RapidArc plan verification using ArcCHECK™ Dr Craig Edwards University Hospital of North Staffordshire The Cancer Centre, Stoke-on-Trent, Staffordshire, ST4 6QG. England The University Hospital of North Staffordshire (UHNS) Radiotherapy Physics; Moved to the new Cancer Centre in June 2009 Cancer Centre features; Eclipse treatment planning system 4 Varian iX linear accelerators HDR Brachytherapy Unit An Acuity simulator GE lightspeed CT scanner IMRT and RapidArc at UHNS Dynamic IMRT 2005 Dynamic IMRT 2010 Commissioned on 600C & 2100C Varian linacs ~ 15 patients per year (2:1 H&N:prostate) 4 matched 2100iX Varian linacs ~ 175 patients per year (8:1 prostate:H&N) have treated uterus, cordoma, breast RapidArc 2011 Attended Varian training course MSc Project Commissioned patient QA system First clinical patient summer 2011 IMRT IMRT = Intensity Modulated RadioTherapy Modulate incident photon beam Highly conformal technique Mould dose around structures Spare radio-sensitive structures Increase delivered dose Arc Therapy - RapidArc • • • • An advanced form of IMRT Uses 177 control points Gantry arm rotates 358° Three forms of modulation Gantry speed: 0.5 – 4.8 °/s Leaf speed: 0 - 3 cm/s Dose rate: 0 - 600 MU/min Patient QA measurements • IMRT patient Every field’s dose fluence measured and compared Gamma analysis used to evaluate (3mm/3%) Point dose measurement compared with expected (+/- 5%) • Arc Therapy Replicate conventional IMRT ? System Choice • Several available ArcCheck Delta4 MatriXX / cube Octavius • Selection – ArcCheck Light Single cable set-up Absolute dose Potentially 3D DVH mapping looked right - cylindrical!! Rotation QA system • ArcCheck Cylindrical with diameter of 26.6 cm Water equivalent 1386 diode detectors Diodes at a depth of 2.9 cm Diodes 1 cm apart Arranged helically Central cavity insert Performance Tests Response reproducibility Response linearity with dose Dose rate dependence Consistency of chamber response with orientation • Stability of calibration with time • • • • Calibration Varian 2100iX linear accelerator Operating energies of 6 and 10 MV Dose rate 600 MU.min-1 Field size 10 x 10 cm2 ArcCheck centre positioned at isocentre (SSD 86.7cm) • 200 MU delivered • Dose at detectors • • • • • 6 MV – 2.47 Gy (1.32 Gy at isocentre) 10 MV – 2.60 Gy (1.54 Gy at isocentre) Response reproducibility • Method Irradiated central diode 100 MU, 10 x 10 cm2 field 5 irradiations Record response & compare with planning system Determine coefficient of variation (CV) • Results Mean reading Comparison with Eclipse Standard deviation: CV: 129.71 cGy - 0.15 % 0.13 cGy 0.10 % Response linearity with dose • Method Irradiated central diode Varied delivered dose from 10 to 500 MU 10x10 cm2 field Record response & compare with planning system Determine CV Response linearity with dose • Results Comparison with Eclipse: max diff 0.3% (@10 MU) Linearity of response (10 MV) above 100 MU, diff -0.1% rec orded dos e (c G y ) 700 600 When 500 normalised against 100 MU response; 2 400 Standard deviation: 0.38 cGy 300 CV: 0.29 % R = 1.00 200 100 0 0 100 200 300 delivered MU 400 500 Dose Rate Dependence • Method Irradiated central diode 100 MU, 10 x 10 cm2 field 100 to 600 MU.min-1 Record response & compare with planning system Determine coefficient of variation (CV) Dose Rate Dependence • Results Dose rate dependence Comparison with Eclipse; 130.0 response (cGy) 129.8 Maximum difference 129.6 Difference at 600MU.min-1 129.4 -1.4% (@100 MU.min-1) -0.1% 129.2 Standard deviation: 129.0 0.70 cGy 128.8 128.6 CV: 0.54 % 128.4 128.2 128.0 0 100 200 300 dose rate (MU.min-1) 400 500 600 Dependence upon irradiated angle • Method Irradiated central diode 100 MU, 10 x 10 cm2 field ArcCheck rotated to each angle Record response Determine coefficient of variation (CV) • Results Mean reading Standard deviation: CV: 129.52 cGy 0.68 cGy 0.50 % Stability of calibration • Method Irradiated central diode 100 MU, 10 x 10 cm2 field Response recorded every 20 mins for 2 hours Determine coefficient of variation (CV) • Results Mean reading Standard deviation: CV: 129.72 cGy 0.23 cGy 0.18 % Clinical tests • Verification of IMRT treatment • Verification of RapidArc treatment Verification of IMRT • Method ArcCheck CT scanned Image used as template for verification plan Export the treatment plan to ArcCheck image Export re-calculated fluences MapCheck Expected dose fluence compared to measured for every field and the sum of all fields Fields evaluated using gamma calculation with setting of 3mm/3% Verification of IMRT • Results Gamma evaluation pass rate 98.5 % Point dose evaluation measured -0.2 % Verification of RapidArc • Method Same as IMRT except there is only 1 field!! Verification of RapidArc • Results – – Gamma evaluation shows 99.4 % pixel pass rate Point dose evaluation measured -0.7 % Conclusions • The ArcCheck dosimetry system; is simple and quick to set-up has a quick and efficient software solution to calibrate diodes in absolute dose detectors demonstrated response characteristic which would make it a suitable dosimetry tool is an excellent solution for the verification of both IMRT and RapidArc treatment