Dosimetry Metrology for IMRT Jean M. Moran, Ph.D. Outline

advertisement
Outline
Dosimetry Metrology for
IMRT
Jean M. Moran, Ph.D.
•
•
•
•
•
Acceptance testing
Detectors for commissioning
Phantoms
Dosimetry analysis tools
Commissioning tests
– Varying complexity and geometry
AAPM Summer School
June 2003
• Potential Pitfalls
• Summary
The Department of Radiation Oncology
University of Michigan
JMM03 2
Acceptance
Collimator Designs
Exit Window
• Prior to purchase and installation,
review manufacturer’s acceptance
tests
• If needed, adapt tests and tolerances
with manufacturer in purchase order
• Test basic functionality of equipment
• Tests may be dependent on the MLC
design
Exit Window
Upper
Jaws
MLC
Diaphragm
Upper
Jaws
Lower
Jaws
MLC
Lower
Jaws
MLC
Isocenter
Isocenter
Elekta
JMM03 3
Exit Window
Isocenter
Siemens
Varian
JMM03 4
Physical MLC Characteristics
Field
Elekta
Siemens
Varian
40x40 cm 2
40x40 cm 2
(40x27 cm 2)
40x40 cm 2
Rounded
Divergent
Rounded
Width
Leaf ends
Effect of Rounded Leaf Ends & Linear Motion
Linear leaf motion
Uncorrected
Corrected
Cold
Xmlc
Leaf width
1 cm
1 and 6.5 cm
0.5 and 1 cm
Length
32.5 cm
31 cm
16 cm (14.5
cm Carriage)
Inter -
No
No
Yes
digitation
=1 cm gap
Uniform
Correction
Hot
JMM03 5
Xlight
Xrad
JMM03 6
Graves 2001
Leaf position offset = Xrad -Xlight
1
MLC Transmission
Static MLC Testing
• Carriage skew
– Orientation of MLC carriages in the accelerator
head
– Determined by measurement with feeler
gauges for very small field size
– Important to prevent collisions especially for
systems with interdigitation of leaves
– Verify with film
• Dependent on
system design
• Values range from
1.5 to 3% for midleaf
and interleaf
transmission
• Alignment of each leaf bank to central axis
• Leaf position reproducibility and accuracy
Film exposed under closed leaves
Evaluate compared to open field
– Verify with graph paper
JMM03 7
JMM03 8
Tongue and Groove Effect
Uniform Square Field Illustration of TG Effect
Caused by overlap of leaves and leaf design
?
Area in Red Never
Directly Exposed
JMM03 9
+
JMM03 10
Cold Spots from T&G Effect
~25% cold spot where
leaves overlap
70
square field
small leaves
60
large leaves
Dose (cGy)
50
40
30
20
10
0
-2.5
-2.0
-1.5
-1.0
-0.5
0.0
0.5
Position (cm)
JMM03 11
1.0
1.5
2.0
2.5
DMLC and/or SMLC Mode Tests
• Leaf speed
• Dose rate evaluation
• Leaf position tolerance and
reproducibility
• Leaf acceleration and deceleration
• Rounded leaf tip transmission
• Beam stability (output, flatness,
symmetry, linearity)
• Interrupted treatments
JMM03 12
2
Equipment and Software for IMRT
Commissioning
Example Test: FenceTest
Multiple incorrect leaf positions
• Ideal measurement system:
• All leaves move
across field and
deliver dose to
small field gap (0.1
or 0.2 cm)
• Sensitive to errors
of 1 mm in leaf
position
– Excellent spatial resolution, accuracy, tissue
equivalent response, provide 3-D data,
portable to multiple phantoms, and easy to use
• Multiple detectors required
– Verifying IMRT fields is more complex than
static fields
– Profiles along major axes in a water phantom
are insufficient for characterizing systems
– Profiles are also time-consuming to obtain in
water phantom for IMRT delivery
JMM03 13
JMM03 14
Equipment
Detector Characterization
• Linearity
• Depth dose and profiles in water
– Ion chamber – required for absolute dose
measurements
