Evaluation of a turbidimetric immunoassay (TIA) for measuring IgG concentrations in mare colostrum. Student Research Project Utrecht University, Faculty of Veterinary Medicine, Department of Equine Sciences, The Netherlands June 2008 Drs. S. Sitters 0052000 Supervisors: Dr. T.A.E. Stout Dr. P.M. McCue Utrecht University, Faculty of Veterinary Medicine, Department of Equine Sciences, The Netherlands Colorado State University, College of Veterinary Medicine and Biomedical Sciences, Equine Reproduction Laboratory, CO, USA Contents Summary 3 Introduction 4 Materials and Methods 7 Results 9 Discussion 15 References 18 2 Summary A commercial turbidimetric immunoassay (ARS Foal-IgG Test; Animal Reproduction Systems, Chino, CA), designed for measuring IgG concentrations in foal serum, was evaluated as a means for quantifying IgG concentrations in the colostrum of 43 mares (28 post-partum/pre-suckle samples and 15 pre-partum samples). This TIA test was found to be unsuitable for determining IgG concentrations in either untreated or centrifuged (fat-free) mare colostrum. The IgG concentrations estimated by TIA were significantly higher than those quantified by the radial immunodiffusion (RID) assay, and there was no overall correlation between the two tests. By contrast, the BRIX (sugar) refractometer proved to be an accurate, fast and easy technique for estimating colostral IgG concentrations. The BRIX percentage score correlated highly with colostral IgG concentrations measured by RID (pre - and postfoaling samples: r=0.78, serial pre-foaling samples: r=0.88, serial post-foaling samples: r=0.99). Pre-partum colostral IgG concentrations were measured daily in 3 mares during the final 14 days of gestation. During this period, 2 mares maintained relatively constant IgG concentrations whereas the third mare showed a rise in colostral IgG starting around 12 days before parturition. Colostral IgG concentrations were also evaluated in 2 mares during the first 24 hours post-foaling. IgG concentrations dropped by 78% in the first 4-6 hours after birth, and then reached a plateau that persisted for the next 10 hours. 3 Introduction Equine neonates are born immunologically naïve because there is no significant in utero transplacental transfer of macromolecules such as maternal circulating antibodies. The absence of transplacental maternal macromolecule transfer in the horse is primarily a factor of the multilayered epitheliochorial type of placenta in which 6 cell layers impede the transfer of large molecules (Blackmer et al., 2002; Hines, 2003; Jeffcott, 1974; McGuire & Crawford, 1973; Steven & Samuel, 1975). On the other hand, newborn foals are immunocompetent and, indeed, an equine fetus is capable of responding to immunological stimulation from relatively early in its development. Nevertheless, because no antigenic stimulation takes place in utero, foals are born virtually agammaglobulinemic (Blackmer et al., 2002; Erhard, 2001; Hines, 2003; Jeffcott, 1974; Massey et al., 1991; McGuire & Crawford, 1973). As a result, adequate uptake of maternal immunoglobulins (primarily IgG) from the dam’s colostrum (passive transfer of immunity) is essential for the initial protection of newborn foals against infectious disease. During the last 2 weeks of pregnancy, antibodies (primarily IgG with some IgM class antibodies) from the mare’s blood are selectively concentrated in the colostrum (Blackmer et al., 2002; Genco et al., 1969; Hines, 2003; Knottenbelt et al., 2004). The average IgG concentration of equine colostrum at foaling is approximately 7,000 mg/dl (70 g/L), with a normal range of approximately 3000 to 12,000 mg/dl (30 to 120 g/L: Curadi et al., 2001; Erhard et al., 2001; Knottenbelt et al., 2004; Kohn et al., 1989; Lavoie et al., 1989; LeBlanc et al., 1992; Massey et al., 1991; Morris et al., 1985; Pearson et al., 1984; Perryman, 1980; Rouse & Ingram, 1970; Shideler et al., 1986; Townsend et al., 1983). Besides providing the newborn foal with antibodies, colostrum is an important source of nutrients, complement and lactoferrin. In addition, colostrum promotes cellular immunity and may have a variety of other physiological activities such as the activation of granulocytes and the provision of a local protective effect in the gastrointestinal tract. (Blackmer et al., 2002; Hines et al., 2003; Knottenbelt et al., 2004). In multiparous mares, the average rate of colostrum secretion is approximately 300 ml/h and the total volume of colostrum produced is 2-5 liters (Knottenbelt at al., 2004; Lavoie et al., 1989; Massey et al., 1991). After a neonatal foal has drunk colostrum, specialized enterocytes in the small intestine non-selectively absorb the colostral immunoglobulins by pinocytosis. In fact, the enterocytes engulf droplets of colostrum from the intestinal lumen, transfer the droplets in small vacuoles across the cell and then discharge the contents into lymphatic vessels. The antibodies are subsequently transported via the lymphatic system to the blood stream (Blackmer et al., 2002; Hines, 2003; Knottenbelt et al., 2004). The absorption of colostral antibodies via the enterocytes is greatest during the first 6-8 hours after birth, and has essentially stopped by 24-36 hours of age because the specialized enterocytes have been replaced by epithelial cells incapable of pinocytosis (‘gut closure’: Blackmer, 2002; Hines, 2003; Jeffcott, 1971, 1974; Knottenbelt et al., 2004; Pearson, 1984). To acquire sufficient antibodies to offer reasonable protection against infectious disease, the average foal needs to ingest approximately 2-4 liters of good quality colostrum within the first few hours of life (Blackmer et al., 2002; Hines, 2003; Knottenbelt et al., 2004; Lavoie et al., 1989; LeBlanc et al., 1992; Massey et al., 1991; Perryman, 1981). 