NMI TR 9 Uncertainties in Photometric Integrals James L. Gardner First edition — November 2005 Bradfield Road, Lindfield, NSW 2070 PO Box 264, Lindfield, NSW 2070 Telephone: (61 2) 8467 3600 Facsimile: (61 2) 8467 3610 Web page: http://www.measurement.gov.au © Commonwealth of Australia 2005 This document was originally prepared for the Korean Institute of Standards and Science in 2003 while James Gardner was there as a guest researcher. It had minor revisions up to November 2003. It was subsequently modified in December 2003 and circulated as Revision 2.2. This document is Revision 3.1. It contains significant changes to the method for f1. CONTENTS 1 Photometric Integrals ............................................................................................................. 1 1.1 Photometric Response ................................................................................................... 1 1.2 Spectral Mismatch Factor.............................................................................................. 1 1.3 Quality Factor of a Photometer ..................................................................................... 2 2 Propagation of Uncertainty .................................................................................................... 2 3 Uncertainty Propagation by Component of the Spectral Measurement ................................. 3 4 Spectral Measurement as a Transfer ...................................................................................... 5 5 Uncertainty Components in Spectral Measurement ............................................................... 5 5.1 Base Uncertainty of the Reference Standard ................................................................. 5 5.2 Offset Components of the Transfer Ratio ..................................................................... 6 5.3 Scaling Components of the Transfer Ratio ................................................................... 7 5.4 Wavelength Components of the Transfer...................................................................... 8 5.5 Multiple Transfers from the Reference ......................................................................... 9 6 Uncertainty Estimates for the Photometric Integrals ............................................................. 9 6.1 Photometric Response for Illuminant A ...................................................................... 10 6.2 Spectral Mismatch Factor............................................................................................ 11 6.3 Quality Factor of a Photometer. .................................................................................. 12 7 Conclusion ............................................................................................................................ 14 8 References ............................................................................................................................ 14 iii SUMMARY Integration of spectral measurements of photometer response and source power distributions is used to calculate illuminance response, spectral mismatch factors and photometer quality factors. Methods are given to estimate uncertainties in these various quantities. The spectral measurements are treated as a transfer from a reference detector or source. Uncertainties are then calculated for effects arising in either the base reference or in the transfer measurement. Uncertainties of the spectral distributions are propagated through the integration process. Various systematic and random effects contributing to the uncertainty are treated separately and then combined into the final uncertainty. Photometric integrals are simple convolutions of two spectral distributions. Many systematic effects can be described by a single parameter. Then for fully-correlated effects, the uncertainty of each integral value is a linear sum of the signed spectral uncertainties, weighted by the appropriate convolving function. Further, pairs of similar integral values are also fully correlated and their covariance is simply obtained. For effects random between wavelengths, the variance of a spectral integral is a linear sum of the spectral variances, weighted by the convolving function. The covariance of integral pairs is the product of the linear sums used to calculate the uncertainty of systematic effects. The uncertainty of combinations of the spectral integrals is simply propagated from their variances and covariances. Representative examples are given for the luminous response of a reference photometer, the spectral-mismatch factor and the photometer quality factor f1'. iv 1 PHOTOMETRIC INTEGRALS While photometric quantities [1] are defined for different geometries, photometers are most commonly used to measure illuminance, and discussion here is in terms of illuminance response, calculated from spectral distributions. The principles used are directly applicable to other luminous quantities provided the response function is measured for the appropriate geometry. The integrals we deal with are photometric response, spectral mismatch factor, and quality factor of a photometer. 