Outline Dosimetry Metrology for IMRT Jean M. Moran, Ph.D. • • • • • Acceptance testing Detectors for commissioning Phantoms Dosimetry analysis tools Commissioning tests – Varying complexity and geometry AAPM Summer School June 2003 • Potential Pitfalls • Summary The Department of Radiation Oncology University of Michigan JMM03 2 Acceptance Collimator Designs Exit Window • Prior to purchase and installation, review manufacturer’s acceptance tests • If needed, adapt tests and tolerances with manufacturer in purchase order • Test basic functionality of equipment • Tests may be dependent on the MLC design Exit Window Upper Jaws MLC Diaphragm Upper Jaws Lower Jaws MLC Lower Jaws MLC Isocenter Isocenter Elekta JMM03 3 Exit Window Isocenter Siemens Varian JMM03 4 Physical MLC Characteristics Field Elekta Siemens Varian 40x40 cm 2 40x40 cm 2 (40x27 cm 2) 40x40 cm 2 Rounded Divergent Rounded Width Leaf ends Effect of Rounded Leaf Ends & Linear Motion Linear leaf motion Uncorrected Corrected Cold Xmlc Leaf width 1 cm 1 and 6.5 cm 0.5 and 1 cm Length 32.5 cm 31 cm 16 cm (14.5 cm Carriage) Inter - No No Yes digitation =1 cm gap Uniform Correction Hot JMM03 5 Xlight Xrad JMM03 6 Graves 2001 Leaf position offset = Xrad -Xlight 1 MLC Transmission Static MLC Testing • Carriage skew – Orientation of MLC carriages in the accelerator head – Determined by measurement with feeler gauges for very small field size – Important to prevent collisions especially for systems with interdigitation of leaves – Verify with film • Dependent on system design • Values range from 1.5 to 3% for midleaf and interleaf transmission • Alignment of each leaf bank to central axis • Leaf position reproducibility and accuracy Film exposed under closed leaves Evaluate compared to open field – Verify with graph paper JMM03 7 JMM03 8 Tongue and Groove Effect Uniform Square Field Illustration of TG Effect Caused by overlap of leaves and leaf design ? Area in Red Never Directly Exposed JMM03 9 + JMM03 10 Cold Spots from T&G Effect ~25% cold spot where leaves overlap 70 square field small leaves 60 large leaves Dose (cGy) 50 40 30 20 10 0 -2.5 -2.0 -1.5 -1.0 -0.5 0.0 0.5 Position (cm) JMM03 11 1.0 1.5 2.0 2.5 DMLC and/or SMLC Mode Tests • Leaf speed • Dose rate evaluation • Leaf position tolerance and reproducibility • Leaf acceleration and deceleration • Rounded leaf tip transmission • Beam stability (output, flatness, symmetry, linearity) • Interrupted treatments JMM03 12 2 Equipment and Software for IMRT Commissioning Example Test: FenceTest Multiple incorrect leaf positions • Ideal measurement system: • All leaves move across field and deliver dose to small field gap (0.1 or 0.2 cm) • Sensitive to errors of 1 mm in leaf position – Excellent spatial resolution, accuracy, tissue equivalent response, provide 3-D data, portable to multiple phantoms, and easy to use • Multiple detectors required – Verifying IMRT fields is more complex than static fields – Profiles along major axes in a water phantom are insufficient for characterizing systems – Profiles are also time-consuming to obtain in water phantom for IMRT delivery JMM03 13 JMM03 14 Equipment Detector Characterization • Linearity • Depth dose and profiles in water – Ion chamber – required for absolute dose measurements – Ion chamber or linear diode array – Note: Depth dose curve measurements must be made with a repeat delivery for each datapoint on the curve (delivery stability) • Detector arrays