see attachment - glucometer.ge

advertisement
INJEX Pharma GmbH
Needle-Free Injection Technology
INJEX Operation
Preparation of Transporter (I)
- turn the green knurled nut of the Transporter, as shown, anticlockwise,
until the cartridge can be placed in the Transporter´s compartment.
Transporter
Preparation of Transporter (II)
- take the cartridge and put it into the Transporter´s compartment.
cartridge
Preparation of Transporter (III)
- for a safe and firm seat of the cartridge, turn the green knurled nut of the
Transporter clockwise, as shown, until the plunger softly touches the
rubber end of the cartridge.
Preparation of Transporter (IV)
- take the Transporter Adapter out of it´s sterile packaging.
- screw the Transporter Adapter on to the Transporter. Ensure that the
adapter needle pierces the rubber seal of the cartridge.
Filling of Ampoule (I)
- take an INJEX-Ampoule out of it´s sterile packaging and press the plunger
completely into the INJEX-Ampoule to force the air out of the Ampoule.
Ampoule
plunger
Filling of Ampoule (II)
- open the protective cap of the Transporter Adapter.
- screw the INJEX-Ampoule, clockwise, as shown, into the open end of the
Adapter until it fits tightly.
Filling of Ampoule (III)
- turn the green knurled nut at the upper end of the Transporter, clockwise,
as shown, until the desired dose of the pharmaceutical agent is in the
INJEX-Ampoule.
Filling of Ampoule (IV)
- unscrew the filled INJEX-Ampoule from the Adapter in an anticlockwise
direction and close the protective cap.
Preparation of the Injector (I)
- take the loaded and safeguarded Injector out of the Reset Box.
Injector
Reset-Box
Preparation of the Injector (II)
- screw the filled INJEX-Ampoule clockwise into the open end of the Injector
(the last round goes a little bit harder, this is normal).
Preparation of the Injector (III)
- push the green safety ring towards the Ampoule into „Safe off“ position.
Now the Injector is ready for use.
The Injection
- hold the Injector under firm pressure perpendicularly to the injection site
(i.e. abdomen), until a clear indentation forms on the skin. Press the
trigger and release the injection. Hold the Injector on the injection site for
another 1 – 2 sec.
After the Injection (I)
- unscrew the INJEX-Ampoule and discard it – domestic waste -.
After the Injection (II)
- push the green safety ring back into the „Safe“ position.
- store the secured and non-reset Injector and its accessories in the INJEX
Case.