1.1 Photometric Response Suppose we have a photometer whose responsivity at wavelength is R(). The relative spectral response is a close match to V(, the relative spectral luminous efficacy function shown in Figure 1. The response of the photometer to a source whose spectral irradiance is E() is: RV k R( ) E ( )d (1) where k is a normalising factor. Most often the response is measured as a power response, not irradiance, and the factor k includes the aperture area. In many instances the response is measured only in relative terms, and k then includes various normalising constants. In discussing uncertainties here, we are concerned only with that of the integral sum itself and not the scaling factors. Components common to all wavelengths affect the absolute value of the integral sum and its uncertainty. Some common factors may be included in the scaling factor k and care must be taken not to include them twice in a complete response uncertainty analysis. The integral in Eq. (1) is important when determining the photometric response of a primary reference standard. This is a ratio of integrated response to the illuminance calculated as: EV km V ( ) E ( )d (2) where V( is normalised to 1 at 555 nm and km = 683 lm·W-1. If the photometer spectral response does not exactly match that of V(, the photometric response will depend on the source distribution. Hence most calibrations are given for response to the defined source CIE Illuminant A. This source has a relative power distribution of a black-body at a temperature of 2856 K, shown in Figure 1. Relative strength 1.2 1.0 0.8 0.6 Response Reference Transfer ratio Illuminant A 0.4 0.2 0.0 400 500 600 700 800 Wavelength /nm Figure1. Transfer calibration of a photometer against a silicon photodiode detector. The response function of the photometer is the V() spectral luminous efficacy function. Also shown is the spectral source distribution CIE Illuminant A NMI TR 9 1 1.2 Spectral Mismatch Factor Most photometers are calibrated by comparison to a primary reference using a source whose spectral energy distribution approximates that of CIE Illuminant A. They may then be used to measure the illuminance of sources with different spectral power distributions. If the photometer spectral response is not an exact match to V(, the illuminance measured by a photometer calibrated for illuminance response to CIE Illuminant A must be corrected by multiplying by the spectral mismatch factor. This is given by: E A ( )R( )d E ( )V ( )d i A (3) F ( E, E ) A E ( ) V ( ) d E ( ) R ( ) d The integrals in the numerator and denominator depending on the common values of measured spectral response R() are correlated, as are those depending on the common source spectral distribution measurements E(). These correlations must be considered when estimating the uncertainty of the spectral mismatch factor. This factor is also known as the colour-correction factor. While we have retained the symbol E for the source, denoting irradiance, it is only the relative spectral irradiance or spectral power distribution that is important in calculating correction factors. 1.3 Quality Factor of a Photometer This is calculated from the relative spectral response of the photometer and used to indicate how well the photometer matches the V( luminous efficacy distribution. It is defined as: R * ( ) V ( ) d ' f1 (4) V ( ) d R * ( ) where 2 E E A ( )V ( )d A ( ) R ( ) d R ( ) (5) PROPAGATION OF UNCERTAINTY We wish to estimate the collective effect that the uncertainty of a particular measured spectral value has on the uncertainty of each of the calculated quantities. Uncertainty propagation is described in detail in the ISO Guide to the Expression of Uncertainty in Measurement [2]. The uncertainty in a quantity X formed by combining measured quantities xi through the relationship X f ( x1 , x2 ,..xN ) is given by: 2 N 1 N f 2 f f u (X ) u ( xi , x j ) (6) u ( xi ) 2 i 1 xi i 1 j i 1 xi x j where u ( xi ) is the uncertainty in xi and u ( xi , x j ) is the covariance of xi and x j . For uncorrelated input quantities, the covariance of pairs of