– Ion chamber or linear diode array
– Note: Depth dose curve measurements must
be made with a repeat delivery for each
datapoint on the curve (delivery stability)
• Detector arrays are useful for profile
measurements in water but 2-D detectors
are still required
• 2-D measurement methods
• Energy dependence
• Stem and cable effects
• Angular response
• Calibrated if for absolute
measurements
• Small field detectors
required for small field
characterization
– Sensitive to position
– Detector should be smaller
than homogeneous region of
dose to be measured
– Film
– Detector arrays
JMM03 15
JMM03 16
1-D Detector Characteristics
Detector
1-D Detectors
Measurement
Volume (cm 3)
Sensitive
Area
(cm2 )
Diameter
(cm)
Thickness
(cm)
Effective Point of
Measurement (cm)
Micro -chamber
0.009
0.24
0.6
NA
p-type Si diode
0.3
0.49
0.4
Stereotactic
diode
NA
0.011
0.015
0.010
Pinpoint
chamber
Dose
(cGy)
70
60
50
40
30
20
10
DETECTOR
DISADVANTAGES
0.2
Micro -chamber
Poorer resolution than diodes
0.06
0.6
p-type Si diode
0.45
0.006
0.07
Stereotactic
diode
0.2
NA
0.06
Pinpoint
chamber
Over-respond to low energy photons
Martens et al. 2000
MOSFET
NA
0.04
NA
0.1
NA
MOSFET
Non-linear dose response for <30 cGyChuang et al 2002
Diamond
0.0019
0.056/0.073
0.73
0.026
0.1
Diamond
< resolution than diodes, expensive, Rustgi et al, Laub et al
JMM03 17
JMM03 18
3
MOSFET System
MOSFET Consistency
4.0
• Excellent spatial
resolution
MOSFET1
MOSFET2
MOSFET3
3.0
2.0
• Automatic and
immediate readout
1.0
Bias Box
0.0
• Can be re-used
immediately
• Linear dose response
Reader
-1.0
-2.0
-3.0
MOSFET
• Response
independent of depth
-4.0
0
C. Chuang, UCSF
JMM03 19
TLDs
JMM03 20
2
4
6
8
10
12
14
16
Number of Measurements
18
20
Cynthia Chuang, UCSF
TLD Holder for Phantom Measurements
• TLDs must be characterized
– Time consuming
• Reusable
• Achievable accuracy: 2-3%
• Automatic reader required for multiple
TLD measurements
D.A. Low et al. “Phantoms for IMRT Dose Distribution Measurement and
Treatment Verification, Int J Radiat Oncol Biol Phys 40: 1231-1235 (1998).
JMM03 21
JMM03 22
2-D Detectors
• Film
Radiographic Film
• Advantages
– Radiographic: XV and EDR
– Radiochromic
• Beam imaging system, CCD, SLIC,
AMFPI
– Excellent spatial resolution
– Readily available
– Less expensive than other 2-D systems
• Disadvantages
– Over-response to low energies
– Dependent on QA of film batch
– EPID systems attached to gantry
– Investigated more for pre-treatment QA
currently
– Dependent on processor and digitizer QA
– Sensitive to storage conditions
– Need to measure the response to dose for
each experiment
JMM03 23
JMM03 24
4
XV vs EDR Film
Depth Dependence of EDR Film
6MV-EDR2-Depth corrected
6MV-EDR2-Regular
3.5
1400
Optical Density
Net Optical Density
3.0
2.5
2.0
XV2, 6 MV
1.5
XV2, 15 MV
1.0
1000
800
600
400
EDR2, 6 MV
200
EDR2, 15 MV
0.5
0
0
0.0
0.0
100.0
200.0
300.0
400.0
500.0
Chetty and Charland “Investigation of Kodak EDR film for
megavoltage photon beam dosimetry” PMB 47: 3629-3641 (2002).
Radiochromic Film: Advantages
• Decreased sensitivity to low energy
photons compared to radiographic
film
• No processing
– Film changes color with irradiation
• Insensitivity to visible light
• High spatial resolution
Niroomand-Rad et al. “Radiochromic film dosimetry: recommendations of
AAPM Radiation Therapy Committee Task Group 55,” Med Phys 25: 2093-2115.
