Polaris Q GC/MSn Ion Trap Technology Steven T. Fannin GC & GC/MS The Column: “heart” of the Instrument 2 Maintaining GC/MS Ruggedness “Extra Column” Effects • Syringes • Septa • Liners • Ferrules • Gas Filters 3 Chromatography: General Overview • Resolution Why Capillary Columns? • Selectivity – Spacing between two peaks – Important role in GC confirmation analyses • Capacity Factor (Relative Retention) The Van Deemter equation: H = A + B/u + C u A: the multipath term (eddy diffusion) B: longitudinal diffusion C: resistance to mass transfer H = A + B/u + C u – Retention relative to an unretained compound • Column Efficiency: H = L/N N.B: Velocity: Pressure regulated vs Flow controlled H2 vs He vs N2 4 Common Mass Analyzers for GC/MS • Time of Flight (TOF) - Ionized compounds/fragments from the source are directed into a flight tube. Ions are separated by virtue of their different flight times over a known distance. • Magnetic Sector - Uses a combination of magnetic and electrical fields to sort ions. The ions are focused and resolved by passing through an electric field then a magnetic field. • Quadrupole - consists of two sets on opposing rods. This mass analyzer uses a combination of RF and DC modulation to sort ions. • Ion Trap - operates on a principle as the quadrupole; however ions can be stored for subsequent analysis. The ions are sorted by changing the electric field inside of the trap by manipulating the RF field and sequentially ejecting the ions from low to high mass to charge. 5 General Mass Spectrometry Characteristics What differentiates mass analyzers is how they perform mass analysis • Mass Analysis - Common to Mass Analyzers – All determine the m/z ratio – All measure gas-phase ions – All operate at low pressure (<10-4 Torr) to allow appropriate mean free path of gas phase ions • General Mass Spectrometry Instrument Characteristics – Sensitivity – Tandem Mass Spectrometry – – – – 6 Mass Range Resolution Mass Accuracy Scan Speed GC/MS Ionization Methods Electron Ionization: EI (“Hard Ionization”) • Transfer of energy to a neutral molecule (in the gaseous state) to eject one of its own electrons and produce an ion (charged molecule), with a mass of m and a charge of z. 8 Example of PFTBA EI+ 9 Chemical Ionization Soft Ionization Techniques Filament EI Ion Volume e- To Mass Analyzer CH4 Removable Ionization Volume 10 Lenses CI Ion Volume Positive Ion Chemical Ionization Reagent gas reactions (methane) m/z 16, 15, 14 m/z 17 m/z 29 m/z 28 m/z 27 m/z 41 11 Positive Ion Chemical Ionization Proton transfer M CH M H CH 4 5 M C 2 H 5 M H C 2 H 4 [M+1]+ Hydride abstraction M C 2 H M H C 2 H 6 5 [M-1]+ Adduct formation M C 2 H 5 M C 2 H 5 [M+29]+ M C 3 H 5 M C 3 H 5 [M+41]+ 12 Common PICI Reagent Gases Reagent Gas • Methane [CH5+ & C2H5+] – – • • Low purity (ion source gets dirty quickly) Anhydrous ammonia [H+(NH3)n=1-3] – – – – Very selective protonation (nitrogen compounds) Forms [M+NH4]+ adduct with many compounds Keeps ion source clean Highly corrosive (short mech. pump lifetime) reaction must be exothermic, i.e., PA (analyte) > PA (reagent gas) * kJ/mol 13 • 549 & 687 Protonates most organic molecules C2H5+ reacts with alkanes primarily by hydride abstraction Isobutane [C4H9+] – Proton Affinity* Hydride Ion Affinity* • 1126 & 1135 • 821 • 976 • 858 • 825 Less fragmentation with higher PA Less [M-H]+ with lower HIA EI vs.PICI for Pesticides EI Spectrum of Heptachlor Intensity is low for any single m/z ion. 14 PICI Spectrum of Heptachlor Intensity is concentrated in [M+H]+ ion. Spectrum is simpler. Adduct Formation in PICI 15 Negative Ion Chemical Ionization (EC-NICI) • Reagent gas reactions (methane) 4 CH4 (70eV )e CH e * Thermal electron • Kinetic energy of electrons reduced by collisions with reagent gas • Resonance electron capture mechanism of ionization AB e * AB Heat [M]- • Reagent gas reacts with electrons to form “plasma” of thermal electrons • Ionization is favored by molecules which have a high electron affinity – electron capture • Useful for selective analysis in heavy matrices, e.g., pesticides in food or waste matrix. 