are useful for profile measurements in water but 2-D detectors are still required • 2-D measurement methods • Energy dependence • Stem and cable effects • Angular response • Calibrated if for absolute measurements • Small field detectors required for small field characterization – Sensitive to position – Detector should be smaller than homogeneous region of dose to be measured – Film – Detector arrays JMM03 15 JMM03 16 1-D Detector Characteristics Detector 1-D Detectors Measurement Volume (cm 3) Sensitive Area (cm2 ) Diameter (cm) Thickness (cm) Effective Point of Measurement (cm) Micro -chamber 0.009 0.24 0.6 NA p-type Si diode 0.3 0.49 0.4 Stereotactic diode NA 0.011 0.015 0.010 Pinpoint chamber Dose (cGy) 70 60 50 40 30 20 10 DETECTOR DISADVANTAGES 0.2 Micro -chamber Poorer resolution than diodes 0.06 0.6 p-type Si diode 0.45 0.006 0.07 Stereotactic diode 0.2 NA 0.06 Pinpoint chamber Over-respond to low energy photons Martens et al. 2000 MOSFET NA 0.04 NA 0.1 NA MOSFET Non-linear dose response for <30 cGyChuang et al 2002 Diamond 0.0019 0.056/0.073 0.73 0.026 0.1 Diamond < resolution than diodes, expensive, Rustgi et al, Laub et al JMM03 17 JMM03 18 3 MOSFET System MOSFET Consistency 4.0 • Excellent spatial resolution MOSFET1 MOSFET2 MOSFET3 3.0 2.0 • Automatic and immediate readout 1.0 Bias Box 0.0 • Can be re-used immediately • Linear dose response Reader -1.0 -2.0 -3.0 MOSFET • Response independent of depth -4.0 0 C. Chuang, UCSF JMM03 19 TLDs JMM03 20 2 4 6 8 10 12 14 16 Number of Measurements 18 20 Cynthia Chuang, UCSF TLD Holder for Phantom Measurements • TLDs must be characterized – Time consuming • Reusable • Achievable accuracy: 2-3% • Automatic reader required for multiple TLD measurements D.A. Low et al. “Phantoms for IMRT Dose Distribution Measurement and Treatment Verification, Int J Radiat Oncol Biol Phys 40: 1231-1235 (1998). JMM03 21 JMM03 22 2-D Detectors • Film Radiographic Film • Advantages – Radiographic: XV and EDR – Radiochromic • Beam imaging system, CCD, SLIC, AMFPI – Excellent spatial resolution – Readily available – Less expensive than other 2-D systems • Disadvantages – Over-response to low energies – Dependent on QA of film batch – EPID systems attached to gantry – Investigated more for pre-treatment QA currently – Dependent on processor and digitizer QA – Sensitive to storage conditions – Need to measure the response to dose for each experiment JMM03 23 JMM03 24 4 XV vs EDR Film Depth Dependence of EDR Film 6MV-EDR2-Depth corrected 6MV-EDR2-Regular 3.5 1400 Optical Density Net Optical Density 3.0 2.5 2.0 XV2, 6 MV 1.5 XV2, 15 MV 1.0 1000 800 600 400 EDR2, 6 MV 200 EDR2, 15 MV 0.5 0 0 0.0 0.0 100.0 200.0 300.0 400.0 500.0 Chetty and Charland “Investigation of Kodak EDR film for megavoltage photon beam dosimetry” PMB 47: 3629-3641 (2002). Radiochromic Film: Advantages • Decreased sensitivity to low energy photons compared to radiographic film • No processing – Film changes color with irradiation • Insensitivity to visible light • High spatial resolution Niroomand-Rad et al. “Radiochromic film dosimetry: recommendations of AAPM Radiation Therapy Committee Task Group 55,” Med Phys 25: 2093-2115. JMM03 27 50 100 150 200 250 300 350 Dose (cGy) 600.0 Dose (cGy) JMM03 25 1200 Dogan et al “Comparative evaluation of Kodak EDR2 and XV2 films for verification of intensity modulated radiation therapy,” PMB 47: 4121-4130 (2002). JMM03 26 Radiochromic Film: Disadvantages • Non-uniform