Session 8 - Display Systems - my Tri

advertisement

Display Modes

A-Mode or Amplitude Mode

B-Mode (Brightness Mode)

M-Mode (Motion Mode)

A-Mode

a one-dimensional display or image each pulse produces a new line of information on the display temporal resolution = PRP an uncommon display, except in ophthalmologic sonography used for precise intraocular length measurements

A-Mode

Height of the spike is proportional to the amplitude

A-Mode

Depth… horizontal axis corresponds to the reflector’s depth or distance

B-Mode (Brightness Mode)

- basis for gray scale,

two-dimensional (2D) imaging

US unit tracks the position of the transducer to place a dot on the screen corresponding to the transducer position (X, Y locations), creating a 2D image

B-Mode (Brightness Mode)

Each pulse from the transducer creates a single scan line from a series of returning echoes

A complete scan line resulting from one emitted pulse. occurs in < 1/1000 of a second (< 1 msec)

Scan Line

One pulse of ultrasound generates a single scan line (from a series of returning echoes).

A complete scan line resulting from one emitted pulse. This is accomplished in < 1/1000 of a second.

Echoes are presented in sequence on a scan line as they return from tissue.

(A) The first echo is displayed. (B) The second echo is added. (C) More echoes are added. (D) All the echoes from a single pulse have been received and displayed as a completed scan line.

2

3

4

0

1

B-Mode

A shade of gray is assigned to the amplitude of the echo

Stronger echo amplitude

= brighter dots

Vertical axis represents depth

M-Mode (Motion Mode)

- one-dimension image used to investigate moving structures with respect to time

Temporal resolution = PRP; each pulse produces a new line of echo information on the display evaluates motion pattern of moving structures such as in the heart

P

T

D

E

H

M Mode

A dot records echo position in relation to time

(horizontal axis) with the vertical axis representing depth

Echo amplitude is represented by the dot’s brightness

TIME >>>>

B Mode

M Mode

A Mode

Scanning Imaging

Static scanning (B scan)

Real-time

Static Scanning (B scan)

An articulated arm scanner scans the patient from many different directions creating a 2Dimage from multiple B-mode pulses

Multiple dots are combined to delineate the echo pattern of internal structures within the body

Superimposition of multiple scan lines creates a two-dimensional image that portrays the general contour of the patient and the internal organs

Compound B-mode scanning produces a static image that can be thought as a stop-action photograph of the reflecting surfaces

B Scan

B Scan

Static Scanner

B Scan

Real-time

Produces a video giving the impression of constant motion of the scanned anatomy

Consists of a series of frames displayed in rapid sequence creating the impression of constant motion

Provides rapid, convenient image acquisition with the display changing continuously as the scan plane is moved through the tissues

Real Time

Who’s that pregnant with twins??

A real time image before computer technology could handle increased lines/frame & faster frame rates.

How many shades of gray can you see?

Temporal Resolution

Resolution related to time & motion

Time from the beginning of one frame to the beginning of the next one (the time required to generate one complete frame)

Expressed in milliseconds (ms)

Ability to accurately determine the position of a structure at a particular instant in time

Depends on extent of movement of the structure & the frame rate

Important in imaging rapidly moving structures

Temporal Resolution

Each frame is made of many scan lines; when you  # of scan lines - the frame rate 

Improves as the frame rate increases (a greater # of frames/ second) because less time elapses from one frame to the next

Temporal Resolution

Depends upon 2 factors:

1.

# of images created/second (frame rate)

- higher frame rates (greater number of frames created/second), the better the temporal resolution. To  temporal resolution, frame rate must be 

2.

Higher frame rates are needed to evaluate motion or moving structures, such as adult, pediatric and fetal hearts

Spatial (detail) resolution

ability to see detail on the image affected by the # of scan lines & focuses

 # of scan lines  spatial resolution

More detail is needed to scan organs in the body so a slower frame rate is tolerated

When temporal resolution  , spatial resolution  !

