Johnson Noise Thermometry at High Pressure Ivan C. Getting (CIRES) and John L. Hall (JILA) University of Colorado at Boulder a COMPRES member institution June 2004 A Johnson Noise Thermometer has been developed which makes it possible to measure temperature accurately at high pressure. All effects of pressure, deformation, chemical reaction, etc., which normally plague thermocouples, are accounted for with a simple measurement of resistance each time temperature is to be measured. Thermocouples and other temperature sensors can be calibrated against this absolute thermometer in each specific type of apparatus and cell design of interest. Such calibration studies need be made only once for each experimental configuration with the results applicable to subsequent runs. This electronic thermometer requires a ~100 ohm probe with four leads which also serve as two thermocouples. For each temperature determination both the electric noise across the resistor and the resistance of the resistor are measured. Slight non-idealities in the electronic amplifiers and the imprecise definition of the band width make it rather difficult to build a Johnson noise thermometer with intrinsic calibration at the ~1 Kelvin level. The thermometer we have built will be calibrated against NIST traceable thermocouples at one atmosphere where the thermocouples are known to be correct. It is anticipated that ITS-90 can be realized to 1 K by this procedure. 1 Figure 1. Layout of the Johnson noise thermometer. A set of low noise relays directs signals from the resistive probe and other calibration resistors to the amplifiers, for noise measurements, or to a commercial LCR meter, for resistance measurements. The digitized noise signals, thermocouple voltages, and resistance values are directed to a computer for analysis. 2 NC Normalized Output vs. RT N.C. Poly. (N.C.) 16 14 NC Normalized Output 12 10 8 6 4 2 0 0 50 100 150 200 250 Sensor RT (ohm*kelvins) 300 350 400 Thousands Figure 2. Johnson noise thermometer output vs. the sensor RT product, where R = sensor resistance, T = sensor temperature. Each point is actually 100 separate measurements. In this room temperature study, the sensor resistance is changed in 200 ohm intervals simulating the RT range intended with a nearly constant 100 ohm sensor over a wide temperature range. With the intended sensor, room temperature corresponds to RT = ~50 thousand. RT = ~350 thousand corresponds to ~2000 K. Amplifier voltage noise is eliminated from the data by taking the correlated output of two amplifier systems in parallel. The data show a slight upward curvature due to current noise in the preamplifier. 3 NC Gain Factor vs. Time 0.990 0.985 NC Gain Factor 0.980 0.975 0.970 0.965 0.960 0.955 2/26/04 2/25/04 2/24/04 2/23/04 2/22/04 2/21/04 0.950 Date Figure 3. Amplifier gain vs. time. The amplifier gain and zero can be checked for each temperature measurement. The gain has recently been made very stable. The drift rate in these data is ~0.01 % per day. This stability permits us to make temperature measurements over several days without checking the gain. 4 Raw Zero Outputs vs. Time N.C. A.C. Linear (A.C.) Linear (N.C.) 0.050 0.040 Output 0.030 0.020 0.010 0.000 2/26/04 2/25/04 2/24/04 2/23/04 2/22/04 2/21/04 -0.010 Date Figure 4. Amplifier zero vs. time. The amplifier zero is now also extremely stable. The drift rate corresponds to 0.003 % of full scale per day. Thus temperature measurements can be made for days without checking the amplifier zero as well. With these recent improvements in stability, each temperature measurement can now be made in about 30 seconds, 15 seconds for the measurements and a similar time for data reduction. 5 N.C. 4 std dev + tol tol + N.C. 4 ave std dev Poly. (N.C. 4 ave) N.C. Norm. Output Fit Residuals vs. RT 0.100 0.080 0.060 Output Residuals 0.040 0.020 0.000 -0.020 -0.040 -0.060 -0.080 -0.100 0 50 100 150 200 250 Sensor RT (ohm*kelvins) 300 350 400 Thousands Figure 5. Thermometer output fit residuals. The output data shown in Figure 2 are fit with a 4th degree polynomial. At each RT value 100 determinations of the Johnson noise are made using one million measurements of each of two amplifier outputs. Three such determinations are shown at each value of RT, blue squares. The plot above represents over four billion samples of Johnson noise. The standard deviation of each group of 100 determinations is shown by the orange squares. The original design Type A standard uncertainty is shown by the dashed black lines. The instrument performs as designed. The Type A relative standard uncertainty of a single measurement, sampling over a 10 s period, is ~0.15 % which corresponds to ~1.5 K at 1000 K. This precision can be improved by making multiple measurement. The red diamonds show the average of each set of 100 measurements. The standard deviation of these means is ~0.015 % or ~0.15 K at 1000 K. 6 NC Relative Std. Dev. of Fit Residuals vs. RT 0.0050 Ave. Z & G meas. Z & G Relative Std. Dev. of Residuals 0.0045 0.0040 0.0035 0.0030 0.0025 0.0020 0.0015 0.0010 0.0005 0.0000 0 50 100 150 200 250 Sensor RT (ohm*kelvins) 300 350 400 Thousands Figure 6. Relative standard deviation of the output fit residuals. This plot shows explicitly the precision of the temperature determinations. Measurement of the amplifier gain and zero contain scatter as do the temperature determinations. The precision with gain and zero measurements is shown in the dashed line. With the newly won stability these two parameters no longer need to be determined for each measurement. The heavy blue lines shows the ~0.15 % precision over the entire measurement range without gain and zero measurements. Having to make gain and zero checks for each measurement would increase uncertainty by a factor of 1.7 and the measurement time by a factor of 3. 7 Planned Testing: The Johnson Noise Thermometer will be migrated to a large volume multianvil device during the winter of 2004/2005. A collaborative program is under development with Yanbin Wang and Mark Rivers of GSECARS at APS, Argonne National Laboratory. We plan to implement the thermometer in a multi-anvil cell in my laboratory in Colorado where we know the electrical ambient is sufficiently quiet and then to migrate the instrument to GSECARS where is will be available to the high pressure community. Uncertainty in temperature measurements made with this thermometer decreases as the number of measurements is increased. It operates at a resolution of ~1.5 K at 1000 K for a single measurement over 10 s. It is thus practical to achieve fractional Kelvin resolution in a few minutes. This is better than the temperature stability in typical large volume high pressure devices. This instrument is capable of resolution higher than the definition of the practical temperature scale, ITS-90, above ~1000 K. The application of Johnson noise thermometry to specific solid medium cells and apparatus type stands to solve the problem of temperature measurement at high pressure, a problem which has been with us for 100 years. 8