film response – Can be minimized by using double-exposure technique • Response dependent on time and temperature • More expensive than radiographic film • Digitizer response is dependent on the light source of the digitizer and may need to be modified Niroomand-Rad et al. “Radiochromic film dosimetry: recommendations of AAPM Radiation Therapy Committee Task Group 55,” Med Phys 25: 2093-2115. JMM03 28 Active Matrix Flat Panel Dosimeter (AMFPD) Example: SMLC Delivery AMFPD Film 512 x 512 pixels 0.508 mm pitch 26.0 x 26.0 cm2 a-Si array Direct Detection -No fluorescent screen Preamplifiers AD converters Non-commercial system AMPFD at 10 cm depth in solid water El-Mohri, et al. “Relative dosimetry using active matrix flat-panel imager (AMFPI) technology,” Medical Physics 26, 1999. JMM03 29 JMM03 30 184 mu 600 mu/min d=10 cm 5 Gel Dosimetry: Advantages Example: SMLC Delivery 4 Film AMFPD 160 120 80 40 0 184 mu 600 mu/min d=10 cm Position (cm) 2 0 150 130 -2 20 30 50 70 80 100 120 -4 -4 -2 0 2 • Obtain 3-D information in one irradiation • Gels can be prepared with different density therefore ideal for heterogeneous measurements • Gel can be used in containers of different shapes • Ideal for anthropomorphic phantoms 4 Position (cm) B4 JMM03 31 JMM03 32 Gel Dosimetry: Disadvantages • Sensitive to many factors such as time, preparation, temperature • Optical reader requires cylindrical container for gel • MR time is often limited and expensive (unless dedicated scanner) – Long scan times are required to increase accuracy of readout – E.g. 5% accuracy over 10 hr scan time (Gum et al 2002) • Interface of gel and container results in less accurate readout at edges of the gel • Not ready for routine use in the community JMM03 33 Gel Dosimetry MR Dose Maps IMRT Plan Slice 1 Slice 2 Slice 3 Dose % JMM03 35 Dose Deviation % 0 20 40 60 80 100 120 -60 -40 -20 0 30 40 60 Gum et al. “Preliminary study on the use of an inhomogeneous anthropomorphic Fricke gel phantom and 3D magnetic resonance dosimetry for verification of IMRT JMM03 34 plans” PMB 47: N67-N77 (2002). treatment Considerations for Phantoms • Fiducials for reproducible setup of phantom and detectors • Flexibility to accommodate different detectors • Simple vs. anthropomorphic • Homogeneous or heterogeneous MR Dose Maps - IMRT Plan Water Tank • Restricted to gantry at 0 degrees – Unless mylar window for 90 degrees • Flexibility in chamber position • Important for basic depth dose and profile measurements • Output, flatness, symmetry, and linearity assessment JMM03 36 6 Water-equivalent Plastics Simple Geometric Phantom • Flat phantoms with custom chamber inserts 20 cm – Accommodate film at multiple depths – Detector position only varies with depth • Cylindrical phantoms (plastic or water filled) – Ion chamber at single position – Possibly hold films • Can be customized to hold various detectors and have multiple positions JMM03 37 Van Esch et al. “Acceptance tests and QC procedures for the clinical implementation of IMRT using inverse planning and the sliding window technique: experience from five radiotherapy departments,” Radiother Oncol 65: 53- 70 (2002). JMM03 38 Cylindrical Phantom: Ion Chamber and MOSFETs IMRT Verification Phantom Calc. 1.37 Gy Meas. 1.42 Gy Diff –3.52% D.A. Low et al. “Phantoms for IMRT Dose Distribution Measurement and Treatment Verification, Int J Radiat Oncol Biol Phys 40: 1231-1235 (1998). JMM03 39 C. Chuang, UCSF JMM03 40 Spiral Phantom