Scanning Speed Limitation

Real-time scanning consists of multiple frames/second that are made up of multiple scan lines per frame

Its advantage is temporal resolution

To create each scan line, the ultrasound unit must wait until all echoes have been received from the selected depth before sending out the next pulse, if not range ambiguity occurs

Range Ambiguity

When structures beyond the indicated range are depicted in an image

Cause – time between the transmitted pulse & detected echo is not coreectly measured

Occurs when an echo (from the previous pulse) is received after the next pulse is transmitted

Scanning Speed Limitation

US unit can’t work any faster than the sound wave can travel, so propagation speed plays a major role in limiting the scanning speed

The # of focuses that the sonographer uses while imaging will slow down the process of obtaining a scan line

Imaging depth controls determine when the next pulse is sent out (PRF)

The # of foci used in imaging will slow down the process of obtaining a scan line

Imaging depth controls determine when the next pulse is sent out (PRF)

Frame Rate

# of frames/second (fps).

Human eye can see flickering (each individual frame being produced) at frame rates < 15 -20 fps

Acceptable frame rates are 30 fps-60 fps

Factors affecting frame rate are:

• depth of field

• # of lines used to create the image

• # of focal zones used

Sound travels 1,540 m/s (154,000 cm/s) in ST

A pulse can travel to & from a depth of 77,000 cm/s

Imaging with multi focus- and annular arrays

(multiple pulses to various depths to create a single composite scan line) requires even more time

Creating a single frame with a large # of scan lines requires TIME. Presenting many frames in rapid sequence requires TIME.

Frame Rate

To avoid misplacing the proper location from returning echoes on the display: imaging depth (cm) X lines/frame

X pulses/line [number of focal zones]

X frame rate  77,000 cm

Frame Rate

Consider the following scenario:

You are imaging a liver that extends to

10 cm deep with a 5.0 MHz probe. What is the maximum PRF permitted to avoid range ambiguity?

10 cm X 13 μs/cm = 130 μs - which means 1 pulse/130 μs

PRF = # pulses/sec = 1 pulse/130 μs = .0077 pulses/μs

7,700 pulses/sec = 7.7 KHz PRF = 7.7 KHz

Maximum Frame Rate

= c = PRF

2 X Distance X # lines/frame [lpf] lpf

= 7.7 KHz = 77 fps

100

Imaging depth

# of pulses (foci) per line lines per frame frame rate

ALL OF THE ABOVE battle over time

Therefore, a compromise to balance these factors must be based on meeting the clinical need.

PRF = # focuses X lines/frame X frame rate

PRF ↑ when you ↑ any of the following:

# of focuses

# of lines per frame

frame rate

Consider this:

FRAME RATE ↓ with:

↑ # of focuses

↑ # of lines per frame

↑ in scanning penetration

Frame Rate

Solution depends on area of interest:

Imaging depth

Multiple focal zones

Line density

Frame rate

Imaging Depth

Complete depth that sound travels per pulse

Controlled by the sonographer to visualize the anatomy to be imaged that may lie superficially or deep in the body

The deeper the system images, the longer the listening time for each pulse

Deeper imaging results in:

longer listening time longer pulse repetition period lower PRF (# pulses/second) more time for each scan line

Multiple Focal Zones

US pulse has only a single focal zone

(region within the beam that provides the finest lateral resolution)

Using multiple sound beams with different focal depths to create a single image line, results in optimal lateral resolution all depths  superior image quality

Multiple Focal Zones

A pulse is required for:

– each focus

– each scan line

– each frame

– More foci/image line = more pulses/image line

Multiple focal zones are controlled by the sonographer & are only used with phased array transducers (linear, curved, and annular)

More foci per image line result in:

more pulses/line superb lateral resolution at all depths more time/image scan line more time needed to create a frame

Line Density

# of scan lines that create a single image

Set automatically by the US system & is not controlled by the sonographer

The greater the line density, the more pulses/image sector

Increased Line Density results in:

greater detail within the image less “space” between image lines more sound pulses per image more time needed to create a frame

Line Density

For a sector scan: lines/degree

For a rectangular scan: lines/ cm

Frame Rate

determined by the US system & is not directly controlled by the sonographer when a rapidly moving structure is imaged at an unsuitably low frame rate, the images are said to ‘flicker’

More Frames per Second result in:

1.

greater accuracy in locating moving structures

2.

less time to make each frame

3.

decreased line density

THE DILEMMA

SO... to optimize these conditions: adjust the maximum imaging depth to the area of interest

Determine the # of foci per scan line.

Superior lateral resolution over a range of depths requires more foci. This determines the number of sound pulses required to make each scan line.

THEN…

The frame rate & line density are determined by the ultrasound system to balance the goals of temporal resolution (frame rate) & image quality (line density).