4 In short, a newborn foal is entirely dependent on the maternal antibodies absorbed following the ingestion of its dam’s colostrum during the first few hours of life for immune protection. Failure of this passive transfer of maternal antibodies significantly increases the risk that a foal will succumb to infectious disease (Crawford et al., 1977; Jeffcott, 1974; Koterba et al., 1984; McGuire et al, 1975; Raidal, 1996; Robinson et al, 1993), although the degree of risk also depends on other factors such as management and level of exposure to pathogens. Nevertheless, in foals developing septicemia, levels of immunoglobulins (IgG) have been shown to be significantly lower than normal (Carter et al, 1986; Koterba et al., 1984). Despite, or maybe because, of its importance, there is considerable discussion with regard to the neonatal circulating IgG values that signify either failure of passive transfer of maternal antibodies (FPT) or partial failure of passive transfer (PFPT). This is partly because it has not been possible to precisely determine the concentration of foal serum IgG required for adequate protection of a healthy vigorous neonate. Nevertheless, it is generally accepted that a foal has suffered complete failure of passive transfer (FPT) if plasma IgG levels are < 200 mg/dl at 18-24 hours after birth. Partial failure of passive transfer (PFPT) has variously been defined as a plasma IgG concentration of 200 -400 mg/dl or 200-800 mg/dl (Crawford et al., 1977; Hines, 2003; Madigan, 1997; McGuire et al, 1977). As a result, it is generally advised that plasma IgG levels ≥400 mg/dl should be adequate for a healthy foal in a clean environment with minimal exposure to pathogens (Kohn et al., 1989; McGuire et al, 1977; Perryman, 1980; Rumbaugh et al, 1979). On the other hand, an IgG concentration ≥ 800 mg/dl is considered optimal because it is rarely possible to ensure ‘minimum exposure to pathogens’ (Blackmer et al., 2002; Hines, 2003; Koterba et al, 1984). Reported incidences of FPT and PFPT range from 10-24 % and 14-19 %, respectively (Crawford et al, 1977; Clabough et al., 1991; Erhard et al., 2001; Kohn et al, 1989; LeBlanc et al., 1986, 1990, 1992; McGuire et al., 1973, 1977; Pemberton et al., 1 980; Perryman & McGuire, 1980; Raidal, 1996; Stoneham et al., 1991; Tyler-McGowan et al., 1997). Variation in the reported incidences may in part be a factor of the different tests used to measure IgG concentrations, different reference ranges used to def ine FPT and PFPT, the age of foal when tested and associated management procedures. In any case, the most important risk factor for the development of FPT is thought to be poor colostrum quality (i.e. IgG concentration of < 1,000 - 3,000 mg/dl: LeBlanc et al., 1986, 1992; Morris et al., 1985; McGuire et al., 1977; Perryman & Crawford, 1979; Tyler McGowan et al., 1997). Other major factors associated with FPT include maternal factors such as agalactia, premature lactation and rejection of the foal, and foal factors like inability or lack of desire to nurse, prematurity, dysmaturity, and failure to absorb colostral immunoglobulins (Blackmer et al., 2002; Crawford et al., 1977; Hines, 2003; Jeffcott, 1971, 1974; LeBlanc et al, 1986; McGuire & Crawford, 1973; Pearson et al., 1984; Perryman & Crawford, 1979; Perryman 1981; Rumbaugh et al., 1979). Little is known about the effects of season, climate or maternal factors on neonatal foal colostral IgG absorption, while reports on the effects of breed, mare age, parit y, gestation length, dystocia, retained placenta, foal sex, and meteorological factors are contradictory (Clabough et al., 1991; LeBlanc et al., 1992; Morris et al, 1985; Shideler et al., 1986). 5 It is, however, clear that serum IgG concentrations in a newborn foal correlate well with the IgG concentrations in its mother’s colostrum (Curadi et al., 2001; Erhard et al., 2001; Kohn et al., 1989; Lavoie et al., 1989; LeBlanc et al., 1986; Morris et al., 1985). And since poor quality colostrum is one of the most important risk factors for the development of FPT, a logical first step in the prevention of FPT is to determine colostrum quality in the immediate postpartum period, prior to nursing. If the colostrum quality is poor, the foal can be supplemented with frozen-thawed colostrum, or a colostrum substitute. An optimal screening test for colostrum quality should therefore be rapid, accurate, repeatable, inexpensive and quantitative. Common qualitative techniques used to assess colostrum quality include the colostrometer (LeBlanc et al, 1986) and the BRIX (or sugar) refractometer (Cash, 1999). Other less widely used techniques include gluteraldehyde precipitation (Blackmer et al., 2002; Knottenbelt et al., 2004), latex agglutination (Knottenbelt et al., 2004 ; Waelchli et al., 1990) and an enzyme-linked immunosorbent assay (ELISA) (Erhard et al., 2001). The colostrometer measures the density or specific gravity of colostrum (Knottenbelt et al., 2004; LeBlanc et al, 1986; Massey et al., 1991; Waelchi et al, 1990). Colostrum with high IgG levels has a higher density and, therefore, specific gravity. However, determination of the specific gravity of equine colostrum with a colostrometer depends on accurate measurement of an exact volume (15 ml) of colostrum; even small errors in volume measurement can lead to significant inaccuracy in specific gravity determination (Chavatte et al., 1998). Refractometry measures the concentration of dissolved solids in a solution. In the case of a BRIX refractometer, a small amount of colostrum is placed on the prism and the light plate is closed so that colostrum is spread evenly across the prism. The refractometer is then pointed at a light source and