for Dosimetric Verification Calc. 0.81 Gy Meas. 0.78 Gy Diff –3.45% RPC Head Phantom Target Volumes Removable Dry Insert Water Water Critical Structure Paliwal et al “A spiral phantom for IMRT and tomotherapy treatment delivery verification” Medical Physics pp. 2503-2507 (2000). JMM03 41 JMM03 42 TLDs in Target Volumes Radiochromic film through multiple plans Delivery is required by RTOG for participation in IMRT trials 7 Anthropomorphic Phantom for Gel Dosimetry Dosimetry Analysis Software • Transfer patient fluence maps and beam geometry to phantom geometry • 2-D dose difference displays with colorwash • DVHs • Highlight of differences • Gamma analysis – % agreement and distance-to-agreement Gum et al. “Preliminary study on the use of an inhomogeneous anthropomorphic Fricke gel phantom and 3D magnetic resonance dosimetry for verification of IMRT treatment plans” PMB 47: N67-N77 (2002). JMM03 43 JMM03 44 Dosimetric Analysis: Overlay of Isodose Lines Phantom Measurements • Calculate IMRT plan for each field in measurement geometry • Film measurement at depth for individual field at gantry=0 70 60 50 20 10 70 cGy 60 cGy 50 cGy 20 cGy 10 cGy Film at depth in water equivalent plastic JMM03 46 Dosimetric Analysis: Dose Difference Display Dosimetric Analysis : Dose Color Wash Calculation Individual Field Film Data Individual Field Dose (cGy) 70 60 50 40 30 20 10 JMM03 47 Calcs Film Intensity Map Calcs Film JMM03 45 cGy cGy cGy cGy cGy Intensity Map JMM03 48 +/- 10% Differences due to tongue -and-groove 8 Dosimetric Analysis: Dose Difference and DTA Dosimetric Analysis: Gamma Evaluation Dose Reference Both 3%/3 mm 5%/5 mm Evaluated Unmod Spatial 3% Dose Difference JMM03 49 3 mm DTA D. A. Low – Washington University D. A. Low – Washington University JMM03 50 Simple Geometry Tests • Leaf position reproducibility in dynamic or step-and-shoot mode • Effect of gravity on leaf position accuracy and reproducibility • Sweeping gap test • Fence test JMM03 51 Simple Geometry Tests • Leaf speed stability • Leaf acceleration/deceleration • Output checks for small to large fields – Including smallest field size – 1 x 1 cm 2 • MLC limits (field size restrictions) • Depth dose curve measurements JMM03 52 Simple Geometric Tests Geometric Tests L. Xing et al. “Dosimetric verification of a commercial inverse treatment planning system” PMB 44: 463-478 (1999). JMM03 53 JMM03 54 9 Geometric Tests Acceptance tests and quality control (QC) procedures for the clinical implementation of intensity modulated radiotherapy (IMRT) using inverse planning and the sliding window technique: experience from five radiotherapy departments Van Esch et al Radiotherapy Oncology 65: 5355 70JMM03 (2002). Complex Cases – Simple Geometry Evaluate dose across field as a function of regional beamlet intensity JMM03 56 Complex Geometry: Anthropomorphic phantom Potential Pitfalls • Limit number of monitor units per segment to verified Linac behavior • Limit field width for IMRT • Overshoot-undershoot phenomenon • Incorrect tolerance value • Integration with record-and-verify system M. A. MacKenzie et al. “Dosimetric verification of inverse planned step and shoot MLC fields from a commercial planning system,” J Appl Clin Med Phys 3: 97-109 (2002) JMM03 57 JMM03 58 Potential Pitfalls: Prior to Measurements Detector Issues • Verify all equipment is functioning properly Volume – Film processor, digitizer – Detectors, cables, electrometers (automatic leakage