Deeper Imaging

Depth longer listening time longer pulse repetition period lower PRF (# of pulses/second)

Amount of time long

THE DILEMMA

Multiple Focal

Zones more sound pulses/line longer

T/R time superb lateral resolution at all depths more time per image scan line more time required for each scan line more time needed to create a frame

Amount of time long

Higher Line

Density greater spatial detail in the image

 space between image lines

 real detail more sound pulses/image

 listening time more time needed to create a frame

Amount of time long

Higher Frame

Rate greater accuracy in locating moving structures

 temporal resolution decreased line density; less detail resolution

Less imaging depth due to speed limitation less time allocated to make each frame/scan line

Amount of time short/frame; long with high frame rates

Recording Techniques

Before an image is recorded, the contrast

& brightness controls of the monitor are adjusted for optimal image quality

Contrast & brightness controls of the recording devices are then matched to the image on the monitor.

– Adjusting the contrast & brightness controls of the monitor without changing the contrast & brightness controls of the recording device, can lead to producing films that appear with increased gain or decreased gain, i.e., too light or too dark

Recording Devices

Hard copy film & paper

Thermal processors

Laser imaging systems

Digital recording devices videotape player

M-mode records

Magneto-optical (MO)

Picture Archiving & Communication

Systems

Teleradiology

Hard Copy Film and Paper

Transparent film uses a single emulsion film

Film is a cellulose acetate sheet coated with a gelatin emulsion containing silver bromide crystals

After exposure to light from the camera monitor, the film is developed in a chemical processor

Changes in the temperature of the chemicals will affect the quality of the film

Age & concentration of the chemicals can also affect the film

Thermal Processors

uses a paper medium to record the image less resolution & shades of gray than single emulsion film less archival stability

Color Thermal Printers

contain a ribbon of colored inks

(black, cyan, yellow, & magenta) printing on photo paper produces good quality color images used for color Doppler hard copy imaging

Laser Imaging Systems

capable of higher resolution, better gray scale uniformity & less image distortion

Initial cost outlay automated film handling & developing – saves sonographer processing time

15+ images available per sheet of film

Digital Recording Devices

- store the image on computer disks or in computer memories for viewing on monitors or later transfer to film

Videotape Player

used to record moving or real-time images record color as well as black & white images tapes store the image information on a magnetic medium

Follow recommended procedures when handling and storing tapes

Most common format is VHS

Super VHS (S-VHS) has better spatial resolution, less distortion & stores more information

Fiber optic recorder

(M-mode recorders)

paper developed by exposure to visible light dry silver paper as recording medium

-better gray scale

Trend

digital archiving & image storage

Magneto-optical (MO)

Magneto-optical (MO) is also known as optical technology. It is a combination of optical & magnetic technologies

Stores lots of information onto disk

(optical portion). MO drives can be rewritten and erased

Not susceptible to magnetic fields like regular diskettes or digital tapes so information storage is safe

Picture Archiving &

Communication Systems (PACS)

a computer network for the acquisition, display, & storage of images. provide for electronically communicating images to work stations, devices, and storage external to the instrument, the examining room,

& even the building where the scanning is done. protocols for communicating images and associated information between imaging devices and workstations have been standardized in the

Digital Imaging & Communications in Medicine

(DICOM) Standard

Picture Archiving &

Communication Systems (PACS)

Other names are digital imaging network

(DIN) & information management archiving and communications systems (IMACS)

Acquisition, display, hard copy, & computer components must be interconnected using a local area network (LAN)

Allows digitized images from multiple imaging modalities to be stored for later retrieval, display, manipulation & interpretation

Picture Archiving &

Communication Systems (PACS)

US scan data is digitized & transferred to the network. Standards for encoding patient file information developed by ACR

(American College of Radiology) & NEMA

(National Electrical Manufacturers Assoc.)

Provides for more centralized processing by reducing the need for multiple hard copy devices for each ultrasound or imaging modality system

Teleradiology

- is the electronic transfer of images from one location to another

Allows multiple hospitals to have quick access to images

Transmission methods include telephone lines, coaxial cables, fiber optic cables, microwave dishes, laser systems, satellites, & T-1 (multiple) telephone lines

Digital information is compressed and converted to transmission methods such as pulsed tone signals for sending over telephone lines

Download