JMM03 27
50
100 150
200 250 300 350
Dose (cGy)
600.0
Dose (cGy)
JMM03 25
1200
Dogan et al “Comparative evaluation of Kodak EDR2 and XV2 films for verification
of intensity modulated radiation therapy,” PMB 47: 4121-4130 (2002).
JMM03 26
Radiochromic Film: Disadvantages
• Non-uniform film response
– Can be minimized by using double-exposure
technique
• Response dependent on time and
temperature
• More expensive than radiographic film
• Digitizer response is dependent on the
light source of the digitizer and may need
to be modified
Niroomand-Rad et al. “Radiochromic film dosimetry: recommendations of
AAPM Radiation Therapy Committee Task Group 55,” Med Phys 25: 2093-2115.
JMM03 28
Active Matrix Flat Panel Dosimeter
(AMFPD)
Example: SMLC Delivery
AMFPD
Film
512 x 512 pixels
0.508 mm pitch
26.0 x 26.0 cm2
a-Si array
Direct Detection
-No fluorescent screen
Preamplifiers
AD converters
Non-commercial system
AMPFD at 10 cm depth in solid water
El-Mohri, et al. “Relative dosimetry using active matrix flat-panel
imager (AMFPI) technology,” Medical Physics 26, 1999.
JMM03 29
JMM03 30
184 mu
600 mu/min
d=10 cm
5
Gel Dosimetry: Advantages
Example: SMLC Delivery
4
Film
AMFPD
160
120
80
40
0
184 mu
600 mu/min
d=10 cm
Position (cm)
2
0
150
130
-2
20 30 50 70
80
100 120
-4
-4
-2
0
2
• Obtain 3-D information in one
irradiation
• Gels can be prepared with different
density therefore ideal for
heterogeneous measurements
• Gel can be used in containers of
different shapes
• Ideal for anthropomorphic phantoms
4
Position (cm)
B4
JMM03 31
JMM03 32
Gel Dosimetry: Disadvantages
• Sensitive to many factors such as time,
preparation, temperature
• Optical reader requires cylindrical container
for gel
• MR time is often limited and expensive
(unless dedicated scanner)
– Long scan times are required to increase
accuracy of readout
– E.g. 5% accuracy over 10 hr scan time (Gum et al
2002)
• Interface of gel and container results in less
accurate readout at edges of the gel
• Not ready for routine use in the community
JMM03 33
Gel Dosimetry
MR Dose Maps IMRT Plan
Slice 1
Slice 2
Slice 3
Dose %
JMM03 35
Dose
Deviation %
0 20 40 60 80 100 120
-60 -40 -20 0 30 40 60
Gum et al. “Preliminary study on the use of an inhomogeneous anthropomorphic
Fricke gel phantom and 3D magnetic resonance dosimetry for verification of IMRT
JMM03 34 plans” PMB 47: N67-N77 (2002).
treatment
Considerations for Phantoms
• Fiducials for reproducible setup of
phantom and detectors
• Flexibility to accommodate different
detectors
• Simple vs. anthropomorphic
• Homogeneous or heterogeneous
MR Dose Maps
- IMRT Plan
Water Tank
• Restricted to gantry at 0 degrees
– Unless mylar window for 90 degrees
• Flexibility in chamber position
• Important for basic depth dose and
profile measurements
• Output, flatness, symmetry, and
linearity assessment
JMM03 36
6
Water-equivalent Plastics
Simple Geometric Phantom
• Flat phantoms with custom chamber
inserts
20 cm
– Accommodate film at multiple depths
– Detector position only varies with depth
• Cylindrical phantoms (plastic or water
filled)
– Ion chamber at single position
– Possibly hold films
• Can be customized to hold various
detectors and have multiple positions
JMM03 37
Van Esch et al. “Acceptance tests and QC procedures for the clinical implementation of
IMRT using inverse planning and the sliding window technique: experience from five
radiotherapy departments,” Radiother Oncol 65: 53- 70 (2002).
JMM03 38
Cylindrical Phantom:
Ion Chamber and MOSFETs
IMRT Verification Phantom
Calc. 1.37 Gy
Meas. 1.42 Gy
Diff –3.52%
D.A. Low et al. “Phantoms for IMRT Dose Distribution Measurement and
Treatment Verification, Int J Radiat Oncol Biol Phys 40: 1231-1235 (1998).