variables is zero and Eq. (6) reduces to: N 2 2 f 2 u (X ) u ( xi ) i 1 xi the ‘sum of squares’ commonly applied. The derivatives f xi are sensitivity coefficients for the dependence of X on the various measured quantities. N 2 NMI TR 9 (7) 2 If we form another quantity Y by combining the measured quantities xi through the relationship Y g ( x1 , x2 ,..xN ) , the uncertainty in Y is given by an expression similar to that of Eq. (6), but now the quantities X and Y are correlated through dependence on the common set xi. The covariance between X and Y is given by: N N f g u ( X , Y ) u ( xi , x j ) (8) i 1 j 1 xi x j Correlation coefficients are normalised covariance values, defined as: u( X , Y ) (9) r( X ,Y ) u ( X )u (Y ) We note that if r ( xi , x j ) 1 for all pairs, Eq. (6) reduces to: f (10) u( xi ) i 1 xi independently of the sign of the sensitivity coefficients. However, while uncertainty components of a spectral measurement, systematic across wavelengths, are fully correlated, the correlations may have mixed signs for different wavelength pairs; this is dealt with in the next section. N u( X ) Just as relative uncertainty is given as a ratio of the uncertainty of a quantity to its value, relative covariance is given by: u( X , Y ) urel ( X , Y ) (11) XY 3 UNCERTAINTY PROPAGATION BY COMPONENT OF THE SPECTRAL MEASUREMENT Individual sources of uncertainty are generally independent. Table 1 shows the two ways that uncertainties in spectral values due to these individual effects can be combined when the spectral values themselves are combined. Because the individual effects are not correlated, we can form the uncertainties of the combined effects at a given wavelength by sum-of-squares, Eq. (7). However, these spectral values are now most often partially correlated, and we must use the full expression of Eq. (6) to calculate the final combined uncertainty, u c . Table 1. Uncertainties, ui , j , in spectral values, Si , due to different effects, j, combined to estimate the uncertainty, u c , in the value of a colour quantity. The total uncertainty of Si at wavelength i is u ic and the uncertainty in the colour value due to effect j is u c | j Wavelength Effect 1 Effect 2 Effect m Combined 1 u1,1 u1,2 u1,m u1c 2 u2,1 u2,2 u2,m u 2c n un ,1 un ,2 un ,m u nc Combined u c |1 u c |2 u c |m uc NMI TR 9 3 The final combined uncertainty u c is also found by combining the u c | j values by sum-of-squares. The advantage of this form of combination is that many of the systematic effects that can contribute to uncertainty across wavelengths are either uncorrelated (random) -which means that the uncertainty for the combination across wavelengths is found by sum-of-squares- or fully correlated, for which simple linear sums can be used. It also allows estimation of the contribution of the individual effects to the final uncertainty. For random components, we have that the variance of a combination X of the spectral values ( xi Si ) is given by Eq. (7). The covariance of the spectral values at different wavelengths is zero, and Eq. (8) for the covariance between two spectral combinations X and Y reduces to: N X Y 2 (12) u( X , Y ) u ( Si ) i 1 Si Si Even though the spectral values are uncorrelated, the common dependence on the one set of spectral values correlates the integrals. Uncertainty components of the spectral measurement systematic across wavelengths are fully correlated between wavelengths, but we have to consider the sign of the correlation. Most uncertainty components (whether random or correlated between wavelengths) can each be described by a single parameter p as: (13) Si Si ( p) Components may have a wavelength dependence, but in the case of systematic effects, uncertainties at different wavelengths scale with the parameter p. Noting that uncertainties are always positive, the uncertainty of Si due to this effect alone is: u ( Si )| p Si u ( p) p S S sgn( i ) i u ( p) p p where sgn( ) is the sign function and Si/p is the sensitivity coefficient of the spectral value for the effect. (14) By substituting these forms into Eq. (6) we have: N X Si (15) u ( X )| p u ( p) i 1 Si p N X or (16) us ( X )| p us ( Si ) i 1 Si where the s subscript indicates that the uncertainty carries the sign of the effect. Similarly, by substituting into Eq. (8), the covariance between two combinations X and Y is given by: N N X Y (17) u( X , Y )| p u s ( Si ) u s ( Sj ) i 1 Si j 1 Si Note that combinations of fully correlated components of the spectral values are themselves fully correlated. The covariance values are a product of terms already calculated for the (signed) uncertainties of the individual combinations. NMI TR 9 4 The variances and covariances of the integrals for all effects are found as sums of those of the independent individual effects. 