16 Common NICI Reagent Gases e- Thermalization Rate* Reagent Gas • Methane • 8.6x10-10 • Isobutane • ~2.1x10-9 – Low purity (ion source get dirty quickly) • Carbon dioxide – Can produce less fragmentation than methane or isobutane • Anhydrous ammonia • 5.8x10-9 • 5.9x10-9 – Keeps ions source clean – Highly corrosive (short mech. pump lifetime) * cm3/s 17 Better sensitivity with higher rate NICI of Carboxy THC - PFPA Negative Ion spectrum of the PFPA/PFPOH derivative of 11-nor-9-Carboxy-D9-THC 18 Analysis of Catecholamines using NICI-MS 753 Pentafluoropropionyl (PFP) Derivatives of Norepinephrine, Epinephrine and Dopamine 19 Ion Trap vs Quadrupole Basic Principles Voltage Relationship During a Mass Scan (Quadrupole) +1500 +Vdc VRF 0 VRF +180° -Vdc RF Potential DC Potential +250 • Ions scanned by varying the DC/Rf voltage across the quadrupoles -1500 -250 +/-(U+Vocoswt) Complete Mass Scan 77001-1380 970608 m/z Ion beam 21 -/+(U+Vocoswt) What is a Quadrupole Ion Trap? Entrance Endcap Ring Electrode ro zo V cost 22 Exit Endcap Potential Energy Surfaces (Ion Traps) R ing V oltage (V ) 150 100 50 0 -50 -100 -150 0 V 90 180 270 360 RF Phase (deg) V V r z r 23 z r z General Principles of Stability Diagrams • Operating line for mass selective instability – 0.4 z stability 0.2 bZ aZ 0.2 0.3 0.4 0.5 0.6 1.0 0.8 0.7 0.1 0.2 0.3 0 Basic Ion Trap Principles q cut-off~.91 0.4 qz 0.5 Mathieu stability diagram and stability/reduced parameters • Ion trap function : “Mass Selective Instability” • Quadrupole: Mass Selective Stability mode of scanning • The (a,q) coordinates are simply related to m/z and the operating voltage - whereas b values are related to ion motion bX Y , -0.2 0.6 r stability U V cost 0.7 -0.4 0.8 Operating line for Mass selective stability U0 0.9 1.0 -0.6 qz = az = 8eV m(r O2+ 2 Z O2 ) 2 16eU 2 0.5 24 2 m(r O+ 2z O) 2 q Z 1.0 1.5 ro zo az 0 Stability Line and Mass Selective Ejection az Mass-selective Instability Scan with Resonant Ejection 476 kHz 0.908 0.0 * qz V m/z az 0 25 qz 1.0 Mass-selective Instability Scanning Ramp RF voltage (V) to sequentially eject ions from low m/z to high m/z. Trapping Injected Ions V cost + Dz eV 2 2 4mz o 2 ro zo + + + + ++ • Correct RF voltage • Helium buffer gas 26 Full Scan MS Scan Function Mass Analysis Mass Analysis Ion Injection Ion Injection V Gate Lens Eject (V’) Multiplier AGC Prescan* One complete scan constitutes a “microscan” 27 Mass Analysis Scan Fixed Ion Injection Time Number of Ions Trapped (arbitrary units) 100000 Space Charge Effects 10000 1000 100 10 1 1 10 100 1000 10000 100000 Amount of Sample (arbitrary units) 28 DYNAMIC RANGE: ~103 1000000 10000000 Ion Injection Time Optimized with AGC Number of Ions Trapped (arbitrary units) 100000 Space Charge Effects 10000 1000 * User Selectable 100 ** 5 µs PolarisQ, 10 µs LCQ, 30 µs GCQ 10 1 1 10 100 1000 10000 100000 Amount of Sample (arbitrary units) DYNAMIC RANGE: >106 29 1000000 10000000 Polaris Q Tune Parameters (AGC and Injection RF) 30 Quadrupole vs. Ion Trap Quadrupole quadrupoles use SIM to enhance sensitivity Transmits one m/z ion at a time Mass-Selective Stability scanning Ion Trap In full scan ion traps are more sensitive than quadrupoles. Trap all m/z ions simultaneously Mass-Selective Instability scanning 31 Quadupoles and Sensitivity Quadrupole • Duty cycle is important for determining mass analyzer efficiency • Efficiency of the mass analyzer: Transmits one m/z ion at a time Mass-Selective Stability scanning Duty Cycle for a Quadrupole Width of transmitted ion total width of m/z range 32 = Duty Cycle EMassAnalyzer ETransmission DutyCycle • Ionization and mass analysis occur simultaneously: Mass resolution and scan range are important when determining duty cycle SIM, MIM and SRM,MRM (Target Compound Techniques) • Single Quadrupole Technology (single-stage MS techniques) – SIM (Selected or Single Ion Monitoring) • Set quadrupole to pass a single characteristic ion during a retention time window in the chromatogram • Increases sensitivity 10-100X • Lose spectral specificity – MIM (Multiple Ion Monitoring) • Monitor 2 to 5 characteristic ions in addition to SIM quanitiation