JMM03 39
C. Chuang, UCSF
JMM03 40
Spiral Phantom for Dosimetric Verification
Calc. 0.81 Gy
Meas. 0.78 Gy
Diff –3.45%
RPC Head Phantom
Target Volumes
Removable Dry
Insert
Water
Water
Critical Structure
Paliwal et al “A spiral phantom for IMRT and tomotherapy treatment
delivery verification” Medical Physics pp. 2503-2507 (2000).
JMM03 41
JMM03 42
TLDs in Target Volumes
Radiochromic film through multiple plans
Delivery is required by RTOG for participation in IMRT trials
7
Anthropomorphic Phantom for Gel Dosimetry
Dosimetry Analysis Software
• Transfer patient fluence maps and beam
geometry to phantom geometry
• 2-D dose difference displays with
colorwash
• DVHs
• Highlight of differences
• Gamma analysis
– % agreement and distance-to-agreement
Gum et al. “Preliminary study on the use of an inhomogeneous anthropomorphic
Fricke gel phantom and 3D magnetic resonance dosimetry for verification of IMRT
treatment plans” PMB 47: N67-N77 (2002).
JMM03 43
JMM03 44
Dosimetric Analysis:
Overlay of Isodose Lines
Phantom Measurements
• Calculate IMRT plan for each field in
measurement geometry
• Film measurement at depth for individual
field at gantry=0
70
60
50
20
10
70 cGy
60 cGy
50 cGy
20 cGy
10 cGy
Film at depth in
water equivalent
plastic
JMM03 46
Dosimetric Analysis:
Dose Difference Display
Dosimetric Analysis : Dose Color Wash
Calculation
Individual Field
Film Data
Individual Field
Dose
(cGy)
70
60
50
40
30
20
10
JMM03 47
Calcs
Film
Intensity Map
Calcs
Film
JMM03 45
cGy
cGy
cGy
cGy
cGy
Intensity Map
JMM03 48
+/- 10%
Differences due to tongue -and-groove
8
Dosimetric Analysis:
Dose Difference and DTA
Dosimetric Analysis: Gamma Evaluation
Dose
Reference
Both
3%/3 mm
5%/5 mm
Evaluated
Unmod
Spatial
3% Dose
Difference
JMM03 49
3 mm
DTA
D. A. Low – Washington University
D. A. Low – Washington University
JMM03 50
Simple Geometry Tests
• Leaf position reproducibility in
dynamic or step-and-shoot mode
• Effect of gravity on leaf position
accuracy and reproducibility
• Sweeping gap test
• Fence test
JMM03 51
Simple Geometry Tests
• Leaf speed stability
• Leaf acceleration/deceleration
• Output checks for small to large fields
– Including smallest field size – 1 x 1 cm 2
• MLC limits (field size restrictions)
• Depth dose curve measurements
JMM03 52
Simple Geometric Tests
Geometric Tests
L. Xing et al. “Dosimetric verification of a commercial inverse
treatment planning system” PMB 44: 463-478 (1999).
JMM03 53
JMM03 54
9
Geometric Tests
Acceptance tests and quality control (QC) procedures for the clinical
implementation of intensity modulated radiotherapy (IMRT) using inverse
planning and the sliding window technique: experience from five
radiotherapy departments Van Esch et al Radiotherapy Oncology 65: 5355
70JMM03
(2002).