4 SPECTRAL MEASUREMENT AS A TRANSFER Spectral responsivity and spectral power distribution are measured by comparison to reference standards. Hence the spectral value Si can be written as a transfer from that of a reference standard, S iRef : Si ti SiRef (18) Uncertainties in the spectral value Si arise both from those of the reference value and those introduced by the spectral transfer. Photometric integrals are formed by combining the spectral measurements at different wavelengths. Systematic effects in a spectral measurement have a cooperative relationship between wavelengths, that is, the values at different wavelengths are correlated. Correlations also exist between the reference values at different wavelengths, arising from systematic effects in the methods used to derive them. For uncertainty components of the reference spectrum, from Eq. (18) we have: u ( Si ) ti u ( SiRef ) (19) whereas for components in the transfer we have: u ( Si ) SiRef u (ti ) (20) In both cases we assume that the system efficiency is constant and calculate the transfer value as the ratio, ti Si SiRef . 5 UNCERTAINTY COMPONENTS IN SPECTRAL MEASUREMENT For each independent component we calculate the transfer ratio at each wavelength and hence the spectral uncertainties using Eq. 20 or 19. We then propagate those uncertainties to the integral values, and from those to the desired integral combination. Discussions here are given in terms of a general spectral quantity Si at the ith wavelength. Uncertainty contributions arise from the uncertainty of the reference values, from offsets in the transfer, scaling effects in the transfer and wavelength effects in the transfer. Each of these effects can have random and systematic causes. Corrections are applied for measurable effects, but these corrections carry uncertainty. Other effects may be considered too small for correction, but there is still an uncertainty in their estimated value. 5.1 Base Uncertainty of the Reference Standard The spectral reference standard is likely to be measured at a limited number of wavelengths, with some systematic errors in the process. The reference values then will be at least partially correlated. In this case the full expressions of Eq. (6) must be used to calculate the integral uncertainties and correlations. Often the contribution of the reference standard can be calculated by taking the correlation coefficient as constant, at its average through the visible range [3]. Then this calculation can be split into fully-correlated and uncorrelated parts and the simpler expressions of Eqs (7) and (10) used to calculate the variance. At the highest levels of accuracy, systematic effects dominate the reference spectra uncertainties and they are highly correlated. NMI TR 9 5 The reference spectrum and the measured spectrum generally are defined at different wavelengths. To avoid introducing correlations through interpolation procedures [4], we use different procedures for reference and transfer uncertainty components. When dealing with reference uncertainty components, we interpolate the measured spectrum (which has no uncertainty if we are only considering a single component of the reference spectrum) to the wavelengths of the reference spectrum. When dealing with transfer uncertainty components, we interpolate the reference spectrum (which has no uncertainty if we are only considering a single component of the transfer) to the wavelengths of the measured spectrum. An alternative to dealing with different wavelength intervals in reference and measured spectra when the correlation coefficient for the reference uncertainties is taken as constant is to interpolate the reference spectrum and its uncertainties to the measurement wavelengths. Propagation of the systematic part of the reference uncertainties will be correct, but propagation of the random part will underestimate the reference contribution by the square root of the ratio of the data wavelength intervals; a correction factor is simply applied for this effect. 5.2 Offset Components of the Transfer Ratio Offsets in signals are additive effects. Hence: S p ti Refi (21) Si pR with p, pR 0 but u ( p), u ( pR ) 0 . It follows that for offsets in the signal channel: u s ( Si ) p u ( p ) and for offsets in the reference channel: u s ( Si ) pR (22) ti u ( p R ) (23) Offsets random across wavelengths arise from background noise in the sample and reference signal channels, which can be related to electronic noise in the measurements. 