correction) – TLD reader, ovens • Input/output to treatment planning system • Standardize measurement setup when possible • Monitor software and hardware changes and QA JMM03 59 Dose (cGy) JMM03 60 10 Detector Position – Example: CAX of Varian MLC is a junction of four 0.5 cm wide leaves Beam 1 Ratio 9/21/02 1.20 Ratio (Calculation/Measurement) • Small ion chambers are very sensitive to position • Position should be considered with respect to MLC design Ratio Calc/Ion Chamber - CAX Beam 1 Ratio - 9/25/02 20 seg Beam 1Ratio 9/25/02 100 seg Beam 2 Ratio 9/21/02 1.15 1.10 1.05 Beam 2 Ratio 9/25/02 20 seg Beam 2 Ratio 9/25/02100 seg 1.00 0.95 0.90 regular spike=cax spike=high spike=low Spikiness values on part of field JMM03 61 JMM03 62 Ratio Calc/Ion Chamber -0.5, -0.5 cm from CAX Effect of Leaf Position Offset on IMRT Ratio (Calculation/Measurement) SMLC Single Field 30 Fractions 1.20 Beam 1 Ratio - Off-axis - 20 seg Beam 1 Ratio off-axis - 100 seg Beam 2 Ratio Off-axis - 20 seg Beam 2 Ratio Off-axis-100 seg 1.15 1.10 1.05 1.00 0.95 0.90 0.85 regular spike=cax spike=high spike=low Spikiness values on part of field Cadman et al “Dosimetric considerations for validation of a sequential IMRT Process with a commercial treatment planning system” PMB: 3001-3010 (2002). JMM03 63 JMM03 64 Effect of Leaf Position Offset on IMRT Overshoot Phenomenon: Example Varian System SMLC Delivery Planned No leaf offset correction -3-12% errors Given With leaf offset corrected +/- 5% Planned Given Planned Given Cadman et al “Dosimetric considerations for validation of a sequential IMRT Process with a commercial treatment planning system” PMB: 3001-3010 (2002). JMM03 65 Ezzell and Chungbin, “The overshoot phenomenon in step -and -shoot IMRT Delivery,” J Appl Clin Med Phys 2: 138-148 (2001). JMM03 66 11 MLC Dynamic Log File (Dynalog) System Communication Delay Accelerator Control Console MLC Workstation Accelerator Control • • • • • MLC Controller Varian 21 EX Expected and actual leaf positions Beam hold-off events Position tolerance setting Recorded approximately every 55 ms • Information can be compared to imported 2-D information ~ 0.055 second JMM03 67 JMM03 68 DMLC: Effect of Incorporating Machine Limitations B1 Deviation in Leaf Trajectory BEFORE AFTER 44 mu, tolerance 0.1 mm no gap: 137 beam hold-offs 54 mu, tolerance 0.25 mm gap 1.1 mm: No beam hold-offs Position (cm) -3.0 -3.5 Expected Trajectory Actual Trajectory MLC Beam Hold-Off -4.0 -4.5 -5.0 1.0 Hold0.0 0.0 0.2 0.4 0.6 0.8 1.0 1.2 1.4 Time (sec) JMM03 69 JMM03 70 Summary • Basic MLC characteristics should be tested in static mode prior to IMRT testing • IMRT tests should be specific to delivery mode and device • Be aware of potential issues with delivery systems that may need further investigation • Multiple detectors and phantoms are typically required for IMRT commissioning • Quantitative dose analysis tools are required for proper evaluation of delivery JMM03 71 Case1, B1 Summary • Measurements may show dosimetric – mechanical differences that planning systems may not model at this time • Need to know the limits of the mechanical systems and combination of software + delivery • Continued need to improve software for delivery system, measurement devices, phantoms, and dose analysis tools JMM03 72 12 Acknowledgements University of Michigan Dale Litzenberg Jeff Radawski Tim Nurushev Benedick Fraass Dan Low – Washington University Cynthia Chuang - UCSF JMM03 73 13