the deviation or refraction of light is evaluated as a percentage score. Colostrum with a low concentration of dissolved solids (i.e. low IgG level) will scatter less light and therefore have a lower percentage score. Colostrum with high amounts of dissolved solids (i.e. high IgG levels) will cause more light scatter and thereby achieve a higher percentage score. BRIX refractometer evaluation of equine colostrum quality has been shown to be both repeatable (r= 0.98) and highly correlated with foal plasma IgG levels (r=0.85) as measured by the radial immunodiffusion assay (Cash, 1999; Chavatte et al., 1998). Similar correlations have been reported for BRIX refractometer evaluation of bovine colostrum (Molla, 1980). Direct quantification of IgG levels in colostrum can be performed using the radial immunodiffusion (RID) assay (Fleenor et al., 1981; Knottenbelt et al., 2004; Mancini et al., 1965). Unfortunately, RID cannot be performed easily on–farm, and it takes approximately 24 hours to obtain a result. Furthermore, the test is expensive. Recently, a commercial turbidimetric immunoassay (TIA) was validated for measuring IgG levels in the plasma of neonatal foals, to detect failure of passive transfer (McCue, unpublished data). The assay (ARS Foal-IgG Test; Animal Reproduction Systems, Chino, CA) is rapid, repeatable and correlates well with RID measurements of plasma IgG concentration. The primary objective of this study was to evaluate the commercial TIA test, designed for measuring IgG concentrations in foal serum, as a means for quantifying IgG concentrations in the colostrum of mares. A second objective was to evaluate colostrum IgG levels during the last days of pregnancy and the first 24 hours post -foaling. 6 Materials and Methods Colostrum samples (2 ml) were collected from 28 mares (various breeds and parities) immediately postpartum and prior to nursing (i.e. pre-suckle). Colostrum samples (2 ml) from another 15 mares (also various breeds and parities) had been previously collected on randomly selected days shortly prior to foaling. All 43 samples were frozen in plastic cryovials (-20 C) until evaluation. Colostrum IgG concentrations were subsequently evaluated using the commercial TIA test for quantifying serum IgG concentrations. Objective I: To evaluate a commercial TIA test for quantifying colostral IgG concentrations. Study 1: Comparing colostral IgG concentrations as measured by TIA with those measured using the BRIX refractometer or by RID. Frozen colostrum samples were thawed in a water bath at room temperature, and thereafter immediately evaluated using a BRIX refractometer, the TIA test and by RID. A commercial refractometer (Equine Colostrum Refractometer, Animal Reproduction Systems, Chino, CA) was used to obtain the BRIX percentage score. The relationship between colostrum quality and the BRIX % refractometer reading, as reported by Cash (1999), is shown in the table below. BRIX (%) <15 15-20 20-30 >30 IgG concentration (g/L) 0-28 28-50 50-80 >80 Colostrum quality Poor Fair Good Very Good Table 1: Equine Colostrum Refractometer Guide, Animal Reproduction Systems, Chino, CA. The TIA test was performed using a commercial kit (ARS Foal-IgG Test; Animal Reproduction Systems, Chino, CA), as follows. All materials were first allowed to equilibrate to room temperature. A colostrum sample was then diluted 1:19 with buffer (diluent containing < 0.1% sodium azide; Midland BioProducts Corporation, Boone, IA), which was expected to ensure that the IgG concentration fell within the working range of the assay. Next, 25 l of the diluted colostrum was added to 1.5 ml of dilution buffer in a cuvette and mixed gently; 180 l of the second dilution was then added to 1.3 ml of a polymer buffer in a second cuvette and again mixed. The second cuvette was then inserted into a spectrophotometer (Densimeter Model 590a, Animal Reproduction Systems, Chino, CA) and the system was zeroed. Next, 180 l of goat anti-equine IgG serum (Midland BioProducts Corporation, Boone, IA) was added and the solution was mixed gently. The cuvette was then replaced in the spectrophotometer and the percentage light transmission (% T) was measured. The spectrophotometer automatically converted the % T score into an estimated IgG concentration in mg/dl. This value had to be multiplied by 20 (initial dilution factor) to obtain the IgG concentration of the original colostrum sample. 7 RID was performed at the Veterinary Diagnostic Laboratory of Colorado State University, using a commercially available kit (Immunocheck, VMRD, Inc., Pullman, WA), as described by Mancini et al (1965). As for the TIA test, colostrum samples were first diluted 20-fold with sterile water to ensure that concentrations would be within the measurable range. The RID assay was then performed using 3 l of the diluted colostrum. Sample precipitation rings were evaluated after approximately 18 hours of incubation at room temperature. Again the value had to be multiplied by 20 to obtain the IgG concentration of the original colostrum sample. Correlation between the IgG concentrations measured using the various tests were examined using SPSS and Microsoft Excel software. Study 2: Validating the TIA test by serial dilution of colostrum samples. Colostrum samples from 4 mares were subjected to halving dilutions using a buffer (diluent containing < 0.1% Sodium azide, Midland BioProducts Corporation, Boone, IA). As previously, colostrum samples were first diluted 20-fold. The resulting solution was then diluted 1:1 with buffer to obtain a 1:40 dilution. Depending on the IgG concentrations measured in these solutions, either a less (1:80) or a more (1:10) concentrated solution was prepared to create a series of 3 dilutions per sample within the