5.2.1 Combining Random Offset Components across Wavelengths Random offsets in the sample and reference channels can be treated separately, or combined in quadrature. The contribution of the two effects to the variance of a combination X is: 2 X 2 2 2 u ( X ) p , pR u ( p ) ti u ( p R ) S i 1 i and the contribution to the covariance between combinations X and Y is: N X Y 2 u( X , Y ) p, pR u ( p) ti2u 2 ( pR ) i 1 Si Si N 2 (24) (25) 5.2.2 Combining Systematic Offset Components across Wavelengths Scattered light in either reference or sample beam can cause an offset systematic across wavelengths, as can electronic offset in an amplifier. Such offsets have positive correlation between wavelengths, in a given signal channel. Offset effects in the reference and sample beams are not equivalent. For systematic offsets in the signal channel: NMI TR 9 6 X (26) u ( p) i 1 Si and for systematic offsets in the reference channel: N X (27) us ( X )| pR ti u( pR ) i 1 Si For each channel, the covariance between two combinations X and Y is given by the product of their signed uncertainties. N us ( X )| p Most systematic offset effects are uncorrelated between the sample and reference channels, and the variance of the two terms are separately added. However, an offset that is common to both channels is correlated. An example is incorrect background subtraction of room light reaching a detector from outside the signal paths. In such a case the (signed) uncertainties of Eqs (26) and (27) are fully correlated with a correlation coefficient of +1; then the two terms are added, and the result squared to obtain the variance in a combination due to this effect. 5.3 Scaling Components of the Transfer Ratio Scaling may arise from a number of instrumental effects. Whether the effect is in the reference or measured spectrum, we have: Si pSi' (28) where Si' is the uncorrected value. We have p 1 (if it is not a correction must be applied) and we estimate its uncertainty u(p), usually in relative terms. Hence: us (Si ) Siurel ( p) (29) 5.3.1 Combining Random Scaling Components across Wavelengths Random noise in the magnitude of the system transfer function is a scaling term. The noise arises from source fluctuations if comparing detectors, response fluctuations if comparing sources and fluctuations in the transmission of the optical system in both cases. For random scaling components, the variance of a combination X is given by: 2 X 2 2 u (X ) p Si urel ( p ) i 1 Si and the covariance between two combinations X and Y is given by: N X Y 2 2 u( X , Y ) p Si urel ( p) i 1 Si Si N 2 (30) (31) 5.3.2 Combining Systematic Scaling Components across Wavelengths Clipping of beams reaching reference and sample detectors through different optical paths, or variation in the distance setting of the reference and sample sources are examples systematic across wavelengths. Non-linearity is a form of scaling systematic over wavelengths, where the scaling factor depends on the spectral value. Corrections must be applied for the linear effects when the absolute value of the transferred value is important. Scaling factors affect the calibration of photometric response, which is an absolute value, but not the uncertainty of ratios such as the spectral mismatch factor. Correlations for systematic scaling factors are generally positive, although it is possible to conceive of non-linearity corrections that may vary NMI TR 9 7 in sign for different wavelength pairs. For scaling effects systematic across wavelengths, the (signed) uncertainty in a combination X is given by: N X (32) us ( X )| p Si urel ( p) i 1 Si and the covariance between two combinations X and Y is given by the product of their signed uncertainties. 5.4 Wavelength Components of the Transfer The uncertainty in spectral value arising from an uncertainty in the wavelength setting u(p) in the sample channel is: S u s ( Si ) i u ( p ) (33) and in the reference channel is: S Ref us ( Si ) ti i u ( p) (34) The derivatives are calculated numerically. 5.4.1 Combining Random Wavelength Offsets across Wavelengths Random errors in the wavelength setting arise from errors in the wavelength selection mechanism, from calibration procedures used to determine line centres, or imprecision in the knowledge of the calibration wavelengths themselves. The treatment of random wavelength effects depends on the method used to record spectra. In step-and-compare mode, the wavelength uncertainties at the one wavelength are correlated between channels; hence: S S Ref (35) us ( Si ) i ti i u ( p) If the reference and sample spectra are recorded separately over a range of wavelengths, the uncertainties at the one wavelength are uncorrelated; hence: 2 Ref S S (36) us ( Si ) i ti i u ( p) For random wavelength offsets, the variance of a combination X is given by Eq. (7) and the covariance between two combinations X and Y is given by Eq. (12), each with the appropriate form of u(Si). 