ion • Set acceptable qualifier ion “ratios” to confirm detection • More qualifier ions boost confidence but reduce sensitivity gains • Triple Quadrupole Technology (MS/MS Techniques) – SRM (Single Reaction Monitoring) • Single product ion monitored – MRM (Multiple Reaction Monitoring) • Multiple product ions monitored 33 Ion Traps and Sensitivity External Source Ion Trap • Efficiency of the mass analyzer: EMassAnalyzer ETransmission DutyCycle Trap all m/z ions simultaneously Mass-Selective Instability scanning Duty Cycle for an Ion Trap Ion Accumulation Time (ion gate time) Total scan time 34 = Duty Cycle • Ionization and mass analysis occur consecutively: Scan time (or rate) relative to ion accumulation is important for determining duty cycle Tandem MS Principles Tandem Mass Spectrometry Why use MS/MS? • Enhanced Selectivity (Qualitative and Quantitative) – TRACE Analyses Criteria for Target Compounds • Sensitivity and Selectivity are important – MS/MS Improves Trace Level Analyses in complex matrices and enhances confirmatory analyses (Enhanced confirmation of identification) • Combined with Soft Ionization techniques – Most signal in [M+H]+ ions; Added selectivity and s/n – Confirmatory assays (MW ions plus 2-3 unique ions) – Qualitative and quantitative with digital reagent gas flow • Structural Characterization Applications – MS/MS provides unique evidence to an unknowns identity providing further information about fragments in the MS spectrum S/N TRACE DSQ uses SIM to increase S Polaris Q uses MS/MS to reduce N 36 MS/MS “Tandem-In-Time”Ion Trap Technology MS/MS and MSn Capability MS/MS “Tandem-In-Space” Triple Stage Quadrupole Technology 37 77001-1383 970602 MS/MS “Tandem-In-Time” Ion Trap Technology MS/MS in an Ion Trap 1. Inject 2. Isolate 38 3. Fragment 4. Detect How Do We Isolate Ions for MS/MS? az Ion we wish to isolate Mass-selective Instability Scan with Resonant Ejection 476 kHz 0.908 0.0 * qz 1.0 V How MS/MS works m/z qz Mp = VRF 77001-1383 970602 az 0 39 qz υ(ion) = (n + β) Ω/2 Isolation Waveforms m/z 1000 Fast Fourier m/z 300 Transform m/z 100 Time Domain 40 Frequency Domain CID using Resonant Excitation qz 0.0 0.908 How MS/MS works t=15 ms qz Mp = VRF υ(ion) = (n + β) Ω/2 Product ions qz 0.0 41 0.908 Polaris Q Excitation Event Characteristics Stability Diagram: Where parent ions reside on the q axis during the excitation event Excitation “q” 0.225 qz 0.300 0.450 P = precursor mass qz V m/z MS/MS qz Mp = VRF 42 The choice of ‘q’ is also a function of the MS/MS lower limit of the product ion m/z range. A ‘q’ of 0.225 is 1/4Mp (where Mp is the m/z of the parent ion), and a ‘q’ of 0.3 is 1/3Mp, and a ‘q’ of 0.45 is 1/2Mp. For example, if a ‘q’ of 0.45 is used, and Mp is m/z 400, then the daughter ion lower limit that can be observed in the spectrum will be m/z 200. If the same ‘q’ is used for an Mp at m/z 800, then the daughter ion lower limit that can be observed will be m/z 400, and so on Higher qz Means Higher Energy D z eV 4mz 2 o Dz qz = 0.225 43 2 mq z 2 2 2 z 2 o 16e 0.30 0.45 Resonant Excitation qz Value Fragment ions not trapped qz 0.0 1/4 x 1/3 2x 1/2 4x 0.908 0.225 qz 0.0 Product Ion Fragmentation m/z Range Energy 0.30 0.908 qz 0.0 44 0.45 0.908 Tandem MS: Polaris Q MS/MS Scan Function Mass Analysis Ion Isolation Ion Injection Ion Injection Mass Analysis Ion Isolation Resonant Excitation V Gate Lens Isolate Excite Eject (V’) How MS/MS works for all RF-Traps qz Mp = VRF υ(ion) = (n + β) Ω/2 45 Multiplier AGC Prescan Mass Analysis Scan MS/MS Example - Chlordane GC/MS Spectrum Isolation of Precursor Ion 46 GC/MS/MS Product Ion Spectrum Fragment Precursor Ion Polaris Q MS/MS Parameters MS/MS Parameters Choice of Excitation q’s 47 MS/MS: Optimizing Conditions ‘q’ is a function of the RF voltage applied to the ring electrode during excitation Dexamethasone Product ion intensity vs Collision Voltage 7.E+05 Excitation Event – higher excitation ‘q’s may provide improved conversion efficiencies (ECID = Fi / P0) Product Ion Intensity 6.E+05 5.E+05 4.E+05 3.E+05 2.E+05 1.E+05 0.E+00 0 0.2 0.4 0.6 0.8 1 1.2 Collision Voltage (p-p) 48 1.4 1.6 1.8 2