working range of the assay; 1:80 dilutions were prepared by 1:1 dilution of the 1:40 solution with buffer, while 1:10 dilutions were made by diluting the original colostrum sample. The IgG concentration of all solutions was then evaluated by both TIA and RID. Correlation analysis was performed. Objective II: To Evaluate pre- and post-partum colostral IgG concentrations. Study 1: Evaluating colostral IgG concentrations during the immediate pre-partum period. Colostrum samples were collected from 3 mares daily for 22 (n=1) or 13 (n=2) days prior to foaling, to monitor colostral antibody concentrations during the pre-foaling period. The IgG levels of these samples were determined using the BRIX (sugar) refractometer and, in the case of one mare, by RID. Study 2: Evaluating post-partum colostral IgG concentrations. Colostrum samples were collected from 2 mares every 2 h after foaling for 22 and 24 h, respectively (i.e. 0, 2, 4, 6, 8, 10, 12, 14, 16, 18, 20, 22 and 24 h), to monitor the decline in colostral antibody concentrations post-foaling. IgG concentrations were measured using both the BRIX refractometer and by RID. Correlation analysis was performed. 8 Results The utility of the commercial TIA test for quantifying colostral IgG concentrations. Pilot study: Comparing colostral IgG concentrations measured by TIA and RID. The TIA test was first performed on 1:20 and 1:40 dilutions of 5 randomly selected colostrum samples, to determine which dilution factor was most appropriate for bringing IgG concentrations into the working range of the assay. It was immediately clear that the IgG concentrations measured by using TIA were not accurate (Table 2) since, for example, the doubling dilution from 1:20 to 1:40 did not result in the expected halving of the measured IgG concentration. Since the measured IgG concentrations were also unfeasibly high, it was concluded that the TIA test is unsuitable for measuring IgG concentrations in untreated equine colostrum. 1 2 3 4 5 Colostrum sample Date of collection A B C D E 29/01/2006 21/02/2006 03/03/2006 20/02/2006 23/02/2006 1:20 dilution IgG (mg/dl) 1871.0 1303.3 1249.8 1294.2 408.2 1:40 dilution IgG (mg/dl) 1402.7 664.2 1412.7 1665.5 - Table 2. Effect of serial dilution of (untreated) colostrum on IgG levels, as determined by TIA. However, because it was possible that the discrepancies in the measured IgG concentrations were due to interference, e.g. from contained fat, the samples were reanalyzed after centrifugation at 1000 g for 15 min to remove cellular debris and fat. The fat-free colostrum located between the superficial fat layer and the underlying precipitate was harvested and examined by TIA (Table 3; Fig. 1). A full 3 sample serial dilution was not performed on sample D, because it was anticipated that other dilutions would result in IgG levels outside the range of the TIA. Serial dilution of fat-free colostrum did result in the desired linear decline in measured IgG concentrations (mg/dl). However, the absolute IgG concentrations recorded were still unrealistically high. Colostrum sample 1 2 3 4 5 A B C D E 1:10 dilution IgG (mg/dl) 1383.4 1530 1:20 dilution IgG (mg/dl) 776.5 750.6 1970.2 1072.8 810.8 1:40 dilution IgG (mg/dl) 367.7 344.2 1109.5 330 541.5 Table 3. IgG concentrations in serial dilutions of fat-free colostrum, as determined by TIA. 9 1:80 dilution IgG (mg/dl) 235.9 546.9 - 2500 IgG (mg/dl) 2000 sample A 1500 sample B sample C 1000 sample E 500 0 Serial dilutions Figure 1. IgG concentrations in serial dilutions of fat-free colostrum, as determined by TIA. To determine whether the IgG concentrations measured by TIA were accurate, 3 dilutions of 3 fat-free colostrum samples were measured by RID and TIA in parallel. In summary, the IgG concentrations measured by RID were significantly lower than those suggested by TIA, and although there was often an obvious relationship between the two tests for a given sample, this relationship varied between samples such that here was no overall correlation (correlation coefficient (r) = 0.43, P = 0.25, Fig. 2). It was concluded that the TIA test is not suitable for determining IgG concentrations in either untreated or centrifuged (fat-free) colostrum. As a result, experiments to examine changes in colostral IgG concentrations during the pre and post-partum periods used alternative tests. IgG concentrations (m g/dl) by TIA and RID 1000 IgG by RID 800 600 400 200 0 0 500 1000 1500 2000 2500 IgG by TIA Figure 2: IgG concentrations in fat-free colostrum as determined by TIA versus RID (n= 3 serial dilutions of 3 samples). Next, the accuracy of BRIX refractometry for estimating IgG concentrations was examined by comparing BRIX scores with RID test scores, because the former is a simple way of estimating colostral IgG concentrations. 10 Evaluating the BRIX (sugar) refractometer as an estimator of colostral IgG concentrations. Study 1: Comparing BRIX percentage scores with IgG concentrations measured by RID. BRIX scores (%) were compared with RID IgG concentrations (mg/dl) (Figs. 3 a, b). Because the maximum value measurable using the BRIX refractometer was 32 %, samples with a score > 32 % were omitted from comparisons. This left 35 pre- and post-foaling colostrum samples with an accurate BRIX score (Fig. 3a). Of these 35 samples, 21 were postfoaling/pre-suckle (Fig. 3b). The mean IgG concentration of the 28 post-foaling/pre-suckle colostrum samples was 11,384 mg/dl (range; 4,660 to 19,520 mg/dl) as measured by RID. A correlation coefficient of 0.78 (P<0.001) was obtained for the relationship between the IgG levels as measured by BRIX and RID in the pre- and post-foaling colostrum samples. For the post-foaling samples a correlation coefficient of 0.72 (P<0.001) was found. Pre- and