2 5.4.2 Combining Systematic Wavelength Offsets across Wavelengths We may also have an offset in the wavelength setting. This can arise if the spectral lamp(s) used for calibrating the wavelength scale have different alignment to the broad-spectrum lamp used for measurement, or as an offset from the calibration process. The uncertainty treatment for each spectral value is identical to that for the random setting but now the uncertainties of different wavelengths are fully correlated. Note that the signs can vary throughout the spectrum, and the correlation between different wavelength pairs can be positive or negative. NMI TR 9 8 For scaling effects systematic across wavelengths, the (signed) uncertainty in a combination X is given by: N SiRef X Si (37) us ( X ) ti u ( p) i 1 Si and the covariance between two combinations X and Y is given by the product of their signed uncertainties. 5.5 Multiple Transfers from the Reference The spectral result of one measurement may be the reference for a subsequent one, as we progress from base standards through various levels of working standards. Here the base reference spectrum may be used to calculate uncertainties provided the transfer uncertainty components are combined for all the transfers. The usual rules for such combination apply; if the uncertainties for a given component are uncorrelated, the combined transfer uncertainty is found by sum-of-squares, Eq. (7). If the effects are correlated, they are added linearly, Eq. (10). Care must be taken here. For example, if the wavelength scale was recalibrated between transfers, and the lamp repositioned, the offset applicable for the second transfer is not correlated to that of the first, whereas it is if the system is undisturbed between transfers. 6 UNCERTAINTY ESTIMATES FOR THE PHOTOMETRIC INTEGRALS Uncertainties were estimated for the photometric integrals listed in section 1. The values of the various independent uncertainty components for detector and source measurements are shown in Table 2. These values were chosen as reasonable representation for typical measurements. Figure 1 shows the various distributions used. Transfer measurements were calculated at 5 nm intervals. Offsets in the transfer were set as fractions of the maximum signal in each channel, here taken as the maxima of the reference and sample spectra. Table 2. Responsivity uncertainty components used to calculate the representative examples Component Base reference – detector Reference signal offset noise Reference signal background offset (drift) Test signal offset noise Test signal background offset Source noise (combined over both channels) Absolute scaling Non-linearity Wavelength setting Wavelength offset NMI TR 9 Type Random Uncertainty Ideal photodetector, 0.1% relative uncertainty with a correlation coefficient of 0.3 between wavelengths 0.1%, relative to max. signal Systematic 0.01%, relative to max. signal Random Systematic 0.1%, relative to max. signal 0.01%, relative to max. signal Random 0.1%, relative Systematic Systematic Random Systematic 0.1%,relative 0.01%, relative to max. signal 0.03 nm 0.1 nm Random and systematic 9 Reference standards usually have relatively smooth spectral distributions and they are derived at limited wavelengths. The results shown here assume reference standards derived at 20 nm intervals. True integrals are required for data recorded at discrete intervals. While end-point integral effects can be ignored because the convolving function tends to zero, not all data are presented at uniform wavelength spacing. Hence in converting from integrals to sum forms we include a weighting i for each term. For N spectral points and trapezoidal integration: 1 (2 1 ) / 2 N (N N 1 ) / 2 i (i 1 i 1 ) / 2 6.1 (38) i 1 or N Photometric Response for Illuminant A The integral part of the response of a photometer to illuminant A is: (39) RV i Ri Ei Here we are concerned only with the contribution of the spectral measurement to the relative value of the integral. Because we are dealing with a broad-band source we use the V( distribution itself for the photometer spectral response The effect of the various independent components is shown in Table 3, along with their combined total. Refer to Table 2 for a more complete description of the uncertainty settings. Table 