Semi – Rigid Isolator Units - Laboratory Animal Welfare Training

advertisement

Semi – Rigid Isolator Units

Charles River Laboratories

Consulting & Staffing Services

Terminology

Flexible Film

SRI

HEPA

Port

Gasket

Magnahelic Gauge

Isolator History & Function

Used since 1950’s

Flexible Film or Rigid contruction

Biocontainment

 Negative Air Pressure

Bioexclusion

Positive Air Pressure

“Biosecurity” at Charles

River

Features

Typical forward facing port

Gasket will change color when sealed properly

Hinge

Gasket

Clamp

Features

Flexible

Front

HEPA

Filter

External

Filter

Glove

Arm

Isolator Supplies

Nitrile Gloves

O-Rings

Electrical

Tape

Leak Detection

Snoop®

Apply liquid to area suspected of leak and watch for bubbles

Alternative

 Mild Soap solution

Air flow

Negative

Positive

Keeping it Clean

Wescodyne®

Clidox

 Bleach Solution

Windex®/Cinch®

Peracetic Acid

Chemical

Incompatibility

 Check MSDS

Ergonomics

Your back should always be as straight as possible.

Hydraulic tables allow the user to adjust the height!

Filter Change

Filters should be checked daily

Filters need to be changed when the magnahelic gauge increases by 0.05

(i.e. from 1.0 to 1.5)

Filter Pop Quiz

 Which filter assembly needs to be replaced?

Testing for deterioration

This rim should appear green even when stretched.

If small holes or cracks are present glove must be replaced.

Punctures from animal bites or instruments also warrant glove replacement.

Semi-Rigid Isolator Daily Checks and

Inspections

Glove and flexible surface checks

•Check the magnahelic gauge

•Pull the gloves inside out, pulling on the cuff only

•Check for holes, punctures, tears, or stretched areas

•Then recheck the magnahelic for proper pressure

•Wipe down the outside of the flexible front surface

Magnahelic and Ball Valve

Magnahelic

Gauge

Ball Valve

Fan Motor

Semi-Rigid Isolator Daily Checks and Inspections

Magnehelic gauge check

•Check gauge and record twice daily

•Gauge reads what the pressure is inside the isolator

(+.10 or -.10)

•Reads in Inches of Water

• Not reading proper pressure report immediately

DO NOT adjust equipment unless properly trained

Filter Locations

HEPA filter

External

Filter

But how do you use it?

Introduction of supplies

Supplies always in unit

Disinfectants

Sharps

Forceps

Scraper

Marking Pens

Absorbent Towels

Paper Bags

Hand Broom

Supplies introduced to unit

Feed

Bedding

Cages

Water

Sterilized vs

Irradiated

Port Door

How to Use the Semi-Rigid Isolator Units

Loading the Isolator:

• Remove all jewelry, watches and especially rings

•Don protective glove liners

•Gather appropriate supplies

•Open the outside port door

•Place supplies into port –Spray down supplies with disinfectant/sterilent

•Allow sufficient contact time

•20 – 40 minutes

•Secure outer door

How to Use the Semi-Rigid Isolator Units

Loading the Isolator: continued

•Push the sleeves into the isolator by holding onto cuff

• Over stretching the gloves will cause damage

•Open the port door

•Remove items from port into the isolator

•After all items have been removed, shut and lock port door clamps

How to Use the Semi-Rigid Isolator Units

Unloading the Isolator:

•Once everything is ready to be removed from the isolator spray the inside of the port completely

•Spray all items before placing them into port

•Once all items are sprayed down with Clidox the port can be loaded

•After everything is loaded the port is then closed

•Set the timer according to SOP/Chemical Manufacturer

How to Use the Semi-Rigid Isolator Units

Unloading the Isolator:

•When the timer goes off you may unload the port

•Contaminated items should be disposed/sanitized appropriately.

How to Use the Semi-Rigid Isolator Units

Label door with time initials

Set exposure timer

Document event in log

Introducing/Removing Animals to Isolator

 Supplies

Biosafety Cabinet

Plastic bottle/containers

Nalgene®

Disinfectant

Forceps

Introducing/Removing Animals to Isolator

 Transfer into bottle/container inside BSC

 Leave lids unsealed until all containers are filled

Use same introduction procedures as supplies

Ensure animals are promptly removed from bottle/container

 Air supply in sealed container must be established

Removing Animals from Isolator

Return to plastic bottles/containers

Double wrap in bag

 Paper or plastic? 

Use regular “spray out” procedures

Air supply for animals limited!

Semi-Rigid Isolator Daily Checks and

Inspections

Inside of the isolator check

•Shelving and all surfaces for cracks or cuts in the plastic

•Light bulb and casing

•Port and clamps to ensure that it seals inside and out

•If any one of these items are wrong or not working report immediately

• DO NOT leave trash inside the isolator

• DO NOT leave the isolator dirty

Animal checks inside the isolators

•Check feed and water levels in each cage

•Observe the animals to ensure health status

•Check for and change any wet cages

Semi-Rigid Isolator weekly

Checks and Inspections

Cage changing

•Before cage changing begins ensure that all of the daily checks have been completed

•Make sure all items for changing are inside the isolator before starting, this will help with being efficient

•Change cages

•After the cages have been changed remove all trash and dirty bedding from the isolator

•Make sure all animals are present and safely in their home cages

•Wash down the inside of the isolator, all surfaces including the inside of the glove sleeves

•Once all trash is removed wipe down the port

Filter checks

•Check the outside filter for any dirt, debris, or damage

•Report to a supervisor if something is found

Isolator #

Egg Timer

Thermo hygrometer

Brush

Scraper

Wescodyne (diluted)

Pen/Sharpie

Paper Bags

Forceps

Towels

Water Bottles/Sipper

Tops

Cage Bottoms

Wire bar lids

Card Holders/ Cage

Cards

TP Rolls and Nestlets

Sharps Container

SRI Supply Checklist

2 1 3 4 5 6 8 7 9

Date Animals

Enter

ISOLATOR # _______

Investigator

Name

Date Animals

Removed

Date Isolator

Deconned

Initials