postfoaling colostrum sam ples 25000 IgG (mg/dl) 20000 15000 10000 5000 0 0 5 10 15 20 25 30 35 Brix score (%) Figure 3a. Comparison of pre- and post-foaling colostrum IgG concentrations using BRIX (sugar) refractometry versus RID. Postfoaling colostrum sam ples 25000 IgG (mg/dl) 20000 15000 10000 5000 0 0 5 10 15 20 25 30 35 Brix score (%) Figure 3b. Comparison of post-foaling colostrum IgG concentrations using BRIX (sugar) refractometry versus RID. 11 Evaluating pre- and postpartum colostral IgG concentrations. Study 1: Evaluating colostral IgG concentrations during the immediate pre-partum period. In the mare from which colostrum samples were collected daily for 22 days pre-foaling, colostral IgG concentrations tended to rise as foaling approached, as measured by both BRIX refractometry and RID (Fig. 4a). In fact, BRIX readings showed a fairly consistent increase from a baseline of approximately 5 % at >12 days prior to foaling to 27 % on the day of foaling. RID measurements gave a broadly similar trend, with an increase from a baseline of approximately 2500 mg/dl at >12 days before foaling to a maximum of 12.000 mg/dl 3 days before foaling, although the increase was less consistently progressive. Nevertheless, the correlation coefficient for the relationship between IgG concentrations measured by TIA and RID was 0.88 (P<0.001) (Fig. 4b). 14000 35 12000 30 10000 25 8000 20 6000 15 4000 10 2000 5 0 Brix score (%) IgG (mg/dl) Mare A RID BRIX 0 1 3 5 7 9 11 13 15 17 19 21 23 Time (days), day 23=day of foaling Figure 4a. Pre-foaling colostral IgG concentrations determined by BRIX (sugar) refractometry and RID: mare A. IgG concentrations by refractom etry and RID 14000 IgG (mg/dl) 12000 10000 8000 6000 4000 2000 0 0 5 10 15 20 25 30 Brix score (%) Figure 4b. Comparison of IgG concentrations determined in pre-foaling colostrum samples using BRIX (sugar) refractometry versus RID: mare A. 12 In the two other mares from which pre-foaling colostrum samples were collected and analyzed by BRIX refractometry (Fig. 5), BRIX values were high (mean = 27 and 19 % for mares B and C, respectively) and relatively constant throughout the period examined (from 13 days pre-foaling). Brix refractom eter 35 Brix score (%) 30 25 20 Mare B 15 Mare C 10 5 0 1 2 3 4 5 6 7 8 9 10 11 12 13 14 Time (days), day 14=day of foaling Figure 5. Pre-foaling colostral IgG concentrations estimated by BRIX (sugar) refractometry: mares B & C. Study 2: Evaluating post-partum colostral IgG concentrations. For the 2 mares examined serially, IgG concentrations in colostrum declined during the post-partum period (Figs. 6 a, b). From an initial IgG concentration of around 9000 mg/dl (BRIX scores of 26 % and 31 %), concentrations fell rapidly during the first 4 -6 hours post-partum to a plateau of approximately 2000 mg/dl (BRIX score of 10%), which was then maintained over the subsequent 10 hours. The correlation coefficient for IgG concentrations measured by TIA and RID was 0.99 (P<0.001) (Fig. 6c). Mare A 10000 35 IgG (mg/dl) 25 6000 20 4000 15 10 2000 Brix score (%) 30 8000 RID BRIX 5 0 0 0 2 4 6 8 10 12 14 16 18 20 22 24 Time (hours) Figure 6a. Post-foaling colostral decline in IgG concentrations determined by BRIX (sugar) refractometry and RID: Mare A. 13 Mare B 10000 35 IgG (mg/dl) 25 6000 20 4000 15 10 2000 Brix score (%) 30 8000 RID BRIX 5 0 0 0 2 4 6 8 10 12 14 16 18 20 22 Time (hours) Figure 6b. Post-foaling colostral decline in IgG concentrations determined by BRIX (sugar) refractometry and RID: Mare B. IgG concentrations by refractom etry and RID 12000 IgG (mg/dl) 10000 8000 6000 4000 2000 0 0 5 10 15 20 25 30 35 Brix score (%) Figure 6c. Comparison of IgG concentrations in post-foaling colostrum samples using BRIX (sugar) refractometry versus RID (n=2 mares). 14 Discussion Although a newborn foal possesses a functional immune system, it is immunologically naïve because there is no significant transplacental transfer of maternal antibodies. As a result, initial immune protection of the newborn foal is critically dependent on adequate passive transfer of immunity (primarily IgG class antibodies) via the dam’s colostrum. Failure of passive transfer therefore significantly increases the risk of a neonatal foa l succumbing to infectious disease. Failure of passive transfer can occur for several reasons, some of which are easily identified and can therefore be managed effectively; examples include maternal agalactia, failure to drink due to neonatal weakness, delayed development of the sucking reflex, maternal non-cooperation, and loss of colostrum due to premature lactation. By contrast, failure of the dam to produce colostrum of sufficient quality is more difficult to identify and tends, therefore, to be less effectively managed (Morris et al., 1985). Moreover, poor colostrum quality is thought to be the most important risk factor for failure of passive transfer (Crawford et al., 1977; McGuire et al., 1977; Perryman, 1979). On the other hand, the reasons for low absolute colostral IgG concentrations are largely unknown, although a genetic defect in a mare’s ability to concentrate IgG from serum into the colostrum in late gestation (Perryman, 1979) has been proposed, and increasing maternal age has been associated with reduced colostral IgG concentration (LeBlanc). And while premature lactation might be an important factor in the total amount of colostral IgG available (Jeffcott, 1974; Morris et al., 1985), other factors may be more important determinants of initial colostrum quality. The postpartum/pre-suckle colostral IgG concentrations recorded in the present study fall within the range of mean values reported previously for mares immediately post -partum (Table 3). Breed Various breeds Various breeds Various breeds Not specified Crossbred Thoroughbred Thoroughbred Thoroughbred Arabian Arabian Arabian Quarter Horse Quarter Horse IgG (mg/dl) 11,384 5,450 11,324 16,583 8,867 5,036 2,354 4,608 9,691 10,847 6,394 5,843 1,080 Sample size 28 139 18 6 10 39 32 22 14 12 25 21 1 Standardbred Standardbred Standardbred Shetland pony Heavy draft mares 3,932 5,128 8,912 4,600 6,438 15 136 36 9 11 Reference Present study Erhard et al., 2001 Massey et al., 1991 Lavoie et al., 1989 Curadi et al., 2001 LeBlanc et al., 1992 McGuire et al., 1977 Pearson et al, 1984 Pearson et al, 1984 Perryman, 1980 LeBlanc et al., 1992 LeBlanc et al., 1992 McGuire & Crawford, 1973 LeBlanc et al., 1992 Morris et al, 1985 Kohn et al., 1989 Rouse & Ingram, 1970 Townsend et al., 1983 Table 4: Average IgG concentrations in colostrum of mares at the time of parturition. 