Charge Diffusion Measurement in Fully Depleted CCD using

advertisement
Charge Diffusion Measurement in Fully
Depleted CCD using X-rays
Ivan Kotov, J. Frank, A. Kotov, P. Kubanek, P. O’Connor , V. Radeka, P. Takacs
Brookhaven National Laboratory, USA
Institute of Physics, Academy of Sciences, Czech Republic
Abstract: Tight requirements on the Large Synoptic Survey Telescope point spread function (PSF) demand sensor contribution to
PSF be both small and well characterized. The sensor PSF is determined by the lateral charge diffusion on the drift path from the
CCD window to the gates. Different techniques for charge diffusion characterization have been developed, each with its own
systematic and measurement difficulties. A new way to measure charge diffusion using an X-ray source is presented. We
demonstrate the effectiveness and limitations of our technique.
Introduction: Specialized CCD sensors are being developed for the LSST camera. They are n-channel, fully depleted, back
illuminated, 100 micron thick devices with pixel size 10 micron. First prototypes have been produced and tested. The LSST
sensors are operated in over depletion mode with an electric field up to 6kV/cm. In thick fully depleted CCDs, charge carrier
transport from the back window to the gates is accompanied by charge diffusion. The charge distribution has a 2D Gaussian
shape. The expected value of the characteristic diffusion width sigma in over depletion mode is 3.2-4.5 micron. The range of
sigma values is due to uncertainties in the behavior of the electron drift velocity and transverse diffusion coefficient at high fields
and low temperatures.
Numerous methods have been developed for lateral diffusion characterization in fully depleted Si detectors. A number of these
techniques have been applied for LSST prototype sensor characterization. The most straight forward technique, the virtual knife
edge, requires high quality optics, precision XYZ stages, and thermal and vibration control of the setup to mitigate systematic
errors. Other methods also require specialized equipment to perform the measurements and have their own sources of
systematic uncertainties as well.
As a part of our suite of test routine, 55Fe data are collected for gain, noise and charge transfer efficiency (CTE) measurements.
X-ray conversions happen at all device depths. The drift time and correspondingly the diffusion sigma depend on the conversion
point depth. X-rays produce compact clusters with conversions close to the gates and wide clusters originating from conversions
close to the window. The sigma value for charges generated at the window surface is used to characterize sensor's PSF. The
window side PSF defines the sensor performance in the visible spectrum where photons convert in the first few microns near the
window. The absorption length of 5.9 keV photons in silicon is ~30 micron. Thus, for a 100 micron thick device, the number of Xrays converted near the window is about 28 times higher than near the gates. The distribution of sigma values in 55Fe sample has
a peak at the ``window'' value, and it is very attractive to use these data for charge diffusion characterization. We have
developed the new method to obtain the characteristic diffusion sigma value from analysis of the charge distribution in the
clusters.
Method: The charge sharing between neighboring pixels depends on the charge cloud size at the gates and the position of the
conversion point in the pixel. The 2D Gaussian charge distribution can be described by 4 parameters: conversion point x- and y-position,
sigma, and total amplitude. These parameters can be determined for an individual X-ray if the cluster contains at least 4 pixels with
amplitudes above the noise (with fewer pixels only an upper estimate on the cluster width can be obtained). The accuracy of these
estimates depends on the signal to noise ratio of the cluster pixels. The low CCD noise allows for measurement of small diffusion sigma
even though the pixel size is 2-3 times larger than sigma.
Consider an example of an X-ray cluster containing 1620e- and positioned at the center of a 10 micron pixel and a diffusion sigma being
4.4 micron. The average signals are 897e- in the center pixel and 157e- in the 4 pixels surrounding the center. With a system noise below
10e-, the signal to noise ratio is more than 15 for these 5 pixels and reliable diffusion estimation is possible. Clusters with a sufficient
signal to noise ratio can be selected by setting a threshold amplitude to count a pixel as ``fired''. Compact clusters, less than 4 ``fired''
pixels, are eliminated from further analysis. To first order, this selection simply cuts off clusters with small diffusion and eliminates clusters
with conversion point positions inside sub pixel zones not sensitive to cluster width determination. The fact that compact clusters are not
used has no negative effects on analysis results when there are enough clusters with more than 4 ``fired'' pixels in the sample. To
investigate in full effects of the cluster selection and to study our analysis efficiency and accuracy for all possible sub pixel cluster
positions, an X-ray simulator was developed and measurements were compared with simulations.
Data were obtained exposing sensors to a 55Fe source. The total system noise for these exposures was 11e-.
The noise power spectrum density is shown in Fig.1. The base line subtraction algorithm performance is illustrated in Fig.2.
There are 6097 Ka,b clusters recorded in the sample and 97% of them are clusters with 4 or more ``fired'' pixels (see Fig.3). Xray clusters with the total amplitude in the range of Ka,b peaks were fit to a model function using MINUIT minimization package.
The model function is a 2D Gaussian integrated over the pixel area. The fit range usually is a 3x3 zone around the pixel with
maximum amplitude. In our case, this zone contains most of the cluster charge and is small enough to keep cluster pile up
negligible. The pixel amplitude error has contributions from both the system noise and signal fluctuations. Errors are calculated
for each pixel, taking both contributions into account. The fit was successful for 94.5% of the clusters. Thus the method total
efficiency is 92% for our data sample. The distributions of fitted parameters are shown in Fig.4. Measured diffusion sigma
values are shown in Fig.5. They peak at 3.6 micron.
Simulation: These measurements have been compared with simulations. Sets of
6000 K_a,b clusters with fixed sigma value and randomly distributed x- and ycoordinates have been generated. Distributions of simulated values are shown in Fig.6.
Simulated clusters have been embedded into zero exposure images and processed
through the same analysis pipeline as real 55Fe frames, see Fig.7. The outcome for a set
with sigma=3.55 micron is shown on Fig.5 in blue. The 95% of these clusters were fit
successfully. The reconstructed sigma values for this set peak at 3.6 micron and r.m.s. of
the distribution is 0.31 micron. Fitted parameters are shown in Fig.8.
Fig.3 X-ray amplitude and cluster size distributions for (left to right): all clusters; one pixel
clusters; two pixel clusters; three pixel clusters; four pixel clusters; five or more pixel
clusters; cluster size. Raw data have been processed using sparse signal base line
subtraction algorithm.
Fig.4 Fitted parameters: total amplitude; x-, y- coordinates and chi2 of the fits.
Conclusion: Measurements and simulation are in good agreement for the ``window''
peak. The characteristic diffusion value for these measurements is estimated
as sigma = 3.55 micron. The statistical accuracy is estimated using the r.m.s. of the blue
histogram as ~0.01micron. More extensive simulations are planned for systematic
uncertainties study.
Fig.6 Simulated: total amplitude; maximum amplitude and local x-, y- coordinates.
Fig.2 Results of application the base line subtraction algorithm to this data set.
Fig.5. The distribution of sigma values measured with a prototype
device is shown in green. The simulated X-rays with diffusion
sigma value fixed at 3.55 micron is shown in blue. Pixel size is
10micron.
Fig.7 Simulated clusters processed through the analysis pipeline. X-ray
amplitude and cluster size distributions for (left to right): all clusters;
one pixel clusters (histogram is empty); two pixel clusters; three pixel
clusters; four pixel clusters; five or more pixel clusters; cluster size.
Fig.1 Noise power spectral density
Fig.8 Fitted parameters for simulated clusters: total amplitude; x-, ycoordinates and chi2 of the fits.