3. Relative uncertainty in response of a typical photometer to CIE Illuminant A arising from the components listed in Table 2 Component uncertainty Relative uncertainty (%) 0.1% (r = 0.3) 0.063 Reference signal offset noise 0.1% 0.028 Reference signal background offset (drift) 0.1% 0.015 Test signal offset noise 0.1% 0.055 Test signal background offset 0.01% 0.042 Source noise (combined over both channels) 0.1% 0.019 Absolute scaling 0.1% 0.100 Non-linearity 0.01% 0.010 Wavelength setting 0.03 nm 0.009 Wavelength offset 0.1 nm 0.051 Component Base reference - detector Total: NMI TR 9 0.151 10 6.2 Spectral Mismatch Factor For a perfect photometer the spectral mismatch factor is 1, independent of the source. Practical photometers can have a response significantly different from ideal at wavelengths where the photometer response is small. Also accurate photometric measurement of narrow-band sources such as quasi-monochromatic LEDs in those spectral regions may require smaller bandwidth and more closely spaced spectral measurements than broad-band sources – errors due to such effects are not considered here. To estimate this quantity and its uncertainty we need to know both the photometer spectral response and the source spectral power distribution. It is sufficient to have these in relative terms only, although in propagating uncertainties from an assumed reference standard we generate the uncertainty for the absolute spectral quantity, and the correlations between quantities at different wavelengths. We treat the measurements of spectral response and spectral power distribution as independent and separately calculate the uncertainty of the two ratios: i EiA Ri (40) FR i Ei Ri and i EV i i (41) FS i Ei Ri For each ratio, we estimate the uncertainties u ( N ) and u ( D) of the sums in the numerator and denominator, respectively, and their covariance u ( N , D ) . The relative uncertainty of the ratio is then given as: N 2 2 2 urel ( ) urel ( N ) urel ( D) 2urel ( N , D) (42) D Because the integrals in the ratios are similar, correlations between the numerator and denominator are strong. This is particularly true for the FS ratio of Eq. (41) because the relative spectral response of any practical photometer is close to a V() distribution, and as a consequence the contribution from source uncertainties is often negligible. Source distributions can be quite different, however, and correlations between the numerator and denominator of the FR ratio of Eq. (40) are reduced compared to those for FS. As the uncertainties of the source and response spectra are independent, the relative uncertainty in the spectral mismatch factor is found combining the relative uncertainties of the two ratios in quadrature. Again we can use V() as a representative photometer spectral response distribution when calculating uncertainties (but not the mismatch value, of course, as this depends on the response distribution itself). Table 4 shows results for uncertainties the spectral mismatch factor of a D65 source for a photometer calibrated against a CIE Source A. The uncertainty treatment is general; it can be applied to any two source distributions used for calibration and subsequent measurement. NMI TR 9 11 Table 4. Spectral mismatch factor uncertainty, by component, for a practical photometer and a source distribution matching CIE Illuminant D Component Component uncertainty Relative uncertainty (%) 0.1% (r = 0.3) 0.008 Reference signal offset noise 0.1% 0.007 Reference signal background offset (drift) 0.1% 0.000 Test signal offset noise 0.1% 0.036 Test signal background offset 0.01% 0.009 Source noise (combined over both channels) 0.1% 0.005 Absolute scaling 0.1% 0.000 Non-linearity 0.01% 0.000 Wavelength setting 0.03 nm 0.004 Wavelength offset 0.1 nm 0.088 Base reference - detector Source contribution (negligible) Total: 6.3 0.096 Quality Factor of a Photometer In practical terms, the quality factor of a photometer is an indicator, rather than an important photometric quantity. Propagation of the uncertainty for the quality factor is complicated first by the renormalisation of the reponsivity of Eq. (44) and secondly by the absolute value of the integrand in Eq. (4). A complete uncertainty propagation is presented here more out of interest than need. In sum terms we have: ' 1 f R V V * i i i i i Ri* Vi i VS (43) with the renormalisation given as: * i R E V E R A i i A i i i i Ri (44) Sensitivity coefficients for f1’ in terms of Ri are: A f1' f1' Ri* i i Ei Vi Ri Ri* Ri VS i EiA Ri E AR * 1 i i Ai , Ri Vi i Ei Ri i i EiAVi i EiA Ri * 1 (45) , Ri Vi A A VS i Ei Ri i Ei Ri 0, Ri* Vi Note that uncertainty