15 The large differences in mean colostral IgG concentrations between studies may in part be breed-related (Kohn et al., 1989; LeBlanc et al., 1992; Pearson et al., 1984), although there is also considerable within-breed variation. Another important factor is the time of colostrum collection post-partum, because colostral IgG levels decrease rapidly after foaling, as will be discussed later. Pre-suckle colostrum samples collected immediately after foaling will therefore have significantly higher IgG concentrations than samples collected several hours post-partum, when the foal has had the opportunity to nurse. Factors affecting individual colostral IgG concentrations (Lavoie et al., 1989; LeBlanc et al., 1992; Morris et al., 1985; Pearson et al., 1984; Shideler et al., 1986) may include repeated vaccination, particularly in the last month before foaling (Pearson et al., 1984; Perryman, 1981; Shideler et al., 1986). Again, however, it is unlikely that vaccination history alone will account for all of the between-mare difference. Age and health of the dam, number of lactations, total colostrum yield and management are other factors that have been suggested to influence colostral IgG content (Knottenbelt et al, 2004). Whatever its determinants, the importance of colostral quality to foal morbidity makes determination in the immediate postpartum period, prior to nursing, highly desirable. Initial postpartum colostral IgG content appears to be a good estimator of total colostral IgG available to a foal (Lavoie et al., 1989), and correlates well with subsequent foal serum IgG concentrations (Curadi et al., 2001; Erhard et al., 2001; Kohn et al., 1989; LeBlanc et al., 1986; Morris et al., 1985). If colostrum quality is found to be poor, a foal can be supplemented with frozen-thawed colostrum, or a colostrum substitute. Oral supplementation is the easiest, safest and most effective way and is possible until the moment of ‘gut closure’, but preferably within 12 hours after birth (LeBlanc et al., 1992; Raidal et al., 2005). An optimal screening test for colostrum quality must therefore be rapid, accurate, repeatable, inexpensive, and quantitative. In the current study, a commercial turbidimetric immunoassay (Foal-IgG Test; Animal Reproduction Systems, Chino, CA) was evaluated for its ability to measure IgG concentrations in mare colostrum. This TIA test had recently been validated for measuring IgG concentrations in the plasma of neonatal foals. Unfortunately, the current study demonstrated that the commercial TIA test is not suitable for determining IgG concentrations in either untreated or centrifuged (fat-free) mare colostrum. Not only were the IgG concentrations estimated by TIA significantly higher than those measured by RID, but there was no overall correlation between the two tests. Although the reason for the differences between the two tests is unclear, they could arise at any stage of the TIA procedure. For example, it is possible that the goat anti-equine IgG-antibodies reacted with colostral substances other than IgG, and that the increased non-specific coagulation was misinterpreted as a raised IgG concentration. In addition, even after centrifugation, colostrum is a thick emulsion containing approximately 25 % fat, 16 % total protein and 22 % dry matter (Csapó-Kiss et al., 1995; Csapó et al., 1995). This would presumably mean that initial light transmission would be much lower than in a serum sample, and thereby reduce the proportional change in transmission after addition of the goat anti-equine IgG in a sample/individual specific fashion. In this respect, there was an apparent correlation between the results of TIA and RID within individual samples (Fig. 2). Another technique for estimating colostral IgG concentrations is the BRIX (sugar) refractometer. Previously, it has been shown that the equine colostrum quality as estimated using the BRIX refractometer correlates highly with foal plasma IgG concentrations (r=0.85), as measured by RID (Cash, 1999; Chavatte et al., 1998). In the current study, we found that, as expected, the BRIX percentage score correlated highly with colostral IgG concentrations measured by RID (pre- and post-foaling samples: r=0.78, post-foaling samples: r=0.72, serial pre-foaling samples: r=0.88, serial post-foaling samples: r=0.99). A similar relationship (r=0.89) has been reported for bovine colostrum (Molla, 1980). 