Complex Cases – Simple Geometry
Evaluate dose across field as a function of regional beamlet intensity
JMM03 56
Complex Geometry:
Anthropomorphic phantom
Potential Pitfalls
• Limit number of monitor units per
segment to verified Linac behavior
• Limit field width for IMRT
• Overshoot-undershoot phenomenon
• Incorrect tolerance value
• Integration with record-and-verify
system
M. A. MacKenzie et al. “Dosimetric verification of inverse planned step and
shoot MLC fields from a commercial planning system,” J Appl Clin Med Phys
3: 97-109 (2002)
JMM03 57
JMM03 58
Potential Pitfalls: Prior to Measurements
Detector Issues
• Verify all equipment is functioning
properly
Volume
– Film processor, digitizer
– Detectors, cables, electrometers (automatic
leakage correction)
– TLD reader, ovens
• Input/output to treatment planning system
• Standardize measurement setup when
possible
• Monitor software and hardware changes
and QA
JMM03 59
Dose (cGy)
JMM03 60
10
Detector Position
– Example: CAX of Varian MLC is a junction
of four 0.5 cm wide leaves
Beam 1 Ratio 9/21/02
1.20
Ratio
(Calculation/Measurement)
• Small ion chambers are very sensitive
to position
• Position should be considered with
respect to MLC design
Ratio Calc/Ion Chamber - CAX
Beam 1 Ratio - 9/25/02 20 seg
Beam 1Ratio 9/25/02 100 seg
Beam 2 Ratio 9/21/02
1.15
1.10
1.05
Beam 2 Ratio 9/25/02 20 seg
Beam 2 Ratio 9/25/02100 seg
1.00
0.95
0.90
regular
spike=cax spike=high spike=low
Spikiness values on part of field
JMM03 61
JMM03 62
Ratio Calc/Ion Chamber -0.5, -0.5 cm from CAX
Effect of Leaf Position Offset on IMRT
Ratio (Calculation/Measurement)
SMLC Single Field 30 Fractions
1.20
Beam 1 Ratio - Off-axis - 20 seg
Beam 1 Ratio off-axis - 100 seg
Beam 2 Ratio Off-axis - 20 seg
Beam 2 Ratio Off-axis-100 seg
1.15
1.10
1.05
1.00
0.95
0.90
0.85
regular
spike=cax
spike=high
spike=low
Spikiness values on part of field
Cadman et al “Dosimetric considerations for validation of a sequential IMRT
Process with a commercial treatment planning system” PMB: 3001-3010 (2002).
JMM03 63
JMM03 64
Effect of Leaf Position Offset on IMRT
Overshoot Phenomenon:
Example Varian System SMLC Delivery
Planned
No leaf offset correction
-3-12% errors
Given
With leaf offset corrected
+/- 5%
Planned
Given
Planned
Given
Cadman et al “Dosimetric considerations for validation of a sequential IMRT
Process with a commercial treatment planning system” PMB: 3001-3010 (2002).
JMM03 65
Ezzell and Chungbin, “The overshoot phenomenon in step -and -shoot IMRT
Delivery,” J Appl Clin Med Phys 2: 138-148 (2001).
JMM03 66
11
MLC Dynamic Log File (Dynalog)
System Communication Delay
Accelerator
Control Console
MLC
Workstation
Accelerator
Control
•
•
•
•
•
MLC
Controller
Varian 21 EX
Expected and actual leaf positions
Beam hold-off events
Position tolerance setting
Recorded approximately every 55 ms
• Information can be compared to imported
2-D information
~ 0.055 second
JMM03 67
JMM03 68
DMLC: Effect of Incorporating Machine
Limitations
B1
Deviation in Leaf Trajectory
BEFORE
AFTER
44 mu, tolerance 0.1 mm
no gap: 137 beam hold-offs
54 mu, tolerance 0.25 mm
gap 1.1 mm: No beam hold-offs
Position (cm)
-3.0
-3.5
Expected Trajectory
Actual Trajectory
MLC Beam Hold-Off
-4.0
-4.5
-5.0
1.0
Hold0.0
0.0
0.2
0.4
0.6
0.8
1.0
1.2
1.4
Time (sec)
JMM03 69
JMM03 70
Summary
• Basic MLC characteristics should be
tested in static mode prior to IMRT testing
• IMRT tests should be specific to delivery
mode and device
• Be aware of potential issues with delivery
systems that may need further
investigation
• Multiple detectors and phantoms are
typically required for IMRT commissioning
• Quantitative dose analysis tools are
required for proper evaluation of delivery
JMM03 71
Case1, B1
Summary
• Measurements may show dosimetric –
mechanical differences that planning
systems may not model at this time
• Need to know the limits of the
mechanical systems and combination
of software + delivery
• Continued need to improve software
for delivery system, measurement
devices, phantoms, and dose analysis
tools
JMM03 72
12
Acknowledgements
University of Michigan
Dale Litzenberg
Jeff Radawski
Tim Nurushev
Benedick Fraass
Dan Low – Washington University
Cynthia Chuang - UCSF
JMM03 73
13
Download