at those wavelengths for which the normalised response equals V() do not contribute to uncertainty in f1' – this is confirmed by calculating sensitivity coefficients directly for positive and negative offsets in responsivity, NMI TR 9 12 where it is seen that the sensitivity changes sign with the offset and the net contribution is zero. This seems a non-physical result at first thought, because it implies that even for a large uncertainty in response, the contribution to uncertainty in f1' can be small. However, the larger the uncertainty, the less likely that the normalised response will equal V(). The sensitivity coefficient in fact changes sign within the uncertainty band for values of normalised responsivity close to V(). We allow for this effect by calculating an average sensitivity coefficient through the uncertainty band. This calculation is simplified by noting that the negative term in the brackets of Eq. (45) is small compared to 1 and can be ignored, and by assuming a rectangular probability distribution within the uncertainty band for a given component of the spectral responsivity values, at each wavelength. The average sensitivity coefficients then become: A f1' i i Ei Vi i EiA Ri V u ( R ) , R i i i i Ri VS i EiA Ri i EiAVi E R V u ( R ) V E R E V E R V R E V E V , otherwise V E R u(R ) i i EiAVi S i A i , Ri i i i A i i A i i i i i S i A i i A i i A i i A i i i i i i (46) i i By directly applying systematic positive and negative offsets to the responsivity data, we can show also that the shift in value of f1' is generally non-symmetric for systematic effects; this asymmetry is ignored here. The photometer response cannot be set to V() to estimate uncertainties, as the sensitivity coefficients all become zero. Table 5 shows the relative uncertainty in f1' calculated for a typical photometer with good matching at red wavelengths, relaxed at blue wavelengths, for which the f1' value was 2.99%. Table 5. Uncertainty in f1' for a typical photometer with f1' = 2.99% arising from the various components Component Base reference - detector Reference signal offset noise Reference signal background offset (drift) Test signal offset noise Test signal background offset Source noise (combined over both channels) Absolute scaling Non-linearity Wavelength setting Wavelength offset NMI TR 9 Component uncertainty 0.1% (r = 0.3) 0.1% 0.1% 0.1% 0.01% 0.1% 0.1% 0.01% 0.03 nm 0.1 nm Total: Uncertainty in f1' (%) 0.17 0.13 0.02 0.17 0.04 0.09 0.00 0.02 0.05 0.01 0.30 13 It can be seen from Eq. (44) that the f1' dependence on systematic scaling of the responsivity values is zero. This in fact can be used to check the accuracy of the uncertainty calculation. For the photometer used here, the propagated uncertainty was 0.1%; this component has been set to zero in the table. 7 CONCLUSION Systematic effects in the measurement system generally dominate the uncertainties of photometric integrals. Calculation of the uncertainty in for combinations of spectral integrals due to the various systematic effects is relatively straightforward. The principles explained here are also useful for estimating uncertainties in the various colour quantities [5]. Individual component uncertainties of combinations scale with their input values. The representative examples given here can be used as a guide to uncertainties to be expected from given experimental conditions. The progression of examples given here also demonstrates the principles of uncertainty propagation in increasing complexity. Photometric response is a single integral and only its uncertainty is required. The spectral mismatch factor requires correlations between integrals to be known. The sensitivity coefficients required to calculate f1' are complex due to taking the absolute value inside an integral. The various uncertainty components introduced by both systematic and random measurement processes can be treated in a consistent manner by applying the methods described in the ISO Guide; however, the f1' example is one where the firstorder calculation of the Guide is only approximate. 8 REFERENCES [1] The Basis of Physical Photometry (1983) Publication CIE 18.2, International Commission on Illumination, Vienna Guide to the Expression of Uncertainty in Measurement (1993) International Organisation for Standardisation JL Gardner (2003) Correlations in Primary Spectral Standards Metrologia 40, S167–S171 JL Gardner (2003) Uncertainties in Interpolated Spectral Data J. Res. Natl Inst. Stand. Technol. 108, 69–78 JL Gardner (2005) NMI TR 8 Uncertainties in Colour Measurements, National Measurement Institute [2] [3] [4] [5] NMI TR 9 14