16 The exact relationship between the BRIX score and IgG concentration (mg/dl) estimated in the current study differed slightly to that provided by the manufacturers of the refractometer (Table 1), with BRIX values in the current study reflecting higher IgG concentrations. The relationship provided by the manufacturers is based on the study by Cash (1999). In that study an immunoturbidimetric assay (Hitachi 717 Automated Biochemistry Analyser) was used to determine colostral IgG (mg/dl) whereas a radial immunodiffusion assay was used in the current study. The difference in techniques most probably explains the difference in the exact relationship between BRIX score and IgG concentration. For this reason, the regression (IgG (mg/dl) = 472.48 * BRIX (%) – 1013.50) found in this study (Fig. 3a) is used in Table 5. BRIX (%) <15 15-20 20-30 >30 IgG Concentration (g/L) 0-60 60-80 80-130 >130 Colostrum quality Poor Fair Good Very Good Table 5. Equine Colostrum Refractometer Guide, as measured in the current study. In any case, BRIX refractometry proved to be an accurate, fast and easy method for estimating colostrum IgG concentrations and was subsequently used to evaluate pre- and post-partum colostral IgG concentrations. In this respect, it has been reported that IgG antibodies from the mare’s blood are selectively concentrated in the colostrum during approximately the last 2 weeks prior to parturition. However, in the current study, mares B and C maintained relatively constant colostral IgG concentrations during the final 14 days of gestation (Fig. 5), and it is assumed that the rate of IgG accumulation was matched by the rate of fluid accumulation in the mammary gland. By contrast, mare A showed a clear rise in colostral IgG concentration starting at around 12 days before parturition (Fig. 4a), and it is assumed that transport of IgG exceeded the rate of fluid accumulation; Pearson et al. (1984) reported similar individual differences in IgG sequestration. The greater variability in IgG concentrations measured by RID than refractometry during the period shortly before foaling in mare A (Fig. 4a) may be explained by greater human error during RID or may reflect real differences in the rate of accumulation of IgG versus sugars. The variability in the timecourse and absolute levels of pre-partum colostral IgG concentrations in mares, means that it would be difficult to accurately predict final colostrum IgG concentration before the foal is born. In the current study, colostral IgG concentrations dropped by 78% in the first 4-6 hours after birth, and then reached a plateau that persisted for at least 10 hours (Fig. 6a, b). Similar results have been reported previously (Knottenbelt et al., 2004; Lavoie et al., 1989; Pearson et al., 1984; Rouse & Ingram, 1970; Erhard et al., 2001; Curadi et al, 2001), and speed of the fall in colostral IgG concentrations emphasize the significance of the first colostrum in the passive transfer of immunity, and explains why the initial colostral IgG concentration correlates well with total available IgG (Lavoie et al., 1989). It is concluded that measuring the IgG concentration of colostrum post-partum, preferably pre-suckle, is of great value for the prevention of failure of passive transfer (Kohn et al., 1989; Morris et al., 1985; Raidal, 1996; Stoneham et al., 1991; Tyler-McGowan et al., 1997). Unfortunately, the large (individual) variation in the rate of increase in colostral IgG concentration in the period leading up to foaling means that estimating colostrum quality in advance is likely to be difficult. Moreover, because it is not clear what factors influence IgG accumulation it is not clear whether it is possible to manipulate (improve) colostral quality. Although, the commercial TIA test proved unsuitable for quantifying colostral IgG concentration, the BRIX (sugar) refractometer proved to be an easy, accurate, rapid, repeatable and inexpensive way of measuring this parameter. 17 References 1. Blackmer JM, Sellon DC, Hines MT. Failure of passive transfer. In: Smith BP (ed), Large Animal Internal Medicine, Third Edition. Mosby, St. Louis, 2002: 15921595. 2. Carter KG, Martens RJ. Septicemia in the neonatal foal. Comp Cont Educ Prac Vet 1986; 8; s256-s270. 3. Cash RSG. Colostral quality determined by refractometry. Equine Veterinary Education 1999; 11(1): 36-38. 4. Chavatte P, Clément F, Cash R, Grongnet JF. Field determination of colostrum quality by using a novel, practical method. In Proceedings 44th Annu Conv Am Assoc Equine Practnr 1998: 206-209. 5. Clabough DL, Levine JF, Grant GL, Conboy HS. Factors associated with failure of passive transfer of colostral antibodies in Standardbred foals. J Vet Intern Med 1991; 5: 335-340. 6. Crawford TB, McGuire TC, Hallowell AL, Macomber LE. Failure of colostral antibody transfer in foals: its effect, diagnosis and treatment. Proceedings 23 rd Annu Conv Am Assoc Equine Practnr 1977: 265-275. 7. Csapó-Kiss Zs, Stefler J, Martin TG, Makray S, Csapó J. Composition of mares’ colostrum and milk. Fat content, fatty acid composition and vitamin content. Int. Dairy Journal 1995; 5: 393-402. 8. Csapó-Kiss ZS, Stefler J, Martin TG, Makray S, Csapó J. Composition of mares’ colostrum and milk. Protein content, amino acid composition and contents of macro- and micro-elements. Int Dairy Journal 1995; 5: 403-415. 9. Curadi MC, Minori D, Demi S, Orlandi M. Immunotransfer in the foal. Annali Facolta di Medicina Veterinaria Pisa 2001; 54: 127-140. 10. Erhard MH, Luft C, Remler HP, Stangassinger M. Assessment of colostral transfer and systemic availability of immunoglobulin G in newborn foals using a newly developed enzyme-linked immunosorbent assay (ELISA) system. J Anim Physiol Nutr (Berl) 2001; 85: 164-173. 11. Fleenor WA, Stott GH. Single radial immunodiffusion analysis for quantitation of colostral immunoglobulin concentration. J Dairy Sci 1981; 64(5): 740-747. 12. Genco RJ, Yecies L, Karush F. The immunoglobulins of equine colostrum and parotid fluid. J Immunol 1969; 103: 437-444. 13. Hines MT. Immunodeficiencies of foals. In: Robinson NE (ed), Current Therapy in Equine Medicine, 5. Saunders, Philadelphia, 2003: 692-697. 14. Jeffcott LB. Duration of permeability of the intestine to macromolecules in the newly-born foal. Vet Rec 1971; 88: 340-341. 