure the switches are set to their normal configurations: On APCS-3: a. First switch (from the left) set to Manual Power Control b. Second switch set to Local Control On the Loading/Tuning Box: c. Loading to Manual d. Tuning to Manual On Black Box Below Source A: e. Multimeter set to “PA Plate Current 1 A” f. PA Tuning set to about 65 Figure 3 (again) 5. Wait 2-3 minutes for the power source to warm up. 7. Close the cryo pump gate valve until it is ¼ open. To open cryo pump gate valve to ¼ open: a. Make sure air valve is open. b. Turn on pneumatic switch (light switch-looking thing). (Cryo pump gate will open.) c. Close air valve. d. Turn off pneumatic switch. (Cryo pump gate handle will go slack.) e. Use the cryo pump gate valve handle to close the gate valve until it is ¼ open. If you want you can wedge the gear in place with a wrench. 13. If making a metal oxide or a metal nitride, add oxygen or nitrogen. The main valve for the oxygen / nitrogen line is the “leak” valve on the chamber, above the cryo pump. [See Fig. 10] If you are using the MPA, open the leak valve until you reach the desired gas partial pressure. (The correct partial pressure will depend on the material you are sputtering. If you don’t know, ask someone.) Note: The MPA is a device which connects to the computer and measures the partial pressures of the different types of gasses in the chamber, where the other pressure gauges just measure the total pressure. [See Fig. 11] How to use the MPA will not be covered in this SOP. Figure 10: The Leak Valve If you are not using the MPA, stabilize the pressure at about 7 or 8 x 10-3 torr. Record how many turns the leak valve was opened on the run sheet. Figure 11: The MPA Note: If you are depositing an oxide onto MOXTEK windows, we have found that the oxygen plasma can react with the windows, causing them to break. We have found better success with depositing a thin layer of the pure metal on the windows before depositing the oxide. This protects the windows from the reactive oxygen plasma. 12. Open the argon valves. Valve 2 should be open until the gauge reads about 15 (you can adjust this as needed). Stabilize the pressure at about 3 x 10-3 torr. (If you are depositing an oxide or a nitride, record the pressure after adding the argon and after adding the oxygen or nitrogen.) To open the argon tank valves, open the three valves on the tank, starting with the one closest to the tank, then moving to the next closest, etc. Note: This is how you open the valves on any gas tank, not just the Argon. [See Fig. 8] To close the valves on the tank, just close the three valves starting with the one furthest from the tank moving in reverse order as how you opened them. Figure 8: Gas tank Remember: Open: “In to out.” Close: valves, numbered in the “Out to in.” order to open. The main valve for the argon line is the Green “Nupro” valve behind the chamber. This line also has a Mass Flow Controller, which can be controlled from the box above the Crystal Monitor Readout. [See Fig. 9] This SOP will not cover how to use the Mass Flow Controller. Figure 9: The Green Nupro Valve (on the left) and the Mass Flow Controller (on the right). The mass flow controller is controlled on the large brown electronics box, next to the power box. 8. Turn RF Power Control Knob (On Source A) to a minimum. [See Fig. 6] 9. Turn the RF Power Button (Red Button On Source A) on. Figure 6: Source A. The Power control knob is the one in the middle. The RF Power Button is the second button from the right. Figure 7: CEX-2. The HV Power Switch is marked. 10. Turn the HV Power Switch (On CEX-2) on. [See Fig. 7] 11. Set Loading and Tuning to about 108.5 and 150.5 respectively. Record the Loading and Tuning values on the run sheet. 14. Once you have stabilized the pressure, record the operating pressure on the run sheet. (If you are depositing an oxide or a nitride, record the pressure after adding the argon and after adding the oxygen or nitrogen.) 15. Input the film material # into the Crystal Monitor Readout. [See Fig. 12] To input film material #: Press the “Film Number” button on the Readout. Press the up and down arrows until the number on the Readout matches the number on the post-it note next to the readout. Leave the Readout and it will set. Figure 12: The Crystal Monitor Readout 16. Turn the RF Power Control Knob (On Source A) to about 200 Watts (0.2 kW) of Incident Power (as read off of the Incident Power readout On Source A). [See Fig. 13] 17. Adjust the RF Power Control Knob and the PA Tuning Knob (PA Tuning Only Slightly!), and most importantly, adjust the working gas pressure, to Figure 13: Source A. The maximize the Incident Power and minimize the RF Power Control Knob and Reflected Power. the Incident Power Readout The Incident and Reflective power represent the have been highlighted. power going into the impedance source (the sputter gun) and the power coming out from this source back into the RF power supply, respectively. By maximizing Incident power and minimizing Reflective power, the output power from the RF power source is efficiently consumed. Eventually, a plasma will be created inside the chamber. When a plasma is first created, the reflected power readout should drop sharply (although it might not be by very much). Also, when there is a plasma in Figure 14: the chamber, you should be Left: The Ion Gauge. able to see a faint purple Right: Looking down in the Ion Gauge. When a glow in the chamber plasma is created, you will be able to see a through the glass Ion faint, purple glow by looking down this tube. Gauge. [See Fig. 14] Once a plasma is created, sputtering off of the target has begun. You can adjust the power until you get a good sputter rate (as read off of the Crystal Monitor Readout). A good sputter rate depends on the material. For pure metals, this is about 1 or 2 Å / sec. For oxides, this is about 0.2 Å / sec. 18. Record the sputter rate on the run sheet as read off of the Crystal Monitor Readout. 19. Once you have established a good rate, open the Sample Shutter and press “Start” on the Crystal Monitor Readout. To open the sample shutter: Unscrew the lower knob. Pull the side knob out. Screw the lower knob back in. To close the sample shutter, simply do the reverse as above. The Crystal Monitor Readout will count the total film thickness deposited after you pressed “Start” as measured Figure 15: The Sample from the crystal monitor (this will not be perfect, but it is a good, rough value). Shutter. 20. Once you reach the desired film thickness, close the Sample Shutter and press “Stop” on the Crystal Monitor Readout. 21. Record the total film thickness on the run sheet as read off of the Crystal Monitor Readout. 22. Once you are done, turn the RF Power Control Knob to a minimum, and turn off all power switches. 23. Close the argon (and oxygen or nitrogen if applicable) tanks. (Remember to close the tank “out to in” as described in Step 12.) 24. Wait a couple of minutes so that the gas in the argon and oxygen or nitrogen lines is vented into the chamber. 25. Tightly close the argon (and oxygen or nitrogen if applicable) valves (these would be the green Nupro valve and the leak valve, respectively). (Also, if applicable, shut off the Mass Flow Controller.) 26. Close the gate valve to the cryo pump. To close cryo pump: a. Remove wrench (if applicable). b. With the air valve still closed, turn on the pneumatic switch (the light switch-looking thing). c. Open the air valve. The gate valve should open completely. d. Turn off the pneumatic switch. The gate valve should close completely. e. You can use the handle to pull the gate valve a little more tightly shut, just in case it did not shut completely. 27. You can leave the turbo pump and the roughing pump running until you are ready to take the sample out of the chamber.