15. Jeffcott LB. Studies on passive immunity in the foal. J Comp Pathol 1974; 84: 93101. 18 16. Jeffcott LB. Some practical aspects of the transfer of passive immunity to newborn foals. Equine Vet J 1974; 6: 109-115 17. Knottenbelt DC, Holdstock N. The role of colostrum in immunity. In: Equine neonatology medicine and surgery. Saunders, Edinburgh, 2004: 15-27. 18. Knottenbelt DC, Holdstock N. Methods of assessing colostrum quality. In: Equine neonatology medicine and surgery. Saunders, Edinburgh, 2004: 393-394. 19. Kohn CW, Knight D, Hueston W, Jacobs R, Reed SM. Colostral and serum IgG, IgA, and IgM concentrations in Standardbred mares and their foals at parturition. J Am Vet Med Assoc 1989; 195: 64-68. 20. Koterba AM, Brewer BD, Tarplee FA. Clinical and clinicopathological characteristics of the septicaemic neonatal foal: review of 38 cases. Equine Vet J 1984; 16: 376-382. 21. Lavoie JP, Spensley MS, Smith BP, Mihalyi J. Colostral Volume and immunoglobulin G and M determination in mares. Am J Vet Res 1989; 50(4): 466470. 22. LeBlanc MM, McLaurin BI, Boswell R. Relationships among serum immunoglobulin concentration in foals, colostral specific gravity, and colostral immunoglobulin concentration. J Am Vet Med Assoc 1986; 189: 57-60. 23. LeBlanc MM, Hurtgen JP, Lyle S. A modified zinc sulfate turbidity test for the detection of immune status in newlyborn foals. Equine Veterinary Science 1990; 10(1); 36-39. 24. LeBlanc MM, Tran T, Baldwin JL, Pritchard EL. Factors that influence passive transfer of immunoglobulins in foals. J Am Vet Med Assoc 1992; 200: 179-183. 25. Madigan JE. Colostrum – Assessment of and sources for foals, Assessment of passive immunity. In: Manual of equine neonatal medicine, third edition. Live Oak Publishing, Woodland, CA. 1997: 32-46. 26. Mancini G, Carbonara AO, Heremans JF. Immunochemical quantitation of antigens by single radial immunodiffusion. Immunochemistry 1965; 2: 235-254. 27. Massey RE, LeBlanc MM, Klapstein EF. Colostrum feeding of foals and colostrum banking. In Proceedings 37 th Annu Conv Am Assoc Equine Practnr 1991: 1-8. 28. McCue PM. Evaluation of a new turbidimetric immunoassay system for measurement of plasma antibody concentration in newborn foals. Unpublished data 2005. 29. McGuire TC, Crawford TB. Passive immunity in the foal: measurement of immunoglobulin classes and of specific antibody. Am J Vet Res 1973; 34: 12991303. 30. McGuire TC, Poppie MJ, Banks KL. Hypogammaglobulinemia predisposing to infection in foals. J Am Vet Med Assoc 1975; 166(1): 71-75. 19 31. McGuire TC, Crawford TB, Hallowell AL, Macomber LE. Failure of colostral immunoglobulin transfer as an explanation for most infections and deaths of neonatal foals. J Am Vet Med Assoc 1977; 170(11): 1302-1304. 32. Molla A. Estimation of bovine colostral immunoglobulins by refractometry. Vet Rec 1980; 107(2): 35-36. 33. Morris DD, Meirs DA, Merryman GS. Passive transfer failure in horses: incidence and causative factors on a breeding farm. Am J Vet Res 1985; 46: 2294-2299. 34. Pearson RC, Hallowell AL, Bayly WM, Torbeck RL, Perryman LE. Times of appearance and disappearance of colostral IgG in the mare. Am J Vet Res 1984; 45(1): 186-190. 35. Pemberton et al., Hyppogammaglobulinaemia in foals: prevalence on Victorian studs and simple methods for detection and correlation in the field. Aust Vet J 1980; 56: 469-473. 36. Perryman LE, Crawford TB. Diagnosis and management of immune system failures of foals. Proceedings 25 th Annu Conv Am Assoc Equine Practnr 1979: 235-245. 37. Perryman LE, McGuire TC. Evaluation for immune system failures in horses and ponies. J Am Vet Med Assoc 1980; 176: 1374-1377. 38. Perryman LE. Immunological management of young foals. Comp Cont Educ Prac Vet 1981; 3: s223-s228. 39. Raidal SL. The incidence and consequences of failure of passive transfer of immunity on a thoroughbred breeding farm. Aust Vet J 1996; 73(6): 201-206. 40. Raidal SL, McTaggert C, Penhale J. Effect of withholding macromolecules on the duration of intestinal permeability to colostral IgG in foals. Aust Vet J 2005; 83(12): 78-81. 41. Robinson JA, Allen GK, Green EM, Fales WH, Loch WE, Wilkerson CG. A prospective study of septicaemia in colostrums-deprived foals. Equine Vet J 1993; 25(3): 214-219. Comment in Equine Vet J 1993; 25(5): 475. 42. Rouse BT, Ingram DG. The total protein and immunoglobulin profile of equine colostrum and milk. Immunology 1970; 19: 901-907. 43. Rumbaugh GE, Ardans AA, Ginno D, Trommerhausen-Smith A. Identification and treatment of colostrum-deficient foals. J Am Vet Med Assoc 1979; 174: 273-276. 44. Shideler RK, Squires EL, Osborne M. Immunoglobulin concentration and blood chemistry parameters in the newborn foal: relationship to sex, type of dam, gestation length and mare serum and colostrums concentrations. Proceedings 32nd Annu Conv Am Assoc Equine Practnr 1986: 145-151. 45. Steven DH, Samuel CA. Anatomy of the placental barrier in the mare. J Reprod Fert, Suppl 1975: 579-582. 20 46. Stoneham SJ, Digby NJ, Ricketts SW. Failure of passive transfer of colostral immunity in the foal: incidence, and the effect of stud management and plasma transfusions. Vet Rec 1991; 128(18): 416-419. 47. Townsend HG, Tabel H, Bristol FM. Induction of parturition in mares: effect on passive transfer of immunity to foals. J Am Vet Med Assoc 1983; 182(3): 255257. 48. Tyler-McGowan CM, Hodgson JL, Hodgson DR. Failure of passive transfer in foals: incidence and outcome on four studs in New South Wales. Aust Vet J 1997; 75: 56-59. 49. Waelchli RO, Hassig M, Eggenberger E, Nussbaumer M. Relationships of total protein, specific gravity, viscosity, refractive index and latex agglutination to immunoglobulin G concentration in mare colostrums. Equine Vet J 1990; 22(1): 39-42. 21